Clinical Benefits of Remote Versus Transtelephonic Monitoring of Implanted Pacemakers
Open accessAbstract
Objectives The purpose of this study was to evaluate remote pacemaker interrogation for the earlier diagnosis of clinically actionable events compared with traditional transtelephonic monitoring and routine in-person evaluation.Background Pacemaker patient follow-up procedures have evolved from evaluating devices with little programmability and diagnostic information solely in person to transtelephonic rhythm strip recordings that allow monitoring of basic device function. More recently developed remote monitoring technology leverages expanded device capabilities, augmenting traditional transtelephonic monitoring to evaluate patients via full device interrogation.Methods The time to first diagnosis of a clinically actionable event was compared in patients who were followed by remote interrogation (Remote) and those who were followed per standard of care with office visits augmented by transtelephonic monitoring (Control). Patients were randomized 2:1. Remote arm patients transmitted pacemaker information at 3-month intervals. Control arm patients with a single-chamber pacemaker transmitted at 2-month intervals. Control arm patients with dual-chamber devices transmitted at 2-month intervals with an office visit at 6 months. All patients were seen in office at 12 months.Results The mean time to first diagnosis of clinically actionable events was earlier in the Remote arm (5.7 months) than in the Control arm (7.7 months). Three (2%) of the 190 events in the Control arm and 446 (66%) of 676 events in the Remote arm were identified remotely.Conclusions The strategic use of remote pacemaker interrogation follow-up detects actionable events that are potentially important more quickly and more frequently than transtelephonic rhythm strip recordings. The use of transtelephonic rhythm strips for pacemaker follow-up is of little value except for battery status determinations. (PREFER [Pacemaker Remote Follow-up Evaluation and Review]; NCT00294645).
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the PREFER trial, clinically actionable events such as atrial arrhythmias, increase in right ventricular pacing, lead function changes, and battery depletion were detected earlier in the remote-monitoring arm.
“In the Pacemaker Remote Follow-up Evaluation and Review (PREFER) trial, which randomly assigned 980 PM patients to follow-up with remote monitoring versus in-office visits plus transtelephonic monitoring, clinically actionable events such as atrial arrhythmias, increase in right ventricular pacing, lead function changes, and battery depletion, were detected earlier in the remote-monitoring arm.57”
From Managing with pacemakers and implantable cardioverter defibrillators (2013) · cites this paper as [57]
The Abstract and participant-flow/results section report that clinically actionable events were detected earlier in the remote arm (mean 5.7 versus 7.7 months). However, the Clinically Actionable Events table says lead-function changes and battery-depletion indicators were not used in the analysis because events were too few to establish earlier detection for those categories. The participant-flow section also says 980 patients were screened, but only 897 were randomized. (LLM, conf=0.91)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities
Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.02.032 · Cited by 2,546 · see citing works
A Class of Rank Test Procedures for Censored Survival Data
Harrington, D. P., & Fleming, T. R. (1982) · Biometrika · doi:10.1093/biomet/69.3.553 · Cited by 1,194 · see citing works
Wilkoff, B. L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K. A., Gillis, A. M., Hayes, D. L., Howlett, J. G., Kautzner, J., Love, C. J., et al. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.04.013 · Cited by 364 · see citing works
