CoChat

Real-world performance of an enhanced atrial fibrillation detection algorithm in an insertable cardiac monitor

Open access
Mittal, Suneet, Rogers, John, Sarkar, Shantanu, Koehler, Jodi, Warman, Eduardo N., Tomson, Todd T., Passman, Rod S. · 2016 · Heart Rhythm
Cited by 93

The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Insertable cardiac monitors (ICMs) are used for long-term ECG monitoring. The Reveal LINQ ICM has an improved atrial fibrillation (AF) detection algorithm. OBJECTIVE: The purpose of this study was to investigate the algorithm's real-world performance in patients with syncope, cryptogenic stroke, and known AF. METHODS: Consecutive patients with implanted ICM and AF detection parameters automatically set and maintained depending on the indication for monitoring were included. A single reviewer annotated all stored episodes after ICM implant. A second reviewer annotated a random sample of 10% of all detected AF episodes. The episode detection positive predictive value as well as true and false detection rates were determined for AF episodes of different durations. RESULTS: The study enrolled 3759 patients (1604 [43%] with syncope, 1049 [28%] with known AF, 1106 [29%] with cryptogenic stroke). Overall, 20,659 AF episodes were detected in 1020 patients. The gross episode detection positive predictive value was 84%, 73%, and 26% for all episodes (≥2 minutes) and improved to 97%, 95%, and 91% for detected AF episodes ≥1 hour in the syncope, known-AF, and cryptogenic stroke patient cohorts, respectively. The true (and false) detection rate was 0.23 (0.05), 3.8 (1.4), and 0.23 (0.65) per patient-month of monitoring for the syncope, known-AF, and cryptogenic stroke patient cohorts, respectively. Limiting ECG storage to the longest detected AF episode significantly reduced the burden of episode adjudication without significantly compromising the identification of patients with true AF. CONCLUSION: The performance of LINQ ICM is dependent on the AF incidence rate in the population being monitored, the programmed sensitivity of AF algorithm, and the duration of detected AF episodes.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.hrthm.2016.05.010

Cites

8 of 8 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 8 references

Cited by

See all 93

Metadata

Authors
Mittal, S., Rogers, J., Sarkar, S., Koehler, J., Warman, E. N., Tomson, T. T., & Passman, R. S.
Venue
Heart Rhythm
Publisher
Elsevier BV
Published
2016-08
Volume
13
Issue
8
Pages
1624-1630
ISSN
1547-5271 (ISSN-L)
OpenAlex
W2346237171
Cited by
93
References
8
FWCI
9.21
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Real-world performance of an enhanced atrial fibrillation detection algorithm in an insertable cardiac monitor. https://app.cochat.ai/papers/10.1016/j.hrthm.2016.05.010
The paper
Mittal, S., Rogers, J., Sarkar, S., Koehler, J., Warman, E. N., Tomson, T. T., & Passman, R. S. (2016). Real-world performance of an enhanced atrial fibrillation detection algorithm in an insertable cardiac monitor. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2016.05.010

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.