Photoplethysmography and its application in clinical physiological measurement
Open accessAbstract
Photoplethysmography (PPG) is a simple and low-cost optical technique that can be used to detect blood volume changes in the microvascular bed of tissue. It is often used non-invasively to make measurements at the skin surface. The PPG waveform comprises a pulsatile ('AC') physiological waveform attributed to cardiac synchronous changes in the blood volume with each heart beat, and is superimposed on a slowly varying ('DC') baseline with various lower frequency components attributed to respiration, sympathetic nervous system activity and thermoregulation. Although the origins of the components of the PPG signal are not fully understood, it is generally accepted that they can provide valuable information about the cardiovascular system. There has been a resurgence of interest in the technique in recent years, driven by the demand for low cost, simple and portable technology for the primary care and community based clinical settings, the wide availability of low cost and small semiconductor components, and the advancement of computer-based pulse wave analysis techniques. The PPG technology has been used in a wide range of commercially available medical devices for measuring oxygen saturation, blood pressure and cardiac output, assessing autonomic function and also detecting peripheral vascular disease. The introductory sections of the topical review describe the basic principle of operation and interaction of light with tissue, early and recent history of PPG, instrumentation, measurement protocol, and pulse wave analysis. The review then focuses on the applications of PPG in clinical physiological measurements, including clinical physiological monitoring, vascular assessment and autonomic function.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Prior research has suggested that PPG sensors used to measure the HR are liable to poor accuracy during activities with increased physical exertion or activities involving repetitive contractions of forearm skeletal muscles [36-38].”
From Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study (2018) · cites this paper as [37]
In the citing paper
“It has been suggested that during activities involving sustained muscle contractions or higher intensity exercises, the contact between the device’s PPG sensor and skin is decreased, leading to a disruption in the signal quality and causing poor quality data [36,37].”
From Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study (2018) · cites this paper as [37]
In the citing paper
“Firstly, surface EMG, increased electrode impedance, respiration induced baseline drift, and electrode contact movement can cause noise and motion artifacts.88”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [88]
In the citing paper
“However, the intrinsic PPG disturbances originating from physiological processes, such as respiration, vasoconstriction due to temperature changes or stress [57], have not been considered.”
From Modeling of artifacts in the wrist photoplethysmogram: Application to the detection of life-threatening arrhythmias (2021) · cites this paper as [57]
In the citing paper
“The photoplethysmogram (PPG) signal is a measure of arterial blood volume, which fluctuates with each heartbeat and is used by many wearables to monitor heart rate (HR).”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [6]
In the citing paper
“The mechanisms underlying the PPG pulse wave shape are not yet fully understood [6].”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [6]
In the citing paper
“The shape is influenced not only by the incident wave transmitted through the arteries from the heart but also by reflected waves from a range of arterial sites [6].”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [6]
In the citing paper
“Cardiac output, monitored in peri-operative and critical care settings, has been estimated from pulse wave shape and low-frequency PPG variations”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [6]
In the citing paper
“Thus, while the pulsating components of PPG are related to arterial blood volume, the non-pulsating component is a function of the average blood volume, respiration, the sympathetic nervous system, and thermoregulation [151].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [151]
In the citing paper
“These more recent wearable physical activity devices, which include a measure of heart rate, utilize photoplethysmography (PPG) to measure heart rate.”
From Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices (2017) · cites this paper as [17]
In the citing paper
“PPG is a simple and low-cost optical technique that detects blood volume changes in the microvascular bed of tissue [17].”
From Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices (2017) · cites this paper as [17]
Citations
Cites
Items 201–277 of 277
Fingertip photoplethysmography and migraine
Komatsu, K.-I., Fukutake, T., & Hattori, T. (2003) · Journal of the Neurological Sciences · doi:10.1016/s0022-510x(03)00208-9 · Cited by 24
Kvernebo, K., Megerman, J. J., Hamilton, G., & Abbott, W. M. (1989) · European Journal of Vascular Surgery · doi:10.1016/s0950-821x(89)80005-2 · Cited by 24
Arterial blood pressure in the skin measured by a photoelectric probe and external counterpressure.
