Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis
Open accessPublished version of a preprint
Abstract
BACKGROUND Photoplethysmography (PPG) is a technology routinely used in clinical practice to assess blood oxygenation (SpO2) and pulse rate (PR). Skin pigmentation may influence accuracy, leading to health outcomes disparities. OBJECTIVE This meta-analysis primarily aimed to evaluate the accuracy of PPG-derived SpO2 and PR by skin pigmentation. Secondarily, we aimed to evaluate statistical biases and the clinical relevance of PPG-derived SpO2 and PR according to skin pigmentation. METHODS We identified 23 pulse oximetry studies (N=59,684; 197,353 paired SpO2-arterial blood observations) and 4 wearable PR studies (N=176; 140,771 paired photoplethysmography-electrocardiography observations). We evaluated accuracy according to skin pigmentation group by comparing SpO2 accuracy root-mean-square (Arms) values to the regulatory threshold of 3% and PR 95% limits of agreement (LoA) values to ±5 bpm, according to the standards of the American National Standards Institute, Advancing Safety in Medical Technology, and the International Electrotechnical Commission. We evaluated biases and clinical relevance using mean bias and 95% confidence intervals (CI). RESULTS For SpO2, Arms were 3.96%, 4.71%, and 4.15% and pooled mean biases were 0.70% (95% CI: 0.17 to 1.22), 0.27% (95% CI: -0.64 to 1.19), and 1.27% (95% CI: 0.58 to 1.95) for light, medium, and dark pigmentation, respectively. For PR, 95% LoAs were -16.02 to 13.54, -18.62 to 16.84, and -33.69 to 32.54 and pooled mean biases were -1.24 bpm (95% CI: -5.31-2.83), -0.89 bpm (95% CI: -3.70-1.93), and -0.57 bpm (95% CI: -9.44-8.29) for light, medium, and dark pigmentation, respectively. CONCLUSIONS SpO2 and PR measurements may be inaccurate across all skin pigmentation groups, breaching FDA guidance and industry standards thresholds. Pulse oximeters significantly overestimate SpO2 for both light and dark skin pigmentation, but this overestimation may not be clinically relevant. PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.
AI snapshot
Written from the paper’s full text.
TL;DR
This systematic review and meta-analysis found that pulse oximetry SpO2 and wearable pulse rate measurements were inaccurate across all skin pigmentation groups, breaching FDA and ANSI/AAMI/IEC thresholds. Pulse oximeters significantly overestimated SpO2 for light and dark skin pigmentation, but this bias may not be clinically relevant; wearable pulse rate showed no statistically or clinically significant bias by skin pigmentation.
Key findings
- For SpO2, pooled accuracy root-mean-square values were 3.96%, 4.71%, and 4.15% for light, medium, and dark pigmentation, respectively, exceeding the FDA threshold of ≤3%; pooled mean biases were 0.70% (95% CI 0.17%-1.22%), 0.27% (95% CI –0.64% to 1.19%), and 1.27% (95% CI 0.58%-1.95%).
- For wearable pulse rate, 95% limits of agreement were –16.02 to 13.54 bpm, –18.62 to 16.84 bpm, and –33.69 to 32.54 bpm for light, medium, and dark pigmentation, respectively, exceeding ±5 bpm; pooled mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm.
- Subgroup analysis found a statistically significant difference in SpO2 bias between light and dark skin pigmentation (estimate=0.596, SE 0.240; 95% CI 0.069-1.123), but no statistically significant difference between light and medium pigmentation; no statistically significant PR bias differences were found between groups.
- In sensitivity analysis excluding Crooks et al, pooled SpO2 Arms dropped to 3.00%, 2.91%, and 3.40% for light, medium, and dark pigmentation, leaving only dark pigmentation inaccurate.
Methods
- PRISMA systematic review and meta-analysis with MEDLINE searches performed between April 2022 and June 2023, plus a final ad hoc search in June 2023; 23 pulse oximetry studies (n=59,684; 197,353 paired SpO2-arterial blood observations) and 4 wearable pulse rate studies (n=176; 140,771 paired PPG-ECG observations) were included.
- Studies were included if they investigated PPG-derived SpO2 or PR, used arterial blood gas (SaO2) or ECG as reference, and reported mean bias and SD, SE, 95% LoA, or 95% CI by race, ethnicity, or skin tone; skin tone, race, and ethnicity were mapped into light, medium, and dark pigmentation groups.
- Pooled mean biases and SDs were estimated using correlated and hierarchical effect models with robust variance estimation (correlation coefficient 0.9; sensitivity analyses at 0.3 and 0.6); accuracy was assessed against Arms≤3% for SpO2 and 95% LoA bounded by ±5 bpm for PR, with statistical significance based on 95% CI excluding 0 and clinical relevance based on FDA/ANSI/AAMI/IEC thresholds.
Limitations
- The paper states that its skin pigmentation categorization reclassified race, ethnicity, and skin tone into a universal schema, but this method classifies most White people in the United States as light rather than medium skin pigmentation; only 4 prior PR studies collected participant skin tone, and only 1 had participants with dark pigmentation.
- The paper states that it used stated FDA and ANSI/AAMI/IEC thresholds that may change in the future, potentially limiting applicability; it also grouped varied patient populations and testing methodologies and did not examine moderators such as sensor type, transmittance versus reflectance, wavelength, sensor placement, SpO2 level, or participant activity level.
- Evident from the text: the PR dark pigmentation analysis rested on a single study (1 study, 6 evaluations, 16 participants, 6,922 paired observations), and the pulse oximetry sample overrepresented light skin pigmentation (40,416/59,684 participants, 67.72%; 131,008/197,353 paired observations, 66.38%).
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
50 of 67 cited sources checked · 13 supported · 8 flagged · accuracy not yet measured
- Findingown data
For light skin pigmentation, pulse rate 95% limits of agreement were -16.02 to 13.54 bpm.
“Our analysis revealed 95% LoA of –16.02 to 13.54 bpm, –18.62 to 16.84 bpm, and –33.69 to 32.54 bpm for light, medium, and dark skin pigmentation groups, respectively.”
Quote in the paperNot a substantial claim4 of 4 sources checked- cites [41] source
The Results section on wearable accuracy across skin tones reports mean directional and absolute errors by Fitzpatrick skin-tone group, but not 95% limits of agreement for light pigmentation. Searches for the stated endpoints and for ‘limits of agreement’ found no passage giving −16.02 to 13.54 bpm. The citing sentence calls these ‘our analysis’ results, and our copy cannot verify those numbers. (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [60] source
The Results section, “Bland-Altman Analysis and 95 Percent Limits of Agreement” (page unavailable), reports limits by device and activity, not by skin pigmentation. Searches for the stated −16.02 to 13.54 bpm range and for light, medium, or dark skin groups found no matching result in our copy. (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [61] source
The Results section (page not available in our copy) reports 95% limits of agreement of −8 to 5 bpm during exercise for Fitzpatrick type II, the lightest skin group studied, rather than −16.02 to 13.54 bpm. Its reported resting limits for type II are −5 to 4 bpm. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [62] source
The Methods and Results report limits of agreement by age group, activity, and device, not by light, medium, and dark skin-pigmentation groups. Searches for “−16.02 to 13.54,” “16.02,” and skin-pigmentation subgroup results found no passage reporting the claimed limits for light skin. (LLM, conf=0.78)
Experimental — accuracy not yet measured
- Findingown data
For dark skin pigmentation, pulse rate 95% limits of agreement were -33.69 to 32.54 bpm.
“Our analysis revealed 95% LoA of –16.02 to 13.54 bpm, –18.62 to 16.84 bpm, and –33.69 to 32.54 bpm for light, medium, and dark skin pigmentation groups, respectively.”
Quote in the paperSubstantial claim4 of 4 sources checked- cites [41] source
The Results section on accuracy across skin tones reports mean directional and absolute errors for six pigmentation groups, but not 95% limits of agreement of −33.69 to 32.54 bpm for a dark-skin group. Searches for those limits and for Bland–Altman limits of agreement found no matching passage in our copy. (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [60] source
The Results section, ‘Bland-Altman Analysis and 95 Percent Limits of Agreement’ (page unavailable), reports limits by wearable device and activity, not by skin-pigmentation group. Searches for the stated limits and for dark-skin pigmentation found no passage reporting −33.69 to 32.54 bpm for that group. (LLM, conf=0.75)
Experimental — accuracy not yet measured - cites [61] source
The Results (section 3; page unavailable) report 95% limits of agreement of −6 to 4 bpm for the darkest skin type studied (type IV) during exercise, and −5 to 4 bpm at rest—not −33.69 to 32.54 bpm. The paper also did not include participants with Fitzpatrick skin types V or VI. (LLM, conf=0.90)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [62] source
The Results (Bland-Altman Analysis and Table 1; page unavailable) report limits of agreement by age group, device, and activity, not by light, medium, or dark skin pigmentation. Searches for the stated limits (−33.69 to 32.54 bpm) and skin-pigmentation subgroups found no matching result in our copy. (LLM, conf=0.78)
Experimental — accuracy not yet measured
- Findingown data
For medium skin pigmentation, pulse rate mean bias was -0.89 bpm (95% CI -3.70 to 1.93).
“Mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for the corresponding groups (Table 4 and Figures 5-7 [41,60-62]).”
Quote in the paperNot a substantial claim4 of 4 sources checked- cites [41] source
The Results section on accuracy across skin tones reports heart-rate errors by six pigmentation categories, but does not report a mean bias of −0.89 bpm or a 95% CI of −3.70 to 1.93 for medium pigmentation. Searches for those numbers and for a medium-pigmentation mean bias found no matching passage in our copy. (LLM, conf=0.73)
Experimental — accuracy not yet measured - cites [60] source
The Methods (Recruitment; page unavailable) describe a single participant with Fitzpatrick skin tone 2, not a medium-pigmentation group. The Results (Bland-Altman Analysis; page unavailable) report device- and activity-specific errors, but not a −0.89 bpm mean bias or the stated 95% confidence interval. Searches for medium skin pigmentation and those figures found no supporting passage. (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [61] source
The Results (page unavailable) report a mean bias of −2 bpm for skin type III during exercise, with 95% limits of agreement of −8 to 4 bpm; at rest, the reported bias is 0 bpm. The cited paper does not report the claimed −0.89 bpm bias or its stated 95% confidence interval for this group. (LLM, conf=0.84)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [62] source
The Results (Bland-Altman Analysis and Table 1; page unavailable) report heart-rate biases by age group, device, and activity, not a bias for a medium-pigmentation group. Searches for −0.89 bpm and the stated confidence interval found no matching result in our copy. (LLM, conf=0.75)
Experimental — accuracy not yet measured
- Findingown data
For medium skin pigmentation, pulse rate 95% limits of agreement were -18.62 to 16.84 bpm.
