Validity and reliability of the Moxy oxygen monitor during incremental cycling exercise
The DOI matches its Crossref record; 16 of the 17 references it cites were found in a registry or an index.
Abstract
Introduction: The Moxy is a novel, cutaneously placed muscle oxygen monitor which claims to measure local oxygen saturation (SmO 2 ) and total haemoglobin (THb) using near‐infrared spectroscopy. If shown to be reliable, its data storage and telemetric capability will be useful for assessing localised O 2 usage during field‐based exercise. This study investigated the reliability of the Moxy during cycling and assessed the correlations between its measurements, whole‐body O 2 consumption (VO 2 ) and heart rate (HR). Methods: Ten highly trained cyclists performed an incremental, step‐wise cycling protocol on two occasions while wearing the Moxy. SmO 2 , THb, VO 2 and HR were recorded in the final minute of each five‐minute stage. Data were analysed using Spearman's Order‐Rank Coefficient (SROC), Intraclass Correlation (ICC), and Coefficient of Variance (COV). Significance was set at p ≤ .05. Results: SmO 2 showed a ‘strong’ or ‘very large’ correlation between trials (SROC: r = 0.842–0.993, ICC: r = 0.773–0.992, p ≤ .01) and was moderately correlated with VO 2 and HR ( r = −0.71–0.73, p ≤ .01). SmO 2 showed a moderate to high reliability at low intensities, but this decreased as relative exercise intensity increased. THb showed poor correlations between tests and with the other measured variables, but was highly reliable at all power outputs. Conclusions: The Moxy is a reliable device to measure SmO 2 at low to moderate intensities, but at higher intensities, greater variation in measurements occurs, likely due to tissue ischaemia or increased movement artefacts due to more frequent muscular contractions. THb has low variation during exercise, and does not appear to be a valid indicator of muscle oxygenation.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“at higher intensity there is a greater standard deviation of SmO2”
From The relationship of muscle oxygen saturation analyzer with other monitoring and quantification tools in a maximal incremental treadmill test (2023) · cites this paper as Crum et al., 2017
In the citing paper
“However, there is some discrepancy since it does state that it is actually valid in an incremental cycle machine test (Crum et al., 2017), and in another incremental race test it is mentioned that it depends on the subject (Austin et al., 2005).”
From The relationship of muscle oxygen saturation analyzer with other monitoring and quantification tools in a maximal incremental treadmill test (2023) · cites this paper as Crum et al., 2017
In the citing paper
“The Moxy Monitor has been compared with alternate NIRS device and evaluated for reliability and validity in previously published papers.7,34”
From Near-infrared spectroscopy-derived muscle oxygen saturation on a 0% to 100% scale: reliability and validity of the Moxy Monitor (2019) · cites this paper as [34]
In the citing paper
“Wearable near-infrared spectroscopy (NIRS) measurements of muscle oxygen saturation (SmO2) demonstrated good test–retest reliability at rest.”
From Comparing the reliability of muscle oxygen saturation with common performance and physiological markers across cycling exercise intensity (2023) · cites this paper as [5]
In the citing paper
“Our ICC scores across workloads were comparable to those by Crum et al., who also looked at the reliability of wearable NIRS during an incremental exercise test (IET) (5).”
From Comparing the reliability of muscle oxygen saturation with common performance and physiological markers across cycling exercise intensity (2023) · cites this paper as [5]
6 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Jöbsis, F. F. (1977) · Science · doi:10.1126/science.929199 · Cited by 3,899 · cited 1×
“The process is based on the differential proportion of wavelengths which are absorbed by oxygenated and deoxygenated Hb (De Blasi, Cope, Elwell, Safoue, & Ferrari, 1993; Jobsis, 1977).”
BackgroundLimiting factors for maximum oxygen uptake and determinants of endurance performance
BASSETT, D. R. (2000) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200001000-00012 · Cited by 2,410 · cited 1×
“In addition, the greater rate of mechanical work requiring the greater VO2 results in an increased concentration of intramuscular adenine diphosphate, which signals a need for more oxidative ATP production and further promotes O2 movement into the muscle (Bassett & Howley, 2000).”
