HRS/EHRA Expert Consensus on the Monitoring of Cardiovascular Implantable Electronic Devices (CIEDs): Description of Techniques, Indications, Personnel, Frequency and Ethical Considerations
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Current recommended minimum frequency of monitoring for patients with cardiovascular implantable electronic devices is once per year in person and every 3 to 6 (ICD) or 3 to 12 (PM) months either remotely or in the office.51”
From Managing with pacemakers and implantable cardioverter defibrillators (2013) · cites this paper as [51]
In the citing paper
“More timely detection of both clinical and device-related issues may improve care, 51”
From Managing with pacemakers and implantable cardioverter defibrillators (2013) · cites this paper as [51]
Citations
Cites
ISO (International Organization for Standardization)
Nollet, L. M. L. (2019) · CRC Press eBooks · doi:10.1201/9781315178592-10 · Cited by 1,902
Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., CHAITMAN, B. R., FROMER, M. A., Gregoratos, G., Klein, G. J., Moss, A. J., Myerburg, R. J., et al. (2006) · EP Europace · doi:10.1093/europace/eul108 · Cited by 741
Self-citationCancer pain relief and palliative care
Akinsanya, C. Y. (1992) · Nurse Education Today · doi:10.1016/0260-6917(92)90050-x · Cited by 693
Management of Implantable Cardioverter Defibrillators in End-of-Life Care
Goldstein, N. E., Lampert, R., Bradley, E., Lynn, J., & Krumholz, H. M. (2004) · Annals of Internal Medicine · doi:10.7326/0003-4819-141-11-200412070-00006 · Cited by 325
ACC/AHA/HRS 2006 Key Data Elements and Definitions for Electrophysiological Studies and Procedures
Buxton, A. E., Calkins, H. G., Callans, D. J., Dimarco, J., Fisher, J. D., Greene, H. L., Haines, D. E., Hayes, D. L., Heidenreich, P. A., Miller, J. M., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.09.020 · Cited by 237
Self-citationClinical Cardiac Pacing, Defibrillation, and Resynchronization Therapy
(2007) · Elsevier eBooks · doi:10.1016/b978-1-4160-2536-8.x5001-8 · Cited by 224
Perceived Quality of Life and Preferences for Life-Sustaining Treatment in Older Adults
Uhlmann, R. F. (1991) · Archives of Internal Medicine · doi:10.1001/archinte.1991.00400030059010 · Cited by 217
Zipes, D. P., Camm, A. J., Borggrefe, M. M., Buxton, A. E., CHAITMAN, B. R., FROMER, M. A., Gregoratos, G., Klein, G. J., Moss, A. J., Myerburg, R. J., et al. (2006) · Revista Española de Cardiología · Cited by 203
Self-citationPatients' Knowledge of Options at the End of Life
Silveira, M. J., DiPiero, A., Gerrity, M., & Feudtner, C. (2000) · JAMA · doi:10.1001/jama.284.19.2483 · Cited by 201
Remote Monitoring of Implantable Cardioverter Defibrillators: A Prospective Analysis
SCHOENFELD, M. H., COMPTON, S. J., MEAD, R. H., WEISS, D. N., SHERFESEE, L., ENGLUND, J., & MONGEON, L. R. (2004) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2004.00524.x · Cited by 194
Self-citation
Flagged first, then by citations
4 of the 66 references couldn’t be matched to a known paper.
