STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 6,701–6,800 of 47,868
Monitoring of Dual-Energy X-ray Absorptiometry Measurement in Clinical Practice
Maghraoui, A. E., Achemlal, L., & Bezza, A. (2006) · Journal of Clinical Densitometry · doi:10.1016/j.jocd.2006.03.014 · Cited by 70
Lopez‐Mattei, J., Ibrahim, H., Shaikh, K. A., Little, S. H., Shah, D. J., Maragiannis, D., & Zoghbi, W. A. (2015) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2015.10.045 · Cited by 73
Agreement between two inertial sensor gait analysis systems for lameness examinations in horses
Pfau, T., Boultbee, H., Davis, H., Walker, A. M., & Rhodin, M. (2015) · Equine Veterinary Education · doi:10.1111/eve.12400 · Cited by 72
Chak, M., Rahi, J. S., & on behalf of the British Congenital Cataract Interest Group (2007) · British Journal of Ophthalmology · doi:10.1136/bjo.2006.109603 · Cited by 76
Subjective outcomes after knee arthroplasty
Dunbar, M. (2001) · Acta Orthopaedica Scandinavica · doi:10.1080/000164701753739124 · Cited by 88
Marques, A. S. P. D. D., Cruz, J., Quina, S., Regêncio, M., & Jácome, C. I. O. (2015) · COPD Journal of Chronic Obstructive Pulmonary Disease · doi:10.3109/15412555.2015.1079816 · Cited by 80
Ruiz, S., Minville, V., Asehnoune, K., Virtos, M., Georges, B., Fourcade, O., & Conil, J. (2015) · Annals of Intensive Care · doi:10.1186/s13613-015-0090-8 · Cited by 72
Spectral characteristics of ventricular response to atrial fibrillation
Hayano, J., Yamasaki, F., Sakata, S., Okada, A., Mukai, S., & Fujinami, T. (1997) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1997.273.6.h2811 · Cited by 73
Ward, S., Hancox, A., Mohammed, M. A., Ismail, T., Griffiths, E. A., Valori, R., & Dunckley, P. (2016) · Gut · doi:10.1136/gutjnl-2015-310443 · Cited by 75
PCI Choice Decision Aid for Stable Coronary Artery Disease: A Randomized Trial
Coylewright, M., Dick, S., Zmolek, B., Askelin, J., Hawkins, E., Branda, M., Inselman, J. W., Zeballos-Palacios, C., Shah, N. D., Hess, E. P., et al. (2016) · Circulation: Cardiovascular Quality and Outcomes · doi:10.1161/circoutcomes.116.002641 · Cited by 70
Leinonen, A., Ahola, R., Kulmala, J., Hakonen, H., Vähä‐Ypyä, H., Herzig, K., AUVINEN, J. P., Keinanen-Kiukaanniemi, S. M., Sievänen, H. T., Tammelin, T., et al. (2017) · Frontiers in Physiology · doi:10.3389/fphys.2016.00681 · Cited by 85
Reinacher, P. C., Krüger, M. T., Coenen, V. A., Shah, M. J., Roelz, R., Jenkner, C., & Egger, K. E. (2017) · American Journal of Neuroradiology · doi:10.3174/ajnr.a5153 · Cited by 70
Benni, P. B., MacLeod, D. B., Ikeda, K., & Lin, H.-M. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-0015-1 · Cited by 77
Scarlata, S., Mancini, D., Laudisio, A., Benigni, A., & Incalzi, R. A. (2018) · Respiration · doi:10.1159/000489229 · Cited by 70
Vujosevic, S., Toma, C., Villani, E., Gatti, V., Brambilla, M., Muraca, A., Ponziani, M. C., Aimaretti, G., Nuzzo, A., Nucci, P., et al. (2019) · Journal of Diabetes Research · doi:10.1155/2019/2547216 · Cited by 70
Bakker, L. A., Schröder, C. D., Tan, H. H. G., Vugts, S. M. A. G., van Eijk, R. P. A., van Es, M. A., Visser‐Meily, J. M. A., & van den Berg, L. H. (2019) · Journal of Neurology Neurosurgery & Psychiatry · doi:10.1136/jnnp-2019-321138 · Cited by 74
