STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 7,501–7,600 of 47,868
Uittenbogaard, L. B., Haak, M. C., Spreeuwenberg, M. D., & van Vugt, J. M. G. (2009) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.6287 · Cited by 70
Reliability of joint count assessment in rheumatoid arthritis: A systematic literature review
Cheung, P., Gossec, L., Mak, A., & MARCH, L. M. (2013) · Seminars in Arthritis and Rheumatism · doi:10.1016/j.semarthrit.2013.11.003 · Cited by 64
Plumb, A., Menys, A., Russo, E., Prezzi, D., Bhatnagar, G., Vega, R., Halligan, S., Orchard, T., & Taylor, S. A. (2015) · Alimentary Pharmacology & Therapeutics · doi:10.1111/apt.13275 · Cited by 65
Aerobic exercise capacity in patients with juvenile dermatomyositis.
Takken, T., Spermon, N., Helders, P. J. M., Prakken, A. B. J., & van der Net, J. J. (2003) · PubMed · Cited by 64
Nagar, G., Vandermeer, B., Campbell, S., & Kumar, M. (2016) · Neonatology · doi:10.1159/000442195 · Cited by 68
Wilenius, M., Tikkakoski, A., Tahvanainen, A., Haring, A., Koskela, J., Huhtala, H., Kähönen, M. A. P., Kööbi, T., Mustonen, J. T., & Pörsti, I. H. (2016) · BMC Cardiovascular Disorders · doi:10.1186/s12872-016-0303-6 · Cited by 75
van de Port, I., Wevers, L., & Kwakkel, G. (2011) · Journal of Rehabilitation Medicine · doi:10.2340/16501977-0881 · Cited by 69
Michelsen, M. M., Mygind, N. D., Peña, A., Olsen, R. H., Christensen, T. E., Ghotbi, A. A., Hasbak, P., Kjær, A., Gustafsson, I., Hansen, P. R., et al. (2016) · International Journal of Cardiology · doi:10.1016/j.ijcard.2016.11.004 · Cited by 70
García-Vaquero, M. P., Barbado, D., Juan‐Recio, C., López‐Valenciano, A., & Vera-García, F. J. (2016) · Journal of sport and health science/Journal of Sport and Health Science · doi:10.1016/j.jshs.2016.08.011 · Cited by 64
Self-gated fetal cardiac MRI with tiny golden angle iGRASP: A feasibility study
Haris, K., Hedström, E., Bidhult, S. L., Testud, F., Maglaveras, N., Heiberg, E., Hansson, S. R., Arheden, H., & Aletras, A. H. (2017) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25599 · Cited by 66
Timon, C., Blain, R. J., McNulty, B. A., Kehoe, L., Evans, K., Walton, J., Flynn, A., & Gibney, E. R. (2017) · Journal of Medical Internet Research · doi:10.2196/jmir.6407 · Cited by 74
Nothnagel, H., Puta, C., Lehmann, T., Baumbach, P., Menard, M. B., Gabriel, B., Gabriel, H. W., Weiß, T., & Musial, F. (2017) · Journal of Pain Research · doi:10.2147/jpr.s137391 · Cited by 75
Real-time human respiration carbon dioxide measurement device for cardiorespiratory assessment
Singh, O. P., Howe, T. A., & Malarvili, M. B. (2017) · Journal of Breath Research · doi:10.1088/1752-7163/aa8dbd · Cited by 66
Cardiac Output Monitoring: Validation Studies–how Results Should be Presented
Odor, P. M., Bampoe, S., & Cecconi, M. (2017) · Current anesthesiology reports · doi:10.1007/s40140-017-0239-0 · Cited by 72
Concurrent Validity of the Zeno Walkway for Measuring Spatiotemporal Gait Parameters in Older Adults
Vallabhajosula, S., Humphrey, S., Cook, A., & Freund, J. E. (2017) · Journal of Geriatric Physical Therapy · doi:10.1519/jpt.0000000000000168 · Cited by 72
Comparative Evaluation of Clinical Methods of Tear Film Stability Assessment