Nielsen, P. E., Hl, P., & Gyntelberg, F. (1973) · PubMed · Cited by 24
A ring-type device for the noninvasive measurement of arterial blood pressure
Zheng, D., & Zhang, Y. (2004) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2003.1280819 · Cited by 23
Optical assessment of recovery of tissue blood supply after removal of externally applied pressure
Murray, A., & Marjanović, D. Z. (1997) · Medical & Biological Engineering & Computing · doi:10.1007/bf02534102 · Cited by 23
Foo, J. Y. A., Wilson, S. J., Williams, G. R., Coates, A. J. S., Harris, M.-A., & Cooper, D. M. (2005) · Physiological Measurement · doi:10.1088/0967-3334/26/3/015 · Cited by 23
Medical electro-optics: measurements in the human microcirculation
Jones, D. P. (1987) · Physics in Technology · doi:10.1088/0305-4624/18/2/305 · Cited by 23
A portable system for monitoring physiological responses to hypoglycaemia
Harris, N. D., Baykouchev, S. B., Marques, J. L. B., Cochrane, T., George, E., Heller, S. R., & Ward, J. D. (1996) · Journal of Medical Engineering & Technology · doi:10.3109/03091909609008998 · Cited by 23
Fourier analysis of the pulse wave in obliterating arteriosclerosis.
Oliva, I. O., Ipser, J., Roztočil, K., & Guttenbergerová, K. (1976) · PubMed · Cited by 23
Analysis of the photoplethysmographic signal by means of the decomposition in principal components
quez, R. H. E. I., Castellanos, M. S., guez, J. F. O. N. R. I., & Hernández-Cáceres, J. L. (2002) · Physiological Measurement · doi:10.1088/0967-3334/23/3/402 · Cited by 22
The quantitative aspect of photoplethysmography revised
Jespersen, L. T., & Pedersen, O. L. (1986) · Heart and Vessels · doi:10.1007/bf02128146 · Cited by 21
Stücker, M., Reich, S., Robak-Pawelczyk, B., Moll, C., Rudolph, T., Altmeyer, P. J., Weindorf, N., Hirche, H., Gambichler, T., & Schultz-Ehrenburg, U. (2005) · Journal of Vascular Surgery · doi:10.1016/j.jvs.2004.11.023 · Cited by 21
Assessment of the healing potentials of ulcers of the skin by photoplethysmography.
Lee, B. Y., Trainor, F. S., Kavner, D., Crisologo, J. A., Shaw, W. W.-L., & Madden, J. L. (1979) · PubMed · Cited by 20
KAWARADA, A., Shimazu, H., Yamakoshi, K., & Kamiya, A. (1986) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446261 · Cited by 20
Belcaro, G., LAURORA, G., Nicolaides, A. N., Agus, G. B., Cesarone, M. R., DeSanctis, M. T., INCANDELA, L., Ricci, A., Cazaubon, M., IPPOLITO, E., et al. (1998) · Angiology · doi:10.1177/000331979804901103 · Cited by 20
Age-related factors that confound peripheral pulse timing characteristics in Caucasian children
Foo, J. Y. A., Wilson, S. J., Williams, G. M., Harris, M., & Cooper, D. M. L. (2005) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001846 · Cited by 20
Effects of External Pressure on Arteries Distal to the Cuff During Sphygmomanometry
Nitzan, M., Rosenfeld, C., Weiss, A. T., Grossman, E., Patron, A., & Murray, A. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.846710 · Cited by 20
Suichies, H. E., Aarnoudse, J. G., Wouda, A. A., Jentink, H. W., de Mul, F. F. M., & Greve, J. (1992) · Angiology · doi:10.1177/000331979204300207 · Cited by 19
Dual-Channel Photoplethysmography to Monitor Local Changes in Vascular Stiffness
Foo, J. Y. A., & Lim, C. S. (2006) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-006-9024-1 · Cited by 19
Pulse Transit Time Based on Piezoelectric Technique at the Radial Artery
Foo, J. Y. A., & Lim, C. S. (2006) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-006-9019-y · Cited by 19
Psychological stress and arterial pulse transit time.
Dutch, J., & Redman, S. (1983) · PubMed · Cited by 19
Observations on the photoplethysmograph pulse derived from a laser Doppler flowmeter
Almond, N. E., & Cooke, E. D. (1989) · Clinical Physics and Physiological Measurement · doi:10.1088/0143-0815/10/2/003 · Cited by 18
Recent advances in non-invasive assessment of human skin blood flow.
