The impact of respiratory motion on tumor quantification and delineation in static PET/CT imaging
Open accessThe DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index.
Abstract
Our aim is to investigate the impact of respiratory motion on tumor quantification and delineation in static PET/CT imaging using a population of patient respiratory traces. A total of 1295 respiratory traces acquired during whole body PET/CT imaging were classified into three types according to the qualitative shape of their signal histograms. Each trace was scaled to three diaphragm motion amplitudes (6 mm, 11 mm and 16 mm) to drive a whole body PET/CT computer simulation that was validated with a physical phantom experiment. Three lung lesions and one liver lesion were simulated with diameters of 1 cm and 2 cm. PET data were reconstructed using the OS-EM algorithm with attenuation correction using CT images at the end-expiration phase and respiratory-averaged CT. The errors of the lesion maximum standardized uptake values (SUV(max)) and lesion volumes between motion-free and motion-blurred PET/CT images were measured and analyzed. For respiration with 11 mm diaphragm motion and larger quiescent period fraction, respiratory motion can cause a mean lesion SUV(max) underestimation of 28% and a mean lesion volume overestimation of 130% in PET/CT images with 1 cm lesions. The errors of lesion SUV(max) and volume are larger for patient traces with larger motion amplitudes. Smaller lesions are more sensitive to respiratory motion than larger lesions for the same motion amplitude. Patient respiratory traces with relatively larger quiescent period fraction yield results less subject to respiratory motion than traces with long-term amplitude variability. Mismatched attenuation correction due to respiratory motion can cause SUV(max) overestimation for lesions in the lower lung region close to the liver dome. Using respiratory-averaged CT for attenuation correction yields smaller mismatch errors than those using end-expiration CT. Respiratory motion can have a significant impact on static oncological PET/CT imaging where SUV and/or volume measurements are important. The impact is highly dependent upon motion amplitude, lesion location and size, attenuation map and respiratory pattern. To overcome the motion effect, motion compensation techniques may be necessary in clinical practice to improve the tumor quantification for determining the response to therapy or for radiation treatment planning.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Accelerated image reconstruction using ordered subsets of projection data
Hudson, H. M., & Larkin, R. S. (1994) · IEEE Transactions on Medical Imaging · doi:10.1109/42.363108 · Cited by 3,711 · cited 1×
“The PET sinogram data were reconstructed into 128 × 128 × 47 image volumes with a voxel size of 5.47 × 5.47 × 3.27 mm using the ordered-subset expectation-maximization (OS-EM) reconstruction algorithm with 4 iterations and 20 subsets (Hudson and Larkin 1994).”
MethodInfluentialSeppenwoolde, Y., Shirato, H., Kitamura, K., Shimizu, S., van Herk, M., Lebesque, J. V., & Miyasaka, K. (2002) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(02)02803-1 · Cited by 1,379 · cited 2×
“These findings are also consistent with other published studies (Segars et al 2007, Seppenwoolde et al 2002).”
BackgroundResult4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT
Pan, T., Lee, T.-Y., Rietzel, E., & Chen, G. T. Y. (2004) · Medical Physics · doi:10.1118/1.1639993 · Cited by 624 · cited 1×
“These methods include respiratory-gated PET/CT (Nehmeh et al 2004a, 2004b, Pan et al 2004, 2007, Rietzel et al 2005, Wolthaus et al 2005, Abdelnour et al 2007, Dawood et al 2007, Guckenberger et al 2007, Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008,…”
Method(2005) · PubMed · Cited by 612
Distance-driven projection and backprojection in three dimensions
Man, B. D., & Basu, S. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/11/024 · Cited by 459 · cited 1×
“Each 3D NCAT phantom had a matrix size of 512 × 512 × 282 and was forward projected with the matched attenuation map to generate 200 sinograms, which had a matrix size of 293 × 280 × 141, using a distance-driven forward projector (De Man and Basu 2004).”
MethodEffect of respiratory gating on quantifying PET images of lung cancer.