Wang, M. T. M., & Craig, J. P. (2018) · JAMA Ophthalmology · doi:10.1001/jamaophthalmol.2017.6489 · Cited by 64
Samušytė, G., Bostock, H., Rothwell, J. C., & Koltzenburg, M. (2018) · Brain stimulation · doi:10.1016/j.brs.2018.03.002 · Cited by 76
Using deep learning to generate synthetic B-mode musculoskeletal ultrasound images
Cronin, N. J., Finni, T. J., & Seynnes, O. (2020) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2020.105583 · Cited by 67
Agreement and repeatability of pupillometry using videokeratography and infrared devices
Wachler, B. S. B., & Krueger, R. R. (2000) · Journal of Cataract & Refractive Surgery · doi:10.1016/s0886-3350(99)00331-4 · Cited by 64
Wong, W. W., Stuff, J. E., Butte, N. F., Smith, E. O., & Ellis, K. J. (2000) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/72.2.348 · Cited by 64
Screening of folate status with use of dried blood spots on filter paper
O’Broin, S., & Gunter, E. W. (1999) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/70.3.359 · Cited by 68
Enhanced brain extraction improves the accuracy of brain atrophy estimation
Battaglini, M., Smith, S. M., Brogi, S., & Stefano, N. D. (2008) · NeuroImage · doi:10.1016/j.neuroimage.2007.10.067 · Cited by 70
Ultrasonography of the hip in developmental hip dysplasia
Bar‐On, E., Meyer, S., Harari, G., & Porat, S. (1998) · Journal of Bone and Joint Surgery - British Volume · doi:10.1302/0301-620x.80b2.0800321 · Cited by 64
Takala, E., Korhonen, I., & Viikari-Juntura, E. R. A. (1997) · Clinical Biomechanics · doi:10.1016/s0268-0033(97)00033-8 · Cited by 64
Sanaka, M., Kuyama, Y., & Yamanaka, M. (1998) · Journal of Gastroenterology · doi:10.1007/s005350050177 · Cited by 67
Airway Closure Measured by a Technegas Bolus and SPECT
King, G. G., Eberl, S., Salome, C. M., Meikle, S. R., & Woolcock, A. J. (1997) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.155.2.9032213 · Cited by 64
How reliable are ultrasound measurements of renal length in adults?
Ablett, M. J., Coulthard, A., Lee, R. E. J., Richardson, D. L., Bellas, T., Owen, J. P., Keir, M. J., & Butler, T. J. (1995) · British Journal of Radiology · doi:10.1259/0007-1285-68-814-1087 · Cited by 64
Acute phase markers for the differentiation of infectious and malignant pleural effusions
Kiropoulos, T. S., Κostikas, Κ., Oikonomidi, S., Tsilioni, I., Nikoulis, D. J., Germenis, A. E., & Gourgoulianis, K. I. (2007) · Respiratory Medicine · doi:10.1016/j.rmed.2006.09.019 · Cited by 64
Agreement and repeatability of infrared pupillometry and the comparison method
Wachler, B. S. B., & Krueger, R. R. (1999) · Ophthalmology · doi:10.1016/s0161-6420(99)90070-2 · Cited by 64
Craig, P. L., Halavatau, V., Comino, E. J., & Caterson, I. D. (2001) · International Journal of Obesity · doi:10.1038/sj.ijo.0801822 · Cited by 73
CONTINUOUS INTRA-ABDOMINAL PRESSURE MONITORING
Balogh, Z. J., Waele, J. J. D., & Malbrain, M. L. N. G. (2007) · Acta Clinica Belgica · doi:10.1179/acb.2007.62.s1.005 · Cited by 64
Postprandial lipemia: effects of intermittent versus continuous exercise
GILL, J. M. R., MURPHY, M. H., & HARDMAN, A. E. (1998) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-199810000-00008 · Cited by 65
Bailey, M. J., Riddoch, M. J., & Crome, P. (2004) · Neuropsychological Rehabilitation · doi:10.1080/09602010343000282 · Cited by 64
