Detection of left ventricular systolic dysfunction from single-lead electrocardiography adapted for portable and wearable devices
Open accessPublished version of a preprint
Abstract
Artificial intelligence (AI) can detect left ventricular systolic dysfunction (LVSD) from electrocardiograms (ECGs). Wearable devices could allow for broad AI-based screening but frequently obtain noisy ECGs. We report a novel strategy that automates the detection of hidden cardiovascular diseases, such as LVSD, adapted for noisy single-lead ECGs obtained on wearable and portable devices. We use 385,601 ECGs for development of a standard and noise-adapted model. For the noise-adapted model, ECGs are augmented during training with random gaussian noise within four distinct frequency ranges, each emulating real-world noise sources. Both models perform comparably on standard ECGs with an AUROC of 0.90. The noise-adapted model performs significantly better on the same test set augmented with four distinct real-world noise recordings at multiple signal-to-noise ratios (SNRs), including noise isolated from a portable device ECG. The standard and noise-adapted models have an AUROC of 0.72 and 0.87, respectively, when evaluated on ECGs augmented with portable ECG device noise at an SNR of 0.5. This approach represents a novel strategy for the development of wearable-adapted tools from clinical ECG repositories.
AI snapshot
Written from the paper’s full text.
TL;DR
A noise-adapted deep-learning model detects left ventricular systolic dysfunction (LVSD) from single-lead ECG and remains more robust to real-world noise than a standard model. It was developed on 385,601 ECG–echocardiogram pairs and tested on noise-augmented ECGs, including noise isolated from a portable ECG device.
Key findings
- On the clean held-out test set, the noise-adapted and standard models performed similarly: AUROC 0.90 (95% CI 0.89–0.91) vs 0.90 (95% CI 0.88–0.91), p = 0.60; AUPRC 0.46 vs 0.48.
- With portable ECG device noise at an SNR of 0.5, the noise-adapted model had AUROC 0.87 (95% CI 0.86–0.88) vs the standard model’s 0.72 (95% CI 0.71–0.74), p < 0.001.
- Across MIT-BIH noises at SNRs 0.5–2.0, the noise-adapted model had AUROC ranges of 0.86–0.89 for electrode motion, 0.87–0.89 for muscle artifact, and 0.88–0.89 for baseline wander; the standard model was lower at 0.79–0.86, 0.81–0.86, and 0.80–0.86, respectively.
- Last-layer prediction shifts were smaller for the noise-adapted model: for portable ECG noise, scaled Euclidean distance was 0.41 (95% CI 0.40–0.42) vs 0.50 (95% CI 0.49–0.51); for baseline wander, 0.36 (95% CI 0.36–0.37) vs 0.52 (95% CI 0.51–0.53), both p < 0.001.
Methods
- Retrospective cohort of 116,210 patients aged 18 or older with clinically indicated ECG and paired transthoracic echocardiogram within 15 days at Yale-New Haven Hospital, 2015–2021; 385,601 ECG–TTE pairs were used, with Lead I isolated from 12-lead ECGs and patient-level splits of 85% training, 5% validation, and 10% test.
- The noise-adapted model was trained with random Gaussian noise in four frequency bands (3–12, 12–50, 50–100, and 100–150 Hz) at SNRs of 0.50, 0.75, 1.00, and 1.25; the standard model was trained on original ECGs. Both used a CNN with seven 2D convolutional layers and dense layers, targeting LVEF < 40%.
- Evaluation used the held-out test set without noise and augmented with MIT-BIH electrode motion, muscle artifact, and baseline wander noise, plus noise extracted from a KardiaMobile 6L portable ECG device, across multiple SNRs; analyses included AUROC, AUPRC, sensitivity, specificity, PPV, NPV, DeLong’s test, UMAP, and scaled Euclidean distances.
Limitations
- The paper states the model was developed in patients with clinically indicated ECG and echocardiogram, not the intended broader screening population, so prospective assessment is warranted; performance may vary by LVSD severity and near the LVEF 40% cutoff, and echocardiographic LVEF is less precise than MRI.
- The paper also states it cannot ascertain whether the model maintains performance on every possible wearable noise type, magnitude, or device-specific noise signature.
- Evident from the text: there was no external validation cohort, the data cannot be shared publicly, and training/evaluation relied on synthetic Gaussian noise bands and selected real-world noise recordings rather than actual wearable ECG datasets.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
20 of 42 cited sources checked · 4 supported · 1 flagged · accuracy not yet measured
- Premisecitation
Models trained on clinical ECGs have been applied to wearable device ECGs, but have traditionally shown significantly lower performance on wearable device ECGs than on held-out clinical ECG test sets.
“Models trained on clinical ECGs have been applied to wearable device ECGs, but have traditionally shown significantly lower performance on wearable device ECGs than on held-out clinical ECG test sets14,15.”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [14] Bachtiger, P. et al. Point-of-care screening for heart failure with reduced ejection fraction using artificial intelligence during ECG-enabled stethoscope examination in London, UK: a prospective, observational, multicentre study. Lancet Digit. Health 4, e117–e125 (2022). source
The Methods (AI-ECG algorithm; page unavailable) describe a model trained on clinical 12-lead ECG data and tested using single-lead recordings from an ECG-enabled stethoscope. The Discussion (page unavailable) reports an internal 12-lead validation AUROC of 0.93, but says the study’s single-lead findings compare favourably with an external 12-lead validation AUROC of 0.82. The paper does not establish a traditional, statistically significant performance drop on wearable-device ECGs relative to held-out clinical ECG test sets. (LLM, conf=0.86)
Experimental — accuracy not yet measured - cites [15] Attia, Z. I. et al. Automated detection of low ejection fraction from a one-lead electrocardiogram: application of an AI algorithm to an electrocardiogram-enabled Digital Stethoscope. Eur. Heart J. Digit. Health https://doi.org/10.1093/ehjdh/ztac030 (2022). source
The Methods (Construction of neural network for use on single-lead electrocardiogram; page unavailable) confirm that a model trained on clinical 12-lead ECGs was adapted for single-lead use and report an AUC of 0.90 on the original clinical test set. The Results (Accuracy of single-lead digital stethoscope-enabled electrocardiogram; page unavailable) report an AUC of 0.91 for the optimal stethoscope ECG at the same EF threshold—not significantly lower performance. This study does not establish a broader historical trend. (LLM, conf=0.86)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured
- Premisecitation
Models which have successfully generalized to wearable device ECGs have relied on automated algorithms to select high-quality subsets of available ECG data, with reduced performance when evaluated on all collected ECGs.
“Additionally, models which have successfully generalized to wearable device ECGs have relied on automated algorithms to select high-quality subsets of available ECG data, with reduced performance when evaluated on all collected ECGs15,26.”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [15] Attia, Z. I. et al. Automated detection of low ejection fraction from a one-lead electrocardiogram: application of an AI algorithm to an electrocardiogram-enabled Digital Stethoscope. Eur. Heart J. Digit. Health https://doi.org/10.1093/ehjdh/ztac030 (2022). source
The Methods section describes an automated algorithm that selects the highest-quality ECG recording from those acquired with a digital stethoscope. The Results section reports higher AUCs for that selection than for scores averaged across all leads. This supports the comparison for this device, but not the broader claim that wearable-device ECG models generally rely on such selection or that performance was evaluated on all collected ECGs without selection. (LLM, conf=0.86)
Experimental — accuracy not yet measured - cites [26] Attia, Z. I. et al. Prospective evaluation of smartwatch-enabled detection of left ventricular dysfunction. Nat. Med. 28, 2497–2503 (2022). source
The Discussion (page not available) reports that an automated filter selected clean smartwatch ECG tracings and improved AUC to 0.88, compared with 0.85 for the ECG closest in time to the echocardiogram. That supports the benefit of quality selection, but the lower-performance comparison was not an evaluation on all collected ECGs. (LLM, conf=0.86)
Experimental — accuracy not yet measured
- Premisecitation
The reliability of AI-based screening tools for LVSD is limited by the presence of noise in data collected from wearable and handheld devices.
“but the reliability of such tools is limited by the presence of noise in data collected from wearable and handheld devices12,13.”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [12] Dörr, M. et al. The WATCH AF Trial: SmartWATCHes for detection of atrial fibrillation. JACC Clin. Electrophysiol. 5, 199–208 (2019). source
The WATCH AF discussion reports that insufficient wrist-signal quality prevented the smartwatch algorithm from producing an AF diagnosis in 21.8% of subjects, and the Study Limitations section identifies a worsening signal-to-noise ratio as a potential problem. This supports a signal-quality limitation for wearable AF screening, but the paper does not establish the claim for AI-based LVSD screening. Page numbers are not available in the supplied passages. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites [13] Mannhart, D. et al. Clinical validation of 5 direct-to-consumer wearable smart devices to detect atrial fibrillation: BASEL wearable study. JACC Clin. Electrophysiol. 9, 232–242 (2023). source
The Discussion says technical artefacts, including noise, affected the devices’ automated interpretation of single-lead ECGs. The study evaluates atrial fibrillation detection, however, not AI-based screening for left ventricular systolic dysfunction (LVSD), so it does not establish that limitation for LVSD tools. (LLM, conf=0.86)
Experimental — accuracy not yet measured
- Premisecitation
Models demonstrate poorer performance when tested on all available wearable ECG data as opposed to selected high-quality subsets.
“as models demonstrate poorer performance when tested on all available wearable ECG data as opposed to selected high-quality subsets15.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [15] Attia, Z. I. et al. Automated detection of low ejection fraction from a one-lead electrocardiogram: application of an AI algorithm to an electrocardiogram-enabled Digital Stethoscope. Eur. Heart J. Digit. Health https://doi.org/10.1093/ehjdh/ztac030 (2022). source
- Premisecitation
Early diagnosis can effectively mitigate the risk of LVSD.