Research works printed but not linked: 6
- Schoenfeld, 2005, Sophisticated remote implantable cardioverter-defibrillator follow-up: a status report, Pacing Clin Electrophysiol, 28, 235, 10.1111/j.1540-8159.2005.09554.x · DOI
- Joseph, 2004, Remote interrogation and monitoring of implantable cardioverter defibrillators, J Interv Card Electrophysiol, 11, 161, 10.1023/B:JICE.0000042356.52369.89 · DOI
- Hayes, 2003, NASPE (HRS) training requirements for cardiac implantable electronic devices: selection, implantation, and follow-up, Pacing Clin Electrophysiol, 26, 1556, 10.1046/j.1460-9592.2003.t01-1-00229.x · DOI
- Berger, 2005, The ethics of deactivating implanted cardioverter defibrillators, Ann Intern Med, 142, 631, 10.7326/0003-4819-142-8-200504190-00012 · DOI
- Lewis, 2006, Withdrawing implantable defibrillator shock therapy in terminally ill patients, Am J Med, 119, 892, 10.1016/j.amjmed.2006.01.017 · DOI
- 2008, ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices), Heart Rhythm, 5
Cited by
Sorted by citations
Subclinical atrial fibrillation and the risk of stroke
Healey, J. S., Connolly, S. J., Gold, M. R., Israel, C. W., Van Gelder, I. C., Capucci, A., Lau, C. P., Fain, E., Yang, S., Bailleul, C., et al. (2012) · New England Journal of Medicine · doi:10.1056/nejmoa1105575 · Cited by 2,100
Epstein, A. E., Dimarco, J., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2012) · Circulation · doi:10.1161/cir.0b013e318276ce9b · Cited by 1,496
Self-citationPoole, J. E., Gleva, M. J., Mela, T., Chung, M. K., Uslan, D. Z., Borge, R., Gottipaty, V. K., Shinn, T., Dan, D., Feldman, L. A., et al. (2010) · Circulation · doi:10.1161/circulationaha.110.976076 · Cited by 778
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2012.11.007 · Cited by 748
Self-citationVarma, N., Epstein, A. E., Irimpen, A., Schweikert, R., & Love, C. (2010) · Circulation · doi:10.1161/circulationaha.110.937409 · Cited by 594
Self-citationUdo, E. O., Zuithoff, N. P. A., van Hemel, N. M., de Cock, C. C., Hendriks, T., Doevendans, P. A., & Moons, K. G. M. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2011.12.014 · Cited by 597
Slotwiner, D. J., Varma, N., Akar, J. G., Annas, G. J., Beardsall, M., Fogel, R. I., Galizio, N. O., Glotzer, T. V., Leahy, R. A., Love, C. J., et al. (2015) · Heart Rhythm · doi:10.1016/j.hrthm.2015.05.008 · Cited by 533
Self-citationThe CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial
Crossley, G. H., Boyle, A., Vitense, H., Chang, Y., & Mead, R. H. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.12.012 · Cited by 526
Saxon, L. A., Hayes, D. L., Gilliam, F. R., Heidenreich, P. A., Day, J., Seth, M., Meyer, T. E., Jones, P. W., & Boehmer, J. P. (2010) · Circulation · doi:10.1161/circulationaha.110.960633 · Cited by 515
Self-citationLampert, R., Hayes, D. L., Annas, G. J., Farley, M. A., Goldstein, N. E., Hamilton, R. M., Kay, G. N., Kramer, D. B., Mueller, P. S., Padeletti, L., et al. (2010) · Heart Rhythm · doi:10.1016/j.hrthm.2010.04.033 · Cited by 434
Self-citationLeft ventricular non-compaction revisited: a distinct phenotype with genetic heterogeneity?
