Effect of EEG electrode number on epileptic source localization in pediatric patients
Open accessAbstract
Objective To investigate the relationship between EEG source localization and the number of scalp EEG recording channels.Methods 128 EEG channel recordings of 5 pediatric patients with medically intractable partial epilepsy were used to perform source localization of interictal spikes. The results were compared with surgical resection and intracranial recordings. Various electrode configurations were tested and a series of computer simulations based on a realistic head boundary element model were also performed in order to further validate the clinical findings.Results The improvement seen in source localization substantially decreases as the number of electrodes increases. This finding was evaluated using the surgical resection, intracranial recordings and computer simulation. It was also shown in the simulation that increasing the electrode numbers could remedy the localization error of deep sources. A plateauing effect was seen in deep and superficial sources with further increasing the electrode number.Conclusion The source localization is improved when electrode numbers increase, but the absolute improvement in accuracy decreases with increasing electrode number.Significance Increasing the electrode number helps decrease localization error and thus can more ably assist the physician to better plan for surgical procedures.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Studies indicate that higher spatial sampling helps improve the precision of ESI (18, 19).”
From Electrophysiological source imaging: A noninvasive window to brain dynamics (2018) · cites this paper as [19]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Standardized low-resolution brain electromagnetic tomography (sLORETA): technical details.
Pascual-Marqui, R. D. (2002) · PubMed · Cited by 3,329
Nunez, P. L., & Srinivasan, R. (2006) · doi:10.1093/acprof:oso/9780195050387.001.0001 · Cited by 2,531
Picton, T. W., Bentin, S., Berg, P., Donchin, E., Hillyard, S. A., Johnson, R., Miller, G. A., Ritter, W., Ruchkin, D. S., Rugg, M. D., et al. (2000) · Psychophysiology · doi:10.1111/1469-8986.3720127 · Cited by 2,157
Dale, A. M., & Sereno, M. I. (1993) · Journal of Cognitive Neuroscience · doi:10.1162/jocn.1993.5.2.162 · Cited by 2,003
Baillet, S., Mosher, J., & Leahy, R. (2001) · IEEE Signal Processing Magazine · doi:10.1109/79.962275 · Cited by 1,846
Michel, C., Murray, M. M., Lantz, G., Gonzalez, S., Spinelli, L., & de Peralta, R. G. (2004) · Clinical Neurophysiology · doi:10.1016/j.clinph.2004.06.001 · Cited by 1,771
Electric Fields of the Brain: The Neurophysics of EEG
Fermaglich, J. (1982) · JAMA · doi:10.1001/jama.1982.03320380071046 · Cited by 1,557
Surgical treatment of the epilepsies
Bloomfield, S. M., & Shetter, A. G. (1988) · Surgical Neurology · doi:10.1016/0090-3019(88)90022-5 · Cited by 1,206
Realistic conductivity geometry model of the human head for interpretation of neuromagnetic data
Hamalainen, M. S., & Sarvas, J. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.16463 · Cited by 989
Spatial sampling of head electrical fields: the geodesic sensor net
Tucker, D. M. (1993) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(93)90121-b · Cited by 756
Statistical control of artifacts in dense array EEG/MEG studies
Junghöfer, M., Elbert, T. R., Tucker, D. M., & Rockstroh, B. S. (2000) · Psychophysiology · doi:10.1111/1469-8986.3740523 · Cited by 577
Towle, V. L., Bolaños, J., Suarez, D., Tan, K. K., Grzeszczuk, R. P., Levin, D. N., Çakmur, R., Frank, S., & Spire, J. (1993) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(93)90061-y · Cited by 443
Intracranial EEG Substrates of Scalp EEG Interictal Spikes
Tao, J. X., Ray, A. K., Hawes‐Ebersole, S., & Ebersole, J. S. (2005) · Epilepsia · doi:10.1111/j.1528-1167.2005.11404.x · Cited by 425
Epileptic source localization with high density EEG: how many electrodes are needed?