“Early diagnosis can effectively mitigate this risk2–4”
Quote in the paperSubstantial claim2 of 3 sources checked- cites [2] Srivastava, P. K. et al. Heart failure hospitalization and guideline-directed prescribing patterns among heart failure with reduced ejection fraction patients. JACC Heart Fail 9, 28–38 (2021). source
The Discussion (page unavailable) examines prescribing changes after heart-failure hospitalization in patients who already have reduced ejection fraction. Searches for early diagnosis, detection, and prevention found no passage saying that early diagnosis mitigates the risk of LVSD. (LLM, conf=0.73)
Experimental — accuracy not yet measured - cites [3] Wolfe, N. K., Mitchell, J. D. & Brown, D. L. The independent reduction in mortality associated with guideline-directed medical therapy in patients with coronary artery disease and heart failure with reduced ejection fraction. Eur. Heart J. Qual Care Clin Outcomes. 7, 416–421 (2021). source
The Introduction describes a study of guideline-directed medical therapy and mortality in patients who already had coronary artery disease and reduced ejection fraction. The Clinical implications section discusses early initiation of therapy, not early diagnosis or a reduced risk of developing LVSD. Searches for early diagnosis and LVSD prevention found no passage addressing the claimed effect. (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [4] Heidenreich, P. A. et al. 2022 AHA/ACC/HFSA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation 145, e895–e1032 (2022). source
- Premisecitation
The performance of models for detecting LVSD from portable device ECGs may degrade in the real-world setting, with lower performance than observed in the original single-lead derivatives of the clinical development studies.
“Consequently, the performance of models for detecting LVSD from portable device ECGs may degrade in the real-world setting, with lower performance than observed in the original single-lead derivatives of the clinical development studies14,15.”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [14] Bachtiger, P. et al. Point-of-care screening for heart failure with reduced ejection fraction using artificial intelligence during ECG-enabled stethoscope examination in London, UK: a prospective, observational, multicentre study. Lancet Digit. Health 4, e117–e125 (2022). source
The Results report an AUROC of 0.85 for the best single-lead stethoscope position in this prospective study. The Discussion gives a higher AUROC of 0.93 for the original internal validation, but that comparison used 12-lead ECGs, not the original single-lead derivatives specified in the claim. It also says the study’s single-lead findings compare favourably with a separate external 12-lead validation. (LLM, conf=0.83)
Experimental — accuracy not yet measured - cites [15] Attia, Z. I. et al. Automated detection of low ejection fraction from a one-lead electrocardiogram: application of an AI algorithm to an electrocardiogram-enabled Digital Stethoscope. Eur. Heart J. Digit. Health https://doi.org/10.1093/ehjdh/ztac030 (2022). source
The Discussion (page unavailable) says poorer-quality stethoscope recordings may produce inaccurate predictions, supporting the possibility of degraded performance with portable-device ECGs. But the paper does not show a general real-world decline relative to the clinical derivation study: Methods reports an AUC of 0.90 for its original-cohort test, while Results reports an AUC of 0.91 for optimally selected stethoscope ECGs. The study was conducted in patients referred for echocardiography, not a routine real-world screening population. (LLM, conf=0.82)
Experimental — accuracy not yet measured
- Premisecitation
LVSD is frequently diagnosed only after patients develop symptomatic disease due to the lack of effective screening strategies.
“but LVSD is frequently diagnosed only after patients develop symptomatic disease due to the lack of effective screening strategies5–7.”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [5] Wang, T. J., Levy, D., Benjamin, E. J. & Vasan, R. S. The epidemiology of “asymptomatic” left ventricular systolic dysfunction: implications for screening. Ann. Intern. Med. 138, 907–916 (2003). source
- cites [6] Vasan, R. S. et al. Plasma natriuretic peptides for community screening for left ventricular hypertrophy and systolic dysfunction: the Framingham heart study. JAMA 288, 1252–1259 (2002). source
- cites [7] McDonagh, T. A. & McDonald, K. Screening for asymptomatic left ventricular dysfunction using B‐type natriuretic peptide. Congest. Heart Fail 14, 5–8 (2008). source
The Conclusions say that screening for asymptomatic left ventricular dysfunction is not yet routine and that health systems focus on treating manifest heart failure. They do not establish that LVSD is frequently diagnosed only after symptoms develop. They also describe BNP screening as acceptably accurate, so the barrier is not simply a lack of effective screening tests. (LLM, conf=0.84)
Experimental — accuracy not yet measured
- Premisecitation
A wearable device-based atrial fibrillation screening study disqualified 22% of patients due to insufficient signal quality.
“with a wearable device-based atrial fibrillation screening study disqualifying 22% of patients due to insufficient signal quality12.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [12] Dörr, M. et al. The WATCH AF Trial: SmartWATCHes for detection of atrial fibrillation. JACC Clin. Electrophysiol. 5, 199–208 (2019). source
- Premiseown data
Artificial intelligence can detect LVSD from ECGs.
“Artificial intelligence (AI) can detect left ventricular systolic dysfunction (LVSD) from electrocardiograms (ECGs)”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [8] Attia, Z. I. et al. Screening for cardiac contractile dysfunction using an artificial intelligence–enabled electrocardiogram. Nat. Med. 25, 70–74 (2019). source
- cites [9] Galasko, G. I. W., Barnes, S. C., Collinson, P., Lahiri, A. & Senior, R. What is the most cost-effective strategy to screen for left ventricular systolic dysfunction: natriuretic peptides, the electrocardiogram, hand-held echocardiography, traditional echocardiography, or their combination? Eur. Heart J. 27, 193–200 (2006). source
The Discussion (page unavailable) reports that ECG screening had high sensitivity for predicting LVSD in community subjects. The Electrocardiography section describes screening based on specified ECG abnormalities, but the paper does not report using artificial intelligence to detect LVSD. (LLM, conf=0.87)
Experimental — accuracy not yet measured
- Premisecitation
Lead I represents the standard lead obtained from wearable devices.
“Lead I was chosen as it represents the standard lead obtained from wearable devices11.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [11] Duarte, R. et al. Lead-I ECG for detecting atrial fibrillation in patients attending primary care with an irregular pulse using single-time point testing: A systematic review and economic evaluation. PLoS ONE 14, e0226671 (2019). source
The Introduction (page unavailable) describes lead-I ECG devices as handheld instruments that simulate Lead I of a 12-lead ECG. A search for wearable devices found no passage saying that Lead I is the standard lead obtained from them. (LLM, conf=0.72)
Experimental — accuracy not yet measured
- Findingown data
The model performance generalizes well to unseen noise.
“this suggests that the model performance generalizes well to unseen noise”
Quote in the paperSubstantial claim - Findingown data
The difference between models at SNR 0.5 for portable ECG device noise was significant with p-value < 0.001.
“at which the noise-adapted model retained an AUROC of 0.87 (95% CI 0.86–0.88) and the standard model had an AUROC of 0.72 (95% CI 0.71–0.74, p-value < 0.001)”
Quote in the paperSubstantial claim - Findingown data
The standard model's performance was significantly lower at each SNR for portable ECG device noise, ranging from 0.72 to 0.83.
“The standard model’s performance was significantly lower at each SNR, ranging from 0.72 to 0.83.”
Quote in the paperSubstantial claim - Findingown data
The scaled Euclidean distances between predictions for noised and standard versions of the ECGs were lower for the noise-adapted model across all four noises.
“Quantitatively, the scaled Euclidean distances between predictions for noised and standard versions of the ECGs were lower for the noise-adapted model across all four noises.”
Quote in the paperSubstantial claim - Findingnone given
A single-lead ECG algorithm accurately identifies LV systolic dysfunction despite significant noisy artifacts.
“Using this approach, we developed a single-lead ECG algorithm that accurately identifies LV systolic dysfunction despite significant noisy artifacts”
Quote in the paperSubstantial claim - Findingown data
The noise-adapted model performs significantly better than the standard model on the same test set augmented with four distinct real-world noise recordings at multiple SNRs, including noise isolated from a portable device ECG.
“The noise-adapted model performs significantly better on the same test set augmented with four distinct real-world noise recordings at multiple signal-to-noise ratios (SNRs), including noise isolated from a portable device ECG.”
Quote in the paperSubstantial claim - Findingown data
The noise-adapted model's performance was comparable to the standard model across subgroups of age, sex, and race.
“The noise-adapted model’s performance was comparable to the standard model across subgroups of age, sex, and race (Table 1).”
Quote in the paperSubstantial claim - Findingown data
When evaluated on ECGs augmented with portable ECG device noise at an SNR of 0.5, the standard and noise-adapted models have AUROC of 0.72 and 0.87, respectively.
“The standard and noise-adapted models have an AUROC of 0.72 and 0.87, respectively, when evaluated on ECGs augmented with portable ECG device noise at an SNR of 0.5.”
- Also stated in: § Standard and noise-adapted performance with portable ECG device noise · plain
“at which the noise-adapted model retained an AUROC of 0.87 (95% CI 0.86–0.88) and the standard model had an AUROC of 0.72 (95% CI 0.71–0.74, p-value < 0.001)”
Quote in the paperSubstantial claim - Also stated in: § Standard and noise-adapted performance with portable ECG device noise · plain
- Findingown data
The algorithm was developed and validated in a diverse population.
“Notably, the algorithm was developed and validated in a diverse population and demonstrates consistent performance across age, sex, and race subgroups.”
Quote in the paperSubstantial claim - Findingown data
Both the noise-adapted and standard models perform comparably on standard ECGs without added noise, with an AUROC of 0.90 each.
“Both models perform comparably on standard ECGs with an AUROC of 0.90.”
- Also stated in: § Detection of LV systolic dysfunction · plain
“the noise-adapted and standard models, with an AUROC of 0.90 (95% CI 0.89–0.91) and 0.90 (95% CI 0.88–0.91), respectively, performed similarly on a held-out test set without added noise (p-value = 0.60).”
Quote in the paperSubstantial claim - Also stated in: § Detection of LV systolic dysfunction · plain
- Findingown data
The noise-adapted model's AUROC ranged from 0.87 to 0.89 across all SNRs for portable ECG device noise.
