Wearable sensors for monitoring the physiological and biochemical profile of the athlete
Open accessAbstract
Athletes are continually seeking new technologies and therapies to gain a competitive edge to maximize their health and performance. Athletes have gravitated toward the use of wearable sensors to monitor their training and recovery. Wearable technologies currently utilized by sports teams monitor both the internal and external workload of athletes. However, there remains an unmet medical need by the sports community to gain further insight into the internal workload of the athlete to tailor recovery protocols to each athlete. The ability to monitor biomarkers from saliva or sweat in a noninvasive and continuous manner remain the next technological gap for sports medical personnel to tailor hydration and recovery protocols per the athlete. The emergence of flexible and stretchable electronics coupled with the ability to quantify biochemical analytes and physiological parameters have enabled the detection of key markers indicative of performance and stress, as reviewed in this paper.
AI snapshot
Written from the paper’s full text.
TL;DR
This narrative review surveys wearable sensors for monitoring athletes’ physiological and biochemical status, with emphasis on noninvasive biomarkers from saliva and eccrine sweat. It argues that continuous, validated multimodal sweat/saliva sensing remains an unmet need for tailoring hydration, recovery, and stress protocols.
Key findings
- Current sports wearables mainly track movement and physiological parameters; biochemical monitoring from saliva/sweat is emerging. Table 4 lists sweat ranges including glucose 10–200 µM, lactate 5–20 mM, Na+ 10–100 mM, Cl− 10–100 mM, K+ 1–18.5 mM, pH 3–8, NH4+ 0.1–1.1 mM, and cortisol 8–140 ng/L; saliva ranges include alpha-amylase 5–17 U/mL, glucose 1 µM, lactate 5–50 µM, and phosphate 3.6–300 µM.
- Mouthguard biosensors showed salivary lactate measurement with a correlation coefficient of 0.988 in human saliva and stable response over 2 h; a salivary uric acid mouthguard measured sustained high salivary UA for 5 h in an untreated hyperuricemia patient.
- Sweat microfluidic/iontophoresis devices showed translational promise: a fluorometric microfluidic device measured sweat Cl− 28–31 mM, Zn2+ ~2.5 µM, and Na+ 35–50 mM, matching laboratory methods (Cl− 28 mM, Zn2+ 3.6 µM, Na+ 36 mM); a waterproof epidermal device adhered for >2 h and measured during the IRONMAN triathlon; an iontophoretic ethanol sensor showed blood-to-sweat lag times of 2.3–11.41 min.
- For stress/mental acuity, a systematic review of 18 studies (16 ECG-HRV, 2 PPG-PRV) found HRV correlation validated at rest but declining with exercise intensity; a stress patch reported skin temperature sensitivity 0.31 Ω/°C, skin conductance sensitivity 0.28 μV/0.02 μS, and pulse wave response time 70 ms; a cortisol sensor correlated with standard ELISA with an RSD of 5%.
Methods
- Narrative review of wearable sensors for athlete physiological and biochemical monitoring, covering scientific literature and the commercial landscape (Tables 1 and 2).
- Commercial device tables were compiled from company websites and social media; the biomarker table was adapted from Malon et al. and Bariya et al. No systematic search strategy, inclusion criteria, or meta-analysis is described.
- The review discusses device examples, biomarker concentration ranges, and validation studies across saliva, eccrine sweat, heart rate variability, skin conductance, and stress biomarkers.
Limitations
- The paper states that many devices are not yet validated in clinical studies for real-time assessment of human performance; it calls for formal clinical studies in nonstationary conditions and larger populations. It also notes saliva is limited to sports requiring mouthguards and current salivary devices do not permit continuous measurements; sweat sensing faces skin contamination, dried sweat, low sampling rates, weather dependence, and motion artifacts; and there remains a need to differentiate physical from mental stress.
- Evident from the text: this is a narrative review without a systematic search or meta-analysis, and the commercial landscape relies on company websites/social media rather than peer-reviewed validation; many cited device studies are proof-of-concept or small-N.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 164 cited sources checked · 0 supported · 0 flagged
- Findingnone given
Analyte concentrations found in saliva are far below those detected in eccrine sweat.
“In addition, analyte concentrations found in saliva are far below those detected in eccrine sweat (Table 4).”
Quote in the paperSubstantial claim - Findingnone given
Companies such as Epicore Biosystems, Halo Wearables, GraphWear, and Kenzen Wear are currently developing epidermal sensors for sweat detection.
“Companies, such as Epicore Biosystems, Halo Wearables, GraphWear, and Kenzen Wear are currently developing epidermal sensors for sweat detection.”
Quote in the paperNot a substantial claim - Premisenone given
Over the last decade, significant progress has been made to develop wearable electrochemical sensors that detect biomarkers non-invasively from biofluids such as saliva and sweat, both of which are easily accessible without impeding the performance of an athlete.
“Over the last decade, significant progress has been made to develop wearable electrochemical sensors that detect biomarkers non-invasively from biofluids such as saliva and sweat, both of which are easily accessible without impeding the performance of an athlete.”
Quote in the paperSubstantial claim - Premisecitation
Electrochemical wearable sensors have received considerable attention recently because of their potential to monitor a wide array of biomarkers in a continuous and non-invasive manner.
“Electrochemical wearable sensors have received considerable attention recently because of their potential to monitor a wide array of biomarkers in a continuous and non-invasive manner.3”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [3] Bandodkar, A. J. & Wang, J. Non-invasive wearable electrochemical sensors: a review. Trends Biotechnol. 32, 363–371 (2014). source
Not checked in the free run.
- Interpretationreasoning
Existing wearable devices that focus on physical or physiological parameters do not permit team trainers and physicians to quantify the biochemical profile of an athlete in a real-time manner with the goal of alleviating soft-tissue injuries, dehydration, or cramping.
“However, such devices, do not permit team trainers and physicians to quantify the biochemical profile of an athlete in a real-time manner with the goal of alleviating soft-tissue injuries, dehydration, or cramping.”
Quote in the paperSubstantial claim
- Premisecitation
Data pertinent to player movement from wearable devices has been used to inform coaches of their players workout load to indicate which players are at a higher risk to suffer a soft-tissue injury or those that should be sidelined to prevent the occurrence or reoccurrence of an injury during high acuity training periods.
“For example, data pertinent to player movement from wearable devices has been used to inform coaches of their players workout load to indicate which players are at a higher risk to suffer a soft-tissue injury or those that should be sidelined to prevent the occurrence or reoccurrence of an injury during high acuity training periods.9”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [9] Cabot, M. K., clevel & .com. Hue Jackson Says There are no Hard Feelings Here in Hardland Between Him and Todd Haley Despite Hard Knocks Scene. cleveland.com. https://www.cleveland.com/browns/index.ssf/2018/08/hue_jackson_says_theres_no_har.html (2018). Accessed 28 Nov 2018.
Not checked in the free run.
- Premisecitation
The use of wearable technology coupled with insightful analytics and necessary athlete-specific recovery protocols have shown to alleviate soft-tissue injuries in an NFL team over a 2-year period.
“The value wearable devices are having in sports to track and correlate player workouts, exertions, and loads to soft-tissue injuries can be traced to an interview given by a coach in the National Football League (NFL) where he affirmed that the use of such technology coupled with insightful analytics and necessary athlete-specific recovery protocols have shown to alleviate soft-tissue injuries in”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [9] Cabot, M. K., clevel & .com. Hue Jackson Says There are no Hard Feelings Here in Hardland Between Him and Todd Haley Despite Hard Knocks Scene. cleveland.com. https://www.cleveland.com/browns/index.ssf/2018/08/hue_jackson_says_theres_no_har.html (2018). Accessed 28 Nov 2018.
Not checked in the free run.
- Premisecitation
Saliva is considered an attractive and emerging option compared to direct blood analysis for quantifying biomarkers related to human performance due to its noninvasive nature and continuous supply.
“Saliva is considered an attractive and emerging option compared to direct blood analysis for quantifying biomarkers related to human performance due to its noninvasive nature and continuous supply11 (Table 3).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [11] Chiappin, S., Antonelli, G., Gatti, R. & De Palo, E. F. Saliva specimen: a new laboratory tool for diagnostic and basic investigation. Clin. Chim. Acta Int. J. Clin. Chem. 383, 30–40 (2007). source
Not checked in the free run.
- Premisecitation
Commercial hand-held sensors, such as ACCU-CHEK® by Roche Diagnostics, iSTAT® by Abbot, and Lactate Scout® by Sports Resource Group, exist for the measurement of metabolites and electrolytes.
“This has led to the development of commercial hand-held sensors, such as ACCU-CHEK® by Roche Diagnostics, iSTAT® by Abbot, and Lactate Scout® by Sports Resource Group for the measurement of metabolites and electrolytes.3”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [3] Bandodkar, A. J. & Wang, J. Non-invasive wearable electrochemical sensors: a review. Trends Biotechnol. 32, 363–371 (2014). source
Not checked in the free run.
- Premiseown data
The mouthguard lactate biosensor demonstrated stability over a similar 2-h period when treated with unspiked human saliva.
“In addition, the biosensor demonstrated stability over a similar 2-h period when treated with unspiked human saliva (Fig. 1c).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] Kim, J. et al. Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites. Analyst 139, 1632 (2014). source
Not checked in the free run.
- Premiseown data
The mouthguard lactate biosensor tested favorably to measuring lactate levels in unstimulated human saliva with good linearity and a correlation coefficient of 0.988.
“Furthermore, the biosensor tested favorably to measuring lactate levels in unstimulated human saliva with good linearity and a correlation coefficient of 0.988 (Fig. 1b).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [14] Kim, J. et al. Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites. Analyst 139, 1632 (2014). source
Not checked in the free run.
- Premiseown data
The efficacy of the salivary UA biosensor was successfully assessed in artificial saliva, undiluted human saliva, and in a hyperuricemia patient with and without medication control.
“The efficacy of the biosensor was successfully assessed in artificial saliva, undiluted human saliva, and in a hyperuricemia patient with and without medication control.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [15] Kim, J. et al. Wearable salivary uric acid mouthguard biosensor with integrated wireless electronics. Biosens. Bioelectron. 74, 1061–1068 (2015). source
Not checked in the free run.
- Premiseown data
The mouthguard lactate biosensor demonstrated a highly stable response over a 2-h duration.
“Results from the corresponding chronoamperogram demonstrated a highly stable response over a 2-h duration.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [14] Kim, J. et al. Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites. Analyst 139, 1632 (2014). source
Not checked in the free run.
- Premisecitation
A collaborative team from PARC, a Xerox Company, NextFlex, and the University of California, San Diego fabricated a smart mouthguard biosensor to detect early signs of dehydration, exhaustion, and mental state based on lactate and glucose measurements from saliva.
“Recently, a collaborative team from PARC, a Xerox Company, NextFlex, and the University of California, San Diego fabricated a smart mouthguard biosensor to detect early signs of dehydration, exhaustion, and mental state based on lactate and glucose measurements from saliva.16,17”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [16] PARC Mouthguard Helps Track Athlete Fatigue Through Saliva. SportTechie https://www.sporttechie.com/xerox-parc-nextflex-ucsd-saliva-mouthguard-athlete-fatigue/ (2018). Accessed 15 Sep 2018.
Not checked in the free run.
- cites [17] Dormehl, L. Spit-checking mouthguard can tell if athletes are tired or mentally drained. Digital Trends. https://www.digitaltrends.com/cool-tech/xerox-parc-mouthguard-saliva/ (2018). Accessed 15 Sep 2018.
Not checked in the free run.
- Premiseown data
The untreated hyperuricemia patient showed a sustained high-SUA level for 5 h.
“The untreated hyperuricemia patient showed a sustained high-SUA level for 5 h (Fig. 1e).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15] Kim, J. et al. Wearable salivary uric acid mouthguard biosensor with integrated wireless electronics. Biosens. Bioelectron. 74, 1061–1068 (2015). source
Not checked in the free run.
- Premisecitation
Biomarkers detected in saliva such as alpha-amylase, glucose, lactate, phosphate, and uric acid have been shown to have a good correlation with that found in blood.
“In addition, biomarkers detected in saliva such as alpha-amylase, glucose, lactate, phosphate, and uric acid (UA) have been shown to have a good correlation with that found in blood12 (Table 4).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [12] Nunes, L. A. S., Mussavira, S. & Bindhu, O. S. Clinical and diagnostic utility of saliva as a non-invasive diagnostic fluid: a systematic review. Biochem. Med. 25, 177–192 (2015). source
Not checked in the free run.
- Premisecitation
Researchers incorporated an amperometric enzymatic biosensor in polymeric mouthguards for monitoring salivary lactate and UA concentrations.
“Researchers incorporated an amperometric enzymatic biosensor in polymeric mouthguards for monitoring salivary lactate (Fig. 1a–c) and UA concentrations14 (Fig. 1d, e).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] Kim, J. et al. Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites. Analyst 139, 1632 (2014). source
Not checked in the free run.
- Interpretationreasoning
The work demonstrated the ability to detect glucose, a key marker indicative of fatigue levels.
“While a sensor embedded in a cavity may be far removed for sports medicine applications, the work demonstrated the ability to detect glucose, a key marker indicative of fatigue levels.”
Quote in the paperSubstantial claim - Premisecitation
The smart mouthguard biosensor enabled replacement of the electrodes to measure other biomarkers such as UA.
“The device enabled replacement of the electrodes to measure other biomarkers such as UA.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [16] PARC Mouthguard Helps Track Athlete Fatigue Through Saliva. SportTechie https://www.sporttechie.com/xerox-parc-nextflex-ucsd-saliva-mouthguard-athlete-fatigue/ (2018). Accessed 15 Sep 2018.
Not checked in the free run.
- cites [17] Dormehl, L. Spit-checking mouthguard can tell if athletes are tired or mentally drained. Digital Trends. https://www.digitaltrends.com/cool-tech/xerox-parc-mouthguard-saliva/ (2018). Accessed 15 Sep 2018.
Not checked in the free run.
- Premisecitation
Wearable devices capable of artificially inducing sweat through the introduction of current (iontophoresis) can overcome hurdles associated with low-collection volumes or skin contamination.
“Wearable devices capable of artificially inducing sweat through the introduction of current (iontophoresis) can overcome hurdles associated with low-collection volumes (e.g., attributed to weather) or skin contamination,20,21 as discussed later.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [20] Bariya, M., Nyein, H. Y. Y. & Javey, A. Wearable sweat sensors. Nat. Electron. 1, 160–171 (2018). source
Not checked in the free run.
- cites [21] Ferrell, W. R. et al. Elimination of electrically induced iontophoretic artefacts: implications for non-invasive assessment of peripheral microvascular function. J. Vasc. Res. 39, 447–455 (2002). source
Not checked in the free run.
- Premisecitation
Epicore Biosystems has established large volume manufacturing for the continuous and noninvasive measurement of various biomarkers from eccrine sweat.
“Epicore Biosystems has established large volume manufacturing for the continuous and noninvasive measurement of various biomarkers from eccrine sweat.24”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [24] Northwestern Engineering Teams with Sports and Rehabilitation Leaders to Advance Sweat Analysis Technologies. Northwestern Eng. http://www.mccormick.northwestern.edu/news/articles/2018/02/northwestern-engineering-teams-with-sports-and-rehabilitation-leaders-to-advance-sweat-analysis-technology.html. Accessed 16 Sep 2018.
Not checked in the free run.
- Premisenone given
Current devices discussed in literature have not shown continuous measurements of biomarkers from saliva to be possible.
“Furthermore, current devices discussed in literature have not shown continuous measurements of biomarkers from saliva to be possible (Table 3).”
Quote in the paperNot a substantial claim - Premisecitation
Replacing Na+ and Cl− levels after high-intensity situations is instrumental in maintaining electrolyte balance due to their role in stimulating hydration.
