Evaluation of Commercial Glucometer Test Strips for Potential Measurement of Glucose in Tears
The DOI matches its Crossref record; 36 of the 37 references it cites were found in a registry or an index.
Abstract
Tear glucose measurements have been suggested as a potential alternative to blood glucose monitoring for diabetic patients. While previous work has reported that there is a correlation between blood and tear glucose levels in humans, this link has not been thoroughly established and additional clinical studies are needed. Herein, we evaluate the potential of using commercial blood glucose test strips to measure glucose in tears. Of several blood glucose strips evaluated, only one brand exhibits the low detection limit required for quantitating glucose in tears. Calibration of these strips in the range of 0-100 μM glucose with an applied potential of 150 mV to the working electrode yields a sensitivity of 0.127 nA/μM and a limit of quantitation (LOQ) of 9 μM. The strips also exhibit ≤13% error (n = 3) for 25, 50, and 75 μM glucose in the presence of 10 μM acetaminophen, 100 μM ascorbic acid, and 100 μM uric acid. Measurements of glucose in tears from nine normal (nondiabetic) fasting human subjects using strips yielded glucose values within the range of 5-148 μM (mean = 47 μM, median = 43 μM), similar to those for human tears reported by others with more complex LC-MS methods. The glucometer strip method could facilitate more clinical studies to determine whether tear glucose and blood glucose levels sufficiently correlate for application to routine measurements in tears to supplement blood glucose testing. This would be especially helpful for children, adolescents, other Type 1 diabetics, and also for Type 2 diabetics who require treatment with insulin and cannot tolerate multiple finger sticks per day.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The biosensor has a detection limit of 0.18 mg/dL and shows good linearity over the typical range of glucose concentrations in a tear film (0.18-10.5 mg/dL).”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [165]
In the citing paper
“To be analytically useful, sensing techniques require a very low detection limit (since glucose is present in tear fluid at levels 50-100 times lower than blood), high analytical sensitivity and selectivity, and the ability to quantitatively measure a very small sample in a short period of time (Figure 4E) 165.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [165]
In the citing paper
“Yao et al. have integrated functional contact lenses consisting of a differential glucose biosensor module, metal interconnections, a readout circuit, an antenna, and a telecommunication circuit to monitor tear glucose levels wirelessly, continuously and non-invasively 165.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [165]
In the citing paper
“Studies designed to establish correlations between blood and tear glucose concentrations are inconclusive from both human (243-245) and animal studies (246).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [245]
Citations
Cites
Danaei, G., Finucane, M. M., Lu, Y., Singh, G. M., Cowan, M. J., Paciorek, C. J., Lin, J. K., Farzadfar, F., Khang, Y.-H., Stevens, G. A., et al. (2011) · The Lancet · doi:10.1016/s0140-6736(11)60679-x · Cited by 3,568
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498
Raman spectroscopy for noninvasive glucose measurements
Enejder, A. M. K., Scecina, T. G., Oh, J., Hunter, M., Shih, W.-C., Sasic, S., Horowitz, G. L., & Feld, M. S. (2005) · Journal of Biomedical Optics · doi:10.1117/1.1920212 · Cited by 271
Current development in non-invasive glucose monitoring
Ferrante do Amaral, C. E., & Wolf, B. (2008) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2007.06.003 · Cited by 269
Glucose monitoring by reverse iontophoresis
Potts, R. O., A. Tamada, J., & J. Tierney, M. (2002) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.210 · Cited by 243
Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus
Baca, J. T., Finegold, D. N., & Asher, S. A. (2007) · The Ocular Surface · doi:10.1016/s1542-0124(12)70094-0 · Cited by 187
Measurement of tear glucose levels with amperometric glucose biosensor/capillary tube configuration
Yan, Q., Peng, B., Su, G., Cohan, B. E., Major, T. C., & Meyerhoff, M. E. (2011) · Analytical Chemistry · doi:10.1021/ac201700c · Cited by 185
Self-citationNovel FAD-Dependent Glucose Dehydrogenase for a Dioxygen-Insensitive Glucose Biosensor
TSUJIMURA, S., KOJIMA, S., KANO, K., IKEDA, T., SATO, M., SANADA, H., & OMURA, H. (2006) · Bioscience, Biotechnology, and Biochemistry · doi:10.1271/bbb.70.654 · Cited by 175
Tear Glucose Dynamics in Diabetes Mellitus
Lane, J. D., Krumholz, D. M., Sack, R. A., & Morris, C. (2006) · Current Eye Research · doi:10.1080/02713680600976552 · Cited by 161
The Technology Behind Glucose Meters: Test Strips
Hönes, J., Müller, P., & Surridge, N. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0005 · Cited by 156
Research works printed but not linked: 2
- Choy C. K. M., 2000, Invest. Ophthalmol. Vis. Sci., 41, 3293
- Kalisz H. M, 2000, Biochem. J., 349, 553
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Electrochemical Glucose Sensing: Is There Still Room for Improvement?