“Our analysis revealed 95% LoA of –16.02 to 13.54 bpm, –18.62 to 16.84 bpm, and –33.69 to 32.54 bpm for light, medium, and dark skin pigmentation groups, respectively.”
Quote in the paperNot a substantial claim4 of 4 sources checked- cites [41] source
The Results section, “Wearables accuracy across skin tones” (page unavailable), reports mean directional and absolute heart-rate errors by Fitzpatrick skin-tone group, not 95% limits of agreement for a medium-pigmentation group. Searches for the stated endpoints and for limits of agreement found no passage verifying −18.62 to 16.84 bpm. (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [60] source
The Methods > Recruitment section (page unavailable) says the study involved a single participant with Fitzpatrick skin tone 2, not light, medium, and dark pigmentation groups. The Results > Bland-Altman Analysis section (page unavailable) reports limits of agreement by device and activity, not the claimed −18.62 to 16.84 bpm for a medium-pigmentation group. (LLM, conf=0.87)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [61] source
The Results section reports 95% limits of agreement of −8 to 4 beats/min for skin type III during exercise, not −18.62 to 16.84 bpm. It reports −5 to 5 beats/min for type III at rest. Our copy does not provide a page number for this passage. (LLM, conf=0.89)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [62] source
The Results section (Bland-Altman Analysis; page unavailable) reports limits of agreement by age group, device, and activity, not by light, medium, and dark skin pigmentation. A search for the stated limits of −18.62 to 16.84 bpm and for skin-pigmentation subgroup analyses found no passage verifying that figure. The citing sentence also describes the figures as the citing authors’ own analysis. (LLM, conf=0.72)
Experimental — accuracy not yet measured
- Findingown data
For dark skin pigmentation, pulse rate mean bias was -0.57 bpm (95% CI -9.44 to 8.29).
“Mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for the corresponding groups (Table 4 and Figures 5-7 [41,60-62]).”
Quote in the paperNot a substantial claim4 of 4 sources checked- cites [41] source
The Results section (‘Wearables accuracy across skin tones’; page unavailable) reports heart-rate measurement error by skin tone, but it does not report a mean bias of −0.57 bpm or the stated 95% confidence interval for dark skin. Its reported values include a mean directional error of −4.25 bpm for FP5 at rest and a mean absolute error of 10.6 bpm for FP6 at rest; these do not establish the claimed estimate. (LLM, conf=0.82)
Experimental — accuracy not yet measured - cites [60] source
The Methods (Recruitment; page unavailable) describe a single participant with Fitzpatrick skin tone 2, not a dark-skin group. The Results (Bland-Altman Analysis; page unavailable) report mean errors of −1.80 bpm for Apple Watch 3 and −3.47 bpm for Fitbit Charge 2 across 24 hours, not the claimed −0.57 bpm with a 95% CI of −9.44 to 8.29. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [61] source
The Results section (page unavailable) reports a mean bias of −1 bpm for the darkest skin type studied (type IV), with 95% limits of agreement of −6 to 4 bpm. It does not report the claimed −0.57 bpm or a 95% confidence interval of −9.44 to 8.29 bpm; the reported limits of agreement are a different measure from a confidence interval. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [62] source
The Results (Reliability of Examined Devices, Table 1) report heart-rate differences by age group, activity, and device, not a mean bias for dark skin pigmentation. The Introduction (Aim of the Study) describes participants with Fitzpatrick skin types III or IV. Searches for the quoted bias and interval found no matching result in our copy. (LLM, conf=0.75)
Experimental — accuracy not yet measured
- Findingown data
For light skin pigmentation, pulse rate mean bias was -1.24 bpm (95% CI -5.31 to 2.83).
“Mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for the corresponding groups (Table 4 and Figures 5-7 [41,60-62]).”
Quote in the paperNot a substantial claim4 of 4 sources checked- cites [41] source
The Results section (“Wearables accuracy across skin tones”; page unavailable) reports heart-rate measurement error by skin tone, including an MDE of −0.53 bpm for the lightest group at rest. It does not report the claimed mean bias of −1.24 bpm or its 95% CI of −5.31 to 2.83; searches for those numbers and for a light-skin bias with a confidence interval found no matching passage. (LLM, conf=0.72)
Experimental — accuracy not yet measured - cites [60] source
The Results section reports mean errors by device and activity, not by skin pigmentation. A search of our copy for light skin pigmentation and the stated bias and confidence interval found no passage reporting those subgroup results; the paper’s overall device estimates do not resolve the claim. (LLM, conf=0.75)
Experimental — accuracy not yet measured - cites [61] source
The Results section (page not available) reports a mean bias of −2 bpm for the lightest skin group studied (type II), with 95% limits of agreement of −8 to 5 bpm. It does not report a bias of −1.24 bpm or a 95% confidence interval of −5.31 to 2.83 bpm; the reported interval is also a different measure. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [62] source
The Results (Reliability of Examined Devices, Table 1; page unavailable) report heart-rate mean differences by age group, activity, and device, not a light-pigmentation group. Searches for −1.24, the stated interval, and skin-pigmentation subgroup results found no passage reporting this estimate. The Introduction (Aim of the Study; page unavailable) describes a study of participants with Fitzpatrick skin types III or IV. (LLM, conf=0.77)
Experimental — accuracy not yet measured
- Premisecitation
There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR or pulse oximeter SpO2.
“There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR [35] or pulse oximeter SpO2 [36-38].”
Quote in the paperSubstantial claim4 of 4 sources checked- cites [35] source
- cites [36] source
- cites [37] source
- cites [38] source
The ‘Skin pigmentation as a limitation of pulse oximetry’ section describes a review of studies comparing pulse-oximeter SpO2 with reference SaO2 across skin pigmentation groups, supporting the subject attributed to this source. But the ‘Statistical significance’ section explicitly calls it a narrative review, not a systematic review; the source therefore does not support that classification. (LLM, conf=0.87)
Experimental — accuracy not yet measured
- Premisecitation
Patients with darker skin pigmentation are more likely than patients with lighter skin color to have overestimated SpO2 readings.
“Patients with darker skin pigmentation are more likely than patients with lighter skin color to have overestimated SpO2 readings”
Quote in the paperSubstantial claim7 of 8 sources checked- cites [11] source
In “Racial Bias in Pulse Oximetry Measurement,” the authors report that Black patients were more likely than White patients to have occult hypoxemia despite SpO2 readings of 92–96%. The study compared self-identified racial groups, not measured skin pigmentation, so it does not directly establish the claim’s darker-versus-lighter-skin comparison. (LLM, conf=0.86)
Experimental — accuracy not yet measured - cites [12] source
The Results (Probability of occult hypoxemia) show that Black inpatients were more likely than white inpatients to have SpO2 readings that masked low arterial oxygen saturation. But the Discussion (Strengths and limitations) says skin color was not recorded and race was used as a surrogate, so the study does not directly establish the comparison by skin pigmentation stated in the claim. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites [13] source
- cites [14] source
- cites [15] source
- cites [16] source
The Results (Association of Race and Ethnicity With Pulse Oximeter Error) find greater SpO2 overestimation among Black, Hispanic, and other racial or ethnic minority patients than among White patients. But the Limitations state that the study used self-reported race and ethnicity as a surrogate for skin tone; it did not directly establish the comparison between darker and lighter skin pigmentation stated in the claim. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites [17] source
- cites [18] source
- Premisecitation
More recent research has highlighted the potential presence of skin pigmentation biases in research-grade and consumer PPG-based wearable devices.
“More recent research has also highlighted the potential presence of these biases in research-grade and consumer PPG-based wearable devices [4,27-29]”
Quote in the paperSubstantial claim1 of 4 sources checked- cites [4] source
The Biobehavioral and demographic data section (page unavailable) reports that darker skin tone may reduce the accuracy of wrist-worn heart-rate wearables, though not all studies find this. The paper focuses on consumer PPG wearables; the passages do not establish the claim for research-grade devices. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [27] source
- cites [28] source
- cites [29] source
- Study design and conductcitation
The skin pigmentation categorization method used in this meta-analysis classifies most White people in the United States as light rather than medium skin pigmentation.
“This method, however, classifies most White people in the United States as light rather than medium skin pigmentation [66].”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [66] source
The Results (p. 163) say that White participants mainly had digital image skin type scores of 2 or 3, on a scale where 1 is lightest and 6 darkest. The paper does not define the meta-analysis’s light-versus-medium categories or establish that most White people fall into its light category rather than its medium category. (LLM, conf=0.82)
Experimental — accuracy not yet measured
- Findingown data
For SpO2, pooled mean percent bias was 0.82% (95% CI 0.29%-1.35%) across all skin pigmentation groups.
“We observed a pooled mean percent bias of 0.82% (95% CI 0.29%-1.35%) across all skin pigmentation groups using the CHE model.”
Quote in the paperSubstantial claim - Findingown data
For SpO2, pooled mean percent bias was 0.70% (95% CI 0.17%-1.22%) for light skin pigmentation.
“the pooled mean percent bias from the CHE model was 0.70% (95% CI 0.17%-1.22%) for light skin”
- Also stated in: § Results · plain
“pooled mean biases were 0.70% (95% CI 0.17%-1.22%)”
Quote in the paperSubstantial claim - Also stated in: § Results · plain
- Findingown data
For SpO2, pooled mean percent bias was 0.27% (95% CI -0.64% to 1.19%) for medium skin pigmentation.
“0.27% (95% CI –0.64% to 1.19%) for medium skin”
- Also stated in: § Results · plain
“0.27% (95% CI –0.64% to 1.19%)”
Quote in the paperNot a substantial claim - Also stated in: § Results · plain
- Findingown data
For SpO2, pooled mean percent bias was 1.27% (95% CI 0.58%-1.95%) for dark skin pigmentation.
“1.27% (95% CI 0.58%-1.95%) for dark skin”
- Also stated in: § Results · plain
“1.27% (95% CI 0.58%-1.95%)”
Quote in the paperSubstantial claim - Also stated in: § Results · plain
- Findingown data
Subgroup analysis found no statistically significant difference in SpO2 bias between light and medium skin pigmentation (estimate=0.113, SE 0.259; 95% CI -0.459 to 0.686).
“Subgroup analyses found no statistically significant difference between light and medium skin pigmentation (estimate=0.113, SE 0.259; 95% CI –0.459 to 0.686)”
Quote in the paperSubstantial claim - Findingown data
There were only 4 prior pulse rate studies that collected participant skin tone and also reported on device accuracy.