“Replacing Na+ and Cl− levels after high-intensity situations is instrumental in maintaining electrolyte balance due to their role in stimulating hydration.29–31”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [29] Baker, L. B. Sweating rate and sweat sodium concentration in athletes: a review of methodology and intra/interindividual variability. Sports Med. Auckl. NZ 47, 111–128 (2017). source
Not checked in the free run.
- cites [30] Baker, L. B., Barnes, K. A., Anderson, M. L., Passe, D. H. & Stofan, J. R. Normative data for regional sweat sodium concentration and whole-body sweating rate in athletes. J. Sports Sci. 34, 358–368 (2016). source
Not checked in the free run.
- cites [31] Lott, M. J. E. & Galloway, S. D. R. Fluid balance and sodium losses during indoor tennis match play. Int. J. Sport Nutr. Exerc. Metab. 21, 492–500 (2011). source
Not checked in the free run.
- Premisecitation
Sweat provides an ideal source toward the continuous and noninvasive measurement of biomarkers, such as sodium, chlorine, potassium, lactate, calcium, glucose, ammonia, ethanol, urea, cortisol, and various neuropeptides and cytokines.
“Sweat provides an ideal source toward the continuous and noninvasive measurement of biomarkers, such as sodium, chlorine, potassium, lactate, calcium, glucose, ammonia, ethanol, urea, cortisol, and various neuropeptides and cytokines.18”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [18] Heikenfeld, J. Non-invasive analyte access and sensing through eccrine sweat: challenges and outlook circa 2016. Electroanalysis 28, 1242–1249 (2016). source
Not checked in the free run.
- Premisecitation
GraphWear Technologies has piloted their sweat sensor technology with a professional football team.
“GraphWear Technologies has piloted their sweat sensor technology with a professional football team.27”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [27] Baum, S. GraphWear Technologies pilots sweat sensor patch with NFL team. MedCity News (2016). https://medcitynews.com/2016/10/graphwear-sweat-sensor-wearable-and-nfl/.
Not checked in the free run.
- Interpretationnone given
The main drawback of saliva being used for real-time measurements of human performance compared to eccrine sweat is that it is limited to sports which require mouthguard devices.
“The main drawback of saliva being used for real-time measurements of human performance compared to eccrine sweat is that it is limited to sports which require mouthguard devices.”
Quote in the paperNot a substantial claim - Premisecitation
Halo H1 has developed the first noninvasive wristband sensor for monitoring hydration levels in athletes by utilizing optical and electrical sensors.
“Halo H1 has developed the first noninvasive wristband sensor for monitoring hydration levels in athletes by utilizing optical and electrical sensors.25”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [25] Miller, D. W. & Miller, D. R. Hydration Monitoring Apparatus. 1–15 (2014). https://patentimages.storage.googleapis.com/89/6c/7a/34703870dde191/US20140221792A1.pdf.
Not checked in the free run.
- Premisecitation
GraphWear has fabricated a graphene-based epidermal sensor, which adheres to the torso, to detect glucose and lactic acid for assessing hydration levels.
“GraphWear has fabricated a graphene-based epidermal sensor, which adheres to the torso, to detect glucose and lactic acid for assessing hydration levels.26”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [26] GraphWear. http://www.graphwear.co/. Accessed 5 May 2017.
Not checked in the free run.
- Premisecitation
Researchers successfully developed a wrist-based potentiometric wearable device capable of detecting Cl− concentrations over time from sweat for cystic fibrosis monitoring.
“While there remains an unmet medical need to measure Cl− levels in real-time for sports, researchers successfully developed a wrist-based potentiometric wearable device capable of detecting Cl− concentrations over time from sweat for cystic fibrosis monitoring.36”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [36] Choi, D.-H., Kim, J. S., Cutting, G. R. & Searson, P. C. Wearable potentiometric chloride sweat sensor: the critical role of the salt bridge. Anal. Chem. 88, 12241–12247 (2016). source
Not checked in the free run.
- Premisecitation
The total Na+ loss from sweat is a function of whole-body sweating rate and sweat Na+ concentration.
“The total Na+ loss from sweat is a function of whole-body (WB) sweating rate and sweat Na+ concentration ([Na+]); thus, quantifying sodium loss from sweat is vital in expediting player recovery and minimizing soft tissue injuries brought about by the onset of dehydration.32”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [32] Allan, J. R. & Wilson, C. G. Influence of acclimatization on sweat sodium concentration. J. Appl. Physiol. 30, 708–712 (1971). source
Not checked in the free run.
- Interpretationcitation
Quantifying sodium loss from sweat is vital in expediting player recovery and minimizing soft tissue injuries brought about by the onset of dehydration.
“The total Na+ loss from sweat is a function of whole-body (WB) sweating rate and sweat Na+ concentration ([Na+]); thus, quantifying sodium loss from sweat is vital in expediting player recovery and minimizing soft tissue injuries brought about by the onset of dehydration.32”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [32] Allan, J. R. & Wilson, C. G. Influence of acclimatization on sweat sodium concentration. J. Appl. Physiol. 30, 708–712 (1971). source
Not checked in the free run.
- Premisecitation
Baker et al. developed a model to calculate the whole-body sweat Na+ concentration using absorbent patches from the forearm.
“Baker et al. developed a model to calculate the WB sweat Na+ concentration using absorbent patches from the forearm33 (Eq. (3)).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [33] Baker, L. B., Stofan, J. R., Hamilton, A. A. & Horswill, C. A. Comparison of regional patch collection vs. whole body washdown for measuring sweat sodium and potassium loss during exercise. J. Appl. Physiol. Bethesda Md 1985 107, 887–895 (2009).
Not checked in the free run.
- Interpretationcitation
There remains an unmet medical need to measure Cl− levels in real-time for sports.
“While there remains an unmet medical need to measure Cl− levels in real-time for sports, researchers successfully developed a wrist-based potentiometric wearable device capable of detecting Cl− concentrations over time from sweat for cystic fibrosis monitoring.36”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [36] Choi, D.-H., Kim, J. S., Cutting, G. R. & Searson, P. C. Wearable potentiometric chloride sweat sensor: the critical role of the salt bridge. Anal. Chem. 88, 12241–12247 (2016). source
Not checked in the free run.
- Premisecitation
This study is the first study pertaining to the quantification of sweat sodium losses to monitor hydration in professional NFL players on a single team utilizing an epidermal patch.
“This study is significant in that it is the first study pertaining to the quantification of sweat sodium losses to monitor hydration in professional NFL players on a single team utilizing an epidermal patch.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [34] Godek, S. F. et al. Sweat rates, sweat sodium concentrations, and sodium losses in 3 groups of professional football players. J. Athl. Train. 45, 364–371 (2010). source
Not checked in the free run.
- Premisecitation
Profuse sweaters required increased dietary consumption of sodium to compensate for such losses during the preseason.
“Profuse sweaters required increased dietary consumption of sodium to compensate for such losses during the preseason.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [34] Godek, S. F. et al. Sweat rates, sweat sodium concentrations, and sodium losses in 3 groups of professional football players. J. Athl. Train. 45, 364–371 (2010). source
Not checked in the free run.
- Premisereasoning
Total whole-body sweat Na+ loss can be estimated from total sweat loss and whole-body sweat Na+ concentration.
“Total WB sweat Na+ loss can be estimated from total sweat loss and WB sweat [Na+] (Eqs (4) and (5)).”
Quote in the paperSubstantial claim - Premisecitation
Bandodkar et al. reported the successful fabrication and analytical performance of an epidermal tattoo potentiometric sodium sensor for continuous noninvasive monitoring of sodium from eccrine sweat.
“Bandodkar et al.35 reported the successful fabrication and analytical performance of an epidermal tattoo potentiometric sodium sensor for continuous noninvasive monitoring of sodium from eccrine sweat.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [35] Bandodkar, A. J. et al. Epidermal tattoo potentiometric sodium sensors with wireless signal transduction for continuous non-invasive sweat monitoring. Biosens. Bioelectron. 54, 603–609 (2014). source
Not checked in the free run.
- Premisecitation
Potassium is eliminated from the blood by a proportional regulator which may be the Na+-K+ pump of the exercising muscle.
“Potassium is eliminated from the blood by a proportional regulator which may be the Na+–K+ pump of the exercising muscle.39”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [39] Medbø, J. I. & Sejersted, O. M. Plasma potassium changes with high intensity exercise. J. Physiol. 421, 105–122 (1990). source
Not checked in the free run.
- Premisecitation
Potassium efflux rate is directly proportional with that of exercise intensity.
“Potassium efflux rate is directly proportional with that of exercise intensity.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [39] Medbø, J. I. & Sejersted, O. M. Plasma potassium changes with high intensity exercise. J. Physiol. 421, 105–122 (1990). source
Not checked in the free run.
- Premisecitation
Sweat [Na+] and [K+] were deemed stable throughout running in euhydration trials (with water intake of 150 ml per 5 min) after the initial [Na+] increase and [K+] decrease.
“Sweat [Na+] and [K+] were deemed stable throughout running in euhydration trials (with water intake of 150 ml per 5 min) after the initial [Na+] increase and [K+] decrease.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [44] Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016). source
Not checked in the free run.
- Premisecitation
An increase in sweat [Na+] and a smaller increase in sweat [K+] were observed in dehydration trials (without water intake) after 80 min when subjects had lost a large amount of water (~2.5% of body weight).
“An increase in sweat [Na+] and a smaller increase in sweat [K+] were observed in dehydration trials (without water intake) after 80 min when subjects had lost a large amount of water (~2.5% of body weight).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [44] Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016). source
Not checked in the free run.
- Premisecitation
Xu et al. devised a low-cost method to mass produce disposable portable sensors for point of care testing of K+ from blood serum.
“Xu et al.43 devised a low-cost method to mass produce disposable portable sensors for point of care testing of K+ from blood serum.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [43] Xu, H. et al. Disposable blood potassium sensors based on screen-printed thick film electrodes. Meas. Sci. Technol. 21, 055802 (2010). source
Not checked in the free run.
- Premisecitation
An increase in K+ concentration during exercise can be explained by the electrical activity in the exercising muscles.
“An increase in K+ concentration ([K+]) during exercise can be explained by the electrical activity in the exercising muscles.39”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [39] Medbø, J. I. & Sejersted, O. M. Plasma potassium changes with high intensity exercise. J. Physiol. 421, 105–122 (1990). source
Not checked in the free run.
- Premisecitation
Potassium (K+) concentrations in plasma predict muscle activity.
“Potassium (K+) concentrations in plasma predict muscle activity.39”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [39] Medbø, J. I. & Sejersted, O. M. Plasma potassium changes with high intensity exercise. J. Physiol. 421, 105–122 (1990). source
Not checked in the free run.
- Premisecitation
Gao et al. developed a flexible sweat sensor for the real-time detection of Na+, lactate, K+, glucose, and skin temperature.
“Gao et al. developed a flexible sweat sensor for the real-time detection of Na+, lactate, K+, glucose, and skin temperature.44”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [44] Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016). source
Not checked in the free run.
- Interpretationreasoning
Measurement of K+ levels could provide tremendous value to gauge and assess the workout intensity, exercise load, and physical exertion of athletes.
“Thus, measurement of K+ levels could provide tremendous value to gauge and assess the workout intensity, exercise load, and physical exertion of athletes.”
Quote in the paperSubstantial claim - Premisecitation
Sodium concentrations were monitored continuously on healthy volunteers during stationary cycling sessions using the device.
“Sodium concentrations were monitored continuously on healthy volunteers during stationary cycling sessions using the device.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [38] Matzeu, G. et al. An integrated sensing and wireless communications platform for sensing sodium in sweat. Anal. Methods 8, 64–71 (2016). source
Not checked in the free run.
- Premisecitation
Final sweat typically has a [K+] similar, albeit with a slightly broader range (~2-8 mmol/L), to that reported for blood plasma.
“Nonetheless, final sweat typically has a [K+] similar, albeit with a slightly broader range (~2–8 mmol/L), to that reported for blood plasma.29,41,42”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [29] Baker, L. B. Sweating rate and sweat sodium concentration in athletes: a review of methodology and intra/interindividual variability. Sports Med. Auckl. NZ 47, 111–128 (2017). source
Not checked in the free run.
- cites [41] Dill, D. B., Hall, F. G. & Van Beaumont, W. Sweat chloride concentration: sweat rate, metabolic rate, skin temperature, and age. J. Appl. Physiol. 21, 99–106 (1966). source
Not checked in the free run.
- cites [42] Patterson, M. J., Galloway, S. D. & Nimmo, M. A. Variations in regional sweat composition in normal human males. Exp. Physiol. 85, 869–875 (2000). source
Not checked in the free run.
- Premisecitation
Sodium and lactate levels increased with an increase in exercise intensity reflecting a rise in anaerobic metabolism.
“Sodium and lactate levels increased with an increase in exercise intensity reflecting a rise in anaerobic metabolism.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Anastasova, S. et al. A wearable multisensing patch for continuous sweat monitoring. Biosens. Bioelectron. 93, 139–145 (2017). source
Not checked in the free run.
- Premisecitation
Abellan-Llobregat's printable stretchable Pt-decorated graphite glucose sensor demonstrated a strong correlation between glucose concentration in perspiration and blood glucose measured by a commercial glucose meter.
“This device was applied on human perspiration samples and demonstrated a strong correlation between glucose concentration in perspiration and glucose concentration in blood, as measured via a commercial glucose meter.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [64] Abellán-Llobregat, A. et al. A stretchable and screen-printed electrochemical sensor for glucose determination in human perspiration. Biosens. Bioelectron. 91, 885–891 (2017). source
Not checked in the free run.
- Premisecitation
Ex situ measurements of [Na+] and [K+] from collected sweat samples demonstrated similar phenomena.
“Ex situ measurements of [Na+] and [K+] from collected sweat samples demonstrated similar phenomena.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [44] Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016). source
Not checked in the free run.
- Premisecitation
Clinical testing showed that perspiration commenced 10-15 min into the warm up period with increasing sweat rate during exercise due to thermoregulation.
“Clinical testing showed that perspiration commenced 10–15 min into the warm up period with increasing sweat rate during exercise due to thermoregulation.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Anastasova, S. et al. A wearable multisensing patch for continuous sweat monitoring. Biosens. Bioelectron. 93, 139–145 (2017). source
Not checked in the free run.
- Premisecitation
The concentration of glucose in human sweat is in the range of 10-200 μM.
“The concentration of glucose in human sweat is in the range of 10–200 µM20, and researchers have sought to assess the correlation between blood glucose and sweat glucose levels52,53 (Table 4).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [20] Bariya, M., Nyein, H. Y. Y. & Javey, A. Wearable sweat sensors. Nat. Electron. 1, 160–171 (2018). source
Not checked in the free run.
- Interpretationown data
These results suggest that cortisol could be a more sensitive marker for stress.
“These results suggest that cortisol could be a more sensitive marker for stress, as is discussed later.”
Quote in the paperSubstantial claim - Premisecitation
Flexible but nonstretchable epidermal electrodes can delaminate due to disparity between nonstretchable electrode mechanics and stretchable skin, especially during high-acuity sporting activities.
“As such, disparity between the nonstretchable mechanics of the electrode coupled with the stretchable mechanics of the skin can lead to electrode delamination especially during high-acuity sporting activities.62”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [62] Choi, J., Ghaffari, R., Baker, L. B. & Rogers, J. A. Skin-interfaced systems for sweat collection and analytics. Sci. Adv. 4, eaar3921 (2018). source
Not checked in the free run.
- Premisecitation
Gao et al.'s epidermal sensor on a PET substrate enabled concurrent and continuous detection of sodium, lactate, potassium, glucose, and skin temperature.
“As previously mentioned, Gao et al.44 fabricated an epidermal sensor on a PET substrate for the concurrent and continuous detection of sodium, lactate, potassium, glucose, and skin temperature.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [44] Gao, W. et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 529, 509–514 (2016). source
Not checked in the free run.
