A finite-element model of the mechanical effects of implantable microelectrodes in the cerebral cortex
Open accessAbstract
The viability of chronic neural microelectrodes for electrophysiological recording and stimulation depends on several factors, including the encapsulation of the implant by a reactive tissue response. We postulate that mechanical strains induced around the implant site may be one of the leading factors responsible for the sustained tissue response in chronic implants. The objectives of this study were to develop a finite-element model of the probe-brain tissue interface and analyze the effects of tethering forces, probe-tissue adhesion and stiffness of the probe substrate on the interfacial strains induced around the implant site. A 3D finite-element model of the probe-brain tissue microenvironment was developed and used to simulate interfacial strains created by 'micromotion' of chronically implanted microelectrodes. Three candidate substrates were considered: (a) silicon, (b) polyimide and (c) a hypothetical 'soft' material. Simulated tethering forces resulted in elevated strains both at the tip and at the sharp edges of the probe track in the tissue. The strain fields induced by a simulated silicon probe were similar to those induced by a simulated polyimide probe, albeit at higher absolute values for radial tethering forces. Simulations of poor probe-tissue adhesion resulted in elevated strains at the tip and delamination of the tissue from the probe. A tangential tethering force results in 94% reduction in the strain value at the tip of the polyimide probe track in the tissue, whereas the simulated 'soft' probe induced two orders of magnitude smaller values of strain compared to a simulated silicon probe. The model results indicate that softer substrates reduce the strain at the probe-tissue interface and thus may also reduce tissue response in chronic implants.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Computational modeling showed that a polyimide material reduced applied strain to surrounding tissue by 94%, while a theoretical 6 MPa material reduced strain by two orders of magnitude.
“the polyimide material demonstrated a 94% reduction in applied strain to the surrounding tissue, while strain with the theoretical 6 MPa material was reduced by two orders of magnitude.”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [169]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Large-scale recording of neuronal ensembles
Buzsáki, G. (2004) · Nature Neuroscience · doi:10.1038/nn1233 · Cited by 1,875 · see citing works
Mechanical Behavior of Materials
Hosford, W. F. (2005) · Cambridge University Press eBooks · doi:10.1017/cbo9780511810930 · Cited by 1,795 · see citing works
Biran, R., Martin, D. C., & Tresco, P. A. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2005.04.020 · Cited by 893 · see citing works
Self-citationBrain responses to micro-machined silicon devices
Szarowski, D. H., Andersen, M. D., Retterer, S. T., Spence, A. J., Isaacson, M. S., Craighead, H. G., Turner, J. N., & Shain, W. G. (2003) · Brain Research · doi:10.1016/s0006-8993(03)03023-3 · Cited by 798 · see citing works
Chronic, multisite, multielectrode recordings in macaque monkeys
Nicolelis, M. A. L., Dimitrov, D., Carmena, J. M., Crist, R., Lehew, G., Kralik, J. D., & Wise, S. P. (2003) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1934665100 · Cited by 740 · see citing works
Flexible polyimide-based intracortical electrode arrays with bioactive capability
Rousche, P. J., Pellinen, D. S., Pivin, D. P., Williams, J. C., Vetter, R. J., & Kipke, D. R. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.914800 · Cited by 644 · see citing works
Self-citationRegional, Directional, and Age-Dependent Properties of the Brain Undergoing Large Deformation
Prange, M. T., & Margulies, S. S. (2002) · Journal of Biomechanical Engineering · doi:10.1115/1.1449907 · Cited by 636 · see citing works
Mechanical properties of brain tissue in tension
Miller, K., & Chinzei, K. (2002) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00234-2 · Cited by 636 · see citing works
Cerebral Astrocyte Response to Micromachined Silicon Implants
Turner, J. N., Shain, W. G., Szarowski, D. H., Andersen, M., Martins, S., Isaacson, M. S., & Craighead, H. G. (1999) · Experimental Neurology · doi:10.1006/exnr.1998.6983 · Cited by 593 · see citing works
Constitutive modelling of brain tissue: Experiment and theory
Miller, K., & Chinzei, K. (1997) · Journal of Biomechanics · doi:10.1016/s0021-9290(97)00092-4 · Cited by 525 · see citing works