Agache, P. G., & Dupond, A.-S. (1994) · PubMed · Cited by 18
Noninvasive Continuous Monitoring of Digital Pulse Waves during Hemodialysis
Burkert, A., Scholze, A., & Tepel, M. (2006) · ASAIO Journal · doi:10.1097/01.mat.0000199892.82612.8e · Cited by 17
A SIMPLE METHOD OF DEMONSTRATING CHANGES IN BLOOD SUPPLY OF THE EAR AND EFFECTS OF SOME MEASURES
HANZLIK, P. J., DeEds, F., & TERADA, B. (1936) · Journal of Pharmacology and Experimental Therapeutics · doi:10.1016/s0022-3565(25)04690-7 · Cited by 17
The reactions of the digital artery and minute pad arteries to local cold
Hertzman, A. B., & Roth, L. W. (1942) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1942.136.4.680 · Cited by 17 · cited 1×
“…vasoconstrictor waves, Traube Hering Mayer (THM) waves and also thermoregulation (Burton 1939, Burton and Taylor 1940, Hertzman and Dillon 1940b, Hertzman and Roth 1942a, 1942b, 1942c, Hertzman and Flath 1963, Hyndman et al 1971, Peñáz 1978, Ahmed et al 1982, Harness and Marjanovic 1989,…”
BackgroundOculopneumoplethysmography, Doppler Examina tion, and Supra orbital Photoplethysmography
Lynch, T. G., Wright, C. B., & MILLER, E. V. (1981) · Annals of Surgery · doi:10.1097/00000658-198112000-00012 · Cited by 16
DC photoplethysmography in the evaluation of sympathetic vasomotor responses
Barron, S. A., Rogowski, Z., KANTER, Y., & Hemli, J. A. (1993) · Clinical Physiology · doi:10.1111/j.1475-097x.1993.tb00472.x · Cited by 16
Bogadi‐Šare, A., & Zavalić, M. (1994) · International Archives of Occupational and Environmental Health · doi:10.1007/bf00383370 · Cited by 16
Tanaka, G., & Sawada, Y. (2003) · International Journal of Psychophysiology · doi:10.1016/s0167-8760(03)00056-4 · Cited by 16
de Trafford, J. C., Lafferty, K. J., Kitney, R. I., et al. (1982) · IEE Proceedings A Physical Science Measurement and Instrumentation Management and Education Reviews · doi:10.1049/ip-a-1:19820102 · Cited by 16
The Applicability of Color-Labeled Microspheres in a Rabbit Model of Ischemia-Reperfusion
Hasdai, D., Fuchs, L., Ohad, D. G., & Battler, A. (1994) · Angiology · doi:10.1177/000331979404500608 · Cited by 15
Development of a Compact Home Health Monitor for Telemedicine
Yoon, G., Lee, J. Y., Jeon, K. J., Park, K. K., & Kim, H. S. (2005) · Telemedicine Journal and e-Health · doi:10.1089/tmj.2005.11.660 · Cited by 15
de Trafford, J. C., Lafferty, K., Kitney, R. I., Cotton, L. T., & Roberts, V. C. (1982) · IEE Proceedings A Physical Science Measurement and Instrumentation Management and Education Reviews · doi:10.1049/ip-a-1.1982.0102 · Cited by 15
Toe pressure determination by audiophotoplethysmography
Fronek, A. S., Blažek, V., & Curran, B. (1994) · Journal of Vascular Surgery · doi:10.1016/0741-5214(94)90015-9 · Cited by 14
Measurement of the variability of the skin blood volume using dynamic spectroscopy
Nitzan, M., Babchenko, A., Milston, A., Turivnenko, S., Khanokh, B., & Mahler, Y. (1996) · Applied Surface Science · doi:10.1016/s0169-4332(96)00387-x · Cited by 14
Corte, M. D., Locchi, F., Spinelli, E., & Scarpelli, P. T. (1979) · Physics in Medicine and Biology · doi:10.1088/0031-9155/24/3/010 · Cited by 13
The reliability of photoelectric technique for measuring systolic blood pressure of the limbs.