Witkowska Nery, E., Kundys, M., Jeleń, P. S., & Jönsson-Niedziółka, M. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03151 · Cited by 292
Pullano, S. A., Greco, M., Bianco, M. G., Foti, D., Brunetti, A., & Fiorillo, A. S. (2022) · Theranostics · doi:10.7150/thno.64035 · Cited by 210
Gabriel, E. F. M., Garcia, P. T., Cardoso, T. M. G., Lopes, F. M., Martins, F. T., & Coltro, W. K. T. (2016) · The Analyst · doi:10.1039/c6an00430j · Cited by 204
Wen, D., Liu, W., Haubold, D., Zhu, C., Oschatz, M., Holzschuh, M., Wolf, A., Simon, F., Kaskel, S., & Eychmüller, A. (2016) · ACS Nano · doi:10.1021/acsnano.5b07505 · Cited by 207
Yang, K., Peretz-Soroka, H., Liu, Y., & Lin, F. (2016) · Lab on a Chip · doi:10.1039/c5lc01524c · Cited by 198
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 203
Organic Bioelectronic Devices for Metabolite Sensing
Koklu, A., Ohayon, D., Wustoni, S., Druet, V., Saleh, A., & Inal, S. (2021) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00395 · Cited by 148
Paper-Based Colorimetric Biosensor for Tear Glucose Measurements
Gabriel, E. F. M., Garcia, P. D. T., Lopes, F. M., & Coltro, W. K. T. (2017) · Micromachines · doi:10.3390/mi8040104 · Cited by 132
“In this way, many research groups have analyzed tears using fluorescence (LIF) [10,11], mass spectrometry (MS) [12,13], electrochemical [1,2,6,7,14,15] and colorimetric [16–18] detectors.”
BackgroundResultInfluentialTang, D., Lin, Y., Zhou, Q., Lin, Y., Li, P., Nießner, R., & Knopp, D. (2014) · Analytical Chemistry · doi:10.1021/ac503616d · Cited by 126
Lab‐on‐a‐Contact Lens: Recent Advances and Future Opportunities in Diagnostics and Therapeutics
Zhu, Y., Li, S., Li, J., Falcone, N., Cui, Q., Shah, S., Hartel, M. C., Yu, N., Young, P., de Barros, N. R., et al. (2022) · Advanced Materials · doi:10.1002/adma.202108389 · Cited by 127
Wireless Integrated Biosensors for Point-of-Care Diagnostic Applications
Ghafar‐Zadeh, E. (2015) · Sensors · doi:10.3390/s150203236 · Cited by 126
“To date, several efforts have been made at developing patch-like biosensors, to detect biomarkers in sweat fluids [19,20] using various techniques, the most promising one being the use of biomarkers that can also be found in the tears fluid (for instance for glucose sensing), saliva and phlegm.”