- Premisecitation
Real-time monitoring showed that the ECG compared to current wearable devices was not affected by concurrent lactate detection.
“Real-time monitoring showed that the ECG compared to current wearable devices was not affected by concurrent lactate detection.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [48] Imani, S. et al. A wearable chemical–electrophysiological hybrid biosensing system for real-time health and fitness monitoring. Nat. Commun. 7, 11650 (2016). source
Not checked in the free run.
- Premisecitation
Lactate levels measured by the biosensor closely approximated the expected sweat-lactate profile for increasing intensity workouts.
“In addition, lactate levels measured by the biosensor closely approximated the expected sweat-lactate profile for increasing intensity workouts.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [48] Imani, S. et al. A wearable chemical–electrophysiological hybrid biosensing system for real-time health and fitness monitoring. Nat. Commun. 7, 11650 (2016). source
Not checked in the free run.
- Premisecitation
Hauke et al. developed and validated a continuous and blood-correlated sweat enzymatic sensor with integrated sweat stimulation to detect ethanol.
“Hauke et al.23 developed and validated a continuous and blood-correlated sweat enzymatic sensor with integrated sweat stimulation to detect ethanol (Fig. 2i).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Hauke, A. et al. Complete validation of a continuous and blood-correlated sweat biosensing device with integrated sweat stimulation. Lab. Chip https://doi.org/10.1039/C8LC01082J (2018). source
Not checked in the free run.
- Rests on an assumption: Ethanol is 1:1 between sweat and blood due to its lipophilic nature.
- Premisecitation
On-body real-time monitoring with Martín et al.'s device showed an increase in amperometric sweat lactate current signal as the sweat sample entered and filled the detector reservoir.
“The continuous monitoring of the amperometric sweat lactate response from the subjects with the LOX-modified flow detector demonstrated an increase of the current signal as the sweat sample entered and filled the detector reservoir.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [67] Martín, A. et al. Epidermal microfluidic electrochemical detection system: enhanced sweat sampling and metabolite detection. ACS Sens. 2, 1860–1868 (2017). source
Not checked in the free run.
- Premiseown data
The same trend was seen when measuring glucose levels using GOX-modified electrodes.
“The same trend was seen when measuring glucose levels using GOX-modified electrodes.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [67] Martín, A. et al. Epidermal microfluidic electrochemical detection system: enhanced sweat sampling and metabolite detection. ACS Sens. 2, 1860–1868 (2017). source
Not checked in the free run.
- Premisecitation
In human subjects during eating, the tattoo-based glucose sensor correlated with a commercial glucose meter.
“Results showed that the sensor correlated with that of a commercial glucose meter.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [65] Bandodkar, A. J. et al. Tattoo-based noninvasive glucose monitoring: a proof-of-concept study. Anal. Chem. 87, 394–398 (2015). source
Not checked in the free run.
- Premisecitation
Koh et al.'s closed microfluidic system proved efficacious when tested on humans during cycling.
“The device proved to be efficacious when tested on humans during cycling.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [55] Koh, A. et al. A soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweat. Sci. Transl. Med. 8, 366ra165 (2016). source
Not checked in the free run.
- Premisecitation
Reeder et al.'s waterproof epidermal microfluidic device adhered to the swimmer's body for greater than two hours under extreme conditions.
“The robust and water-tight bonding to the skin, under extreme conditions, enabled the device to adhere to the swimmer’s body for greater than two hours.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [69] Reeder, J. T. et al. Waterproof, electronics-enabled, epidermal microfluidic devices for sweat collection, biomarker analysis, and thermography in aquatic settings. Sci. Adv. 5, eaau6356 (2019). source
Not checked in the free run.
- Premisecitation
The trend in measuring glucose levels from eccrine sweat matched that of blood glucose values.
“The trend in measuring glucose levels from eccrine sweat matched that of blood glucose values; however, the need to prevent sample contamination remains the next step for long-term non-invasive glucose measurements utilizing such technology.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [67] Martín, A. et al. Epidermal microfluidic electrochemical detection system: enhanced sweat sampling and metabolite detection. ACS Sens. 2, 1860–1868 (2017). source
Not checked in the free run.
- Premisecitation
Bariya et al.'s roll-to-roll gravure printed electrochemical electrodes on flexible PET achieved devices with uniform redox kinetics for detection of pH, K+, Na+, Cu2+, glucose, and caffeine.
“Bariya et al.66 developed roll-to-roll (R2R) gravure printed electrochemical electrodes on 150 m flexible PET substrate rolls for the detection of pH, K+, Na+, Cu2+, glucose, and caffeine (Fig. 2a, b). The team utilized inks and electrode morphologies designed for electrochemical and mechanical stability to achieve devices with uniform redox kinetics.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [66] Bariya, M. et al. Roll-to-roll gravure printed electrochemical sensors for wearable and medical devices. ACS Nano 12, 6978–6987 (2018). source
Not checked in the free run.
- Premisecitation
There are three primary methods to monitor stress levels for athletes: self-reporting, multimodal physiological analysis, and body-fluid analysis.
“There are three primary methods to monitor stress levels for athletes: (1) self-reporting (current method), (2) multimodal physiological analysis, and (3) body-fluid analysis74 (Table 6).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [74] Yoon, S., Sim, J. K. & Cho, Y.-H. A flexible and wearable human stress monitoring patch. Sci. Rep. 6, 1–11 (2016). source
Not checked in the free run.
- Premisecitation
The training status of athletes may affect HRV.
“The training status of athletes may affect HRV.83,84”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [83] Pichot, V. et al. Relation between heart rate variability and training load in middle-distance runners. Med. Sci. Sports Exerc. 32, 1729–1736 (2000). source
Not checked in the free run.
- cites [84] Uusitalo, A. L., Uusitalo, A. J. & Rusko, H. K. Heart rate and blood pressure variability during heavy training and overtraining in the female athlete. Int. J. Sports Med. 21, 45–53 (2000). source
Not checked in the free run.
- Premisecitation
During physical activity, the parasympathetic activity is first withdrawn and then sympathetic nerve activity is augmented to meet the metabolic demands of the behavior.
“During physical activity, the parasympathetic activity is first withdrawn and then sympathetic nerve activity is augmented to meet the metabolic demands of the behavior.81”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [81] Mitchell, J. H. Neural circulatory control during exercise: early insights. Exp. Physiol. 98, 867–878 (2013). source
Not checked in the free run.
- Premisecitation
Continuous data from Hauke et al.'s sensor in two human subjects allowed analysis of blood-to-sweat lag times ranging between 2.3 and 11.41 min for the onset of the ethanol signal.
“The continuous nature of the data when tested on two human subjects allowed for analysis of blood-to-sweat lag times that ranged between 2.3 and 11.41 min for the onset of the ethanol signal.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Hauke, A. et al. Complete validation of a continuous and blood-correlated sweat biosensing device with integrated sweat stimulation. Lab. Chip https://doi.org/10.1039/C8LC01082J (2018). source
Not checked in the free run.
- Rests on an assumption: Ethanol is 1:1 between sweat and blood due to its lipophilic nature.
- Premisecitation
The magnitude of the neuroendocrine response reflects the metabolic and physiological demands required for a given activity.
“The magnitude of the neuroendocrine response reflects the metabolic and physiological demands required for a given activity.77”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [77] Koolhaas, J. M. et al. Stress revisited: a critical evaluation of the stress concept. Neurosci. Biobehav. Rev. 35, 1291–1301 (2011). source
Not checked in the free run.
- Premisecitation
Heterogeneity in the QRS complex often poses challenges to identify the RR interval.
“Secondly, heterogeneity in the QRS complex often poses challenges to identify the RR interval.89”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [89] Gambarotta, N., Aletti, F., Baselli, G. & Ferrario, M. A review of methods for the signal quality assessment to improve reliability of heart rate and blood pressures derived parameters. Med. Biol. Eng. Comput. 54, 1025–1035 (2016). source
Not checked in the free run.
- Premisecitation
A negative correlation exists between peripheral skin temperature and chronic stress levels.
“A negative correlation exists between peripheral skin temperature and chronic stress levels,96,97”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [96] Lee, Y., Lee, B. & Lee, M. Wearable sensor glove based on conducting fabric using electrodermal activity and pulse-wave sensors for e-health application. Telemed. J. E Health 16, 209–217 (2010). source
Not checked in the free run.
- cites [97] Torii, M., Yamasaki, M., Sasaki, T. & Nakayama, H. Fall in skin temperature of exercising man. Br. J. Sports Med. 26, 29–32 (1992). source
Not checked in the free run.
- Premisecitation
A recent systematic review found that the correlation between ECG-derived HRV and wearable RV was validated during rest but declined as exercise intensity was increased.
“The correlation between ECG-derived HRV and the wearable RV was validated during rest but declined as exercise intensity was increased.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [94] Georgiou, K. et al. Can wearable devices accurately measure heart rate variability? A systematic review. Folia Med. 60, 7–20 (2018).
Not checked in the free run.
- Premisecitation
A reported drawback in most ECG-based wearable devices that do not record standard ECG derivations is their inability to distinguish some arrhythmias and ectopic beats.
“Lastly, a reported drawback in most ECG-based wearable devices that do not record standard ECG derivations is their inability to distinguish some arrhythmias and ectopic beats.90”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [90] Wallén, M. B., Hasson, D., Theorell, T., Canlon, B. & Osika, W. Possibilities and limitations of the Polar RS800 in measuring heart rate variability at rest. Eur. J. Appl. Physiol. 112, 1153–1165 (2012). source
Not checked in the free run.
- Premisecitation
Quantification of ANS function is possible by calculating parameters pertinent to HRV according to time-domain, frequency-domain, and nonlinear analysis of consecutive RR intervals of an ECG waveform.
“The quantification of ANS function is possible by calculating parameters pertinent to HRV according to time-domain, frequency-domain, and nonlinear analysis of consecutive RR intervals of an ECG waveform.86”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [86] Liu, A.-B., Wu, H.-T., Liu, C.-C., Hsu, C.-H. & Chen, D.-Y. The factors influence compatibility of pulse-pulse intervals with R-R intervals. Conf. Proc. IEEE Eng. Med. Biol. Soc. 2013, 2068–2071 (2013).
Not checked in the free run.
- Interpretationreasoning
HRV is context dependent so the overall environment the athlete is in must be factored in prior to monitoring and subsequent diagnosis.
“Thus, HRV is context dependent so the overall environment the athlete is in must be factored in prior to monitoring and subsequent diagnosis.”
Quote in the paperSubstantial claim - Premisecitation
A positive correlation exists between skin conductance on the palm and volar wrist with chronic and acute stress levels.
“while a positive correlation exists between skin conductance on the palm and volar wrist with chronic and acute stress levels.95”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [95] Kreibig, S. D., Wilhelm, F. H., Roth, W. T. & Gross, J. J. Cardiovascular, electrodermal, and respiratory response patterns to fear- and sadness-inducing films. Psychophysiology 44, 787–806 (2007). source
Not checked in the free run.
- Premisecitation
Prior work to measure stress-related biomarkers has utilized high-performance liquid chromatography (HPLC), enzyme-linked immunosorbent assay, radioimmunoassay kit, or a HPLC mass spectrometry (MS) system.
“Prior work to measure stress-related biomarkers has utilized high-performance liquid chromatography (HPLC),99 enzyme-linked immunosorbent assay,100 radioimmunoassay kit,101 or a HPLC mass spectrometry (MS) system.102”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [99] Arihara, Z. et al. Immunoreactive orexin-A in human plasma. Peptides 22, 139–142 (2001). source
Not checked in the free run.
- cites [100] Heinonen, M. V. et al. Apelin, orexin-A and leptin plasma levels in morbid obesity and effect of gastric banding. Regul. Pept. 130, 7–13 (2005). source
Not checked in the free run.
- cites [101] Abdo, W. F. et al. CSF hypocretin-1 levels are normal in multiple-system atrophy. Park. Relat. Disord. 14, 342–344 (2008). source
Not checked in the free run.
- cites [102] Jia, M., Chew, W. M., Feinstein, Y., Skeath, P. & Sternberg, E. M. Quantification of cortisol in human eccrine sweat by liquid chromatography–tandem mass spectrometry. Analyst 141, 2053–2060 (2016). source
Not checked in the free run.
- Premisecitation
The stress patch categorized four types of human emotions (surprise, anger, stress, and sadness) based on a singular vector machine algorithm.
“The patch categorized the four types of human emotions (surprise, anger, stress, and sadness) based on a singular vector machine algorithm.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [74] Yoon, S., Sim, J. K. & Cho, Y.-H. A flexible and wearable human stress monitoring patch. Sci. Rep. 6, 1–11 (2016). source
Not checked in the free run.
- Interpretationreasoning
Existing methods for measuring stress-related biomarkers are limited for use in sports applications due to the significant cost and technical expertize required.
“These existing methods are limited for use in sports applications due to the significant cost and technical expertize required.”
Quote in the paperSubstantial claim - Premisecitation
Jia et al.'s study detected one isomer in eccrine sweat that had a similar hydrophobicity, retention time, and fragmentation patterns to that of cortisol.
“The study detected one isomer that had a similar hydrophobicity, retention time, and fragmentation patterns to that of cortisol found in eccrine sweat.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [102] Jia, M., Chew, W. M., Feinstein, Y., Skeath, P. & Sternberg, E. M. Quantification of cortisol in human eccrine sweat by liquid chromatography–tandem mass spectrometry. Analyst 141, 2053–2060 (2016). source
Not checked in the free run.
- Premisecitation
Prior studies have shown that the levels of certain molecular markers in eccrine sweat are comparable to those found in human plasma.
“Prior studies have shown that the levels of certain molecular markers in eccrine sweat are comparable to those found in human plasma.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [112] Marques-Deak, A. et al. Measurement of cytokines in sweat patches and plasma in healthy women: Validation in a controlled study. J. Immunol. Methods 315, 99–109 (2006). source
Not checked in the free run.
- Premisecitation
EEG has been utilized as a noninvasive means to measure stress levels by placing electrodes on the human scalp which measure oscillations of the brain's electric potential.
“In addition, EEG has been utilized as a noninvasive means to measure stress levels by placing electrodes on the human scalp which measure oscillations of the brain’s electric potential.103,104”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [103] Peng, H. et al. A method of identifying chronic stress by EEG. Pers. Ubiquitous Comput. 17, 1341–1347 (2013). source
Not checked in the free run.
- cites [104] Vanitha, V. & Krishnan, P. Real time stress detection system based on EEG signals. Biomed. Res. S271–S275 (2016).
Not checked in the free run.
- Premisecitation
Marques-Deak et al. compared baseline levels of cytokines in plasma and eccrine sweat and showed that the measured cytokine levels were comparable to that of circulating levels in plasma.
“Marques-Deak et al.112 compared baseline levels of cytokines, such as IL-1α, IL-1β, IL-6, TNF-α, IL-8, and TGF-β in plasma and eccrine sweat, and showed that the measured cytokine levels were comparable to that of circulating levels in plasma.112”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [112] Marques-Deak, A. et al. Measurement of cytokines in sweat patches and plasma in healthy women: Validation in a controlled study. J. Immunol. Methods 315, 99–109 (2006). source
Not checked in the free run.
- Premisecitation
The OPTIMI project recorded EEG and ECG activity, voice analysis, electronic diaries, and cortisol sampling to monitor an individual's mental state.
“Collaborative research undertakings, such as the Online Predictive Tools for Intervention in Mental Illness (OPTIMI), have recorded EEG and ECG activity, voice analysis, electronic diaries, and cortisol sampling to monitor an individual’s mental state.105”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [105] OPTIMI | Online Predictive Tools for Intervention in Mental Illness.
Not checked in the free run.