BackgroundValidation of near-infrared spectroscopy in humans
Mancini, D. M., Bolinger, L., Li, H., Kendrick, K., Chance, B., & Wilson, J. R. (1994) · Journal of Applied Physiology · doi:10.1152/jappl.1994.77.6.2740 · Cited by 669 · cited 2×
“…the NIRS method has been demonstrated through its highly significant correlations with venous O2 saturation at rest and during exercise (r= 0.92) (Mancini et al., 1994), and from the reproducibility of its measurements during repeated incremental running (r = 0.87–0.88) and cycling (r =…”
BackgroundMethodInfluentialMuscle coordination of maximum-speed pedaling
Raasch, C. C., Zajac, F. E., Ma, B., & Levine, W. S. (1997) · Journal of Biomechanics · doi:10.1016/s0021-9290(96)00188-1 · Cited by 362
Hamaoka, T., McCully, K. K., Quaresima, V., Yamamoto, K., & Chance, B. (2007) · Journal of Biomedical Optics · doi:10.1117/1.2805437 · Cited by 329
Noninvasive measurement of human forearm oxygen consumption by near infrared spectroscopy
De Blasi, R. A., Cope, M., Elwell, C., Safoue, F., & Ferrari, M. (1993) · European Journal of Applied Physiology and Occupational Physiology · doi:10.1007/bf00377698 · Cited by 238
Davis, S. L., Fadel, P. J., Cui, J., Thomas, G. D., & Crandall, C. G. (2006) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00867.2005 · Cited by 179
Muscle Oxygenation Trends During Dynamic Exercise Measured by Near Infrared Spectroscopy
Bhambhani, Y. N. (2004) · Canadian Journal of Applied Physiology · doi:10.1139/h04-033 · Cited by 149 · cited 4×
“Minute ventilation (VE) and concentrations of O2 and CO2 values were used to calculate the volume of inspired air (VI) using the Haldane transformation, where VE was corrected for barometric pressure, ambient temperature and atmospheric water saturation.”
BackgroundInfluentialO2 transport and its interaction with metabolism; a systems view of aerobic capacity
HONIG, C. R., CONNETT, R. J., & GAYESKI, T. E. J. (1992) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-199201000-00009 · Cited by 91
Tew, G. A., Klonizakis, M., Moss, J., Ruddock, A. D., Saxton, J., & Hodges, G. J. (2010) · Microvascular Research · doi:10.1016/j.mvr.2010.12.001 · Cited by 74 · cited 1×
“In accordance with Tew et al. (2011), COVs of ≤10%, 10–25% and >25% were considered good, moderate and poor.”
Background
Research works printed but not linked: 2
- Cornachione K., 2014, Science and Skiing VI, 369
- McCully K. K., 2000, Near‐infrared spectroscopy: What can it tell us about oxygen saturation in skeletal muscle, Exercise and Sport Sciences Reviews, 28, 123
Cited by
Items 101–200 of 228
Oatyimprai, K. (2020) · International Journal of Geomate · doi:10.21660/2020.72.5636 · Cited by 5
The inclusion of wildcard players during small-sided games causes alterations on players’ workload
Bujalance-Moreno, P., Román, P. Á. L., Ramírez‐Campillo, R., Martínez‐Amat, A., & García‐Pinillos, F. (2020) · Isokinetics and Exercise Science · doi:10.3233/ies-202109 · Cited by 5
“The current results are consistent with those observed in the previous study [21], as SmO 2 decreased more when the exercise intensity was higher (with significant differences in bout 3), such as during the 4-a-side SSG without wildcard players, without changes throughout four bouts nor during…”
InfluentialVasquez-Bonilla, A. A., González-Custodio, A., Timón, R., & Olcina, G. J. (2022) · Sport Sciences for Health · doi:10.1007/s11332-022-00985-2 · Cited by 6
“A greater SmO2 decrease is related to power output increase [24, 34].”
BackgroundAvila, L., Bentley, R. F., Bastas, D., Brandão, L. R., Schneiderman, J. E., Ward, L. C., Wong, G., Stephens, S., Liu, K., & Thomas, S. G. (2023) · Blood Advances · doi:10.1182/bloodadvances.2022009599 · Cited by 5
“Ejection fraction (%) 36 (29-56) 45 (33-59) .”
THE EFFECTS OF THAI MASSAGE ON LEG MUSCLE OXYGENATION AND TIME TO FATIGUE IN HEALTHY MALE SUBJECTS
Sucharit, W. (2019) · International Journal of Geomate · doi:10.21660/2019.61.4527 · Cited by 5
Chien, K., Kan, N., Liao, Y., Yang, W.-T., & Yang, Y. (2020) · Scientific Reports · doi:10.1038/s41598-020-70599-6 · Cited by 5
Brochhagen, J., Barroso, M. T. C., Baumgart, C., Freiwald, J., & Hoppe, M. W. (2020) · BMC Cardiovascular Disorders · doi:10.1186/s12872-020-01661-4 · Cited by 5
Cycling Athlete Performance: Analysis of Muscle Oxygen Saturation through Moxy Measurement
Negara, J. D. K. (2023) · JURNAL PENDIDIKAN JASMANI DAN OLAHRAGA · doi:10.17509/jpjo.v8i1.56177 · Cited by 7
Overarm Training Tolerance: A Pilot Study on the Use of Muscle Oxygen Saturation as a Biomarker
Gorti, B., Stephenson, C., Sethi, M., Gross, K., Ramos, M., Seshadri, D. R., & Drummond, C. K. (2024) · Sensors · doi:10.3390/s24144710 · Cited by 4
“For the current study, the Moxy sensor was chosen for SmO2 measurements based on device validation clinical trials [12,22–24].”