Oechslin, E. N., & Jenni, R. (2011) · European Heart Journal · doi:10.1093/eurheartj/ehq508 · Cited by 423
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2012.07.009 · Cited by 412
Self-citationBongiorni, M. G., Burri, H., Deharo, J., Starck, C. T., Kennergren, C. E. H., Sághy, L., Rao, A., Tascini, C., Lever, N. A., Kutarski, A., et al. (2018) · EP Europace · doi:10.1093/europace/euy050 · Cited by 404
Self-citationPadeletti, L., Arnar, D. O., Boncinelli, L., Brachman, J., Camm, A. J., Daubert, J., Kassam, S., Deliëns, L., Glikson, M., Hayes, D. L., et al. (2010) · EP Europace · doi:10.1093/europace/euq275 · Cited by 382
Self-citationWhellan, D. J., Ousdigian, K. T., Al‐Khatib, S. M., Pu, W., Sarkar, S., Porter, C., Pavri, B. B., & O’Connor, C. M. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.11.089 · Cited by 373
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Circulation · doi:10.1161/cir.0b013e3182618569 · Cited by 373
Self-citationLandolina, M., Perego, G. B., Lunati, M., Curnis, A., Guenzati, G., Vicentini, A., Parati, G., Borghi, G., Zanaboni, P., Valsecchi, S., et al. (2012) · Circulation · doi:10.1161/circulationaha.111.088971 · Cited by 344
Crossley, G. H., Poole, J. E., Rozner, M. A., Asirvatham, S. J., Cheng, A., Chung, M. K., Ferguson, T. B., Gallagher, J. D., Gold, M. R., HOYT, R. H., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2010.12.023 · Cited by 337
Guédon-Moreau, L., Lacroix, D., Sadoul, N., Clémenty, J., Kouakam, C., Js, H., Aliot, E., Boursier, M., Bizeau, O., Kacet, S., et al. (2012) · European Heart Journal · doi:10.1093/eurheartj/ehs425 · Cited by 307
Crossley, G. H., Poole, J. E., Rozner, M. A., Asirvatham, S. J., Cheng, A., Chung, M. K., Ferguson, T. B., Gallagher, J. D., Gold, M. R., HOYT, R. H., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2011.05.010 · Cited by 300
Remote Monitoring of Implantable Cardioverter-Defibrillators
Parthiban, N., Esterman, A., Mahajan, R., Twomey, D. J., Pathak, R. K., Lau, D. H., Roberts‐Thomson, K. C., Young, G. D., Sanders, P., & Ganesan, A. N. (2015) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2015.04.029 · Cited by 247
Varma, N., Piccini, J. P., Snell, J. D., Fischer, A., Dalal, N., & Mittal, S. (2015) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2015.04.033 · Cited by 232
Clinical Benefits of Remote Versus Transtelephonic Monitoring of Implanted Pacemakers
Crossley, G. H., Chen, J., Choucair, W., Cohen, T. J., Gohn, D. C., Johnson, W. B., Kennedy, E. E., Mongeon, L. R., Serwer, G. A., Qiao, H., et al. (2009) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.10.001 · Cited by 202
Self-citationHome Monitoring for Heart Failure Management
Bui, A. L., & Fonarow, G. C. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.09.044 · Cited by 204
Schwamm, L. H., Chumbler, N. R., Brown, E., Fonarow, G. C., Berube, D., Nyström, K. V., Suter, R. E., Zavala, M., Polsky, D. E., Radhakrishnan, K., et al. (2016) · Circulation · doi:10.1161/cir.0000000000000475 · Cited by 212
Charitos, E. I., Stierle, U., Ziegler, P. D., Baldewig, M., Robinson, D. R., Sievers, H.-H., & Hanke, T. (2012) · Circulation · doi:10.1161/circulationaha.112.098079 · Cited by 196