“with the noise-adapted model’s AUROC ranging from 0.87 to 0.89.”
Quote in the paperSubstantial claim - Interpretationown data
We developed a novel strategy that automates the detection of hidden signatures of cardiovascular diseases, such as LVSD, adapted for noisy single-lead ECGs obtained on wearable and portable devices.
“We developed a novel strategy that automates the detection of hidden signatures of cardiovascular diseases, such as LVSD, adapted for noisy single-lead ECGs obtained on wearable and portable devices.”
Quote in the paperSubstantial claim - Premisecitation
3-12 Hz models the motion artifact noises attributable to tremors.
“3–12 Hz models the motion artifact noises attributable to tremors, which occur within this frequency range20,21.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [20] Charles, P. D., Esper, G. J., Davis, T. L., Maciunas, R. J. & Robertson, D. Classification of tremor and update on treatment. Am. Fam. Physician 59, 1565–1572 (1999).
- cites [21] Friesen, G. M. et al. A comparison of the noise sensitivity of nine QRS detection algorithms. IEEE Trans. Biomed. Eng. 37, 85–98 (1990). source
- Study design and conductreasoning
No exclusion criteria were applied to ensure generalizability of the models.
“To ensure the generalizability of our models, we applied no exclusion criteria, including patients of all sexes, races, and ethnicities (Table S1).”
Quote in the paperSubstantial claim - Study design and conductnone given
The model was developed using ECGs from patients who had both an ECG and a clinically indicated echocardiogram.
“First, this model was developed using ECGs from patients who had both an ECG and a clinically indicated echocardiogram.”
Quote in the paperSubstantial claim - Interpretationreasoning
The noise-adapted approach enables reliable modelling of noisy, wearable-derived, single-lead ECG signals.
“First, it augments clinical ECG datasets in such way that it enables reliable modelling of noisy, wearable-derived, single-lead ECG signals.”
Quote in the paperSubstantial claim - Interpretationown data
The consistent performance across demographic subgroups suggests robustness and generalizability of the model's performance.
“the consistent performance across demographic subgroups suggests robustness and generalizability of the model’s performance.”
Quote in the paperSubstantial claim - Interpretationown data
The predictions of the noise-adapted model are more resilient to noise than those of the standard model.
“indicating that the predictions of the noise-adapted model are more resilient than those of the standard model”
Quote in the paperSubstantial claim - Interpretationreasoning
Accounting for noise is a prerequisite to develop broadly accessible models that will form the basis of effective screening programs for LVSD in the community.
“Accounting for this noise is a prerequisite to develop broadly accessible models that will form the basis of effective screening programs for LVSD in the community.”
Quote in the paperNot a substantial claim - Premisecitation
Several sources of noise exist in wearable data, arising from factors such as poor electrode contact with the skin, movement and muscle contraction during the ECG, and external electrical interference.
“Indeed, several sources of noise exist in wearable data, arising from factors such as poor electrode contact with the skin, movement and muscle contraction during the ECG, and external electrical interference16–19.”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [16] Satija, U., Ramkumar, B. & Manikandan, M. S. A review of signal processing techniques for electrocardiogram signal quality assessment. IEEE Rev. Biomed. Eng. 11, 36–52 (2018). source
- cites [17] Moody, G. B., Muldrow, W. & Mark, R. A noise stress test for arrhythmia detectors. Comput. Cardiol. 11, 381–384 (1984).
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [18] Satija, U., Ramkumar, B. & Manikandan, M. S. A new automated signal quality-aware ecg beat classification method for unsupervised ECG diagnosis environments. IEEE Sens. J. 19, 277–286 (2019). source
- cites [19] Satija, U., Ramkumar, B. & Manikandan, M. S. A simple method for detection and classification of ECG noises for wearable ECG monitoring devices. in 2015 2nd International Conference on Signal Processing and Integrated Networks (SPIN) 164–169 (2015). source
- Interpretationown data
The algorithm demonstrates excellent discriminatory performance even on ECGs containing twice as much noise as signal.
“Specifically, the algorithm demonstrates excellent discriminatory performance even on ECGs containing twice as much noise as signal”
Quote in the paperSubstantial claim
- Classified as probably not a substantial claim. 21
- May combine two claims. 18
Classified as probably not a substantial claim. May combine two claims.
PremisecitationLVEF < 40% is a threshold present in most heart failure diagnosis guidelines and consistent with prior work.
“The cutoff for low LVEF was set as LVEF < 40%, a threshold present in most heart failure diagnosis guidelines4, and consistent with prior work in this space8,22.”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [4] Heidenreich, P. A. et al. 2022 AHA/ACC/HFSA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation 145, e895–e1032 (2022). source
- cites [8] Attia, Z. I. et al. Screening for cardiac contractile dysfunction using an artificial intelligence–enabled electrocardiogram. Nat. Med. 25, 70–74 (2019). source
- cites [22] Sangha, V. et al. Detection of left ventricular systolic dysfunction from electrocardiographic images. Preprint at https://www.medrxiv.org/content/10.1101/2022.06.04.22276000v2 (2022). source
May combine two claims.
PremisecitationStudies have shown high rates of poor signal quality among data recorded by wearable devices, and significant variation in quality across different manufacturers.
“Studies have shown high rates of poor signal quality among data recorded by wearable devices, and significant variation in quality across different manufacturers13.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [13] Mannhart, D. et al. Clinical validation of 5 direct-to-consumer wearable smart devices to detect atrial fibrillation: BASEL wearable study. JACC Clin. Electrophysiol. 9, 232–242 (2023). source
The Discussion reports that 17%–26% of recordings were labeled inconclusive by device algorithms and says these rates differed significantly across devices. But inconclusive does not necessarily mean poor signal quality: the Results say cardiologists could interpret 95% of those tracings, and only 1% of 825 recordings were noninterpretable because of low quality or noise. The paper therefore supports variation in algorithm outcomes, not high rates of poor-quality signals across manufacturers. (LLM, conf=0.87)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationEchocardiography LVEF measurement is less precise relative to other approaches such as magnetic resonance imaging.
“This might also be attributable to a lack of precision in LVEF measurement by echocardiography, which has shown to be less precise relative to other approaches, such as magnetic resonance imaging27,28.”
Quote in the paperNot a substantial claim2 of 2 sources checked- cites [27] Pellikka, P. A. et al. Variability in ejection fraction measured by echocardiography, gated single-photon emission computed tomography, and cardiac magnetic resonance in patients with coronary artery disease and left ventricular dysfunction. JAMA Netw. Open 1, e181456 (2018). source
The Results (Comparison of Imaging Modalities; page unavailable) show greater disagreement between echocardiography and CMR than between CMR and SPECT, supporting a concern about variability in echocardiographic LVEF measurements. But the Discussion and Limitations say there is no gold standard and do not establish that echocardiography itself is less precise than CMR; the study primarily compares agreement between modalities. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [28] Jenkins, C. et al. Left ventricular volume measurement with echocardiography: a comparison of left ventricular opacification, three-dimensional echocardiography, or both with magnetic resonance imaging. Eur. Heart J. 30, 98–106 (2009). source
The Introduction and ‘Limitations of each technique’ sections (page not specified) describe poor reproducibility and observer variability for non-contrast 2D echocardiography, supporting the concern about its precision. But the Introduction also says contrast enhancement can bring inter-observer reproducibility to a level comparable with MRI, so the claim is too broad for echocardiography generally. (LLM, conf=0.85)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationLVEF below 40% is a threshold present in most heart failure diagnosis guidelines.
“Both models were trained to detect LVEF below 40%, a threshold present in most heart failure diagnosis guidelines4, and consistent with prior work8,22.”
Quote in the paperNot a substantial claim1 of 3 sources checked- cites [4] Heidenreich, P. A. et al. 2022 AHA/ACC/HFSA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation 145, e895–e1032 (2022). source
- cites [8] Attia, Z. I. et al. Screening for cardiac contractile dysfunction using an artificial intelligence–enabled electrocardiogram. Nat. Med. 25, 70–74 (2019). source
- cites [22] Sangha, V. et al. Detection of left ventricular systolic dysfunction from electrocardiographic images. Preprint at https://www.medrxiv.org/content/10.1101/2022.06.04.22276000v2 (2022). source
May combine two claims.
Findingown dataThe noise-adapted model had specificity and sensitivity of 0.68 and 0.92, respectively, and a PPV and NPV of 0.20 and 0.99, respectively, in the held-out test set.
“the noise-adapted model had specificity and sensitivity of 0.68 and 0.92, respectively, and a PPV and NPV of 0.20 and 0.99, respectively, in the held-out test set.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe standard model had lower performance across all SNRs for every noise combination, with AUROC between 0.77-0.85, 0.77-0.85, and 0.78-0.86 for portable ECG device noise combined with electrode motion, muscle artifact, and baseline wander noise, respectively.
“The standard model had lower performance across all SNRs for every noise combination, with AUROC between 0.77–0.85, 0.77–0.85, and 0.78–0.86 for portable ECG device noise combined with electrode motion, muscle artifact, and baseline wander noise, respectively (Table S3).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor portable ECG noise at SNR 0.5, the average scaled Euclidean distance was 0.50 for the standard model and 0.41 for the noise-adapted model.
“In the ECG with noise derived from portable ECGs, for instance, the average scaled Euclidean distance for the standard model was 0.50 (95% CI 0.49–0.51) and for the noise-adapted model was 0.41 (95% CI 0.40–0.42).”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenThe noise-adapted approach resulted in selective removal of noise from signal, even for noises the model had not encountered before, while preserving the model's robustness in discerning complex hidden labels.
“Compared with the current approach of developing models, our noise-adapted approach resulted in selective removal of noise from signal, even for noises the model had not encountered before, while preserving the model’s robustness in discerning complex hidden labels.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor the standard model, predictions for noised ECGs were visually distinct from those of standard ECGs in UMAP projections, whereas for the noise-adapted model there was no visual separation.