BackgroundMethodUpper Extremity Exercise Capacity and Muscle Oxygenation in Patients With Primary Ciliary Dyskinesia
Mutlu, Ş., Güçlü, M. B., Eyüboğlu, T. Ş., & Aslan, A. T. (2025) · Pediatric Pulmonology · doi:10.1002/ppul.27470 · Cited by 4
“The Moxy oxygen monitor is a reliable and valid method [23, 36].”
Whyte, E., Thomas, S. G., & Marzolini, S. (2022) · PM&R · doi:10.1002/pmrj.12927 · Cited by 5
Shirai, Y., Ito, T., Ito, Y., Matsunaga, N., Noritake, K., Ochi, N., & Sugiura, H. (2023) · Healthcare · doi:10.3390/healthcare11020221 · Cited by 4
Meireles, A. A. F. D. M., de Souza, H. L. R., Arriel, R. A., Ribeiro, G. G. S., Rodrigues, A. B., de Oliveira, G. T., Ide, B. N., Chagas, M. H., & Marocolo, M. (2023) · International journal of exercise science · doi:10.70252/ityd2629 · Cited by 4
Behrendt, T., Bielitzki, R., Behrens, M., Jahns, L.-M., Boersma, M., & Schega, L. (2024) · PeerJ · doi:10.7717/peerj.18027 · Cited by 4
“…using a muscle near-infrared spectroscopy (mNIRS) monitor (MOXY; Fortiori Design LLC, Hutchinson, MN, USA), which recorded changes in total tissue hemoglobin concentration (tHb) and oxygenated hemoglobin as a percentage of tHb ( i.e ., muscle oxygenation saturation, S m O 2 ) ( Crum et al., 2017 ).”
MethodCarreño-Román, M., Ramos‐López, D., Rapaport, B., Caulier‐Cisterna, R., Espinosa‐Ramírez, M., & Briceño, F. A. C. (2024) · Journal of Visualized Experiments · doi:10.3791/67609 · Cited by 5
Vasquez-Bonilla, A. A., Sepulveda, R. A. Y., Monsalves‐Álvarez, M., Tuesta, M., Bastías, D. M. D., Cortés-Roco, G., Arancibia, J. E. O., Guzmán-Muñoz, E., & López‐Gil, J. F. (2025) · Journal of Functional Morphology and Kinesiology · doi:10.3390/jfmk10020136 · Cited by 5
“Local SmO 2 assessment was carried out using a NIRS sensor (Moxy, Fortiori Design LLC, Minneapolis, MN, USA) with a sampling rate of 1 Hz, which is reliability for measuring SmO 2 (ICC = 0.773–0.992) [6].”
InfluentialSchober, P. R., & Schwarte, L. A. (2019) · Air Medical Journal · doi:10.1016/j.amj.2019.10.002 · Cited by 4
NEAR INFRARED SPECTROSCOPY AND SPIROERGOMETRY TESTING IN CROSSFIT
Schlegel, P., Hiblbauer, J., & Agricola, A. (2020) · Studia sportiva · doi:10.5817/sts2020-1-1 · Cited by 4
“The validity and reliability of the device has been confirmed on a sample of high performance cyclists (Crum, 2017).”
BackgroundMyoglobinemia, Peripheral Arterial Disease, and Patient Mortality
Scrivner, O., Fletcher, E., Hoffmann, C., Li, F., Wilkinson, T. C. I., Miserlis, D., SMITH, R. S., Bohannon, W. T., Sutliff, R. L., Jordan, W. D., et al. (2023) · Journal of the American College of Surgeons · doi:10.1097/xcs.0000000000000554 · Cited by 4
“Hemoglobin oxygen saturation was measured using 2 Moxy Monitors (Fortiori Design LLC, Hutchinson, MN), as published.(28) The monitors were placed on the belly of the medial gastrocnemius and secured with cover-roll tape strips.”
MethodLacto-resistance training: a method to facilitate muscle hypertrophy in professional bodybuilders
Hatami, M., Nikooie, R., & Enhesari, A. (2023) · Sport Sciences for Health · doi:10.1007/s11332-023-01106-3 · Cited by 4
“SmO 2 , also known as muscle oxygenation, represents the balance between oxygen delivery and oxygen consumption in the working muscles [30, 31].”
Repeated Short-Term Bouts of Hyperoxia Improve Aerobic Performance in Acute Hypoxia
Faulhaber, M., Schneider, S., Rausch, L. K., Dünnwald, T., Menz, V., Gatterer, H., Kennedy, M. D., & Schobersberger, W. (2023) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000004502 · Cited by 5
“(SmO2) and total hemoglobin content (THb) during the TTs in hypoxia (TT2 and TT3) (9).”