Mechanical properties of brain tissue in-vivo: experiment and computer simulation
Miller, K., Chinzei, K., Orssengo, G. J., & Bednarz, P. (2000) · Journal of Biomechanics · doi:10.1016/s0021-9290(00)00120-2 · Cited by 501 · see citing works
Factors influencing the biocompatibility of insertable silicon microshafts in cerebral cortex
Edell, D. J., Toi, V. V., McNeil, V. M., & Clark, L. D. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.141202 · Cited by 479 · see citing works
Chronic Neural Recording Using Silicon-Substrate Microelectrode Arrays Implanted in Cerebral Cortex
Vetter, R. J., Williams, J. C., Hetke, J. F., Nunamaker, E. A., & Kipke, D. R. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.826680 · Cited by 477 · see citing works
Self-citationA three-dimensional microelectrode array for chronic neural recording
Hoogerwerf, A. C., & Wise, K. D. (1994) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.335862 · Cited by 372 · see citing works
Kipke, D. R., Vetter, R. J., Williams, J. C., & Hetke, J. F. (2003) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2003.814443 · Cited by 364 · see citing works
Self-citationDeformation and fracture of small silicon crystals
Pearson, G. L., Read, W. T., & Feldmann, W. L. (1957) · Acta Metallurgica · doi:10.1016/0001-6160(57)90164-5 · Cited by 300 · see citing works
Age-Dependent Changes in Material Properties of the Brain and Braincase of the Rat
Gefen, A., Gefen, N., Zhu, Q., Raghupathi, R., & Margulies, S. S. (2003) · Journal of Neurotrauma · doi:10.1089/089771503770802853 · Cited by 299 · see citing works
A multichannel neural probe for selective chemical delivery at the cellular level
Chen, J., Wise, K. D., Hetke, J. F., & Bledsoe, S. C. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.605435 · Cited by 285 · see citing works
Zhang, M., Zheng, Y., & Mak, A. F. T. (1997) · Medical Engineering & Physics · doi:10.1016/s1350-4533(97)00017-9 · Cited by 282 · see citing works
Shear Properties of Human Brain Tissue
Donnelly, B. R., & Medige, J. (1997) · Journal of Biomechanical Engineering · doi:10.1115/1.2798289 · Cited by 239 · see citing works
Reassessment of brain elasticity for analysis of biomechanisms of hydrocephalus
Taylor, Z. A., & Miller, K. (2004) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2003.11.027 · Cited by 235 · see citing works
3D flexible multichannel neural probe array
Takeuchi, S., Suzuki, T. J., Mabuchi, K., & Fujita, H. (2003) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/14/1/014 · Cited by 221 · see citing works
Chronic response of adult rat brain tissue to implants anchored to the skull
Kim, Y. T., Hitchcock, R. W., Bridge, M. J., & Tresco, P. A. (2003) · Biomaterials · doi:10.1016/j.biomaterials.2003.09.010 · Cited by 214 · see citing works
A 32-site neural recording probe fabricated by DRIE of SOI substrates
Peter Norlin, A., Kindlundh, M., Mouroux, A., Yoshida, K., & Hofmann, U. (2002) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/12/4/312 · Cited by 184 · see citing works
Histopathological evaluation of materials implanted in the cerebral cortex
Stensaas, S. S., & Stensaas, L. J. (1978) · Acta Neuropathologica · doi:10.1007/bf00689766 · Cited by 177 · see citing works
Metz, S., Bertsch, A., Bertrand, D., & Renaud, P. (2004) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2003.11.021 · Cited by 167 · see citing works
Mechanical properties of tissues of the nervous system
Ommaya, A. K. (1968) · Journal of Biomechanics · doi:10.1016/0021-9290(68)90015-8 · Cited by 164 · see citing works
High density interconnects and flexible hybrid assemblies for active biomedical implants
Meyer, J.-U., Stieglitz, T., Scholz, O., Haberer, W., & Beutel, H. (2001) · IEEE Transactions on Advanced Packaging · doi:10.1109/6040.938305 · Cited by 145 · see citing works
Repeated voltage biasing improves unit recordings by reducing resistive tissue impedances
Johnson, M. D., Otto, K. J., & Kipke, D. R. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847373 · Cited by 131 · see citing works
Self-citationStieglitz, T., Beutel, H., & Meyer, J.-U. (1997) · Sensors and Actuators A: Physical · doi:10.1016/s0924-4247(97)01494-5 · Cited by 125 · see citing works
Brands, D. W. A., Peters, G. G., & Bovendeerd, P. H. M. (2003) · Journal of Biomechanics · doi:10.1016/s0021-9290(03)00243-4 · Cited by 115 · see citing works
Tissue response to potential neuroprosthetic materials implanted subdurally
Yuen, T. G. H., Agnew, W. F., & Bullara, L. A. (1987) · Biomaterials · doi:10.1016/0142-9612(87)90103-7 · Cited by 92 · see citing works
The reaction of the cerebral cortex to chronically implanted plastic needles.
