Abstract
The Ingenuity TF PET-MRI is a newly released whole-body hybrid PET-MR imaging system with a Philips time-of-flight GEMINI TF PET and Achieva 3T X-series MRI system. Compared to PET-CT, modifications to the positron emission tomography (PET) gantry were made to avoid mutual system interference and deliver uncompromising performance which is equivalent to the standalone systems. The PET gantry was redesigned to introduce magnetic shielding for the photomultiplier tubes (PMTs). Stringent electromagnetic noise requirements of the MR system necessitated the removal of PET gantry electronics to be housed in the PET-MR equipment room. We report the standard NEMA measurements for the PET scanner. PET imaging and performance measurements were done at Geneva University Hospital as described in the NEMA Standards NU 2-2007 manual. The scatter fraction (SF) and noise equivalent count rate (NECR) measurements with the NEMA cylinder (20 cm diameter) were repeated for two larger cylinders (27 cm and 35 cm diameter), which better represent average and heavy patients. A NEMA/IEC torso phantom was used for overall assessment of image quality. The transverse and axial resolution near the center was 4.7 mm. Timing and energy resolution of the PET-MR system were measured to be 525 ps and 12%, respectively. The results were comparable to PET-CT systems demonstrating that the effect of design modifications required on the PET system to remove the harmful effect of the magnetic field on the PMTs was negligible. The absolute sensitivity of this scanner was 7.0 cps kBq(-1), whereas SF was 26%. NECR measurements performed with cylinders having three different diameters, and image quality measurements performed with IEC phantom yielded excellent results. The Ingenuity TF PET-MRI represents the first commercial whole-body hybrid PET-MRI system. The performance of the PET subsystem was comparable to the GEMINI TF PET-CT system using phantom and patient studies. It is conceived that advantages of hybrid PET-MRI will become more evident in the near future.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Simultaneous PET-MRI: a new approach for functional and morphological imaging
Judenhofer, M. S., Wehrl, H. F., Newport, D. F., Catana, C., Siegel, S. B., Becker, M., Thielscher, A., Kneilling, M., Lichy, M. P., Eichner, M., et al. (2008) · Nature Medicine · doi:10.1038/nm1700 · Cited by 981
Surti, S., Kuhn, A., Werner, M. E., Perkins, A. E., Kolthammer, J. A., & Karp, J. S. (2007) · PubMed · Cited by 639
Benefit of Time-of-Flight in PET: Experimental and Clinical Results
Karp, J. S., Surti, S., Daube-Witherspoon, M. E., & Muehllehner, G. (2008) · Journal of Nuclear Medicine · doi:10.2967/jnumed.107.044834 · Cited by 584
Simultaneous MR/PET Imaging of the Human Brain: Feasibility Study
Schlemmer, H.-P. W., Pichler, B. J., Schmand, M., Burbar, Z., Michel, C., Ladebeck, R., Jattke, K., Townsend, D., Nahmias, C., Jacob, P. K., et al. (2008) · Radiology · doi:10.1148/radiol.2483071927 · Cited by 474
PET/MRI: Paving the Way for the Next Generation of Clinical Multimodality Imaging Applications
Pichler, B. J., Kolb, A., Nägele, T., & Schlemmer, H.-P. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.061853 · Cited by 418
Simultaneous PET and MR imaging
Shao, Y., Cherry, S. R., Farahani, K., Meadors, K., Siegel, S., Silverman, R. W., & Marsden, P. K. (1997) · Physics in Medicine and Biology · doi:10.1088/0031-9155/42/10/010 · Cited by 374
Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques
Hofmann, M., Pichler, B., Schölkopf, B., & Beyer, T. (2009) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-008-1007-7 · Cited by 343
Simultaneous in vivo positron emission tomography and magnetic resonance imaging
Catana, C., Procissi, D., Wu, Y., Judenhofer, M. S., Qi, J., Pichler, B. J., Jacobs, R. E., & Cherry, S. R. (2008) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0711622105 · Cited by 272
Pre-clinical PET/MR: technological advances and new perspectives in biomedical research
Wehrl, H. F., Judenhofer, M. S., Wiehr, S., & Pichler, B. J. (2009) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-009-1078-0 · Cited by 181
PET/MR Images Acquired with a Compact MR-compatible PET Detector in a 7-T Magnet
Judenhofer, M. S., Catana, C., Swann, B. K., Siegel, S. B., Jung, W.-I., Nutt, R. E., Cherry, S. R., Claussen, C. D., & Pichler, B. J. (2007) · Radiology · doi:10.1148/radiol.2443061756 · Cited by 173
A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner
Slates, R., Farahani, K., Shao, Y., Marsden, P. K., Taylor, J., Summers, P. E., Williams, S., Beech, J. S., & Cherry, S. R. (1999) · Physics in Medicine and Biology · doi:10.1088/0031-9155/44/8/312 · Cited by 112
Is MR-guided Attenuation Correction a Viable Option for Dual-Modality PET/MR Imaging?