BackgroundInfluentialBriceño, F. A. C., Espinosa‐Ramírez, M., Rivera-Greene, A., Guerra-Venegas, C., Lungenstrass-Poulsen, A., Villagra-Reyes, V., Caulier‐Cisterna, R., Araneda, O. F., & Viscor, G. (2023) · Biosensors · doi:10.3390/bios13110985 · Cited by 4
“…2 ‐ m.intercostales ) 2 m.intercostales m.vastus lateralis m.vastus lateralis protocols [22,23,25,26,34 and m.vastus lateralis (SmO 2 ‐ m.vastus –36,46]. lateralis ), according to previ ‐ ous protocols [22,23,25,26,34–36, Briefly, to record SmO 2 - m. Briefly, to record SmO 2 ‐ m.intercostales, a…”
Intra-Reliability of a Wearable Near-Infrared Sensor for Monitoring the Intensity of Exercise
Michalik, K., Smolarek, M., Nowak, M., Pueo, B., & Żmijewski, P. (2024) · Applied Sciences · doi:10.3390/app14135856 · Cited by 3
Vasquez-Bonilla, A. A., Rojas‐Valverde, D., Feliu-Ilvonen, J. M., Timón, R., & Olcina, G. J. (2024) · Sport Sciences for Health · doi:10.1007/s11332-024-01218-4 · Cited by 3
Sendra‐Pérez, C., Martínez, A. E., & Priego‐Quesada, J. I. (2025) · Sensors · doi:10.3390/s25216798 · Cited by 5
Muscle oxygen saturation and cutting movements in Women's Futsal
Vasquez-Bonilla, A. A., Fuentes, I. C., Ugalde-Ramírez, A., Trejos-Montoya, J. A., Rojas‐Valverde, D., Clemente‐Suárez, V. J., & Sepulveda, R. A. Y. (2025) · Apunts Sports Medicine · doi:10.1016/j.apunsm.2025.100505 · Cited by 3
Brochhagen, J., Barroso, M. T. C., Baumgart, C., Wasmus, D. T., Freiwald, J., & Hoppe, M. (2023) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2023.1218948 · Cited by 3
“Further details on the measurement principle can be found elsewhere (49).”
BackgroundMethodInfluentialAidiel, L., Lim, D. Z. Y., Chow, K. M., Ihsan, M., Chia, M., & Choo, H. C. (2024) · Temperature · doi:10.1080/23328940.2024.2302772 · Cited by 3
“The Moxy monitor has four light emitting diodes (630 to 850 nm) and 2 detectors which record the amount of back-scattered light at 12.5 and 25 mm from the light source [32].”
BackgroundInfluentialKol, B. K., Güçlü, M. B., Baytok, E., & Demirci, N. Y. (2024) · Physiotherapy Theory and Practice · doi:10.1080/09593985.2024.2327534 · Cited by 3
“The muscle oxygen saturation (SmO 2 ) was evaluated using the Moxy ® device (Moxy ® , Fortiori Design LLC, Minnesota, USA), which is a valid and reliable noninvasive method (Crum et al., 2017).”
Pelvan, O., Akbaş, S., & Çotuk, H. B. (2024) · International Journal of Sport Exercise & Training Sciences · doi:10.18826/useeabd.1481068 · Cited by 4
“Crum et al. (2017) reported a strong negative correlation (r=-0.730) between SmO 2 depletion and oxygen consumption during a gradually increasing cycle ergometer test.”
BackgroundInfluentialSousa, A., Chambion-Diaz, M., Pialoux, V., Carin, R., Viana, J. L., Milheiro, J., Reis, V. M., & Millet, G. P. (2025) · Biology of Sport · doi:10.5114/biolsport.2025.139851 · Cited by 4
“…and in total haemoglobin (THb) were assessed in the capillaries (vastus lateralis muscle) through a near-infrared spectroscopy (NIRS) mon - itor (Moxy monitor, Fortiori Design, Minnesota, USA), positioned as previously suggested [18], and demonstrated to be valid and reliable to measure SmO 2 [19].”
Bentley, R. F., Bernal, J. B., Basile, D. C., Salvo, A. N. D., & Schwartz, J. L. (2024) · Experimental Physiology · doi:10.1113/ep092256 · Cited by 3
“This device has a coefficient of variation during 100 W of cycling of 4% (Crum et al., 2017) and a penetration depth of 12.5 mm based upon light detector and emitter distances.”
InfluentialEvaluating Arterial Blood Flow Limitation Using Muscle Oxygenation and Cycling Power
Arnold, J., Yogev, A., & Koehle, M. S. (2021) · Clinical Journal of Sport Medicine · doi:10.1097/jsm.0000000000000942 · Cited by 3
“SmO2 was analyzed as the NIRS-derived muscle oxygenation signal for the present experiment.(11,12,22) The 2 NIRS sensors were positioned on the left and right vastus lateralis (VL) muscle belly, at 1⁄3 the distance from the superolateral patellar pole to the greater trochanter.”