Tarakji, K. G., Wazni, O. M., Harb, S. C., Hsu, A., Saliba, W. I., & Wilkoff, B. L. (2014) · EP Europace · doi:10.1093/europace/euu147 · Cited by 185
Self-citationDubner, S., Auricchio, A., Steinberg, J. S., Vardas, P. E., Stone, P. H., Brugada, J., Piotrowicz, R., Hayes, D. L., Kirchhof, P. F., BREITHARDT, G. E., et al. (2012) · EP Europace · doi:10.1093/europace/eur303 · Cited by 178
Self-citationAutomatic Remote Monitoring of Implantable Cardioverter-Defibrillator Lead and Generator Performance
Varma, N., Michalski, J., Epstein, A. E., & Schweikert, R. A. (2010) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.110.951962 · Cited by 164
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.08.021 · Cited by 163
Self-citationShah, M. J., Silka, M. J., Silva, J. N. A., Balaji, S., Beach, C. M., Benjamin, M. N., Berul, C. I., Cannon, B. C., Cecchin, F., Cohen, M. I., et al. (2021) · Heart Rhythm · doi:10.1016/j.hrthm.2021.07.038 · Cited by 155
Long-term patient acceptance of and satisfaction with implanted device remote monitoring
Ricci, R. P., Morichelli, L., Quarta, L., Sassi, A. D. G., Porfili, A., Laudadio, M. T., Gargaro, A., & Santini, M. (2010) · EP Europace · doi:10.1093/europace/euq046 · Cited by 138
Ricci, R. P., Morichelli, L., D’Onofrio, A., Calò, L., Vaccari, D., Zanotto, G., Curnis, A., Buja, G., Rovai, N., & Gargaro, A. (2013) · EP Europace · doi:10.1093/europace/eus440 · Cited by 135
Ferrick, A. M., Raj, S. R., Deneke, T., Kojodjojo, P., López-Cabanillas, N., Abe, H., Bovéda, S., Chew, D. S., Choi, J., Dagres, N., et al. (2023) · Heart Rhythm · doi:10.1016/j.hrthm.2023.03.1525 · Cited by 122
Klersy, C., Boriani, G., De Silvestri, A., Mairesse, G. H., Braunschweig, F., Scotti, V., Balduini, A., Cowie, M. R., Leyva, F., & for the Health Economics Committee of the European Heart Rhythm Association (2016) · European Journal of Heart Failure · doi:10.1002/ejhf.470 · Cited by 120
Self-citationFernandez, F., Osca, J., Romero, R., Lozano, I. F., Larrazabal, J. M., Ferrer, J. M., Ortiz, R., Pombo, M., Tornés, F. J., & Kolbolandi, M. M. (2019) · European Heart Journal · doi:10.1093/eurheartj/ehz067 · Cited by 117
Klersy, C., De Silvestri, A., Gabutti, G., Raisaro, A., Curti, M., Regoli, F., & Auricchio, A. (2011) · European Journal of Heart Failure · doi:10.1093/eurjhf/hfq232 · Cited by 150
Self-citationRecommendations from the Heart Rhythm Society Task Force on Lead Performance Policies and Guidelines
Maisel, W. H., Hauser, R. G., Hammill, S. C., Hauser, R. G., Ellenbogen, K. A., Epstein, A. E., Hayes, D. L., Alpert, J. S., Berger, R. D., Curtis, A. B., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.04.024 · Cited by 114
Self-citationBraunschweig, F., Boriani, G., Bauer, A., Hatala, R., Herrmann‐Lingen, C., Kautzner, J., Pedersen, S. S., Pehrson, S. M., Ricci, R. P., & Schalij, M. J. (2010) · EP Europace · doi:10.1093/europace/euq316 · Cited by 116
Self-citationIsolated Noncompaction of the Left Ventricle in Adults
Hussein, A. A., Karimianpour, A., Collier, P. H., & Krasuski, R. A. (2015) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2015.06.017 · Cited by 105
Telemedicine and cardiac implants: what is the benefit?
