Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices
Open accessAbstract
Deep tendon reflexes are one of the main components of the clinical nervous system examinations. These assessments are inexpensive and quick. However, evaluation can be subjective and qualitative. This study aimed to objectively evaluate hyperreflexia of the patellar tendon reflex using portable mechanomyography (MMG) and electromyography (EMG) devices. This study included 10 preoperative patients (20 legs) who had a pathology that could cause bilateral patellar tendon hyperreflexia and 12 healthy volunteers (24 legs) with no prior history of neurological disorders. We attached MMG/EMG sensors onto the quadriceps and tapped the patellar tendon with maximal and constant force. Our results showed a significantly high amplitude of the root mean square (RMS) and low frequency of the mean power frequency (MPF) in the rectus femoris, vastus medialis, and vastus lateralis muscles in both EMG and MMG with both maximal and constant force. Especially in the patients with cervical and thoracic myelopathy, the receiver operating characteristic (ROC) curve for diagnosing hyperreflexia of the patellar tendon showed a moderate to very high area under the curve for all EMG–RMS, EMG–MPF, MMG–RMS, and MMG–MPF values. The use of EMG and MMG for objectively quantifying the patellar tendon reflex is simple and desirable for future clinical applications and could help diagnose neurological disorders.
Overview
CoChat Summary
TL;DR
Portable mechanomyography (MMG) and electromyography (EMG) sensors detected higher root mean square (RMS) and lower mean power frequency (MPF) in the quadriceps of patients with patellar tendon hyperreflexia compared with healthy controls. The authors conclude that combined EMG/MMG may objectively quantify the patellar tendon reflex and aid diagnosis of neurological disorders.
Key findings
- In 10 preoperative patients (20 legs; 9 cervical myelopathy, 1 thoracic myelopathy) versus 12 healthy volunteers (24 legs), the hyperreflexia group had NINDS scores of 3+ in 15 legs and 4+ in 5 legs, while controls had 2+ in 6 legs and 1+ in 18 legs.
- Both maximal tap force (Experiment 1) and constant tap force (Experiment 2; mean tap force 142.6 ± 7.0 N) produced significantly higher RMS and lower MPF in the rectus femoris, vastus medialis, and vastus lateralis for both EMG and MMG in the hyperreflexia group than in controls.
- ROC analysis: in Experiment 1, AUC was over 0.9 for each parameter in the rectus femoris and vastus lateralis; in the vastus medialis, EMG–RMS and MMG–RMS were over 0.9, while EMG–MPF was 0.82 and MMG–MPF was 0.79. In Experiment 2, AUC was over 0.9 for all parameters in the three muscles.
- Power analysis showed effect sizes and power over 0.8 for both experiments; ICC was almost perfect for most parameters in the hyperreflexia group, except moderate reliability for EMG–MPF of the vastus lateralis, and controls showed substantial to almost perfect reliability depending on parameter.
Methods
- Participants were 10 preoperative patients (20 legs, age 62.0 ± 8.23 years, seven men, three women) with pathologies that could cause bilateral patellar tendon hyperreflexia and 12 healthy volunteers (24 legs, age 33.83 ± 4.71 years, nine men, three women) with no prior neurological disorders; reflex was scored with the NINDS scale.
- Sensors were placed on the rectus femoris, vastus medialis, and vastus lateralis; Experiment 1 used a clinical hammer for maximal patellar tendon tap, and Experiment 2 used a custom tool with an iron ball (diameter 2 cm; weight 45 g) dropped at 90° for constant force.
- Data processing included band-pass filtering, automatic peak detection with visual checks, RMS and MPF calculation, and fast Fourier transformation in MATLAB R2019b; statistical analysis used Kolmogorov–Smirnov, Mann–Whitney U, Cohen’s d, power analysis, ICC, and SEM.
Limitations
- Stated: the relationships between frequency-domain analyses and muscle contraction need further study; patients’ diseases were limited to spinal pathologies, and atypical cases were not examined, making it difficult to distinguish pathological hyperreflexia from healthy hyperreflexia and vice versa.
- Stated: the Jendrassik maneuver was not used, and the study included older patients with central nervous pathology and younger healthy volunteers; aging decreases reflex response, and further evidence including age, height, and sex is needed.
- Stated: quadriceps muscle volume and subcutaneous thigh fat may affect MMG results, and although BMI did not differ significantly between groups, local thigh fat and muscle were not assessed. Evident from the text: EMG latency was difficult to evaluate, and the timing of peak EMG and MMG during patellar tendon reflex was not evaluated.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures







Verification
Claims the paper makes
1 Valid · 0 Not valid · 14 Not checked · 25 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The H group showed significantly higher amplitudes in the root mean square (RMS) and lower frequencies in the mean power frequency (MPF) in the rectus femoris (RF), vastus medialis (VM), and vastus lateralis (VL) for both EMG and MMG than the C group (Fig. 1).”
§ Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices > Results > Experiment 1: PTR movement with maximal tap force
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The mean tap force was 142.6 ± 7.0 N using our tool.”
§ Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices > Results > Experiment 2: PTR movement with constant tap force
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“When tapped by our tool, the H group showed significantly higher amplitudes in RMS and lower frequencies in MPF in all muscles for both EMG and MMG than the C group (Fig. 2).”
§ Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices > Results > Experiment 2: PTR movement with constant tap force
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In experiment 1, the AUC of each parameter was over 0.9 in the RF and VL.”
§ Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices > Results > ROC curve
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In experiment 2, the AUC of each parameter was over 0.9 in the three muscles.”
§ Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices > Results > ROC curve
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
8 more claims
Create a free account to see all 40 claims and to verify the 19 references still open.
- Probably not a substantial claim. 14
- It may state two separate claims as one. 21
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated major: it would touch a result — a value a result rests on, reported and recomputed p-values on opposite sides of the significance level, two passages that conflict with no sign which is right, or a cited work retracted before this paper was published.
In the paper
§ Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices > Discussion
“the range of 0.5 s from the peak was analyzed.”
§ Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices > Methods > Data processing
“We analyzed the waveforms at 0.5 ms after each peak, and the RMS and MPF values of this period were calculated.”
Citations
Cites
Clinical Methods: The History, Physical, and Laboratory Examinations
Kruse, J. A. (1990) · JAMA: The Journal of the American Medical Association · doi:10.1001/jama.1990.03450210108045 · Cited by 2,029
Clinical Methods: The History, Physical, and Laboratory Examinations
(1990) · Annals of Internal Medicine · doi:10.7326/0003-4819-113-7-563_2 · Cited by 1,396
Muscle sound: bases for the introduction of a mechanomyographic signal in muscle studies.
Orizio, C. (1993) · PubMed · Cited by 523
Alternatives to P value: confidence interval and effect size
Lee, D. K. (2016) · Korean Journal of Anesthesiology · doi:10.4097/kjae.2016.69.6.555 · Cited by 479
Stratford, P. W., & Goldsmith, C. H. (1997) · Physical Therapy · doi:10.1093/ptj/77.7.745 · Cited by 455
Acoustic myography: A noninvasive monitor of motor unit fatigue
Barry, D. T., Geiringer, S. R., & Ball, R. D. (1985) · Muscle & Nerve · doi:10.1002/mus.880080303 · Cited by 208
Harrop, J. S., Naroji, S., Maltenfort, M., Anderson, D. G., Albert, T., Ratliff, J. K., Ponnappan, R. K., Rihn, J. A., Smith, H. E., Hilibrand, A., et al. (2010) · Spine · doi:10.1097/brs.0b013e3181b723af · Cited by 130
Xie, H.-B., Zheng, Y.-P., & Guo, J.-Y. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/5/002 · Cited by 118
Acoustic myography for investigating human skeletal muscle fatigue
Stokes, M. J., & Dalton, P. A. (1991) · Journal of Applied Physiology · doi:10.1152/jappl.1991.71.4.1422 · Cited by 106
The clinical identification of peripheral neuropathy among older persons
Richardson, J. K. (2002) · Archives of Physical Medicine and Rehabilitation · doi:10.1053/apmr.2002.35656 · Cited by 95
Cited by
Dwivedi, A., Groll, H., & Beckerle, P. (2022) · Sensors · doi:10.3390/s22176319 · Cited by 22
Fukuhara, S., Kawashima, T., & Oka, H. (2021) · Scientific Reports · doi:10.1038/s41598-021-00671-2 · Cited by 21
A Wireless Multi-Layered EMG/MMG/NIRS Sensor for Muscular Activity Evaluation
Kimoto, A., Fujiyama, H., & Machida, M. (2023) · Sensors · doi:10.3390/s23031539 · Cited by 14
Benlidayı, İ. C., Deniz, V., Ornek, C., & Sarıyıldız, A. (2025) · Rheumatology International · doi:10.1007/s00296-025-05808-4 · Cited by 9
Combined tendon reflex and motor evoked potential recordings in amyotrophic lateral sclerosis
Truffert, A., Sukockienė, E., Desmaison, A., Ališauskienė, M., Ferfoglia, R. I., & Guy, N. (2023) · Clinical Neurophysiology · doi:10.1016/j.clinph.2022.12.013 · Cited by 5
Metadata
- Authors
- Tsuji, H., Misawa, H., Takigawa, T., Tetsunaga, T., Yamane, K., Oda, Y., & Ozaki, T.
- Affiliations
- Okayama University (JP)
- Okayama University Hospital (JP)
- Japanese Red Cross Kobe Hospital (JP)
- Corresponding author
- Tsuji, Hironori
- Venue
- Scientific Reports
- Published
- 2021-01-27
- Volume
- 11
- Issue
- 1
- Pages
- 2284-2284
- ISSN
- 2045-2322 (ISSN-L)
- PMC
- PMC7840930
- OpenAlex
- W3125845873
- Cited by
- 16
- References
- 42
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 2 · 2023: 5 · 2022: 3 · 2021: 3
- FWCI
- 0.99
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- Scientific Reports (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Institutional Repositories DataBase (IRDB) (submittedVersion)
- Okayama University Scientific Achievement Repository (Okayama University) (submittedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- NCP Inc.
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Health Sciences › Medicine › Clinical Neurology › Peripheral Nerve Disorders
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Muscle Physiology and Disorders
- Keywords
- patellar reflex, hyperreflexia, mechanomyography, electromyography, cervical myelopathy, thoracic myelopathy, quadriceps muscle activity, root mean square amplitude, mean power frequency, neurological disease diagnosis
- MeSH terms
- Aged, Electromyography, Female, Humans, Male, Middle Aged, Myography, Neurosciences, Patella, ROC Curve, Patellar Ligament
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices. https://app.cochat.ai/papers/10.1038/s41598-021-81874-5
- The paper
- Tsuji, H., Misawa, H., Takigawa, T., Tetsunaga, T., Yamane, K., Oda, Y., & Ozaki, T. (2021). Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices. Scientific Reports. https://doi.org/10.1038/s41598-021-81874-5
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.