Lantz, G., Grave de Peralta, R., Spinelli, L., Seeck, M., & Michel, C. M. (2003) · Clinical Neurophysiology · doi:10.1016/s1388-2457(02)00337-1 · Cited by 413
Spatial spectra of scalp EEG and EMG from awake humans
Freeman, W., Holmes, M., Burke, B. C., & Vanhatalo, S. (2003) · Clinical Neurophysiology · doi:10.1016/s1388-2457(03)00045-2 · Cited by 379
Focal neuronal migration disorders and intractable partial epilepsy: Results of surgical treatment
Palmini, A., Andermann, F., Olivier, A., Tampieri, D. I., & ROBITAILLE, Y. (1991) · Annals of Neurology · doi:10.1002/ana.410300603 · Cited by 344
Electric Dipole Tracing in the Brain by Means of the Boundary Element Method and Its Accuracy
He, B., Musha, T., Okamoto, Y., Homma, S., Nakajima, Y., & Sato, T. (1987) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1987.326056 · Cited by 334
Self-citationElectrophysiological Imaging of Brain Activity and Connectivity—Challenges and Opportunities
He, B., Yang, L., Wilke, C., & Yuan, H. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2139210 · Cited by 300
Self-citationEstimating the spatial Nyquist of the human EEG
Srinivasan, R., Tucker, D., & Murias, M. (1998) · Behavior Research Methods, Instruments, & Computers · doi:10.3758/bf03209412 · Cited by 270
Update on the Epidemiology and Prognosis of Pediatric Epilepsy
Shinnar, S., & Pellock, J. M. (2002) · Journal of Child Neurology · doi:10.1177/08830738020170010201 · Cited by 257
EEG source localization in focal epilepsy: Where are we now?
Plummer, C., Harvey, A. S., & Cook, M. J. (2007) · Epilepsia · doi:10.1111/j.1528-1167.2007.01381.x · Cited by 252
Spatial sampling and filtering of EEG with spline Laplacians to estimate cortical potentials
Srinivasan, R., Nunez, P. L., Tucker, D. M., Silberstein, R. B., & Cadusch, P. J. (1996) · Brain Topography · doi:10.1007/bf01186911 · Cited by 245
128-Channel EEG Source Imaging in Epilepsy: Clinical Yield and Localization Precision
Michel, C., Lantz, G., Spinelli, L., de Peralta, R. G., Landis, T., & Seeck, M. (2004) · Journal of clinical neurophysiology · doi:10.1097/00004691-200403000-00001 · Cited by 222
Validation of ICA-based myogenic artifact correction for scalp and source-localized EEG
McMenamin, B. W., Shackman, A. J., Maxwell, J. S., Bachhuber, D. R. W., Koppenhaver, A. M., Greischar, L. L., & Davidson, R. J. (2009) · NeuroImage · doi:10.1016/j.neuroimage.2009.10.010 · Cited by 214
Ictal source analysis: Localization and imaging of causal interactions in humans
Ding, L., Worrell, G. A., Lagerlund, T. D., & He, B. (2007) · NeuroImage · doi:10.1016/j.neuroimage.2006.09.042 · Cited by 183
Self-citationA critical analysis of linear inverse solutions to the neuroelectromagnetic inverse problem
de Peralta-Menendez, R. G., & Andino, S. L. G. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.664200 · Cited by 175
Lantz, G., Spinelli, L., Seeck, M., Menéndez, R. G. D. P., Sottas, C. C., & Michel, C. M. (2003) · Journal of Clinical Neurophysiology · doi:10.1097/00004691-200309000-00003 · Cited by 172
L1-Norm-Based Common Spatial Patterns
Wang, H., Tang, Q., & Zheng, W. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2177523 · Cited by 144
EEG Source Imaging in Pediatric Epilepsy Surgery: A New Perspective in Presurgical Workup
Sperli, F., Spinelli, L., Seeck, M., Kurian, M., Michel, C. M., & Lantz, G. (2006) · Epilepsia · doi:10.1111/j.1528-1167.2006.00550.x · Cited by 137
Spatiotemporal EEG Analysis and Distributed Source Estimation in Presurgical Epilepsy Evaluation
Michel, C. M., de Peralta, R. G., Lantz, G., Andino, S. L. G., Spinelli, L., Blanke, O., Landis, T., & Seeck, M. (1999) · Journal of Clinical Neurophysiology · doi:10.1097/00004691-199905000-00005 · Cited by 118
