Skin photoplethysmography — a review
Abstract
The photoplethysmograph has been used for over 50 years but there are still misconceptions in how and what is the information obtained. A photoplethysmograph signal from any site on the skin can be separated into an oscillating (a.c.) and a steady-state (d.c.) component, their amplitudes dependent upon the structure and flow in the vascular bed. Many simple applications are available: pulse counters, using the a.c. component, skin colour and haemoglobin saturation meters, using the d.c. component. The d.c. component of the photoplethysmograph signal is a function of the blood flux beneath the device. A good emitter for use in a photoplethysmograph of skin blood flow is one in the frequency range 600-700 nm and the best signal for a.c. analysis is obtained from the finger pulp. The frequency range of the electronic circuitry should be from 0.01 to 15 Hz, then all the information in the signal can be extracted about the autonomic nervous system control of the cardiovascular system, particularly between 0.01 and 2 Hz. Comparative studies may be drawn between similar skin sites on a subject or between subjects if the afferent inputs to the brain stem are controlled or driven at a known frequency. These afferents, inputs, will modulate the efferents, outputs, which generate variations in the a.c. component of the detected photoplethysmograph signal.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Without this type of real-time, in-sensor processing (IR_Non in Fig. 6D), robust operation would be impossible: PPG signals would quickly drift outside of the narrow operating range of the ADC as a result of patient-to-patient variations in skin pigmentation and unavoidable, time-dependent fluctuations in optical scattering that result from micromotions relative to underlying blood vessels and sub”
From Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care (2019) · cites this paper as [28]
Citations
Cites
McDonald, D. A. (1963) · The American Journal of the Medical Sciences · doi:10.1097/00000441-196301000-00017 · Cited by 1,568
Analysis of Heart Rate Variability
SAYKRS, B. M. (1973) · Ergonomics · doi:10.1080/00140137308924479 · Cited by 957
In vivo evaluation of microcirculation by coherent light scattering
Stern, M. D. (1975) · Nature · doi:10.1038/254056a0 · Cited by 866
Arterial assessment by Doppler-shift ultrasound.
Gosling, R. G., & King, D. H. (1974) · PubMed · Cited by 667
LASER DOPPLER MEASUREMENT OF CUTANEOUS BLOOD FLOW
Holloway, G. A., & Watkins, D. (1977) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12507665 · Cited by 543
THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH
Hertzman, A. B. (1938) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1938.124.2.328 · Cited by 460
Bini, G., Hagbarth, K. E., Hynninen, P., & Wallin, B. G. (1980) · The Journal of Physiology · doi:10.1113/jphysiol.1980.sp013413 · Cited by 421
A theoretical and experimental study of light absorption and scattering by in vivo skin
Dawson, J. B., Barker, D., Ellis, D. J., COTTERILL, J. A., Grassam, E., Fisher, G. W., & Feather, J. W. (1980) · Physics in Medicine and Biology · doi:10.1088/0031-9155/25/4/008 · Cited by 355
Burton, A. C. (1939) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1939.127.3.437 · Cited by 333
Sympathetic Control of Human Blood Vessels
(1953) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.1953.tb84977.x · Cited by 259
Research works printed but not linked: 8
- Burton, 1940, Direct measurement of thermal conductance of skin as ideal index of peripheral blood flow, Am. J. Physiol., 129, 389
- Challoner, 1979, Photoelectric plethysmography for estimating cutaneous blood flow, Vol. 1, 127
- Fairs, 1985, Optical methods for measuring blood flow in skin, Vol. 1, 11
- Hertzman, 1937, Observations on the finger volume pulse recorded photoelectrically, Am. J. Physiol., 119, 334
- Mearns, 1985, Forcing frequency testing, theory and practice with reference to skin blood flow, Vol. 1, 23
- Molitor, 1936, A new continuous recording of peripheral circulatory changes, J. Pharmacol. Exp. Ther., 57, 6
- Sejrsen, 1971, Measurement of cutaneous bloodflow by freely diffusible radioactive isotopes, Dan. Med. Bull., 18, 9
- Thune, 1970, Plethysmographic recordings of skin pulses with particular reference to the piezoelectric method, Acta Dermatovener., 50, 27
Cited by
Sorted by citations
Items 301–328 of 328
Chen, Y. (2012) · doi:10.32657/10356/50792
“These optical processes have been related to PPG measurements in several studies and investigations [159-170].”