“Second, there were only 4 prior PR studies that collected participant skin tone that also reported on device accuracy and only 1 that had participants with dark pigmentation.”
Quote in the paperSubstantial claim - Findingown data
For combined skin pigmentation, pulse oximetry Arms was 3.60%.
“| | Combined | 23 (95) | 59,684 (197,353) | Percent | 0.82 (0.29 to 1.35) | 3.50 (1.28) | –6.04 to 7.68 | 3.60e | 98.55% (14.53% and 84.02%) |”
Quote in the paperNot a substantial claim - Findingown data
Subgroup analyses found no statistically significant difference between light and dark skin pigmentation pooled pulse rate bias (P≥.05).
“Subgroup analyses found no statistically significant difference between light and medium or between light and dark skin pigmentation pooled bias (P≥.05; Table S2 in Multimedia Appendix 1).”
Quote in the paperSubstantial claim - Findingown data
For medium skin pigmentation, pulse oximetry pooled mean bias was 0.27% (95% CI -0.64 to 1.19).
“| | Medium | 9 (15) | 9,967 (32,095) | Percent | 0.27 (–0.64 to 1.19) | 4.71 (1.71) | –8.95 to 9.50 | 4.71e | 95.31% (81.16% and 14.14%) |”
Quote in the paperNot a substantial claim - Findingown data
After removing Crooks et al as an outlier, the pooled Arms for pulse oximetry SpO2 in the dark skin pigmentation group was 3.40%.
“The pooled mean percent biases and 95% LoA were essentially unchanged, while the pooled Arms across all skin pigmentation groups was 3.00%, 2.91%, and 3.40% for light, medium, and dark skin pigmentation, respectively.”
Quote in the paperNot a substantial claim - Findingnone given
There was a lack of consumer-device reporting despite growing use in clinical settings.
“The study revealed a paucity of studies properly assessing skin pigmentation, a tendency for uneven distribution across different skin pigmentation groups (overrepresentation of light skin for SpO2 and a single study with representation of dark skin for PR), and lack of consumer-device reporting despite growing use in clinical setting.”
Quote in the paperNot a substantial claim - Findingown data
No systematic or clinically relevant bias was found in estimation of pulse rate.
“No systematic or clinically relevant bias was found in estimation of PR.”
Quote in the paperSubstantial claim - Findingown data
There was a tendency for uneven distribution across different skin pigmentation groups, with overrepresentation of light skin for SpO2.
“The study revealed a paucity of studies properly assessing skin pigmentation, a tendency for uneven distribution across different skin pigmentation groups (overrepresentation of light skin for SpO2 and a single study with representation of dark skin for PR), and lack of consumer-device reporting despite growing use in clinical setting.”
Quote in the paperSubstantial claim - Findingown data
Sensitivity analyses using CHE models with a range of correlation coefficients yielded similar conclusions to those with ρ=0.9, except for the pulse rate analysis in the light skin pigmentation group.
“Sensitivity analyses using CHE models with a range of correlation coefficients (including small and moderate values) yielded similar conclusions to those obtained with a high correlation coefficient (ρ=0.9), except for the PR analysis in the light skin pigmentation group.”
Quote in the paperNot a substantial claim - Findingown data
There was only 1 prior pulse rate study that had participants with dark pigmentation.
“Second, there were only 4 prior PR studies that collected participant skin tone that also reported on device accuracy and only 1 that had participants with dark pigmentation.”
Quote in the paperSubstantial claim - Findingown data
After removing Crooks et al as an outlier, the pooled Arms for pulse oximetry SpO2 in the medium skin pigmentation group was 2.91%.
“The pooled mean percent biases and 95% LoA were essentially unchanged, while the pooled Arms across all skin pigmentation groups was 3.00%, 2.91%, and 3.40% for light, medium, and dark skin pigmentation, respectively.”
Quote in the paperNot a substantial claim - Findingown data
Pulse oximeter SpO2 was only found to be inaccurate for dark skin pigmentation in sensitivity analyses.
“This systematic review and meta-analysis found that pulse oximeter SpO2 and wearable PR were inaccurate across all skin pigmentation groups as the resulting accuracy values breached FDA guidance and ANSI/AAMI/IEC standard thresholds, respectively, although pulse oximeter SpO2 was only found to be inaccurate for dark skin pigmentation in sensitivity analyses.”
Quote in the paperSubstantial claim - Findingown data
Subgroup analyses found no statistically significant difference between light and medium skin pigmentation pooled pulse rate bias (P≥.05).
“Subgroup analyses found no statistically significant difference between light and medium or between light and dark skin pigmentation pooled bias (P≥.05; Table S2 in Multimedia Appendix 1).”
Quote in the paperSubstantial claim - Findingown data
There was a single study with representation of dark skin for pulse rate.
“The study revealed a paucity of studies properly assessing skin pigmentation, a tendency for uneven distribution across different skin pigmentation groups (overrepresentation of light skin for SpO2 and a single study with representation of dark skin for PR), and lack of consumer-device reporting despite growing use in clinical setting.”
Quote in the paperNot a substantial claim - Findingown data
In the pulse rate analysis for light skin pigmentation, a statistically significant bias was revealed when a low correlation coefficient (ρ=0.3) was used.
“This analysis revealed a statistically significant bias when a low correlation coefficient (ρ=0.3) was used.”
Quote in the paperNot a substantial claim - Findingown data
After removing Crooks et al as an outlier, the pooled Arms for pulse oximetry SpO2 in the light skin pigmentation group was 3.00%.
“The pooled mean percent biases and 95% LoA were essentially unchanged, while the pooled Arms across all skin pigmentation groups was 3.00%, 2.91%, and 3.40% for light, medium, and dark skin pigmentation, respectively.”
Quote in the paperNot a substantial claim - Findingown data
Despite not exceeding clinically relevant bias thresholds, pulse oximeters were found to significantly overestimate SpO2 for light and dark skin pigmentation.
“In addition, despite not exceeding clinically relevant bias thresholds, pulse oximeters were found to significantly overestimate SpO2 for light and dark skin pigmentation.”
Quote in the paperSubstantial claim - Findingown data
The study revealed a paucity of studies properly assessing skin pigmentation.
“The study revealed a paucity of studies properly assessing skin pigmentation, a tendency for uneven distribution across different skin pigmentation groups (overrepresentation of light skin for SpO2 and a single study with representation of dark skin for PR), and lack of consumer-device reporting despite growing use in clinical setting.”
Quote in the paperSubstantial claim - Findingown data
After removing Crooks et al as an outlier, the pooled mean percent biases and 95% limits of agreement for pulse oximetry SpO2 were essentially unchanged.
“The pooled mean percent biases and 95% LoA were essentially unchanged, while the pooled Arms across all skin pigmentation groups was 3.00%, 2.91%, and 3.40% for light, medium, and dark skin pigmentation, respectively.”
Quote in the paperNot a substantial claim - Findingown data
For combined skin pigmentation, pulse oximetry pooled mean bias was 0.82% (95% CI 0.29 to 1.35).
“| | Combined | 23 (95) | 59,684 (197,353) | Percent | 0.82 (0.29 to 1.35) | 3.50 (1.28) | –6.04 to 7.68 | 3.60e | 98.55% (14.53% and 84.02%) |”
Quote in the paperSubstantial claim - Interpretationown data
The overestimation of SpO2 by pulse oximeters may not be clinically relevant.
“but this overestimation may not be clinically relevant.”
Quote in the paperSubstantial claim - Interpretationown data
Pulse oximeters significantly overestimate SpO2 for both light and dark skin pigmentation.
“Pulse oximeters significantly overestimate SpO2 for both light and dark skin pigmentation”
Quote in the paperSubstantial claim - Interpretationown data
PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.
“PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.”
Quote in the paperSubstantial claim - Interpretationreasoning
When pulse oximetry devices are deployed in their setting of intended use (uncontrolled settings such as real-world medical settings and home environments with diverse patient populations), the performance observed in analytical validation studies may not generalize.
“Overall, these findings suggest that when pulse oximetry devices are deployed in their setting of intended use (ie, uncontrolled settings, such real-world medical settings and home environments with diverse patient populations), the performance observed in analytical validation studies may not generalize.”
Quote in the paperSubstantial claim - Interpretationown data
The studies included in the SpO2 and pulse rate analyses were largely consistent with one another.
“Lastly, it should be noted that both SpO2 and PR studies had lower between-study heterogeneity and higher within-study heterogeneity, indicating that the studies that were included in these analyses were largely consistent with one another and that most of the variation came from within studies potentially due to higher variability across devices used or participants enrolled within each study.”
Quote in the paperNot a substantial claim - Interpretationcitation
The 2013 FDA guidance may be insufficient to ensure accuracy in pulse oximeters across all skin pigmentation and settings of intended use.
“The 2013 FDA guidance may be insufficient to ensure accuracy in pulse oximeters across all skin pigmentation and settings of intended use [19].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [19] Pulse oximeters - premarket notification submissions [510(k)s] guidance for industry and Food and Drug Administration staffU.S. Food and Drug Administration201303042024-08-24https://www.fda.gov/media/72470/download
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationnone given
The overestimation of SpO2 by pulse oximeters for light and dark skin pigmentation is without clinically relevant bias.
“Pulse oximeters may also overestimate SpO2 significantly for light and dark skin pigmentation, but without clinically relevant bias.”
Quote in the paperSubstantial claim - Interpretationreasoning
Wearable accuracy varies considerably depending on the model.
“Results suggest inaccurate SpO2 and PR measurements across all skin pigmentation groups as they breach FDA and ANSI/AAMI/IEC standards, respectively, with wearable accuracy varying considerably depending on the model, which may be due to date of model production or algorithm development.”
Quote in the paperNot a substantial claim
- May combine two claims. 34
- Classified as probably not a substantial claim. 22
- The quote only partly states this claim. 7
May combine two claims.
PremisecitationOnly 1 meta-analysis focused on pulse oximetry for this topic is available, and an additional 7 studies have been published recently, providing an additional 50,980 participants and 182,369 paired observations.
“Only 1 meta-analysis focused on pulse oximetry for this topic is available [36], and an additional 7 studies have been published recently, providing an additional 50,980 participants and 182,369 paired observations.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [36] source
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationFunnel plots evaluated publication bias.
“Funnel plots evaluated publication bias using the metafor package [63] (Figures S2 and S3 in Multimedia Appendix 1).”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [63] Viechtbauer, W, 2010, J Stat Soft source
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationThe study followed PRISMA guidelines.