Stensaas, S. S., & Stensaas, L. J. (1976) · PubMed · Cited by 88 · see citing works
Flexible biomedical microdevices with double-sided electrode arrangements for neural applications
Stieglitz, T. (2001) · Sensors and Actuators A Physical · doi:10.1016/s0924-4247(01)00520-9 · Cited by 75 · see citing works
Quantitative analysis of methods for reducing physiological brain pulsations
Britt, R. H., & Rossi, G. T. (1982) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(82)90085-1 · Cited by 64 · see citing works
Mechanical factors in the design of chronic recording intracortical microelectrodes
Goldstein, S. R., & Salcman, M. (1973) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1973.324190 · Cited by 53 · see citing works
34th Annual Meeting of Society for Neuroscience
Calzavara, R., Mailly, P., Bennett, S. R., Kim, K. J., Yung, Y. W., & Haber, S. N. (2004) · HAL · Cited by 40 · see citing works
(2001) · The International Journal of Artificial Organs · doi:10.1177/039139880102400822 · Cited by 40 · see citing works
3-D silicon probe array with hybrid polymer interconnect for chronic cortical recording
Hetke, J. F., Williams, J. C., Pellinen, D. S., Vetter, R. J., & Kipke, D. R. (2003) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/cne.2003.1196787 · Cited by 39 · see citing works
Self-citationA shuttered probe with in-line flowmeters for chronic in-vivo drug delivery
Papageorgiou, D., Bledsoe, S. C., Gulari, M. N., Hetke, J. F., Anderson, D. J., & Wise, K. D. (2002) · IEEE International Conference on Micro Electro Mechanical Systems (MEMS) · doi:10.1109/memsys.2001.906516 · Cited by 33 · see citing works
Microactuated neural probes to compensate for brain micromotion
Muthuswamy, J., Gilletti, A., Jain, T., & Okandan, M. (2004) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2003.1279819 · Cited by 29 · see citing works
Koeneman, B. A., Lee, K., Singh, A., He, J., Raupp, G. B., Panitch, A., & Capco, D. G. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.02.033 · Cited by 18 · see citing works
Using region-of-interest based finite element modelling for brain-surgery simulation
Hansen, K. V., & Larsen, O. V. (1998) · Lecture notes in computer science · doi:10.1007/bfb0056214 · Cited by 11 · see citing works
Chemical sensing capability of MEMS implantable multichannel neural microelectrode arrays
Johnson, M. D., Williams, J. C., Holecko, M. M., & Kipke, D. R. (2004) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2003.1280858 · Cited by 10 · see citing works
Self-citationChronically implantable neural information transducers
Meyer, J.-U., Edell, D. J., Rutten, W. L. C., Pine, J., & Edell, D. J. (1996) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1996.646468 · Cited by 10 · see citing works
Tomorrow's implantable electronic systems
Edell, D. J. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1996.646483 · Cited by 4 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Design of biodegradable, implantable devices towards clinical translation
Li, C., Guo, C., Fitzpatrick, V., Ibrahim, A. G. A., Zwierstra, M. J., Hanna, P. C., Lechtig, A., Nazarian, A., Lin, S. J., & Kaplan, D. L. (2019) · Nature Reviews Materials · doi:10.1038/s41578-019-0150-z · Cited by 819 · see citing works
Kozai, T. D. Y., Langhals, N. B., Patel, P. R., Deng, X., Zhang, H., Smith, K. L., Lahann, J., Kotov, N. A., & Kipke, D. R. (2012) · Nature Materials · doi:10.1038/nmat3468 · Cited by 721 · see citing works
Self-citationLudwig, K. A., Uram, J. D., Yang, J., Martin, D. C., & Kipke, D. R. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/1/007 · Cited by 638 · see citing works
Self-citationBrain Tissue Responses to Neural Implants Impact Signal Sensitivity and Intervention Strategies
Kozai, T. D. Y., Jaquins-Gerstl, A. S., Vazquez, A. L., Michael, A. C., & Cui, X. T. (2015) · ACS Chemical Neuroscience · doi:10.1021/cn500256e · Cited by 622 · see citing works
Nanomaterials for Neural Interfaces
Kotov, N. A., Winter, J. O., Clements, I. P., Jan, E., Timko, B. P., Campidelli, S., Pathak, S., Mazzatenta, A., Lieber, C. M., Prato, M., et al. (2009) · Advanced Materials · doi:10.1002/adma.200801984 · Cited by 504 · see citing works
Wireless Optofluidic Systems for Programmable In Vivo Pharmacology and Optogenetics
Jeong, J., McCall, J. G., Shin, G., Zhang, Y., Al‐Hasani, R., Kim, M. K., Li, S., Sim, J. Y., Jang, K., Shi, Y., et al. (2015) · Cell · doi:10.1016/j.cell.2015.06.058 · Cited by 493 · see citing works
Liu, J., Qu, S., Suo, Z., & Yang, W. (2020) · National Science Review · doi:10.1093/nsr/nwaa254 · Cited by 485 · see citing works
Neural probe design for reduced tissue encapsulation in CNS
Seymour, J. P., & Kipke, D. R. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.03.024 · Cited by 462 · see citing works
Self-citationA Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389 · see citing works
Naturally sourced hydrogels: emerging fundamental materials for next-generation healthcare sensing
Wang, Z., Wei, H. Z., Huang, Y., Wei, Y., & Chen, J. (2023) · Chemical Society Reviews · doi:10.1039/d2cs00813k · Cited by 400 · see citing works
Progress towards biocompatible intracortical microelectrodes for neural interfacing applications
Jorfi, M., Skousen, J. L., Weder, C., & Capadona, J. R. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/011001 · Cited by 384 · see citing works
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347 · see citing works
Bjornsson, C. S., Oh, S., Al‐Kofahi, Y., Lim, Y.-J., Smith, K. L., Turner, J. N., De, S., Roysam, B., Shain, W. G., & Kim, S. J. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/3/002 · Cited by 330 · see citing works
Park, S., Yuk, H., Zhao, R., Yim, Y. S., Woldeghebriel, E. W., Kang, J., Canales, A., Fink, Y., Choi, G. B., Zhao, X., et al. (2021) · Nature Communications · doi:10.1038/s41467-021-23802-9 · Cited by 312 · see citing works
Brain micromotion around implants in the rodent somatosensory cortex
Gilletti, A., & Muthuswamy, J. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/3/001 · Cited by 307 · see citing works
Mechanically-compliant intracortical implants reduce the neuroinflammatory response
Nguyen, J. K., Park, D. J., Skousen, J. L., Hess-Dunning, A. E., Tyler, D. J., Rowan, S. J., Weder, C., & Capadona, J. R. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/5/056014 · Cited by 274 · see citing works
Partial pressure of oxygen in the human body: a general review.