Hirai, M., & Kawai, S. (1977) · PubMed · Cited by 12
Allen, J. (2002) · OpenGrey (Institut de l'Information Scientifique et Technique) · Cited by 12
Self-citationA practice of vascular surgery
Kester, R. C., & Leveson, S. H. (1981) · Medical Entomology and Zoology · Cited by 11
Morris, A. M. (1990) · The Lancet · doi:10.1016/0140-6736(90)91798-f · Cited by 10
Stern, R. M. (1974) · Psychophysiology · doi:10.1111/j.1469-8986.1974.tb00824.x · Cited by 10
The effect of hematocrit on photoelectric plethysmogram.
OCHOA, W., & Ohara, I. (1980) · The Tohoku Journal of Experimental Medicine · doi:10.1620/tjem.132.413 · Cited by 10
Photoelectric plethysmography; methods, normal standards, and clinical application.
Simonson, E. (1956) · PubMed · Cited by 10
Reproducibility of noninvasive tests of peripheral occlusive arterial disease
Osmundson, P. J., O’Fallon, W. M., Clements, I. P., et al. (1985) · Journal of Vascular Surgery · doi:10.1067/mva.1985.avs0020678 · Cited by 10
Vascular Tone in Patients with Hemorrhagic Shock
Miyagatani, Y., Yukioka, T., Ohta, S., Ohta, S., Matsuda, H., Shimazu, H., & Shimazaki, S. (1999) · doi:10.1097/00005373-199908000-00010 · Cited by 9
Quantitative measurement of sympathetic neuropathy in patients with diabetes mellitus
Nasimi, S. G. A. A., Mearns, A. J., Harness, J. B., & Heath, I. (1991) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(91)90127-s · Cited by 9
Tordoir, J. H. M., Haeck, L., Winterkamp, H., & Dekkers, W. J. M. (1986) · Journal of Vascular Surgery · doi:10.1067/mva.1986.avs0030456 · Cited by 9
Brumen, V., Horvat, D., & Bonić, I. (1994) · Microvascular Research · doi:10.1006/mvre.1994.1020 · Cited by 8
On a non-invasive evaluation of pulse wave velocity in human peripheral arteries
Gizdulich, P. (1984) · Clinical Physics and Physiological Measurement · doi:10.1088/0143-0815/5/1/006 · Cited by 8
The use of photoconductive cells in photoplethysmography
Fine, S., & Weinman, J. (1973) · Medical & Biological Engineering & Computing · doi:10.1007/bf02483691 · Cited by 8
Photoplethysmography: Selecting optoelectronic components
Burke, M. J., & Whelan, M. (1986) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446270 · Cited by 8
Handbook of Noninvasive Diagnostic Techniques in Vascular Surgery.
(1983) · Annals of Internal Medicine · doi:10.7326/0003-4819-98-6-1046_1 · Cited by 7
Rasmussen, B., Peura, R. A., Wheeler, H. B., & Cutler, B. S. (1981) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1981.324830 · Cited by 7
Simultaneous Measurement of the Photoplethysmographic Signal Variability in the Right and Left Hands
Nitzan, M., Vatine, J., Babchenko, A., Khanokh, B., Tsenter, J., & Stessman, J. (1998) · Lasers in Medical Science · doi:10.1007/s101030050073 · Cited by 7
Effect of external cuff pressure on arterial compliance
Zheng, D., Allen, J., & Murray, A. (2005) · Computers in cardiology · doi:10.1109/cic.2005.1588100 · Cited by 7
Self-citationApplying Neural Network Techniques to Plethysmographic Pulse Shape Analysis
Weng, J., Knoop, P., Hornberger, C., & Konecny, E. (2009) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmte.1998.43.s1.92 · Cited by 6
Periodic posture stimulation of the baroreceptors and the local vasomotor reflexes
Nasimi, S. G. A. A., Harness, J. B., Marjanović, D. Z., Knight, T., & Mearns, A. J. (1992) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(92)90005-6 · Cited by 6
Low-frequency photoplethysmograph signals
Harness, J. B., & Marjanovic, D. Z. (1989) · Clinical Physics and Physiological Measurement · doi:10.1088/0143-0815/10/4/009 · Cited by 6
Jones, M. E., Withey, S., Grover, R., & Smith, P. (2000) · British Journal of Plastic Surgery · doi:10.1054/bjps.2000.3359 · Cited by 6
Heck, A. F., & Vr, H. (1975) · PubMed · Cited by 6
The use of light reflection rheography in diagnosing venous disease and arterial microcirculation.