“We followed the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) guidelines [39]”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [39] source
May combine two claims.
PremisecitationOverestimated SpO2 readings in patients with darker skin pigmentation lead to lower hospital admission, higher occult hypoxemia, delayed or no access to dexamethasone, therapeutic oxygen, and COVID-19 therapies, resulting in increased hospital readmission, organ dysfunction, and mortality.
“leading to lower hospital admission; higher occult hypoxemia; and delayed or no access to dexamethasone, therapeutic oxygen, and COVID-19 therapies, resulting in increased hospital readmission, organ dysfunction, and mortality [11-18].”
Quote in the paperSubstantial claim7 of 8 sources checked- cites [11] source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [12] source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [13] source
- cites [14] source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [15] source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [16] source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [17] source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [18] source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured
Classified as probably not a substantial claim. May combine two claims.
PremisecitationNo established methodologies exist for providing pooled 95% LoA and Arms, so the study followed an analytical approach used in prior meta-analyses.
“When it comes to providing pooled 95% LoA and Arms, no established methodologies exist, so we followed an analytical approach used in prior meta-analyses within the discipline [36].”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [36] source
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataSensitivity analyses for pulse oximetry SpO2 were performed by removing Crooks et al [48] as an outlier due to substantially large standard error.
“Sensitivity analyses were performed for pulse oximetry SpO2 by removing Crooks et al [48] as an outlier due to substantially large SE (Tables S6 and S7 in Multimedia Appendix 1).”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [48] source
May combine two claims.
Findingown dataIn sensitivity analyses without the outlier study of Crooks et al, all pooled Arms values dropped, resulting in inaccurate pulse oximeter SpO2 only for dark skin pigmentation and no group exceeding international thresholds.
“Additionally, in the sensitivity analyses without the outlier study of Crooks et al [48], all pooled Arms values dropped, resulting in inaccurate pulse oximeter SpO2 only for dark skin pigmentation and no group exceeding international thresholds.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [48] source
May combine two claims.
Findingown dataFor PR, 95% limits of agreement were -16.02 to 13.54 for light, -18.62 to 16.84 for medium, and -33.69 to 32.54 for dark skin pigmentation.
“For PR, 95% limits of agreement values were from –16.02 to 13.54, from –18.62 to 16.84, and from –33.69 to 32.54”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor SpO2, pooled accuracy root-mean-square values were 3.6% combined, 3.96% light, 4.71% medium, and 4.15% dark skin pigmentation.
“The pooled Arms across different skin pigmentation groups was 3.6%, 3.96%, 4.71%, and 4.15% for combined, light, medium, and dark skin pigmentation, respectively”
- Also stated in: § Results · plain
“For SpO2, accuracy root-mean-square values were 3.96%, 4.71%, and 4.15%”
Quote in the paperSubstantial claim- Also stated in: § Results · plain
May combine two claims.
Findingown dataThe systematic review included 23 pulse oximetry studies with 59,684 participants and 197,353 paired SpO2-SaO2 observations.
“We identified 23 pulse oximetry studies (n=59,684; 197,353 paired SpO2-arterial blood observations)”
- Also stated in: § Study Selection and Characteristics · plain
“A total of 23 pulse oximetry studies involving 59,684 participants with 197,353 paired SpO2-SaO2 observations were included.”
Quote in the paperSubstantial claim- Also stated in: § Study Selection and Characteristics · plain
May combine two claims.
Findingown dataThe systematic review included 4 wearable pulse rate studies with 176 participants and 140,771 paired PPG-ECG observations.
“4 wearable PR studies (n=176; 140,771 paired PPG-electrocardiography observations)”
- Also stated in: § Study Selection and Characteristics · plain
“Additionally, 4 PR studies with 176 participants and 140,771 paired PR-ECG observations were analyzed (Table 3).”
Quote in the paperSubstantial claim- Also stated in: § Study Selection and Characteristics · plain
Classified as probably not a substantial claim. May combine two claims.
Findingown data27 studies were selected for full review (23 pulse oximetry, 4 PR).
“We selected 27 studies for full review (pulse oximetry: n=23, PR: n=4; Figure 1, Table 1, and Table 2).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn pulse oximetry studies, there were 40,416 (67.72%) patients with light skin pigmentation and 131,008 (66.38%) paired observations; 9967 (16.70%) patients with medium skin pigmentation and 32,095 (16.26%) paired observations; and 9301 (15.58%) patients with dark skin pigmentation and 34,250 (17.35%) paired observations.
“Of the 59,684 patients with 197,353 paired observations, there were a total of 40,416 (67.72%) patients with light skin pigmentation with 131,008 (66.38%) paired observations, 9967 (16.70%) patients with medium skin pigmentation with 32,095 (16.26%) paired observations, and 9301 (15.58%) patients with dark skin pigmentation with 34,250 (17.35%) paired observations.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn PR studies, there were 31 (17.61%) participants with light skin pigmentation and 43,116 (30.60%) paired observations; 129 (73.30%) participants with medium skin pigmentation and 90,733 (64.50%) paired observations; and 16 (9.09%) participants with dark skin pigmentation and 6922 (4.92%) paired observations.
“Of the 176 patients with 140,771 paired observations, there were a total of 31 (17.61%) participants with light skin pigmentation with 43,116 (30.60%) paired observations, 129 (73.30%) participants with medium skin pigmentation with 90,733 (64.50%) paired observations, and 16 (9.09%) participants with dark skin pigmentation with 6922 (4.92%) paired observations.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataSkin pigmentation was classified by race in 13 (57%) of 23 pulse oximetry studies, ethnicity in 3 (13%), skin tone in 4 (17%), and both race and ethnicity in 3 (13%).
“Skin pigmentation was classified by race in 13 (57%) out of 23 studies, ethnicity in 3 (13%) studies, skin tone in 4 (17%) studies, and both race and ethnicity in 3 (13%) studies (Table 3).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown data22 (96%) of 23 pulse oximeter studies used a transmittance device and only one used an ear clip.
“We used this threshold as 22 (96%) out of 23 of pulse oximeter studies used a transmittance device and only one used an ear clip”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn the pulse rate meta-analysis, 26.76% of the variation in bias stemmed from between-study differences and 73.34% originated from within-study variation.
“Heterogeneity analysis demonstrated that 26.76% of the variation in bias (I2Level3) stemmed from between-study differences, while 73.34% (I2Level2) originated from within-study variation.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe search strategy resulted in 8582 records.
“Search strategy resulted in 8582 records.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor PR, pooled mean biases were -1.24 (95% CI -5.31 to 2.83) bpm for light, -0.89 (95% CI -3.70 to 1.93) bpm for medium, and -0.57 (95% CI -9.44 to 8.29) bpm for dark skin pigmentation.
“pooled mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for light, medium, and dark pigmentation, respectively.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataIn PR studies, skin pigmentation was classified by skin tone in 4 (100%) studies and by race or ethnicity in 0 (0%) studies.
“Skin pigmentation was classified by race or ethnicity in 0 (0%) out of 4 studies and skin tone in 4 (100%) studies (Table 3).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataAcross all pulse rate studies, the pooled mean bias was -0.29 bpm (95% CI -3.87 to 3.29) and the 95% limits of agreement were -17.23 to 16.65 bpm.
“Analysis using the CHE model revealed LoA values of –17.23 to 16.65 bpm and a mean bias of –0.29 (95% CI –3.87 to 3.29) bpm across all studies.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataBetween-study heterogeneity accounted for 14.53% of total variation, and within-study heterogeneity explained 84.02% of total variation.
“Between-study heterogeneity (I2Level3) accounted for 14.53% of the total variation, while within-study heterogeneity (I2Level2) explained 84.02% of total variation.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataSubgroup analysis found a statistically significant difference in SpO2 bias between light and dark skin pigmentation (estimate=0.596, SE 0.240; 95% CI 0.069-1.123), with higher pooled bias for darker skin.
“but there was a statistically significant difference between light and dark skin pigmentation (estimate=0.596, SE 0.240; 95% CI 0.069-1.123), such that pooled bias for those with darker skin was higher as compared to those with lighter skin”
Quote in the paperSubstantial claimMay combine two claims. The quote only partly states this claim.
Findingown dataFor dark skin pigmentation, pulse oximetry Arms was 4.15%, exceeding FDA guidance.
“| | Dark | 19 (36) | 9,301 (34,250) | Percent | 1.27 (0.58 to 1.95) | 3.96 (1.30) | –6.49 to 9.02 | 4.15e | 98.46% (0.00% and 98.46%) |”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims. The quote only partly states this claim.
Findingown dataFor light skin pigmentation, pulse oximetry Arms was 3.96%, exceeding FDA guidance.
“| | Light | 20 (44) | 40,416 (131,008) | Percent | 0.70 (0.17 to 1.22) d | 3.90 (1.36) | –6.94 to 8.34 | 3.96e | 97.45% (46.44% and 51.01%) |”
Quote in the paperNot a substantial claimMay combine two claims. The quote only partly states this claim.
Findingown dataFor dark skin pigmentation, pulse oximetry pooled mean bias was 1.27% (95% CI 0.58 to 1.95), and this was statistically significant.
“| | Dark | 19 (36) | 9,301 (34,250) | Percent | 1.27 (0.58 to 1.95) | 3.96 (1.30) | –6.49 to 9.02 | 4.15e | 98.46% (0.00% and 98.46%) |”
Quote in the paperSubstantial claimMay combine two claims.
FindingreasoningBoth SpO2 and pulse rate studies had lower between-study heterogeneity and higher within-study heterogeneity.
“Lastly, it should be noted that both SpO2 and PR studies had lower between-study heterogeneity and higher within-study heterogeneity, indicating that the studies that were included in these analyses were largely consistent with one another and that most of the variation came from within studies potentially due to higher variability across devices used or participants enrolled within each study.”
Quote in the paperNot a substantial claimMay combine two claims. The quote only partly states this claim.
Findingown dataFor light skin pigmentation, pulse oximetry pooled mean bias was 0.70% (95% CI 0.17 to 1.22), and this was statistically significant.
“| | Light | 20 (44) | 40,416 (131,008) | Percent | 0.70 (0.17 to 1.22) d | 3.90 (1.36) | –6.94 to 8.34 | 3.96e | 97.45% (46.44% and 51.01%) |”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataNo statistically significant or clinically relevant bias was found in wearable pulse rate devices.
“We did not find statistically significant or clinically relevant bias in wearable PR devices.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims. The quote only partly states this claim.