Faucher, M., Poiraudeau, S., Lefèvre‐Colau, M., Rannou, F., Fermanian, J., & Revel, M. (2003) · Joint Bone Spine · doi:10.1016/s1297-319x(03)00070-8 · Cited by 70
Interobserver Reliability of the Prism and Alternate Cover Test in Children With Esotropia
Pediatric Eye Disease Investigator Group* (2009) · Archives of Ophthalmology · doi:10.1001/archophthalmol.2008.548 · Cited by 71
Guan, K., Hudson, C., & Flanagan, J. G. (2003) · Microvascular Research · doi:10.1016/s0026-2862(03)00007-4 · Cited by 70
mann, E. M. H., Grus, F., & Pfeiffer, N. (2004) · BMC Ophthalmology · doi:10.1186/1471-2415-4-4 · Cited by 70
Gugger, M. (1997) · European Respiratory Journal · doi:10.1183/09031936.97.10030587 · Cited by 71
Dean, D. A., Jia, C.-X., Cabreriza, S. E., ALESSANDRO, D. A. D., Dickstein, M. L., Sardo, M. J., Chalik, N., & Spotnitz, H. M. (1996) · ASAIO Journal · doi:10.1097/00002480-199609000-00072 · Cited by 70
Mh, C., Jd, L., Gutiérrez, E., Pacheco‐Tena, C., Vázquez‐Mellado, J., & Burgos‐Vargas, R. (2003) · PubMed · Cited by 70
BAILEY, S. M., SARMANDAL, P., & Grant, J. M. (1989) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.1989.tb03223.x · Cited by 70
Roland, N. J., CASLIN, A. W., Bowie, G., & Jones, A. S. (1994) · Clinical Otolaryngology · doi:10.1111/j.1365-2273.1994.tb01140.x · Cited by 73
Reliability of stationary dynamometer muscle strength testing in community-dwelling older adults
Ford-Smith, C. D., Wyman, J. F., Elswick, R. K., & Fernandez, T. M. (2001) · Archives of Physical Medicine and Rehabilitation · doi:10.1053/apmr.2001.24291 · Cited by 70
Caries Detection Methods: Can They Aid Decision Making for Invasive Sealant Treatment?
Pereira, A. C., Verdonschot, E. H., & Huysmans, M. C. D. N. J. M. (2001) · Caries Research · doi:10.1159/000047437 · Cited by 76
Harrenstien, L. A., Tornquist, S. J., Miller-Morgan, T. J., Fodness, B. G., & Clifford, K. E. (2005) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.2005.226.255 · Cited by 70
Braithwaite, J. M., Morris, R. W., & Economides, D. L. (1996) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.1996.tb09629.x · Cited by 70
Lin, X., Blumhardt, L. D., & Constantinescu, C. S. (2003) · Acta Neurologica Scandinavica · doi:10.1034/j.1600-0404.2003.00160.x · Cited by 79
Laser scanning for the identification of repeatable landmarks of the ears and face
Coward, T. J., Watson, R. M., & Scott, B. J. J. (1997) · British Journal of Plastic Surgery · doi:10.1016/s0007-1226(97)90538-5 · Cited by 70
Angeloni, S., Nicolini, G., Merli, M., Nicolao, F., Pinto, G., Aronne, T., Attili, A. F., & Riggio, O. (2003) · The American Journal of Gastroenterology · doi:10.1111/j.1572-0241.2003.07620.x · Cited by 70
Variables influencing Platelet Function Analyzer‐100TM closure times in healthy individuals
Haubelt, H., Anders, C., Vogt, A., Hoerdt, P., Seyfert, U. T., & Hellstern, P. (2005) · British Journal of Haematology · doi:10.1111/j.1365-2141.2005.05680.x · Cited by 70
Wuyts, B., Bernard, D. R., Noortgate, N. J. V. D., Walle, J. V., Vlem, B. V., Smet, R. D., Geeter, F. D., Vanholder, R. C., & Delanghe, J. (2003) · Clinical Chemistry · doi:10.1373/49.6.1011 · Cited by 70
The Relation Between Visual Acuity and Other Spatial Vision Measures