- Premisecitation
Ex situ testing with artificial sweat and real-time testing during exercise demonstrated the utility of the device to measure cortisol.
“Ex situ testing, performed by spraying artificial sweat with increased cortisol concentrations on the forearm, and real-time testing during exercise, demonstrated the utility of the device to measure cortisol (Fig. 3e–g).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [111] Parlak, O., Keene, S. T., Marais, A., Curto, V. F. & Salleo, A. Molecularly selective nanoporous membrane-based wearable organic electrochemical device for noninvasive cortisol sensing. Sci. Adv. 4, eaar2904 (2018). source
Not checked in the free run.
- Premisecitation
A multifunctional layered wearable organic electrochemical sensor was developed for the non-invasive detection of cortisol from eccrine sweat.
“et al. developed a multifunctional layered wearable organic electrochemical sensor for the non-invasive detection of cortisol from eccrine sweat (Fig. 3e–g).111”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [111] Parlak, O., Keene, S. T., Marais, A., Curto, V. F. & Salleo, A. Molecularly selective nanoporous membrane-based wearable organic electrochemical device for noninvasive cortisol sensing. Sci. Adv. 4, eaar2904 (2018). source
Not checked in the free run.
- Premisecitation
Transcranial direct current stimulation at low currents (e.g., <2 mA) applied over the scalp induced changes in brain excitability for an extended duration to result in synaptic and nonsynaptic functional changes.
“Research showed that transcranial direct current stimulation at low currents (e.g., <2 mA) applied over the scalp induced changes in brain excitability for an extended duration to result in synaptic and nonsynaptic functional changes.121,122”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [121] Reis, J. et al. Noninvasive cortical stimulation enhances motor skill acquisition over multiple days through an effect on consolidation. Proc. Natl Acad. Sci. USA 106, 1590–1595 (2009). source
Not checked in the free run.
- cites [122] Cogiamanian, F., Marceglia, S., Ardolino, G., Barbieri, S. & Priori, A. Improved isometric force endurance after transcranial direct current stimulation over the human motor cortical areas. Eur. J. Neurosci. 26, 242–249 (2007). source
Not checked in the free run.
- Premisecitation
Measuring serum levels of Orexin A can aid in predicting mood and cognitive performance in athletes.
“Measuring serum levels of Orexin A can aid in predicting mood and cognitive performance in athletes.113”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [113] Hagen, J. et al. Detection of orexin A neuropeptide in biological fluids using a zinc oxide field effect transistor. ACS Chem. Neurosci. 4, 444–453 (2013). source
Not checked in the free run.
- Premisecitation
The binding was transmitted to an electrical signal.
“The binding was transmitted to an electrical signal”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [113] Hagen, J. et al. Detection of orexin A neuropeptide in biological fluids using a zinc oxide field effect transistor. ACS Chem. Neurosci. 4, 444–453 (2013). source
Not checked in the free run.
- Premisecitation
Hagen et al. utilized interdigitated zinc oxide field effect transistors (ZnO FETs) to detect Orexin A by binding a bifunctional peptide to both the ZnO semiconductor and the neuropeptide.
“Hagen et al. utilized interdigitated zinc oxide field effect transistors (ZnO FETs) to detect Orexin A by binding a bifunctional peptide to both the ZnO semiconductor and the neuropeptide.113”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [113] Hagen, J. et al. Detection of orexin A neuropeptide in biological fluids using a zinc oxide field effect transistor. ACS Chem. Neurosci. 4, 444–453 (2013). source
Not checked in the free run.
- Premisecitation
The sensor platform demonstrated the potential of an FET device to measure a wide array of biomarkers in complex biofluids.
“The sensor platform demonstrated the potential of an FET device to measure a wide array of biomarkers in complex biofluids; however, development of improved sensitivity and stability of biosensors is required for the real-time detection of such biomarkers.113”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [113] Hagen, J. et al. Detection of orexin A neuropeptide in biological fluids using a zinc oxide field effect transistor. ACS Chem. Neurosci. 4, 444–453 (2013). source
Not checked in the free run.
- Premisecitation
Devices such as those by Halo Neuroscience can potentially improve cognition during athletic training sessions by utilizing neuropriming to increase the excitability of motor neurons.
“In addition, devices such as those by Halo Neuroscience can potentially improve cognition during athletic training sessions by utilizing neuropriming to increase the excitability of motor neurons.120”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [120] Halo Neuroscience. https://www.haloneuro.com. Accessed 18 May 2017.
Not checked in the free run.
- Premisecitation
Noninvasive measurement of several neuropeptides utilizing functionalized antibodies or peptide-decorated semiconductors has been reported.
“Noninvasive measurement of several neuropeptides utilizing functionalized antibodies or peptide-decorated semiconductors has been reported.117,118”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [117] North, J. R. Immunosensors: antibody-based biosensors. Trends Biotechnol. 3, 180–186 (1985). source
Not checked in the free run.
- cites [118] Kuang, Z., Kim, S. N., Crookes-Goodson, W. J., Farmer, B. L. & Naik, R. R. Biomimetic chemosensor: designing peptide recognition elements for surface functionalization of carbon nanotube field effect transistors. ACS Nano 4, 452–458 (2010). source
Not checked in the free run.
- Premisecitation
The sensor combined with a microcapillary channel array provided precise sample delivery to the sensor interface.
“The sensor, combined with a microcapillary channel array, was integrated in the sensor thereby providing precise sample delivery to the sensor interface.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [111] Parlak, O., Keene, S. T., Marais, A., Curto, V. F. & Salleo, A. Molecularly selective nanoporous membrane-based wearable organic electrochemical device for noninvasive cortisol sensing. Sci. Adv. 4, eaar2904 (2018). source
Not checked in the free run.
- Premisecitation
Clinical trials by Halo Neuroscience have shown that this effect can significantly improve player performance and training.
“Clinical trials by Halo Neuroscience have shown that this effect can significantly improve player performance and training.120”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [120] Halo Neuroscience. https://www.haloneuro.com. Accessed 18 May 2017.
Not checked in the free run.
- Classified as probably not a substantial claim. 36
- May combine two claims. 42
Probably not a substantial claim: may not pass S4 (based).
Interpretationnone givenThe ability to monitor biomarkers from saliva or sweat in a noninvasive and continuous manner remains the next technological gap for sports medical personnel to tailor hydration and recovery protocols per the athlete.
“The ability to monitor biomarkers from saliva or sweat in a noninvasive and continuous manner remain the next technological gap for sports medical personnel to tailor hydration and recovery protocols per the athlete.”
Quote in the paperNot a substantial claim- Rests on an assumption: Noninvasive continuous monitoring of biomarkers from saliva or sweat would allow sports medical personnel to tailor hydration and recovery protocols per athlete.
Probably not a substantial claim: may not pass S4 (based).
Premisenone givenThe majority of existing wearable devices utilized in the sporting community currently focus on monitoring physical or physiological parameters (e.g., motion, HR, respiration rate, RR).
“The majority of existing wearable devices utilized in the sporting community currently focus on monitoring physical or physiological parameters (e.g., motion, HR, respiration rate, RR).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
Interpretationnone givenThe use of wearable sensors for sports is at its infancy, with the majority of devices currently used to measure movement-based parameters such as distance, velocity, and acceleration.
“The use of wearable sensors for sports is at its infancy, with the majority of devices currently used to measure movement-based parameters such as distance, velocity, and acceleration.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenOne issue plaguing the wearables field is the translation of the data to create actionable insight in its respective clinical domain.
“While beyond the scope of this review, one issue plaguing the wearables field is the translation of the data to create actionable insight in its respective clinical domain.”
Quote in the paperNot a substantial claim- Rests on an assumption: Wearable sensor data can be translated into actionable insight to improve athlete health and performance.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationProper hydration is key to success as under-drinking can lead to hypohydration and over-drinking can lead to hyponatremia (low-serum sodium concentration).
“Proper hydration is key to success as under-drinking can lead to hypohydration and over-drinking can lead to hyponatremia (low-serum sodium concentration).10”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [10] Hew-Butler, T. et al. Statement of the Third International Exercise-Associated Hyponatremia Consensus Development Conference, Carlsbad, California, 2015. Clin. J. Sport Med. 25, 303 (2015). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe development of electrochemical transducers has been especially promising due to their low cost, simplicity, and portability.
“The development of electrochemical transducers has been especially promising due to their low cost, simplicity, and portability.7,8”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [7] Barry, R. C., Lin, Y., Wang, J., Liu, G. & Timchalk, C. A. Nanotechnology-based electrochemical sensors for biomonitoring chemical exposures. J. Expo. Sci. Environ. Epidemiol. 19, 1–18 (2009). source
Not checked in the free run.
- cites [8] Couto, R. A. S. & Quinaz, M. B. Development of a Nafion/MWCNT-SPCE-based portable sensor for the voltammetric analysis of the anti-tuberculosis drug ethambutol. Sensors 16, 1–13 (2016). https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4970065/pdf/sensors-16-01015.pdf. source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S4 (based).
Interpretationnone givenThere remains an unmet medical need by the sports community to gain further insight into the internal workload of the athlete to tailor recovery protocols to each athlete.
“However, there remains an unmet medical need by the sports community to gain further insight into the internal workload of the athlete to tailor recovery protocols to each athlete.”
Quote in the paperNot a substantial claim- Rests on an assumption: Noninvasive continuous monitoring of biomarkers from saliva or sweat would allow sports medical personnel to tailor hydration and recovery protocols per athlete.
Probably not a substantial claim: may not pass S2 (checkable), S4 (based).
Interpretationnone givenThere remains a major need to quantify the athlete by measuring biochemical markers such as electrolytes, analytes, and neuropeptides all of which are indicative of physical exertion and fitness, fatigue, and mental acuity.
“There remains a major need to “quantify the athlete” by measuring biochemical markers such as electrolytes, analytes, and neuropeptides all of which are indicative of physical exertion and fitness, fatigue, and mental acuity.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningThese commercial hand-held sensors require blood samples, posing a barrier to their utility in real-time monitoring for sports medicine.
“However, these sensors require blood samples thus posing as a barrier to their utility in real-time monitoring for sports medicine.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable), S4 (based).
InterpretationreasoningThe emergence of wearable biosensors to measure analytes from eccrine sweat to assess the performance and mental acuity of the athlete serve as next steps to assessing human performance.
“The emergence of wearable biosensors to measure analytes from eccrine sweat to assess the performance and mental acuity of the athlete serve as next steps to assessing human performance.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenThe emergence of flexible and stretchable electronics coupled with the ability to quantify biochemical analytes and physiological parameters has enabled the detection of key markers indicative of performance and stress.
“The emergence of flexible and stretchable electronics coupled with the ability to quantify biochemical analytes and physiological parameters have enabled the detection of key markers indicative of performance and stress, as reviewed in this paper.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Premisenone givenRecent advancements in wearable sensor technology from a device to systems standpoint have provided new avenues to change this paradigm and are currently being implemented by teams worldwide.
“Recent advancements in wearable sensor technology from a device to systems standpoint have provided new avenues to change this paradigm and are currently being implemented by teams worldwide.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Interpretationnone givenReal-time testing of the presented devices is needed to assess the true clinical efficacy of such technology for sports.
“Real-time testing of the devices presented are needed to assess the true clinical efficacy of such technology for sports.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
InterpretationreasoningThere remains a clinical need to validate the sensitivity and reliability of mouthguard biosensors in real-time during athletic scenarios.
“While the application of mouthguards to quantify biomarkers is relatively new from a commercial standpoint, there remains a clinical need to validate the sensitivity and reliability of these devices in real-time during athletic scenarios.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationInitial work pertaining to electrochemical salivary sensors was conducted by Graf in the 1960s to measure pH and fluoride ion levels on a partial denture.
“Initial work pertaining to electrochemical salivary sensors was conducted by Graf in the 1960s to measure pH and fluoride ion levels on a partial denture.13”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [13] Graf, H. & Mühlemann, H. R. Telemetry of plaque pH from interdental area. Helv. Odontol. Acta 10, 94–101 (1966).
Not checked in the free run.
May combine two claims.
Premiseown dataThe chronoamperometric response of lactate in the presence and absence of physiological concentrations of ascorbic acid and UA showed that these potential interferences had a negligible effect upon the lactate response and that the biosensor system provided high selectivity to measure lactate in a noninvasive manner.
“The chronoamperometric response of lactate in the presence and absence of physiological concentrations of ascorbic acid and UA showed that these potential interferences had a negligible effect upon the lactate response and that the biosensor system provided high selectivity to measure lactate in a noninvasive manner.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [14] Kim, J. et al. Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites. Analyst 139, 1632 (2014). source
Not checked in the free run.
May combine two claims.
Premiseown dataThe mouthguard lactate biosensor displayed a very high sensitivity toward lactate, with current-signals proportional to the lactate concentration.
“Data from chronoamperograms for increasing concentrations of lactate in phosphate-buffered saline medium showed that the biosensor displayed a very high sensitivity toward lactate, with current-signals proportional to the lactate concentration.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [14] Kim, J. et al. Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites. Analyst 139, 1632 (2014). source
Not checked in the free run.
May combine two claims.
PremisecitationDisadvantages of using eccrine sweat include skin contamination, existence of dried sweat on the glands thereby skewing analyte measurements, and low-sampling rates.
“Disadvantages of using eccrine sweat, include skin contamination, existence of dried sweat on the glands thereby skewing analyte measurements, and low-sampling rates.18”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [18] Heikenfeld, J. Non-invasive analyte access and sensing through eccrine sweat: challenges and outlook circa 2016. Electroanalysis 28, 1242–1249 (2016). source
Not checked in the free run.
May combine two claims.
PremisecitationDetection of biomarkers from eccrine sweat glands poses tremendous advantages over those from urine, blood, tears, and apocrine sweat glands, including their abundance on the body (>100 glands/cm2), ease of access, sampling and detection efficiency without foreign contamination during testing, and the inability to degrade analytes.
“Detection of biomarkers from eccrine sweat glands pose tremendous advantages over those from urine, blood, tears, and apocrine sweat glands. These, include their abundance on the body (>100 glands/cm2), ease of access, sampling and detection efficiency without foreign contamination during testing, and the inability to degrade analytes19 (Table 3).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [19] Sonner, Z. et al. The microfluidics of the eccrine sweat gland, including biomarker partitioning, transport, and biosensing implications. Biomicrofluidics 9, 031301 (2015). source
Not checked in the free run.
May combine two claims.
PremisecitationThe Halo H1 sensor tracks hydration levels at a cellular level in the bloodstream and utilizes an algorithm to rank the levels out of 100; green (68-100), fully hydrated, yellow (35-67), caution to hydrate soon, and red (1-34), need to hydrate immediately.
“The sensor tracks hydration levels at a cellular level in the bloodstream and utilizes an algorithm to rank the levels out of 100; green (68–100), fully hydrated, yellow (35–67), caution to hydrate soon, and red (1–34), need to hydrate immediately.25”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [25] Miller, D. W. & Miller, D. R. Hydration Monitoring Apparatus. 1–15 (2014). https://patentimages.storage.googleapis.com/89/6c/7a/34703870dde191/US20140221792A1.pdf.
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationEpicore Biosystems has partnered with the Gatorade Sports Science Institute, Seattle Mariners (MLB), US Air Force (USAF), and the Shirley Ryan AbilityLab to further test and validate their devices.
“The company has partnered with the Gatorade Sports Science Institute, Seattle Mariners (MLB), US Air Force (USAF), and the Shirley Ryan AbilityLab to further test and validate their devices.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [24] Northwestern Engineering Teams with Sports and Rehabilitation Leaders to Advance Sweat Analysis Technologies. Northwestern Eng. http://www.mccormick.northwestern.edu/news/articles/2018/02/northwestern-engineering-teams-with-sports-and-rehabilitation-leaders-to-advance-sweat-analysis-technology.html. Accessed 16 Sep 2018.
