STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 20,501–20,600 of 47,868
Evaluation of ammonia measurements in dogs with two analysers for use in veterinary practice
Sterczer, Á., Meyer, H. P., Boswijk, H., & Rothuizen, J. J. (1999) · Veterinary Record · doi:10.1136/vr.144.19.523 · Cited by 22
Gall, O., Annequin, D., Ravault, N., & Murat, I. (1999) · Pediatric Anesthesia · doi:10.1046/j.1460-9592.1999.00354.x · Cited by 22
Davey, R., Edwards, S. M., & Cochrane, T. (2003) · Advances in Physiotherapy · doi:10.1080/1403819038190310017075 · Cited by 23
BMR in a Brazilian adult probability sample: the Nutrition, Physical Activity and Health Survey
dos Anjos, L. A., Wahrlich, V., & de Vasconcellos, M. T. L. (2013) · Public Health Nutrition · doi:10.1017/s1368980012005381 · Cited by 24
Comparison of PET with radioactive microspheres to assess pulmonary blood flow.
Richard, J., Janier, M. F., Decailliot, F., Bars, D. L., Lavenne, F., Berthier, V., Lionnet, M., Cinotti, L., Annat, G. J., & Guérin, C. (2002) · PubMed · Cited by 22
Bocksch, W., Wellnhofer, E., Schartl, M., Dreysse, S., Klimek, W. J., Franke, R., Musci, M., Hetzer, R., & Fleck, E. (2000) · Coronary Artery Disease · doi:10.1097/00019501-200010000-00007 · Cited by 22
The pressure phosphene tonometer—a clinical evaluation
Chew, G., SANDERSON, G. F., & Molteno, A. C. (2004) · Eye · doi:10.1038/sj.eye.6701600 · Cited by 22
Using the Larsen index to assess radiographic progression in rheumatoid arthritis.
Kirwan, J. (2000) · PubMed · Cited by 26
Effects of Acute Gastric Distention and Recovery on Tendency for Ventricular Arrhythmia in Dogs
Michau, T. M., Schwartz, D. S., Nakayama, T., & Hamlin, R. L. (2000) · Journal of Veterinary Internal Medicine · doi:10.1111/j.1939-1676.2000.tb02253.x · Cited by 22
Surface area in children: a simple formula.
Furqan, M., & Haque, A. (2009) · PubMed · Cited by 26
Semi-automatic Volumetric Measurement of Lung Cancer Using Multi-detector CT
Iwano, S., Okada, T., Koike, W., Matsuo, K., Toya, R., Yamazaki, M., Ito, S., Ito, J., & Naganwa, S. (2009) · Academic Radiology · doi:10.1016/j.acra.2009.04.007 · Cited by 22
Giorgetti, A., Marzullo, P., Sambuceti, G., Quirico, S. D., Kusch, A., Landi, P., Salvadori, P. A., Pisani, P., & L’Abbate, A. (2004) · Journal of Nuclear Cardiology · doi:10.1016/j.nuclcard.2003.12.011 · Cited by 22
Treatment of spondyloarthropathies. Recent advances and prospects in 2001
Dougados, M. (2001) · Joint Bone Spine · doi:10.1016/s1297-319x(01)00328-1 · Cited by 22
Accuracy of Three Pulse Oximeters During Exercise and Hypoxemia in Patients With Cystic Fibrosis
Orenstein, D. M., Curtis, S. E., Nixon, P. A., & Hartigan, E. R. (1993) · CHEST Journal · doi:10.1378/chest.104.4.1187 · Cited by 22
Cornish, B., Ward, L. C., & Thomas, B. J. (1994) · Nutrition Research · doi:10.1016/s0271-5317(05)80207-6 · Cited by 22
The effect of patient technique and training on the accuracy of self-recorded peak expiratory flow
Gannon, P. F. G., Belcher, J., Pantin, C. F. A., & Burge, P. S. (1999) · European Respiratory Journal · doi:10.1034/j.1399-3003.1999.14a07.x · Cited by 22
Spadea, L., Giammaria, D., Ferrante, R., & Balestrazzi, E. (2005) · Ophthalmology · doi:10.1016/j.ophtha.2004.12.031 · Cited by 22
Bruining, N., Hamers, R., Teo, T., de Feijter, P. J., Serruys, P. W. J. C., & Roelandt, J. R. T. C. (2004) · International journal of cardiac imaging · doi:10.1023/b:caim.0000014046.63648.1c · Cited by 22
Moral, R. R., Castillejo, J. A. P., Jurado, M. A., Bellón, J. Á., & de Torres, L. Á. P. (2001) · Atención Primaria · doi:10.1016/s0212-6567(01)78837-3 · Cited by 22
Vanhove, C., & Franken, P. R. (2001) · Journal of Nuclear Cardiology · doi:10.1067/mnc.2001.115518 · Cited by 22
Evaluation of calcaneus bone densitometry against hip and spine for diagnosis of osteoporosis
Williams, E. D., & Daymond, T. J. (2003) · British Journal of Radiology · doi:10.1259/bjr/56105358 · Cited by 22
Does adjustment of GFR to extracellular fluid volume improve the clinical utility of cystatin C?
