Functional and Biological Test of a 20 Channel Implantable Stimulator in Sheep in View of Functional Electrical Stimulation Walking for Spinal Cord Injured Persons
Abstract
Abstract: A newly developed implantable stimulator with 20 output channels, mainly intended for the stimulation of lower extremities in paraplegics, was implanted in 6 sheep over a time period of 26 weeks. Five epineural electrodes each were used to contact various nerves at different locations to elicit hip and knee extension and flexion and to make carrousel and selective stimulation possible. Different electrode application strategies in view of paraplegic standing and walking were investigated. Additional implanted electrodes allowed M‐wave monitoring for selectivity investigations in 3 sheep. Stimulator, electrode leads, and electrodes proved to be reliable. Selective stimulation with electrodes placed on the trunk of the sciatic nerve could be demonstrated but with bad reproducibility. Histological investigation of the tissues surrounding electrodes and leads showed the expected stable foreign body response. Strong hip and knee extension could be gained in all cases while only weak flexion forces could be elicited in most cases. Muscle biopsies showed that daily stimulation for 8 h at threshold level caused an increase in muscle Type I fibers and a decrease in Type IIc fibers. Implants and electrodes fulfill the most important functional and biological criteria for their clinical application for paraplegic walking. The intention to provide selective flexion functions via epineural stimulation could not be demonstrated sufficiently in this animal model.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In a 26-week animal study with 6 sheep, epineural electrodes increased type I muscle fiber distribution compared to the contralateral side, with no change in type II fiber distribution.
“Muscle biopsies at the end showed increased type I muscle fiber distribution compared to the contralateral side, with no change in type II fiber distribution [46].”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [46]
Cites
Smith, B., Tang, Z., Johnson, M. W., Pourmehdi, S., Gazdik, M. M., Buckett, J. R., & Peckham, P. H. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.664202 · Cited by 270 · see citing works
Muscle selection and walking performance of multichannel FES systems for ambulation in paraplegia
Kobetic, R., Triolo, R. J., & Marsolais, E. B. (1997) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.559346 · Cited by 185 · see citing works
Kobetic, R., Sharma, M., Triolo, R. J., Uhlir, J. P., Bieri, C., Wibowo, M., Polando, G., Marsolais, E. B., Davis, J. A., & Ferguson, K. A. (1999) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.808942 · Cited by 155 · see citing works
Long-term follow-up of patients with sacral anterior root stimulator implants
Brindley, G. S., & Rushton, D. N. (1990) · Spinal Cord · doi:10.1038/sc.1990.63 · Cited by 140 · see citing works
Micromodular implants to provide electrical stimulation of paralyzed muscles and limbs
Cameron, T., Loeb, G. E., Peck, R. A., Schulman, J. H., Strojnik, P., & Troyk, P. R. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.623047 · Cited by 136 · see citing works
Development of a Practical Electrical Stimulation System for Restoring Gait in the Paralyzed Patient
Marsolais, E. B., & Kobetic, R. (1988) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-198808000-00009 · Cited by 121 · see citing works
Solomonow, M., Aguilar, E., Reisin, E., Baratta, R. V., Best, R., Coetzee, T., & D'Ambrosia, R. D. (1997) · Orthopedics · doi:10.3928/0147-7447-19970401-08 · Cited by 118 · see citing works
HOLLE, J., Frey, M., Gruber, H., Kern, H., Stöhr, H., & THOMA, H. (1984) · Orthopedics · doi:10.3928/0147-7447-19840701-09 · Cited by 89 · see citing works
Self-citation(1996) · PubMed · Cited by 79 · see citing works
Programmed Six-Channel Electrical Stimulator for Complex Stimulation of Leg Muscles During Walking
Strojnik, P., Kralj, A. R., & Ursic, I. (1979) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1979.326520 · Cited by 77 · see citing works
Electronic design of a multichannel programmable implant for neuromuscular electrical stimulation
Arabi, K., & Sawan, M. (1999) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.769411 · Cited by 75 · see citing works
Satisfaction with and usage of a hand neuroprosthesis
Wuolle, K. S., Van Doren, C. L., Bryden, A. M., Hunter Peckham, P., Keith, M. W., Kilgore, K. L., & Grill, J. H. (1999) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(99)90123-5 · Cited by 70 · see citing works
Clinical use of percutaneous intramuscular electrodes for functional electrical stimulation
Shimada, Y., Satô, K., Kagaya, H., et al. (1996) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(96)90061-1 · Cited by 59 · see citing works
Davis, R., Houdayer, T., Andrews, B., Emmons, S. E., & Patrick, J. F. (1998) · Stereotactic and Functional Neurosurgery · doi:10.1159/000099889 · Cited by 46 · see citing works
Technology and Long-term Application of an Epineural Electrode
THOMA, H., CIRSCH, W., HOLLE, J., & Mayr, W. (1989) · ASAIO Transactions · doi:10.1097/00002480-198907000-00103 · Cited by 43 · see citing works
Self-citationNeuroprostheses for leg function after spinal-cord injury
Donaldson, N., Rushton, D., & Tromans, T. (1997) · The Lancet · doi:10.1016/s0140-6736(97)24036-5 · Cited by 29 · see citing works
Mayr, W., Bijak, M., Girsch, W., et al. (1994) · PubMed · Cited by 24 · see citing works
Self-citationImplantable stimulation system dedicated for neural selective stimulation
Robin, S., Sawan, M., Abdel-Gawad, M., AbdelBaky, T. M., & ELHILALI, M. M. (1998) · Medical & Biological Engineering & Computing · doi:10.1007/bf02523220 · Cited by 23 · see citing works
Myelinated sensory and alpha motor axon regeneration in peripheral nerve neuromas
Macias, M. Y., Lehman, C. T., Sanger, J. R., & Riley, D. A. (1998) · Muscle & Nerve · doi:10.1002/(sici)1097-4598(199812)21:12 · Cited by 23 · see citing works
Winter, K.-F., Hartmann, R., & Klinke, R. (1998) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(97)00160-x · Cited by 21 · see citing works
Useful applications and limits of battery powered implants in functional electrical stimulations.