Findingown dataFor medium skin pigmentation, pulse oximetry Arms was 4.71%, exceeding FDA guidance.
“| | Medium | 9 (15) | 9,967 (32,095) | Percent | 0.27 (–0.64 to 1.19) | 4.71 (1.71) | –8.95 to 9.50 | 4.71e | 95.31% (81.16% and 14.14%) |”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningDevices with accuracy measures breaching these accuracy thresholds can produce questionable results.
“Devices with accuracy measures breaching these accuracy thresholds can produce questionable results.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationANSI/AAMI/IEC recommend that ECG devices have mean bias of +5 or -5 bpm or mean absolute percentage error≤10%, whichever is greater.
“The American National Standards Institute (ANSI), Association for the Advancement of Medical Instrumentation (AAMI), and International Electrotechnical Commission (IEC) have set the recommendation that electrocardiography (ECG) devices have mean bias of +5 or –5 beats per minute (bpm) or mean absolute percentage error≤10%, whichever is greater [33,34].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [33] ANSI/AAMI/IECIEC 60601-2-27 particular requirements for the basic safety and essential performance of electro cardiographic monitoringITC India2024-08-24https://www.itcindia.org/iec-60601-2-27-particular-requirements-for-the-basic-safety-and-essential-performance-of-electro-cardiographic-monitoring/#:~:text=What%20is%20IEC%2060601%2D2,processing%2C%20alarms%2C%20and%20displays
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [34] ANSI/AAMI EC13-2002 cardiac monitors, heart rate meters, and alarmsANSI20022023-10-30https://webstore.ansi.org/standards/aami/ansiaamiec132002
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
PremisecitationFDA guidances since 2013 have called to address darker skin pigmentation bias in pulse oximeters but lacked standards for assessing skin tone.
“Starting in 2013, a series of U.S. Food and Drug Administration (FDA) guidances, safety communications, and more recently, statements from attorneys general have called to address darker skin pigmentation bias in pulse oximeters [19]; however, these guidances lacked standards for assessing skin tone [20].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [19] Pulse oximeters - premarket notification submissions [510(k)s] guidance for industry and Food and Drug Administration staffU.S. Food and Drug Administration201303042024-08-24https://www.fda.gov/media/72470/download
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [20] Review of pulse oximeters and factors that can impact their accuracyU.S. Food and Drug Administration20222024-08-24https://www.fda.gov/media/162709/download
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe first author verified the quality of manuscripts.
“The first author verified the quality of manuscripts.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationFDA guidance has a threshold of Arms≤3% for transmittance devices and Arms≤3.5% for ear clip and reflectance devices, while international thresholds are set at Arms≤4%.
“FDA guidance has a threshold of Arms≤3% [19] for transmittance devices and a threshold of Arms≤3.5% [19] for ear clip and reflectance devices, while international thresholds are set at Arms≤4% [32].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [19] Pulse oximeters - premarket notification submissions [510(k)s] guidance for industry and Food and Drug Administration staffU.S. Food and Drug Administration201303042024-08-24https://www.fda.gov/media/72470/download
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [32] Medical electrical equipment particular requirements for basic safety and essential performance of pulse oximeter equipment (British standard)ANSI2023-10-23https://tinyurl.com/ayfrdpau
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataData were independently extracted by two reviewers, with a third and fourth reviewer adjudicating differences.
“Data were independently extracted from published manuscripts (BWN and SS). A third and fourth reviewer (MRB and HG) adjudicated any differences between the initial 2 reviewers and resolved any disagreements by checking manuscripts.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataSpO2 and PR measurements may be inaccurate across all skin pigmentation groups, breaching FDA guidance and industry standard thresholds.
“SpO2 and PR measurements may be inaccurate across all skin pigmentation groups, breaching U.S. Food and Drug Administration guidance and industry standard thresholds.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisenone givenFor heart rate or PR devices, overall accuracy can be assessed with 95% limits of agreement.
“For heart rate or PR devices, overall accuracy can be assessed with 95% limits of agreement (LoAs).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenQUADAS-2 tool was used to evaluate risk of bias and applicability.
“QUADAS-2 tool was used to evaluate risk of bias and applicability (Figure S1 in Multimedia Appendix 1).”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
PremisecitationOverall accuracy of pulse oximeters can be assessed with accuracy root-mean-square (Arms), which combines mean bias and precision into a single metric.
“Overall accuracy of pulse oximeters can be assessed with accuracy root-mean-square (Arms), which combines mean bias and precision (SD of bias) into a single metric [32].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [32] Medical electrical equipment particular requirements for basic safety and essential performance of pulse oximeter equipment (British standard)ANSI2023-10-23https://tinyurl.com/ayfrdpau
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim.
Study design and conductreasoningThe level of SpO2 was not examined given the study's goal of gathering the largest sample upon which to study the effect of skin tone.
“The level of SpO2 was not examined given this study’s goal of gathering the largest sample upon which to study the effect of skin tone.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataDespite not meeting FDA guidance across all groups, pulse oximeter SpO2 was inaccurate only across medium and dark skin pigmentation groups when compared to the more liberal international thresholds.
“Despite not meeting FDA guidance across all groups, pulse oximeter SpO2 was inaccurate only across medium and dark skin pigmentation groups when compared to the more liberal international thresholds.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningMost of the variation in SpO2 and pulse rate studies came from within studies, potentially due to higher variability across devices used or participants enrolled within each study.
“Lastly, it should be noted that both SpO2 and PR studies had lower between-study heterogeneity and higher within-study heterogeneity, indicating that the studies that were included in these analyses were largely consistent with one another and that most of the variation came from within studies potentially due to higher variability across devices used or participants enrolled within each study.”
Quote in the paperNot a substantial claimThe quote only partly states this claim.
Interpretationown dataPulse oximeters may overestimate SpO2 significantly for light skin pigmentation.
“Pulse oximeters may also overestimate SpO2 significantly for light and dark skin pigmentation, but without clinically relevant bias.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductreasoningThe potential moderating effects of sensor type, transmittance versus reflectance sensors, wavelength of light used by sensors, and location of sensor placement were not examined because the core sensor technology studied was PPG across all included studies.
“We did not examine the potential moderating effects of sensor type, transmittance versus reflectance sensors, wavelength of light used by sensors, and location of sensor placement given the core sensor technology studied was PPG across all included studies.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningParticipant activity level was not evaluated as a moderator variable because many included studies did not delineate participant activity level explicitly.
“Participant activity level was not evaluated as a moderator variable because many included studies did not delineate participant activity level explicitly.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductown dataSubgroup analyses for pulse rate were not conducted because most studies did not specify whether subjects were healthy or part of medical populations.
“We did not conduct these analyses for pulse rate because most studies did not specify whether subjects were healthy or part of medical populations.”
Quote in the paperSubstantial claimThe quote only partly states this claim.
Interpretationown dataPulse oximeters may overestimate SpO2 significantly for dark skin pigmentation.
“Pulse oximeters may also overestimate SpO2 significantly for light and dark skin pigmentation, but without clinically relevant bias.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThis systematic review and meta-analysis found that pulse oximeter SpO2 and wearable pulse rate were inaccurate across all skin pigmentation groups as the resulting accuracy values breached FDA guidance and ANSI/AAMI/IEC standard thresholds, respectively, although pulse oximeter SpO2 was only found to be inaccurate for dark skin pigmentation in sensitivity analyses.
“This systematic review and meta-analysis found that pulse oximeter SpO2 and wearable PR were inaccurate across all skin pigmentation groups as the resulting accuracy values breached FDA guidance and ANSI/AAMI/IEC standard thresholds, respectively, although pulse oximeter SpO2 was only found to be inaccurate for dark skin pigmentation in sensitivity analyses.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataSubgroup analyses were conducted to examine the potential moderating variables of medical versus healthy populations for pulse oximeters.
“Despite this, we conducted subgroup analyses to examine the potential moderating variables of medical versus healthy populations for pulse oximeters (see Table S8 in Multimedia Appendix 1).”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataResults suggest that SpO2 and pulse rate measurements are inaccurate across all skin pigmentation groups because they breach FDA and ANSI/AAMI/IEC standards, respectively.
“Results suggest inaccurate SpO2 and PR measurements across all skin pigmentation groups as they breach FDA and ANSI/AAMI/IEC standards, respectively, with wearable accuracy varying considerably depending on the model, which may be due to date of model production or algorithm development.”
Quote in the paperSubstantial claim
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Darker skin, with higher melanin levels, absorbs more light, reducing signal strength and limiting light penetration to blood vessels.
“Darker skin, with higher melanin levels, absorbs more light, reducing signal strength and limiting light penetration to blood vessels [19].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [19]
The Discussion’s Strengths and Limitations section says skin tone is determined by the amount of melanin in the epidermis. But neither that section nor the Principal Findings discussion states that darker skin absorbs more light, weakens the signal, or limits light penetration to blood vessels; the discussion only suggests that melanin levels could affect the PPG signal. No page numbers are provided in our copy. (LLM, conf=0.82)
Experimental — accuracy not yet measuredWeaker signals and increased scattering from darker skin degrade measurement accuracy.
“This results in weaker signals and increased scattering, which degrade measurement accuracy [19,20].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [19]
The Results (PPG Accuracy and Bias by Skin Pigmentation; page not provided) report higher pooled pulse-oximetry bias for darker than lighter skin, supporting a difference in measurement accuracy. The paper does not establish weaker signals or increased scattering as the cause of that difference. (LLM, conf=0.85)
Experimental — accuracy not yet measuredSocioeconomic disparities in access to smartwatches may interact with skin-tone-related bias, raising additional equity concerns.
“However, regional variations may exist, and socioeconomic disparities in access to these devices may further interact with skin-tone-related bias, raising additional equity concerns [17-23].”
From Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring (2025) · cites this paper as [19]
The Introduction (page unavailable) links potential skin-pigmentation bias in wearable devices to concerns about inequity and discrimination. It does not discuss socioeconomic disparities in smartwatch access or whether those disparities interact with skin-tone-related bias. (LLM, conf=0.84)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Medical-combined setting percentage is miscalculated
MinorTable 3 lists one medical-combined study among 23 pulse-oximetry studies as 5%; 1 out of 23 is approximately 4.35%, which rounds to 4% at the table's whole-percent precision.
Reported 5 · recomputed 4.347826
“| | Medical combined | 1 (5) | 0 (0) |”
Pulse-rate within-study heterogeneity conflicts with Table 4
MinorThe results text attributes 73.34% of pulse-rate variation to within-study differences, whereas the combined pulse-rate row of Table 4 reports within-study heterogeneity of 3.90% and overall heterogeneity of 30.66%.