Kilpatrick, E. S. (2000) · Archives of Disease in Childhood · doi:10.1136/adc.82.6.499 · Cited by 22
Evaluation of a disposable prism for applanation tonometry
Desai, S. P., Sivakumar, S., & Fryers, P. T. (2002) · Eye · doi:10.1038/sj.eye.6700129 · Cited by 22
Atherosclerosis and left ventricular hypertrophy
Gamble, G. D., MacMahon, S. W., Culpan, A., Ciobo, C., WHALLEY, G. A., & Sharpe, N. (1998) · Journal of Hypertension · doi:10.1097/00004872-199816090-00020 · Cited by 22
Singer, M., & Bennett, D. (1989) · BMJ · doi:10.1136/bmj.298.6684.1350 · Cited by 22
Municinò, A., SIMONE, G. D., Roman, M. J., Cody, R. J., Ganau, A., Hahn, R. T., & Devereux, R. B. (1996) · The American Journal of Cardiology · doi:10.1016/s0002-9149(96)00362-1 · Cited by 25
Backer, H., & COLLINS, S. (1999) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(99)70306-9 · Cited by 22
Reproducibility of the acetylene rebreathe technique for determining cardiac output
Warburton, D. E. R., Gledhill, N., & Jamnik, V. (1998) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-199806000-00027 · Cited by 22
Migaleddu, V., Virgilio, G., Turilli, D., Conti, M., Campisi, G., Canu, N. A., Sirigu, D., & Vincentelli, I. (2004) · American Journal of Roentgenology · doi:10.2214/ajr.182.6.1821505 · Cited by 22
Evaluation of two new high-sensitivity methods for C-reactive protein
Kos, S., Bekers, O., Gubbels, A., Drent, M., Schmitz, M. P., & van Dieijen‐Visser, M. P. (2003) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1258/000456303766477057 · Cited by 23
Lee, D. S., Cheon, G. J., Ahn, J., Chung, J., & Lee, M. C. (2000) · Nuclear Medicine Communications · doi:10.1097/00006231-200012000-00006 · Cited by 22
Automatic backscatter analysis of regional left ventricular systolic function using color kinesis
Schwartz, S. L., Cao, Q., Vannan, M. A., & Pandian, N. G. (1996) · The American Journal of Cardiology · doi:10.1016/s0002-9149(96)00203-2 · Cited by 22
Interrupter technique for evaluation of exercise-induced bronchospasm in children
Kannisto, S., Vanninen, E. J., Remes, K., & Korppi, M. O. (1999) · Pediatric Pulmonology · doi:10.1002/(sici)1099-0496(199903)27:3<203::aid-ppul9>3.0.co;2-g · Cited by 27
Lebecque, P., Shango, P., Stijns, M., Vliers, A., & Coates, A. L. (1991) · Pediatric Pulmonology · doi:10.1002/ppul.1950100216 · Cited by 22
Electrodiagnostic Automation: Principles and Practice
Gozani, S. N., Fisher, M. A., Kong, X., Megerian, J. T., & Rutkove, S. B. (2005) · Physical Medicine and Rehabilitation Clinics of North America · doi:10.1016/j.pmr.2005.08.005 · Cited by 22
Ozaki, Y., Keane, D., HERRMAN, J. P., Foley, D., Haase, J., den Boer, A., Mario, C. D., & Serruys, P. W. J. C. (1995) · American Heart Journal · doi:10.1016/0002-8703(95)90270-8 · Cited by 27
O’Sullivan, M., Lewis, P. A., Newcombe, R. G., Broderick, N. J., Robinson, D. A., Coles, E. C., & Jessop, J. D. (1990) · Annals of the Rheumatic Diseases · doi:10.1136/ard.49.5.286 · Cited by 22
Puskás, C., Kosch, M., Kerber, S., Jonas, M., Weyand, M., BREITHARDT, G. E., Scheld, H. H., & Schober, O. (1997) · PubMed · Cited by 22