BackgroundNanomaterial based electrochemical sensors for in vitro detection of small molecule metabolites
Xiao, F., Wang, L., & Duan, H. (2016) · Biotechnology Advances · doi:10.1016/j.biotechadv.2016.01.006 · Cited by 117
Li, J., Li, Y., Shahzad, S. A., Chen, J., Yang, C., Wang, Y., Yang, M., & Yu, C. (2015) · Chemical Communications · doi:10.1039/c4cc10381e · Cited by 84
Naked-Eye Detection of Glucose in Saliva with Bienzymatic Paper-Based Sensor
Santana-Jiménez, L. A., Márquez‐Lucero, A., Osuna, V., Estrada-Moreno, I. A., & Dominguez, R. B. (2018) · Sensors · doi:10.3390/s18041071 · Cited by 69
Circular dichroism glucose biosensor based on chiral cadmium sulfide quantum dots
Ngamdee, K., & Ngeontae, W. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2018.08.005 · Cited by 67
Wearable eye health monitoring sensors based on peacock tail-inspired inverse opal carbon
Gao, B., He, Z., He, B., & Gu, Z. (2019) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2019.03.029 · Cited by 65
Optical glucose biosensor built-in disposable strips and wearable electronic devices
Reda, A., El‐Safty, S. A., Selim, M. M., & Shenashen, M. A. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113237 · Cited by 64
Zhang, J., Tang, Y. C., Teng, L., Lu, M., & Tang, D. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.01.001 · Cited by 56
Jusík, T., Podešva, P., Farka, Z., Kovar, D. R., Skládal, P., & Foret, F. (2015) · Electrochimica Acta · doi:10.1016/j.electacta.2015.12.009 · Cited by 51
Rezki, M., Septiani, N. L. W., Iqbal, M., Adhika, D. R., Wenten, I. G., & Yuliarto, B. (2021) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ac3713 · Cited by 49
Noninvasive Self-diagnostic Device for Tear Collection and Glucose Measurement
Lee, S. H., Cho, Y. C., & Choy, Y. B. (2019) · Scientific Reports · doi:10.1038/s41598-019-41066-8 · Cited by 49
Intelligent point of care testing for medicine diagnosis
Gao, Q., & Li, S. (2024) · Interdisciplinary medicine · doi:10.1002/inmd.20230031 · Cited by 53
Highly specific and rapid glycan based amperometric detection of influenza viruses
Cui, X., Das, A., Dhawane, A. N., Sweeney, J. A., Zhang, X., Chivukula, V., & Iyer, S. S. (2017) · Chemical Science · doi:10.1039/c6sc03720h · Cited by 47
Mao, X., Tian, W., Hatton, T. A., & Rutledge, G. C. (2015) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-015-9209-x · Cited by 37
Fakhri, N., Salehnia, F., Beigi, S. M., Aghabalazadeh, S., Hosseini, M., & Ganjali, M. R. (2019) · Microchimica Acta · doi:10.1007/s00604-019-3489-3 · Cited by 37
Sampling of fluid through skin with magnetohydrodynamics for noninvasive glucose monitoring
Hakala, T. A., García Pérez, A., Wardale, M., Ruuth, I. A., Vänskä, R. T., Nurminen, T. A., Kemp, E., Boeva, Z. A., Alakoskela, J.-M., Pettersson-Fernholm, K., et al. (2021) · Scientific Reports · doi:10.1038/s41598-021-86931-7 · Cited by 32
Enhancing glucose flux into sweat by increasing paracellular permeability of the sweat gland
Jajack, A., Brothers, M., Kasting, G., & Heikenfeld, J. (2018) · PLOS ONE · doi:10.1371/journal.pone.0200009 · Cited by 35
Duong, H. D., Sohn, O.-J., & Rhee, J. I. (2020) · Biosensors · doi:10.3390/bios10080086 · Cited by 34
Performance of the Prototype NovioSense Noninvasive Biosensor for Tear Glucose in Type 1 Diabetes
Geelhoed-Duijvestijn, P. H. L. M., Vegelyte, D., Kownacka, A., Anton, N., Joosse, M. V., & Wilson, C. J. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820964844 · Cited by 34
Li, W., Zhang, L., Li, Q., Wang, S., Luo, X., Deng, H., & Liu, S. (2018) · Carbohydrate Polymers · doi:10.1016/j.carbpol.2018.10.084 · Cited by 26
Infection Screening in Biofluids with Glucose Test Strips
Bekhit, M., Wang, H.-Y., McHardy, S. F., & Górski, W. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.9b05313 · Cited by 23
Electron transfer from FAD-dependent glucose dehydrogenase to single-sheet graphene electrodes
Filipiak, M. S., Vetter, D., Thodkar, K., Gutiérrez‐Sanz, Ó., Jönsson‐Niedziółka, M., & Tarasov, A. (2019) · Electrochimica Acta · doi:10.1016/j.electacta.2019.134998 · Cited by 24
Review of point-of-care platforms for diabetes: (1) sensing
Liu, Y., Luo, X., Yu, Q., Ye, L., Yang, L., & Cui, Y. (2022) · Sensors and Actuators Reports · doi:10.1016/j.snr.2022.100113 · Cited by 20
Han, J. H., Cho, Y. C., Koh, W., & Choy, Y. B. (2020) · PLoS ONE · doi:10.1371/journal.pone.0239317 · Cited by 17
“The extensions of the glucose sensor electrodes were connected to a potentiostat at 150 mV to measure the electric currents, and an equation was generated for calibration to the corresponding known glucose concentrations [29, 30].”