Surgical Treatment for Epilepsy
Engel, J. (2008) · JAMA · doi:10.1001/jama.2008.756 · Cited by 116
Noninvasive Localization of Epileptogenic Foci by EEG Source Modeling
Ebersole, J. S. (2000) · Epilepsia · doi:10.1111/j.1528-1157.2000.tb01531.x · Cited by 104
Geodesic photogrammetry for localizing sensor positions in dense-array EEG
Russell, G. S., Eriksen, K. J., Poolman, P., Luu, P., & Tucker, D. M. (2005) · Clinical Neurophysiology · doi:10.1016/j.clinph.2004.12.022 · Cited by 104
Lu, Y., Yang, L., Worrell, G. A., & He, B. (2012) · Clinical Neurophysiology · doi:10.1016/j.clinph.2011.11.007 · Cited by 98
Self-citationHe, B., Zhang, X., Lian, J., Sasaki, H., Wu, D., & Towle, V. L. (2002) · NeuroImage · doi:10.1006/nimg.2002.1127 · Cited by 95
Self-citationDynamic imaging of ictal oscillations using non-invasive high-resolution EEG
Yang, L., Wilke, C., Brinkmann, B., Worrell, G. A., & He, B. (2011) · NeuroImage · doi:10.1016/j.neuroimage.2011.03.043 · Cited by 93
Self-citationResective reoperation for failed epilepsy surgery
Siegel, A. M., Cascino, G. D., Meyer, F. B., McClelland, R. L., So, E. L., Marsh, W. R., Scheithauer, B. W., & Sharbrough, F. W. (2004) · Neurology · doi:10.1212/01.wnl.0000147476.86575.a7 · Cited by 92
A method for determining optimal interelectrode spacing for cerebral topographic mapping
Spitzer, A. R., Cohen, L. G., Fabrikant, J., & Hallett, M. (1989) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(89)90072-2 · Cited by 82
Dipole localization in patients with epilepsy using the relistically shaped head model
Roth, B. J., Ko, D., von Albertini-Carletti, I., Scaffidi, D., & Sato, S. (1997) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/s0013-4694(96)95111-5 · Cited by 79
Interictal spike analysis of high-density EEG in patients with partial epilepsy
Wang, G., Worrell, G. A., Yang, L., Wilke, C. T., & He, B. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.10.043 · Cited by 77
Self-citationHuppertz, H., Hof, E., Klisch, J., Wagner, M., Lücking, C. H., & Kristeva-Feige, R. (2001) · NeuroImage · doi:10.1006/nimg.2000.0680 · Cited by 74
Grand Challenges in Mapping the Human Brain: NSF Workshop Report
He, B., Coleman, T. P., Genin, G. M., Glover, G. H., Hu, X. P., Johnson, N., Liu, T., Makeig, S., Sajda, P., & Ye, K. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2283970 · Cited by 69
Self-citationWang, Y., & Gotman, J. (2001) · Clinical Neurophysiology · doi:10.1016/s1388-2457(01)00594-6 · Cited by 69
A practical method for determining electrode positions on the head
de Munck, J. C., Vijn, P. C. M., & Spekreijse, H. (1991) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(91)90023-w · Cited by 60
Evaluation of EEG localization methods using realistic simulations of interictal spikes
Grova, C., Daunizeau, J., Lina, J.-M., Bénar, C., Benali, H., & Gotman, J. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2005.08.053 · Cited by 59
Spatial mislocalization of EEG electrodes – effects on accuracy of dipole estimation
Khosla, D., Don, M., & Kwong, B. (1999) · Clinical Neurophysiology · doi:10.1016/s0013-4694(98)00121-7 · Cited by 57
Dipole‐Source Analysis in a Realistic Head Model in Patients with Focal Epilepsy
Herrendorf, G., Steinhoff, B. J., Kolle, R., Baudewig, J., Waberski, T. D., Büchner, H., & Paulus, W. (2000) · Epilepsia · doi:10.1111/j.1528-1157.2000.tb01508.x · Cited by 55
Salmelin, R., & Baillet, S. (2009) · Human Brain Mapping · doi:10.1002/hbm.20795 · Cited by 54
Gevins, A. S. (1993) · Brain Topography · doi:10.1007/bf01128684 · Cited by 51
Krings, T., Chiappa, K. H., Cuffin, B. N., Buchbinder, B. R., & Cosgrove, G. R. (1998) · Annals of Neurology · doi:10.1002/ana.410440114 · Cited by 51
