A Dual-Chamber Leadless Pacemaker
Open accessAbstract
BACKGROUND: Single-chamber ventricular leadless pacemakers do not support atrial pacing or consistent atrioventricular synchrony. A dual-chamber leadless pacemaker system consisting of two devices implanted percutaneously, one in the right atrium and one in the right ventricle, would make leadless pacemaker therapy a treatment option for a wider range of indications. METHODS: We conducted a prospective, multicenter, single-group study to evaluate the safety and performance of a dual-chamber leadless pacemaker system. Patients with a conventional indication for dual-chamber pacing were eligible for participation. The primary safety end point was freedom from complications (i.e., device- or procedure-related serious adverse events) at 90 days. The first primary performance end point was a combination of adequate atrial capture threshold and sensing amplitude at 3 months. The second primary performance end point was at least 70% atrioventricular synchrony at 3 months while the patient was sitting. RESULTS: Among the 300 patients enrolled, 190 (63.3%) had sinus-node dysfunction and 100 (33.3%) had atrioventricular block as the primary pacing indication. The implantation procedure was successful (i.e., two functioning leadless pacemakers were implanted and had established implant-to-implant communication) in 295 patients (98.3%). A total of 35 device- or procedure-related serious adverse events occurred in 29 patients. The primary safety end point was met in 271 patients (90.3%; 95% confidence interval [CI], 87.0 to 93.7), which exceeded the performance goal of 78% (P<0.001). The first primary performance end point was met in 90.2% of the patients (95% CI, 86.8 to 93.6), which exceeded the performance goal of 82.5% (P<0.001). The mean (±SD) atrial capture threshold was 0.82±0.70 V, and the mean P-wave amplitude was 3.58±1.88 mV. Of the 21 patients (7%) with a P-wave amplitude of less than 1.0 mV, none required device revision for inadequate sensing. At least 70% atrioventricular synchrony was achieved in 97.3% of the patients (95% CI, 95.4 to 99.3), which exceeded the performance goal of 83% (P<0.001). CONCLUSIONS: The dual-chamber leadless pacemaker system met the primary safety end point and provided atrial pacing and reliable atrioventricular synchrony for 3 months after implantation. (Funded by Abbott Medical; Aveir DR i2i ClinicalTrials.gov number, NCT05252702.).
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab364 · Cited by 2,019
Self-citationMOND, H. G., & PROCLEMER, A. (2011) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2011.03150.x · Cited by 984
A Leadless Intracardiac Transcatheter Pacing System
Reynolds, D., Duray, G. Z., Omar, R., Soejima, K., Neuzil, P., Zhang, S., Narasimhan, C., Steinwender, C., Brugada, J., Lloyd, M., et al. (2016) · New England Journal of Medicine · doi:10.1056/nejmoa1511643 · Cited by 933
Self-citationUdo, E. O., Zuithoff, N. P. A., van Hemel, N. M., de Cock, C. C., Hendriks, T., Doevendans, P. A., & Moons, K. G. M. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2011.12.014 · Cited by 597
Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker
Reddy, V. Y., Exner, D. V., Cantillon, D. J., Doshi, R., Bunch, T. J., Tomassoni, G. F., Friedman, P. A., Estes, N. A. M., Ip, J., Niazi, I., et al. (2015) · New England Journal of Medicine · doi:10.1056/nejmoa1507192 · Cited by 539
Self-citationEl-Chami, M. F., Al-Samadi, F., Clementy, N., Garweg, C., Martinez-Sande, J. L., Piccini, J. P., Iacopino, S., Lloyd, M., Viñolas Prat, X., Jacobsen, M. D., et al. (2018) · Heart Rhythm · doi:10.1016/j.hrthm.2018.08.005 · Cited by 355
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.043 · Cited by 337
Wilkoff, B. L., Bello, D., Taborsky, M., Vymazal, J., Kanal, E., Heuer, H., Hecking, K., Johnson, W. B., Young, W., Ramza, B., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2010.10.002 · Cited by 266
Atrioventricular Synchronous Pacing Using a Leadless Ventricular Pacemaker
Steinwender, C., Khelae, S. K., Garweg, C., Chan, J. Y. S., Ritter, P., Johansen, J. B., Sagi, V., Epstein, L. M., Piccini, J. P., Pascual, M., et al. (2020) · JACC: Clinical Electrophysiology · doi:10.1016/j.jacep.2019.10.017 · Cited by 257
Complications arising after implantation of DDD pacemakers: the MOST experience
Ellenbogen, K. A., Hellkamp, A. S., Wilkoff, B. L., Camunãs, J. L., Love, J. C., Hadjis, T. A., Lee, K. L., & Lamas, G. A. (2003) · The American Journal of Cardiology · doi:10.1016/s0002-9149(03)00844-0 · Cited by 210
Cited by
Lampert, R., Chung, E. H., Ackerman, M. J., Arroyo, A. R., Darden, D., Deo, R., Dolan, J., Etheridge, S. P., Gray, B. R., Harmon, K. G., et al. (2024) · Heart Rhythm · doi:10.1016/j.hrthm.2024.05.018 · Cited by 167
Metadata
- Authors
- Knops, R. E., Reddy, V. Y., Ip, J. E., Doshi, R., Exner, D. V., Defaye, P., Canby, R., Bongiorni, M. G., Shoda, M., Hindricks, G., Neužil, P., Rashtian, M., Breeman, K. T. N., Nevo, J. R., Ganz, L., Hubbard, C., & Cantillon, D. J.
- Venue
- New England Journal of Medicine
- Publisher
- Massachusetts Medical Society
- Published
- 2023-06-22
- Volume
- 388
- Issue
- 25
- Pages
- 2360-2370
- ISSN
- 0028-4793 (ISSN-L)
- 1533-4406
- OpenAlex
- W4377140679
- Cited by
- 283
- References
- 20
- FWCI
- 49.78
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A Dual-Chamber Leadless Pacemaker. https://app.cochat.ai/papers/10.1056/nejmoa2300080
- The paper
- Knops, R. E., Reddy, V. Y., Ip, J. E., Doshi, R., Exner, D. V., Defaye, P., Canby, R., Bongiorni, M. G., Shoda, M., Hindricks, G., Neužil, P., Rashtian, M., Breeman, K. T. N., Nevo, J. R., Ganz, L., Hubbard, C., & Cantillon, D. J. (2023). A Dual-Chamber Leadless Pacemaker. New England Journal of Medicine. https://doi.org/10.1056/nejmoa2300080
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.