“For the standard model, the predictions for each of the noised ECGs were visually distinct from those of the standard ECGs, despite being for the same set of 1000 ECGs, and differing only on the added noise. However, for the noise-adapted model, there was no visual separation in the model’s predictions between standard and noised ECGs”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor baseline wander at SNR 0.5, the average scaled Euclidean distance was 0.52 for the standard model and 0.36 for the noise-adapted model.
“Similarly for the baseline wander it was 0.52 (0.51–0.53) and 0.36 (0.36–0.37), respectively (Table 3).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataAUPRC on the held-out test set without added noise was 0.46 for the noise-adapted model and 0.48 for the standard model.
“Area under the precision-recall curve (AUPRC) on this held-out test set was 0.46 and 0.48, respectively.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor the noise-adapted model, performance was comparable across all SNRs for combinations of portable ECG device noise and electrode motion, muscle artifact, and baseline wander noise with AUROC between 0.86-0.89, 0.87-0.89, and 0.88-0.89, respectively, for SNRs from 0.5 to 2.0.
“For the noise-adapted model, model performance was comparable across all SNRs for the combinations of portable ECG device noise and electrode motion, muscle artifact, and baseline wander noise with AUROC between 0.86–0.89, 0.87–0.89, and 0.88–0.89, respectively, for SNRs from 0.5 to 2.0.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe standard model had lower performance across all SNRs for every MIT-BIH noise, with AUROC between 0.79-0.86, 0.81-0.86, and 0.80-0.86 for electrode motion, muscle artifact, and baseline wander noise, respectively.
“The standard model had lower performance across all SNRs for every noise, with AUROC between 0.79–0.86, 0.81–0.86, and 0.80–0.86 for electrode motion, muscle artifact, and baseline wander noise, respectively (Table 2 and Fig. 1).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe standard model had sensitivity and specificity of 0.69 and 0.91, respectively, and a PPV and NPV of 0.21 and 0.99, respectively.
“The standard model had sensitivity and specificity of 0.69 and 0.91, respectively, and a PPV and NPV of 0.21 and 0.99, respectively.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor the noise-adapted model, performance was comparable across all SNRs for each MIT-BIH noise (electrode motion, muscle artifact, baseline wander) with AUROC between 0.86-0.89, 0.87-0.89, and 0.88-0.89 respectively for SNRs from 0.5 to 2.0.
“For the noise-adapted model, model performance was comparable across all SNRs for each MIT-BIH noise with AUROC between 0.86–0.89, 0.87–0.89, and 0.88–0.89 for electrode motion, muscle artifact, and baseline wander noise, respectively, for SNRs from 0.5 to 2.0.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataThe selected architecture yielded the highest AUROC on the validation set with the fewest number of parameters for the standard model.
“We selected the architecture yielding the highest area under the receiver operator characteristic (AUROC) curve on the validation set with the fewest number of parameters for the standard model.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningThe approach suggests a novel approach for the development of wearable-adapted tools from clinical ECG repositories.
“suggesting a novel approach for the development of wearable-adapted tools from clinical ECG repositories.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductown dataThreshold values were chosen based on cutoffs that achieved sensitivity of 0.90 in validation data.
“chose threshold values based on cutoffs that achieved sensitivity of 0.90 in validation data.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThe noise-adapted deep-learning algorithm accurately identifies LV systolic dysfunction from single-lead ECG data and is resilient to significant noisy artifacts, despite not encountering the specific noises in the model development process.
“we developed a noise-adapted deep-learning algorithm that accurately identifies LV systolic dysfunction from single-lead ECG data and was resilient to significant noisy artifacts, despite not encountering the specific noises in the model development process.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningThe noise-adapted approach defines a novel paradigm on how to build robust, wearable-ready, single-lead ECG cardiovascular screening models from clinical ECG repositories.
“The noise-adapted approach defines a novel paradigm on how to build robust, wearable-ready, single-lead ECG cardiovascular screening models from clinical ECG repositories”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe MIT-BIH dataset noises represent three types of noises frequently encountered in ECGs: baseline wander, muscle artifact, and electrode motion artifact.
“The MIT-BIH dataset noises, each obtained at a sampling frequency of 360 Hz, represent three types of noises frequently encountered in ECGs: baseline wander noise, a low-frequency noise produced by lead or subject movement17, muscle artifact, caused by muscle contractions21, and electrode motion artifact, which is caused by irregular movement of the electrodes during ECG recordings17,21.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [17] Moody, G. B., Muldrow, W. & Mark, R. A noise stress test for arrhythmia detectors. Comput. Cardiol. 11, 381–384 (1984).
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [21] Friesen, G. M. et al. A comparison of the noise sensitivity of nine QRS detection algorithms. IEEE Trans. Biomed. Eng. 37, 85–98 (1990). source
Classified as probably not a substantial claim.
Study design and conductown dataEchocardiograms were evaluated by expert cardiologists.
“These echocardiograms had been evaluated by expert cardiologists, and the LVEF defined in their interpretation was identified.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
InterpretationreasoningThe current 12-lead ECG-based models are limited to investments by health systems to incorporate tools into digital ECG repositories, and thereby limited to individuals who already seek care in those systems.
“The current 12-lead ECG-based models are limited to investments by health systems to incorporate tools into digital ECG repositories, and thereby limited to individuals who already seek care in those systems.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation100-150 Hz contains higher-frequency muscle noises.
“the 12–50 Hz and 100–150 Hz ranges contain the lower and higher-frequency muscle noises, respectively17,21.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [17] Moody, G. B., Muldrow, W. & Mark, R. A noise stress test for arrhythmia detectors. Comput. Cardiol. 11, 381–384 (1984).
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [21] Friesen, G. M. et al. A comparison of the noise sensitivity of nine QRS detection algorithms. IEEE Trans. Biomed. Eng. 37, 85–98 (1990). source
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
InterpretationreasoningEven this technology may not be available or cost-effective for smaller hospitals and clinics with limited access to digital ECGs.
“In addition to the clinically indicated ECGs limiting the scope of screening, even this technology may not be available or cost-effective for smaller hospitals and clinics with limited access to digital ECGs.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductreasoningThe study was reviewed and approved by the Yale Institutional Review Board, which waived the need for informed consent.
“The study was reviewed by the Yale Institutional Review Board, which approved the study protocol and waived the need for informed consent as the study represents secondary analysis of existing data.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningModel performance may vary by the severity of LV systolic dysfunction.
“Second, the model performance may vary by the severity of LV systolic dysfunction.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenModel performance among patients with an LVEF near 40% may differ compared with those with LVEF significantly higher or lower than 40%.
“it is possible that the model performance among patients with an LVEF near to this cutoff differed compared with those individuals with LVEF significantly higher or lower than 40%.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningA model which generalizes to ECGs with varying types and levels of noise may be necessary for implementation in community-wide screening settings.
“A model which generalizes to ECGs with varying types and levels of noise may be necessary for implementation in such settings.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation50-100 Hz reflects consistent electrode contact noise.
“The 50–100 Hz domain reflects consistent electrode contact noise21”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [21] Friesen, G. M. et al. A comparison of the noise sensitivity of nine QRS detection algorithms. IEEE Trans. Biomed. Eng. 37, 85–98 (1990). source
Classified as probably not a substantial claim.
Study design and conductcitationModel weights were calculated using the effective number of samples class re-weighting scheme so that learning was not affected by the lower frequency of LVEF < 40%.
“Model weights were calculated for the loss function such that learning was not affected by the lower frequency of LVEF < 40% compared to the incidence of LVEF ≥ 40% using the effective number of samples class re-weighting scheme33 (Fig. S2).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [33] Cui, Y., Jia, M., Lin, T.-Y., Song, Y. & Belongie, S. Class-balanced loss based on effective number of samples. https://arxiv.org/abs/1901.05555 (2019). source
Probably not a substantial claim: may not pass S3 (load-bearing).
Premisecitation12-50 Hz contains lower-frequency muscle noises.
“the 12–50 Hz and 100–150 Hz ranges contain the lower and higher-frequency muscle noises, respectively17,21.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [17] Moody, G. B., Muldrow, W. & Mark, R. A noise stress test for arrhythmia detectors. Comput. Cardiol. 11, 381–384 (1984).
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [21] Friesen, G. M. et al. A comparison of the noise sensitivity of nine QRS detection algorithms. IEEE Trans. Biomed. Eng. 37, 85–98 (1990). source
Probably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningThe approach has significant potential to expand the screening of LV structural cardiac disorders to low-resource settings with limited access to hospital-grade equipment.
“with significant potential to expand the screening of LV structural cardiac disorders to low-resource settings with limited access to hospital-grade equipment.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
InterpretationreasoningWearable devices allow obtaining ECGs that are more accessible and allow for community-wide screening.
“Wearable devices allow obtaining ECGs that are more accessible and allow for community-wide screening, an important next step in the early detection of common and rare cardiomyopathies.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe approach avoids the unnecessary exclusion of collected data, which increases its generalizability across different device platforms.
“Additionally, this approach avoids the unnecessary exclusion of collected data, which increases its generalizability across different device platforms”
Quote in the paperSubstantial claim
The paper relies on this without support.
There is a lack of publicly available wearable device ECG datasets, making training models using wearable ECG data directly challenging.
“Due to the lack of publicly available wearable device ECG datasets, training models using wearable ECG data directly is challenging.”
The paper relies on this without support.
In the absence of large, labelled datasets of wearable ECGs, the development of algorithms that can detect underlying structural heart disease on wearable devices relies on single-lead information specifically adapted from 12-lead ECGs extracted from clinical ECG libraries.
“In the absence of large, labelled datasets of wearable ECGs, the development of algorithms that can detect underlying structural heart disease on wearable devices relies on single-lead information specifically adapted from 12-lead ECGs extracted from clinical ECG libraries.”
The paper relies on this without support.
AI-ECG algorithms have been designed in clinically obtained 12-lead ECGs.