Ding, L., Lai, Y. L., & He, B. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/50/1/004 · Cited by 49
Self-citationNoninvasive Imaging of the High Frequency Brain Activity in Focal Epilepsy Patients
Lu, Y., Worrell, G., Zhang, H., Yang, L., Brinkmann, B., Nelson, C., & He, B. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2297332 · Cited by 46
Self-citationHigh-resolution EEG: Cortical potential imaging of interictal spikes
Zhang, X., van Drongelen, W., Hecox, K. E., Towle, V. L., Frim, D. M., McGee, A., & He, B. (2003) · Clinical Neurophysiology · doi:10.1016/s1388-2457(03)00194-9 · Cited by 38
Self-citationA Sparse and Spatially Constrained Generative Regression Model for fMRI Data Analysis
Oikonomou, V. P., Blekas, K., & Astrakas, L. G. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2104321 · Cited by 38
Dynamic imaging of seizure activity in pediatric epilepsy patients
Lu, Y., Yang, L., Worrell, G. A., Brinkmann, B., Nelson, C., & He, B. (2012) · Clinical Neurophysiology · doi:10.1016/j.clinph.2012.04.021 · Cited by 32
Self-citationInterictal electromagnetic source imaging in focal epilepsy: practices, results and recommendations
Leijten, F. S. S., & Huiskamp, G. J. M. (2008) · Current Opinion in Neurology · doi:10.1097/wco.0b013e3283081e23 · Cited by 30
A State-Space Modeling Approach for Localization of Focal Current Sources From MEG
Fukushima, M., Yamashita, O., Kanemura, A., Ishii, S., Kawato, M., & Sato, M.-A. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2189713 · Cited by 29
Electrophysiological Mapping and Neuroimaging
He, B., & Ding, L. (2013) · Neural Engineering · doi:10.1007/978-1-4614-5227-0_12 · Cited by 26
Self-citation3-D Source Localization of Epileptic Foci Integrating EEG and MRI Data
Mirkovic, N., Adjouadi, M., Yaylali, I., & Jayakar, P. (2003) · Brain Topography · doi:10.1023/b:brat.0000006335.75534.78 · Cited by 22
Noninvasive cortical imaging of epileptiform activities from interictal spikes in pediatric patients
Lai, Y., Zhang, X., van Drongelen, W., Korhman, M., Hecox, K., Ni, Y., & He, B. (2011) · NeuroImage · doi:10.1016/j.neuroimage.2010.07.026 · Cited by 22
Self-citationGenetic Localization of the Ca2+ Channel Gene CACNG2 Near SCA10 on Chromosome 22q13
Burgess, D. L., Matsuura, T., Ashizawa, T., & Noebels, J. L. (2000) · Epilepsia · doi:10.1111/j.1528-1157.2000.tb01500.x · Cited by 22
3D source localization of interictal spikes in epilepsy patients with MRI lesions
Ding, L., Worrell, G. A., Lagerlund, T. D., & He, B. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/16/011 · Cited by 19
Self-citationEfficient Dipole Parameter Estimation in EEG Systems With Near-ML Performance
Wu, S. C., Swindlehurst, A. L., Wang, P. T., & Nenadić, Z. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2187336 · Cited by 16
Electric Fields and Currents in Biological Tissue
Nunez, P. L., & Srinivasan, R. (2006) · Oxford University Press eBooks · doi:10.1093/acprof:oso/9780195050387.003.0004 · Cited by 11
Lai, D., Liu, C., Eggen, M. D., Iaizzo, P. A., & He, B. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/13/027 · Cited by 9
Self-citation
Cited by
Sorted by citations
The standardized EEG electrode array of the IFCN
Seeck, M., Koessler, L., Bast, T., Leijten, F., Michel, C., Baumgartner, C., He, B., & Beniczky, S. (2017) · Clinical Neurophysiology · doi:10.1016/j.clinph.2017.06.254 · Cited by 489
“A recent interictal study using 128-electrodes recordings in pediatric and adult epilepsy patients indicated that at least 64–76 electrodes are desirable to avoid significant source localization errors (Sohrabpour et al., 2015).”