Not checked in the free run.
May combine two claims.
PremisecitationKenzen Wear's Echo patch is an epidermal sensor, adhered to the torso, which monitors sodium and potassium from sweat in addition to measuring pH and skin temperature.
“Similarly, Kenzen Wear’s Echo patch is an epidermal sensor, adhered to the torso, which monitors sodium and potassium from sweat in addition to measuring pH and skin temperature.28”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [28] ECHO Smart Patch. Kenzen https://www.kenzen.com/. Accessed 5 May 2017.
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationcitationThere remains a need to ensure that the administered current intensity does not cause cathodal vasodilation or erythema, leading to discomfort to the athlete.
“However, there remains a need to ensure that the administered current intensity does not cause cathodal vasodilation or erythema, leading to discomfort to the athlete.22”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22] Roustit, M., Blaise, S. & Cracowski, J.-L. Trials and tribulations of skin iontophoresis in therapeutics. Br. J. Clin. Pharmacol. 77, 63–71 (2014). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based). May combine two claims.
Interpretationnone givenThe need to minimize body motion artifacts is crucial for sports to prevent false sensor readings or cause the sweat flow to accelerate or even reverse.
“The need to minimize body motion artifacts are crucial for sports to prevent false sensor readings or cause the sweat flow to accelerate or even reverse.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe wrist-based Cl− sensor was placed on human subjects by a wristband or adhesive tape and tested during exercise to demonstrate the feasibility of this technology as a wearable device.
“The sensor was placed on human subjects by a wristband or adhesive tape and tested during exercise to demonstrate the feasibility of this technology as a wearable device.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [36] Choi, D.-H., Kim, J. S., Cutting, G. R. & Searson, P. C. Wearable potentiometric chloride sweat sensor: the critical role of the salt bridge. Anal. Chem. 88, 12241–12247 (2016). source
Not checked in the free run.
May combine two claims.
PremisecitationThe study showed large variations in sweat Na+ concentration between backs and receivers and linebackers and quarterbacks but not between linebackers and quarterbacks and linemen.
“The study showed large variations in sweat Na+ concentration between BKs and LB/QB but not between LB/QB and LM.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [34] Godek, S. F. et al. Sweat rates, sweat sodium concentrations, and sodium losses in 3 groups of professional football players. J. Athl. Train. 45, 364–371 (2010). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenMonitoring the change in Cl− concentration in a noninvasive manner along with or independent of sodium measurements utilizing wearables can expedite treatment and recovery to mitigate soft-tissue injuries.
“Monitoring the change in Cl− concentration in a noninvasive manner along with or independent of sodium measurements utilizing wearables can expedite treatment and recovery to mitigate soft-tissue injuries.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe screen printed epidermal tattoo sodium sensor withstood mechanical deformation without impeding analyte detection and wireless transmission, highlighting its translational potential for the sporting community.
“The screen printed device withstood mechanical deformation without impeding analyte detection and wireless transmission thereby highlighting its translational potential for the sporting community.35”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [35] Bandodkar, A. J. et al. Epidermal tattoo potentiometric sodium sensors with wireless signal transduction for continuous non-invasive sweat monitoring. Biosens. Bioelectron. 54, 603–609 (2014). source
Not checked in the free run.
May combine two claims.
Premisenone givenTotal whole-body sweat loss and whole-body sweating rate can be estimated.
“Total whole-body sweat loss and WB sweating rate can be estimated.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationExtracellular K+ is indirectly linked to the pump stimulus and the rate of reuptake is proportional to that of extracellular accumulation.
“Extracellular K+ is indirectly linked to the pump stimulus and the rate of reuptake is proportional to that of extracellular accumulation.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [39] Medbø, J. I. & Sejersted, O. M. Plasma potassium changes with high intensity exercise. J. Physiol. 421, 105–122 (1990). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenWhile such core-sensor technology is promising, it remains impractical for sports medicine.
“However, while such core-sensor technology is promising, it remains impractical for sports medicine.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThe rate of sweat K+ loss has been reported to be indirectly related to sweat flow rate, but the underlying mechanism is unclear and requires further investigation.
“The rate of sweat K+ loss has been reported to be indirectly related to sweat flow rate, but the underlying mechanism is unclear and requires further investigation.40”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [40] Sato, K. The physiology, pharmacology, and biochemistry of the eccrine sweat gland. Rev. Physiol. Biochem. Pharmacol. 79, 51–131 (1977). source
Not checked in the free run.
May combine two claims.
PremisecitationThe K+ sensor demonstrated a smaller and faster response compared to current standard of care (e.g., clinical laboratory electrolyte analyzer) and required a smaller blood sample volume.
“The sensor demonstrated a smaller and faster response compared to current standard of care (e.g., clinical laboratory electrolyte analyzer) and required a smaller blood sample volume.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [43] Xu, H. et al. Disposable blood potassium sensors based on screen-printed thick film electrodes. Meas. Sci. Technol. 21, 055802 (2010). source
Not checked in the free run.
May combine two claims.
InterpretationreasoningThe use of such adhesion platforms is not appropriate for long-term use in sports, but the electrochemical performance and stability of the device demonstrates promise for athletics.
“While the use of such adhesion platforms is not appropriate for long-term use in sports, the electrochemical performance and stability of the device demonstrates promise for athletics.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationA luminol-based biosensor for lactate detection achieved a detection limit of 8.9 × 10−12 mol/L and an average recovery of 101.3% using athlete sweat samples during a training course.
“A detection limit of 8.9 × 10−12 mol/L and an average recovery of 101.3% was obtained when utilizing athlete sweat samples during a training course.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [49] Cai, X., Yan, J., Chu, H., Wu, M. & Tu, Y. An exercise degree monitoring biosensor based on electrochemiluminescent detection of lactate in sweat. Sens. Actuators B Chem. 143, 655–659 (2010). source
Not checked in the free run.
May combine two claims.
PremisecitationSodium and lactate levels from saliva demonstrated the same correlation as noted with sweat; salivary cortisol levels exhibited the largest variation among the subjects.
“Sodium and lactate levels demonstrated the same correlation as noted with sweat; however, salivary cortisol levels exhibited the largest variation among the subjects.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [50] Anastasova, S. et al. A wearable multisensing patch for continuous sweat monitoring. Biosens. Bioelectron. 93, 139–145 (2017). source
Not checked in the free run.
May combine two claims.
PremisecitationPlasma lactate concentrations closely approximate those of sweat lactate and provide an indication of body exertion and exercise intensity.
“Plasma lactate concentrations closely approximate those of sweat lactate and provide an indication of body exertion and exercise intensity.47”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [47] Jia, W. et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85, 6553–6560 (2013). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationTemperature readings of the sensor varied between 20 and 40 °C across subjects.
“Temperature readings of the sensor varied between 20 and 40 °C across subjects.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [50] Anastasova, S. et al. A wearable multisensing patch for continuous sweat monitoring. Biosens. Bioelectron. 93, 139–145 (2017). source
Not checked in the free run.
May combine two claims.
PremisecitationLa Count et al.'s glucose model demonstrated a ten-minute blood-to-sweat lag time and a sweat/blood glucose level ranging from 0.001 to 0.02, depending on the sweat rate.
“The glucose model, calibrated under a variety of experimental conditions including electrical enhancement, demonstrated a ten-minute blood-to-sweat lag time and a sweat/blood glucose level ranging from 0.001 to 0.02, depending on the sweat rate.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [54] Count, T. D. L., Jajack, A., Heikenfeld, J. & Kasting, G. B. Modeling glucose transport from systemic circulation to sweat. J. Pharm. Sci. 108, 364–371 (2019). source
Not checked in the free run.
May combine two claims.
Interpretationown dataThe preliminary study suggested that the tattoo-based iontophoresis-sensor platform could be efficacious for diabetes management and could be relevant to monitor biomarkers from biofluids indicative of human performance.
“The preliminary study suggested that the tattoo-based iontophoresis-sensor platform could be efficacious for diabetes management and could be relevant to monitor biomarkers from biofluids indicative of human performance.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationA printed flexible tattoo-based glucose sensor showed a linear response toward physiologically relevant glucose levels with negligible responses from electroactive species in vitro.
“In vitro studies using the biosensor demonstrated a linear response toward physiologically relevant glucose levels with negligible responses from electroactive species.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [65] Bandodkar, A. J. et al. Tattoo-based noninvasive glucose monitoring: a proof-of-concept study. Anal. Chem. 87, 394–398 (2015). source
Not checked in the free run.
May combine two claims.
PremisecitationSweat rates in swimmers are lower than that of athletes on land and range from 0.33 to 1.6 l/h, depending on workout intensity and water temperature.
“The sweat rates in swimmers is lower than that of athletes on land and has been shown to range from 0.33 to 1.6 l/h, depending on the workout intensity and water temperature.70–72”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [70] Cox, G. R., Broad, E. M., Riley, M. D. & Burke, L. M. Body mass changes and voluntary fluid intakes of elite level water polo players and swimmers. J. Sci. Med. Sport 5, 183–193 (2002). source
Not checked in the free run.
- cites [71] Macaluso, F. et al. Effects of three different water temperatures on dehydration in competitive swimmers. Sci. Sports 26, 265–271 (2011). source
Not checked in the free run.
- cites [72] Maughan, R. J., Dargavel, L. A., Hares, R. & Shirreffs, S. M. Water and salt balance of well-trained swimmers in training. Int. J. Sport Nutr. Exerc. Metab. 19, 598–606 (2009). source
Not checked in the free run.
May combine two claims.
PremisecitationClinical studies demonstrated that Reeder et al.'s device measured local sweat chloride concentrations, local sweat loss, sweat rate, and skin temperature during intense physical activity in controlled indoor conditions and in open-ocean swimming during the IRONMAN triathlon.
“Clinical studies demonstrated the ability of the device to measure local sweat chloride concentrations, local sweat loss, sweat rate, and skin temperature during intense physical activity in controlled, indoor conditions and in open-ocean swimming during the IRONMAN triathlon.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [69] Reeder, J. T. et al. Waterproof, electronics-enabled, epidermal microfluidic devices for sweat collection, biomarker analysis, and thermography in aquatic settings. Sci. Adv. 5, eaau6356 (2019). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe sympathetic and parasympathetic nervous systems maintain cardiovascular homeostasis by responding to beat-to-beat perturbations that are sensed by baroreceptors and chemoreceptors.
“The sympathetic and parasympathetic nervous systems maintain cardiovascular homeostasis by responding to beat-to-beat perturbations that are sensed by baroreceptors and chemoreceptors.80”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [80] Heartbeat Analytics. Firstbeat. https://www.firstbeat.com/en/. Accessed 13 Jan 2019.
Not checked in the free run.
May combine two claims.
PremisecitationThe self-reported method is disadvantageous to measure human stress levels due to the lack of standards for stress status and the inability to assess the wellness and mental acuity of the athlete in a real-time manner.
“The self-reported method is disadvantageous to measure human stress levels due to the lack of standards for stress status and the inability to assess the wellness and mental acuity of the athlete in a real-time manner.75”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [75] Lee, J.-H., Hwang, Y., Cheon, K.-A. & Jung, H.-I. Emotion-on-a-chip (EOC): evolution of biochip technology to measure human emotion using body fluids. Med. Hypotheses 79, 827–832 (2012). source
Not checked in the free run.
May combine two claims.
Interpretationown dataThe work confirmed that sweat can be stimulated and the desired ions or analytes can be measured in a continuous manner to correlate with that in blood.
“Nonetheless, the work confirmed that sweat can be stimulated and the desired ions or analytes can be measured in a continuous manner to correlate with that in blood.”
Quote in the paperSubstantial claim- Rests on an assumption: Ethanol is 1:1 between sweat and blood due to its lipophilic nature.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationFluctuations in HR caused by respiration are referred to as respiratory sinus arrhythmia (RSA); HR increases during inspiration and decreases during expiration.
“Fluctuations in HR caused by respiration are referred to as respiratory sinus arrhythmia (RSA).79 Specifically, HR increases during inspiration and decreases during expiration.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [79] Yasuma, F. & Hayano, J.-I. Respiratory sinus arrhythmia: why does the heartbeat synchronize with respiratory rhythm? Chest 125, 683–690 (2004). source
Not checked in the free run.
May combine two claims.
PremisecitationMonitoring HRV provides useful knowledge for observing the interplay between the sympathetic and parasympathetic nervous systems, and is reflective of ANS activity.
“Monitoring HRV provides useful knowledge for observing the interplay between the sympathetic and parasympathetic nervous systems, and is reflective of ANS activity.82”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [82] Heart rate variability: a review | SpringerLink. https://link.springer.com/article/10.1007/s11517-006-0119-0. Accessed 10 Jan 2019.
Not checked in the free run.
May combine two claims.
PremisecitationHRV is affected by the training load of individual exercise sessions; the higher the training load, the lower the HRV after exercise.
“HRV is also affected by the training load of individual exercise sessions; the higher the training load, the lower the HRV after exercise.85”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [85] Kaikkonen, P., Hynynen, E., Mann, T., Rusko, H. & Nummela, A. Heart rate variability is related to training load variables in interval running exercises. Eur. J. Appl. Physiol. 112, 829–838 (2012). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationA stress reaction causes the activation of the ANS and the production of stress hormones along with an increased HR and an increased force of heart contractions.
“A stress reaction causes the activation of the ANS and the production of stress hormones along with an increased HR and an increased force of heart contractions.76”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [76] Chrousos, G. P. & Gold, P. W. The concepts of stress and stress system disorders. Overview of physical and behavioral homeostasis. J. Am. Med. Assoc. 267, 1244–1252 (1992). source
Not checked in the free run.
May combine two claims.
PremisecitationA low HRV (short-time interval) is suggestive of a stressful period whereas a prolonged HRV (long-time interval) is suggestive of a calm period.
“Thus, a low HRV (short-time interval) is suggestive of a stressful period whereas a prolonged HRV (long-time interval) is suggestive of a calm period.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [82] Heart rate variability: a review | SpringerLink. https://link.springer.com/article/10.1007/s11517-006-0119-0. Accessed 10 Jan 2019.
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe normal resting HR in a sitting position ranges between 60-80 beats/min due to the sympathetic and parasympathetic nervous systems, hormonal factors, and reflexive factors.
“The normal resting HR in a sitting position ranges between 60–80 beats/min due to the sympathetic and parasympathetic nervous systems, hormonal factors, and reflexive factors.78”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [78] 2 easy, accurate ways to measure your heart rate. Mayo Clinic. https://www.mayoclinic.org/healthy-lifestyle/fitness/expert-answers/heart-rate/faq-20057979. Accessed 13 Jan 2019.
Not checked in the free run.
May combine two claims.
PremisecitationConcurrently, HR increases and HRV decreases with increasing exercise intensity.
“Concurrently, HR increases and HRV decreases with increasing exercise intensity.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [81] Mitchell, J. H. Neural circulatory control during exercise: early insights. Exp. Physiol. 98, 867–878 (2013). source
Not checked in the free run.
May combine two claims.
PremisecitationTo characterize stress, arterial pulsewave signals are transformed into HRV, which represents the chronic stress level and an individual's stress vulnerability.
“To characterize stress, arterial pulsewave signals are transformed into HRV, which represents the chronic stress level and an individual’s stress vulnerability.98”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [98] Gil, E. et al. Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions. Physiol. Meas. 31, 1271–1290 (2010). source
Not checked in the free run.
May combine two claims.
PremisecitationAn epidermal sensor for multimodal physiological data analysis of stress measured skin temperature with a sensitivity of 0.31 Ω/°C, skin conductance sensitivity of 0.28 μV/0.02 μS, and a pulse wave response time of 70 ms.
