Point-of-care screening for heart failure with reduced ejection fraction using artificial intelligence during ECG-enabled stethoscope examination in London, UK: a prospective, observational, multicentre study
Open accessAbstract
BACKGROUND: Most patients who have heart failure with a reduced ejection fraction, when left ventricular ejection fraction (LVEF) is 40% or lower, are diagnosed in hospital. This is despite previous presentations to primary care with symptoms. We aimed to test an artificial intelligence (AI) algorithm applied to a single-lead ECG, recorded during ECG-enabled stethoscope examination, to validate a potential point-of-care screening tool for LVEF of 40% or lower. METHODS: We conducted an observational, prospective, multicentre study of a convolutional neural network (known as AI-ECG) that was previously validated for the detection of reduced LVEF using 12-lead ECG as input. We used AI-ECG retrained to interpret single-lead ECG input alone. Patients (aged ≥18 years) attending for transthoracic echocardiogram in London (UK) were recruited. All participants had 15 s of supine, single-lead ECG recorded at the four standard anatomical positions for cardiac auscultation, plus one handheld position, using an ECG-enabled stethoscope. Transthoracic echocardiogram-derived percentage LVEF was used as ground truth. The primary outcome was performance of AI-ECG at classifying reduced LVEF (LVEF ≤40%), measured using metrics including the area under the receiver operating characteristic curve (AUROC), sensitivity, and specificity, with two-sided 95% CIs. The primary outcome was reported for each position individually and with an optimal combination of AI-ECG outputs (interval range 0-1) from two positions using a rule-based approach and several classification models. This study is registered with ClinicalTrials.gov, NCT04601415. FINDINGS: Between Feb 6 and May 27, 2021, we recruited 1050 patients (mean age 62 years [SD 17·4], 535 [51%] male, 432 [41%] non-White). 945 (90%) had an ejection fraction of at least 40%, and 105 (10%) had an ejection fraction of 40% or lower. Across all positions, ECGs were most frequently of adequate quality for AI-ECG interpretation at the pulmonary position (979 [93·3%] of 1050). Quality was lowest for the aortic position (846 [80·6%]). AI-ECG performed best at the pulmonary valve position (p=0·02), with an AUROC of 0·85 (95% CI 0·81-0·89), sensitivity of 84·8% (76·2-91·3), and specificity of 69·5% (66·4-72·6). Diagnostic odds ratios did not differ by age, sex, or non-White ethnicity. Taking the optimal combination of two positions (pulmonary and handheld positions), the rule-based approach resulted in an AUROC of 0·85 (0·81-0·89), sensitivity of 82·7% (72·7-90·2), and specificity of 79·9% (77·0-82·6). Using AI-ECG outputs from these two positions, a weighted logistic regression with l2 regularisation resulted in an AUROC of 0·91 (0·88-0·95), sensitivity of 91·9% (78·1-98·3), and specificity of 80·2% (75·5-84·3). INTERPRETATION: A deep learning system applied to single-lead ECGs acquired during a routine examination with an ECG-enabled stethoscope can detect LVEF of 40% or lower. These findings highlight the potential for inexpensive, non-invasive, workflow-adapted, point-of-care screening, for earlier diagnosis and prognostically beneficial treatment. FUNDING: NHS Accelerated Access Collaborative, NHSX, and the National Institute for Health Research.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Figures




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Models trained on clinical ECGs have been applied to wearable device ECGs, but have traditionally shown significantly lower performance on wearable device ECGs than on held-out clinical ECG test sets14,15.”
From Detection of left ventricular systolic dysfunction from single-lead electrocardiography adapted for portable and wearable devices (2023) · cites this paper as [14]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Consequently, the performance of models for detecting LVSD from portable device ECGs may degrade in the real-world setting, with lower performance than observed in the original single-lead derivatives of the clinical development studies14,15.”
From Detection of left ventricular systolic dysfunction from single-lead electrocardiography adapted for portable and wearable devices (2023) · cites this paper as [14]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Separately, a prospective multicenter study showed that a deep-learning model applied to single-lead ECGs obtained during examination with an ECG-enabled stethoscope could identify patients with a left ventricular ejection fraction of 40% or lower [64].”