Scalfaro, P., Pillow, J. J., Sly, P. D., & Cotting, J. (2001) · Critical Care Medicine · doi:10.1097/00003246-200110000-00013 · Cited by 64
Montgomery, C., Reilly, J. J., Jackson, D. M., Kelly, L. A., Slater, C., Paton, J. Y., & Grant, S. (2005) · British Journal Of Nutrition · doi:10.1079/bjn20051405 · Cited by 79
Lanigan, J., Wells, J. C. K., Lawson, M. S., & Lucas, A. (2001) · European Journal of Clinical Nutrition · doi:10.1038/sj.ejcn.1601128 · Cited by 74
Pulse Contour Analysis for Cardiac Output Monitoring in Cardiac Surgery for Congenital Heart Disease
Mahajan, A., Shabanie, A., Turner, J., Sopher, M., & Marijic, J. (2003) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000081797.61469.12 · Cited by 64
Tseng, W. I., Liao, T.-Y., & Wang, J. (2002) · Journal of Cardiovascular Magnetic Resonance · doi:10.1081/jcmr-120016383 · Cited by 68
Bellmann, C., Jorzik, J., Spital, G., Unnebrink, K., Pauleikhoff, D., & Holz, F. G. (2002) · Archives of Ophthalmology · doi:10.1001/archopht.120.5.579 · Cited by 71
Nanas, S., Nanas, J., Kassiotis, C., Nikolaou, C., Tsagalou, E., Sakellariou, D., Terovitis, I., Papazachou, O., Drakos, S., Papamichalopoulos, A., et al. (2001) · European Journal of Heart Failure · doi:10.1016/s1388-9842(01)00187-8 · Cited by 64
Chee, W. S. S., Suriah, A. R., Zaitun, Y., Chan, S., Yap, S. L., & Chan, Y. (2002) · Asia Pacific Journal of Clinical Nutrition · doi:10.1046/j.1440-6047.2002.00276.x · Cited by 64
Medial plantar nerve conduction studies in healthy controls and diabetics
Løseth, S., Nebuchennykh, M., Stålberg, E., & Mellgren, S. I. (2007) · Clinical Neurophysiology · doi:10.1016/j.clinph.2007.01.008 · Cited by 64
Diagnostic accuracy of computer-assisted sperm motion analysis
Geyter, C. D., Geyter, M. D., Koppers, B., & Nieschlag, E. (1998) · Human Reproduction · doi:10.1093/humrep/13.9.2512 · Cited by 64
Kinematic MRI Assessment of McConnell Taping before and after Exercise
Pfeiffer, R. P., DeBeliso, M., Shea, K. G., Kelley, L. L., Irmischer, B. S., & Harris, C. (2004) · The American Journal of Sports Medicine · doi:10.1177/0363546503261693 · Cited by 64
Statistical analyses in the physiology of exercise and kinanthropometry
Winter, E., Eston, R. G., & Lamb, K. (2001) · Journal of Sports Sciences · doi:10.1080/026404101317015429 · Cited by 74
Branson, K., WAGNER‐MANN, C. C., & Mann, F. A. (1997) · Veterinary Surgery · doi:10.1111/j.1532-950x.1997.tb01510.x · Cited by 71
Zucconi, M., Calori, G., Castronovo, V., & Ferini‐Strambi, L. (2003) · CHEST Journal · doi:10.1378/chest.124.2.602 · Cited by 64
Estimation of pupil size by digital photography
Twa, M. D., Bailey, M. D., Hayes, J., & Bullimore, M. A. (2004) · Journal of Cataract & Refractive Surgery · doi:10.1016/s0886-3350(03)00619-9 · Cited by 64
Eron, J. J., Smeaton, L. M., Fiscus, S. A., Gulick, R. M., Currier, J. S., Lennox, J. L., D’Aquila, R. T., Rogers, M. D., Tung, R. D., & Murphy, R. L. (2000) · The Journal of Infectious Diseases · doi:10.1086/315447 · Cited by 64
Point-of-Care Testing of Hemostasis in Cardiac Surgery
Prisco, D., & Paniccia, R. (2003) · Thrombosis Journal · doi:10.1186/1477-9560-1-1 · Cited by 75
Alfakih, K., Thiele, H., Plein, S., Bainbridge, G. J., Ridgway, J. P., & Sivananthan, M. U. (2002) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.10164 · Cited by 69