“The stress patch measured skin temperature with a sensitivity of 0.31 Ω/°C, skin conductance sensitivity of 0.28 μV/0.02 μS, and a pulse wave response time of 70 ms.74”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [74] Yoon, S., Sim, J. K. & Cho, Y.-H. A flexible and wearable human stress monitoring patch. Sci. Rep. 6, 1–11 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe high-frequency component derived by the frequency domain analysis denotes the parasympathetic activity.
“For example, the high-frequency component derived by the frequency domain analysis denotes the parasympathetic activity.87”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [87] Georgiou, K., Larentzakis, A. & Papavassiliou, A. G. Surgeons’ and surgical trainees’ acute stress in real operations or simulation: a systematic review. Surgeon 15, 355–365 (2017). source
Not checked in the free run.
May combine two claims.
PremisecitationECG monitoring via wearable sensors poses challenges for athletics: surface EMG, increased electrode impedance, respiration induced baseline drift, and electrode contact movement can cause noise and motion artifacts.
“Firstly, surface EMG, increased electrode impedance, respiration induced baseline drift, and electrode contact movement can cause noise and motion artifacts.88”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [88] Allen, J. Photoplethysmography and its application in clinical physiological measurement. Physiol. Meas. 28, R1–R39 (2007). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationCortisol is responsible for maintaining homeostasis in the body via the regulation of neural, immune, cardiovascular, metabolic, and endocrine systems.
“Cortisol is responsible for maintaining homeostasis in the body via the regulation of neural, immune, cardiovascular, metabolic, and endocrine systems.102”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [102] Jia, M., Chew, W. M., Feinstein, Y., Skeath, P. & Sternberg, E. M. Quantification of cortisol in human eccrine sweat by liquid chromatography–tandem mass spectrometry. Analyst 141, 2053–2060 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationPrior work using biosensors to measure stress and cognition has focused on the measurement of a wide array of biomarkers (10−6-10−12 M) such as Orexin-A, cortisol, dopamine, neuropeptide Y, and interleukin-6 (IL-6).
“Prior work using biosensors to measure stress and cognition has focused on the measurement of a wide array of biomarkers (10−6–10−12 M) such as Orexin-A,113 cortisol,110 dopamine,114 neuropeptide Y,115 and interleukin-6 (IL-6).116”
Quote in the paperNot a substantial claim0 of 4 sources checked- cites [113] Hagen, J. et al. Detection of orexin A neuropeptide in biological fluids using a zinc oxide field effect transistor. ACS Chem. Neurosci. 4, 444–453 (2013). source
Not checked in the free run.
- cites [114] Roychoudhury, A., Basu, S. & Jha, S. K. Dopamine biosensor based on surface functionalized nanostructured nickel oxide platform. Biosens. Bioelectron. 84, 72–81 (2016). source
Not checked in the free run.
- cites [115] HAGEN, J. A. et al. Sweat Simulation, Collection and Sensing Systems (2013). https://patents.google.com/patent/WO2013152087A2/en.
Not checked in the free run.
- cites [116] Kumar, L. S. S. et al. Label free nano-aptasensor for interleukin-6 in protein-dilute bio fluids such as sweat. Anal. Methods 8, 3440–3444 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationCortisol is a steroid hormone secreted from the adrenal glands in response to stress as a product in the hypothalamic-pituitary adrenal pathway.
“Cortisol is a steroid hormone secreted from the adrenal glands in response to stress as a product in the hypothalamic–pituitary adrenal pathway.102”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [102] Jia, M., Chew, W. M., Feinstein, Y., Skeath, P. & Sternberg, E. M. Quantification of cortisol in human eccrine sweat by liquid chromatography–tandem mass spectrometry. Analyst 141, 2053–2060 (2016). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationCortisol has been measured in human blood, serum, urine, saliva, hair, interstitial fluid, and most recently eccrine sweat.
“Cortisol has been measured in human blood,106 serum,107 urine,107 saliva,107,108 hair,109 interstitial fluid,110 and most recently eccrine sweat.111”
Quote in the paperNot a substantial claim0 of 6 sources checked- cites [106] Port, K. Serum and saliva cortisol responses and blood lactate accumulation during incremental exercise testing. Int. J. Sports Med. 12, 490–494 (1991). source
Not checked in the free run.
- cites [107] Turpeinen, U. & Hämäläinen, E. Determination of cortisol in serum, saliva and urine. Best. Pract. Res. Clin. Endocrinol. Metab. 27, 795–801 (2013). source
Not checked in the free run.
- cites [108] Manetti, L. et al. Usefulness of salivary cortisol in the diagnosis of hypercortisolism: comparison with serum and urinary cortisol. Eur. J. Endocrinol. 168, 315–321 (2013). source
Not checked in the free run.
- cites [109] Wester, V. L. & Rossum, E. F. Cvan Clinical applications of cortisol measurements in hair. Eur. J. Endocrinol. 173, M1–M10 (2015). source
Not checked in the free run.
- cites [110] Venugopal, M., Arya, S. K., Chornokur, G. & Bhansali, S. A realtime and continuous assessment of cortisol in ISF using electrochemical impedance. Spectrosc. Sens. Actuators Phys. 172, 154–160 (2011). source
Not checked in the free run.
- cites [111] Parlak, O., Keene, S. T., Marais, A., Curto, V. F. & Salleo, A. Molecularly selective nanoporous membrane-based wearable organic electrochemical device for noninvasive cortisol sensing. Sci. Adv. 4, eaar2904 (2018). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationOrexin peptides and their corresponding receptors in the brain contribute to autonomic control, attention, feeding, memory, sleep, and stress.
“Orexin peptides and their corresponding receptors in the brain contribute to autonomic control, attention, feeding, memory, sleep, and stress.113”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [113] Hagen, J. et al. Detection of orexin A neuropeptide in biological fluids using a zinc oxide field effect transistor. ACS Chem. Neurosci. 4, 444–453 (2013). source
Not checked in the free run.
May combine two claims.
PremisecitationThe sensor selectivity was able to detect concentrations of approximately 100 aM in water, 10 fM in filtered human saliva, and 1 nM in filtered fetal bovine serum.
“The binding was transmitted to an electrical signal and the sensor selectivity was able to detect concentrations of approximately 100 aM in water, 10 fM in filtered human saliva, and 1 nM in filtered fetal bovine serum.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [113] Hagen, J. et al. Detection of orexin A neuropeptide in biological fluids using a zinc oxide field effect transistor. ACS Chem. Neurosci. 4, 444–453 (2013). source
Not checked in the free run.
May combine two claims.
PremisecitationThe sensor integrated an electrochemical transistor and a biomimetic polymeric membrane, which permitted the detection of cortisol.
“The team integrated an electrochemical transistor and a biomimetic polymeric membrane, which permitted the detection of cortisol.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [111] Parlak, O., Keene, S. T., Marais, A., Curto, V. F. & Salleo, A. Molecularly selective nanoporous membrane-based wearable organic electrochemical device for noninvasive cortisol sensing. Sci. Adv. 4, eaar2904 (2018). source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable).
InterpretationcitationDevelopment of improved sensitivity and stability of biosensors is required for the real-time detection of such biomarkers.
“The sensor platform demonstrated the potential of an FET device to measure a wide array of biomarkers in complex biofluids; however, development of improved sensitivity and stability of biosensors is required for the real-time detection of such biomarkers.113”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [113] Hagen, J. et al. Detection of orexin A neuropeptide in biological fluids using a zinc oxide field effect transistor. ACS Chem. Neurosci. 4, 444–453 (2013). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationKitman Labs utilizes machine learning platforms and motion technology to diagnose and uncover an athlete's unique stress response and various movements.
“Kitman Labs utilizes machine learning platforms and motion technology to diagnose and uncover an athlete’s unique stress response and various movements.133”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [133] Labs, K. Kitman Labs Releases First Annual Injury Assessment Report. http://www.prnewswire.com/news-releases/kitman-labs-releases-first-annual-injury-assessment-report-300281383.html. Accessed 15 May 2017.
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenDevices seeking to measure biosignals, biomarkers, and biomechanical parameters are challenged by the device technology itself and systems level issues associated with data analytics/data mining.
“However, as previously mentioned, devices seeking to measure biosignals, biomarkers, and biomechanical parameters are challenged by the device technology itself and systems level issues associated with data analytics/data mining.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationData mining techniques such as wavelet analysis for artifact reduction and data compression, rule-based methods for data summary and transmission, and Gaussian processing for secure authentication have been implemented.
“To solve such problems, data mining techniques such as wavelet analysis for artifact reduction and data compression, rule-based methods for data summary and transmission, and Gaussian processing for secure authentication have been implemented.129”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [129] Banaee, H., Ahmed, M. U. & Loutfi, A. Data mining for wearable sensors in health monitoring systems: a review of recent trends and challenges. Sensors 13, 17472–17500 (2013). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe emergence of IC fabrication strategies, flexible electronics, device design, and e-garments has revolutionized the development of soft electronics and biological and chemical sensor technologies toward advancing sports medicine.
“The emergence of IC fabrication strategies, flexible electronics, device design, and e-garments has revolutionized the development of soft electronics and biological and chemical sensor technologies toward advancing sports medicine.123–127”
Quote in the paperNot a substantial claim0 of 5 sources checked- cites [123] Bandodkar, A. J., Jeerapan, I. & Wang, J. Wearable chemical sensors: present challenges and future prospects. ACS Sens. 1, 464–482 (2016). source
Not checked in the free run.
- cites [124] Makin, T. R., Vignemont, Fde & Faisal, A. A. Neurocognitive barriers to the embodiment of technology. Nat. Biomed. Eng. 1, 0014 (2017). source
Not checked in the free run.
- cites [125] Bansal, A. K., Hou, S., Kulyk, O., Bowman, E. M. & Samuel, I. D. W. Wearable organic optoelectronic sensors for medicine. Adv. Mater. 27, 7638–7644 (2015). source
Not checked in the free run.
- cites [126] Choi, S., Lee, H., Ghaffari, R., Hyeon, T. & Kim, D.-H. Recent advances in flexible and stretchable bio-electronic devices integrated with nanomaterials. Adv. Mater. 28, 4203–4218 (2016). source
Not checked in the free run.
- cites [127] Parrilla, M., Cánovas, R., Jeerapan, I., Andrade, F. J. & Wang, J. A textile-based stretchable multi-ion potentiometric sensor. Adv. Healthc. Mater. 5, 996–1001 (2016). source
Not checked in the free run.
The paper relies on this without support.
Noninvasive continuous monitoring of biomarkers from saliva or sweat would allow sports medical personnel to tailor hydration and recovery protocols per athlete.
“The ability to monitor biomarkers from saliva or sweat in a noninvasive and continuous manner remain the next technological gap for sports medical personnel to tailor hydration and recovery protocols per the athlete.”
The paper relies on this without support.
Wearable sensor data can be translated into actionable insight to improve athlete health and performance.
“one issue plaguing the wearables field is the translation of the data to create actionable insight in its respective clinical domain”
The paper relies on this without support.
Ethanol is 1:1 between sweat and blood due to its lipophilic nature.
“Ethanol was selected because it is 1:1 between sweat and blood due to its lipophilic nature.”
The paper relies on this without support.
Human performance is a function of the physical demands of the sport coupled with the mental acuity exerted on the athlete.
“Human performance is a function of the physical demands of the sport coupled with the mental acuity exerted on the athlete.”
The paper relies on this without support.
Overtraining without adequate rest may lead to the overtraining condition, characterized by decreased performance and subsequent detriments to long-term health.
“Overtraining without adequate rest may lead to the overtraining condition, which is characterized by decreased performance and subsequent detriments to long-term health.”
The authors state this as an assumption.
Stress can be physiologically characterized by a reduced recovery of the neuroendocrine reaction and sympathetic dominance of the ANS function.
“While there currently lacks a formal definition of stress in scientific literature, we postulate that stress can be physiologically characterized by a reduced recovery of the neuroendocrine reaction and sympathetic dominance of the ANS function.77”
The authors state this as an assumption.
Recovery can be characterized as parasympathetic dominance.
“Recovery can be characterized as parasympathetic dominance.”
The paper relies on this without support.
The body should adapt to the demands put upon it in various situations.
“In other words, the body should adapt to the demands put upon it in various situations.”
The paper relies on this without support.
Parasympathetic activity should be dominating the ANS activity during the sleep period.
“Parasympathetic activity should be dominating the ANS activity during the sleep period.”
The paper relies on this without support.
HRV remains the most common method to determine the stress.
“HRV remains the most common method to determine the stress.”
The paper relies on this without support.
Overtraining is caused by long-term stress or exhaustion due to imbalance between training, other external/internal stressors, and recovery.
“Overtraining is caused by long-term stress or exhaustion due to imbalance between training, other external/internal stressors, and recovery.”
The paper relies on this without support.
Successful derivation of HRV parameters is dependent on the recording quality, the subject's activity during the recording, the removal of artifacts, the detection of arrhythmic beats, and the recording duration.
“Successful derivation of these HRV parameters is dependent on the recording quality, the subject’s activity during the recording, the removal of artifacts, the detection of arrhythmic beats, and the recording duration (seconds to days).”
The paper relies on this without support.
A majority of devices presented in both scientific and commercial literature are not yet validated in clinical studies for real-time assessment of human performance.
“However, a majority of the devices presented in both scientific and commercial literature are not yet validated in clinical studies for real-time assessment of human performance.”
The paper relies on this without support.
There remains an unmet medical need to differentiate between physical and mental (psychological) stress to understand the physical and mental demands of the athlete.
“Lastly, there remains an unmet medical need to differentiate between physical and mental (psychological) stress to understand the physical and mental demands of the athlete.”
The paper relies on this without support.
Current acute-to-chronic workload ratio models from wearable devices focus on only select sports such as Australian Football, Cricket, or Rugby.
“Current models focus in on only select sports such as Australian Football, Cricket, or Rugby.”
The paper relies on this without support.
The wearables field has recently seen a variety of devices for detecting position and motion, coupled with the emergence of clinical studies to assess their validity in professional or collegiate sports teams.
“The wearables field has recently seen a variety of devices for detecting position and motion, coupled with the emergence of clinical studies to assess their validity in professional or collegiate sports teams.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Cortisol concentration responds to the induction of physical stress in the body.
“In addition, cortisol concentration responds to the induction of physical stress in the body.[ 35 ]”
From MicroSweat: A wearable microfluidic patch for noninvasive and reliable sweat collection enables human stress monitoring (2023) · cites this paper as [35]
Non-invasive sensors of sweat and saliva might be more practical to integrate into wearables but still need to be carefully evaluated.
“Non-invasive sensors of sweat and saliva might be more practical to integrate into wearables but still need to be carefully evaluated32.”
From Smart wearable devices in cardiovascular care: Where we are and how to move forward (2021) · cites this paper as [32]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Table 6 denies commercial skin-conductance sensors listed in Table 2
MinorTable 6 says there are no commercial sensors for skin conductance, whereas Table 2 lists a company product whose functionality includes skin conductivity measurements.