From Deep Learning Applied to 12-Lead ECGs for Detection of Structural, Metabolic and Systemic Disease (2026) · cites this paper as [64]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
DeLong, E. R., DeLong, D. M., & Clarke-Pearson, D. L. (1988) · Biometrics · doi:10.2307/2531595 · Cited by 22,649
High-performance medicine: the convergence of human and artificial intelligence
Topol, E. J. (2019) · Nature Medicine · doi:10.1038/s41591-018-0300-7 · Cited by 10,450
Forecasting the Impact of Heart Failure in the United States
Heidenreich, P. A., Albert, N. M., Allen, L. A., Bluemke, D. A., Butler, J., Fonarow, G. C., Ikonomidis, J. S., Khavjou, O., Konstam, M. A., Maddox, T. M., et al. (2013) · Circulation Heart Failure · doi:10.1161/hhf.0b013e318291329a · Cited by 2,884
Global Public Health Burden of Heart Failure
Savarese, G., & Lund, L. H. (2017) · Cardiac failure review · doi:10.15420/cfr.2016:25:2 · Cited by 2,818
Angiotensin–Neprilysin Inhibition in Heart Failure with Preserved Ejection Fraction
Solomon, S. D., McMurray, J. J. V., Anand, I. S., Ge, J., Lam, C. S. P., Maggioni, A. P., Martinez, F., Packer, M., Pfeffer, M. A., Pieske, B., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1908655 · Cited by 2,430
The Global Health and Economic Burden of Hospitalizations for Heart Failure
Ambrosy, A. P., Fonarow, G. C., Butler, J., Chioncel, O. D., Greene, S. J., Vaduganathan, M., Nodari, S., Lam, C. S. P., Sato, N., Shah, A. N., et al. (2014) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2013.11.053 · Cited by 2,110
Bozkurt, B., Coats, A. J. S., Tsutsui, H., Abdelhamid, C. M., Adamopoulos, S., Albert, N., Anker, S. D., Atherton, J., Böhm, M., Butler, J., et al. (2021) · European Journal of Heart Failure · doi:10.1002/ejhf.2115 · Cited by 1,665
Attia, Z. I., Noseworthy, P. A., Lopez-Jimenez, F., Asirvatham, S. J., Deshmukh, A. J., Gersh, B. J., Carter, R. E., Yao, X., Rabinstein, A. A., Erickson, B. J., et al. (2019) · The Lancet · doi:10.1016/s0140-6736(19)31721-0 · Cited by 1,613
Attia, Z. I., Kapa, S., Lopez-Jimenez, F., McKie, P. M., Ladewig, D. J., Satam, G., Pellikka, P. A., Enriquez-Sarano, M., Noseworthy, P. A., Munger, T. M., et al. (2019) · Nature Medicine · doi:10.1038/s41591-018-0240-2 · Cited by 1,389
Nagendran, M., Chen, Y., Lovejoy, C. A., Gordon, A. C., Komorowski, M., Harvey, H., Topol, E. J., Ioannidis, J. P. A., Collins, G. S., & Maruthappu, M. (2020) · BMJ · doi:10.1136/bmj.m689 · Cited by 1,152
Artificial intelligence and the future of global health
Schwalbe, N., & Wahl, B. (2020) · The Lancet · doi:10.1016/s0140-6736(20)30226-9 · Cited by 956
Wahl, B., Cossy-Gantner, A., Germann, S., & Schwalbe, N. R. (2018) · BMJ Global Health · doi:10.1136/bmjgh-2018-000798 · Cited by 796
Yao, X., Rushlow, D. R., Inselman, J. W., McCoy, R. G., Thacher, T. D., Behnken, E. M., Bernard, M. E., Rosas, S. L., Akfaly, A., Misra, A., et al. (2021) · Nature Medicine · doi:10.1038/s41591-021-01335-4 · Cited by 424
Heart failure with reduced ejection fraction
Bloom, M. W., Greenberg, B., Jaarsma, T., Januzzi, J. L., Lam, C. S. P., Maggioni, A. P., Trochu, J., & Butler, J. (2017) · Nature Reviews Disease Primers · doi:10.1038/nrdp.2017.58 · Cited by 239