Self-citationEpstein, A. E., Dimarco, J., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Circulation · doi:10.1161/circualtionaha.108.189741 · Cited by 363 · see citing works
Lazarus, A. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00595.x · Cited by 227 · see citing works
Clinical Cardiac Pacing, Defibrillation, and Resynchronization Therapy
(2007) · Elsevier eBooks · doi:10.1016/b978-1-4160-2536-8.x5001-8 · Cited by 224 · see citing works
Remote Monitoring of Implantable Cardioverter Defibrillators: A Prospective Analysis
SCHOENFELD, M. H., COMPTON, S. J., MEAD, R. H., WEISS, D. N., SHERFESEE, L., ENGLUND, J., & MONGEON, L. R. (2004) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2004.00524.x · Cited by 194 · see citing works
Self-citationOrlov, M. V., Ghali, J. K., Araghi‐Niknam, M., Sherfesee, L., Sahr, D. M., Hettrick, D. A., & for the Atrial High Rate Trial Investigators (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00682.x · Cited by 119 · see citing works
Sheard, L., Wright, N. M. J., Adams, C. E., Bound, N., Rushforth, B., Hart, R., & Tompkins, C. N. E. (2009) · Trials · doi:10.1186/1745-6215-10-53 · Cited by 117 · see citing works
Transtelephone pacemaker clinic
Furman, S., Parker, B., & Escher, D. J. W. (1971) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/s0022-5223(19)42190-9 · Cited by 61 · see citing works
Practical Aspects of Implantable Defibrillator Therapy in Patients with Congenital Heart Disease
Walsh, E. P. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.00954.x · Cited by 44 · see citing works
Transtelephonic Monitoring of 25,919 Implanted Pacemakers
Dreifus, L. S., ZINBERG, A., Hurzeler, P., PUZIAK, A. D., Pennock, R. S., Feldman, M., & Morse, D. P. (1986) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1986.tb04491.x · Cited by 44 · see citing works
Lunati, M., Gasparini, M., Santini, M., Landolina, M., Perego, G. B., Pappone, C., Marzegalli, M., ARGIOLAS, C., Murthy, A., Valsecchi, S., et al. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00923.x · Cited by 41 · see citing works
Sophisticated Remote Implantable Cardioverter‐Defibrillator Follow‐Up: A Status Report
Schoenfeld, M. H., & REYNOLDS, D. W. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.09554.x · Cited by 33 · see citing works
Chen, J., Wilkoff, B. L., Choucair, W. K., Cohen, T. J., Crossley, G. H., Johnson, W. B., Mongeon, L., Serwer, G. A., & Sherfesee, L. (2008) · Trials · doi:10.1186/1745-6215-9-18 · Cited by 30 · see citing works
Self-citationGhali, J. K., Orlov, M. V., Araghi‐Niknam, M., Sherfesee, L., & Hettrick, D. A. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00772.x · Cited by 13 · see citing works
Pacemaker Troubleshooting and Follow-up
Love, C. J. (2011) · Elsevier eBooks · doi:10.1016/b978-1-4377-1616-0.00029-1 · Cited by 12 · see citing works
Instrumentation for the Follow-Up of Pacemade Patients.
Federico, A., Giori, F. A., Bhayana, J. N., Chardack, W. M., & MICHALEK, S. (1979) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1979.tb03651.x · Cited by 5 · see citing works
Flagged first, then by citations
1 of the 19 references couldn’t be matched to a known paper.
Research works printed but not linked: 1
- [1]1. Furman S, Parker B, Escher DJ. Trans-telephone pacemaker clinic. J Thorac Cardiovasc Surg 1971;61:827-34.
Cited by
Sorted by citations
Epstein, A. E., Dimarco, J., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2012) · Circulation · doi:10.1161/cir.0b013e318276ce9b · Cited by 1,495 · see citing works
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2012.11.007 · Cited by 738 · see citing works
Varma, N., Epstein, A. E., Irimpen, A., Schweikert, R., & Love, C. (2010) · Circulation · doi:10.1161/circulationaha.110.937409 · Cited by 594 · see citing works
Slotwiner, D. J., Varma, N., Akar, J. G., Annas, G. J., Beardsall, M., Fogel, R. I., Galizio, N. O., Glotzer, T. V., Leahy, R. A., Love, C. J., et al. (2015) · Heart Rhythm · doi:10.1016/j.hrthm.2015.05.008 · Cited by 533 · see citing works