“No. Currently there are no commercial sensors (sampling of devices exists in literature)”
“| Vinaya | Zenta | Wrist-based device | Optical, bio-impedance, and skin conductivity measurements are translated via machine learning to detect stress | London, UK |”
Cites
High-performance medicine: the convergence of human and artificial intelligence
Topol, E. J. (2019) · Nature Medicine · doi:10.1038/s41591-018-0300-7 · Cited by 10,450 · see citing works
Large-scale pattern growth of graphene films for stretchable transparent electrodes
Kim, K., Zhao, Y., Jang, H., Lee, S., Kim, J. M., Kim, K. S., Ahn, J., Kim, P., Choi, J., & Hong, B. H. (2009) · Nature · doi:10.1038/nature07719 · Cited by 10,399 · see citing works
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087 · see citing works
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
Electrochemical Glucose Biosensors
Wang, J. (2008) · Chemical Reviews · doi:10.1021/cr068123a · Cited by 3,498 · see citing works
Big data analytics in healthcare: promise and potential
Raghupathi, W., & Raghupathi, V. (2014) · Health Information Science and Systems · doi:10.1186/2047-2501-2-3 · Cited by 3,034 · see citing works
Heart rate variability: A review
Rajendra Acharya, U., Paul Joseph, K., Kannathal, N., Lim, C. M., & Suri, J. S. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0119-0 · Cited by 2,820 · see citing works
The Concepts of Stress and Stress System Disorders: Overview of Physical and Behavioral Homeostasis
Chrousos, G. P. (1992) · JAMA · doi:10.1001/jama.1992.03480090092034 · Cited by 2,689 · see citing works
Stress revisited: a critical evaluation of the stress concept
Koolhaas, J. M., Bartolomucci, A., Buwalda, B., de Boer, S. F., Flügge, G., Korte, S. M., Meerlo, P., Murison, R., Olivier, B., Palanza, P., et al. (2011) · Neuroscience & Biobehavioral Reviews · doi:10.1016/j.neubiorev.2011.02.003 · Cited by 1,626 · see citing works
Bariya, M., Nyein, H. Y. Y., & Javey, A. (2018) · Nature Electronics · doi:10.1038/s41928-018-0043-y · Cited by 1,486 · see citing works
Reis, J., Schambra, H. M., Cohen, L. G., Buch, E. R., Fritsch, B., Zarahn, E., Celnik, P. A., & Krakauer, J. W. (2009) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0805413106 · Cited by 1,401 · see citing works
A soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweat
Koh, A., Kang, D., Xue, Y., Lee, S., Pielak, R. M., Kim, J., Hwang, T., Min, S., Banks, A., Bastien, P., et al. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf2593 · Cited by 1,308 · see citing works
Non-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247 · see citing works
Lee, H., Song, C., Hong, Y. S., Kim, M., Cho, H. R., Kang, T., Shin, K., Choi, S. H., Hyeon, T., & Kim, D.-H. (2017) · Science Advances · doi:10.1126/sciadv.1601314 · Cited by 1,118 · see citing works
Recent Advances in Flexible and Stretchable Bio‐Electronic Devices Integrated with Nanomaterials
Choi, S., Lee, H., Ghaffari, R., Hyeon, T., & Kim, D. (2016) · Advanced Materials · doi:10.1002/adma.201504150 · Cited by 1,104 · see citing works
Yasuma, F., & Hayano, J.-I. (2004) · Chest · doi:10.1378/chest.125.2.683 · Cited by 907 · see citing works
Electrochemical Tattoo Biosensors for Real-Time Noninvasive Lactate Monitoring in Human Perspiration
Jia, W., Bandodkar, A. J., Valdés-Ramírez, G., Windmiller, J. R., Yang, Z., Ramírez, J., Chan, G., & Wang, J. (2013) · Analytical Chemistry · doi:10.1021/ac401573r · Cited by 862 · see citing works
Imani, S., Bandodkar, A. J., Mohan, A. M. V., Kumar, R., Yu, S., Wang, J., & Mercier, P. P. (2016) · Nature Communications · doi:10.1038/ncomms11650 · Cited by 861 · see citing works
Emaminejad, S., Gao, W., Wu, E., Davies, Z. A., Yin Yin Nyein, H., Challa, S., Ryan, S. P., Fahad, H. M., Chen, K., Shahpar, Z., et al. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1701740114 · Cited by 839 · see citing works
Saliva specimen: A new laboratory tool for diagnostic and basic investigation
Chiappin, S., Antonelli, G., Gatti, R., & De Palo, E. F. (2007) · Clinica Chimica Acta · doi:10.1016/j.cca.2007.04.011 · Cited by 828 · see citing works
Wearable Chemical Sensors: Present Challenges and Future Prospects
Bandodkar, A. J., Jeerapan, I., & Wang, J. (2016) · ACS Sensors · doi:10.1021/acssensors.6b00250 · Cited by 778 · see citing works
Urinary Indices of Hydration Status
Armstrong, L. E., Maresh, C. M., Castellani, J. W., Bergeron, M. F., Kenefick, R. W., LaGasse, K. E., & Riebe, D. A. (1994) · International Journal of Sport Nutrition · doi:10.1123/ijsn.4.3.265 · Cited by 768 · see citing works
Sonner, Z., Wilder, E., Heikenfeld, J., Kasting, G., Beyette, F., Swaile, D., Sherman, F., Joyce, J., Hagen, J., Kelley-Loughnane, N., et al. (2015) · Biomicrofluidics · doi:10.1063/1.4921039 · Cited by 724 · see citing works
Tattoo-Based Noninvasive Glucose Monitoring: A Proof-of-Concept Study
Bandodkar, A. J., Jia, W., Yardımcı, C., Wang, X., Ramirez, J., & Wang, J. (2015) · Analytical Chemistry · doi:10.1021/ac504300n · Cited by 700 · see citing works
Wearable salivary uric acid mouthguard biosensor with integrated wireless electronics
Kim, J., Imani, S., de Araujo, W. R., Warchall, J., Valdés-Ramírez, G., Paixão, T. R. L. C., Mercier, P. P., & Wang, J. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.07.039 · Cited by 678 · see citing works
Nyein, H. Y. Y., Gao, W., Shahpar, Z., Emaminejad, S., Challa, S., Chen, K., Fahad, H. M., Tai, L.-C., Ota, H., Davis, R. W., et al. (2016) · ACS Nano · doi:10.1021/acsnano.6b04005 · Cited by 669 · see citing works
Inside the brain of an elite athlete: the neural processes that support high achievement in sports
Yarrow, K., Brown, P., & Krakauer, J. W. (2009) · Nature reviews. Neuroscience · doi:10.1038/nrn2672 · Cited by 643 · see citing works
Parlak, O., Keene, S. T., Marais, A., Curto, V. F., & Salleo, A. (2018) · Science Advances · doi:10.1126/sciadv.aar2904 · Cited by 603 · see citing works
A BioMEMS Review: MEMS Technology for Physiologically Integrated Devices
Grayson, A. C. R., Shawgo, R. S., Johnson, A. M., Flynn, N. T., Li, Y., Cima, M. J., & Langer, R. (2004) · Proceedings of the IEEE · doi:10.1109/jproc.2003.820534 · Cited by 563 · see citing works
A contact lens with embedded sensor for monitoring tear glucose level
Yao, H., Shum, A. J., Cowan, M., Lähdesmäki, I., & Parviz, B. A. (2011) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.12.042 · Cited by 513 · see citing works
Cardiovascular, electrodermal, and respiratory response patterns to fear‐ and sadness‐inducing films
Kreibig, S. D., Wilhelm, F. H., Roth, W. T., & Gross, J. J. (2007) · Psychophysiology · doi:10.1111/j.1469-8986.2007.00550.x · Cited by 495 · see citing works
Bandodkar, A. J., Molinnus, D., Mirza, O., Guinovart, T., Windmiller, J. R., Valdés-Ramírez, G., Andrade, F. J., Schöning, M. J., & Wang, J. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2013.11.039 · Cited by 490 · see citing works
Martín, A., Kim, J., Kurniawan, J. F., Sempionatto, J. R., Moreto, J. R., Tang, G., Campbell, A. S., Shin, A., Lee, M. Y., Liu, X., et al. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00729 · Cited by 475 · see citing works
Gil, E., Orini, M., Bailón, R., Vergara, J. M., Mainardi, L., & Laguna, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/9/015 · Cited by 461 · see citing works
Banaee, H., Ahmed, M., & Loutfi, A. (2013) · Sensors · doi:10.3390/s131217472 · Cited by 452 · see citing works
A wearable multisensing patch for continuous sweat monitoring
Anastasova, S., Crewther, B., Bembnowicz, P., Curto, V., Ip, H. M., Rosa, B., & Yang, G.-Z. (2017) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2016.09.038 · Cited by 443 · see citing works
Variations in Regional Sweat Composition in Normal Human Males
Patterson, M. J., Galloway, S. D. R., & Nimmo, M. A. (2000) · Experimental Physiology · doi:10.1111/j.1469-445x.2000.02058.x · Cited by 438 · see citing works
Skin-interfaced systems for sweat collection and analytics
Choi, J., Ghaffari, R., Baker, L. B., & Rogers, J. A. (2018) · Science Advances · doi:10.1126/sciadv.aar3921 · Cited by 433 · see citing works
Li, X., Dunn, J., Salins, D., Zhou, G., Zhou, W., Schüssler-Fiorenza Rose, S. M., Perelman, D., Colbert, E., Runge, R., Rego, S., et al. (2017) · PLOS Biology · doi:10.1371/journal.pbio.2001402 · Cited by 431 · see citing works
Roll-to-Roll Gravure Printed Electrochemical Sensors for Wearable and Medical Devices
Bariya, M., Shahpar, Z., Park, H., Sun, J., Jung, Y., Gao, W., Nyein, H. Y. Y., Liaw, T. S., Tai, L.-C., Ngo, Q. P., et al. (2018) · ACS Nano · doi:10.1021/acsnano.8b02505 · Cited by 417 · see citing works
Adhesive RFID Sensor Patch for Monitoring of Sweat Electrolytes
Rose, D. P., Ratterman, M. E., Griffin, D. K., Hou, L., Kelley-Loughnane, N., Naik, R. R., Hagen, J. A., Papautsky, I., & Heikenfeld, J. C. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2369991 · Cited by 416 · see citing works
Amyotrophic Lateral Sclerosis: A Focus on Disease Progression
Calvo, A. C., Manzano, R., de Mendonça, D. M. F., Muñoz, M. J., Zaragoza, P., & Osta, R. (2014) · BioMed Research International · doi:10.1155/2014/925101 · Cited by 404 · see citing works
Baker, L. B. (2017) · Sports Medicine · doi:10.1007/s40279-017-0691-5 · Cited by 403 · see citing works
The physiology, pharmacology, and biochemistry of the eccrine sweat gland
Sato, K. (1977) · Reviews of Physiology, Biochemistry and Pharmacology · doi:10.1007/bfb0037089 · Cited by 384 · see citing works
Non-invasive mouthguard biosensor for continuous salivary monitoring of metabolites
Kim, J., Valdés-Ramírez, G., Bandodkar, A. J., Jia, W., Martinez, A. G., Ramírez, J., Mercier, P., & Wang, J. (2014) · The Analyst · doi:10.1039/c3an02359a · Cited by 367 · see citing works
Abellán-Llobregat, A., Jeerapan, I., Bandodkar, A., Vidal, L., Canals, A., Wang, J., & Morallón, E. (2017) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2017.01.058 · Cited by 348 · see citing works
A potentiometric tattoo sensor for monitoring ammonium in sweat
Guinovart, T., Bandodkar, A. J., Windmiller, J. R., Andrade, F. J., & Wang, J. (2013) · The Analyst · doi:10.1039/c3an01672b · Cited by 343 · see citing works
Cogiamanian, F., Marceglia, S., Ardolino, G., Barbieri, S., & Priori, A. (2007) · European Journal of Neuroscience · doi:10.1111/j.1460-9568.2007.05633.x · Cited by 334 · see citing works
Clinical and diagnostic utility of saliva as a non-invasive diagnostic fluid: a systematic review
Soares Nunes, L. A., Mussavira, S., & Sukumaran Bindhu, O. (2015) · Biochemia Medica · doi:10.11613/bm.2015.018 · Cited by 333 · see citing works
Recent advances in cortisol sensing technologies for point-of-care application
Kaushik, A., Vasudev, A., Arya, S. K., Pasha, S. K., & Bhansali, S. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2013.09.060 · Cited by 316 · see citing works
Recent developments in blood glucose sensors
Wang, H.-C., & Lee, A. (2015) · Journal of Food and Drug Analysis · doi:10.1016/j.jfda.2014.12.001 · Cited by 316 · see citing works
Relation between heart rate variability and training load in middle-distance runners
PICHOT, V., ROCHE, F., GASPOZ, J.-M., ENJOLRAS, F., ANTONIADIS, A., MININI, P., COSTES, F., BUSSO, T., LACOUR, J.-R., & BARTH??L??MY, J. C. (2000) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200010000-00011 · Cited by 315 · see citing works
Reeder, J. T., Choi, J., Xue, Y., Gutruf, P., Hanson, J., Liu, M., Ray, T., Bandodkar, A. J., Avila, R., Xia, W., et al. (2019) · Science Advances · doi:10.1126/sciadv.aau6356 · Cited by 308 · see citing works
Towards Measuring Stress with Smartphones and Wearable Devices During Workday and Sleep
Muaremi, A., Arnrich, B., & Tröster, G. (2013) · BioNanoScience · doi:10.1007/s12668-013-0089-2 · Cited by 302 · see citing works
Non‐invasive Analyte Access and Sensing through Eccrine Sweat: Challenges and Outlook circa 2016
Heikenfeld, J. (2016) · Electroanalysis · doi:10.1002/elan.201600018 · Cited by 294 · see citing works
A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration
Schazmann, B., Morris, D., Slater, C., Beirne, S., Fay, C., Reuveny, R., Moyna, N., & Diamond, D. (2010) · Analytical Methods · doi:10.1039/b9ay00184k · Cited by 286 · see citing works
Flexible glucose sensor using CVD-grown graphene-based field effect transistor
Kwak, Y. H., Choi, D. S., Kim, Y. N., Kim, H., Yoon, D. H., Ahn, S., Yang, J., Yang, W. S., & Seo, S. (2012) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2012.04.042 · Cited by 285 · see citing works
Plasma potassium changes with high intensity exercise.