Jahnsen, J. (1999) · Scandinavian Journal of Gastroenterology · doi:10.1080/003655299750025723 · Cited by 22
Continuous intravascular blood gas monitoring
Wahr, J. A., & Tremper, K. K. (1994) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1016/1053-0770(94)90250-x · Cited by 22
Braggion, C., Pradal, U., & Mastella, G. (1992) · Acta Paediatrica · doi:10.1111/j.1651-2227.1992.tb12163.x · Cited by 22
Jc, H., Rabellino, E. M., Gewirtz, A., Lm, G., Rucinski, B., & Niewiarowski, S. (1988) · PubMed · Cited by 22
Cavelaars, M., Tulen, J. H. M., Veld, A. J. M. ʼ., Gelsema, E. S., & van den Meiracker, A. H. (2000) · Journal of Hypertension · doi:10.1097/00004872-200018120-00006 · Cited by 22
A Comparison of Chromogen Test Strip (Chemstrip bG) and Serum Glucose Values in Newborns
Holtrop, P. C., Madison, K. A., Kiechle, F. L., Karcher, R. E., & Batton, D. G. (1990) · Archives of Pediatrics and Adolescent Medicine · doi:10.1001/archpedi.1990.02150260063029 · Cited by 24
Assessment of health-enhancing physical activity at population level
Hagströmer, M. (2007) · Figshare · Cited by 37
Localization and determination of infarct size by Gd-Mesoporphyrin enhanced MRI in dogs
Herijgers, P., Laycock, S. K., Ni, Y., Marchal, G., Bogaert, J., Bosmans, H., ́, C. P., & Flameng, W. J. (1997) · International journal of cardiac imaging · doi:10.1023/a:1005715426172 · Cited by 22
Scott, D. J. A., Vowden, P., Beard, J. D., & Horrocks, M. (1990) · British journal of surgery · doi:10.1002/bjs.1800770412 · Cited by 22
IOP measurement and central corneal thickness
Stodtmeister, R., Kron, M., & Gaus, W. (2002) · British Journal of Ophthalmology · doi:10.1136/bjo.86.1.120 · Cited by 22
Objective morphological assessment of macular hole surgery by scanning laser tomography
Hudson, C., Charles, S., Flanagan, J. G., Brahma, A. K., Turner, G. S., & McLeod, D. (1997) · British Journal of Ophthalmology · doi:10.1136/bjo.81.2.107 · Cited by 22
Reproducibility of topographic measures of the glaucomatous optic nerve head
Geyer, O., Michaeli‐Cohen, A., Silver, D. M., Versano, D., Neudorfer, M., Dzhanov, R., & Lazar, M. (1998) · British Journal of Ophthalmology · doi:10.1136/bjo.82.1.14 · Cited by 25
On-line urea kinetics in haemodiafiltration
Santoro, A., Tetta, C., Mandolfo, S., Arrigo, S., Berti, M., Colasanti, G., D'Amico, G., Imbasciati, E., Mazzocchi, C., Pacini, G., et al. (1996) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/11.6.1084 · Cited by 22
Analytical Error of Home Glucose Monitors: A Comparison of 18 Systems
Johnson, R. N., & Baker, J. R. (1999) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456329903600110 · Cited by 23
Non-invasive fetal electrocardiogram:analysis and interpretation
Vullings, R. (2010) · TU/e Research Portal · doi:10.6100/ir692881 · Cited by 34
McLaren, J. W., & Bourne, W. M. (1999) · PubMed · Cited by 22
Costa, R. F. D., Silva, A. M., Masset, K. V. D. S. B., Cesário, T. D. M., Cabral, B. G. D. A. T., Ferrari, G., & Dantas, P. M. S. (2022) · Frontiers in Nutrition · doi:10.3389/fnut.2022.820736 · Cited by 21