Optical non-invasive blood pressure : PPG waveform-based estimation and verification
정인철 (2011) · YUHSpace (Yonsei University Medical Library)
Principle and Equipment of Resonance LIL Photobiostimulation
Grimblatov, V. M. (1996) · Springer eBooks · doi:10.1007/978-3-642-80264-5_116
OSTEOPOROSIS, CIRCULATION, AND FLUID DYNAMICS
ZHANG, B., Mateus, J., & Hargens, A. R. (2016) · WORLD SCIENTIFIC eBooks · doi:10.1142/9789814713764_0012
van Haaren, J. H. L., Smit, J., Verhulst, M. M. L. H., & Vlaskamp, L. B. (2016)
Vahdani-Manaf, N., & Kayıkçıoğlu, T. (2015) · International Congress on Technology, Communication and Knowledge · doi:10.1109/ictck.2015.7582673
“The complex interplay between light and biological tissue includes some optical processes such as scattering, absorption, reflection, transmission and fluorescence [2].”
BackgroundRakurthi, A. (2012) · OhioLink ETD Center (Ohio Library and Information Network)
Pedapati, C. (2014) · Arizona State University Library Digital Repository (Arizona State University)
Recommendation for Measuring Digital Volume Pulse in Mobile Application: For Healthy Normal Subject
Seok, H. S., & Shin, H. (2021) · IEEE Access · doi:10.1109/access.2021.3077639
Predict Blood Pressure by Photoplethysmogram with the Fluid-Structure Interaction Modeling
Chen, J. C. J., Hao, W., Sun, P., & Zhang, L. Z. L. (2022) · Communications in Computational Physics · doi:10.4208/cicp.oa-2021-0135
Multi-Wavelength Transmittance Photoplethysmography with Near Infrared Laser Diodes during Exercise
Silva, S. M. L. (2005) · Diagnostic Optical Spectroscopy in Biomedicine · doi:10.1364/ecbo.2005.tuh6
Recommendation in Measuring Digital Volume Pulse for Mobile Application (Preprint)Preprint
Seok, H. S., & Shin, H. (2020) · doi:10.2196/preprints.24159
Rusterholz, F., Rodriguez, A., Ortiz, V., & Peacock, C. A. (2023) · Research Directs in Health Sciences · doi:10.53520/rdhs2023.10475
Imaging photoplethysmography for assessing stimulus-evoked hemodynamics in mouse barrel cortex
He, Q., Li, Y., & Wang, R. K. (2023) · Optics in the Life Sciences · doi:10.1364/brain.2023.bth1b.5
Nishidate, I. (2022) · Conference on Lasers and Electro-Optics Pacific Rim (CLEO-PR) · doi:10.1364/cleopr.2022.ctua15d_01
Nishidate, I. (2022) · Conference on Lasers and Electro-Optics Pacific Rim (CLEO-PR) · doi:10.1109/cleo-pr62338.2022.10432017
Wavelet Analysis of Photoplethysmogram in Zebrafish Larvaе Cardiovascular System Study
Slavin, A. E., Guryleva, A. V., Bukova, V. I., Shuman, L. A., & Burlakov, A. B. (2024) · International Conference on Digital Signal Processing and its Applications (DSPA) · doi:10.1109/dspa60853.2024.10510097
“The photoplethysmogram (PPG), which is acquired using this method, is a periodic signal that is modulated by heart muscle activity [2].”