Ortiz‐Prado, E., Dunn, J. F., Vásconez-González, J., Castillo, D., & Viscor, G. (2019) · Europe PMC (PubMed Central) · Cited by 263 · see citing works
High‐Adhesion Stretchable Electrodes Based on Nanopile Interlocking
Liu, Z., Wang, X., Qi, D., Xu, C. Y., Yu, J., Liu, Y., Jiang, Y., Liedberg, B., & Chen, X. (2016) · Advanced Materials · doi:10.1002/adma.201603382 · Cited by 225 · see citing works
A carbon-fiber electrode array for long-term neural recording
Guitchounts, G., Markowitz, J. E., Liberti, W. A., & Gardner, T. J. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/4/046016 · Cited by 224 · see citing works
Elastocapillary self-assembled neurotassels for stable neural activity recordings
Guan, S., Wang, J., Gu, X., Zhao, Y., Hou, R., Fan, H., Zou, L., Gao, L., Du, M., Li, C., et al. (2019) · Science Advances · doi:10.1126/sciadv.aav2842 · Cited by 222 · see citing works
Kozai, T. D. Y., Catt, K. A., Li, X., Gugel, Z. V., Olafsson, V. T., Vazquez, A. L., & Cui, X. T. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.10.040 · Cited by 214 · see citing works
State-of-the-art MEMS and microsystem tools for brain research
Seymour, J. P., Wu, F., Wise, K. D., & Yoon, E. (2017) · Microsystems & Nanoengineering · doi:10.1038/micronano.2016.66 · Cited by 216 · see citing works
Kozai, T. D. Y., Vazquez, A. L., Weaver, C. L., Kim, S.-G., & Cui, X. T. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/6/066001 · Cited by 215 · see citing works
Softening gold for elastronics
Zhu, B., Gong, S. P., & Cheng, W. (2018) · Chemical Society Reviews · doi:10.1039/c8cs00609a · Cited by 213 · see citing works
Ultrasoft microwire neural electrodes improve chronic tissue integration
Du, Z. J., Kolarcik, C. L., Kozai, T. D. Y., Luebben, S. D., Sapp, S. A., Zheng, X. S., Nabity, J., & Cui, X. T. (2017) · Acta Biomaterialia · doi:10.1016/j.actbio.2017.02.010 · Cited by 208 · see citing works
Kim, D., Wiler, J. A., Anderson, D. J., Kipke, D. R., & Martin, D. C. (2009) · Acta Biomaterialia · doi:10.1016/j.actbio.2009.07.034 · Cited by 206 · see citing works
Self-citationA review on mechanical considerations for chronically-implanted neural probes
Lecomte, A., Descamps, E., & Bergaud, C. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa8b4f · Cited by 205 · see citing works
The Foreign Body Response Demystified
Chandorkar, Y., K, R., & Basu, B. (2019) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.8b00252 · Cited by 201 · see citing works
Implantable microscale neural interfaces
Cheung, K. C. (2007) · Biomedical Microdevices · doi:10.1007/s10544-006-9045-z · Cited by 202 · see citing works
Electroactive polymers for neural interfaces
Asplund, M., Nyberg, T., & Inganäs, O. (2010) · Polymer Chemistry · doi:10.1039/c0py00077a · Cited by 195 · see citing works
Electrochemical polymerization of conducting polymers in living neural tissue
Richardson-Burns, S. M., Hendricks, J. L., & Martin, D. C. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/2/l02 · Cited by 196 · see citing works
Self-citationKozai, T. D. Y., Gugel, Z., Li, X., Gilgunn, P. J., Khilwani, R., Ozdoganlar, O. B., Fedder, G. K., Weber, D. J., & Cui, X. T. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.07.039 · Cited by 193 · see citing works
Kozai, T. D. Y., & Kipke, D. R. (2009) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2009.08.002 · Cited by 186 · see citing works
Self-citationMechanically adaptive intracortical implants improve the proximity of neuronal cell bodies
Harris, J. P., Capadona, J. R., Miller, R. H., Healy, B. C., Shanmuganathan, K., Rowan, S. J., Weder, C., & Tyler, D. J. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/6/066011 · Cited by 190 · see citing works
Mechanically Adaptive Organic Transistors for Implantable Electronics
Reeder, J. T., Kaltenbrunner, M., Ware, T. H., Arreaga‐Salas, D., Avendaño-Bolívar, A., Yokota, T., Inoue, Y., Sekino, M., Voit, W. E., Sekitani, T., et al. (2014) · Advanced Materials · doi:10.1002/adma.201400420 · Cited by 185 · see citing works