“while 73.34% (I2Level2) originated from within-study variation.”
“| | Combined | 4 (24) | 176 (140,771) | bpm | –0.29 (–3.87 to 3.29) | 8.64 (1.67) | –17.23g to 16.65g | — | 30.66% (26.76% and 3.90%) |”
Pulse-rate clinical-relevance rule contradicts the stated conclusion
MinorTable 2 calls a pulse-rate bias clinically relevant if the upper 95% CI bound is at most 5 bpm. The reported upper bounds for the light and medium groups are 2.83 and 1.93 bpm, so those groups meet the stated rule despite the conclusion of no clinically relevant bias.
“Clinically relevant bias if either upper bound of 95% CI ≤5 bpm or lower bound of 95% CI >5 bpm”
“PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.”
Cites
The PRISMA 2020 statement: An updated guideline for reporting systematic reviews
Page, M. J., McKenzie, J. E., Bossuyt, P. M., Boutron, I., Hoffmann, T. C., Mulrow, C. D., Shamseer, L., Tetzlaff, J. M., Akl, E. A., Brennan, S. E., et al. (2021) · BMJ · doi:10.1136/bmj.n71 · Cited by 108,283 · see citing works
Conducting Meta-Analyses in <i>R</i> with the <b>metafor</b> Package
Viechtbauer, W. (2010) · Journal of Statistical Software · doi:10.18637/jss.v036.i03 · Cited by 20,464 · cited 2× · see citing works
“All analyses were conducted using R (version 4.3.1; R Foundation for Statistical Computing) with statistical models based on R packages metafor [63] and clubSandwich [65].”
InfluentialRacial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114 · cited 3× · see citing works
“…with lighter skin color to have overestimated SpO 2 readings, leading to lower hospital admission; higher occult hypoxemia; and delayed or no access to dexamethasone, therapeutic oxygen, and COVID-19 therapies, resulting in increased hospital readmission, organ dysfunction, and mortality [11-18].”
InfluentialMeta-analysis with robust variance estimation: Expanding the range of working models
Pustejovsky, J. E., & Tipton, E. (2022) · Prevention Science · doi:10.1007/s11121-021-01246-3 · Cited by 737 · cited 1× · see citing works
“To account for this dependence, the CHE model used the robust variance estimation (RVE) method, which allowed us to combine data from single-measure design studies with multiple dependent estimates of effect size, even when the dependence is unknown [64].”
Investigating sources of inaccuracy in wearable optical heart rate sensors
Bent, B., Goldstein, B. A., Kibbe, W. A., & Dunn, J. P. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0226-6 · Cited by 673 · cited 2× · see citing works
“Mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for the corresponding groups (Table 4 and Figures 5-7 [41,60-62]).”
InfluentialGoldsack, J. C., Coravos, A., Bakker, J. P., Bent, B., Dowling, A. V., Fitzer-Attas, C., Godfrey, A., Godino, J. G., Gujar, N., Izmailova, E., et al. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0260-4 · Cited by 557 · cited 2× · see citing works
“It is vitally important for medical pulse oximetry devices as well as nonregulated research and consumer devices to incorporate the V3 framework [74] for sensor verification and analytical validation of derived SpO 2 and PR values for regulatory submission before these metrics can be responsibly…”
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427 · cited 1× · see citing works
“Notably, darker skin pigmentation may influence PPG-based SpO 2 and PR readings [7-10].”
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400 · cited 1× · see citing works
“The pooled A rms across different skin pigmentation groups was 3.6%, 3.96%, 4.71%, and 4.15% for combined, light, medium, and dark skin pigmentation, respectively (Table 4 and Figures 2-4 [8-11,17,42-51,54-56,58,59]).”
InfluentialSources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · cited 1× · see citing works
“Researchers have identified several factors that influence the accuracy of PPG-based measurements, including skin temperature, sensor contact pressure, skin thickness, and hydration level [5,6].”
Fawzy, A., Wu, T. D., Wang, K., Robinson, M. L., Farha, J., Bradke, A., Golden, S. H., Xu, Y., & Garibaldi, B. T. (2022) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2022.1906 · Cited by 326 · cited 2× · see citing works
“…showing pulse oximeters significantly overestimating SpO 2 were expected for dark pigmentation and supported by findings on patient outcomes [11-15, 17 ], but Lastly, it should be noted that both SpO 2 and PR studies had lower between-study heterogeneity and higher within-study heterogeneity,…”
InfluentialNelson, B. W., & Allen, N. B. (2019) · JMIR mHealth and uHealth · doi:10.2196/10828 · Cited by 310 · cited 1× · see citing works
“Mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for the corresponding groups (Table 4 and Figures 5-7 [41,60-62]).”
InfluentialSelf-citationJubran, A., & Tobin, M. J. (1990) · Chest · doi:10.1378/chest.97.6.1420 · Cited by 296 · cited 1× · see citing works
“The pooled A rms across different skin pigmentation groups was 3.6%, 3.96%, 4.71%, and 4.15% for combined, light, medium, and dark skin pigmentation, respectively (Table 4 and Figures 2-4 [8-11,17,42-51,54-56,58,59]).”
InfluentialWong, A.-K. I., Charpignon, M., Kim, H., Josef, C., de Hond, A. A. H., Fojas, J. J., Tabaie, A., Liu, X., Mireles-Cabodevila, E., Carvalho, L., et al. (2021) · JAMA Network Open · doi:10.1001/jamanetworkopen.2021.31674 · Cited by 268 · see citing works
Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research
Nelson, B. W., Low, C. A., Jacobson, N., Areán, P., Torous, J., & Allen, N. B. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0297-4 · Cited by 241 · cited 6× · see citing works
“This has enabled consumers to continuously and unobtrusively track health status, generating information that is becoming commonly used in health care settings and as a source of research end points [3,4].”
InfluentialSelf-citationImproving Representation in Clinical Trials and Research
Affairs, P. A. G., & National Academies of Sciences, E. (2022) · National Academies Press eBooks · doi:10.17226/26479 · Cited by 202 · see citing works
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · cited 8× · see citing works
“To reduce heterogeneity, this meta-analysis reclassified race, ethnicity, and skin tone into a universal schema for skin pigmentation based on the system used by Shi et al [36].”
InfluentialRies, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145 · see citing works
Chow, H.-W., & Yang, C.-C. (2020) · JMIR mHealth and uHealth · doi:10.2196/14707 · Cited by 140 · cited 1× · see citing works
“Mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for the corresponding groups (Table 4 and Figures 5-7 [41,60-62]).”
InfluentialLimiting racial disparities and bias for wearable devices in health science research
Colvonen, P. J., DeYoung, P. N., Bosompra, N.-O. A., & Owens, R. L. (2020) · Sleep · doi:10.1093/sleep/zsaa159 · Cited by 137 · cited 3× · see citing works
“More recent research has also highlighted the potential presence of these biases in research-grade and consumer PPG-based wearable devices [4,27-29], leading to calls to address inequity, bias, and discrimination in wearable health technology and clinical practice algorithms [30,31].”
Gottlieb, E. R., Ziegler, J., Morley, K., Rush, B., & Celi, L. A. (2022) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2022.2587 · Cited by 134 · see citing works
Utility of pulse oximetry in the detection of arterial hypoxemia in liver transplant candidates
Abrams, G. A., Sanders, M. K., & Fallon, M. B. (2002) · Liver Transplantation · doi:10.1053/jlts.2002.32252 · Cited by 127 · cited 2× · see citing works
“It was calculated as , where precision is the SD of bias [42,43].”
InfluentialValbuena, V. S. M., Seelye, S., Sjoding, M. W., Valley, T. S., Dickson, R. P., Gay, S. E., Claar, D., Prescott, H. C., & Iwashyna, T. J. (2022) · BMJ · doi:10.1136/bmj-2021-069775 · Cited by 127 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
Racial discrepancy in pulse oximeter accuracy in preterm infants
Vesoulis, Z., Tims, A., Lodhi, H., Lalos, N., & Whitehead, H. (2022) · Journal of Perinatology · doi:10.1038/s41372-021-01230-3 · Cited by 120 · cited 1× · see citing works
“The pooled A rms across different skin pigmentation groups was 3.6%, 3.96%, 4.71%, and 4.15% for combined, light, medium, and dark skin pigmentation, respectively (Table 4 and Figures 2-4 [8-11,17,42-51,54-56,58,59]).”
InfluentialA review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119 · cited 1× · see citing works
“There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR [35] or pulse oximeter SpO 2 [36-38].”
Reliability of noninvasive oximetry in black subjects during exercise and hypoxia
Zeballos, R. J., & Weisman, I. M. (1991) · American Review of Respiratory Disease · doi:10.1164/ajrccm/144.6.1240 · Cited by 108 · see citing works
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102 · cited 1× · see citing works
“It was calculated as , where precision is the SD of bias [42,43].”
The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia
Foglia, E. E., Whyte, R. K., Chaudhary, A., Mott, A., Chen, J., Propert, K. J., & Schmidt, B. (2017) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2016.11.043 · Cited by 96 · see citing works
Pulse Oximeter Performance, Racial Inequity, and the Work Ahead
Okunlola, O. E., Lipnick, M. S., Batchelder, P. B., Bernstein, M., Feiner, J. R., & Bickler, P. E. (2022) · Respiratory Care · doi:10.4187/respcare.09795 · Cited by 96 · see citing works
Ebmeier, S. J., Barker, M., Bacon, M., Beasley, R. C., Bellomo, R., Chong, C. K., Eastwood, G. M., Gilchrist, J., Kagaya, H., Pilcher, J., et al. (2018) · Anaesthesia and Intensive Care · doi:10.1177/0310057x1804600307 · Cited by 92 · see citing works
Koerber, D., Khan, S., Shamsheri, T., Kirubarajan, A., & Mehta, S. (2023) · Journal of Racial and Ethnic Health Disparities · doi:10.1007/s40615-022-01446-9 · Cited by 91 · cited 1× · see citing works
“There have been systematic reviews on PPG skin pigmentation bias, but these have been limited to either consumer device PR [35] or pulse oximeter SpO 2 [36-38].”
Crooks, C. J., West, J., Morling, J. R., Simmonds, M., Juurlink, I., Briggs, S., Cruickshank, S., Hammond-Pears, S., Shaw, D., Card, T. R., et al. (2022) · European Respiratory Journal · doi:10.1183/13993003.03246-2021 · Cited by 68 · cited 3× · see citing works
“Sensitivity analyses were performed for pulse oximetry SpO 2 by removing Crooks et al [48] as an outlier due to substantially large SE (Tables S6 and S7 in Multimedia Appendix 1).”