MethodFu, X., Xu, K., Ye, J., Chen, J., & Feng, X. (2014) · Analytical Methods · doi:10.1039/c4ay02527j · Cited by 16
Cha, K. H., Qin, Y., & Meyerhoff, M. E. (2015) · Electroanalysis · doi:10.1002/elan.201400576 · Cited by 21
Self-citationHaque, E., Noman, A. A., Mithu, A. M., & Ahmed, F. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3383314 · Cited by 15
“Cancer cells are also facilitated in such high glucose conditions as cancerous cells absorb higher levels of glucose [1], [2].”
BackgroundMi, Y., Tao, X., Zhang, X., & Si, Y. (2019) · Royal Society Open Science · doi:10.1098/rsos.181769 · Cited by 15
“In recent years, some researches have indicated that heavy metals and other pollutants inhibited the glucose oxidase or glucose dehydrogenase of bacteria, which were related with glucose production [30,38,39].”
BackgroundResultKim, H., Lee, S., Lee, K. W., Kim, E. S., Kim, H., Im, H., Yoon, H. C., Ko, J., & Kim, E. (2023) · Dyes and Pigments · doi:10.1016/j.dyepig.2023.111287 · Cited by 14
Bioinspired assemblies and plasmonic interfaces for electrochemical biosensing
Hinman, S. S., & Cheng, Q. J. (2016) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2016.05.014 · Cited by 14
“While blood glucose has been the medium of choice for glucose testing, tear glucose levels have seen a recent increase in popularity due to their high correlation with blood glucose values and lower invasiveness for sample collection [4-6].”
BackgroundManbohi, A., & Ahmadi, S. H. (2020) · Chemical Papers · doi:10.1007/s11696-020-01105-5 · Cited by 14
Mohammadinejad, A., Heydari, M., Oskuee, R. K., & Rezayi, M. (2021) · Critical Reviews in Analytical Chemistry · doi:10.1080/10408347.2021.1919986 · Cited by 13
“Nowadays, enzyme-based electrochemical glucometer strips have been known as a routine and standard method which is commercially available and frequently used as self-patient care.([133]) Unfortunately, due to unreliability of glucose measurement which fluctuated by lifestyle, exercise, diet, stress, taking medicine an…”
MethodHossain, F., Shen, Q., Balasuriya, N., Law, J. L., Logan, M. R., Houghton, M., Tyrrell, D. L., Joyce, M., & Serpe, M. J. (2023) · Analytical Chemistry · doi:10.1021/acs.analchem.3c00481 · Cited by 12
Oh, J.-M., & Chow, K. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b00552 · Cited by 12
Paper integrated microfluidic contact lens for colorimetric glucose detection
İşgör, P. K., Abbasiasl, T., Das, R., Istif, E., Yener, U. C., & Beker, L. (2024) · Sensors & Diagnostics · doi:10.1039/d4sd00135d · Cited by 13
“15 21 Moreover, based on blood and tear sample collection from 30 subjects, there is a strong correlation between tear and plasma glucose concentrations.”
