Histological assessment of nerve lesions caused by epineurial electrode application in rat sciatic nerve
Abstract
✓ The left sciatic nerve of 36 rats was exposed and four ring-shaped stainless steel wire electrodes were sutured to the epineurium of each nerve in the same manner as performed clinically for “carousel stimulation” in man. The rats were sacrificed 10 days (Group 1), 3 weeks (Group 2), or 3 months (Group 3) after implantation. The electrodes were excised, the nerves were embedded in Epon, and semithin sections were obtained for histological and planimetric assessment of lesions caused by the epineurially sutured electrodes. The right sciatic nerves served as controls. The total area of neural tissue within the perineurium was determined at three levels: at the site of the electrodes, 8 mm proximal, and 8 mm distal. The area of neural tissue damaged by the surgical procedure was expressed as a percentage of the total area. In Group 1, nine of 12 nerves showed lesions ranging from 0.39% to 25.39% of the total area of neural tissue, in Group 2 eight of 11 sciatic nerves showed lesions ranging from 0.24% to 13.03% of the total area, and in Group 3 five of 12 nerves showed lesions ranging from 0.21% to 4.96% of the total area. The pathologically altered areas in Groups 2 and 3 exhibited distinct signs of nerve fiber regeneration. The reasons for the decrease in damage from Group 1 to Group 3 and the clinical implications of the results for long-term electrical stimulation are discussed.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In rat sciatic nerve, altered fascicular area after cuff electrode implantation gradually recovered from 4.74% at 10 days to 0.57% at 3 months.
“Girsch et al. assessed nerve lesions in the rat sciatic nerve 10 days, 3 weeks, and 3 months after cuff electrode implantation, and noted a gradual recovery from 4.74 to 0.57% of altered fascicular area.”
From Safety of long-term electrical peripheral nerve stimulation: review of the state of the art (2019) · cites this paper as [53]
Neural damage induced by implantation and presence of an extraneural electrode is reversible.
“Girsch et al. and Larsen et al. showed that neural damage induced by the implantation and presence of an extraneural electrode is reversible [53, 54].”
From Safety of long-term electrical peripheral nerve stimulation: review of the state of the art (2019) · cites this paper as [53]
Cites
Fiber composition of the rat sciatic nerve
Schmalbruch, H. (1986) · The Anatomical Record · doi:10.1002/ar.1092150111 · Cited by 402 · see citing works
Kempe, L. G. (1979) · Military Medicine · doi:10.1093/milmed/144.9.613 · Cited by 368 · see citing works
Altered ratio between axon diameter and myelin sheath thickness in regenerated nerve fibers
̈der, J. M. S. (1972) · Brain Research · doi:10.1016/0006-8993(72)90215-6 · Cited by 192 · see citing works
Agnew, W. F., McCreery, D. B., Yuen, T. G. H., & Bullara, L. A. (1989) · Annals of Biomedical Engineering · doi:10.1007/bf02364272 · Cited by 185 · see citing works
Experimental correction of footdrop by electrical stimulation of the peroneal nerve
Waters, R. L., McNeal, D. R., & Perry, J. (1975) · Journal of Bone and Joint Surgery · doi:10.2106/00004623-197557080-00002 · Cited by 152 · see citing works
Total Ventilatory Support in a Quadriplegic Patient with Radiofrequency Electrophrenic Respiration
Glenn, W. W. L., Holcomb, W. G., Aj, M., O'Hare, J. M., Hogan, J., & Yasuda, R. (1972) · New England Journal of Medicine · doi:10.1056/nejm197203092861004 · Cited by 116 · see citing works