Hægerström-Portnoy, G., Schneck, M. E., Lott, L. A., & BRABYN, A. J. A. (2000) · Optometry and Vision Science · doi:10.1002/j.1538-9235.2000.tb01158.x · Cited by 83
Energy expenditure and severity of injury and illness indices in multiple trauma patients
Brandi, L. S., Santini, L., Bertolini, R., Malacarne, P., Casagli, S., & Baraglia, A. M. (1999) · Critical Care Medicine · doi:10.1097/00003246-199912000-00013 · Cited by 71
A Rapid, Specific Membrane Filtration Procedure for Enumeration of Enterococci in Recreational Water
Messer, J. W., & Dufour, A. P. (1998) · Applied and Environmental Microbiology · doi:10.1128/aem.64.2.678-680.1998 · Cited by 70
Barnes, E., Dutka, D. P., Khan, M., Camici, P. G., & Hall, R. J. C. (2002) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00786.2001 · Cited by 70
Kozàkovà, M., Galetta, F., Gregorini, L., Bigalli, G., Franzoni, F., Giusti, C., & Palombo, C. (2000) · Hypertension · doi:10.1161/01.hyp.36.3.343 · Cited by 70
Pini, R., Giannazzo, G., Bari, M. D., Innocenti, F., Rega, L., Casolo, G., & Devereux, R. B. (1997) · American Heart Journal · doi:10.1016/s0002-8703(97)70212-6 · Cited by 70
A new use for an old Holter monitor: an ambulatory cough meter
Chang, A. B., Newman, R., Phelan, P. D., & Robertson, C. (1997) · European Respiratory Journal · doi:10.1183/09031936.97.10071637 · Cited by 70
BRAY, M. S., WONG, W. W., MORROW, J. R., BUTTE, N. F., & PIVARNIK, J. M. (1994) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-199412000-00017 · Cited by 70
Bellenger, N. G., MARCUS, N. J., Davies, C., Yacoub, M. H., Banner, N. R., & Pennell, D. J. (2000) · The Journal of Heart and Lung Transplantation · doi:10.1016/s1053-2498(00)00079-6 · Cited by 75
Relationship of gated SPECT ventricular function parameters to angiographic measurements
Nichols, K., Tamis, J., DePuey, E. G., Mieres, J. H., Malhotra, S., & Rozanski, A. (1998) · Journal of Nuclear Cardiology · doi:10.1016/s1071-3581(98)90131-3 · Cited by 70
Determination of Right Ventricular Structure and Function in Normoxic and Hypoxic Mice
Scherrer‐Crosbie, M., Steudel, W., Hunziker, P., Foster, G. P., Garrido, L., Liel‐Cohen, N., Zapol, W. M., & Picard, M. H. (1998) · Circulation · doi:10.1161/01.cir.98.10.1015 · Cited by 70
Comparison of body composition and bone mineral measurements from two DXA instruments in young men
Modlesky, C., Lewis, R., Yetman, K. A., Rose, B. S., Rosskopf, L. B., Snow, T., & Sparling, P. B. (1996) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.5.669 · Cited by 74
BROWN, R., Wright, G. W. J., & Royston, D. (1993) · Anaesthesia · doi:10.1111/j.1365-2044.1993.tb07184.x · Cited by 73
Verdery, R. B., Benham, D. F., Baldwin, H. L., Goldberg, A. P., & Nichols, A. V. (1989) · Journal of Lipid Research · doi:10.1016/s0022-2275(20)38295-x · Cited by 70
Heinonen, P., Jousela, I. T., Blomqvist, K., Olkkola, K. T., & Takkunen, O. S. (1997) · Intensive Care Medicine · doi:10.1007/s001340050368 · Cited by 70
Vilagut, G. (2014) · Springer eBooks · doi:10.1007/978-94-007-0753-5_3001 · Cited by 84
Bordenave, S., Flavier, S., Fédou, C., Brun, J. F., & Mercier, J. (2007) · Diabetes & Metabolism · doi:10.1016/j.diabet.2007.04.003 · Cited by 69