Kozai, T. D. Y., Marzullo, T., Hooi, F. M., Langhals, N. B., Majewska, A. K., Brown, E. B., & Kipke, D. R. (2010) · Journal of Neural Engineering · doi:10.1088/1741-2560/7/4/046011 · Cited by 179 · see citing works
Self-citationFlexible, Penetrating Brain Probes Enabled by Advances in Polymer Microfabrication
Weltman, A., Yoo, J. J., & Meng, E. F. (2016) · Micromachines · doi:10.3390/mi7100180 · Cited by 180 · see citing works
Spencer, K. C., Sy, J. C., Ramadi, K. B., Graybiel, A. M., Langer, R., & Cima, M. J. (2017) · Scientific Reports · doi:10.1038/s41598-017-02107-2 · Cited by 184 · see citing works
Large enhancement in neurite outgrowth on a cell membrane-mimicking conducting polymer
Zhu, B., Luo, S., Zhao, H., Lin, H., Sekine, J., Nakao, A., Chen, C., Yamashita, Y., & Yu, H. (2014) · Nature Communications · doi:10.1038/ncomms5523 · Cited by 173 · see citing works
Li, Y., Li, N., Liu, W., Promiński, A., Kang, S., Dai, Y., Liu, Y., Hu, H., Wai, S., Dai, S., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-40191-3 · Cited by 171 · see citing works
Wen, X., Wang, B., Huang, S., Liu, T. “., Lee, M., Chung, P., Chow, Y. T., Huang, I.-W., Monbouquette, H. G., Maidment, N. T., et al. (2019) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2019.01.060 · Cited by 162 · see citing works
Semiconducting Polymers for Neural Applications
Dimov, I. B., Moser, M., Malliaras, G. G., & McCulloch, I. A. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00685 · Cited by 160 · see citing works
Sharp, A. A., Ortega, A. M., Restrepo, D., Curran-Everett, D., & Gall, K. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.2003261 · Cited by 157 · see citing works
Rao, L., Zhou, H., Li, T., Li, C., & Duan, Y. Y. (2012) · Acta Biomaterialia · doi:10.1016/j.actbio.2012.03.001 · Cited by 154 · see citing works
In vivodeployment of mechanically adaptive nanocomposites for intracortical microelectrodes
Harris, J. P., Hess‐Dunning, A., Rowan, S. J., Weder, C., Zorman, C. A., Tyler, D. J., & Capadona, J. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/4/046010 · Cited by 153 · see citing works
Kozai, T. D. Y., & Vazquez, A. L. (2015) · Journal of Materials Chemistry B · doi:10.1039/c5tb00108k · Cited by 151 · see citing works
Luan, L., Robinson, J. T., Aazhang, B., Chi, T., Yang, K., Li, X., Rathore, H., Singer, A., Yellapantula, S., Fan, Y., et al. (2020) · Neuron · doi:10.1016/j.neuron.2020.10.011 · Cited by 150 · see citing works
Implantable intracortical microelectrodes: reviewing the present with a focus on the future
Wang, Y., Yang, X., Zhang, X., Wang, Y., & Pei, W. (2023) · Microsystems & Nanoengineering · doi:10.1038/s41378-022-00451-6 · Cited by 145 · see citing works
Kozai, T. D. Y., Li, X., Bodily, L. M., Caparosa, E. M., Zenonos, G. A., Carlisle, D. L., Friedlander, R. M., & Cui, X. T. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.08.006 · Cited by 136 · see citing works
Tien, L. W., Wu, F., Tang‐Schomer, M. D., Yoon, E., Omenetto, F. G., & Kaplan, D. L. (2013) · Advanced Functional Materials · doi:10.1002/adfm.201203716 · Cited by 136 · see citing works
Long-term changes in the material properties of brain tissue at the implant–tissue interface
Sridharan, A., Rajan, S. D., & Muthuswamy, J. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/6/066001 · Cited by 133 · see citing works
Elastic and viscoelastic characterization of agar
Nayar, V. T., Weiland, J. D., Nelson, C. S., & Hodge, A. M. (2011) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2011.05.027 · Cited by 133 · see citing works
Repeated voltage biasing improves unit recordings by reducing resistive tissue impedances
Johnson, M. D., Otto, K. J., & Kipke, D. R. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847373 · Cited by 131 · see citing works
Self-citationSU-8 based microprobes for simultaneous neural depth recording and drug delivery in the brain
Altuna, A., Bellistri, E., Cid, E., Aivar, P., Gal, B., Berganzo, J., Gabriel, G., Guimerà‐Brunet, A., Villa, R., Fernández, L. J., et al. (2013) · Lab on a Chip · doi:10.1039/c3lc41364k · Cited by 129 · see citing works