Zaidi, H. (2007) · Radiology · doi:10.1148/radiol.2443070092 · Cited by 80
Self-citationAn integrated MR/PET system: prospective applications
Schlemmer, H.-P., Pichler, B. J., Krieg, R., & Heiss, W. (2008) · Abdominal Imaging · doi:10.1007/s00261-008-9450-2 · Cited by 74
MR-based attenuation correction for a whole-body sequential PET/MR system
Hu, Z., Ojha, N., Renisch, S., Schulz, V., Torres, I., Buhl, A., Pal, D., Muswick, G. J., Penatzer, J., Guo, T., et al. (2009) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2009.5401802 · Cited by 69
Self-citationThe Clinical Role of Fusion Imaging Using PET, CT, and MR Imaging
Zaidi, H., Montandon, M., & Alavi, A. A. (2009) · Magnetic Resonance Imaging Clinics of North America · doi:10.1016/j.mric.2009.09.010 · Cited by 64
Self-citation3D-FRP: direct Fourier reconstruction with Fourier reprojection for fully 3-D PET
Matej, S., & Lewitt, R. M. (2001) · IEEE Transactions on Nuclear Science · doi:10.1109/23.958359 · Cited by 64
Attenuation correction for MR table and coils for a sequential PET/MR system
Zhang, B., Pal, D., Hu, Z., Ojha, N., Guo, T., Muswick, G. J., Tung, C., & Kaste, J. (2009) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2009.5401737 · Cited by 29
Self-citation
Cited by
Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
Delso, G., Fürst, S., Jakoby, B., Ladebeck, R., Ganter, C., Nekolla, S. G., Schwaiger, M., & Ziegler, S. I. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.092726 · Cited by 874
Drzezga, A., Souvatzoglou, M., Eiber, M., Beer, A. J., Fürst, S., Martinez-Möller, A., Nekolla, S. G., Ziegler, S., Ganter, C., Rummeny, E. J., et al. (2012) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.098608 · Cited by 508
Recent developments in time-of-flight PET
Vandenberghe, S., Mikhaylova, E., D’Hoe, E., Mollet, P., & Karp, J. S. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0138-3 · Cited by 368
History and future technical innovation in positron emission tomography
Jones, T., & Townsend, D. (2017) · Journal of Medical Imaging · doi:10.1117/1.jmi.4.1.011013 · Cited by 303
PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging
Vandenberghe, S., & Marsden, P. K. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/4/r115 · Cited by 258
Metadata
- Authors
- Zaidi, H., Ojha, N., Morich, M., Griesmer, J., Hu, Z., Maniawski, P., Ratib, O., Izquierdo-Garcia, D., Fayad, Z. A., & Shao, L.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2011-04-20
- Volume
- 56
- Issue
- 10
- Pages
- 3091-3106
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W1971098086
- Cited by
- 402
- References
- 17
- FWCI
- 32.86
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.