Hohlfeld, J. M., Holland‐Letz, T., Larbig, M., Lavae‐Mokhtari, M., Wierenga, E., Kapsenberg, M., Van Ree, R., Krug, N., & Bufe, A. (2010) · Clinical & Experimental Allergy · doi:10.1111/j.1365-2222.2010.03498.x · Cited by 69
Celis-Morales, C., Livingstone, K. M., Woolhead, C., Forster, H., O’Donovan, C. B., Macready, A. L., Fallaize, R., Marsaux, C. F. M., Tsirigoti, L., Efstathopoulou, E., et al. (2015) · Genes & Nutrition · doi:10.1007/s12263-015-0476-0 · Cited by 69
Analysis of a Pulse Rate Variability Measurement Using a Smartphone Camera
Bánhalmi, A., Borbás, J., Fidrich, M., Bilicki, V., Gingl, Z., & Rudas, L. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/4038034 · Cited by 69
Weisstanner, C., Gratz, P. P., Schroth, G., Verma, R. K., Kochl, A., Jung, S., Arnold, M., Gralla, J., Zubler, C., Hsieh, K., et al. (2014) · European Radiology · doi:10.1007/s00330-014-3200-3 · Cited by 69
Rooney, A. G., McNamara, S., Mackinnon, M., Fraser, M., Rampling, R., Carson, A. J., & Grant, R. (2013) · Journal of Neuro-Oncology · doi:10.1007/s11060-013-1088-4 · Cited by 75
Artificial acrylic finger nails may alter pulse oximetry measurement
Hinkelbein, J., Koehler, H., Genzwuerker, H. V., & Fiedler, F. (2007) · Resuscitation · doi:10.1016/j.resuscitation.2006.11.018 · Cited by 69
van den Berg, E., Ruis, C., Biessels, G. J., Kappelle, L. J., & van Zandvoort, M. J. E. (2012) · Journal of Clinical and Experimental Neuropsychology · doi:10.1080/13803395.2012.667066 · Cited by 80
Chao, T.-F., Sung, S., Wang, K.-L., Lin, Y.-J., Chang, S., Lo, L.-W., Hu, Y., TUAN, T., Suenari, K., Li, C., et al. (2010) · Heart · doi:10.1136/hrt.2010.212373 · Cited by 69
Contrast-Based Fully Automatic Segmentation of White Matter Hyperintensities: Method and Validation
Samaille, T., Fillon, L., Cuingnet, R., Jouvent, É., Chabriat, H., Dormont, D., Colliot, O., & Chupin, M. (2012) · PLoS ONE · doi:10.1371/journal.pone.0048953 · Cited by 79
Menting, M. E., McGhie, J. S., Koopman, L. P., Vletter, W. B., Helbing, W. A., van den Bosch, A. E., & Roos‐Hesselink, J. W. (2016) · Echocardiography · doi:10.1111/echo.13323 · Cited by 76
O’Connell, R. L., Khabra, K., Bamber, J. C., deSouza, N. M., Meybodi, F., Barry, P. A., & Rusby, J. E. (2018) · Breast Cancer Research and Treatment · doi:10.1007/s10549-018-4843-6 · Cited by 69
Graybeal, A. J., Moore, M. L., Cruz, M. R., & Tinsley, G. M. (2020) · Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000002831 · Cited by 59
Mass transfer coefficient for water evaporation by theoretical and empirical correlations
Poós, T., & Varju, E. (2020) · International Journal of Heat and Mass Transfer · doi:10.1016/j.ijheatmasstransfer.2020.119500 · Cited by 69
Gustafson, L. W., Gabel, P., Hammer, A., Lauridsen, H. H., Petersen, L. K., Andersen, B. S., Bor, P., & Larsen, M. B. (2020) · BMC Medical Research Methodology · doi:10.1186/s12874-020-00982-4 · Cited by 69
Adaptive Checkpoint Adjoint Method for Gradient Estimation in Neural ODE
Zhuang, J., Dvornek, N. C., Li, X., Tatikonda, S., Papademetris, X., & Duncan, J. S. (2020) · PubMed · doi:10.48550/arxiv.2006.02493 · Cited by 158
Utembe, W., M.A., H. J. C. I., Sanabria, N. M., Doganis, P., & Gulumian, M. (2020) · Nanomaterials · doi:10.3390/nano10071267 · Cited by 78