Nehmeh, S. A., Erdi, Y. E., Ling, C. C., Rosenzweig, K. E., Schöder, H. M., Larson, S. M., Macapinlac, H. A., Squire, O. D., & Humm, J. L. (2002) · PubMed · Cited by 406
Noise properties of the EM algorithm. I. Theory
Barrett, H. H., Wilson, D. W., & Tsui, B. M. W. (1994) · Physics in Medicine and Biology · doi:10.1088/0031-9155/39/5/004 · Cited by 390 · cited 1×
“As shown by Barrett et al (1994), the noiseless ML-EM reconstruction results in the same image as the average of reconstructed images from multiple noise realizations of the same object.”
BackgroundFour‐dimensional (4D) PET/CT imaging of the thorax
Nehmeh, S. A., Erdi, Y. E., Pan, T., Pevsner, A., Rosenzweig, K. E., Yorke, E., Mageras, G. S., Schoder, H., Vernon, P., Squire, O., et al. (2004) · Medical Physics · doi:10.1118/1.1809778 · Cited by 349 · cited 1×
“These methods include respiratory-gated PET/CT (Nehmeh et al 2004a, 2004b, Pan et al 2004, 2007, Rietzel et al 2005, Wolthaus et al 2005, Abdelnour et al 2007, Dawood et al 2007, Guckenberger et al 2007, Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008,…”
MethodCorrelation of lung tumor motion with external surrogate indicators of respiration
Hoisak, J. D. P., Sixel, K. E., Tirona, R., Cheung, P. C. F., & Pignol, J.-P. (2004) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2004.07.681 · Cited by 318 · cited 1×
“As several studies demonstrated, there is a good correlation between the internal tissue motion and external marker motion in general, with different amplitudes (Beddar et al 2007, Gierga et al 2005, Hoisak et al 2004, Ionascu et al 2007).”
BackgroundRespiratory Motion in Positron Emission Tomography/Computed Tomography: A Review
Nehmeh, S. A., & Erdi, Y. E. (2008) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2008.01.002 · Cited by 317 · cited 1×
“…generally affected by patient respiratory motion, which can lead to underestimation of SUVmax of a region of interest, overestimation of tumor volume and mismatched PET and CT images that yield attenuation correction (AC) errors, registration errors and tumor mislocalization (Nehmeh and Erdi 2008).”
BackgroundBeyer, T., Antoch, G., Blodgett, T., Freudenberg, L. F., Akhurst, T., & Mueller, S. (2003) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-002-1097-6 · Cited by 298 · cited 1×
“For reconstructing the motion-blurred data with attenuation correction, an attenuation map at the end-expiration phase was chosen, as using attenuation maps other than the end-expiration phase results in greater SUVmax underestimation and motion artifacts (Beyer et al 2003, Erdi et al 2004).”
MethodEffect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer
Nehmeh, S. A., Erdi, Y. E., Ling, C. C., Rosenzweig, K. E., Squire, O. D., Braban, L. E., Ford, E., Sidhu, K., Mageras, G. S., Larson, S. M., et al. (2002) · Medical Physics · doi:10.1118/1.1448824 · Cited by 297 · cited 1×
“…that various respiratory motion correction methods can cause the tumor SUVmax to increase as much as 159% and tumor volume to decrease as much as 43% compared to those of the conventional measurement in patient studies (Kawano et al 2008, Meirelles et al 2007, Nehmeh et al 2002a, 2002b, 2007).”
ResultThe CT motion quantitation of lung lesions and its impact on PET-measured SUVs.
Erdi, Y. E., Nehmeh, S. A., Pan, T., Pevsner, A., Rosenzweig, K. E., Mageras, G. S., Yorke, E. D., Schöder, H. M., Hsiao, W., Squire, O. D., et al. (2004) · PubMed · Cited by 282
Biehl, K. J., Kong, F., Dehdashti, F., Jin, J., Mutic, S., Naqa, I. E., Siegel, B. A., & Bradley, J. D. (2006) · PubMed · Cited by 260
Lamare, F., Carbayo, M. J. L., Cresson, T., Kontaxakis, G., Santos, A., Rest, C. C. L., Reader, A. J., & Visvikis, D. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/17/006 · Cited by 259 · cited 1×
“…Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008, Meirelles et al 2007, Nehmeh et al 2007), postprocessing methods (Dawood et al 2006, Yamazaki et al 2006) and motion-corrected PET reconstruction (Lamare et al 2007, Li et al 2006, Qiao et al 2006, 2007).”