Novel flexible Parylene neural probe with 3D sheath structure for enhancing tissue integration
Kuo, J. T. W., Kim, B., Hara, S. A., Lee, C. D., Gutierrez, C. A., Hoang, T. Q., & Meng, E. F. (2012) · Lab on a Chip · doi:10.1039/c2lc40935f · Cited by 127 · see citing works
Thiol‐ene/acrylate substrates for softening intracortical electrodes
Ware, T. H., Simon, D., Liu, H. C., Musa, T., Vasudevan, S., Sloan, A. M., Keefer, E. W., Rennaker, R. L., & Voit, W. E. (2013) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbmb.32946 · Cited by 126 · see citing works
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500317 · Cited by 125 · see citing works
Eles, J. R., Vazquez, A. L., Snyder, N. R., Lagenaur, C. F., Murphy, M. C., Kozai, T. D. Y., & Cui, X. T. (2016) · Biomaterials · doi:10.1016/j.biomaterials.2016.10.054 · Cited by 116 · see citing works
Toward a self-deploying shape memory polymer neuronal electrode
Sharp, A. A., Panchawagh, H. V., Ortega, A., Artale, R., Richardson-Burns, S., Finch, D. S., Gall, K., Mahajan, R. L., & Restrepo, D. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/4/l02 · Cited by 113 · see citing works
Fluidic Microactuation of Flexible Electrodes for Neural Recording
Vitale, F., Vercosa, D. G., Rodriguez, A. V., Pamulapati, S. S., Seibt, F., Lewis, E., Yan, J. S., Badhiwala, K. N., Adnan, M., Carfagni, G. R., et al. (2017) · Nano Letters · doi:10.1021/acs.nanolett.7b04184 · Cited by 114 · see citing works
Lee, H. C., Ejserholm, F., Gaire, J., Currlin, S., Schouenborg, J., Wallman, L., Bengtsson, M., Park, K., & Otto, K. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa68f0 · Cited by 110 · see citing works
A Novel Anti‐inflammatory Surface for Neural Electrodes
He, W. J., McConnell, G. C., Schneider, T. M., & Bellamkonda, R. V. (2007) · Advanced Materials · doi:10.1002/adma.200700943 · Cited by 111 · see citing works
Compliant intracortical implants reduce strains and strain rates in brain tissuein vivo
Sridharan, A., Nguyen, J. K., Capadona, J. R., & Muthuswamy, J. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/3/036002 · Cited by 112 · see citing works
Karumbaiah, L., Norman, S. E., Rajan, N. B., Anand, S., Saxena, T. K., Betancur, M. I., Patkar, R., & Bellamkonda, R. V. (2012) · Biomaterials · doi:10.1016/j.biomaterials.2012.05.021 · Cited by 107 · see citing works
Chen, Y., Lai, H.-Y., Lin, S.-H., Chao, W.-H., Chao, W.-H., Liao, C., Tsang, S., Chen, Y., & Lin, S.-Y. (2009) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2009.05.010 · Cited by 110 · see citing works
The sinusoidal probe: a new approach to improve electrode longevity
Sohal, H. S., Jackson, A., Jackson, R., Clowry, G. J., Vassilevski, K. V., O’Neill, A., & Baker, S. N. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00010 · Cited by 104 · see citing works
3D Electrodes for Bioelectronics
Cho, Y. H., Park, Y., Kim, S., & Park, J. (2021) · Advanced Materials · doi:10.1002/adma.202005805 · Cited by 104 · see citing works
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500318 · Cited by 102 · see citing works
Ultraflexible Neural Electrodes for Long-Lasting Intracortical Recording
He, F., Lycke, R., Ganji, M., Xie, C., & Luan, L. (2020) · iScience · doi:10.1016/j.isci.2020.101387 · Cited by 98 · see citing works
Mechanically adaptive nanocomposites for neural interfacing
Capadona, J. R., Tyler, D. J., Zorman, C. A., Rowan, S. J., & Weder, C. (2012) · MRS Bulletin · doi:10.1557/mrs.2012.97 · Cited by 101 · see citing works
Miniaturized optogenetic neural implants: a review
Fan, B., & Li, W. (2015) · Lab on a Chip · doi:10.1039/c5lc00588d · Cited by 99 · see citing works
Conformable polyimide-based μECoGs: Bringing the electrodes closer to the signal source
Vomero, M., Porto Cruz, M. F., Zucchini, E., Ciarpella, F., Delfino, E., Carli, S., Boehler, C., Asplund, M., Ricci, D., Fadiga, L., et al. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2020.120178 · Cited by 95 · see citing works