Glenn, W. W. L., Holcomb, W. G., Hogan, J., Matano, I., Gee, J. B. L., Motoyama, E. K., Kim, C. S., Pokier, R. S., & Forbes, G. S. (1973) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1016/s0022-5223(19)40583-7 · Cited by 104 · see citing works
Bowman, B. R., & Erickson, R. C. (1985) · Annals of Biomedical Engineering · doi:10.1007/bf02371251 · Cited by 66 · see citing works
Kim, J. H., Manuelidis, E. E., Glenn, W. W. L., & Kaneyuki, T. (1976) · PubMed · Cited by 54 · see citing works
Light and electron microscopic studies of phrenic nerves after long-term electrical stimulation
Kim, J. H., Manuelidis, E. E., Glenn, W. W. L., Fukuda, Y., Cole, D. S., & Hogan, J. F. (1983) · Journal of Neurosurgery · doi:10.3171/jns.1983.58.1.0084 · Cited by 47 · see citing works
Functional electrical stimulation
Grant, L. J. (1988) · IEE Review · doi:10.1049/ir:19880186 · Cited by 37 · see citing works
The Swedish Experience in Phrenic Nerve Stimulation
Fodstad, H. (1987) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1987.tb05957.x · Cited by 30 · see citing works
Hughes, G. B., Bottomy, M. B., Dickins, J. R. E., Jackson, C. G., Sismanis, A. A., & Glasscock, M. E. (1980) · American Journal of Otolaryngology · doi:10.1016/s0196-0709(80)80018-4 · Cited by 28 · see citing works
Yuen, T. G. H., Agnew, W. F., & Bullara, L. A. (1984) · Neurosurgery · doi:10.1227/00006123-198404000-00010 · Cited by 23 · see citing works
Hughes, G. B., Bottomy, M. B., Jackson, C. G., Glasscock, M. E., & Sismanis, A. A. (1981) · Otolaryngology · doi:10.1177/019459988108900515 · Cited by 19 · see citing works
HOLLE, J., Mortiz, E., THOMA, H., & Lischka, A. R. (1974) · PubMed · Cited by 18 · see citing works
Self-citationHistologic Changes in Continuous, Long-Term Electrical Stimulation of a Peripheral Nerve
Hershberg, P. I., Sohn, D., Agrawal, G. P., & Kantrowitz, Á. (1967) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1967.4502479 · Cited by 15 · see citing works
Diaphragm Pacing as an Adjunct in Respiratory Insufficiency
Young, R. F. (1978) · Neurosurgery · doi:10.1227/00006123-197801000-00009 · Cited by 8 · see citing works
Influence of Long-Term Low Direct Current on Rat Ischiadic Nerves
Rosenkranz, D., Fenzl, G., HOLLE, J., Lack, W., Losert, U. M., & THOMA, H. (1986) · Stereotactic and Functional Neurosurgery · doi:10.1159/000100128 · Cited by 5 · see citing works
Self-citationElectrical stimulation of motor nerves: results of an animal experiment.
Rs, B. (1985) · PubMed · Cited by 2 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Safety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198 · see citing works
Paradigms for restoration of somatosensory feedback via stimulation of the peripheral nervous system
Pasluosta, C. F., Kiele, P., & Stieglitz, T. (2017) · Clinical Neurophysiology · doi:10.1016/j.clinph.2017.12.027 · Cited by 97 · see citing works
Raspopović, S., Petrini, F. M., Żelechowski, M., & Valle, G. (2016) · Proceedings of the IEEE · doi:10.1109/jproc.2016.2600560 · Cited by 84 · see citing works
Analysis of current density and related parameters in spinal cord stimulation
Wesselink, W. A., Holsheimer, J., & Boom, H. B. K. (1998) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.681186 · Cited by 64 · see citing works
Mayr, W., Bijak, M., Rafolt, D., Sauermann, S., Unger, E., & Lanmüller, H. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00014-5 · Cited by 57 · see citing works