BackgroundPaper and Other Fibrous Materials—A Complete Platform for Biosensing Applications
Flores-Hernandez, D. R., Santamaría‐García, V. J., Melchor-Martínez, E. M., Sosa‐Hernández, J. E., Parra‐Saldívar, R., & Bonilla‐Ríos, J. (2021) · Biosensors · doi:10.3390/bios11050128 · Cited by 11
Karim, Z., Khan, M. J., Hussain, A., Ahmed, F., & Khan, Z. H. (2024) · Scientific Reports · doi:10.1038/s41598-024-74899-z · Cited by 10
“While in developed countries, screening blood for irregularities is typically performed as part of a routine checkup, measurements of blood glucose are accessible only in the 50% of the primary care settings in low-income countries 7–9 .”
BackgroundA Review: Non Invasive Sensing System for Detection Glucose Level
Bachache, L. N., Hasan, J. A., & Al-Neami, A. Q. (2021) · Journal of Physics Conference Series · doi:10.1088/1742-6596/1963/1/012125 · Cited by 12
“The commercial glucometer is a painful invasive device and could be used in limited times a day due to its desolation and depression [6].”
BackgroundMethodPortable Infrared-Based Glucometer Reinforced with Fuzzy Logic
Nazha, H. M., Darwich, M. A., Ismaiel, E., Shahen, A., Nasser, T., Assaad, M. M., & Juhre, D. (2023) · Biosensors · doi:10.3390/bios13110991 · Cited by 9
Fang, D., Gao, G., Yu, Y., Shen, J., & Zhi, J. (2016) · The Analyst · doi:10.1039/c5an02478a · Cited by 8
Quantification of tear glucose levels and their correlation with blood glucose levels in dogs
Lee, E., Kang, S., Shim, J., Jeong, D., Jeong, Y., Ahn, J., & Seo, K. (2022) · Veterinary Medicine and Science · doi:10.1002/vms3.788 · Cited by 8
“For example, in previous studies using 1- µl tear samples in humans, TG was quantified using electrospray ionisation mass spectrometry (Baca et al., 2007; Taormina et al., 2007) or by measuring output current from an electrochemical blood glucose glucometer strip (Cha et al., 2014).”
MethodResultPortable multimodal sensing system for screening of diabetes mellitus from human urine
Martinez, M. S., Tsyban, Y., Przybysz, A., Bukhamsin, A., Sivalingam, Y., & Salama, K. N. (2025) · View · doi:10.1002/viw.20250010 · Cited by 9
Changes in tear glucose and insulin concentrations following an oral glucose tolerance test
Chao, C., Lema, C., Redfern, R. L., & Richdale, K. (2022) · Clinical and Experimental Optometry · doi:10.1080/08164622.2022.2111204 · Cited by 6
“5–1 μl to determine the glucose level in tears.(11,12) Figure 5.”
BackgroundMethodInfluentialNumerical investigation of PCF-based biosensor for pathologies detection in blood
Haque, E., Noman, A. A., Islam, S. M. R., Islam, M. R., Wang, R., Wang, D., & Ahmed, F. (2026) · Sensing and Bio-Sensing Research · doi:10.1016/j.sbsr.2026.100961 · Cited by 4
Lu, X., Wang, P., Wang, Y., Liu, C., & Li, Z. (2016) · Advanced Materials Technologies · doi:10.1002/admt.201600024 · Cited by 4
Sensor Ecosystems for Health Self‐Monitoring: Focus on Diabetes Management
Ganguly, A., Churcher, N. K. M., Upasham, S., Dhamu, V. N., & Prasad, S. (2023) · Wiley eBooks · doi:10.1002/9783527834266.ch44 · Cited by 4
Espinosa-Hernandez, M. A., Reveles-Huizar, S., & Hosseini, S. (2020) · Lecture notes in bioengineering · doi:10.1007/978-981-15-6382-9_2 · Cited by 3
“This provides an alternative for diabetic patients pricking their fingers with a lower potential interference, non-invasive, and pain-free sample (Cha et al. 2014).”