Varma, N., & Ricci, R. P. (2013) · European Heart Journal · doi:10.1093/eurheartj/ehs388 · Cited by 102
Swerdlow, C. D., Gunderson, B. D., Ousdigian, K. T., Abeyratne, A., Sachanandani, H., & Ellenbogen, K. A. (2010) · Circulation · doi:10.1161/circulationaha.110.962407 · Cited by 99
Self-citationVijgen, J. M., Botto, G., Camm, J. A., Höijer, C. J., Jung, W., Heuzey, J. L., Lubiński, A., Norekvål, T. M., Santomauro, M., Schalij, M. J., et al. (2009) · EP Europace · doi:10.1093/europace/eup112 · Cited by 94
Self-citationRemote monitoring of cardiovascular devices: a time and activity analysis
Cronin, E. M., Ching, E. A., Varma, N., Martin, D. O., Wilkoff, B. L., & Lindsay, B. D. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.08.002 · Cited by 95
Self-citationHowlett, J. G., McKelvie, R. S., Arnold, J. M. O., Costigan, J., Dorian, P., Ducharme, A., Estrella-Holder, E., Ezekowitz, J. A., Giannetti, N., Haddad, H. A., et al. (2009) · Canadian Journal of Cardiology · doi:10.1016/s0828-282x(09)70477-5 · Cited by 90
Self-citationFerrick, A. M., Raj, S. R., Deneke, T., Kojodjojo, P., López-Cabanillas, N., Abe, H., Bovéda, S., Chew, D. S., Choi, J., Dagres, N., et al. (2023) · EP Europace · doi:10.1093/europace/euad123 · Cited by 89
Use of Remote Monitoring of Newly Implanted Cardioverter-Defibrillators
Akar, J. G., Bao, H., Jones, P. W., Wang, Y., Chaudhry, S. I., Varosy, P. D., Masoudi, F. A., Steín, K. M., Saxon, L. A., & Curtis, J. P. (2013) · Circulation · doi:10.1161/circulationaha.113.002481 · Cited by 86
Hindricks, G., Elsner, C., Piorkowski, C., Táborský, M., Geller, J. C., Schumacher, B., BYTESNIK, J., & Kottkamp, H. (2013) · European Heart Journal · doi:10.1093/eurheartj/eht207 · Cited by 82
Kramer, D. B., Kesselheim, A. S., Brock, D. W., & Maisel, W. H. (2010) · Heart Rhythm · doi:10.1016/j.hrthm.2010.07.018 · Cited by 77
Worldwide Randomized Antibiotic EnveloPe Infection PrevenTion Trial (WRAP-IT)
Tarakji, K. G., Mittal, S., Kennergren, C. E. H., Corey, R. G., Poole, J. E., Stromberg, K. D., Lexcen, D. R., & Wilkoff, B. L. (2016) · American Heart Journal · doi:10.1016/j.ahj.2016.06.010 · Cited by 72
Self-citationShah, M. J., Silka, M. J., Silva, J. N. A., Balaji, S., Beach, C. M., Benjamin, M. N., Berul, C. I., Cannon, B. C., Cecchin, F., Cohen, M. I., et al. (2021) · Cardiology in the Young · doi:10.1017/s1047951121003413 · Cited by 78
Simone, A. D., Leoni, L., Luzi, M., Amellone, C., Stabile, G., Rocca, V. L., Capucci, A., D’Onofrio, A., Ammendola, E., Accardi, F., et al. (2015) · EP Europace · doi:10.1093/europace/euu318 · Cited by 69
Akar, J. G., Bao, H., Jones, P. W., Wang, Y., Varosy, P. D., Masoudi, F. A., Steín, K. M., Saxon, L. A., Normand, S. T., & Curtis, J. P. (2015) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.114.003030 · Cited by 67
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2012.08.032 · Cited by 67
Self-citationGuasti, L., Dilaveris, P., Mamas, M. A., Richter, D. J., Christodorescu, R., Lumens, J., Schuuring, M. J., Carugo, S., Afilalo, J., Ferrini, M., et al. (2022) · ESC Heart Failure · doi:10.1002/ehf2.14022 · Cited by 74
Self-citationHeart Failure Management in Skilled Nursing Facilities
Jurgens, C. Y., Goodlin, S. J., Dolansky, M. A., Ahmed, A., Fonarow, G. C., Boxer, R. S., Arena, R. A., Blank, L., Buck, H. G., Cranmer, K. W., et al. (2015) · Circulation Heart Failure · doi:10.1161/hhf.0000000000000005 · Cited by 66
Boriani, G., Burri, H., Mantovani, L. G., Maniadakis, N., Leyva, F., Kautzner, J., Lubiński, A., Braunschweig, F., Jung, W., Lozano, I. F., et al. (2011) · EP Europace · doi:10.1093/europace/eur080 · Cited by 62
Self-citationDigital Health: Implications for Heart Failure Management
Singhal, A., & Cowie, M. (2021) · Cardiac failure review · doi:10.15420/cfr.2020.28 · Cited by 60
Self-citationSudden Death in the Young: Information for the Primary Care Provider
Erickson, C. C., Salerno, J. C., Berger, S., Campbell, R., Cannon, B. C., Christiansen, J., Moffatt, K. A., Pflaumer, A., Snyder, C. S., Srinivasan, C., et al. (2021) · PEDIATRICS · doi:10.1542/peds.2021-052044 · Cited by 60
End‐of‐Life Care in Patients with Implantable Cardioverter Defibrillators: A MADIT‐II Substudy
Sherazi, S., McNitt, S. A., Aktaş, M. K., Polonsky, B., Shah, A. H., Moss, A. J., Daubert, J. P., & Zaręba, W. (2013) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12188 · Cited by 57
Patient satisfaction and suggestions for improvement of remote ICD monitoring
Petersen, H. H., Larsen, M. C. J., Nielsen, O. W., Kensing, F., & Svendsen, J. H. (2012) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-012-9675-4 · Cited by 57
Burri, H., Sticherling, C., Wright, D. M., Makino, K., Smala, A. M., & Tilden, D. (2013) · EP Europace · doi:10.1093/europace/eut070 · Cited by 56
Applying health economics for policy decision making: do devices differ from drugs?