Functional Rating for Knee Arthroplasty: Comparison of Three Scoring Systems
Liow, R., Walker, K. A., Wajid, M. A., Bedi, G., & Lennox, C. M. E. (2003) · Orthopedics · doi:10.3928/0147-7447-20030201-15 · Cited by 64
Freezer, N., Croasdell, H., Doull, I. J. M., & Holgate, S. (1995) · European Respiratory Journal · doi:10.1183/09031936.95.08091488 · Cited by 65
Ellis, K. J., Shypailo, R. J., Hergenroeder, A. C., Pérez, M. T., & Abrams, S. A. (1996) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.5650110616 · Cited by 64
Cognitive correlates of supratentorial atrophy on MRI in multiple sclerosis
Edwards, S. G. M., Liu, C., & Blumhardt, L. D. (2001) · Acta Neurologica Scandinavica · doi:10.1034/j.1600-0404.2001.00270.x · Cited by 79
The effect of fetal neck position on nuchal translucency measurement
Whitlow, B., Chatzipapas, I., & Economides, D. L. (1998) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.1998.tb10232.x · Cited by 64
Mustonen, J. N., Uusitupa, M. I. J., Tahvanainen, K. U. O., Talwar, S., Laakso, M., Länsimies, E., Kuikka, J. T., & Pyörälà, K. (1988) · The American Journal of Cardiology · doi:10.1016/0002-9149(88)90273-1 · Cited by 67
Continuous noninvasive cardiac output as estimated from the pulse contour curve
Gratz, I., Kraidin, J. L., Jacobi, A. G. M., deCastro, N. G., Spagna, P., & Larijani, G. E. (1992) · Journal of Clinical Monitoring and Computing · doi:10.1007/bf01618083 · Cited by 64
Polosa, R., Phillips, G., Rajakulasingam, K., & Holgate, S. T. (1991) · Journal of Allergy and Clinical Immunology · doi:10.1016/0091-6749(91)90415-k · Cited by 64
Lindower, P. D., Rath, L., Preslar, J. S., Burns, T. L., Rezai, K., & Vandenberg, B. F. (1994) · The American Journal of Cardiology · doi:10.1016/0002-9149(94)90214-3 · Cited by 64
Shi, K., Schellenberger, S., Will, C., Steigleder, T., Michler, F., Fuchs, J., Weigel, R., Ostgathe, C., & Koelpin, A. (2020) · Scientific Data · doi:10.1038/s41597-020-0390-1 · Cited by 67
Italian validation of INQoL, a quality of life questionnaire for adults with muscle diseases
Sansone, V. A., Panzeri, M., Montanari, M., Apolone, G., Gandossini, S., Rose, M. R., Politano, L., Solimene, C., Siciliano, G., Volpi, L., et al. (2010) · European Journal of Neurology · doi:10.1111/j.1468-1331.2010.02992.x · Cited by 64
Pleus, S., Baumstark, A., Jendrike, N., Mende, J., Link, M., Zschornack, E., Haug, C., & Freckmann, G. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-001067 · Cited by 63
Nussbaumer, Y., Bloch, K. E., Genser, T., & Thurnheer, R. (2006) · Chest · doi:10.1378/chest.129.3.638 · Cited by 63
ŠRÁMEK, A., BOSCH, J. G., REIBER, J. H. C., VAN OOSTAYEN, J. A., & ROSENDAAL, F. R. (2000) · Investigative Radiology · doi:10.1097/00004424-200012000-00001 · Cited by 63
Castelletti, S., Dagradi, F., Goulene, K., Danza, A. I., Baldi, E., Stramba-Badiale, M., & Schwartz, P. J. (2018) · International Journal of Cardiology · doi:10.1016/j.ijcard.2018.03.097 · Cited by 63
Fontana, F. Y., Keir, D. A., Bellotti, C., De Roia, G. F., Murias, J. M., & Pogliaghi, S. (2015) · Journal of Science and Medicine in Sport · doi:10.1016/j.jsams.2014.07.016 · Cited by 63