Thin microelectrodes reduce GFAP expression in the implant site in rodent somatosensory cortex
Stice, P., Gilletti, A., Panitch, A., & Muthuswamy, J. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/2/005 · Cited by 97 · see citing works
Hess‐Dunning, A., Capadona, J. R., Shanmuganathan, K., Hsu, L., Rowan, S. J., Weder, C., Tyler, D. J., & Zorman, C. A. (2011) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/21/5/054009 · Cited by 96 · see citing works
Widge, A. S., Jeffries‐EL, M., Cui, X. T., Lagenaur, C. F., & Matsuoka, Y. (2006) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2006.08.011 · Cited by 94 · see citing works
Hydrogels for next generation neural interfaces
Cheng, S., Zhu, R., & Xu, X. (2024) · Communications Materials · doi:10.1038/s43246-024-00541-0 · Cited by 95 · see citing works
Elastomeric and soft conducting microwires for implantable neural interfaces
Kolarcik, C. L., Luebben, S. D., Sapp, S. A., Hanner, J., Snyder, N. R., Kozai, T. D. Y., Chang, E., Nabity, J., Nabity, S. T., Lagenaur, C. F., et al. (2015) · Soft Matter · doi:10.1039/c5sm00174a · Cited by 87 · see citing works
Bioinspired Materials for In Vivo Bioelectronic Neural Interfaces
Woods, G. A., Rommelfanger, N. J., & Hong, G. (2020) · Matter · doi:10.1016/j.matt.2020.08.002 · Cited by 87 · see citing works
Sharafkhani, N., Kouzani, A. Z., Adams, S. D., Long, J. M., Lissorgues, G., Rousseau, L., & Orwa, J. O. (2021) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2021.109388 · Cited by 85 · see citing works
Insertion of Flexible Neural Probes Using Rigid Stiffeners Attached with Biodissolvable Adhesive
Felix, S. H., Shah, K. G., Tolosa, V. M., Sheth, H. J., Tooker, A., Delima, T. L., Jadhav, S. P., Frank, L. M., & Pannu, S. S. (2013) · Journal of Visualized Experiments · doi:10.3791/50609 · Cited by 87 · see citing works
Tissue-Compliant Neural Implants from Microfabricated Carbon Nanotube Multilayer Composite
Zhang, H., Patel, P. R., Xie, Z., Swanson, S. D., Wang, X., & Kotov, N. A. (2013) · ACS Nano · doi:10.1021/nn402074y · Cited by 88 · see citing works
In Vivo Chemical Sensors: Role of Biocompatibility on Performance and Utility
Soto, R. J., Hall, J. R., Brown, M. D., Taylor, J. B., & Schoenfisch, M. H. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b04251 · Cited by 85 · see citing works
Sung, C., Jeon, W., Nam, K. S., Kim, Y., Butt, H., & Park, S. (2020) · Journal of Materials Chemistry B · doi:10.1039/d0tb00872a · Cited by 85 · see citing works
A Parylene Neural Probe Array for Multi-Region Deep Brain Recordings
Wang, X., Weltman, A., Xu, H., Slingsby-Smith, Z., Lecomte, A., Scholten, K., Song, D. B., & Meng, E. F. (2020) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2020.3000235 · Cited by 83 · see citing works
Lai, H., Liao, L., Lin, C., Hsu, J.-H., He, X., Chen, Y., Chang, J.-Y., Chen, H.-F., Tsang, S., & Shih, Y. I. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/3/036001 · Cited by 83 · see citing works
A silk-based self-adaptive flexible opto-electro neural probe
Zhou, Y., Gu, C., Liang, J., Zhang, B., Yang, H., Zhou, Z., Li, M., Sun, L., Tao, T. H., & Wei, X. (2022) · Microsystems & Nanoengineering · doi:10.1038/s41378-022-00461-4 · Cited by 81 · see citing works
Flexible, diamond-based microelectrodes fabricated using the diamond growth side for neural sensing
Fan, B., Rusinek, C. A., Thompson, C. H., Setien, M. B., Guo, Y., Rechenberg, R., Gong, Y. P., Weber, A. J., Becker, M., Purcell, E. K., et al. (2020) · Microsystems & Nanoengineering · doi:10.1038/s41378-020-0155-1 · Cited by 78 · see citing works
Layered Nanocomposites from Gold Nanoparticles for Neural Prosthetic Devices
Zhang, H., Shih, J., Zhu, J. G., & Kotov, N. A. (2012) · Nano Letters · doi:10.1021/nl3015632 · Cited by 78 · see citing works
Melatonin improves quality and longevity of chronic neural recording
Golabchi, A., Wu, B., Li, X., Carlisle, D. L., Kozai, T. D. Y., Friedlander, R. M., & Cui, X. T. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.07.026 · Cited by 74 · see citing works