Sénéchal, M., O'Connor, K., Deblois, J., Magne, J., Dumesnil, J., Pibarot, P., Bergeron, S., & Poirier, P. (2008) · Echocardiography · doi:10.1111/j.1540-8175.2007.00555.x · Cited by 21
Xie, F., Xu, J., Xia, L.-L., Luo, X., Jiang, Z., Wu, Y., & Su, H. (2020) · Hypertension Research · doi:10.1038/s41440-019-0386-4 · Cited by 21
Underreporting of energy intake in weight loss maintainers
Dahle, J. H., Ostendorf, D. M., Zaman, A., Pan, Z., Melanson, E. L., & Catenacci, V. A. (2021) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/nqab012 · Cited by 21
Wan, V., Reddy, S., Thomas, A., Issa, N., Posluszny, J. A., Schwulst, S. J., Shapiro, M. B., Alam, H. B., Bilimoria, K. Y., & Stey, A. M. (2022) · The Journal of Trauma: Injury, Infection, and Critical Care · doi:10.1097/ta.0000000000003656 · Cited by 21
Reyes-Hernandez, M. E., Bischoff, A. R., Giesinger, R. E., Rios, D. R., Stanford, A. H., & McNamara, P. J. (2023) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2023.08.013 · Cited by 23
Measurement properties of cervical joint position error in people with and without chronic neck pain
AlDahas, A., Devecchi, V., Deane, J. A., & Falla, D. (2023) · PLoS ONE · doi:10.1371/journal.pone.0292798 · Cited by 26
Accuracy and Reliability of Grip Strength Measurements: A Comparative Device Analysis
Gränicher, P., Maurer, Y., Spörri, J., Haller, B., Swanenburg, J., de Bie, R. A., Lenssen, A. F., & Scherr, J. (2024) · Journal of Functional Morphology and Kinesiology · doi:10.3390/jfmk9040274 · Cited by 27
Influence of the Camera Viewing Angle on OpenPose Validity in Motion Analysis
Baldinger, M., Reimer, L. M., & Senner, V. (2025) · Sensors · doi:10.3390/s25030799 · Cited by 24
Heyard, R., Pawel, S., Frese, J., Voelkl, B., Würbel, H., McCann, S., Held, L., Wever, K. E., Hartmann, H., Townsin, L., et al. (2025) · Royal Society Open Science · doi:10.1098/rsos.242076 · Cited by 18
Souka, A. P., Pilalis, A., Papastefanou, I., Salamalekis, G., & Kassanos, D. (2012) · Prenatal Diagnosis · doi:10.1002/pd.3976 · Cited by 21
Validation of multiple‐breath washout equipment for infants and young children
Schmidt, A., Yammine, S., Proietti, E., Frey, U., Latzin, P., Riedel, T., & Singer, F. (2014) · Pediatric Pulmonology · doi:10.1002/ppul.23010 · Cited by 22
Xia, Y., Kongstad, O., Hertervig, E. J., Li, Z., Holm, M., Olsson, B. S., & Yuan, S. (2005) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2004.09.010 · Cited by 21
Methods of Developing UWIN, the Modified American Urological Association Symptom Score
Barqawi, A. B., Sullivan, K. F., Crawford, E. D., Roehrborn, C. G., Hughes, A., O’Leary, M. P., Stone, N. N., & O’Donnell, C. I. (2011) · The Journal of Urology · doi:10.1016/j.juro.2011.04.057 · Cited by 21
Inaccuracy of the FloTrac/Vigileo™ System in Patients With Low Cardiac Index
Maeda, T., Yoshitani, K., Inatomi, Y., & Ohnishi, Y. (2014) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2014.04.013 · Cited by 22