BackgroundCerda-Kipper, A. S., & Hosseini, S. (2020) · Lecture notes in bioengineering · doi:10.1007/978-981-15-6382-9_4 · Cited by 2
Development of lactate sensors and transfer to printed electronics
Scheiblin, G. (2016) · theses.fr (ABES) · doi:10.70675/59b1550ez11e3z448fzaf37z7322b04ddb1d
Repurposing blood glucose test strips for identification of the antimicrobial colistin
Lopez, C., Raykova, M. R., Corrigan, D. K., Knapp, C. W., & Ward, A. C. (2022) · Sensors and Actuators Reports · doi:10.1016/j.snr.2022.100119
The Clinical Accuracy of Non-Invasive Glucose Monitoring for ex vivo Artificial Pancreas
Levy, M. M., & Levy, Y. M. (2016) · Journal of Emerging Investigators · doi:10.59720/16-008
Ahmed, H., Edris, S. S., Sarg, N., Mhlab, E. M., & Mousa, H. R. (2024) · Benha Medical Journal · doi:10.21608/bmfj.2024.266022.2007
Desagani, D., Cetó, X., del Valle, M., & Ben‐Yoav, H. (2024) · Elsevier eBooks · doi:10.1016/b978-0-443-15684-7.00025-7
de Tijuana, I. T., Beltrán‐Gastélum, M., Gutiérrez-Gutiérrez, R., de Tijuana, I. T., Godínez-Dávalos, B. U., de Tijuana, I. T., Salazar‐Gastélum, M. I., de Tijuana, I. T., Cárdenas-Valdez, J. R., de Tijuana, I. T., et al. (2026) · Revista Mexicana de Ingeniería Biomédica · doi:10.17488/rmib.47.si-taih.1521
Quantification of avian influenza A virus (H5N1) via enzymatic reactions and glucometer test strips
Islam, M. R., Hossain, F., Joyce, M., Elaish, M., Tyrrell, D. L., Hobman, T. C., & Serpe, M. J. (2026) · Sensors & Diagnostics · doi:10.1039/d6sd00003g
Design features of glucose biosensor active layer based on insoluble redox mediator
Kleinikova, S. A., Afanasyeva, P., Levchenko, M., Zavorotnaya, U., Tomilina, M., Gor’kov, K. V., Gerasimova, E., & Zolotukhina, E. V. (2026) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2026.120554
Bionanofactory-Enabled Point-of-Care Biosensors
Saravanan, D., Rajamohan, R., Sundaram, T., & Mohan, M. (2026) · Interdisciplinary biotechnological advances · doi:10.1007/978-981-92-1553-9_9
Smart contact lens for personalized glucose level sensing and forecasting
Paek, J., Park, W., Seo, H., Song, H., Kim, D., Chung, W. G., An, S., Hong, Y., Kim, J., Lee, Y.-H., et al. (2026) · Device · doi:10.1016/j.device.2026.101330
Metadata
- Authors
- Cha, K. H., Jensen, G. C., Balijepalli, A. S., Cohan, B. E., & Meyerhoff, M. E.
- Affiliations
- University of Michigan (US)
- EyeLab Group LLC, 2350 Washtenaw Avenue No. 4, Ann Arbor, Michigan, 48104-4525, United States
- Corresponding author
- Meyerhoff, Mark E.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2014-01-24
- Volume
- 86
- Issue
- 3
- Pages
- 1902-1908
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2094203185
- Semantic Scholar
- fab7c610aa10576f9c465e749ce19ca2dc3d222f
- Cited by
- 67
- References
- 37
- Citations by year
- 2026: 5 · 2025: 1 · 2024: 6 · 2023: 6 · 2022: 5 · 2021: 8 · 2020: 7 · 2019: 5 · 2018: 4 · 2017: 2 · 2016: 10 · 2015: 6 · 2014: 2
- FWCI
- 3.54
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Analytical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- tear glucose, blood glucose monitoring, diabetes mellitus, non-invasive glucose measurement, limit of quantification, acetaminophen interference, ascorbic acid interference, uric acid interference
- MeSH terms
- Electrochemistry, Glucose, Humans, Reagent Strips, Tears, Blood Glucose Self-Monitoring, Limit of Detection
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.