InfluentialSelf-reported Race/Ethnicity and Intraoperative Occult Hypoxemia: A Retrospective Cohort Study
Burnett, G. W., Stannard, B., Wax, D. B., Lin, H.-M., Pyram-Vincent, C., DeMaria, S., & Levin, M. A. (2022) · Anesthesiology · doi:10.1097/aln.0000000000004153 · Cited by 58 · see citing works
Association of Race With Pulse Oximetry Accuracy in Hospitalized Children
Andrist, E., Nuppnau, M., Barbaro, R. P., Valley, T. S., & Sjoding, M. W. (2022) · JAMA Network Open · doi:10.1001/jamanetworkopen.2022.4584 · Cited by 58 · see citing works
Sudat, S. E. K., Wesson, P., Rhoads, K. F., Brown, S., Aboelata, N., Pressman, A. R., Mani, A., & Azar, K. M. J. (2022) · American Journal of Epidemiology · doi:10.1093/aje/kwac164 · Cited by 58 · cited 1× · see citing works
“The pooled A rms across different skin pigmentation groups was 3.6%, 3.96%, 4.71%, and 4.15% for combined, light, medium, and dark skin pigmentation, respectively (Table 4 and Figures 2-4 [8-11,17,42-51,54-56,58,59]).”
InfluentialWearable devices—addressing bias and inequity
Zinzuwadia, A., & Singh, J. P. (2022) · The Lancet Digital Health · doi:10.1016/s2589-7500(22)00194-7 · Cited by 58 · cited 1× · see citing works
“More recent research has also highlighted the potential presence of these biases in research-grade and consumer PPG-based wearable devices [4,27-29], leading to calls to address inequity, bias, and discrimination in wearable health technology and clinical practice algorithms [30,31].”
Accuracy and reliability of pulse oximetry at different arterial carbon dioxide pressure levels
Muñoz, X., Torres, F., Sampol, G., Ríos, J., Martí, S., & Escrich, E. (2008) · European Respiratory Journal · doi:10.1183/09031936.00126507 · Cited by 57 · see citing works
Photoplethysmography: Technology, Signal Analysis and Applications
Kyriacou, P. A., & Allen, J. (2021) · Cited by 56 · see citing works
McGovern, J., Sasse, S. A., Stansbury, D. W., Causing, L. A., & Light, R. W. (1996) · CHEST Journal · doi:10.1378/chest.109.5.1151 · Cited by 55 · see citing works
Photoplethysmography: Technology, Signal Analysis and Applications
Kyriacou, P. A., & Allen, J. (2021) · Newcastle University ePrints (Newcastle Univesity) · doi:10.25405/ncl.34438614 · Cited by 55
Prevention of Bias and Discrimination in Clinical Practice Algorithms
Shachar, C., & Gerke, S. (2023) · JAMA · doi:10.1001/jama.2022.23867 · Cited by 53 · see citing works
Accuracy of pulse oximetry in pigmented patients.
Bothma, P. A., Joynt, G. M., Lipman, J., Hon, H., Mathala, B., Scribante, J., & Kromberg, J. G. R. (1996) · PubMed · Cited by 50 · see citing works
Pilcher, J., Ploen, L., McKinstry, S., Bardsley, G., Chien, J., Howard, L., Lee, S., Beckert, L., Swanney, M. P., Weatherall, M., et al. (2020) · BMC Pulmonary Medicine · doi:10.1186/s12890-019-1007-3 · Cited by 50 · cited 1× · see citing works
“The pooled A rms across different skin pigmentation groups was 3.6%, 3.96%, 4.71%, and 4.15% for combined, light, medium, and dark skin pigmentation, respectively (Table 4 and Figures 2-4 [8-11,17,42-51,54-56,58,59]).”
InfluentialFawzy, A., Wu, T. D., Wang, K., Sands, K. E., Fisher, A. M., Arnold Egloff, S. A., DellaVolpe, J. D., Iwashyna, T. J., Xu, Y., & Garibaldi, B. T. (2023) · JAMA Network Open · doi:10.1001/jamanetworkopen.2023.30856 · Cited by 50 · see citing works
Accuracy of Pulse Oximetry During Hypoxemia
Thrush, D., & HODGES, M. R. (1994) · Southern Medical Journal · doi:10.1097/00007611-199404000-00019 · Cited by 49 · see citing works
Sañudo, B., De Hoyo, M., Muñoz-López, A., Perry, J., & Abt, G. (2019) · Journal of Medical Systems · doi:10.1007/s10916-019-1325-2 · Cited by 48 · see citing works
Clinical Algorithms, Antidiscrimination Laws, and Medical Device Regulation
Goodman, K. E., Morgan, D. J., & Hoffmann, D. E. (2023) · JAMA · doi:10.1001/jama.2022.23870 · Cited by 48 · cited 1× · see citing works
“More recent research has also highlighted the potential presence of these biases in research-grade and consumer PPG-based wearable devices [4,27-29], leading to calls to address inequity, bias, and discrimination in wearable health technology and clinical practice algorithms [30,31].”
Wiles, M. D., El‐Nayal, A., Elton, G., Malaj, M., Winterbottom, J., Gillies, C., Moppett, I. K., & Bauchmuller, K. (2022) · Anaesthesia · doi:10.1111/anae.15581 · Cited by 45 · cited 1× · see citing works
“The pooled A rms across different skin pigmentation groups was 3.6%, 3.96%, 4.71%, and 4.15% for combined, light, medium, and dark skin pigmentation, respectively (Table 4 and Figures 2-4 [8-11,17,42-51,54-56,58,59]).”
InfluentialInvestigating the effects of temperature on photoplethysmography
Khan, M., Pretty, C. G., Amies, A. C., Elliott, R., Shaw, G. M., & Chase, J. G. (2015) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2015.10.166 · Cited by 41 · cited 1× · see citing works
“Researchers have identified several factors that influence the accuracy of PPG-based measurements, including skin temperature, sensor contact pressure, skin thickness, and hydration level [5,6].”
Skin colour affects the accuracy of medical oxygen sensors
Keller, M. D., Harrison-Smith, B., Patil, C., & Arefin, M. S. (2022) · Nature · doi:10.1038/d41586-022-03161-1 · Cited by 38 · cited 1× · see citing works
“…with lighter skin color to have overestimated SpO 2 readings, leading to lower hospital admission; higher occult hypoxemia; and delayed or no access to dexamethasone, therapeutic oxygen, and COVID-19 therapies, resulting in increased hospital readmission, organ dysfunction, and mortality [11-18].”
Racial effects on Masimo pulse oximetry: a laboratory study
Barker, S. J., & Wilson, W. C. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-022-00927-w · Cited by 36 · see citing works
Evaluating the Accuracy of Pulse Oximetry in Children According to Race
Ruppel, H., Makeneni, S., Faerber, J., Lane-Fall, M. B., Foglia, E. E., O'Byrne, M. L., & Bonafide, C. P. (2023) · Archives of Pediatrics and Adolescent Medicine · doi:10.1001/jamapediatrics.2023.0071 · Cited by 36 · see citing works
Sjoding, M. W., Iwashyna, T. J., & Valley, T. S. (2022) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.202209-1773ed · Cited by 29 · cited 2× · see citing works
“To generate fit-for-purpose evidence applicable to diverse population, here, we propose 5 recommendations based on FDA guidance and literature (Textbox 1) [4,30,31,67-73]: Textbox 1.”
Trends in Consumer Wearable Devices With Cardiac Sensors in a Primary Care Cohort
Al‐Alusi, M. A., Khurshid, S. S., Wang, X., Venn, R. A., Pipilas, D. C., Ashburner, J. M., Ellinor, P. T., Singer, D. E., Atlas, S. J., & Lubitz, S. A. (2022) · Circulation Cardiovascular Quality and Outcomes · doi:10.1161/circoutcomes.121.008833 · Cited by 28 · cited 2× · see citing works
“This has enabled consumers to continuously and unobtrusively track health status, generating information that is becoming commonly used in health care settings and as a source of research end points [3,4].”
Keiser, E., Linos, E., Kanzler, M., Lee, W., Sainani, K. L., & Tang, J. Y. (2012) · Journal of the American Academy of Dermatology · doi:10.1016/j.jaad.2011.02.044 · Cited by 20 · cited 2× · see citing works
“This is particularly problematic as skin tone is a physiological concept (determined by the melanin amount in the basal layer of the epidermis), while ethnicity and race are largely social constructs, with high underlying physiological heterogeneity [66].”
Flagged first, then by citations
Research works printed but not linked: 1
- Bothma, P A, 1996, S Afr Med J, 86, 594
Cited by
Sorted by citations
Wearable Devices in Healthcare Beyond the One-Size-Fits All Paradigm
Bignami, E. G., Fornaciari, A., Fedele, S., Madeo, M., Panizzi, M., Marconi, F., Cerdelli, E., & Bellini, V. (2025) · Sensors · doi:10.3390/s25206472 · Cited by 16 · see citing works
“The optical sensors that are part of WDs, based on photoplethysmography (PPG) in relation to skin pigmentation, can overestimate oxygen saturation in subjects with dark phototypes, exposing them to a greater risk of undetected hypoxemia and possible clinical disparities [34,35].”
Khan, W. U., Alissa, M., Ma, H., Bhatti, U. A., Alghamdi, A. A., Alshehri, M. A., & Albelasi, A. (2025) · Materials Today Bio · doi:10.1016/j.mtbio.2025.102259 · Cited by 12 · see citing works
“A graphical representation of various magnetic sensors, including HES, AMR, GMR, TMR, and GMI sensors. consistency across various skin types, fostering equitable healthcare solutions [53,54].”
Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health Monitoring
Asif, S., AlSaafeen, A., Nadar, S., Nambiar, S., Dannawi, J., Korrapati, H., & Wilkhoo, H. S. (2025) · Cureus · doi:10.7759/cureus.94074 · Cited by 9 · see citing works
“This results in weaker signals and increased scattering, which degrade measurement accuracy [19,20].”
Combating cardiovascular disease disparities: The potential role of artificial intelligence
Orakwue, C. J., Tajrishi, F. Z., Gistand, C. M., Feng, H., & Ferdinand, K. C. (2025) · American Journal of Preventive Cardiology · doi:10.1016/j.ajpc.2025.100954 · Cited by 9 · see citing works
“Nevertheless, ongoing research aims to validate these technologies in diverse clinical settings, including cohorts with varying skin tones [31, 32].”