Spearman's Footrule as a Measure of Agreement
Berry, K. J., & Mielke, P. W. (1997) · Psychological Reports · doi:10.2466/pr0.1997.80.3.839 · Cited by 23
Coghlan, J. A., Bell, S., Forbes, A. B., & Buchbinder, R. (2008) · Journal of Shoulder and Elbow Surgery · doi:10.1016/j.jse.2007.11.020 · Cited by 21
Dénes, M., Farkas, K., Erdei, T., & Lengyel, M. (2010) · Echocardiography · doi:10.1111/j.1540-8175.2009.01018.x · Cited by 21
Azizian, H., Kramer, J. K. G., & Mossoba, M. M. (2012) · Journal of the American Oil Chemists Society · doi:10.1007/s11746-012-2116-9 · Cited by 21
Biostatistics Primer for the Radiologist
Psoter, K. J., Roudsari, B. S., Dighe, M., Richardson, M. L., Katz, D. S., & Bhargava, P. (2014) · American Journal of Roentgenology · doi:10.2214/ajr.13.11657 · Cited by 21
Bachner-Hinenzon, N., Ertracht, O., Lysiansky, M., Binah, O., & Adam, D. (2010) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-010-0662-6 · Cited by 21
Automatic segmentation of cardiac MRI cines validated for long axis views
Tsadok, Y., Petrank, Y. F., Sarvari, S. I., Edvardsen, T., & Adam, D. R. (2013) · Computerized Medical Imaging and Graphics · doi:10.1016/j.compmedimag.2013.09.002 · Cited by 21
Guo, L., Cui, Y., Pharis, S., Walsh, M., Atallah, J., Meng-wei, T., Rutledge, J. M., Coe, J. Y., & Adatia, I. (2013) · Pediatric Cardiology · doi:10.1007/s00246-013-0855-5 · Cited by 21
Reliability of the Primary Triage Process after the Volendam Fire Disaster
Welling, L., van Harten, S. M., Henny, C. P., Mackie, D. P., Ubbink, D. T., Kreis, R. W., & Trouwborst, A. (2008) · Journal of Emergency Medicine · doi:10.1016/j.jemermed.2007.06.009 · Cited by 21
Finger counting: an alternative method for estimating pediatric weights
Young, T. P., Chen, B. G., Kim, T. Y., Thorp, A. W., & Brown, L. (2013) · The American Journal of Emergency Medicine · doi:10.1016/j.ajem.2013.11.034 · Cited by 21
Reproducibility of Ultrasonographic Measurements of the Ulnar Nerve at the Cubital Tunnel
Fink, A., Teggeler, M., Schmitz, M., Janssen, J., & Pisters, M. F. (2016) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2016.09.022 · Cited by 21
Hong, J. P., Nam, S. M., Kim, T., Seo, K. Y., Lee, S. Y., Meduri, A., & Kim, E. K. (2012) · Yonsei Medical Journal · doi:10.3349/ymj.2012.53.3.634 · Cited by 21
Robinson, D., Oelke, M., Khullar, V., Wijkstra, H., Tretter, R., Stow, B., Compion, G., & Tubaro, A. (2015) · Neurourology and Urodynamics · doi:10.1002/nau.22808 · Cited by 21
Quantification of Dynamic 11 C-Phenytoin PET Studies
Mansor, S., Boellaard, R., Froklage, F. E., Bakker, E. D. M., Yaqub, M. M., Voskuyl, R. A., Schwarte, L. A., Verbeek, J., Windhorst, A. D., & Lammertsma, A. A. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.158055 · Cited by 21
Shinoda, T., Kanai, M., Nakamura, R., Murata, S., SAITO, T. K., Sawa, R., Ono, R., & Izawa, K. P. (2016) · Aging Clinical and Experimental Research · doi:10.1007/s40520-016-0573-y · Cited by 21
Weight Estimation Tool for Children Aged 6 to 59 Months in Limited-Resource Settings