Sorenson, C., Tarricone, R., Siebert, M., & Drummond, M. (2011) · EP Europace · doi:10.1093/europace/eur089 · Cited by 55
Timmermans, I., Meine, M., Szendey, I., Aring, J., Roldán, J. R., van Erven, L., Kahlert, P., Zitron, E., Mabo, P., Denollet, J., et al. (2018) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.13574 · Cited by 53
Remote Monitoring for Advisories: Automatic Early Detection of Silent Lead Failure
Varma, N. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02314.x · Cited by 52
Shah, M. J., Silka, M. J., Silva, J. N. A., Balaji, S., Beach, C. M., Benjamin, M. N., Berul, C. I., Cannon, B. C., Cecchin, F., Cohen, M. I., et al. (2021) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2021.07.009 · Cited by 49
Remote monitoring: a cost or an investment?
Burri, H., Heidbuchel, H., Jung, W., & Brugada, P. (2011) · EP Europace · doi:10.1093/europace/eur082 · Cited by 49
Home Monitoring of Cardiac Devices in the Era of COVID-19
Miller, J. C., Skoll, D., & Saxon, L. A. (2020) · Current Cardiology Reports · doi:10.1007/s11886-020-01431-w · Cited by 48
Dubner, S., Auricchio, A., Steinberg, J. S., Vardas, P., Stone, P., Brugada, J., Piotrowicz, R., Hayes, D. L., Kirchhof, P., Breithardt, G., et al. (2012) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2011.00484.x · Cited by 47
Self-citationMittal, S., Piccini, J. P., Snell, J., Prillinger, J. B., Dalal, N., & Varma, N. (2016) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-016-0112-y · Cited by 47
Follow-up of Patients With New Cardiovascular Implantable Electronic Devices
Al‐Khatib, S. M., Mi, X., Wilkoff, B. L., Qualls, L. G., Frazier‐Mills, C. G., Setoguchi, S., Hess, P. L., & Curtis, L. H. (2012) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.112.974337 · Cited by 47
Self-citationRemote monitoring of cardiac implantable electronic devices (CIED)
Zeitler, E. P., & Piccini, J. P. (2016) · Trends in Cardiovascular Medicine · doi:10.1016/j.tcm.2016.03.012 · Cited by 46
Zellner, R. A., Aulisio, M. P., & Lewis, W. R. (2009) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.109.848523 · Cited by 43
Varma, N., Love, C. J., Michalski, J., & Epstein, A. E. (2021) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2021.01.008 · Cited by 44
Self-citationChiu, C. S. L., Timmermans, I., Versteeg, H., Zitron, E., Mabo, P., Pedersen, S. S., Meine, M., & for the REMOTE-CIED Trial Investigators (2021) · EP Europace · doi:10.1093/europace/euab221 · Cited by 42
Qintar, M., Zardkoohi, O., Hammadah, M., HSU, A., Wazni, O. M., Wilkoff, B. L., & Tarakji, K. G. (2014) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12514 · Cited by 42
Self-citationMalloy-Walton, L. E., Gingerich, J. C., Olson, M. D., & Atkins, D. L. (2013) · Pediatric Cardiology · doi:10.1007/s00246-013-0774-5 · Cited by 42
Long-term outcome of cardiac pacing in octogenarians and nonagenarians
Udo, E. O., van Hemel, N. M., Zuithoff, N. P. A., Kelder, J. C., Crommentuijn, H. A., Koopman-Verhagen, A. M., Voskuil, T. L. H. M., Doevendans, P. A., & Moons, K. G. M. (2011) · EP Europace · doi:10.1093/europace/eur329 · Cited by 41