MethodHudgins, J. H., Schoeberlein, M. I., & Wilkins, B. W. (2025) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000003875 · Cited by 3
Physiological Response to Non-Traditional High-Intensity Interval Training
Schlegel, P., Hiblbauer, J., & Agricola, A. (2020) · Acta Facultatis Educationis Physicae Universitatis Comenianae · doi:10.2478/afepuc-2020-0001 · Cited by 3
Haddad, T., Spence, A. L., Peiffer, J. J., Blain, G. M., Brisswalter, J., & Abbiss, C. R. (2024) · Journal of Sports Science and Medicine · doi:10.52082/jssm.2024.46 · Cited by 2
Marocolo, M., Arriel, R. A., Ribeiro, G. G. S., Meireles, A. A. F. D. M., Krzysztofik, M., Chycki, J., Zając, A., & de Souza, H. L. R. (2024) · Journal of Human Kinetics · doi:10.5114/jhk/194066 · Cited by 4
“The NIRS device operates by sequentially sending light waves (630–850 nm) from four light-emitting diodes into the tissue beneath it and recording the amount of scattered light returned to two detectors positioned 12.5 mm and 25 mm from the light source (Crum et al., 2017).”
InfluentialGüneş, M., Güçlü, M. B., Güvenir, A., & Tokçaer, A. B. (2025) · Physiotherapy Theory and Practice · doi:10.1080/09593985.2025.2602854 · Cited by 4
“The SmO 2 and THb values were measured at rest, during the test, and recovery (Crum et al., 2017).”
Lahade, S., Ross, N., & O’Sullivan, T. D. (2026) · Journal of Physics Photonics · doi:10.1088/2515-7647/ae6ae4 · Cited by 2
Lillquist, T. J., Mahoney, S. J., Kotarsky, C. J., McGrath, R. P., Jarajapu, Y. P. R., Scholten, S. D., & Hackney, K. J. (2023) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000004492 · Cited by 2
“With 2 pairs of testing sessions for both PEIC and control, there were 4 conditions of treatment, including (1) PEIC-Direct, (2) PEICRemote, (3) SHAM-Direct, and (4) SHAM-Remote, and 2 conditions of time (pre-exercise and 24 hours postexercise).”
MethodKol, B. K. (2024) · Eurasian Journal of Pulmonology · doi:10.14744/ejp.2024.1102 · Cited by 2
“[26] Before the 6-MWT, the Moxy ® monitor was placed on the 1/3 lower motor point of the QF muscle with the help of an anti-allergic plaster.”
Characterization of muscle oxygenation response in well-trained handcyclists
Puglia, V. F., Paquette, M., & Bergdahl, A. (2024) · European Journal of Applied Physiology · doi:10.1007/s00421-024-05524-0 · Cited by 2
“…to other traditional NIRS devices, and was previously validated in sports science, research has identified that the reliability of the SmO 2 measurements decreases as the intensity of exercise increases due to movement artifact and tissue ischemia (Feldmann et al. 2019; Crum et al. 2017).”
Lim, Z. X., Peos, J., Ostojic, S., Arthur, P. G., & Fournier, P. A. (2025) · Antioxidants · doi:10.3390/antiox14060667 · Cited by 2
Vasquez-Bonilla, A. A., Trejos-Montoya, J. A., Rojas‐Valverde, D., Ugalde-Ramírez, A., Timón, R., Ureña, B. S., & Olcina, G. J. (2025) · Sport Sciences for Health · doi:10.1007/s11332-025-01479-7 · Cited by 2
“Correlations between SmO ₂ , ▽%SmO ₂ , and key physiological markers ( V O ₂ max, HR, RQ, VE, and RPE) were consistent with previous findings [30, 48, 49], although the Fig.”
InfluentialLower Limb Compression Socks with Biomechanical Concepts
Sergio, D. (2020) · Biomedical Journal of Scientific & Technical Research · doi:10.26717/bjstr.2020.32.005301 · Cited by 2
Romero‐Arenas, S., Quero-Calero, C. D., Abellán‐Aynés, O., Andreu‐Caravaca, L., Fernandez-Calero, M., Manonelles, P., & López-Plaza, D. (2023) · Sports · doi:10.3390/sports11110212 · Cited by 2
“Furthermore, the validity and reliability of NIRS devices have been corroborated not only in activities such as running but also in cycling [12,15,26].”
BackgroundA polyphenol-rich cranberry supplement improves muscle oxidative capacity in healthy adults
Parenteau, F., Denis, A., Roberts, M., Comtois, A. S., & Bergdahl, A. (2024) · Applied Physiology, Nutrition, and Metabolism · doi:10.1139/apnm-2023-0633 · Cited by 3
Forot, J., Schorderet, F., Burdet, C., Brocherie, F., & Millet, G. P. (2026) · Experimental Physiology · doi:10.1113/ep093311 · Cited by 1
“However, despite its potential, the integration of NIRS into training routines remains uncertain, requiring further research to validate its effectiveness and reliability in field settings (Crum et al., 2017; Feldmann et al., 2019; Klusiewicz et al., 2021; McManus et al., 2018).”