Al-Halawani, R., Qassem, M., & Kyriacou, P. A. (2025) · Sensors · doi:10.3390/s25020559 · Cited by 10 · see citing works
Wearable technologies in clinical trials for drug development: trends and emerging opportunities
Fayad, Z. A., Hirten, R. P., Nadkarni, G. N., & Kim, Y. (2026) · Nature Reviews Drug Discovery · doi:10.1038/s41573-026-01403-9 · Cited by 8 · see citing works
“In particular, darker skin pigmentation can reduce the amount of returning red and NIR light — both longer wavelengths used for SpO 2 estimation — which may bias saturation estimates upward by 1.5% 46 .”
InfluentialKim, D.-H., Han, J. Y., Jung, H., Song, D., Lee, C., Ryu, G., Hong, S. D., Kim, H. Y., & Jung, Y. G. (2025) · Nature and Science of Sleep · doi:10.2147/nss.s540460 · Cited by 8 · see citing works
“Recent meta-analyses have confirmed that this bias is statistically significant in participants with darker skin tones, 28 which is due to the optical characteristics where increased melanin reduces the alternating current to direct current ratio of PPG signals.”
Towett, G., Snead, R. S., Marczika, J., Ambalavanan, R., Mbogori-Kairichi, M., & Malioukis, A. (2025) · Digital Health · doi:10.1177/20552076251406654 · Cited by 10 · see citing works
Biostrap Kairos Wristband Versus Electrocardiography for Resting Heart Rate Variability Assessment
Flatt, A. A., Blalock, A. C. E., Wade, A. N., & Riemann, B. L. (2025) · Sensors · doi:10.3390/s25103165 · Cited by 4 · see citing works
“A MAPE <10% was considered good absolute agreement, while a MAPE >10% was considered poor absolute agreement [34].”
Cuffless blood pressure in 2025: from promise to practice: a narrative review
Wehbe, F., & Hiremath, S. S. (2025) · Current Opinion in Nephrology & Hypertension · doi:10.1097/mnh.0000000000001150 · Cited by 4 · see citing works
Li, W., Feng, J., Feng, G., & Feng, S. (2026) · Microchimica Acta · doi:10.1007/s00604-025-07792-9 · Cited by 4 · see citing works
“Physiological and anatomical differences, such as skin tone, hair density, age, gender, and body composition, can influence sensor accuracy, especially for optical technologies used in heart rate or oxygen saturation monitors [ 142, 143 ].”
Respiratory Depression in Non-Operating Room Anesthesia: An Overview
Royz, I. E., Clevenger, N. B., Bochenek, A., et al. (2025) · Journal of Clinical Medicine · doi:10.3390/jcm14134528 · Cited by 2 · see citing works
“However, some evidence suggests that pulse oximetry accuracy may not differ significantly across different skin tones [68].”
Srestha, S. D., & Kim, S. (2026) · PLoS ONE · doi:10.1371/journal.pone.0325340 · Cited by 2 · see citing works
“Unlike cPPG, which relies on optical measurement of blood volume changes and is susceptible to skin pigmentation effects [41, 42], the respiration belt used in this study operates via mechanical sensing of thoracic expansion.”
Martín-Escudero, P., & Cabanas, A. M. (2025) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2025.117213 · Cited by 2 · see citing works
Beydon, N., Simaga, B., Hammiche, H., et al. (2025) · Archives of Disease in Childhood · doi:10.1136/archdischild-2025-328723 · Cited by 1 · see citing works
“%) by the experts is probably due to the SpO 2 measurement’s imprecision, 4 which is variable depending on the oximeter that has been used.”
Al-Daraghmeh, M. Y., Stone, R., Mumani, A. A., & Jaskowiak, N. (2026) · International Journal of Industrial Ergonomics · doi:10.1016/j.ergon.2026.104023 · Cited by 1 · see citing works
Uc-Martín, J. A., Cortés, A., Castellón-Pérez, I., & Ramírez-Chavarría, R. G. (2026) · Biosensors · doi:10.3390/bios16080425 · Cited by 1 · see citing works
“Furthermore, the influence of skin tone on its accuracy remains a matter of debate [20,21], and its implementation requires comparatively complex architectures by integrating optical hardware with dedicated signal processing software [22].”
Mobile Health (mHealth) Apps in Sport Training: A Scoping Review
LIU, J., Dong, Y., Brooks, I. S., Cheung, W. C., Nguyen, V. L., & Jan, Y. (2026) · Sensors · doi:10.3390/s26175394 · Cited by 1 · see citing works
“…that optical sensing, the modality underlying photoplethysmography-based heart-rate and camera-based pose-estimation apps that together account for a substantial share of this review’s evidence base (Section 3.2), is systematically less accurate in users with darker skin pigmentation [146,147].”
InfluentialZhang, P., Zhuang, J., Li, T., Fan, D., Lei, Q., & Lin, D. (2025) · Scientific Reports · doi:10.1038/s41598-025-12927-2 · Cited by 1 · see citing works
“The agreement between systems for pulse rate was suboptimal, with 95% LoA (-8.57 to 7.60 beats/min) slightly exceeding the ± 5 beats/min threshold for critical care 14 .”
InfluentialOximetry at the crossroads of innovation and equity in sleep apnea diagnosis
Cruz, M. G. M. E., & Kryger, M. H. (2025) · Sleep Medicine Reviews · doi:10.1016/j.smrv.2025.102144 · Cited by 1 · see citing works
Schipper, R. M., Djibrillah, F. O., Roosink, M., Winkens, L. H. H., John, A., Hazebroek, E. J., Witteveen, A., & Berendsen, A. A. M. (2025) · doi:10.2196/preprints.87324 · Cited by 1 · see citing works
Elston, D. M. (2025) · Journal of the American Academy of Dermatology · doi:10.1016/j.jaad.2025.02.020
Evaluation of Two Asynchronous Envelope Detection Techniques for Photoplethysmography
Kovacs, R., Barbieri, R., & Brown, E. N. (2025) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc58623.2025.11251797
“For information on the latter three areas, the reader is referred to the following review articles [9]–[12].”
Wang, C.-F., Wang, H., Kuo, P.-H., Li, S.-Z., Chen, B., Li, S., Chen, Y., & Lin, S.-H. (2025) · American Journal of Hypertension · doi:10.1093/ajh/hpaf238
Ribeiro, R., Calvi, I. P., Ferreira, F., Lima, V., Bernardi, F. A., Conceição, L., Marreiros, G., & Freitas, A. (2026) · doi:10.2196/preprints.91812
Adaptive Hybrid Model for Enhanced PPG Signal Processing (HARMONY)
Ivković, R., Spalević, P., Kunicina, N., & Dragić, R. (2025) · IEEE International Scientific Conference on Power and Electrical Engineering of Riga Technical University (RTUCON) · doi:10.1109/rtucon67996.2025.11414961
“Skin pigmentation and thickness also affect light penetration and signal amplitude, necessitating adaptive algorithms for diverse populations [23], which has been shown to systematically impact pulse oximetry accuracy across different demographic groups [26].”
Lee, S.-K., Maryani, D., Lee, Y. Z., Phuna, Z. X., Ng, E. S. C., Bustami, N. A., Thiagarajah, S., Tan, C. K., & Tan, E. S. S. (2026) · Discover Public Health · doi:10.1186/s12982-026-01901-6
Pulse Oximetry, Skin Pigmentation, and Occult Hypoxemia in Pediatric Cardiac Critical Care
Savorgnan, F., Prabhu, J., Pilla, P., Flores, S., Loomba, R. S., & Acosta, S. (2026) · Pediatric Cardiology · doi:10.1007/s00246-026-04292-2
Monitoring and assessment for obstructive sleep apnea
Xu, Y., Lin, F., Chen, S., You, Z., Zhu, S., Zhao, W., & Zhao, Q. (2026) · FlexMat. · doi:10.1002/flm2.70092
Schipper, R. M., Djibrillah, F. O., Roosink, M., Winkens, L. H. H., John, A., Hazebroek, E. J., Witteveen, A., & Berendsen, A. A. M. (2026) · JMIR mhealth and uhealth · doi:10.2196/87324 · Cited by 1 · see citing works
“However, the influence of characteristics such as sex, age, BMI, and skin tone on measurement validity remains largely unclear, and studies investigating this influence appear to be scarce [13-15].”
Opioid overdose detection and reversal with closed-loop systems
Angelone, L. M., & Koustova, E. (2026) · Nature Reviews Bioengineering · doi:10.1038/s44222-026-00492-w
“Finally, algorithmic bias risks undermining both clinical validity and user trust: training data that over-represent certain populations can yield poor performance for specific groups, as illustrated by pulse oximetry across skin tones 33,34 .”
Lucas, E. (2026) · Nursing Older People · doi:10.7748/nop.2026.e1564
Dante Gil, Z., & Giglio, A. (2026) · Revista Médica Clínica Las Condes · doi:10.1016/j.rmclc.2026.08.002
Ţiţoc, O.-A., & Tudose, D. (2026) · Results in Engineering · doi:10.1016/j.rineng.2026.113072
Physical characterization of sensor-skin coupling for wrist photoplethysmography signal acquisition
Cao, X., Liu, Y., Zhu, H., Zhang, G., Wang, C., Tao, R., Zhao, H., & Liu, J. (2026) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2026.118546
Sleep Data Repositories in Practice
Maghsoudi, A., Razjouyan, J., Sharafkhaneh, A., & Werner, J. K. (2026) · Sleep Medicine Clinics · doi:10.1016/j.jsmc.2026.08.003
- Authors
- Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W.
- Affiliations
- University of Michigan (US)
- Verily Life Sciences LLC (United States) (US)
- Beth Israel Deaconess Medical Center (US)
- Venue
- Journal of Medical Internet Research
- Publisher
- JMIR Publications Inc.
- Published
- 2024-10-10
- Volume
- 26
- Pages
- e62769
- ISSN
- 1438-8871 (ISSN-L)
- PMC
- PMC11502980
- OpenAlex
- W4401647034
- Cited by
- 36
- References
- 54
- Citations by year
- 2026: 17 · 2025: 19
- FWCI
- 1.68
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- Journal of Medical Internet Research (publishedVersion, cc-by)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- pulse oximetry, photoplethysmography, oxygen saturation, wearable devices, measurement bias, health disparities, dark skin, arterial oxygenation, electrocardiography comparison, limits of agreement, root mean square error, measurement reliability, measurement validity
- MeSH terms
- Wearable Electronic Devices, Female, Heart Rate, Humans, Male, Oximetry, Oxygen, Skin Pigmentation, Photoplethysmography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-17.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-17.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review