CoChat

Targeted quantitative analysis of eicosanoid lipids in biological samples using liquid chromatography–tandem mass spectrometry

Open access
Mesaros, Clementina, Lee, Seon Hwa, Blair, Ian A. · 2009 · Journal of Chromatography B
Cited by 102

The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.

Abstract

The eicosanoids are a large family of arachidonic acid oxidation products that contain twenty carbon atoms. Cyclooxygenase (COX)-derived eicosanoids have important roles as autacoids involved in the regulation of cardiovascular function and tumor progression. Lipoxygenase (LO)-derived eicosanoids have been implicated as important mediators of inflammation, asthma, cardiovascular disease and cancer. Cytochrome P-450 (P450)-derived eicosanoids are both vasodilators and vasoconstrictors. There is intense interest in the analysis of reactive oxygen species (ROS)-derived isoprostanes (isoPs) because of their utility as biomarkers of oxidative stress. Enzymatic pathways of eicosanoid formation are regioselective and enantioselective, whereas ROS-mediated eicosanoid formation proceeds with no stereoselectivity. Many of the eicosanoids are also present in only pM concentrations in biological fluids. This presents a formidable analytical challenge because methodology is required that can separate enantiomers and diastereomers with high sensitivity and specificity. However, the discovery of atmospheric pressure ionization (API)/MS methodology of electrospray ionization (ESI), atmospheric pressure chemical ionization (APCI), and electron capture (EC) APCI has revolutionized our ability to analyze endogenous eicosanoids. LC separations of eicosanoids can now be readily coupled with API ionization, collision induced dissociation (CID) and tandem MS (MS/MS). This makes it possible to efficiently conduct targeted eicosanoid analyses using LC-multiple reaction motoring (MRM)/MS. Several examples of targeted eicosanoid lipid analysis using conventional LC-ESI/MS have been discussed and some new data on the analysis of eicosanoids using chiral LC-ECAPCI/MS has been presented.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.jchromb.2009.03.011

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 7 references

Cited by

See all 102

Metadata

Authors
Mesaros, C., Lee, S. H., & Blair, I. A.
Venue
Journal of Chromatography B
Publisher
Elsevier BV
Published
2009-09
Volume
877
Issue
26
Pages
2736-2745
ISSN
1570-0232 (ISSN-L)
OpenAlex
W2070048233
Cited by
102
References
7
FWCI
6.00
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Targeted quantitative analysis of eicosanoid lipids in biological samples using liquid chromatography–tandem mass spectrometry. https://app.cochat.ai/papers/10.1016/j.jchromb.2009.03.011
The paper
Mesaros, C., Lee, S. H., & Blair, I. A. (2009). Targeted quantitative analysis of eicosanoid lipids in biological samples using liquid chromatography–tandem mass spectrometry. Journal of Chromatography B. https://doi.org/10.1016/j.jchromb.2009.03.011

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.