“Even though AI-ECG is a promising screening tool for detecting LVSD, the algorithms have been designed in clinically obtained 12-lead ECGs.”
The paper relies on this without support.
The intended broader real-world use of this algorithm is as a screening method for LV systolic dysfunction among individuals with no clinical disease.
“this population differs from the intended broader real-world use of this algorithm as a screening method for LV systolic dysfunction among individuals with no clinical disease”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Previously, applications of artificial intelligence for electrocardiograms have shown potential to detect signatures of structural heart diseases.
“Previously applications of artificial intelligence for electrocardiograms (AI-ECG) have shown potential to detect signatures of SHDs.10-18”
From Ensemble Deep Learning Algorithm for Structural Heart Disease Screening Using Electrocardiographic Images (2025) · cites this paper as [14]
Recent work using 1-lead ECGs to detect cardiomyopathies supports the notion of scalability for HF screening.
“Recent work using 1-lead ECGs to detect cardiomyopathies supports the notion of scalability for HF screening.”
From Unlocking Tomorrow’s Health Care: Expanding the Clinical Scope of Wearables by Applying Artificial Intelligence (2024) · cites this paper as [33]
The algorithm in question showed robust performance even when the clinical ECG signal was augmented with typical noises from 1 wearable ECG sensor.
“The algorithm in question showed robust performance even when the clinical ECG signal was augmented with typical noises from 1 wearable ECG sensor.”
From Unlocking Tomorrow’s Health Care: Expanding the Clinical Scope of Wearables by Applying Artificial Intelligence (2024) · cites this paper as [33]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Standard-model sensitivity and specificity are reversed in the text
MinorThe text reports sensitivity of 0.69 and specificity of 0.91, while Table 2 reports sensitivity of 0.910 and specificity of 0.688 for the standard model without noise.
“The standard model had sensitivity and specificity of 0.69 and 0.91, respectively, and a PPV and NPV of 0.21 and 0.99, respectively.”
“| Standard | None | N/A | 0.207 | 0.988 | 0.688 | 0.910 | 0.895 (0.884–0.905) | 0.475 |”
Cites
Batch Normalization: Accelerating Deep Network Training by Reducing Internal Covariate Shift
Ioffe, S., & Szegedy, C. (2015) · arXiv · doi:10.48550/arxiv.1502.03167 · Cited by 27,818 · see citing works
DeLong, E. R., DeLong, D. M., & Clarke-Pearson, D. L. (1988) · Biometrics · doi:10.2307/2531595 · Cited by 22,649 · see citing works
UMAP: Uniform Manifold Approximation and Projection
McInnes, L., Healy, J., Saul, N., & Großberger, L. (2018) · Journal of Open Source Software · doi:10.21105/joss.00861 · Cited by 9,826 · see citing works
UMAP: Uniform Manifold Approximation and Projection for Dimension Reduction
McInnes, L., Healy, J., & Melville, J. (2018) · arXiv · doi:10.48550/arxiv.1802.03426 · Cited by 7,101 · see citing works
Improving neural networks by preventing co-adaptation of feature detectors
Hinton, G. E., Srivastava, N., Krizhevsky, A., Sutskever, I., & Salakhutdinov, R. R. (2012) · arXiv · doi:10.48550/arxiv.1207.0580 · Cited by 6,382 · see citing works
Heidenreich, P. A., Bozkurt, B., Aguilar, D., Allen, L. A., Byun, J. J., Colvin, M. M., Deswal, A., Drazner, M. H., Dunlay, S. M., Evers, L. R., et al. (2022) · Circulation · doi:10.1161/cir.0000000000001063 · Cited by 4,922 · cited 4× · see citing works
“Though the LVEF threshold of 40% was selected due to its therapeutic implications 4 , it is possible that the model performance among patients with an LVEF near to this cutoff differed compared with those individuals with LVEF signi fi cantly higher or lower than 40%.”
BackgroundMethodInfluentialClass-Balanced Loss Based on Effective Number of Samples
Cui, Y., Jia, M., Lin, T.-Y., Song, Y., & Belongie, S. (2019) · 2019 IEEE/CVF Conference on Computer Vision and Pattern Recognition (CVPR) · doi:10.1109/cvpr.2019.00949 · Cited by 3,094 · see citing works
Attia, Z. I., Kapa, S., Lopez-Jimenez, F., McKie, P. M., Ladewig, D. J., Satam, G., Pellikka, P. A., Enriquez-Sarano, M., Noseworthy, P. A., Munger, T. M., et al. (2019) · Nature Medicine · doi:10.1038/s41591-018-0240-2 · Cited by 1,389 · see citing works
A comparison of the noise sensitivity of nine QRS detection algorithms
Friesen, G. M., Jannett, T. C., Jadallah, M. A., Yates, S. L., Quint, S. R., & Nagle, H. T. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.43620 · Cited by 1,043 · see citing works
Sun, X., & Xu, W. (2014) · IEEE Signal Processing Letters · doi:10.1109/lsp.2014.2337313 · Cited by 638 · see citing works
Natural history of asymptomatic left ventricular systolic dysfunction in the community
Wang, T. J., Evans, J. C., Benjamin, E. J., Levy, D., LeRoy, E. C., & Vasan, R. S. (2003) · Circulation · doi:10.1161/01.cir.0000085166.44904.79 · Cited by 633 · cited 1× · see citing works
“This architecture consisted of a (5000, 1, 1) input layer, corresponding to a 10-s, 500 Hz, Lead I ECG, followed by seven two-dimensional convolutional layers, each of which were followed by a batch normalization layer(31), ReLU activation layer, and a two-dimensional max-pooling layer.”
MethodVasan, R. S., Benjamin, E. J., Larson, M. G., Leip, E. P., Wang, T. J., Wilson, P. W. F., & Levy, D. (2002) · JAMA · doi:10.1001/jama.288.10.1252 · Cited by 493 · see citing works
A Review of Signal Processing Techniques for Electrocardiogram Signal Quality Assessment
Satija, U., Ramkumar, B., & Manikandan, M. S. (2018) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2018.2810957 · Cited by 295 · see citing works
Johnson, D. H. (2006) · Scholarpedia · doi:10.4249/scholarpedia.2088 · Cited by 284 · see citing works
Jenkins, C., Moir, S., Chan, J., Rakhit, D., Haluska, B., & Marwick, T. H. (2008) · European Heart Journal · doi:10.1093/eurheartj/ehn484 · Cited by 253 · see citing works
The WATCH AF trial: SmartWATCHes for detection of atrial fibrillation
Dörr, M., Nohturfft, V., Brasier, N., Bosshard, E., Djurdjevic, A., Gross, S., Raichle, C. J., Rhinisperger, M., Stöckli, R., & Eckstein, J. (2019) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2018.10.006 · Cited by 245 · see citing works
Pellikka, P. A., She, L., Holly, T. A., Lin, G., Varadarajan, P., Pai, R. G., Bonow, R. O., Pohost, G. M., Panza, J. A., Berman, D. S., et al. (2018) · JAMA Network Open · doi:10.1001/jamanetworkopen.2018.1456 · Cited by 238 · see citing works
The Epidemiology of Asymptomatic Left Ventricular Systolic Dysfunction: Implications for Screening
Wang, T. J., Levy, D., Benjamin, E. J., & Vasan, R. S. (2003) · Annals of Internal Medicine · doi:10.7326/0003-4819-138-11-200306030-00012 · Cited by 229 · see citing works
Clinical Validation of 5 Direct-to-Consumer Wearable Smart Devices to Detect Atrial Fibrillation
Mannhart, D., Lischer, M., Knecht, S., du Fay de Lavallaz, J., Strebel, I., Serban, T., Vögeli, D., Schaer, B., Osswald, S., Mueller, C., et al. (2023) · JACC: Clinical Electrophysiology · doi:10.1016/j.jacep.2022.09.011 · Cited by 175 · see citing works
Bachtiger, P., Petri, C. F., Scott, F. E., Ri Park, S., Kelshiker, M. A., Sahemey, H. K., Dumea, B., Alquero, R., Padam, P. S., Hatrick, I. R., et al. (2022) · The Lancet Digital Health · doi:10.1016/s2589-7500(21)00256-9 · Cited by 167 · see citing works
Prospective evaluation of smartwatch-enabled detection of left ventricular dysfunction
Attia, Z. I., Harmon, D. M., Dugan, J., Manka, L., Lopez-Jimenez, F., Lerman, A., Siontis, K. C., Noseworthy, P. A., Yao, X., Klavetter, E. W., et al. (2022) · Nature Medicine · doi:10.1038/s41591-022-02053-1 · Cited by 154 · see citing works
Detection of Left Ventricular Systolic Dysfunction From Electrocardiographic Images
Sangha, V., Nargesi, A. A., Dhingra, L. S., Khunte, A., Mortazavi, B. J., Ribeiro, A. H., Banina, E., Adeola, O., Garg, N., Brandt, C. A., et al. (2023) · Circulation · doi:10.1161/circulationaha.122.062646 · Cited by 150 · see citing works
Self-citationDetection of Left Ventricular Systolic Dysfunction from Electrocardiographic Images
Sangha, V., Nargesi, A. A., Dhingra, L. S., Khunte, A., Mortazavi, B. J., Ribeiro, A. H., Banina, E., Adeola, O., Garg, N., Brandt, C. A., et al. (2022) · doi:10.1101/2022.06.04.22276000 · Cited by 150
Self-citationClass-Balanced Loss Based on Effective Number of SamplesPreprint
Cui, Y., Jia, M., Lin, T.-Y., et al. (2019) · arXiv (Cornell University) · doi:10.48550/arxiv.1901.05555 · Cited by 128 · see citing works
Satija, U., Ramkumar, B., & Manikandan, M. S. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2877055 · Cited by 102 · see citing works
Galasko, G. I. W., Barnes, S. C., Collinson, P., Lahiri, A., & Senior, R. (2006) · European Heart Journal · doi:10.1093/eurheartj/ehi559 · Cited by 94 · see citing works
The Use of Wearable ECG Devices in the Clinical Setting: A Review
Kamga, P., Mostafa, R., & Zafar, S. (2022) · Current Emergency and Hospital Medicine Reports · doi:10.1007/s40138-022-00248-x · Cited by 94 · see citing works
(1999) · PubMed · Cited by 77 · see citing works
Srivastava, P. K., DeVore, A. D., Hellkamp, A. S., Thomas, L., Albert, N. M., Butler, J., Patterson, J. H., Spertus, J. A., Williams, F. B., Duffy, C. I., et al. (2021) · JACC: Heart Failure · doi:10.1016/j.jchf.2020.08.017 · Cited by 71 · see citing works
Attia, Z. I., Dugan, J., Rideout, A., Maidens, J. N., Venkatraman, S., Guo, L., Noseworthy, P. A., Pellikka, P. A., Pham, S. L., Kapa, S., et al. (2022) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztac030 · Cited by 46 · see citing works
Efficient QRS complex detection algorithm based on Fast Fourier Transform
Kumar, A., Ranganatham, R., Komaragiri, R., et al. (2018) · Biomedical Engineering Letters · doi:10.1007/s13534-018-0087-y · Cited by 45 · see citing works
Screening for Asymptomatic Left Ventricular Dysfunction Using B‐Type Natriuretic Peptide
McDonagh, T. A., McDonald, K., & Maisel, A. S. (2008) · Congestive Heart Failure · doi:10.1111/j.1751-7133.2008.tb00002.x · Cited by 35 · see citing works
A simple method for detection and classification of ECG noises for wearable ECG monitoring devices
Satija, U., Ramkumar, B., & Manikandan, M. S. (2015) · 2015 2nd International Conference on Signal Processing and Integrated Networks (SPIN) · doi:10.1109/spin.2015.7095425 · Cited by 34 · see citing works
Duarte, R., Stainthorpe, A., Mahon, J., Greenhalgh, J., Richardson, M., Nevitt, S., Kotas, E., Boland, A., Thom, H., Marshall, T., et al. (2019) · PLOS ONE · doi:10.1371/journal.pone.0226671 · Cited by 28 · see citing works
Wolfe, N. K., Mitchell, J. D., & Brown, D. L. (2021) · European Heart Journal - Quality of Care and Clinical Outcomes · doi:10.1093/ehjqcco/qcaa032 · Cited by 27 · see citing works
Abstract 14057: Detection of Left Ventricular Systolic Dysfunction From Electrocardiographic Images
Sangha, V., Nargesi, A. A., Dhingra, L. S., Mortazavi, B. J., Ribeiro, A. H., Banina, E., Brandt, C., Miller, E. J., Ribeiro, A. L. P., Velazquez, E. J., et al. (2022) · Circulation · doi:10.1161/circ.146.suppl_1.14057 · Cited by 2 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 1
- Moody, G. B., Muldrow, W. & Mark, R. A noise stress test for arrhythmia detectors. Comput. Cardiol. 11, 381–384 (1984).