Compliance with Remote Monitoring of ICDS/CRTDS in a Real‐World Population
Rosenfeld, L. E., Patel, A. S., Ajmani, V. B., Holbrook, R., & BRAND, T. A. (2014) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12358 · Cited by 40
Boriani, G., Auricchio, A., Klersy, C., Kirchhof, P. F., Brugada, J., Morgan, J., & Vardas, P. E. (2011) · EP Europace · doi:10.1093/europace/eur026 · Cited by 40
Self-citationVOGTMANN, T., Stiller, S., Marek, A., Kespohl, S., Gomer, M., Kühlkamp, V., Zach, G., Löscher, S., & Baumann, G. (2012) · EP Europace · doi:10.1093/europace/eus252 · Cited by 39
Silka, M. J., Shah, M. J., Silva, J. N. A., Balaji, S., Beach, C. M., Benjamin, M. N., Berul, C. I., Cannon, B. C., Cecchin, F., Cohen, M. I., et al. (2021) · Heart Rhythm · doi:10.1016/j.hrthm.2021.07.051 · Cited by 39
Remote follow-up and continuous remote monitoring, distinguished
Burri, H. (2013) · EP Europace · doi:10.1093/europace/eut071 · Cited by 38
Benezet‐Mazuecos, J., Campal, J. M. R., Cortés, M., Iglesias, J. A., Calle, S., de la Vieja, J. J., Quinones, M. A., Sanchez-Borque, P., de la Cruz, E., Espejo, A., et al. (2014) · EP Europace · doi:10.1093/europace/euu267 · Cited by 37
(2018) · Europe PMC (PubMed Central) · Cited by 38
Burri, H., Quesada, A., Ricci, R. P., Boriani, G., Davinelli, M., Favale, S., Costa, A. D., Kautzner, J., Moser, R., Navarro, X., et al. (2010) · American Heart Journal · doi:10.1016/j.ahj.2010.04.005 · Cited by 36
Self-citationVarma, N., Love, C. J., Schweikert, R. A., Moll, P., Michalski, J., Epstein, A. E., & TRUST Investigators (2017) · EP Europace · doi:10.1093/europace/eux059 · Cited by 36
Self-citationManaging with pacemakers and implantable cardioverter defibrillators
Lampert, R. (2013) · Circulation · doi:10.1161/circulationaha.113.001555 · Cited by 34
Agostinho, A. M., James, G. A., Wazni, O. M., Citron, M., & Wilkoff, B. L. (2009) · Clinical and Translational Science · doi:10.1111/j.1752-8062.2009.00123.x · Cited by 35
Self-citationRemote monitoring and follow-up of cardiovascular implantable electronic devices in the Netherlands
de Cock, C. C., Elders, J., van Hemel, N. M., van den Broek, K., van Erven, L., de Mol, B. A. J. M., Talmon, J. L., Theuns, D. A. M. J., & de Voogt, W. (2012) · Netherlands Heart Journal · doi:10.1007/s12471-011-0239-5 · Cited by 34
Heart Failure Management in Skilled Nursing Facilities
Jurgens, C. Y., Goodlin, S. J., Dolansky, M. A., Ahmed, A., Fonarow, G. C., Boxer, R. S., Arena, R. A., Blank, L., Buck, H. G., Cranmer, K. W., et al. (2015) · Journal of Cardiac Failure · doi:10.1016/j.cardfail.2015.02.007 · Cited by 36
Patients with Pacemaker or Implantable Cardioverter-Defibrillator
Schulman, P. M., Rozner, M. A., Sera, V., & Stecker, E. C. (2013) · Medical Clinics of North America · doi:10.1016/j.mcna.2013.05.004 · Cited by 35
Kypta, A., Blessberger, H., SALEH, K., Hönig, S., Kammler, J., Neeser, K., & Steinwender, C. (2014) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12488 · Cited by 33
Remote monitoring costs, benefits, and reimbursement: a European perspective