Effects of chronic obstructive pulmonary disease stage on muscle oxygenation during exercise
Gökmen, G. Y., & Demir, C. (2023) · Anatolian Current Medical Journal · doi:10.38053/acmj.1366947 · Cited by 2
Bradley, A., Roehl, A. S., Smith, J., McGrath, R. P., & Hackney, K. J. (2023) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2023.1210880 · Cited by 1
“Greater dynamic muscle activation and aerobic energy metabolism can subsequently decrease in SmO 2 , as local oxygen delivery and uptake is below the capabilities of aerobic metabolism in the muscle (34).”
BackgroundChoi, W., Hwang, G., & Lee, S. (2024) · Physical therapy rehabilitation science · doi:10.14474/ptrs.2024.13.2.187 · Cited by 1
Succi, P. J., Benitez, B., Kwak, M., Mitchinson, C. J., Butterfield, T. A., Samaan, M. A., Clasey, J. L., & Bergstrom, H. C. (2025) · European Journal of Applied Physiology · doi:10.1007/s00421-024-05698-7 · Cited by 2
Suicmez, S., Yazıcı, G., Altıparmak, T., Cagatay, C., Çağlayan, H. Z. B., & Nazlıel, B. (2025) · Topics in Stroke Rehabilitation · doi:10.1080/10749357.2025.2538520 · Cited by 1
“SmO 2 showed, contrary to expectations, 37 an increase rather than a decrease during 6MWT, and a decrease in SmO 2 was observed during the recovery phase.”
Benincasa, M. T., Coiro, F., Coppola, S., Serra, E., Celentano, E., Costa, C., Albano, D., & Vastola, R. (2025) · Journal of Functional Morphology and Kinesiology · doi:10.3390/jfmk10030317 · Cited by 1
“Thanks to advancements in non-invasive technologies, such as near-infrared spectroscopy (NIRS), it is now possible to monitor muscle oxygen saturation (SmO 2 ), expressed as a percentage, in real time during various types of exercise [54–58].”
Lesmana, H. S., Schroeder, B., Marume, K., Rodrigues, P., & Lawley, J. S. (2025) · Experimental Physiology · doi:10.1113/ep092908 · Cited by 1
“This NIRS sensor sent light waves (630–850 nm) from four light-emitting diodes into the tissue beneath it and recorded the amount of returned scattered light at two detectors positioned 12.5 and 25.0 mm from the light source (Crum et al., 2017).”
InfluentialChan, W. W. Y., Sun, E. R., Fu, S.-N., Zheng, Y., Parent, É., Cheung, J. P. Y., & Wong, A. Y. L. (2025) · European Spine Journal · doi:10.1007/s00586-025-09313-x · Cited by 2
“Bilateral paraspinal muscle oxygenation (SmO 2 ) was measured using a validated NIRS device (Moxy muscle sensor, Fortiori Design LLC, Minneapolis, Minnesota, USA) [13].”
Wu, X. V., Tam, W., Yap, X. Y., Zhang, J., Mok, W. Y. J., Lim, Z. X., & Goh, J. (2025) · BMC Geriatrics · doi:10.1186/s12877-025-06796-7 · Cited by 1
“The device measured muscle O 2 saturation (SmO 2 ) every 2 seconds throughout the trial and recovery and has been validated and described in detail in a study by Crum et al. [37].”
Güneş, M., Güçlü, M. B., Güvenir, A., & Tokçaer, A. B. (2025) · Scientific Reports · doi:10.1038/s41598-025-33413-9 · Cited by 1
“Muscle oxygen saturation (SmO 2 ), minimum (SmO 2min ) and maximum muscle oxygen saturation (SmO 2max ), total hemoglobin (THb), minimum (THb min ) and maximum (THb max ) total hemoglobin values at rest, during the test, and during recovery were measured and recorded 31 .”
Brown, J. B., Lauver, J. D., Rotarius, T. R., & Guilkey, J. P. (2026) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2025.1737095 · Cited by 1
Yoleri, B., Güçlü, M. B., Eyüboğlu, T. Ş., & Aslan, A. T. (2026) · European Journal of Applied Physiology · doi:10.1007/s00421-026-06130-y · Cited by 1
“Muscle oxygenation was measured using the “Moxy ® ” monitor (Moxy, Fortiori Design LLC, Minnesota, USA) (Crum et al. 2017).”