Incze, A., Lázár, I., Abraham, E., Copotoiu, M., & Cotoi, S. (2003) · PubMed · Cited by 5
Vibration syndrome diagnosis using a cooling test verified by computerized photoplethysmography
Dyszkiewicz, A. J., & Tendera, M. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/4/003 · Cited by 5
Cardiovascular Monitoring of Elective Aortic Aneurysm Repair using Methods of Chaos Analysis
Christ, F., Abicht, J., Athelogou, M., Basehnegger, H., Niklas, M., Peter, K., & Meßmer, K. (2008) · International Journal of Microcirculation · doi:10.1159/000179253 · Cited by 5
Zheng, D., Allen, J., & Murray, A. (2006) · Cited by 5
Self-citationSilver, D. K., Malvezzi, L. K., Castronuovo, J. J., Swayne, L. C., & Cone, D. C. (1992) · Journal of Vascular Surgery · doi:10.1067/mva.1992.37088 · Cited by 5
到, 高. (1984) · The Annals of physiological anthropology · Cited by 4
CONTINUOUS SIMULTANEOUS REGISTRATION OF SWEATING AND BLOOD FLOW IN A SMALL SKIN AREA
Hertzman, A. B., & Flath, F. (1963) · doi:10.21236/ad0402762 · Cited by 4 · cited 1×
“…Mayer (THM) waves and also thermoregulation (Burton 1939, Burton and Taylor 1940, Hertzman and Dillon 1940b, Hertzman and Roth 1942a, 1942b, 1942c, Hertzman and Flath 1963, Hyndman et al 1971, Peñáz 1978, Ahmed et al 1982, Harness and Marjanovic 1989, Nitzan et al 1996b, 1996a, general…”
BackgroundDetectivities of photoconductive and silicon p-i-n light sensors in photoplethysmography.
Weinman, J., & Fine, S. (1972) · PubMed · Cited by 4
Marston, W. A. (2002) · Seminars in Vascular Surgery · doi:10.1053/svas.2002.30451 · Cited by 4
MATSUURA, M. (2002) · Perceptual and Motor Skills · doi:10.2466/pms.2002.94.2.607 · Cited by 3
Sophromadze, Z., Chabashvili, N., & Kakhabrishvili, Z. (2006) · PubMed · Cited by 3
Evaluation of vasodilator effects by means of changes of photoplethysmographic pulse waveshape.
Ruiz-Vega, H., Benitez, G., Carranza-Madrigal, J., & Ms, H.-A. (1998) · PubMed · Cited by 3
Grossmann, K. F., Rolka, H., Reinhardt, H., & Hc, R. (1987) · PubMed · Cited by 2
Jaron, D., Moore, T. E., Reddy, B. R. S., Hrebien, L., & Kepics, F. (1987) · PubMed · Cited by 2
McGuigan, M. R., Bronks, R., Newton, R. U., Graham, J. C., & Cody, D. V. (2002) · Journal of Aging and Physical Activity · doi:10.1123/japa.10.1.28 · Cited by 2
MATSUURA, M. (2002) · Perceptual and Motor Skills · doi:10.2466/pms.94.2.607-622 · Cited by 1
Research works printed but not linked: 37
- Abramowitz H B, 1979, Surgery, 86, 434
- Agache P G, 1994, Acta. Derm. Venereol. Suppl. (Stockholm), 185, 47
- Aoyagi T Kiahi M Yamaguchi K Watanabe S, 1974, Abstracts of the 13th Annual Meeting of the Japanese Society of Medical Electronics and Biological Engineering, 90
- Aoyagi T, 2002, Anesth. Analg., 94, 51
- Barnes R W, 1991, Circulation, 83, I20
- Dutch J, 1983, N. Z. Med. J., 96, 607
- Evans M L, 1988, Med. Instrum., 22, 29
- Grohmann G, 1996, Perfusion, 9, 268
- Grohmann G, 1996, Perfusion, 9, 300
- Grossmann K, 1987, VASA, 16, 324
- Hanzlik P J, 1936, J. Pharmac. Exp. Ther., 56, 194
- Heck A F, 1975, Med. Instrum., 9, 88
- Hertzer N R, 1991, Circulation, 83, I12
- Hertzman A B, 1963, Aerospace Med., 34, 710
- Hertzman A B, 1937, Am. J. Physiol., 119, 334
- Hirai M, 1977, VASA, 6, 215