Ralston, M. E., & Myatt, M. A. (2016) · PLoS ONE · doi:10.1371/journal.pone.0159260 · Cited by 23
Nickerson, B. S., Esco, M. R., Bishop, P. A., Schumacker, R. E., Richardson, M. T., Fedewa, M. V., Wingo, J. E., & Welborn, B. A. (2016) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000001648 · Cited by 25
Aleksandric, S. B., Djordjevic‐Dikic, A., Beleslin, B., Parapid, B. J., Teofilovski‐Parapid, G., Stepanović, J., Šimić, D., Nedeljković, I., Petrović, M., Dobric, M. R., et al. (2016) · International Journal of Cardiology · doi:10.1016/j.ijcard.2016.09.101 · Cited by 21
Hamood, R., Hamood, H., Merhasin, I., & Keinan‐Boker, L. (2016) · Israel Journal of Health Policy Research · doi:10.1186/s13584-016-0111-6 · Cited by 21
Ramírez‐Vélez, R., Correa‐Bautista, J. E., González‐Ruíz, K., Vivas, A., García‐Hermoso, A., & Triana-Reina, H. R. (2016) · Nutrients · doi:10.3390/nu8120737 · Cited by 23
Podlesnikar, T., Prihadi, E. A., van Rosendael, P. J., Vollema, E. M., van der Kley, F., de Weger, A., Marsan, N. A., Naji, F. H., Fras, Z., Bax, J. J., et al. (2017) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2017.09.016 · Cited by 23
Knešaurek, K., Warnock, G., Kostakoğlu, L., Burger, C., & Initiative, A. D. N. (2018) · World Journal of Nuclear Medicine · doi:10.4103/wjnm.wjnm_5_17 · Cited by 24
Dekkers, I. A., de Heer, P., Bizino, M. B., de Vries, A. P. J., & Lamb, H. J. (2018) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26003 · Cited by 21
Determining the reliability of craniocervical flexion test in asymptomatic individuals
Kotwani, S., Bid, D. D., Ghatamaneni, D., Alahmari, K. A., Ramalingam, T., & Samuel, P. S. (2018) · Hong Kong Physiotherapy Journal · doi:10.1142/s101370251850004x · Cited by 29
Mercado, C. I., Cogswell, M. E., Loria, C. M., Liu, K., Allen, N. B., Gillespie, C., Wang, C., de Boer, I. H., & Wright, J. D. (2018) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/nqy138 · Cited by 22
Laza‐Cagigas, R., Goss-Sampson, M. A., Larumbe‐Zabala, E., Termkolli, L., & Naclerio, F. (2018) · Journal of Sports Sciences · doi:10.1080/02640414.2018.1527673 · Cited by 27
van Grootel, R. W. J., Strachinaru, M., Menting, M. E., McGhie, J. S., Roos‐Hesselink, J. W., & van den Bosch, A. E. (2018) · Echocardiography · doi:10.1111/echo.14174 · Cited by 22
Validation of the Danish version of the musculoskeletal tumour society score questionnaire
Sæbye, C., Keller, J. O., & Baad‐Hansen, T. (2019) · World Journal of Orthopedics · doi:10.5312/wjo.v10.i1.23 · Cited by 23
Larsen, J. B., Madeleine, P., & Arendt‐Nielsen, L. (2019) · Scandinavian Journal of Pain · doi:10.1515/sjpain-2018-0353 · Cited by 21
Peake, M., Pan, K. S., Rotatori, R. M., Powell, H. M., Fowler, L. A., James, L. E., & Dale, E. L. (2019) · Burns · doi:10.1016/j.burns.2019.05.005 · Cited by 25
A clinical method for estimating the modulus of elasticity of the human cornea in vivo
Pye, D. C. (2020) · PLoS ONE · doi:10.1371/journal.pone.0224824 · Cited by 21
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.