Kol, B. K., Güçlü, M. B., Baytok, E., & Demirci, N. Y. (2026) · BMC Pulmonary Medicine · doi:10.1186/s12890-026-04249-4 · Cited by 1
Muscle Oxygen Saturation as a Marker of Internal Load in Elite Soccer: A Case Series
Herzog, S., Philippi, M., Feldmann, A., Asian, J., Fílter, A., Requena, B., Batterson, P. M., & Bogdanis, G. C. (2026) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000005483 · Cited by 1
Güneş, M., & Avcı, R. H. (2026) · Respiratory Medicine · doi:10.1016/j.rmed.2026.109024 · Cited by 1
Pagán-Lassalle, P., Palamar, A. J., DeBlois, J. P., Lefferts, W. K., & Heffernan, K. S. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04170-1 · Cited by 1
Buxton, J., Prins, P., Ryan, E. J., Jones, D. W., Thrasher, I., Faulkner, M. N., Robertson, E., Welton, G. L., & Ault, D. L. (2022) · Research Directs in Strength and Performance · doi:10.53520/rdsp2022.10529 · Cited by 1
Meireles, A. A. F. D. M., de Oliveira, G. T., de Souza, H. L. R., Arriel, R. A., Leitão, L., dos Santos, M. H. P., & Marocolo, M. (2022) · Motriz Revista de Educação Física · doi:10.1590/s1980-657420220004122 · Cited by 1
Rojas‐Valverde, D., Oliva‐Lozano, J. M., Gutiérrez-Vargas, R., Pino‐Ortega, J., Muyor, J. M., & Gómez‐Carmona, C. D. (2023) · International Journal of Performance Analysis in Sport · doi:10.1080/24748668.2023.2185744 · Cited by 1
“To investigate the interaction of internal and external loads variables, mechanisms that allow real-time and minimally invasive monitoring have been recently developed (Crum et al., 2017; Rojas-Valverde et al., 2019; Tan et al., 2008).”
BackgroundThe EMG, NIRS, and RPE responses to two turn transition techniques in alpine skiing
Seifert, J. G., Kipp, R. W., Griffin, S., & Graham, D. (2023) · JSAMS Plus · doi:10.1016/j.jsampl.2023.100037 · Cited by 1
Reply to Drs. Henni and Abraham: Muscle oxygen content at exercise in patients with claudication
Luck, J. C., Aziz, F., Sinoway, L. I., & Muller, M. D. (2017) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00706.2017
“We used NIRS because it is able to measure at the depth of skeletal muscle and has been shown to track changes in muscle oxygen saturation reliably across different exercise intensities (1).”
MethodLozano, M. F., & Gómez-López, P. J. (2021) · Journal of Physical Education and Human Movement · doi:10.24310/jpehmjpehmjpehm.v3i213035
Soelkner, S., Jennewein, S., Hovorka, M., Prinz, B., & Nimmerichter, A. (2023) · Advances in intelligent systems and computing · doi:10.1007/978-3-031-31772-9_18
ŘEHOŘ, A., Novotný, J., Bernacikovà, M., & Paludo, A. C. (2023) · The Journal of Sports Medicine and Physical Fitness · doi:10.23736/s0022-4707.23.15438-7 · Cited by 1
Sepúlveda, R. Y. (2023) · Buleria (Universidad de León) · doi:10.18002/10612/18621
Non-invasive optical methods for perioperative flap monitoring in plastic and reconstructive surgery
Lindelauf, A. A. M. A. (2024) · Research Publications (Maastricht University) · doi:10.26481/dis.20240521al
Choi, W., Hwang, G., & Lee, S. (2024) · Physical therapy rehabilitation science · doi:10.14474/ptrs.2024.13.2.45
Muscular Oxygen Saturation and Hemoglobin Concentration during Freediving: A Case Study
Vrdoljak, D., Dujić, Ž., & Foretić, N. (2024) · Oxygen · doi:10.3390/oxygen4030016
“In particular, the decrease in SmO 2 corresponded with an increase in tHb. Additionally, tHb is derived as the sum of oxygenated (O 2 Hb) and deoxygenated hemoglobin (HHb) in the measured region (tHb = O 2 Hb + HHb), and it indicates a change in the blood flow to the muscle [27].”
BackgroundEffects of Radiochemotherapy on Muscle Oxygen Saturation in Locally Advanced Lung Cancer Patients
Şahin, Y., Zortul, A., & Öztürk, D. (2025) · Sylwan · doi:10.59879/1bykq
Interplay Among Muscle Oxygen Saturation, Activation, and Power on a Swim-Bench
Coloretti, V., Quagliarotti, C., Gatta, G., Piacentini, M. F., Cortesi, M., & Fantozzi, S. (2025) · Sensors · doi:10.3390/s25134148 · Cited by 1
“Although MOXY has shown high reliability and validity [56] and moderate inverse correlation with metabolic oxygen consumption VO 2 [57,58], when compared to other CW-NIRS devices (i.e., Portamon, Hu-mon Hex and NIMO), it has shown both favorable [52] and unfavorable [23,51] outcomes.”