Self-citationThe CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial
Crossley, G. H., Boyle, A., Vitense, H., Chang, Y., & Mead, R. H. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.12.012 · Cited by 526 · see citing works
Self-citationDaubert, J.-C., Saxon, L., Adamson, P. B., Auricchio, A., Berger, R. D., Beshai, J. F., Breithard, O., Brignole, M., Cleland, J., DeLurgio, D. B., et al. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.07.025 · Cited by 467 · see citing works
Self-citationTracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2012.07.009 · Cited by 411 · see citing works
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Circulation · doi:10.1161/cir.0b013e3182618569 · Cited by 369 · see citing works
Landolina, M., Perego, G. B., Lunati, M., Curnis, A., Guenzati, G., Vicentini, A., Parati, G., Borghi, G., Zanaboni, P., Valsecchi, S., et al. (2012) · Circulation · doi:10.1161/circulationaha.111.088971 · Cited by 344 · see citing works
Task Force Chairs, Daubert, J., Saxon, L. A., Adamson, P. B., Auricchio, A., Berger, R. D., Beshai, J. F., Breithard, O., Brignole, M., Cleland, J. G. F., et al. (2012) · EP Europace · doi:10.1093/europace/eus222 · Cited by 256 · see citing works
Self-citationGörenek, B., Bax, J. J., Boriani, G., Chen, S., Dagres, N., Glotzer, T. V., Healey, J. S., Israel, C. W., Kudaiberdieva, G., Levin, L., et al. (2017) · EP Europace · doi:10.1093/europace/eux163 · Cited by 249 · see citing works
Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study
Ritter, P., Duray, G. Z., Steinwender, C., Soejima, K., Omar, R., Mont, L., Boersma, L. V. A., Knops, R. E., Chinitz, L. A., Zhang, S., et al. (2015) · European Heart Journal · doi:10.1093/eurheartj/ehv214 · Cited by 230 · see citing works
Home Monitoring for Heart Failure Management
Bui, A. L., & Fonarow, G. C. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.09.044 · Cited by 204 · see citing works
Schwamm, L. H., Chumbler, N. R., Brown, E., Fonarow, G. C., Berube, D., Nyström, K. V., Suter, R. E., Zavala, M., Polsky, D. E., Radhakrishnan, K., et al. (2016) · Circulation · doi:10.1161/cir.0000000000000475 · Cited by 200 · see citing works
Dubner, S., Auricchio, A., Steinberg, J. S., Vardas, P. E., Stone, P. H., Brugada, J., Piotrowicz, R., Hayes, D. L., Kirchhof, P. F., BREITHARDT, G. E., et al. (2012) · EP Europace · doi:10.1093/europace/eur303 · Cited by 178 · see citing works
Automatic Remote Monitoring of Implantable Cardioverter-Defibrillator Lead and Generator Performance
Varma, N., Michalski, J., Epstein, A. E., & Schweikert, R. A. (2010) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.110.951962 · Cited by 164 · see citing works
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.08.021 · Cited by 163 · see citing works
Pandor, A., Thokala, P., Gomersall, T., Baalbaki, H., Stevens, J., Wang, J., Wong, R., Brennan, A., & Fitzgerald, P. (2013) · Health Technol Assess · doi:10.3310/hta17320 · Cited by 159 · see citing works
Ferrick, A. M., Raj, S. R., Deneke, T., Kojodjojo, P., López-Cabanillas, N., Abe, H., Bovéda, S., Chew, D. S., Choi, J., Dagres, N., et al. (2023) · Heart Rhythm · doi:10.1016/j.hrthm.2023.03.1525 · Cited by 122 · see citing works
Fernandez, F., Osca, J., Romero, R., Lozano, I. F., Larrazabal, J. M., Ferrer, J. M., Ortiz, R., Pombo, M., Tornés, F. J., & Kolbolandi, M. M. (2019) · European Heart Journal · doi:10.1093/eurheartj/ehz067 · Cited by 117 · see citing works
Klersy, C., De Silvestri, A., Gabutti, G., Raisaro, A., Curti, M., Regoli, F., & Auricchio, A. (2011) · European Journal of Heart Failure · doi:10.1093/eurjhf/hfq232 · Cited by 150 · see citing works
Boriani, G., Costa, A. D., Ricci, R. P., Quesada, A., Favale, S., Iacopino, S., Romeo, F., Risi, A., Stefano, L. M. D. S., Navarro, X., et al. (2013) · Journal of Medical Internet Research · doi:10.2196/jmir.2608 · Cited by 107 · see citing works
Telemedicine and cardiac implants: what is the benefit?