- Incze A, 2003, Rom. J. Int. Med., 41, 35
- Lane I F, 1984, Hosp. Update, 861
- 10.1023/A:1016942504084 · DOI
- Lee B Y, 1979, Surg. Gynecol. Obstet., 148, 233
- Miwa Z, 2002, Pediatr. Dent., 24, 594
- Molitor H, 1936, J. Pharmacol. Exp. Ther., 57, 6
- Nielsen P E, 1973, VASA, 2, 65
- Ochoa W, 1980, J. Exp. Med., 132, 413
- Oliva I, 1976, VASA, 5, 95
- Peñáz J, 1973, Digest 10th Int Conf. Med. Biol. Eng., 104
- Peñáz J, 1978, Automedica, 2, 135
- Porret C A, 1995, Am. J. Physiol., 269, H1852
- Ruiz-Vega H, 1998, Proc. West Pharmacol. Soc., 41, 73
- Shusterman V, 1997, Am. J. Physiol., 273, R1173
- Simonson E, 1956, Geriatrics, 11, 425
- Sophromadze Z, 2006, Georgian Med. News, 133, 72
- 10.1023/A:1016936206694 · DOI
- Weinman J, 1972, T-I-T J. Life Sci., 2, 121
- Weng J, 1998, Biomed. Tech., 43, 92
- Zheng D, 2006, Comput. Cardiol., 33, 289
- Zheng D C Zhang Y T, 2003, Proc. 25th Annual International Conf. of the IEEE EMBC, 4, 3184
Cited by
Sorted by citations
Remote plethysmographic imaging using ambient light
Verkruysse, W., Svaasand, L. O., & Nelson, J. S. (2008) · Optics Express · doi:10.1364/oe.16.021434 · Cited by 1,728
Non-contact, automated cardiac pulse measurements using video imaging and blind source separation
Poh, M., McDuff, D., & Picard, R. W. (2010) · Optics Express · doi:10.1364/oe.18.010762 · Cited by 1,652
Advancements in Noncontact, Multiparameter Physiological Measurements Using a Webcam
Poh, M., McDuff, D., & Picard, R. W. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2086456 · Cited by 1,514
Wearable sensors: modalities, challenges, and prospects
Heikenfeld, J., Jajack, A., Rogers, J., Gutruf, P., Tian, L., Pan, T., Li, R., Khine, M., Kim, J., Wang, J., et al. (2018) · Lab on a Chip · doi:10.1039/c7lc00914c · Cited by 1,434
Wearable Sensors for Remote Health Monitoring
Majumder, S., Mondal, T., & Deen, M. (2017) · Sensors · doi:10.3390/s17010130 · Cited by 1,375
Metadata
- Authors
- Allen, J.
- Affiliations
- Freeman Hospital (GB)
- Corresponding author
- Allen, John
- ORCID
- Allen, John
- Venue
- Physiological Measurement
- Publisher
- IOP Publishing
- Published
- 2007-02-20
- Volume
- 28
- Issue
- 3
- Pages
- R1-R39
- ISSN
- 0967-3334 (ISSN-L)
- 1361-6579
- OpenAlex
- W2083872334
- Cited by
- 3,967
- References
- 275
- Citations by year
- 2026: 308 · 2025: 292 · 2024: 310 · 2023: 317 · 2022: 323 · 2021: 314 · 2020: 291 · 2019: 304 · 2018: 299 · 2017: 227 · 2016: 185 · 2015: 216 · 2014: 159 · 2013: 132 · 2012: 86
- FWCI
- 3.44
- Type
- journal article
- Language
- en
- Open access
- green
- doi.org
- Locations
- Physiological Measurement (publishedVersion)
- PubMed (publishedVersion)
- Newcastle University ePrints (Newcastle Univesity)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- photoplethysmography, blood volume changes, pulse wave analysis, non-invasive physiological monitoring, oxygen saturation, blood pressure measurement, cardiac output, autonomic function, peripheral vascular disease, vascular assessment, light-tissue interaction
- MeSH terms
- Autonomic Nervous System, Blood Vessels, Humans, Monitoring, Physiologic, Vascular Diseases, Photoplethysmography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review