Chronaki, C. E., & Vardas, P. E. (2013) · EP Europace · doi:10.1093/europace/eut110 · Cited by 33
Self-citationVersteeg, H., Pedersen, S. S., Mastenbroek, M. H., Redekop, W. K., Schwab, J. O., Mabo, P., & Meine, M. (2014) · Netherlands Heart Journal · doi:10.1007/s12471-014-0587-z · Cited by 32
Effective communication and ethical consent in decisions related to ICDs
Clark, A. M., Jaarsma, T., Strachan, P. H., Davidson, P. M., Jerke, M., Beattie, J. M., Duncan, A. S., Ski, C. F., & Thompson, D. R. (2011) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2011.101 · Cited by 32
Castellanos, J. M., Smith, L. M., Varosy, P. D., Dehlendorf, C. E., & Marcus, G. M. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.01.029 · Cited by 32
Braunschweig, F., Anker, S. D., Proff, J., & Varma, N. (2019) · EP Europace · doi:10.1093/europace/euz011 · Cited by 31
Causes of Failure to Capture in Pacemakers and Implantable Cardioverter-defibrillators
Sabbagh, E., Abdelfattah, T., Karim, M., Farah, A., Grubb, B. P., & Karim, S. (2020) · Journal of Innovations in Cardiac Rhythm Management · doi:10.19102/icrm.2020.110207 · Cited by 32
Remote monitoring for active cardiovascular implantable electronic devices: a European survey
Halimi, F., & Cantù, F. (2010) · EP Europace · doi:10.1093/europace/euq399 · Cited by 31
Metadata
- Authors
- Wilkoff, B. L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K. A., Gillis, A. M., Hayes, D. L., Howlett, J. G., Kautzner, J., Love, C. J., Morgan, J. M., Priori, S. G., Reynolds, D. W., Schoenfeld, M. H., & Vardas, P. E.
- Affiliations
- Cleveland Clinic (US)
- Istituto Cardiocentro Ticino (CH)
- Virginia Commonwealth University (US)
- University of Calgary (CA)
- Mayo Clinic (US)
- Dalhousie University (CA)
- Institute of Clinical and Experimental Medicine (CZ)
- The Ohio State University (US)
- Istituti Clinici Scientifici Maugeri (IT)
- University of Oklahoma (US)
- Yale University (US)
- University Hospital of Heraklion (GR)
- ORCID
- Venue
- Heart Rhythm
- Publisher
- Elsevier BV
- Published
- 2008-06
- Volume
- 5
- Issue
- 6
- Pages
- 907-925
- ISSN
- 1547-5271 (ISSN-L)
- OpenAlex
- W2116854993
- Cited by
- 364
- References
- 66
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 4 · 2023: 10 · 2022: 11 · 2021: 20 · 2020: 11 · 2019: 14 · 2018: 14 · 2017: 22 · 2016: 18 · 2015: 30 · 2014: 22 · 2013: 40 · 2012: 44
- FWCI
- 23.37
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Heart Rhythm (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Infective Endocarditis Diagnosis and Management
- Keywords
- cardiac implantable electronic devices, cardiac pacemaker, implantable cardioverter-defibrillator, device follow-up, data management, regulatory compliance, reimbursement, ethical considerations, monitoring frequency, patient safety
- MeSH terms
- American Heart Association, Electric Countershock, Europe, Heart Failure, Humans, Monitoring, Physiologic, Societies, Medical, United States, Defibrillators, Implantable, Consensus
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review