Varma, N., & Ricci, R. P. (2013) · European Heart Journal · doi:10.1093/eurheartj/ehs388 · Cited by 102 · see citing works
Remote monitoring of cardiovascular devices: a time and activity analysis
Cronin, E. M., Ching, E. A., Varma, N., Martin, D. O., Wilkoff, B. L., & Lindsay, B. D. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.08.002 · Cited by 95 · see citing works
Self-citationFerrick, A. M., Raj, S. R., Deneke, T., Kojodjojo, P., López-Cabanillas, N., Abe, H., Bovéda, S., Chew, D. S., Choi, J., Dagres, N., et al. (2023) · EP Europace · doi:10.1093/europace/euad123 · Cited by 89 · see citing works
Advances and Future Directions in Cardiac Pacemakers
Madhavan, M., Mulpuru, S. K., McLeod, C. J., Cha, Y., & Friedman, P. A. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2016.10.064 · Cited by 86 · see citing works
Simone, A. D., Leoni, L., Luzi, M., Amellone, C., Stabile, G., Rocca, V. L., Capucci, A., D’Onofrio, A., Ammendola, E., Accardi, F., et al. (2015) · EP Europace · doi:10.1093/europace/euu318 · Cited by 69 · see citing works
Cardiac resynchronization therapy: state of the art 2013
Yu, C.-M., & Hayes, D. L. (2013) · European Heart Journal · doi:10.1093/eurheartj/ehs454 · Cited by 68 · see citing works
Remote Management of Pacemaker Patients With Biennial In-Clinic Evaluation
Watanabe, E., Yamazaki, F., Goto, T., Asai, T., Yamamoto, T., Hirooka, K., Sato, T., Kasai, A., Ueda, M., Yamakawa, T., et al. (2020) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.119.007734 · Cited by 67 · see citing works
Tracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/j.jtcvs.2012.08.032 · Cited by 67 · see citing works
Svennberg, E., Caiani, E. G., Bruining, N., Desteghe, L., Han, J. K., Narayan, S. M., Rademakers, F. E., Sanders, P., & Duncker, D. (2023) · EP Europace · doi:10.1093/europace/euad176 · Cited by 64 · see citing works
JCS/JHRS 2021 Guideline Focused Update on Non-Pharmacotherapy of Cardiac Arrhythmias
Nogami, A., Kurita, T., KUSANO, K. F., Goya, M., Shoda, M., Tada, H., Naito, S., Yamane, T., Kimura, M., Shiga, T., et al. (2022) · Circulation Journal · doi:10.1253/circj.cj-21-0162 · Cited by 55 · see citing works
Yee, R., Verma, A., Beardsall, M., Fraser, J. A., Philippon, F., & Exner, D. V. (2013) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2012.11.036 · Cited by 54 · see citing works
Remote monitoring and telemedicine in heart failure: implementation and benefits
Imberti, J. F., Tosetti, A., Mei, D. A., Maisano, A., & Boriani, G. (2021) · Current Cardiology Reports · doi:10.1007/s11886-021-01487-2 · Cited by 53 · see citing works
Amara, W., Montagnier, C., Cheggour, S., Boursier, M., Gully, C., BARNAY, C., Georger, F., Deplagne, A., Fromentin, S., Młotek, M., et al. (2017) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.13062 · Cited by 49 · see citing works
Home Telemonitoring in Patients With Chronic Heart Failure
Schmidt, S., Schuchert, A., Krieg, T., & Oeff, M. (2010) · Deutsches Ärzteblatt international · doi:10.3238/arztebl.2010.0131 · Cited by 48 · see citing works
Home Monitoring of Cardiac Devices in the Era of COVID-19
Miller, J. C., Skoll, D., & Saxon, L. A. (2020) · Current Cardiology Reports · doi:10.1007/s11886-020-01431-w · Cited by 48 · see citing works
Dubner, S., Auricchio, A., Steinberg, J. S., Vardas, P., Stone, P., Brugada, J., Piotrowicz, R., Hayes, D. L., Kirchhof, P., Breithardt, G., et al. (2012) · Annals of Noninvasive Electrocardiology · doi:10.1111/j.1542-474x.2011.00484.x · Cited by 47 · see citing works
Varma, N., Pavri, B. B., Stambler, B. S., Michalski, J., & TRUST Investigators (2012) · EP Europace · doi:10.1093/europace/eus410 · Cited by 47 · see citing works
Follow-up of Patients With New Cardiovascular Implantable Electronic Devices
Al‐Khatib, S. M., Mi, X., Wilkoff, B. L., Qualls, L. G., Frazier‐Mills, C. G., Setoguchi, S., Hess, P. L., & Curtis, L. H. (2012) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.112.974337 · Cited by 47 · see citing works
Self-citationRemote monitoring of cardiac implantable electronic devices (CIED)