Self-citationYoshida, K., Farina, D., Akay, M., & Jensen, W. (2010) · Proceedings of the IEEE · doi:10.1109/jproc.2009.2038613 · Cited by 53 · see citing works
Functional and Histological Effects of Chronic Neural Electrode Implantation
Sahyouni, R., Chang, D. T., Moshtaghi, O., Mahmoodi, A., Djalilian, H. R., & Lin, H. W. (2017) · Laryngoscope Investigative Otolaryngology · doi:10.1002/lio2.66 · Cited by 27 · see citing works
KOLLER, R., GIRSCH, W., LIEGL, C., CRUDER, H., HOLLE, J., LOSERT, U., MAYR, W., & THOMA, H. (1992) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1992.tb02906.x · Cited by 17 · see citing works
Self-citationFunctional electrical stimulation on paraplegic patients
Kern, H. (2014) · European Journal of Translational Myology · doi:10.4081/ejtm.2014.2940 · Cited by 15 · see citing works
Evaluation of a Thin-film Peripheral Nerve Cuff Electrode
Walter, J. S., McLane, J. A., Cai, W., Khan, T. A., & Cogan, S. F. (1995) · Journal of Spinal Cord Medicine · doi:10.1080/10790268.1995.11719377 · Cited by 9 · see citing works
Bijak, M., Mayr, W., Girsch, W., Lanmüller, H., Unger, E., Stöhr, H., Thoma, H., & Plenk, H. (2001) · Artificial Organs · doi:10.1046/j.1525-1594.2001.025006467.x · Cited by 8 · see citing works
Self-citationROLLER, R., Girsch, W., Huber, L., Rab, M., Stöhr, H., Schima, H., Losert, U. M., Wolner, E., & THOMA, H. (1994) · Artificial Organs · doi:10.1111/j.1525-1594.1994.tb03371.x · Cited by 8 · see citing works
Self-citationKoller, R., Girsch, W., Huber, L., Rab, M., Stöhr, H., Schima, H., ALEXANDER, M., ROKITANSKY, Losert, U. M., & Wolner, E. (1996) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1996.tb03314.x · Cited by 7 · see citing works
Self-citationVideo-Assisted Application of “Epineurial” Electrodes for Functional Electrical Stimulation
Girsch, W., Kamolz, L., Lanmüller, H., & Windberger, U. B. (2001) · Springer eBooks · doi:10.1007/978-3-7091-6209-5_12
Self-citationExamination of epimysial leads used to stimulate the latissimus dorsi muscle
Hansen, S. B., Slater, A. D., Rodriguez, C. A. R., Spinale, F. G., & Santamore, W. P. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.495285
Praeklinische Testung eines implantierbaren 20-Kanal Nervstimulators
Bijak, M., Girsch, W., HOLLE, J., Lanmüller, H., Mayr, W., Plenk, H., Jr., Schmutterer, C., Stöhr, H., THOMA, H., & Unger, E. (1996) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmte.1996.41.s1.390
Self-citationAdvancing Peripheral Nerve Interfaces in a Large Animal Model
Andreis, F. R. (2022) · doi:10.54337/aau478976327
Jeong (2000) · doi:10.1002/(sici)1097-4636(200005)50:2
- Authors
- Girsch, W., Koller, R., Gruber, H., Holle, J., Liegl, C., Losert, U., Mayr, W., & Thoma, H.
- Affiliations
- University of Vienna (AT)
- Venue
- Journal of Neurosurgery
- Published
- 1991-04
- Volume
- 74
- Issue
- 4
- Pages
- 636-642
- ISSN
- 0022-3085 (ISSN-L)
- OpenAlex
- W2130809133
- Cited by
- 18
- References
- 20
- Citations by year
- 2022: 1 · 2019: 1 · 2017: 2 · 2016: 1 · 2014: 1
- FWCI
- 0.26
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of neurosurgery (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Nerve injury and regeneration
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- sciatic nerve, nerve injury, perineurium, nerve fiber regeneration, nerve histology, rat model
- MeSH terms
- Animals, Electric Stimulation, Electrodes, Implanted, Female, Rats, Inbred Strains, Sciatic Nerve, Time Factors, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review