Cited by
Sorted by citations
Transforming Cardiovascular Care With Artificial Intelligence: From Discovery to Practice
Khera, R., Oikonomou, E. K., Nadkarni, G. N., Morley, J., Wiens, J., Butte, A. J., & Topol, E. J. (2024) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2024.05.003 · Cited by 309 · see citing works
Self-citationMachine learning in precision diabetes care and cardiovascular risk prediction
Oikonomou, E. K., & Khera, R. (2023) · Cardiovascular Diabetology · doi:10.1186/s12933-023-01985-3 · Cited by 190 · see citing works
Self-citationCardiovascular care with digital twin technology in the era of generative artificial intelligence
Thangaraj, P., Benson, S., Oikonomou, E. K., Asselbergs, F. W., & Khera, R. (2024) · European Heart Journal · doi:10.1093/eurheartj/ehae619 · Cited by 185 · see citing works
Self-citationArtificial Intelligence for Cardiovascular Care—Part 1: Advances
Elias, P. A., Jain, S. S., Poterucha, T. J., Randazzo, M. J., Jimenez, F. L., Khera, R., Perez, M. V., Ouyang, D., Pirruccello, J. P., Salerno, M., et al. (2024) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2024.03.400 · Cited by 134 · see citing works
Self-citationKhera, R., Asnani, A. H., Krive, J., Addison, D., Zhu, H., Vasbinder, A. L., Fleming, M. R., Arnaout, R., Razavi, P., & Okwuosa, T. M. (2025) · Circulation Genomic and Precision Medicine · doi:10.1161/hcg.0000000000000097 · Cited by 77 · see citing works
Self-citationLi, H., Le, J., Tan, H., et al. (2024) · Advanced Materials · doi:10.1002/adma.202411989 · Cited by 72 · see citing works
Artificial Intelligence–Enabled Prediction of Heart Failure Risk From Single-Lead Electrocardiograms
Dhingra, L. S., Aminorroaya, A., Pedroso, A. F., Khunte, A., Sangha, V., McIntyre, D., Chow, C. K., Asselbergs, F. W., Brant, L. C. C., Barreto, S. M., et al. (2025) · JAMA Cardiology · doi:10.1001/jamacardio.2025.0492 · Cited by 65 · see citing works
Self-citationDhingra, L. S., Aminorroaya, A., Sangha, V., Pedroso, A. F., Asselbergs, F. W., Brant, L. C. C., Barreto, S. M., Ribeiro, A. L. P., Krumholz, H. M., Oikonomou, E. K., et al. (2025) · European Heart Journal · doi:10.1093/eurheartj/ehae914 · Cited by 64 · see citing works
Self-citationDhingra, L. S., Aminorroaya, A., Sangha, V., Pedroso, A. F., Shankar, S. V., Coppi, A., Foppa, M., Brant, L. C. C., Barreto, S. M., Ribeiro, A. L. P., et al. (2025) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2025.01.030 · Cited by 53 · see citing works
Self-citationDing, C., Yao, T., Wu, C., & Ni, J. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2024.117073 · Cited by 52 · see citing works
Stamate, E., Piraianu, A. I., Ciobotaru, O. R., Crassas, R., Duca, O.-M., Fulga, A., Grigore, I., Vintila, V., Fulga, I., & Ciobotaru, O. C. (2024) · Diagnostics · doi:10.3390/diagnostics14111103 · Cited by 49 · see citing works
Pedroso, A. F., & Khera, R. (2025) · npj Cardiovascular Health · doi:10.1038/s44325-025-00071-9 · Cited by 46 · see citing works
Self-citationCardiovascular Care Innovation through Data-Driven Discoveries in the Electronic Health Record
Dhingra, L. S., Shen, M., Mangla, A., & Khera, R. (2023) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2023.06.104 · Cited by 40 · see citing works
Self-citationArtificial intelligence-enhanced patient evaluation: bridging art and science
Oikonomou, E. K., & Khera, R. (2024) · European Heart Journal · doi:10.1093/eurheartj/ehae415 · Cited by 34 · see citing works
Self-citationMarvasti, T. B., Gao, Y., Murray, K. R., Hershman, S., McIntosh, C., & Moayedi, Y. (2024) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2024.07.009 · Cited by 35 · see citing works
Jia, H., Gao, Y., Zhou, J., Li, J., Yiu, C. K., Park, W., Yang, Z., & Yu, X. (2024) · Nano Energy · doi:10.1016/j.nanoen.2024.109796 · Cited by 34 · see citing works
Biometric contrastive learning for data-efficient deep learning from electrocardiographic images
Sangha, V., Khunte, A., Holste, G., Mortazavi, B. J., Wang, Z., Oikonomou, E. K., & Khera, R. (2024) · Journal of the American Medical Informatics Association · doi:10.1093/jamia/ocae002 · Cited by 32 · see citing works
Self-citationAminorroaya, A., Dhingra, L. S., Pedroso, A. F., Shankar, S. V., Coppi, A. C., Khunte, A., Foppa, M., Brant, L. C., Barreto, S. M., Ribeiro, A. L. P., et al. (2025) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztaf034 · Cited by 29 · see citing works
Self-citationCoatings on Lithium Battery Separators: A Strategy to Inhibit Lithium Dendrites Growth
Cheng, H., Tan, R., Li, J., et al. (2023) · Molecules · doi:10.3390/molecules28237788 · Cited by 28 · see citing works
Artificial Intelligence in Diagnosis of Heart Failure
Xie, Y., Zhang, L., Sun, W., Zhu, Y. C., Zhang, Z., Chen, L., Xie, M., & Zhang, L. (2025) · Journal of the American Heart Association · doi:10.1161/jaha.124.039511 · Cited by 27 · see citing works
Oikonomou, E. K., & Khera, R. (2024) · Hellenic Journal of Cardiology · doi:10.1016/j.hjc.2024.07.003 · Cited by 26 · see citing works
Self-citationArtificial intelligence in cardiovascular pharmacotherapy: applications and perspectives
Costa, F., Gómez‐Doblas, J. J., Expósito, A. D., Adamo, M., D’Ascenzo, F., Kołtowski, Ł., Saba, L., Mendieta, G., Gragnano, F., Calabrò, P., et al. (2025) · European Heart Journal · doi:10.1093/eurheartj/ehaf474 · Cited by 26 · see citing works
Barriers to translating continuous monitoring technologies for preventative medicine
Chen, J., Jastrzebska‐Perfect, P., Chai, P. R., Say, M. G., Tu, J., Gao, W. F., Halperin, F., Korzenik, J., Huang, H., Katabi, D., et al. (2025) · Nature Biomedical Engineering · doi:10.1038/s41551-025-01520-7 · Cited by 25 · see citing works
Artificial intelligence and digital tools for design and execution of cardiovascular clinical trials
Hu, J., Power, J. R., Zannad, F. M., & Lam, C. S. P. (2024) · European Heart Journal · doi:10.1093/eurheartj/ehae794 · Cited by 23 · see citing works
Sau, A., Zeidaabadi, B., Patlatzoglou, K., Pastika, L., Ribeiro, A. H., Sabino, É. C., Peters, N. S., Ribeiro, A. L. P., Kramer, D. B., Waks, J. W., et al. (2024) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztae090 · Cited by 20 · see citing works
Automated Detection of Reduced Ejection Fraction Using an ECG-Enabled Digital Stethoscope
Guo, L. T., Pressman, G. S., Kieu, S. N., Marrus, S. B., Mathew, G., Prince, J., Lastowski, E., McDonough, R. V., Currie, C., Tiwari, U., et al. (2025) · JACC Advances · doi:10.1016/j.jacadv.2025.101619 · Cited by 17 · see citing works
Shankar, S. V., Dhingra, L. S., Aminorroaya, A., Adejumo, P., Nadkarni, G. N., Xu, H., Brandt, C., Oikonomou, E. K., Pedroso, A. F., & Khera, R. (2025) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztaf030 · Cited by 17 · see citing works
Self-citationArtificial Intelligence in the Management of Heart Failure
Cheema, B. S., Hourmozdi, J., Kline, A. S., Ahmad, F. S., & Khera, R. (2025) · Journal of Cardiac Failure · doi:10.1016/j.cardfail.2025.02.020 · Cited by 16 · see citing works
Self-citationWearable devices and cardiovascular health: revolutionizing remote monitoring and disease prevention
Hughes, A. M., Taylor, D. J., Morris, P. D., & Brittain, E. L. (2026) · European Heart Journal · doi:10.1093/eurheartj/ehag189 · Cited by 23 · see citing works
A single-lead ECG based cardiotoxicity detection in patients on polychemotherapy
Фератовна, М. Д., Fashafsha, Z. Z. A., Poltavskaya, M. G., et al. (2024) · IJC Heart & Vasculature · doi:10.1016/j.ijcha.2024.101336 · Cited by 15 · see citing works
Artificial Intelligence for Cardiovascular Care in Action
Biswas, D., Thangaraj, P., Pedroso, A. F., Ogunniyi, M. O., Maddox, T. M., Ahmad, F. S., & Khera, R. (2025) · JACC Advances · doi:10.1016/j.jacadv.2025.102307 · Cited by 15 · see citing works
Self-citationDeep Learning-enabled Detection of Aortic Stenosis from Noisy Single Lead ElectrocardiogramsPreprint
Aminorroaya, A., Dhingra, L. S., Sangha, V., Oikonomou, E. K., Khunte, A., Shankar, S. V., Pedroso, A. F., Haynes, N., Hofer, I. S., Ouyang, D., et al. (2023) · medRxiv · doi:10.1101/2023.09.29.23296310 · Cited by 13 · see citing works