Baseline quality of life before intensive care: A comparison of patient versus proxy responses
Gifford, J. M., Husain, N., Dinglas, V. D., Colantuoni, E., & Needham, D. M. (2010) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181cd10c7 · Cited by 72
van Til, J. A., Groothuis‐Oudshoorn, C. G. M., Lieferink, M., Dolan, J. G., & Goetghebeur, M. M. (2014) · Cost Effectiveness and Resource Allocation · doi:10.1186/1478-7547-12-22 · Cited by 78
Blohm, M. E., Obrecht, D., Hartwich, J., Mueller, G. C., Kersten, J. F., Weil, J., & Singer, D. (2014) · Critical Care · doi:10.1186/s13054-014-0603-0 · Cited by 63
Veqar, Z., & Hussain, M. E. (2017) · International Journal of Adolescent Medicine and Health · doi:10.1515/ijamh-2016-0090 · Cited by 73
Validity of the EQ-5D-5L and EQ-5D-3L in patients with Crohn’s disease
Rencz, F., Lakatos, P. L., Gulàcsi, L., Brodszky, V., Kürti, Z., Lovas, S., Banai, J., Herszényi, L., Cserni, T., Molnár, T., et al. (2018) · Quality of Life Research · doi:10.1007/s11136-018-2003-4 · Cited by 69
Castaldo, R., Pane, K., Nicolai, E., Salvatore, M., & Franzese, M. (2020) · Cancers · doi:10.3390/cancers12020518 · Cited by 72
Balakrishnan, R., Valdés Hernández, M. D. C., & Farrall, A. J. (2021) · Computerized Medical Imaging and Graphics · doi:10.1016/j.compmedimag.2021.101867 · Cited by 63
Oefner, C., Herrmann, S., Kebbach, M., Lange, H.-E., Kluess, D., & Woiczinski, M. (2021) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2021.03.011 · Cited by 71
Takamura, M., Kobayashi, T., Nikkuni, Y., Katsura, K., Yamazaki, M., Maruyama, S., Tanuma, J., & Hayashi, T. (2021) · Oral Radiology · doi:10.1007/s11282-021-00533-7 · Cited by 63
Takeuchi, T., Kawanishi, M., Nakanishi, M., Yamasaki, H., & Tanaka, Y. (2022) · Arthritis & Rheumatology · doi:10.1002/art.42273 · Cited by 65
Lahkar, B. K., Muller, A., Dumas, R., Revéret, L., & Robert, T. (2022) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2022.939980 · Cited by 82
van Hulst, K., Lindeboom, R., van der Burg, J. J. W., & Jongerius, P. H. (2012) · Developmental Medicine & Child Neurology · doi:10.1111/j.1469-8749.2012.04428.x · Cited by 63
On stability issues in deriving multivariable regression models
Sauerbrei, W., Buchholz, A., Boulesteix, A., & Binder, H. (2014) · Biometrical Journal · doi:10.1002/bimj.201300222 · Cited by 72
Jia, X., Craig, L. C. A., Aucott, L. S., Milne, A. C., & Mcneill, G. (2008) · The Journal of nutrition, health and aging · doi:10.1007/bf03028622 · Cited by 65
Needleman, I. G., Hirsch, N. P., Leemans, M., Moles, D. R., Wilson, M., Ready, D. R., Ismail, S., Ciric, L., Shaw, M. J., Smith, M., et al. (2011) · Journal Of Clinical Periodontology · doi:10.1111/j.1600-051x.2010.01688.x · Cited by 77
Freehand three-dimensional ultrasound system for assessment of scoliosis
Cheung, J. C., Zhou, G., Law, S.-Y., Lai, K.-L., Jiang, W., & Zheng, Y. (2015) · Journal of Orthopaedic Translation · doi:10.1016/j.jot.2015.06.001 · Cited by 67
Quantifying image distortion of orthopedic materials in magnetic resonance imaging
Koff, M. F., Shah, P. H., Koch, K. M., & Potter, H. G. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.23991 · Cited by 64
Test-Retest Reliability of the 6-Minute Walk Test in Individuals With Down Syndrome