Burnett, H., Earley, A., Voors, A. A., Senni, M., McMurray, J. J. V., Deschaseaux, C., & Cope, S. (2017) · Circulation Heart Failure · doi:10.1161/circheartfailure.116.003529 · Cited by 238
Rim, T. H., Lee, C. J., Tham, Y.-C., Cheung, N., Yu, M., Lee, G., Kim, Y., Ting, D. S. W., Chong, C. C. Y., Choi, Y. S., et al. (2021) · The Lancet Digital Health · doi:10.1016/s2589-7500(21)00043-1 · Cited by 234
Regulation of predictive analytics in medicine
Parikh, R. B., Obermeyer, Z., & Navathe, A. S. (2019) · Science · doi:10.1126/science.aaw0029 · Cited by 223
Attia, Z. I., Kapa, S., Yao, X., López-Jiménez, F. A., Mohan, T. L., Pellikka, P. A., Carter, R. E., Shah, N. D., Friedman, P. A., & Noseworthy, P. A. (2019) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.13889 · Cited by 199
Computer-aided diagnosis of atrial fibrillation based on ECG Signals: A review
Hagiwara, Y., Fujita, H., Oh, S. L., Tan, J. H., Tan, R., Ciaccio, E. J., & Acharya, U. R. (2018) · Information Sciences · doi:10.1016/j.ins.2018.07.063 · Cited by 197
Adedinsewo, D., Carter, R. E., Attia, Z., Johnson, P., Kashou, A. H., Dugan, J. L., Albus, M., Sheele, J. M., Bellolio, F., Friedman, P. A., et al. (2020) · Circulation: Arrhythmia and Electrophysiology · doi:10.1161/circep.120.008437 · Cited by 169
Deep Learning Algorithm for Automated Cardiac Murmur Detection via a Digital Stethoscope Platform
Chorba, J. S., Shapiro, A. M., Le, L., Maidens, J., Prince, J., Pham, S., Kanzawa, M. M., Barbosa, D. N., Currie, C., Brooks, C., et al. (2021) · Journal of the American Heart Association · doi:10.1161/jaha.120.019905 · Cited by 149
Routes to diagnosis of heart failure: observational study using linked data in England
Bottle, A., Kim, D. J., Aylin, P. P., Cowie, M., Majeed, A., & Hayhoe, B. W. J. (2017) · Heart · doi:10.1136/heartjnl-2017-312183 · Cited by 141
Early detection of myocardial dysfunction and heart failure
de Couto, G., Ouzounian, M., & Liu, P. P. (2010) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2010.51 · Cited by 119
External validation of a deep learning electrocardiogram algorithm to detect ventricular dysfunction
Attia, I. Z., Tseng, A. S., Benavente, E. D., Medina‐Inojosa, J. R., Clark, T. G., Malyutina, S. K., Kapa, S., Schirmer, H., Kudryavtsev, A. V., Noseworthy, P. A., et al. (2021) · International Journal of Cardiology · doi:10.1016/j.ijcard.2020.12.065 · Cited by 80
O’Neal, W. T., Mazur, M. M., Bertoni, A. G., Bluemke, D. A., Al-Mallah, M. H., Lima, J. A. C., Kitzman, D. W., & Soliman, E. Z. (2017) · Journal of the American Heart Association · doi:10.1161/jaha.117.006023 · Cited by 76
The heart failure epidemic: a UK perspective
Cowie, M. (2017) · Echo Research and Practice · doi:10.1530/erp-16-0043 · Cited by 67
An awakening in medicine: the partnership of humanity and intelligent machines
Celi, L. A., Fine, B., & Stone, D. J. (2019) · The Lancet Digital Health · doi:10.1016/s2589-7500(19)30127-x · Cited by 45
Artificial Intelligence ECG to Detect Left Ventricular Dysfunction in COVID-19
Attia, Z. I., Kapa, S., Noseworthy, P. A., López-Jiménez, F. A., & Friedman, P. A. (2020) · Mayo Clinic Proceedings · doi:10.1016/j.mayocp.2020.09.020 · Cited by 39
Orchard, J. J., Li, J., Gallagher, R. D., Freedman, B., Lowres, N., & Neubeck, L. (2019) · BMC Family Practice · doi:10.1186/s12875-019-1058-9 · Cited by 36