Self-citationClinical care of family members of patients with dilated cardiomyopathy
Verdonschot, J. A. J., Kaski, J. P., Asselbergs, F. W., Behr, E. R., Charron, P., Dawson, D. K., Haugaa, K. H., Kuchynka, P., Lopes, L. R., Mazzanti, A., et al. (2025) · European Heart Journal · doi:10.1093/eurheartj/ehaf571 · Cited by 14 · see citing works
Scalia, I. G., Pathangey, G., Abdelnabi, M. H., Ibrahim, O., Abdelfattah, F. E., Pereyra, M., Ibrahim, R., Farina, J., Banerjee, I., Tamarappoo, B. K., et al. (2025) · Cancers · doi:10.3390/cancers17040605 · Cited by 12 · see citing works
Shankar, S. V., Oikonomou, E. K., & Khera, R. (2023) · medRxiv · doi:10.1101/2023.10.02.23296404 · Cited by 11 · see citing works
Self-citationCroon, P. M., Boonstra, M., Allaart, C. P., Arends, B. K. O., Dhingra, L. S., Huang, Y.-C., Mast, T. P., Khera, R., Kuo, C.-F., Kwon, J., et al. (2026) · JACC Advances · doi:10.1016/j.jacadv.2025.102572 · Cited by 11 · see citing works
Self-citationKim, S., Lim, J., & Jang, J. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3411056 · Cited by 10 · see citing works
Dhingra, L. S., Aminorroaya, A., Sangha, V., Pedroso, A. F., Asselbergs, F. W., Brant, L. C. C., Barreto, S. M., Ribeiro, A. L. P., Krumholz, H. M., Oikonomou, E. K., et al. (2024) · medRxiv · doi:10.1101/2024.04.02.24305232 · Cited by 9 · see citing works
Self-citationAdvancements in Incident Heart Failure Risk Prediction and Screening Tools
Matasic, D. S., Zeitoun, R., Fonarow, G. C., Razavi, A. C., Blumenthal, R. S., & Gulati, M. (2024) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2024.07.014 · Cited by 8 · see citing works
Zhu, Y., Luo, D., Shen, X., Shi, Q., Lv, H., Zhang, S., Ye, F., & Kong, N. (2025) · BMC Medical Informatics and Decision Making · doi:10.1186/s12911-025-03114-4 · Cited by 8 · see citing works
Aminorroaya, A., Dhingra, L. S., Pedroso, A. F., Shankar, S. V., Khunte, A., Sangha, V., Sen, S., McNamara, R. L., Haynes, N., Oikonomou, E. K., et al. (2024) · medRxiv · doi:10.1101/2024.03.18.24304477 · Cited by 7 · see citing works
Self-citationDhingra, L. S., Aminorroaya, A., Sangha, V., Pedroso, A. F., Shankar, S. V., Coppi, A. C., Foppa, M., Brant, L. C. C., Barreto, S. M., Ribeiro, A. L. P., et al. (2024) · medRxiv · doi:10.1101/2024.10.06.24314939 · Cited by 7 · see citing works
Self-citationLeveraging the full potential of wearable devices in cardiomyopathies
Oikonomou, E. K., & Khera, R. (2024) · Journal of Cardiac Failure · doi:10.1016/j.cardfail.2024.02.011 · Cited by 6 · see citing works
Self-citationTiwari, A., Shah, P. C., Kumar, H., Borse, T., Arun, A. R., Chekragari, M., Ochani, S., Shah, Y. R., Ganesh, A., Ahmed, R., et al. (2025) · Current Cardiology Reviews · doi:10.2174/011573403x351048250329170744 · Cited by 8 · see citing works
Khurshid, S. S., Friedman, S. F., Al‐Alusi, M. A., Kany, S., Sommers, T. F., Anderson, C. D., Ho, J. E., McManus, D. D., Borowsky, L. H., Ashburner, J. M., et al. (2025) · npj Digital Medicine · doi:10.1038/s41746-025-02164-2 · Cited by 7 · see citing works
Leveraging ECG images for predicting ejection fraction using machine learning algorithms
Swamy, A. K., Rajagopal, V., Krishnan, D., et al. (2025) · Indian Heart Journal · doi:10.1016/j.ihj.2025.03.009 · Cited by 4 · see citing works
Interpretable manifold learning for T-wave alternans assessment with electrocardiographic imaging
Sánchez-Carballo, E., Melgarejo-Meseguer, F.-M., Vijayakumar, R., et al. (2025) · Engineering Applications of Artificial Intelligence · doi:10.1016/j.engappai.2024.109996 · Cited by 4 · see citing works
Park, S., Lee, H. J., Song, S., Woo, K., Kim, J., Kim, J., Kim, J. Y., Park, S., On, Y. K., & Park, K. (2025) · Journal of Clinical Medicine · doi:10.3390/jcm14124257 · Cited by 4 · see citing works
Summertime for Cardiovascular AI
Khera, R., & Wiens, J. (2024) · Circulation Cardiovascular Quality and Outcomes · doi:10.1161/circoutcomes.123.010404 · Cited by 4 · see citing works
Self-citationZeidaabadi, B., Patlatzoglou, K., Barker, J., Pastika, L., Khattak, G. R., Gurnani, M., Machado, X. D. S. A., Peters, N. S., Kramer, D. B., Waks, J. W., et al. (2025) · Heart Rhythm · doi:10.1016/j.hrthm.2025.10.024 · Cited by 3 · see citing works
Biswas, D., Aminorroaya, A., Croon, P. M., Batinica, B., Pedroso, A. F., & Khera, R. (2025) · Current Atherosclerosis Reports · doi:10.1007/s11883-025-01337-4 · Cited by 3 · see citing works
Self-citationSynthetic generation of cardiac tissue motion from surface electrocardiograms
Radhakrishnan, A., Yanamala, N. V. K., Jamthikar, A. D., Wang, Y., East, S.-A., Hamirani, Y. S., Maganti, K., & Sengupta, P. P. (2025) · Nature Cardiovascular Research · doi:10.1038/s44161-025-00629-x · Cited by 5 · see citing works
Khera, R. (2024) · Heart · doi:10.1136/heartjnl-2024-324299 · Cited by 3 · see citing works
Self-citationPande, A., Kumar, A., & Anjankar, A. P. (2025) · Multidisciplinary Reviews · doi:10.31893/multirev.2025218 · Cited by 2 · see citing works
Flexible Bioelectrodes-Integrated Miniaturized System for Unconstrained ECG Monitoring
Zhan, Y., Wang, X., & Yang, J. (2025) · Sensors · doi:10.3390/s25134213 · Cited by 2 · see citing works
Brar, S. S., Yathindra, M. R., Arango, J. S. A., Gutierrez, E. A., Aldoohan, F., Chahal, P. S., Youssef, A., Narra, M. S., Tatineni, M. B., Saldaña, J. M. A., et al. (2025) · Current Cardiology Reviews · doi:10.2174/011573403x369978250818060357 · Cited by 2 · see citing works
Safdar, M. F., Nowak, R. M., Pałka, P., et al. (2025) · Scientific Reports · doi:10.1038/s41598-025-18910-1 · Cited by 2 · see citing works
Hempel, P., Steinbrinker, T. F. A., Graf, L., et al. (2025) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc58623.2025.11254282 · Cited by 2 · see citing works
An in-sensor communication electronic textile for imperceptible and ultrarobust silent speech
Lin, Y., Yu, S., Liu, L., Wang, H., Wang, Y., Jiang, M., Su, G., Yu, Y., Chen, H., Luo, Y., et al. (2026) · Nature Communications · doi:10.1038/s41467-026-74035-7 · Cited by 2 · see citing works
Boshu, 孙. S., Chunlei, 孙. S., Tang, R., Yiting, 王. W., & Li, L. (2024) · ACTA PHOTONICA SINICA · doi:10.3788/gzxb20245305.0553102 · Cited by 2 · see citing works
Augmenting reality in echocardiography
Sangha, V. (2023) · Heart · doi:10.1136/heartjnl-2023-323443 · Cited by 2 · see citing works
Self-citationPrinciples of telehealth and telemedicine in cardio-oncology
Saleh, A., Nona, P. J., Price, E. B., Rickards, J., & Andrikopoulou, E. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-29829-5.00027-x · Cited by 1 · see citing works
Arends, B. K. O., Zwetsloot, P.-P. M., Heeres, P., van Amsterdam, W. A. C., Cramer, M. J., Reenen, E. T. K., van der Harst, P., van Osch, D., & van Es, R. (2025) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2025.154048 · Cited by 1 · see citing works
Divya, N. J., Kumar, N. S., & Devi, R. K. (2025) · Transactions on Emerging Telecommunications Technologies · doi:10.1002/ett.70229 · Cited by 1 · see citing works
Batinica, B., Oikonomou, E. K., Pedroso, A. F., Aminorroaya, A., Biswas, D., Barreto, S. M., Brant, L. C. C., Ribeiro, A. L. P., Dhingra, L. S., & Khera, R. (2026) · American Journal of Preventive Cardiology · doi:10.1016/j.ajpc.2026.101794 · Cited by 1 · see citing works
Self-citationDhingra, L. S., Croon, P. M., Batinica, B., Aminorroaya, A., Pedroso, A. F., & Khera, R. (2026) · Current Heart Failure Reports · doi:10.1007/s11897-026-00748-x · Cited by 2 · see citing works
Self-citationWearable-Echo-FM: an ECG echo foundation model for 1-lead electrocardiography
Knight, E., Oikonomou, E. K., Aminorroaya, A., Pedroso, A. R., & Khera, R. (2026) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztag049 · Cited by 2 · see citing works
Self-citationArtificial Intelligence in Heart Failure with Preserved Ejection Fraction
Li, X., Xu, C., Deng, W., Zhang, Y., Liu, C. X., Gao, L. L., Ji, M., He, Q., Wu, Z., Qin, S., et al. (2026) · Diagnostics · doi:10.3390/diagnostics16111597 · Cited by 1 · see citing works
Pre-Heart Failure Risk Assessment: Don't Get Lost in an Echo Chamber!