Association between muscle oxygenation dynamics and cardiac workload and ventilation during exercise
Nemoto, S., Nakabo, T., Yoshikawa, A., Tashiro, N., Kuroyama, Y., Nakamura, D., & Geshi, E. (2025) · The Journal of Sports Medicine and Physical Fitness · doi:10.23736/s0022-4707.25.17189-2 · Cited by 1
Real-Time Nitric Oxide Bioactivity Response to a Bout of High- vs Low-Load Resistance Exercise
Gonzalez, A. M., Krawczyk, M. P., Mangine, G. T., Pinzone, A. G., & Harhash, T. S. (2025) · Journal of Fitness Wellness and Human Performance · doi:10.70834/jfwhp-25-0001
Babity, M., Zámodics, M., Kovács, É., Bucskó-Varga, Á., Kulcsár, P., Boroncsok, D., Benkő, R., Fábián, A., Horváth, M., Balla, D., et al. (2025) · Journal of Functional Morphology and Kinesiology · doi:10.3390/jfmk10040464
“Muscle oxygen saturation measurements were performed using a commercially available, validated NIRS device (Moxy, Fortiori Design LLC, Hutchinson, MN, USA) during cardiopulmonary exercise testing [14,22].”
del-Blanco-Muñiz, J. Á., Ladriñán-Maestro, A., Vergara-Gatica, G., Orellana-García, C., García-Pérez-de-Sevilla, G., Martín-Vera, D., & Sánchez-Sierra, A. (2025) · Healthcare · doi:10.3390/healthcare13243211
“[17].”
Measuring muscle oxygenation in endurance sports with wearable nirs - from validation to application
Yogev, A. (2023) · cIRcle (University of British Columbia) · doi:10.14288/1.0431178 · Cited by 1
Apnea. Una alternativa a la hipoxia para mejorar el rendimiento en ciclismo
Salas-Montoro, J.-A. (2023) · Institutional Repository of the University of Granada (University of Granada) · doi:10.13140/rg.2.2.31471.00163
Walsh, J. (2025) · Scholarship@Western (Western University)
Study of the local oxygen saturation in cycling and differences between sexes
Sendra‐Pérez, C. (2025) · Repository of Digital Objects for Teaching Research and Culture (University of Valencia)
Tapio, J. (2024) · Theseus (Ammattikorkeakoulujen)
Muhammad Fathur Rachman (2024) · Repository at Universitas Pendidikan Indonesia (Universitas Pendidikan Indonesia) · Cited by 1
Whyte, E. (2021) · TSpace (University of Toronto)
Lalande-Gauthier, M. (2018) · Archipel (Université du Québec à Montréal)
Archambault, A. (2022) · Archipel (Université du Québec à Montréal)
Oleksy, Ł., Mika, A., Daszkiewicz, M., Sopa, M., Szczudło, M., Kuchciak, M., Stolarczyk, A., Adamska, O., Reichert, P., Dzięcioł-Anikiej, Z., et al. (2026) · Journal of Clinical Medicine · doi:10.3390/jcm15020794
Ammar, A., Salem, A., Kerkeni, M., Mera, T. A., Baťalík, L., Toral-Noristz, M., Jahrami, H., Tounsi, S., Żmijewski, P., Ghazzawi, H., et al. (2026) · International Journal of Food Sciences and Nutrition · doi:10.1080/09637486.2026.2625827
“SME values were derived from previously published reliability studies (SmO 2 : 1.89%, threshold ±3.79% (Crum et al. 2017); peak HR: 1.18 bpm, threshold ±2.01 bpm (Coswig et al. 2020); RMSSD: 0.65 ms, threshold ±1.3 ms (Shaffer and Ginsberg 2017)).”
Influential
Metadata
- Authors
- Crum, E. M., O'Connor, W. J., Van Loo, L., Valckx, M., & Stannard, S. R.
- Affiliations
- Massey University (NZ)
- The Hague University of Applied Sciences (NL)
- Corresponding author
- Crum, E. M.
- Venue
- European Journal of Sport Science
- Publisher
- Wiley
- Published
- 2017-05-30
- Volume
- 17
- Issue
- 8
- Pages
- 1037-1043
- ISSN
- 1536-7290 (ISSN-L)
- 1746-1391
- OpenAlex
- W2618079828
- Semantic Scholar
- 72b53146cd8a530b27c47cad8a68824947aca5a4
- Cited by
- 206
- References
- 17
- Citations by year
- 2026: 30 · 2025: 25 · 2024: 32 · 2023: 31 · 2022: 27 · 2021: 19 · 2020: 14 · 2019: 20 · 2018: 3 · 2017: 5
- FWCI
- 17.03
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- European Journal of Sport Science (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Massey University
- Topics
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Sports Performance and Training
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- muscle oxygen saturation, total hemoglobin, near-infrared spectroscopy, incremental cycling exercise, measurement reliability, whole-body oxygen consumption, heart rate, exercise intensity, muscle oxygenation
- MeSH terms
- Adolescent, Adult, Bicycling, Exercise Test, Female, Heart Rate, Hemoglobins, Humans, Male, Oximetry, Oxygen Consumption, Reproducibility of Results, Muscle, Skeletal, Spectroscopy, Near-Infrared, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.