Lanmüller, H., Bijak, M., Mayr, W., et al. (1997) · PubMed · Cited by 21 · see citing works
Self-citationMultichannel Stimulation of Phrenic Nerves by Epineural Electrodes
Mayr, W., Bijak, M., Girsch, W., et al. (1993) · ASAIO Journal · doi:10.1097/00002480-199339030-00111 · Cited by 19 · see citing works
Self-citationGirsch, W., Koller, R., Gruber, H., Holle, J., Liegl, C., Losert, U., Mayr, W., & Thoma, H. (1991) · Journal of Neurosurgery · doi:10.3171/jns.1991.74.4.0636 · Cited by 18 · see citing works
Self-citationVienna Phrenic Pacemaker - Experience with Diaphragm Pacing in Children
Girsch, W., Koller, R., Holle, J., et al. (1996) · European Journal of Pediatric Surgery · doi:10.1055/s-2008-1066491 · Cited by 17 · see citing works
Self-citationImplant provision of key, pinch and power grips in a C6 tetraplegic
Perkins, T. A., Brindley, G. S., Donaldson, N., et al. (1994) · Medical & Biological Engineering & Computing · doi:10.1007/bf02524686 · Cited by 16 · see citing works
Lanmüller, H., Sauermann, S., Unger, E., et al. (1999) · Artificial Organs · doi:10.1046/j.1525-1594.1999.06359.x · Cited by 12 · see citing works
Self-citationMultifunctional Implantable Nerve Stimulator for Cardiac Assistance by Skeletal Muscle
Lanmüller, H., Sauermann, S., Unger, E., Schnetz, G., Mayr, W., Bijak, M., & Girsch, W. (1999) · Artificial Organs · doi:10.1046/j.1525-1594.1999.06258.x · Cited by 12 · see citing works
Self-citationPersonal Computer Supported Eight Channel Surface Stimulator for Paraplegic Walking: First Results
Bijak, M., Höfer, C., Lanmüller, H., et al. (1999) · Artificial Organs · doi:10.1046/j.1525-1594.1999.06370.x · Cited by 10 · see citing works
Self-citationPourmehdi, S., Strojnik, P., Peckham, H., et al. (1999) · Artificial Organs · doi:10.1046/j.1525-1594.1999.06358.x · Cited by 8 · see citing works
Monitoring of Fes-Induced Muscle Activity by Continuous Emg-Recording
Girsch, W., Bijak, M., Heger, G., et al. (1995) · The International Journal of Artificial Organs · doi:10.1177/039139889501800608 · Cited by 7 · see citing works
Self-citationState of the art of implanted multichannel devices to mobilize paraplegics
THOMA, H., Frey, M., Holle, J., et al. (1987) · International Journal of Rehabilitation Research · doi:10.1097/00004356-198712005-00018 · Cited by 6 · see citing works
Self-citationState of the art of implanted multichannel devices to mobilize paraplegics
Thoma, H., Frey, M., HOLLE, J., et al. (1987) · International Journal of Rehabilitation Research · doi:10.1097/00004356-198700105-00018 · Cited by 4 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113 · see citing works
- Authors
- Bijak, M., Mayr, W., Girsch, W., Lanmüller, H., Unger, E., Stöhr, H., Thoma, H., & Plenk, H.
- Affiliations
- University of Vienna (AT)
- Corresponding author
- Bijak, Manfred
- Venue
- Artificial Organs
- Publisher
- Wiley
- Published
- 2001-06
- Volume
- 25
- Issue
- 6
- Pages
- 467-474
- ISSN
- 0160-564X (ISSN-L)
- 1525-1594
- OpenAlex
- W2006496847
- Cited by
- 8
- References
- 32
- Citations by year
- 2020: 1 · 2019: 1 · 2017: 1 · 2014: 1
- FWCI
- 0.37
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Artificial Organs (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Pathology and Forensic Medicine › Spinal Cord Injury Research
- Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
- Keywords
- implantable neural stimulator, functional electrical stimulation, spinal cord injury, paraplegia, sciatic nerve, selective stimulation, foreign body reaction, muscle biopsy, gait assistance, sheep model
- MeSH terms
- Animals, Electrodes, Electric Stimulation Therapy, Leg, Sheep, Spinal Cord Injuries, Muscle, Skeletal
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review