Self-citationEEG source localization: Sensor density and head surface coverage
Song, J., Davey, C., Poulsen, C., Luu, P., Turovets, S. I., Anderson, E. W., Li, K., & Tucker, D. M. (2015) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2015.08.015 · Cited by 347
Electrophysiological source imaging: A noninvasive window to brain dynamics
He, B., Sohrabpour, A., Brown, E., & Liu, Z. (2018) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-062117-120853 · Cited by 297
“Studies indicate that higher spatial sampling helps improve the precision of ESI (18, 19).”
BackgroundSelf-citationTMS combined with EEG: Recommendations and open issues for data collection and analysis
Hernandez-Pavon, J. C., Veniero, D., Bergmann, T. O., Belardinelli, P., Bortoletto, M., Casarotto, S., Casula, E. P., Farzan, F., Fecchio, M., Julkunen, P., et al. (2023) · Brain Stimulation · doi:10.1016/j.brs.2023.02.009 · Cited by 238
“16 that with optimal placement on the scalp, 30–50 electrodes would be enough to gather the spatial 429 information that is available to EEG [92-94].”
Electrophysiological Brain Connectivity: Theory and Implementation
He, B., Astolfi, L., Valdes-Sosa, P. A., Marinazzo, D., Palva, S., Bénar, C., Michel, C. M., & Koenig, T. W. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2019.2913928 · Cited by 242
Self-citation
Metadata
- Authors
- Sohrabpour, A., Lu, Y., Kankirawatana, P., Blount, J., Kim, H., & He, B.
- Affiliations
- University of Minnesota (US)
- University of Alabama at Birmingham (US)
- Corresponding author
- He, Bin
- Venue
- Clinical Neurophysiology
- Publisher
- Elsevier BV
- Published
- 2015-03
- Volume
- 126
- Issue
- 3
- Pages
- 472-480
- ISSN
- 1388-2457 (ISSN-L)
- PMC
- PMC4289666
- OpenAlex
- W2020479538
- Cited by
- 149
- References
- 65
- Citations by year
- 2026: 2 · 2025: 14 · 2024: 9 · 2023: 29 · 2022: 13 · 2021: 16 · 2020: 10 · 2019: 19 · 2018: 8 · 2017: 9 · 2016: 10 · 2015: 7
- FWCI
- 5.21
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Clinical Neurophysiology (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- NIH (EB006433, EY023101, HL117664)
- NSF
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Health Sciences › Medicine › Clinical Neurology › Epilepsy research and treatment
- Physical Sciences › Computer Science › Signal Processing › Direction-of-Arrival Estimation Techniques
- Keywords
- EEG source localization, electrode number, pediatric epilepsy, interictal spikes, electrode configuration, surgical planning, intracranial recordings
- MeSH terms
- Adolescent, Brain, Brain Mapping, Child, Child, Preschool, Computer Simulation, Electrodes, Electroencephalography, Epilepsies, Partial, Female, Humans, Male
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review