Mailloux, C., Beaulieu, L., Wideman, T. H., & Massé‐Alarie, H. (2021) · PLoS ONE · doi:10.1371/journal.pone.0245278 · Cited by 86
Schiano‐Lomoriello, D., Hoffer, K. J., Abicca, I., & Savini, G. (2021) · Scientific Reports · doi:10.1038/s41598-020-79674-4 · Cited by 73
van Hoek, W. J., Wang, J., Vilmin, L., Beusen, A. H. W., Mogollón, J. M., Müller, G., Píka, P., Liu, X., Langeveld, J. J., Bouwman, L., et al. (2021) · Environmental Science & Technology · doi:10.1021/acs.est.1c04605 · Cited by 72
Kaptein, B. L., Pijls, B. G. C. W., Koster, L. A., Kärrholm, J., Hull, M. L., Niesen, A., Heesterbeek, P. J. C., Callary, S. A., Teeter, M. G., Gascoyne, T. C., et al. (2024) · Acta Orthopaedica · doi:10.2340/17453674.2024.40709 · Cited by 77
Fromentin, O., Lassauzay, C., Nader, S. A., Feine, J. S., & de Albuquerque, R. F. (2009) · Journal of Oral Rehabilitation · doi:10.1111/j.1365-2842.2009.02020.x · Cited by 69
Relationship between the pivot shift and the configuration of the lateral tibial plateau
Kujala, U. M., Nelimarkka, O., & Koskinen, S. K. (1992) · Archives of Orthopaedic and Trauma Surgery · doi:10.1007/bf00571483 · Cited by 69
Qi, Y., Soh, C. B., Gunawan, E., Low, K., & Thomas, R. (2015) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2015.2409123 · Cited by 69
Kampen, A. H., Tollersrud, T., Larsen, S. E., Roth, J. A., Frank, D. Ε., & Lund, A. (2003) · Veterinary Immunology and Immunopathology · doi:10.1016/j.vetimm.2003.08.018 · Cited by 84
Jacobson, P., Torgerson, J. S., Sjostrom, L., & Bouchard, C. (2006) · American Journal of Epidemiology · doi:10.1093/aje/kwj342 · Cited by 70
Psychomotor Development of Children from an Iodine-Deficient Region
Costeira, M. J., Oliveira, P., Santos, N. C., Segura, S. A., de Santiago, B. S.-R., de Escobar, G. M., & Palha, J. A. (2011) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2011.02.034 · Cited by 75
Wiens, A. D., Etemadi, M., Roy, S., Klein, L., & Inan, O. T. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2359937 · Cited by 69
Evaluation of image analysis methods used for quantification of particle angularity
Tafesse, S., Fernlund, J. M. R., Sun, W., & Bergholm, F. (2012) · Sedimentology · doi:10.1111/j.1365-3091.2012.01367.x · Cited by 73
Okubo, H., Miyake, Y., Tanaka, K., Sasaki, S., & Hirota, Y. (2015) · Nutrition Research · doi:10.1016/j.nutres.2015.02.009 · Cited by 82
Diagnosing Sensory Abnormalities with Either Normal Values or Values from Contralateral Skin
Kemler, M. A., Schouten, H. J. A., & Gracely, R. H. (2000) · Anesthesiology · doi:10.1097/00000542-200009000-00021 · Cited by 69
Within-Person Variability in High-Sensitivity C-Reactive Protein
Bower, J. K., Lazo, M., Juraschek, S. P., & Selvin, E. (2012) · Archives of Internal Medicine · doi:10.1001/archinternmed.2012.3712 · Cited by 72
Body composition in adolescent girls with Type 1 diabetes
Ingberg, C. M., Särnblad, S., Palmér, M., Schvarcz, E., Berne, C., & Åman, J. (2003) · Diabetic Medicine · doi:10.1046/j.1464-5491.2003.01055.x · Cited by 81
Lu, T.-W., Hsu, H., Chang, L. Y., & Chen, H. (2007) · Acta Dermato Venereologica · doi:10.2340/16501977-0107 · Cited by 69
Al-Qattan, M. M., Djordjevic, J., Zhurov, A. I., & Richmond, S. (2013) · European Journal of Orthodontics · doi:10.1093/ejo/cjt075 · Cited by 69
