Integration of High-Charge-Injection-Capacity Electrodes onto Polymer Softening Neural Interfaces
The DOI matches its Crossref record; 20 of the 20 references it cites were found in a registry or an index.
Abstract
Softening neural interfaces are implanted stiff to enable precise insertion, and they soften in physiological conditions to minimize modulus mismatch with tissue. In this work, a high-charge-injection-capacity iridium electrode fabrication process is detailed. For the first time, this process enables integration of iridium electrodes onto softening substrates using photolithography to define all features in the device. Importantly, no electroplated layers are utilized, leading to a highly scalable method for consistent device fabrication. The iridium electrode is metallically bonded to the gold conductor layer, which is covalently bonded to the softening substrate via sulfur-based click chemistry. The resulting shape-memory polymer neural interfaces can deliver more than 2 billion symmetric biphasic pulses (100 μs/phase), with a charge of 200 μC/cm(2) and geometric surface area (GSA) of 300 μm(2). A transfer-by-polymerization method is used in combination with standard semiconductor processing techniques to fabricate functional neural probes onto a thiol-ene-based, thin film substrate. Electrical stability is tested under simulated physiological conditions in an accelerated electrical aging paradigm with periodic measurement of electrochemical impedance spectra (EIS) and charge storage capacity (CSC) at various intervals. Electrochemical characterization and both optical and scanning electron microscopy suggest significant breakdown of the 600 nm-thick parylene-C insulation, although no delamination of the conductors or of the final electrode interface was observed. Minor cracking at the edges of the thin film iridium electrodes was occasionally observed. The resulting devices will provide electrical recording and stimulation of the nervous system to better understand neural wiring and timing, to target treatments for debilitating diseases, and to give neuroscientists spatially selective and specific tools to interact with the body. This approach has uses for cochlear implants, nerve cuff electrodes, penetrating cortical probes, spinal stimulators, blanket electrodes for the gut, stomach, and visceral organs and a host of other custom nerve-interfacing devices.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Neuronal ensemble control of prosthetic devices by a human with tetraplegia
Hochberg, L. R., Serruya, M. D., Friehs, G. M., Mukand, J. A., Saleh, M., Caplan, A. H., Branner, A., Chen, D., Penn, R. D., & Donoghue, J. P. (2006) · Nature · doi:10.1038/nature04970 · Cited by 3,295 · cited 1×
“volitionally control robotic arms for short periods of time.(2) Despite the remarkable developments Page 2 of 41”
BackgroundBiran, R., Martin, D. C., & Tresco, P. A. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2005.04.020 · Cited by 893
Kozai, T. D. Y., Langhals, N. B., Patel, P. R., Deng, X., Zhang, H., Smith, K. L., Lahann, J., Kotov, N. A., & Kipke, D. R. (2012) · Nature Materials · doi:10.1038/nmat3468 · Cited by 721
Carbon nanotube coating improves neuronal recordings
Keefer, E. W., Botterman, B. R., Romero, M. I., Rossi, A. F., & Gross, G. W. (2008) · Nature Nanotechnology · doi:10.1038/nnano.2008.174 · Cited by 721
Conducting polymers for neural interfaces: Challenges in developing an effective long-term implant
Green, R. A., Lovell, N. H., Wallace, G. G., & Poole-Warren, L. A. (2008) · Biomaterials · doi:10.1016/j.biomaterials.2008.04.047 · Cited by 691
McConnell, G. C., Rees, H. D., Levey, A. I., Gutekunst, C.-A., Gross, R. E., & Bellamkonda, R. V. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/056003 · Cited by 480
Subbaroyan, J., Martin, D. C., & Kipke, D. R. (2005) · Journal of Neural Engineering · doi:10.1088/1741-2560/2/4/006 · Cited by 368
Micromachined, Polyimide-Based Devices for Flexible Neural Interfaces
Stieglitz, T., Beutel, H., Schuettler, M., & Meyer, J.-U. (2000) · Biomedical Microdevices · doi:10.1023/a:1009955222114 · Cited by 343 · cited 1×
“and the foreign implanted device.(9-11) Kim et al.”
BackgroundMechanically-compliant intracortical implants reduce the neuroinflammatory response
Nguyen, J. K., Park, D. J., Skousen, J. L., Hess-Dunning, A. E., Tyler, D. J., Rowan, S. J., Weder, C., & Capadona, J. R. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/5/056014 · Cited by 274
Electrodeposited iridium oxide for neural stimulation and recording electrodes
Meyer, R. D., Cogan, S. F., Nguyen, T. H., & Rauh, R. D. (2001) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/7333.918271 · Cited by 255
Cited by
Bioinspired Polymer Systems with Stimuli-Responsive Mechanical Properties
Montero de Espinosa, L., Meesorn, W., Moatsou, D., & Weder, C. (2017) · Chemical Reviews · doi:10.1021/acs.chemrev.7b00168 · Cited by 415
The future of neuroscience: flexible and wireless implantable neural electronics
McGlynn, E., Nabaei, V., Ren, E., Galeote‐Checa, G., Das, R., Curia, G., & Heidari, H. (2021) · Advanced Science · doi:10.1002/advs.202002693 · Cited by 115
Recent advances in neural electrode–tissue interfaces
Woeppel, K., Yang, Q., & Cui, X. T. (2017) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2017.09.003 · Cited by 99
Thin film multi-electrode softening cuffs for selective neuromodulation
González-González, M. A., Kanneganti, A., Joshi-Imre, A., Hernandez-Reynoso, A. G., Bendale, G., Modi, R., Ecker, M., Khurram, A., Cogan, S. F., Voit, W. E., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-34566-6 · Cited by 97
Self-citationComputational assessment of neural probe and brain tissue interface under transient motion.
Polanco, M., Bawab, S., & Yoon, H. (2016) · Biosensors · doi:10.3390/bios6020027 · Cited by 36
Metadata
- Authors
- Arreaga-Salas, D. E., Avendaño-Bolívar, A., Simon, D., Reit, R., Garcia-Sandoval, A., Rennaker, R. L., & Voit, W.
- Venue
- ACS Applied Materials & Interfaces
- Publisher
- American Chemical Society (ACS)
- Published
- 2015-11-25
- Volume
- 7
- Issue
- 48
- Pages
- 26614-26623
- ISSN
- 1944-8244 (ISSN-L)
- 1944-8252
- OpenAlex
- W2236281078
- Cited by
- 41
- References
- 20
- FWCI
- 2.20
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Integration of High-Charge-Injection-Capacity Electrodes onto Polymer Softening Neural Interfaces. https://app.cochat.ai/papers/10.1021/acsami.5b08139
- The paper
- Arreaga-Salas, D. E., Avendaño-Bolívar, A., Simon, D., Reit, R., Garcia-Sandoval, A., Rennaker, R. L., & Voit, W. (2015). Integration of High-Charge-Injection-Capacity Electrodes onto Polymer Softening Neural Interfaces. ACS Applied Materials & Interfaces. https://doi.org/10.1021/acsami.5b08139