MethodMisra, G., Wegerif, S., Fairlie, L., Kapoor, M., Fok, J., Salt, G., Halbert, J., Maconochie, I. K., & Mullen, N. (2024) · JMIR Research Protocols · doi:10.2196/58334
“…measurement of VS. Photoplethysmography (PPG) is an optical measurement technique that records changes in the light reflected from the skin surface due to volumetric changes in the facial blood vessels; small variations in perfusion provide valuable information about the cardiovascular system [9].”
Heart Rate Estimation From Photoplethysmography Signals Using Fast Fourier Transform
Mohan, N., Awwad, F. R., Bastaki, N., & Atef, M. (2024) · International Conference on Signal Processing and Information Security (ICSPIS) · doi:10.1109/icspis63676.2024.10812645
Development of a Software Algorithm for Processing and Analysis of Video Photoplethysmogram Signals
Derevesnikova, D., Ziganuriva, D., Luzhnov, P. V., Guryleva, A. V., & Machikhin, A. S. (2025) · Systems and Technologies of the Digital HealthCare (STDH) · doi:10.1109/stdh66836.2025.11227476
“Firstly, these are due to physiological characteristics – the presence of second-and third-order waves reflecting respiratory activity and fluctuations in vascular tone; secondly, limitations and artifacts of the recording equipment [1,7].”
Wrist Photoplethysmography Pulse Waves: Morphological Classes and Physiological Influences
Dendorfer, A., & Charlton, P. H. (2025) · International Electronic Conference on Sensors and Applications (ECSA) · doi:10.3390/ecsa-12-26556
“The shape of the PPG pulse wave is influenced by both the heart and the blood vessels, making it a rich source of information on cardiovascular physiology [1].”
Lee, W., Chang, M. H., Kim, S., Lee, B., Jung, C. H., & Kim, K. (2025) · IEEE Access · doi:10.1109/access.2025.3646742
James, J., Gudigar, A., Raghavendra, U., Samanth, J., Maithri, M., Kaprekar, A., Prabhu, M. A., Salvi, M., Molinari, F., Ciaccio, E. J., et al. (2025) · Informatics in Medicine Unlocked · doi:10.1016/j.imu.2025.101730
Research on Risk Assessment of Arterial Stiffness by Using Photoplethysmogram in Young people
Sun, Y., & Abe, M. (2025) · IEEJ Transactions on Electronics Information and Systems · doi:10.1541/ieejeiss.146.19
Avina-Bravo, E. G. (2024) · theses.fr (ABES) · doi:10.70675/ad625489z7b4dz4f64z81ebz3bcf5655f757
Lapitan, D. G., & Рогаткин, Д. А. (2026) · Ural Symposium on Biomedical Engineering, Radioelectronics and Information Technology (USBEREIT) · doi:10.1109/usbereit70063.2026.11581011
“The physical principle of the method consists of probing the tissue with optical radiation (usually incoherent) and recording the signal scattered from the tissue, which consists of a slowly changing component (DC), onto which a pulsating component (AC) is superimposed [2].”
Metadata
- Authors
- Kamal, A. A. R., Harness, J. B., Irving, G., & Mearns, A. J.
- Affiliations
- University of Bradford (GB)
- Corresponding author
- Harness, J.B.
- Venue
- Computer Methods and Programs in Biomedicine
- Publisher
- Elsevier BV
- Published
- 1989-04
- Volume
- 28
- Issue
- 4
- Pages
- 257-269
- ISSN
- 0169-2607 (ISSN-L)
- OpenAlex
- W2054173774
- Cited by
- 328
- References
- 56
- Citations by year
- 2026: 1 · 2025: 13 · 2024: 16 · 2023: 21 · 2022: 31 · 2021: 15 · 2020: 19 · 2019: 26 · 2018: 15 · 2017: 22 · 2016: 14 · 2015: 17 · 2014: 13 · 2013: 8 · 2012: 12
- FWCI
- 0.25
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Computer Methods and Programs in Biomedicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- photoplethysmography, skin blood flow, pulse counting, hemoglobin oxygen saturation, finger pulp, vascular bed
- MeSH terms
- Equipment Design, Humans, Monitoring, Physiologic, Plethysmography, Regional Blood Flow, Skin
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review