Casey, A., Wang, X., & Osterling, K. (2012) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2012.04.022 · Cited by 76
Furlanetto, T. S., Candotti, C. T., Comerlato, T., & Loss, J. F. (2012) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2012.03.012 · Cited by 63
The Repeatability of Discrete and Continuous Anterior Segment Grading Scales
Chong, T., SIMPSON, T. L., & Fonn, D. (2000) · Optometry and Vision Science · doi:10.1002/j.1538-9235.2000.tb01087.x · Cited by 63
Performance of Automated Drusen Detection by Polarization-Sensitive Optical Coherence Tomography
Schlanitz, F. G., Baumann, B., Spalek, T., Schütze, C., Ahlers, C., Pircher, M., Götzinger, E., Hitzenberger, C. K. F., & Schmidt-Erfurth, U. M. (2011) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.10-6846 · Cited by 64
Development and Psychometric Properties of the Ambulatory Self-Confidence Questionnaire
Asano, M., Miller, W. C., & Eng, J. J. (2007) · Gerontology · doi:10.1159/000104830 · Cited by 70
Frankenfield, D. C., Sarson, G. Y., Blosser, S. A., Cooney, R. N., & Smith, J. S. (1996) · Journal of the American College of Nutrition · doi:10.1080/07315724.1996.10718615 · Cited by 64
The influence of cycloplegia in objective refraction
Jorge, J., Queirós, A., González‐Méijome, J. M., Fernandes, P. R. B., Almeida, J. B., & Parafita, M. Á. (2005) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2005.00277.x · Cited by 72
A Simple Method for Obtaining Original Data from Published Graphs and Plots
Sistrom, C. L., & Mergo, P. J. (2000) · American Journal of Roentgenology · doi:10.2214/ajr.174.5.1741241 · Cited by 65
Measures of Reliability in the Kinematics of Maximal Undulatory Underwater Swimming
Connaboy, C., Coleman, S. G. S., Moir, G. L., & Sanders, R. H. (2010) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e3181badc68 · Cited by 63
Majida, M., BrÆkken, I. H., Bø, K., Šaltytė Benth, J., & Engh, M. E. (2010) · Ultrasound in Obstetrics & Gynecology · doi:10.1002/uog.7587 · Cited by 63
Egle, D., Strobl, I., Weiskopf-Schwendinger, V., Grubinger, E., Kraxner, F., Mutz‐Dehbalaie, I., Strasak, A. M., & Scheier, M. (2011) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.8957 · Cited by 64
Measurement Error in Estimates of Sprint Velocity from a Laser Displacement Measurement Device
Bezodis, N. E., Salo, A., & Trewartha, G. (2012) · International Journal of Sports Medicine · doi:10.1055/s-0031-1301313 · Cited by 76
Bastarrika, G., Zudaire, B., Ferreira, M. M., Arraiza, M., Saiz-Mendiguren, R., & Rábago, G. (2010) · Investigative Radiology · doi:10.1097/rli.0b013e3181c4f535 · Cited by 65
Blood pressure estimation from pulse wave velocity measured on the chest
Puke, S., Suzuki, T., Nakayama, K., Tanaka, H., & Minami, S. (2013) · 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6610946 · Cited by 60
Clinical validation of the AMTAS automated audiometer
Eikelboom, R. H., Swanepoel, D. W., Motakef, S., & Upson, G. S. (2013) · International Journal of Audiology · doi:10.3109/14992027.2013.769065 · Cited by 63
Methodological issues in the assessment of satiety
Livingstone, M. B. E., Robson, P. J., Welch, R., Burns, A. A., Burrows, M. S., & McCormack, C. (2000) · Näringsforskning · doi:10.3402/fnr.v44i0.1776 · Cited by 77
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.