Zeitler, E. P., & Piccini, J. P. (2016) · Trends in Cardiovascular Medicine · doi:10.1016/j.tcm.2016.03.012 · Cited by 46 · see citing works
Remote Monitoring of Cardiac Implantable Electronic Devices
Cheung, C. C., & Deyell, M. W. (2018) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2018.01.003 · Cited by 45 · see citing works
Varma, N., Love, C. J., Michalski, J., & Epstein, A. E. (2021) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2021.01.008 · Cited by 44 · see citing works
JCS/JHRS 2021 guideline focused update on non‐pharmacotherapy of cardiac arrhythmias
Nogami, A., Kurita, T., KUSANO, K. F., Goya, M., Shoda, M., Tada, H., Naito, S., Yamane, T., Kimura, M., Shiga, T., et al. (2022) · Journal of Arrhythmia · doi:10.1002/joa3.12649 · Cited by 41 · see citing works
Performance of first pacemaker to use smart device app for remote monitoring
Tarakji, K. G., Zaidi, A. M., ZWEIBEL, S. L., Varma, N., Sears, S. F., Allred, J. D., Roberts, P. R., Shaik, N. A., Silverstein, J. R., Maher, A. R., et al. (2021) · Heart Rhythm O2 · doi:10.1016/j.hroo.2021.07.008 · Cited by 42 · see citing works
Compliance with Remote Monitoring of ICDS/CRTDS in a Real‐World Population
Rosenfeld, L. E., Patel, A. S., Ajmani, V. B., Holbrook, R., & BRAND, T. A. (2014) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12358 · Cited by 40 · see citing works
VOGTMANN, T., Stiller, S., Marek, A., Kespohl, S., Gomer, M., Kühlkamp, V., Zach, G., Löscher, S., & Baumann, G. (2012) · EP Europace · doi:10.1093/europace/eus252 · Cited by 39 · see citing works
(2018) · Europe PMC (PubMed Central) · Cited by 38 · see citing works
Managing with pacemakers and implantable cardioverter defibrillators
Lampert, R. (2013) · Circulation · doi:10.1161/circulationaha.113.001555 · Cited by 34 · see citing works
Maines, M., Palmisano, P., Greco, M. D., Melissano, D., Bonis, S. D., Baccillieri, S., Zanotto, G., D’Onofrio, A., Ricci, R. P., Ponti, R. D., et al. (2021) · Journal of Clinical Medicine · doi:10.3390/jcm10184086 · Cited by 33 · see citing works
Braunschweig, F., Anker, S. D., Proff, J., & Varma, N. (2019) · EP Europace · doi:10.1093/europace/euz011 · Cited by 31 · see citing works
Piotrowicz, R., Krzesiński, P., Balsam, P., Piotrowicz, E., Kempa, M., Lewicka, E., Główczyńska, R., Grabowski, M., Kołtowski, Ł., Peller, M., et al. (2021) · Kardiologia Polska · doi:10.33963/kp.15824 · Cited by 31 · see citing works
Bradyarrhythmias in the Elderly
Kumar, P., Kusumoto, F. M., & Goldschlager, N. (2012) · Clinics in Geriatric Medicine · doi:10.1016/j.cger.2012.08.004 · Cited by 31 · see citing works
Digital monitoring and care: Virtual medicine
Shinbane, J. S., & Saxon, L. A. (2016) · Trends in Cardiovascular Medicine · doi:10.1016/j.tcm.2016.05.007 · Cited by 31 · see citing works
Dechert, B. E., Serwer, G. A., Bradley, D. J., Dick, M., & LaPage, M. J. (2014) · Heart Rhythm · doi:10.1016/j.hrthm.2014.10.009 · Cited by 30 · see citing works
Self-citationRuvo, E. D., Sciarra, L., Martino, A. M., Rebecchi, M., Iulianella, R. V., Sebastiani, F., Fagagnini, A., Borrelli, A., Scarà, A., Grieco, D., et al. (2015) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-015-0067-4 · Cited by 30 · see citing works
Transparent sharing of digital health data: A call to action
Slotwiner, D. J., Tarakji, K. G., Al‐Khatib, S. M., PASSMAN, R. S., Saxon, L. A., Peters, N. S., McCall, D., Turakhia, M. P., Schaeffer, J., Mendenhall, G. S., et al. (2019) · Heart Rhythm · doi:10.1016/j.hrthm.2019.04.042 · Cited by 29 · see citing works
Postmarket surveillance of medical devices: current capabilities and future opportunities
Blake, K. (2013) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-013-9778-6 · Cited by 29 · see citing works
Remote patient management using implantable devices
Movsowitz, C., & Mittal, S. (2011) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-011-9548-2 · Cited by 28 · see citing works
Remote monitoring of implantable devices: should we continue to ignore it?