Khan, S. S., & Shah, S. J. (2023) · Journal of Cardiac Failure · doi:10.1016/j.cardfail.2023.07.007 · Cited by 1 · see citing works
Antoniades, C., & Chan, K. (2025) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztaf048 · Cited by 1 · see citing works
Croon, P. M., Boonstra, M., Allaart, C. P., Arends, B. K. O., Dhingra, L. S., Huang, Y., Mast, T. P., Khera, R., Kuo, C., Kwon, J., et al. (2025) · medRxiv · doi:10.1101/2025.07.08.25331129 · Cited by 1 · see citing works
Self-citationZhang, H., Zhang, P., Ma, F., Chao, L., Shan, W., & Li, Q. (2024) · International Conference on Bioinformatics and Computational Biology (ICBCB) · doi:10.1109/icbcb61507.2024.11012075
Wearable-Echo-FM: An ECG-echo foundation model for single lead electrocardiographyPreprint
Knight, E., Oikonomou, E. K., Aminorroaya, A., Pedroso, A. F., & Khera, R. (2025) · medRxiv · doi:10.1101/2025.06.10.25329163
Self-citationHigh-quality triage and diagnosis of gynecological diseases via artificial intelligence
Fu, L., Wang, C., Liu, Z., Pan, C., Du, Z., Sun, Z., Zhu, L., & Deng, K. (2025) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2025.103267
Carbon-Based Nanosensors for the Detection of Pollutants
Bibi, I., Rauf, I., Shad, N. A., et al. (2025) · Lecture notes in nanoscale science and technology · doi:10.1007/978-3-031-96895-2_12
Kılıç, M. E., Arayıcı, M. E., Akbilgiç, O., & Yılmaz, M. B. (2025) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2025.12.022
Nakano, T., Okayama, K., Yasuoka, M., Shigeta, H., Kawamura, A., Sekihara, T., Ozu, K., Oka, T., Seno, S., & Sakata, Y. (2026) · doi:10.2196/preprints.91960
Borde, D. P., Joshi, S. S., Chidambaram, K., Jasapara, A., Thimmarayappa, A., Devarakonda, B. V., Kamat, S. S., Gandhe, U., Muniraj, G., Madanaik, S. S., et al. (2026) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2026.03.002
AI-enabled clinically aligned fewer-lead ECG approaches for detecting cardiovascular diseases
Sikandar, T., Hasan, M. T., Azad, A. K. M., Khan, A., & Moni, M. A. (2026) · Artificial Intelligence Review · doi:10.1007/s10462-026-11577-8
Emerging Artificial Intelligence Tools for the Screening of Structural and Valvular Heart Disease
Abbaoui, Y., Nolin-Lapalme, A., Morisset, J., Adib, I. E., Généreux, P., Poterucha, T. J., Elias, P. A., Yao, X., & Avram, R. (2026) · Current Heart Failure Reports · doi:10.1007/s11897-026-00757-w
Croon, P. M., Dhingra, L. S., Pedroso, A. F., & Khera, R. (2026) · Current Heart Failure Reports · doi:10.1007/s11897-026-00764-x · Cited by 1 · see citing works
Self-citationAlvaro, G., & Tjahyadi, H. (2026) · International Conference on Current Research in Artificial Intelligence and Data Science (ICCRAIDS) · doi:10.1109/iccraids67816.2026.11519685
AI in cardiovascular disease classification—a systematic literature review
Karandikar, M., Sawant, S., Kumbhar, S., & Bidwe, R. V. (2026) · Artificial Intelligence Review · doi:10.1007/s10462-026-11592-9
Prasad, K., Swamy, A. K., Ghorai, P. A., Umredkar, P. N., Padmanabhan, D., Rajagopal, V., & Narayan, P. (2026) · Indian Journal of Thoracic and Cardiovascular Surgery · doi:10.1007/s12055-026-02226-4
Su, H., Wang, S., Wang, H., & Qiu, K. (2026) · Sensors · doi:10.3390/s26123753
Nakano, T., Okayama, K., Yasuoka, M., Shigeta, H., Kawamura, A., Sekihara, T., Ozu, K., Oka, T., Seno, S., & Sakata, Y. (2026) · JMIR Medical Informatics · doi:10.2196/91960
Advancements in wearable technology for heart failure patients
Perlman, G., Leroux, J., & Sharma, A. (2026) · Current Opinion in Cardiology · doi:10.1097/hco.0000000000001326
Lin, C., Liu, W., Chen, K., Tsai, D., Chang, D.-W., Tsai, T.-N., Cheng, C.-C., Lin, C.-Y., Chen, Y.-H., Chien-Sung, T., et al. (2026) · npj Digital Medicine · doi:10.1038/s41746-026-02919-5
Kattamuri, L. P. V., Sharma, K., & Mukherjee, D. (2026) · International Journal of Angiology · doi:10.1055/a-2904-8847
Alavi, R., Li, J., Matthews, R. V., Pahlevan, N. M., Kloner, R. A., & Gharib, M. (2026) · medRxiv · doi:10.64898/2026.07.08.26357521
Tom-Ayegunle, K., Oine, A. A., & Oche, I. D. (2026) · International Journal of Scientific Research and Modern Technology. · doi:10.38124/ijsrmt.v5i7.1575
Batinica, B., Pedroso, A. F., & Khera, R. (2026) · Journal of the American Heart Association · doi:10.1161/jaha.125.046267
Self-citationECG-Based Emotion Recognition Using Beat-Level Complementary Feature Fusion
Peng, G., & Guo, Y. (2026) · Sensors · doi:10.3390/s26196169
Leveraging the wearable 1-lead ECG signal. From cardiac rhythm to cardiac function assessment
van der Valk, V., Atsma, D. E., Scherptong, R., et al. (2026) · npj Cardiovascular Health · doi:10.1038/s44325-026-00153-2
Choi, J., Seyam, O., Biswas, D., Sivalokanathan, S., Tai, H.-H., Kong, D., & Farhan, S. (2026) · Sensors · doi:10.3390/s26206503
- Authors
- Khunte, A., Sangha, V., Oikonomou, E. K., Dhingra, L. S., Aminorroaya, A., Mortazavi, B. J., Coppi, A., Brandt, C. A., Krumholz, H. M., & Khera, R.
- Affiliations
- Yale University (US)
- Yale New Haven Hospital (US)
- Corresponding author
- Khera, Rohan
- ORCID
- Venue
- npj Digital Medicine
- Published
- 2023-07-11
- Volume
- 6
- Issue
- 1
- Pages
- 124-124
- ISSN
- 2398-6352 (ISSN-L)
- PMC
- PMC10336107
- OpenAlex
- W4383911793
- Cited by
- 95
- References
- 33
- Citations by year
- 2026: 26 · 2025: 38 · 2024: 24 · 2023: 7
- FWCI
- 18.86
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- npj Digital Medicine (publishedVersion, cc-by)
- medRxiv (acceptedVersion, cc-by-nd)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Doris Duke Charitable Foundation (2022060)
- U.S. Department of Health & Human Services | NIH | National Heart, Lung, and Blood Institute (K23HL153775)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Keywords
- left ventricular systolic dysfunction, wearable devices, portable ECG, ECG noise, signal augmentation, artificial intelligence, cardiovascular screening, signal-to-noise ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-28.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-28.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review