Bagaswoto, H. P., Krisdinarti, L., & Erika, M. (2016) · Cardiology Research · doi:10.14740/cr473w · Cited by 25
Alhamaydeh, M., Gregg, R. E., Ahmad, A., Faramand, Z., Saba, S., & Al‐Zaiti, S. S. (2020) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2020.06.003 · Cited by 17
Cited by
Sorted by citations
Armoundas, A. A., Narayan, S. M., Arnett, D. K., Spector-Bagdady, K., Bennett, D. A., Celi, L. A., Friedman, P. A., Gollob, M. H., Hall, J. L., Kwitek, A. E., et al. (2024) · Circulation · doi:10.1161/cir.0000000000001201 · Cited by 347
Artificial intelligence in clinical medicine: catalyzing a sustainable global healthcare paradigm
Krishnan, G., Singh, S., Pathania, M., Gosavi, S., Abhishek, S., Parchani, A., & Dhar, M. (2023) · Frontiers in Artificial Intelligence · doi:10.3389/frai.2023.1227091 · Cited by 339
Transforming Cardiovascular Care With Artificial Intelligence: From Discovery to Practice
Khera, R., Oikonomou, E. K., Nadkarni, G. N., Morley, J., Wiens, J., Butte, A. J., & Topol, E. J. (2024) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2024.05.003 · Cited by 324
The Integration of Artificial Intelligence into Clinical Practice
Karalis, V. D. (2024) · Applied Biosciences · doi:10.3390/applbiosci3010002 · Cited by 255
Huang, J.-D., Wang, J., Ramsey, E., Leavey, G., Chico, T., & Condell, J. V. (2022) · Sensors · doi:10.3390/s22208002 · Cited by 203
Metadata
- Authors
- Bachtiger, P., Petri, C. F., Scott, F. E., Ri Park, S., Kelshiker, M. A., Sahemey, H. K., Dumea, B., Alquero, R., Padam, P. S., Hatrick, I. R., Ali, A., Ribeiro, M., Cheung, W.-S., Bual, N., Rana, B., Shun-Shin, M., Kramer, D. B., Fragoyannis, A., Keene, D., Plymen, C. M., & Peters, N. S.
- Affiliations
- Imperial College London (GB)
- Imperial College Healthcare NHS Trust (GB)
- Beth Israel Deaconess Medical Center (US)
- Corresponding author
- Peters, Nicholas S
- ORCID
- Venue
- The Lancet Digital Health
- Publisher
- Elsevier BV
- Published
- 2022-02
- Volume
- 4
- Issue
- 2
- Pages
- e117-e125
- ISSN
- 2589-7500 (ISSN-L)
- PMC
- PMC8789562
- OpenAlex
- W4206756110
- Cited by
- 167
- References
- 31
- Citations by year
- 2026: 40 · 2025: 48 · 2024: 46 · 2023: 18 · 2022: 14
- FWCI
- 21.94
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- The Lancet Digital Health (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Spiral (Imperial College London) (submittedVersion, cc-by)
- Funders
- National Institute for Health Research
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Health Sciences › Health Professions › Health Information Management › Artificial Intelligence in Healthcare
- Health Sciences › Medicine › Health Informatics › Artificial Intelligence in Healthcare and Education
- Keywords
- heart failure with reduced ejection fraction, point-of-care screening, artificial intelligence, single-lead ECG, convolutional neural networks, left ventricular ejection fraction, transthoracic echocardiography, cardiac auscultation, primary care, early diagnosis, model generalization, sex differences
- MeSH terms
- Aged, Artificial Intelligence, Electrocardiography, Female, United Kingdom, Heart Failure, Humans, Male, Middle Aged, Physical Examination, Prospective Studies, Neural Networks, Computer, Stethoscopes, Point-of-Care Systems
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review