Bertini, M., Marcantoni, L., Toselli, T., & Ferrari, R. (2016) · International Journal of Cardiology · doi:10.1016/j.ijcard.2015.09.033 · Cited by 28 · see citing works
Maines, M., Tomasi, G., Moggio, P., Peruzza, F., Catanzariti, D., Angheben, C., Simoncelli, M., Degiampietro, M., Piffer, L., Valsecchi, S., et al. (2020) · Journal of Cardiovascular Medicine · doi:10.2459/jcm.0000000000001011 · Cited by 28 · see citing works
Remote monitoring of implantable cardiac devices
Ganeshan, R., Enriquez, A. D., & Freeman, J. V. (2017) · Current Opinion in Cardiology · doi:10.1097/hco.0000000000000471 · Cited by 27 · see citing works
Kelly, S., Campbell, D., Duhn, L., Giddens, K., Gillis, A. M., AbdelWahab, A. M., Nault, I., Raj, S. R., Lockwood, E. E., Basta, J., et al. (2020) · CJC Open · doi:10.1016/j.cjco.2020.11.010 · Cited by 27 · see citing works
Disease management: atrial fibrillation and Home Monitoring
Ricci, R. P. (2013) · EP Europace · doi:10.1093/europace/eut114 · Cited by 26 · see citing works
Current clinical evidence for remote patient management
Guédon-Moreau, L., Mabo, P., & Kacet, S. (2013) · EP Europace · doi:10.1093/europace/eut119 · Cited by 25 · see citing works
Douarin, Y.-M. L., Traversino, Y., Graciet, A., Josseran, A., Bili, A.-B., Blaise, L., Châtellier, G., Coulonjou, H., Delval, C., Detournay, B., et al. (2019) · Therapies · doi:10.1016/j.therap.2019.12.009 · Cited by 25 · see citing works
Recommendations on the utilization of telemedicine in cardiology
Working Group Rhythmology of the Austrian Cardiological Society, Gruska, M., Aigner, G., Altenberger, J., Burkart-Küttner, D., Fiedler, L., Gwechenberger, M., Lercher, P., Martinek, M., Nürnberg, M., et al. (2020) · Wiener klinische Wochenschrift · doi:10.1007/s00508-020-01762-2 · Cited by 24 · see citing works
Helms, T. M., Stockburger, M. A., Koehler, F., Leonhardt, V., Muller, A. M., Rybak, K., Sack, S., Schwab, J. O., Störk, S., Zippel‐Schultz, B., et al. (2019) · Herzschrittmachertherapie + Elektrophysiologie · doi:10.1007/s00399-018-0606-7 · Cited by 23 · see citing works
Varma, N. (2013) · EP Europace · doi:10.1093/europace/eut116 · Cited by 23 · see citing works
Perings, C. A., Bauer, W., Bondke, H., Mewis, C., James, M., Böcker, D., Broadhurst, P., Korte, T., Toft, E., Hintringer, F., et al. (2011) · EP Europace · doi:10.1093/europace/euq447 · Cited by 22 · see citing works
Li, Y., Pastori, D., Miyazawa, K., Shahid, F., & Lip, G. Y. H. (2021) · Journal of the American Heart Association · doi:10.1161/jaha.120.017519 · Cited by 22 · see citing works
ROSENTHAL, L. S., MESTER, S., Rakovec, P., Peñaranda, J. B., Sherman, J. R., Sheldon, T., Zeng, C., Wang, P. J., & for the CAPTURE Trial Investigators (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2010.02809.x · Cited by 22 · see citing works
Ferrick, A. M., Raj, S. R., Deneke, T., Kojodjojo, P., López-Cabanillas, N., Abe, H., Bovéda, S., Chew, D. S., Choi, J., Dagres, N., et al. (2023) · Journal of Arrhythmia · doi:10.1002/joa3.12851 · Cited by 21 · see citing works
Ruvo, E. D., Gargaro, A., Sciarra, L., Luca, L. D., Zuccaro, L. M., Stirpe, F., Rebecchi, M., Sette, A., Lioy, E., & Calò, L. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2010.02932.x · Cited by 20 · see citing works
Mascioli, G., Curnis, A., Landolina, M., Klersy, C., Gelmini, G., Ruffa, F., on behalf of ATHENS Investigators, Mascioli, G., Lucca, E., Bakhtadze, N., et al. (2011) · EP Europace · doi:10.1093/europace/eur233 · Cited by 20 · see citing works
Cardiac rhythm management devices
Stevenson, I., & Voskoboinik, A. (2018) · Australian Journal of General Practice · doi:10.31128/ajgp-12-17-4439 · Cited by 19 · see citing works
Emerging role of digital technology and remote monitoring in the care of cardiac patients.