Martin, D. C. (2015) · MRS Communications · doi:10.1557/mrc.2015.17 · Cited by 77 · see citing works
Self-citationSU-8 based microprobes with integrated planar electrodes for enhanced neural depth recording
Altuna, A., de la Prida, L. M., Bellistri, E., Gabriel, G., Guimerà‐Brunet, A., Berganzo, J., Villa, R., & Fernández, L. J. (2012) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2012.03.039 · Cited by 76 · see citing works
Study of functional viability of SU-8-based microneedles for neural applications
Fernández, L. J., Altuna, A., Tijero, M., Gabriel, G., Villa, R., Rodríguez, M. J., Batlle, M., Vilares, R., Berganzo, J., & Blanco, F. J. (2009) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/19/2/025007 · Cited by 76 · see citing works
Smart Polymers for Neural Interfaces
Ware, T. H., Simon, D., Rennaker, R. L., & Voit, W. (2013) · Journal of macromolecular science. Part C, Reviews in macromolecular chemistry and physics/Journal of macromolecular science. Reviews in macromolecular chemistry and physics · doi:10.1080/15583724.2012.751924 · Cited by 74 · see citing works
Fekete, Z., & Pongracz, A. (2016) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2016.12.096 · Cited by 72 · see citing works
Metallo, C., White, R. D., & Trimmer, B. A. (2010) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2010.12.005 · Cited by 72 · see citing works
Welkenhuysen, M., Andrei, A., Ameye, L., Eberle, W., & Nuttin, B. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2166963 · Cited by 71 · see citing works
Ji, B., Xie, Z., Hong, W., Jiang, C., Guo, Z., Wang, L., Wang, X., Yang, B. T., & Liu, J. (2019) · Journal of Materiomics · doi:10.1016/j.jmat.2019.11.006 · Cited by 66 · see citing works
Sohal, H. S., Clowry, G. J., Jackson, A., O'Neill, A. G., & Baker, S. N. (2016) · PLoS ONE · doi:10.1371/journal.pone.0165606 · Cited by 67 · see citing works
Subcellular Neural Probes from Single-Crystal Gold Nanowires
Kang, M., Jung, S., Zhang, H., Kang, T., Kang, H., Yoo, Y., Hong, J.-P., Ahn, J.-P., Kwak, J., Jeon, D., et al. (2014) · ACS Nano · doi:10.1021/nn5024522 · Cited by 64 · see citing works
Brain–computer interfaces: an overview of the hardware to record neural signals from the cortex
Stieglitz, T., Rubehn, B., Henle, C., Kisban, S., Herwik, S., Ruther, P., & Schüettler, M. (2009) · Progress in brain research · doi:10.1016/s0079-6123(09)17521-0 · Cited by 63 · see citing works
- Authors
- Subbaroyan, J., Martin, D. C., & Kipke, D. R.
- Affiliations
- University of Michigan (US)
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2005-10-11
- Volume
- 2
- Issue
- 4
- Pages
- 103-113
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W1982562840
- Cited by
- 368
- References
- 46
- Citations by year
- 2026: 13 · 2025: 20 · 2024: 14 · 2023: 17 · 2022: 15 · 2021: 21 · 2020: 27 · 2019: 19 · 2018: 29 · 2017: 15 · 2016: 24 · 2015: 23 · 2014: 20 · 2013: 24 · 2012: 14
- FWCI
- 3.85
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- Deep Blue (University of Michigan) (submittedVersion)
- Funders
- National Science Foundation (9986866)
- National Aeronautics and Space Administration
- National Institutes of Health
- National Nanotechnology Coordinating Office
- National Institute of Biomedical Imaging and Bioengineering (P41 EB-00 230-10)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Cell Biology › Cellular Mechanics and Interactions
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Keywords
- implantable microelectrodes, cerebral cortex, interfacial strain, silicon probes, tissue encapsulation, electrophysiological recording, neurostimulation
- MeSH terms
- Animals, Biomechanical Phenomena, Cerebral Cortex, Computer Simulation, Elasticity, Electrodes, Implanted, Humans, Microelectrodes, Models, Neurological, Motion, Physical Stimulation, Stress, Mechanical, Finite Element Analysis, Mechanotransduction, Cellular
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review