MethodAttenuation correction of PET images with respiration-averaged CT images in PET/CT.
Pan, T., Mawlawi, O. R., Nehmeh, S. A., Erdi, Y. E., Luo, D., Liu, H. H., Castillo, R., Mohan, R., Liao, Z. X., & Macapinlac, H. A. (2005) · PubMed · Cited by 251
A motion-incorporated reconstruction method for gated PET studies
Qiao, F., Pan, T., Clark, J. W., & Mawlawi, O. R. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/15/012 · Cited by 239 · cited 1×
“…Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008, Meirelles et al 2007, Nehmeh et al 2007), postprocessing methods (Dawood et al 2006, Yamazaki et al 2006) and motion-corrected PET reconstruction (Lamare et al 2007, Li et al 2006, Qiao et al 2006, 2007).”
MethodQuantitation of respiratory motion during 4D‐PET/CT acquisition
Nehmeh, S. A., Erdi, Y. E., Pan, T., Yorke, E., Mageras, G. S., Rosenzweig, K. E., Schoder, H., Mostafavi, H., Squire, O., Pevsner, A., et al. (2004) · Medical Physics · doi:10.1118/1.1739671 · Cited by 232 · cited 1×
“These methods include respiratory-gated PET/CT (Nehmeh et al 2004a, 2004b, Pan et al 2004, 2007, Rietzel et al 2005, Wolthaus et al 2005, Abdelnour et al 2007, Dawood et al 2007, Guckenberger et al 2007, Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008,…”
MethodDawood, M., Büther, F., Lang, N., Schober, O., & Schäfers, K. P. (2007) · Medical Physics · doi:10.1118/1.2748104 · Cited by 211 · cited 1×
“…include respiratory-gated PET/CT (Nehmeh et al 2004a, 2004b, Pan et al 2004, 2007, Rietzel et al 2005, Wolthaus et al 2005, Abdelnour et al 2007, Dawood et al 2007, Guckenberger et al 2007, Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008, Meirelles…”
BackgroundModel‐based image reconstruction for four‐dimensional PET
Li, T., Thorndyke, B., Schreibmann, E., Yang, Y., & Xing, L. (2006) · Medical Physics · doi:10.1118/1.2192581 · Cited by 210 · cited 1×
“…Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008, Meirelles et al 2007, Nehmeh et al 2007), postprocessing methods (Dawood et al 2006, Yamazaki et al 2006) and motion-corrected PET reconstruction (Lamare et al 2007, Li et al 2006, Qiao et al 2006, 2007).”
MethodInternal‐external correlation investigations of respiratory induced motion of lung tumors
Ionascu, D., Jiang, S. B., Nishioka, S., Shirato, H., & Berbeco, R. I. (2007) · Medical Physics · doi:10.1118/1.2779941 · Cited by 209 · cited 1×
“As several studies demonstrated, there is a good correlation between the internal tissue motion and external marker motion in general, with different amplitudes (Beddar et al 2007, Gierga et al 2005, Hoisak et al 2004, Ionascu et al 2007).”
BackgroundLung motion correction on respiratory gated 3-D PET/CT images
Dawood, M., Lang, N., Jiang, X., & Schafers, K. P. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.870892 · Cited by 184 · cited 1×
“…Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008, Meirelles et al 2007, Nehmeh et al 2007), postprocessing methods (Dawood et al 2006, Yamazaki et al 2006) and motion-corrected PET reconstruction (Lamare et al 2007, Li et al 2006, Qiao et al 2006, 2007).”
MethodDeep-inspiration breath-hold PET/CT of the thorax.