Banchs, J., & Scher, D. (2015) · The Medical clinics of North America · doi:10.1016/j.mcna.2015.02.013 · Cited by 19 · see citing works
Device Features for Managing Patients with Heart Failure
Varma, N., & Wilkoff, B. L. (2011) · Heart Failure Clinics · doi:10.1016/j.hfc.2010.12.005 · Cited by 18 · see citing works
Self-citationMaines, M., Tomasi, G., Moggio, P., Poian, L., Peruzza, F., Catanzariti, D., Angheben, C., Cont, N., Valsecchi, S., & Greco, M. D. (2021) · International Journal of Cardiology · doi:10.1016/j.ijcard.2021.04.043 · Cited by 17 · see citing works
Cronin, E. M., & Varma, N. (2012) · Expert Review of Medical Devices · doi:10.1586/erd.12.18 · Cited by 17 · see citing works
Ethics and the cardiac pacemaker: more than just end-of-life issues
Hutchison, K., & Sparrow, R. (2017) · EP Europace · doi:10.1093/europace/eux019 · Cited by 17 · see citing works
Pron, G. E., Ieraci, L., & Kaulback, K. (2012) · Europe PMC (PubMed Central) · Cited by 17 · see citing works
HRS Policy Statement: Clinical Cardiac Electrophysiology Fellowship Curriculum: Update 2011
Link, M. S., Exner, D. V., Anderson, M. E., Ackerman, M., Al‐Ahmad, A., Knight, B. P., Markowitz, S. M., Kaufman, E. S., Haines, D. E., Asirvatham, S. J., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2011.06.008 · Cited by 16 · see citing works
Clinical recognition of pacemaker battery depletion and automatic reprogramming
Sinha, S. K., Chrispin, J., Barth, A. S., Rickard, J., SPRAGG, D. D., Berger, R. D., Calkins, H. G., Tomaselli, G. F., & Marine, J. E. (2017) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.13135 · Cited by 16 · see citing works
Holtzman, J. N., Wadhera, R. K., Choi, E. H., Zhao, T., Secemsky, E. A., Fraiche, A. M., Shen, C., & Kramer, D. B. (2020) · Heart Rhythm · doi:10.1016/j.hrthm.2020.05.044 · Cited by 16 · see citing works
Lim, P., Lee, A., Chua, K., Lim, E., Chong, D. W. S., Tan, B., Ho, K., Teo, W. S., & Ching, C. (2016) · Singapore Medical Journal · doi:10.11622/smedj.2016120 · Cited by 16 · see citing works
Jamil, H. A., Mohammed, S., Gierula, J., Paton, M. F., Lowry, J. E., Cubbon, R. M., Kearney, M. T., & Witte, K. K. (2018) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2018.10.040 · Cited by 15 · see citing works
Gramegna, L., Tomasi, C., Gasparini, G. M., Scaboro, G., Zanon, F., Boaretto, G., Tomei, R., & Tomasi, L. (2011) · EP Europace · doi:10.1093/europace/eur334 · Cited by 14 · see citing works
López-Villegas, A., Catalán-Matamoros, D., Saborido, C. M., Villegas-Tripiana, I., & Robles‐Musso, E. (2015) · Revista Española de Cardiología · doi:10.1016/j.recesp.2015.06.021 · Cited by 14 · see citing works
Technology-Assisted Congestive Heart Failure Care
Iyngkaran, P., Toukhsati, S. R., Biddagardi, N., Zimmet, H., Atherton, J. J., & Hare, D. L. (2015) · Current Heart Failure Reports · doi:10.1007/s11897-014-0251-3 · Cited by 14 · see citing works
The Use of App-based Follow-up of Cardiac Implantable Electronic Devices
Roberts, P. R., & ElRefai, M. (2020) · Cardiac failure review · doi:10.15420/cfr.2019.13 · Cited by 14 · see citing works
Zoppo, F., Facchin, D., Molon, G., Zanotto, G., Catanzariti, D., Rossillo, A., Baccillieri, S., MENARD, C., COMISSO, J., Gentili, A., et al. (2014) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12474 · Cited by 14 · see citing works
Pacemaker Remote Monitoring in the Pediatric Population: Is It A Real Solution?