Fischbach, R., Juergens, K. U., Özgün, M., Maintz, D. C., Grude, M., Seifarth, H., Heindel, W. L., & Wichter, T. (2006) · European Radiology · doi:10.1007/s00330-006-0438-4 · Cited by 74
The repeatability and validity of dynamic retinoscopy in assessing the accommodative response
McClelland, J. F., & Saunders, K. J. (2003) · Ophthalmic and Physiological Optics · doi:10.1046/j.1475-1313.2003.00113.x · Cited by 85
Chappard, C., Basillais, A., Benhamou, L., Bonassie, A., Brunet-Imbault, B., Bonnet, N., & Peyrin, F. (2006) · Medical Physics · doi:10.1118/1.2256069 · Cited by 81
Kruidenier, L. M., Nicolaï, S. P., Willigendael, E. M., de Bie, R. A., Prins, M. H., & Teijink, J. A. W. (2009) · BMC Cardiovascular Disorders · doi:10.1186/1471-2261-9-9 · Cited by 69
Validity of accelerometry in assessing the duration of the sit-to-stand movement
Janssen, W. G. M., Bussmann, J. B. J., Horemans, H. L. D., & Stam, H. J. (2008) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-008-0366-3 · Cited by 78
Calibration and Validation of an Innovative Approach for Estimating General Cognitive Performance
Gross, A. L., Jones, R. N., Fong, T. G., Tommet, D., & Inouye, S. K. (2014) · Neuroepidemiology · doi:10.1159/000357647 · Cited by 71
Parker, M. V., O'Donnell, S. D., Chang, A. S., Johnson, C. A., Gillespie, D. L., Goff, J. M., Rasmussen, T. E., & Rich, N. M. (2005) · Journal of Vascular Surgery · doi:10.1016/j.jvs.2004.12.010 · Cited by 69
A study of inter-arm blood pressure differences in primary care
Cassidy, P., & Jones, K. (2001) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001224 · Cited by 69
Changes in gait patterns and muscle activity following total hip arthroplasty: A six-month follow-up
Horstmann, T., Listringhaus, R., Haase, G.-B., Grau, S., & Mündermann, A. (2013) · Clinical Biomechanics · doi:10.1016/j.clinbiomech.2013.07.001 · Cited by 71
Dried Blood Spot Sampling for Hepatitis B Virus Serology and Molecular Testing
Mohamed, S., Raimondo, A., Pénaranda, G., Camus, C., Ouzan, D., Ravet, S., Bourlière, M., Khiri, H., Dukan, P., Olive, D., et al. (2013) · PLoS ONE · doi:10.1371/journal.pone.0061077 · Cited by 69
Determination of stroke volume by means of electrical impedance tomography
Noordegraaf, A. V., Janse, A., Marcus, J. T., Bronzwaer, J. G. F., Postmus, P. E., Faes, T. J. C., & VRIES, P. M. J. M. D. (2000) · Physiological Measurement · doi:10.1088/0967-3334/21/2/308 · Cited by 69
Riche, M. A. (2006) · Aquaculture · doi:10.1016/j.aquaculture.2006.12.018 · Cited by 69
Victor, S., Appleton, R. E., Beirne, M., Marson, A. G., & Weindling, A. M. (2005) · Pediatric Research · doi:10.1203/01.pdr.0000169402.13435.09 · Cited by 69
Reference values of spirometry for F innish adults
Kainu, A., Timonen, K. L., Toikka, J., Qaiser, B., Pitkäniemi, J. M., Kotaniemi, J., Lindqvist, A. E., Vanninen, E. J., Länsimies, E., & Sovijärvi, A. R. A. (2015) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12237 · Cited by 88
Alfonso‐Rosa, R. M., Cruz, B. D. P., del Pozo-Cruz, J., Sañudo, B., & Rogers, M. E. (2013) · Rehabilitation Nursing · doi:10.1002/rnj.111 · Cited by 82
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.