Nehmeh, S. A., Erdi, Y. E., Meirelles, G. D. S. P., Squire, O. D., Larson, S. M., Humm, J. L., & Schöder, H. M. (2007) · PubMed · Cited by 152
Phase and amplitude binning for 4D-CT imaging
Abdelnour, A. F., Nehmeh, S. A., Pan, T., Humm, J. L., Vernon, P., Schöder, H. M., Rosenzweig, K., Mageras, G. S., Yorke, E. D., Larson, S. M., et al. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/12/012 · Cited by 150 · cited 1×
“These methods include respiratory-gated PET/CT (Nehmeh et al 2004a, 2004b, Pan et al 2004, 2007, Rietzel et al 2005, Wolthaus et al 2005, Abdelnour et al 2007, Dawood et al 2007, Guckenberger et al 2007, Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008,…”
MethodWolthaus, J., van Herk, M., Muller, S. H., Belderbos, J. S. A., Lebesque, J. V., de Bois, J. A., Rossi, M. M. G., & Damen, E. M. F. (2005) · Physics in Medicine and Biology · doi:10.1088/0031-9155/50/7/017 · Cited by 124 · cited 1×
“These methods include respiratory-gated PET/CT (Nehmeh et al 2004a, 2004b, Pan et al 2004, 2007, Rietzel et al 2005, Wolthaus et al 2005, Abdelnour et al 2007, Dawood et al 2007, Guckenberger et al 2007, Lu et al 2006, Wink et al 2006), deep-inspiration-breathhold (DIBH) PET/CT (Kawano et al 2008,…”
MethodPark, S., Ionascu, D., Killoran, J. H., Mamede, M. H., Gerbaudo, V. H., Chin, L. M., & Berbeco, R. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/13/018 · Cited by 109 · cited 1×
“…experiment with known true tumor SUVmax in motion-free images showed that the SUVmax can be underestimated by as much as 75% and the recovery coefficient can be as low as 0.2 in motion-blurred images, depending on different lesion size and motion amplitude (Park et al 2008, Pevsner et al 2005).”
BackgroundInfluentialMeirelles, G. D. S. P., Erdi, Y. E., Nehmeh, S. A., Squire, O. D., Larson, S. M., Humm, J. L., & Schöder, H. M. (2007) · Journal of Nuclear Medicine · doi:10.2967/jnumed.106.038034 · Cited by 103 · cited 2×
“…that various respiratory motion correction methods can cause the tumor SUVmax to increase as much as 159% and tumor volume to decrease as much as 43% compared to those of the conventional measurement in patient studies (Kawano et al 2008, Meirelles et al 2007, Nehmeh et al 2002a, 2002b, 2007).”
MethodResultKawano, T., Ohtake, E., & Inoue, T. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.049296 · Cited by 85 · cited 3×
“…that various respiratory motion correction methods can cause the tumor SUVmax to increase as much as 159% and tumor volume to decrease as much as 43% compared to those of the conventional measurement in patient studies (Kawano et al 2008, Meirelles et al 2007, Nehmeh et al 2002a, 2002b, 2007).”
BackgroundMethodResultInfluentialApplication of a spatially variant system model for 3-D whole-body pet image reconstruction
Alessio, A. M., & Kinahan, P. E. (2008) · IEEE International Symposium on Biomedical Imaging (ISBI) · doi:10.1109/isbi.2008.4541246 · Cited by 15 · cited 3×
“The detector point-spread function model used in this study only incorporated radial blurring for the major source of error from parallax and did not account for axial blurring (Alessio and Kinahan 2008).”
MethodInfluentialSelf-citation
Cited by
Vaquero, J. J., & Kinahan, P. (2015) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-071114-040723 · Cited by 383
Self-citationPET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging
Vandenberghe, S., & Marsden, P. K. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/4/r115 · Cited by 258
Gating Approaches in Cardiac PET Imaging
Lassen, M. L., Kwiecinski, J., & Slomka, P. J. (2019) · PET Clinics · doi:10.1016/j.cpet.2018.12.008 · Cited by 38
Metadata
- Authors
- Liu, C., Pierce II, L. A., Alessio, A. M., & Kinahan, P. E.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2009-11-20
- Volume
- 54
- Issue
- 24
- Pages
- 7345-7362
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W1965519237
- Cited by
- 232
- References
- 29
- FWCI
- 6.34
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.