Leoni, L., Padalino, M. A., Biffanti, R., Ferretto, S., Vettor, G., Corrado, D., Stellin, G., Milanesi, O., & Iliceto, S. (2015) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12600 · Cited by 14 · see citing works
Incidence and predictors of pacemaker reprogramming: potential consequences for remote follow-up
Udo, E. O., van Hemel, N. M., Zuithoff, N. P. A., Barrett, M. J., RUITER, J. H., Doevendans, P. A., & Moons, K. G. M. (2013) · EP Europace · doi:10.1093/europace/eut002 · Cited by 14 · see citing works
Giannola, G., Torcivia, R., Farulla, R. A., & Cipolla, T. (2018) · JMIR Cardio · doi:10.2196/cardio.9815 · Cited by 13 · see citing works
In‐home controls of pacemakers in debilitated elderly patients
Folino, A. F., Breda, R., Calzavara, P., Migliore, F., Iliceto, S., & Buja, G. (2011) · Geriatrics and gerontology international/Geriatrics & gerontology international · doi:10.1111/j.1447-0594.2011.00723.x · Cited by 13 · see citing works
Facchin, D., Baccillieri, S., GASPARINI, G., Zoppo, F., Allocca, G., Brieda, M., Verlato, R., & Proclemer, A. (2016) · International Journal of Cardiology · doi:10.1016/j.ijcard.2016.06.162 · Cited by 13 · see citing works
Mittal, S., Movsowitz, C., & Varma, N. (2014) · Cardiology Clinics · doi:10.1016/j.ccl.2014.01.001 · Cited by 12 · see citing works
Selby, P. J., Banks, R. E., Gregory, W. M., Hewison, J., Rosenberg, W. M., Altman, D. G., Deeks, J. J., McCabe, C. J., Parkes, J., Sturgeon, C. M., et al. (2018) · Programme Grants for Applied Research · doi:10.3310/pgfar06030 · Cited by 12 · see citing works
Home Monitoring for Cardiovascular Implantable Electronic Devices
Leahy, R. A., & Davenport, E. E. (2015) · AACN Advanced Critical Care · doi:10.1097/nci.0000000000000110 · Cited by 11 · see citing works
- Authors
- Crossley, G. H., Chen, J., Choucair, W., Cohen, T. J., Gohn, D. C., Johnson, W. B., Kennedy, E. E., Mongeon, L. R., Serwer, G. A., Qiao, H., & Wilkoff, B. L.
- Affiliations
- University of Tennessee at Knoxville (US)
- Washington University in St. Louis (US)
- Winthrop-University Hospital (US)
- Arkansas Heart Hospital (US)
- Medtronic (United States) (US)
- University of Michigan (US)
- Cleveland Clinic (US)
- Corresponding author
- Crossley, George H.
- Venue
- Journal of the American College of Cardiology
- Publisher
- Elsevier BV
- Published
- 2009-11
- Volume
- 54
- Issue
- 22
- Pages
- 2012-2019
- ISSN
- 0735-1097 (ISSN-L)
- OpenAlex
- W1989653793
- Cited by
- 202
- References
- 19
- Citations by year
- 2026: 3 · 2025: 3 · 2024: 2 · 2023: 11 · 2022: 5 · 2021: 13 · 2020: 12 · 2019: 9 · 2018: 11 · 2017: 11 · 2016: 14 · 2015: 18 · 2014: 11 · 2013: 27 · 2012: 27
- FWCI
- 11.21
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- Locations
- Journal of the American College of Cardiology (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Medtronic
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Telemedicine and Telehealth Implementation
- Keywords
- transtelephonic monitoring, implantable pacemaker, pacemaker follow-up, time to diagnosis, remote monitoring, device interrogation, battery status, single-chamber pacemaker, dual-chamber pacemaker
- MeSH terms
- Aged, Aged, 80 and over, Arrhythmias, Cardiac, Coronary Artery Disease, Diabetic Angiopathies, Female, Heart Failure, Humans, Male, Middle Aged, Monitoring, Physiologic, Office Visits, Pacemaker, Artificial, Prospective Studies, Telemetry, Telephone, Time Factors, Comorbidity
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review