Medbø, J. I., & Sejersted, O. M. (1990) · The Journal of Physiology · doi:10.1113/jphysiol.1990.sp017935 · Cited by 275 · see citing works
A Textile‐Based Stretchable Multi‐Ion Potentiometric Sensor
Parrilla, M., Cánovas, R., Jeerapan, I., Andrade, F. J., & Wang, J. (2016) · Advanced Healthcare Materials · doi:10.1002/adhm.201600092 · Cited by 269 · see citing works
Saliva-Based Biosensors: Noninvasive Monitoring Tool for Clinical Diagnostics
Malon, R. S. P., Sadir, S., Balakrishnan, M., & Córcoles, E. P. (2014) · BioMed Research International · doi:10.1155/2014/962903 · Cited by 260 · see citing works
Can Wearable Devices Accurately Measure Heart Rate Variability? A Systematic Review
Georgiou, K., Larentzakis, A. V., Khamis, N. N., Alsuhaibani, G. I., Alaska, Y. A., & Giallafos, E. J. (2018) · Folia Medica · doi:10.2478/folmed-2018-0012 · Cited by 244 · see citing works
Sekine, Y., Kim, S. B., Zhang, Y., Bandodkar, A. J., Xu, S., Choi, J., Irie, M., Ray, T. R., Kohli, P., Kozai, N., et al. (2018) · Lab on a Chip · doi:10.1039/c8lc00530c · Cited by 242 · see citing works
Apelin, orexin-A and leptin plasma levels in morbid obesity and effect of gastric banding
Heinonen, M. V., Purhonen, A. K., Miettinen, P., Pääkkönen, M., Pirinen, E., Alhava, E., Åkerman, K., & Herzig, K. H. (2005) · Regulatory Peptides · doi:10.1016/j.regpep.2005.05.003 · Cited by 232 · see citing works
Highly flexible, wearable, and disposable cardiac biosensors for remote and ambulatory monitoring
Lee, S. P., Ha, G., Wright, D. E., Ma, Y., Sen-Gupta, E., Haubrich, N. R., Branche, P. C., Li, W., Huppert, G. L., Johnson, M. D., et al. (2018) · npj Digital Medicine · doi:10.1038/s41746-017-0009-x · Cited by 229 · see citing works
Making the case for mobile cognition: EEG and sports performance
Park, J. L., Fairweather, M. M., & Donaldson, D. I. (2015) · Neuroscience & Biobehavioral Reviews · doi:10.1016/j.neubiorev.2015.02.014 · Cited by 218 · see citing works
Determination of cortisol in serum, saliva and urine
Turpeinen, U., & Hämäläinen, E. (2013) · Best Practice & Research Clinical Endocrinology & Metabolism · doi:10.1016/j.beem.2013.10.008 · Cited by 214 · see citing works
Kubota, K. J., Chen, J. A., & Little, M. A. (2016) · Movement Disorders · doi:10.1002/mds.26693 · Cited by 210 · see citing works
Clinical applications of cortisol measurements in hair
Wester, V. L., & van Rossum, E. F. C. (2015) · European Journal of Endocrinology · doi:10.1530/eje-15-0313 · Cited by 210 · see citing works
Urine osmolality and conductivity as indices of hydration status in athletes in the heat
Shirreffs, S. M., & Maughan, R. J. (1998) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-199811000-00007 · Cited by 209 · see citing works
Wearable Organic Optoelectronic Sensors for Medicine
Bansal, A. K., Hou, S., Kulyk, O., Bowman, E. M., & Samuel, I. D. W. (2015) · Advanced Materials · doi:10.1002/adma.201403560 · Cited by 205 · see citing works
Hew-Butler, T., Rosner, M. H., Fowkes-Godek, S., Dugas, J. P., Hoffman, M. D., Lewis, D. P., Maughan, R. J., Miller, K. C., Montain, S. J., Rehrer, N. J., et al. (2015) · Clinical Journal of Sport Medicine · doi:10.1097/jsm.0000000000000221 · Cited by 205 · see citing works
Munje, R. D., Muthukumar, S., & Prasad, S. (2017) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2016.07.088 · Cited by 190 · see citing works
Baker, L. B., Stofan, J. R., Hamilton, A. A., & Horswill, C. A. (2009) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00197.2009 · Cited by 189 · see citing works
Kuang, Z., Kim, S. N., Crookes-Goodson, W. J., Farmer, B. L., & Naik, R. R. (2010) · ACS Nano · doi:10.1021/nn901365g · Cited by 184 · see citing works
Dopamine biosensor based on surface functionalized nanostructured nickel oxide platform
Roychoudhury, A., Basu, S., & Jha, S. K. (2016) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.11.061 · Cited by 184 · see citing works
Lee, Y., Howe, C., Mishra, S., Lee, D. S., Mahmood, M., Piper, M., Kim, Y., Tieu, K., Byun, H.-S., Coffey, J. P., et al. (2018) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1719573115 · Cited by 183 · see citing works
A flexible and wearable biosensor for tear glucose measurement
Iguchi, S., Kudo, H., Saito, T., Ogawa, M., Saito, H., Otsuka, K., Funakubo, A., & Mitsubayashi, K. (2007) · Biomedical Microdevices · doi:10.1007/s10544-007-9073-3 · Cited by 173 · see citing works
Influence of acclimatization on sweat sodium concentration.
Allan, J. R., & Wilson, C. G. (1971) · Journal of Applied Physiology · doi:10.1152/jappl.1971.30.5.708 · Cited by 171 · see citing works
Neurocognitive barriers to the embodiment of technology
Makin, T. R., de Vignemont, F., & Faisal, A. A. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-016-0014 · Cited by 165 · see citing works
Hauke, A., Simmers, P., Ojha, Y. R., Cameron, B. D., Ballweg, R., Zhang, T., Twine, N., Brothers, M., Gomez, E., & Heikenfeld, J. (2018) · Lab on a Chip · doi:10.1039/c8lc01082j · Cited by 156 · see citing works
Uusitalo, Uusitalo, & Rusko (2000) · International Journal of Sports Medicine · doi:10.1055/s-2000-8853 · Cited by 155 · see citing works
Analysis of metabolites in sweat as a measure of physical condition
Mitsubayashi, K., Suzuki, M., Tamiya, E., & Karube, I. (1994) · Analytica Chimica Acta · doi:10.1016/0003-2670(94)80004-9 · Cited by 153 · see citing works
A Flexible and Wearable Human Stress Monitoring Patch
Yoon, S., Sim, J. K., & Cho, Y.-H. (2016) · Scientific Reports · doi:10.1038/srep23468 · Cited by 151 · see citing works
Fall in skin temperature of exercising man.
Torii, M., Yamasaki, M., Sasaki, T., & Nakayama, H. (1992) · British Journal of Sports Medicine · doi:10.1136/bjsm.26.1.29 · Cited by 145 · see citing works
Trials and tribulations of skin iontophoresis in therapeutics
Roustit, M., Blaise, S., & Cracowski, J. (2014) · British Journal of Clinical Pharmacology · doi:10.1111/bcp.12128 · Cited by 136 · see citing works
Possibilities and limitations of the polar RS800 in measuring heart rate variability at rest
Wallén, M. B., Hasson, D., Theorell, T., Canlon, B., & Osika, W. (2012) · European Journal of Applied Physiology · doi:10.1007/s00421-011-2079-9 · Cited by 135 · see citing works
Normative data for regional sweat sodium concentration and whole-body sweating rate in athletes
Baker, L. B., Barnes, K. A., Anderson, M. L., Passe, D. H., & Stofan, J. R. (2016) · Journal of Sports Sciences · doi:10.1080/02640414.2015.1055291 · Cited by 133 · see citing works
Marques-Deak, A., Cizza, G., Eskandari, F., Torvik, S., Christie, I. C., Sternberg, E. M., & Phillips, T. M. (2006) · Journal of Immunological Methods · doi:10.1016/j.jim.2006.07.011 · Cited by 129 · see citing works
Immunosensors: Antibody-based biosensors
North, J. R. (1985) · Trends in Biotechnology · doi:10.1016/0167-7799(85)90119-2 · Cited by 123 · see citing works
Jia, M., Chew, W. M., Feinstein, Y., Skeath, P., & Sternberg, E. M. (2016) · The Analyst · doi:10.1039/c5an02387d · Cited by 113 · see citing works
The Role of Intrinsic Pathway in Apoptosis Activation and Progression in Peyronie’s Disease
Loreto, C., La Rocca, G., Anzalone, R., Caltabiano, R., Vespasiani, G., Castorina, S., Ralph, D. J., Cellek, S., Musumeci, G., Giunta, S., et al. (2014) · BioMed Research International · doi:10.1155/2014/616149 · Cited by 108 · see citing works
Port, K. (1991) · International Journal of Sports Medicine · doi:10.1055/s-2007-1024720 · Cited by 105 · see citing works
Immunoreactive orexin-A in human plasma
Arihara, Z., Takahashi, K., Murakami, O., Totsune, K., Sone, M., Satoh, F., Ito, S., & Mouri, T. (2001) · Peptides · doi:10.1016/s0196-9781(00)00369-7 · Cited by 101 · see citing works
A realtime and continuous assessment of cortisol in ISF using electrochemical impedance spectroscopy
Venugopal, M., Arya, S. K., Chornokur, G., & Bhansali, S. (2011) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2011.04.028 · Cited by 100 · see citing works
Wearable Potentiometric Chloride Sweat Sensor: The Critical Role of the Salt Bridge
Choi, D.-H., Kim, J. S., Cutting, G. R., & Searson, P. C. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03391 · Cited by 100 · see citing works
Telemetry of plaque pH from interdental area.
Graf, H., & Hr, M. (1966) · PubMed · Cited by 82 · see citing works
An integrated sensing and wireless communications platform for sensing sodium in sweat
Matzeu, G., O'Quigley, C., McNamara, E., Zuliani, C., Fay, C., Glennon, T., & Diamond, D. (2016) · Analytical Methods · doi:10.1039/c5ay02254a · Cited by 80 · see citing works
Real time stress detection system based on EEG signals
Vanitha, V., & Krishnan, P. (2016) · Biomedical Research-tokyo · Cited by 79 · see citing works
Heart rate variability is related to training load variables in interval running exercises
Kaikkonen, P., Hynynen, E., Mann, T., Rusko, H., & Nummela, A. (2012) · European Journal of Applied Physiology · doi:10.1007/s00421-011-2031-z · Cited by 78 · see citing works
Ferrell, W. R., Ramsay, J. E., Brooks, N., Lockhart, J. C., Dickson, S., McNeece, G. M., Greer, I. A., & Sattar, N. (2002) · Journal of Vascular Research · doi:10.1159/000064515 · Cited by 76 · see citing works
The Roles of Ionic Processes in Muscular Fatigue During Intense Exercise1
McKenna, M. J. (1992) · Sports Medicine · doi:10.2165/00007256-199213020-00009 · Cited by 73 · see citing works
Godek, S. F., Peduzzi, C., Burkholder, R., Condon, S., Dorshimer, G., & Bartolozzi, A. R. (2010) · Journal of Athletic Training · doi:10.4085/1062-6050-45.4.364 · Cited by 73 · see citing works
Nanotechnology-based electrochemical sensors for biomonitoring chemical exposures
Barry, R. C., Lin, Y., Wang, J., Liu, G., & Timchalk, C. A. (2009) · Journal of Exposure Science & Environmental Epidemiology · doi:10.1038/jes.2008.71 · Cited by 73 · see citing works
Cai, X., Yan, J., Chu, H., Wu, M., & Tu, Y. (2010) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2009.10.002 · Cited by 71 · see citing works
Manetti, L., Rossi, G., Grasso, L., Raffaelli, V., Scattina, I., Del Sarto, S., Cosottini, M., Iannelli, A., Gasperi, M., Bogazzi, F., et al. (2013) · European Journal of Endocrinology · doi:10.1530/eje-12-0685 · Cited by 69 · see citing works
Surgeons' and surgical trainees' acute stress in real operations or simulation: A systematic review
Georgiou, K., Larentzakis, A., & Papavassiliou, A. G. (2017) · The Surgeon · doi:10.1016/j.surge.2017.06.003 · Cited by 69 · see citing works
Body mass changes and voluntary fluid intakes of elite level water polo players and swimmers
Cox, G., Broad, E., Riley, M., & Burke, L. (2002) · Journal of Science and Medicine in Sport · doi:10.1016/s1440-2440(02)80003-2 · Cited by 66 · see citing works
Modeling Glucose Transport From Systemic Circulation to Sweat
La Count, T. D., Jajack, A., Heikenfeld, J., & Kasting, G. B. (2019) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2018.09.026 · Cited by 60 · see citing works
Label free nano-aptasensor for interleukin-6 in protein-dilute bio fluids such as sweat
Kumar, L. S. S., Wang, X., Hagen, J., Naik, R., Papautsky, I., & Heikenfeld, J. (2016) · Analytical Methods · doi:10.1039/c6ay00331a · Cited by 59 · see citing works
Ion Sensor for the Quantification of Sodium in Sweat Samples
Wujcik, E. K., Blasdel, N. J., Trowbridge, D., & Monty, C. N. (2013) · IEEE Sensors Journal · doi:10.1109/jsen.2013.2257168 · Cited by 56 · see citing works
A method of identifying chronic stress by EEG
Peng, H., Hu, B., Zheng, F., Fan, D., Zhao, W., Chen, X., Yang, Y., & Cai, Q. (2013) · Personal and Ubiquitous Computing · doi:10.1007/s00779-012-0593-3 · Cited by 55 · see citing works
Neural circulatory control during exercise: early insights
Mitchell, J. H. (2013) · Experimental Physiology · doi:10.1113/expphysiol.2012.071001 · Cited by 54 · see citing works
Sweat chloride concentration: sweat rate, metabolic rate, skin temperature, and age.
Dill, D. B., Hall, F. G., & Van Beaumont, W. (1966) · Journal of Applied Physiology · doi:10.1152/jappl.1966.21.1.99 · Cited by 51 · see citing works
Amperometric biosensor for determining human salivary phosphate
Kwan, R. C. H., Leung, H. F., Hon, P. Y. T., Cheung, H. C. F., Hirota, K., & Renneberg, R. (2005) · Analytical Biochemistry · doi:10.1016/j.ab.2005.05.021 · Cited by 50 · see citing works
Glucose detection in human sweat using an electronic nose
Olarte, O., Chilo, J., Pelegrí-Sebastiá, J., Barbé, K., & Moer, W. V. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6609787 · Cited by 49 · see citing works
Zhou, X., Xu, H., Cheng, J., et al. (2015) · Scientific Reports · doi:10.1038/srep10402 · Cited by 47 · see citing works
A disposable biosensor for the determination of alpha-amylase in human saliva
Mahosenaho, M., Caprio, F., Micheli, L., Sesay, A. M., Palleschi, G., & Virtanen, V. (2010) · Microchimica Acta · doi:10.1007/s00604-010-0360-y · Cited by 45 · see citing works
Gambarotta, N., Aletti, F., Baselli, G., & Ferrario, M. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-016-1453-5 · Cited by 45 · see citing works
Textile Integrated Contactless EMG Sensing for Stress Analysis
Taelman, J., Adriaensen, T., van der Horst, C., et al. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4353202 · Cited by 44 · see citing works
American Heart Association Precision Medicine Platform
Kass‐Hout, T., Stevens, L. M., & Hall, J. L. (2018) · Circulation · doi:10.1161/circulationaha.117.032041 · Cited by 44 · see citing works
Water and Salt Balance of Well-Trained Swimmers in Training
Maughan, R. J., Dargavel, L. A., Hares, R., & Shirreffs, S. M. (2009) · International Journal of Sport Nutrition and Exercise Metabolism · doi:10.1123/ijsnem.19.6.598 · Cited by 39 · see citing works
A review of wearable technology: Moving beyond the hype: From need through sensor implementation
Seshadri, D. R., Rowbottom, J. R., Drummond, C. K., Voos, J. E., & Craker, J. (2016) · Cairo International Biomedical Engineering Conference · doi:10.1109/cibec.2016.7836118 · Cited by 37 · see citing works
Self-citationCSF hypocretin-1 levels are normal in multiple-system atrophy
Abdo, W. F., Bloem, B. R., Kremer, H. P. H., Lammers, G. J., Verbeek, M. M., & Overeem, S. (2008) · Parkinsonism & Related Disorders · doi:10.1016/j.parkreldis.2007.08.012 · Cited by 34 · see citing works
Effects of three different water temperatures on dehydration in competitive swimmers
Macaluso, F., Di Felice, V., Boscaino, G., Bonsignore, G., Stampone, T., Farina, F., & Morici, G. (2011) · Science & Sports · doi:10.1016/j.scispo.2010.10.004 · Cited by 32 · see citing works
Couto, R., & Quinaz, M. (2016) · Sensors · doi:10.3390/s16071015 · Cited by 28 · see citing works
Fluid Balance and Sodium Losses During Indoor Tennis Match Play
Lott, M. J. E., & Galloway, S. D. R. (2011) · International Journal of Sport Nutrition and Exercise Metabolism · doi:10.1123/ijsnem.21.6.492 · Cited by 27 · see citing works
Detection of Orexin A Neuropeptide in Biological Fluids Using a Zinc Oxide Field Effect Transistor
Hagen, J., Lyon, W., Chushak, Y., Tomczak, M., Naik, R., Stone, M., & Kelley-Loughnane, N. (2013) · ACS Chemical Neuroscience · doi:10.1021/cn300159e · Cited by 22 · see citing works
Emotion-on-a-chip (EOC): Evolution of biochip technology to measure human emotion using body fluids
Lee, J.-H., Hwang, Y., Cheon, K.-A., & Jung, H.-I. (2012) · Medical Hypotheses · doi:10.1016/j.mehy.2012.09.002 · Cited by 20 · see citing works
Online Predictive Tools for Intervention in Mental Illness: The OPTIMI Project
Botella, C., Moragrega, I., Baños, R., & García‐Palacios, A. (2011) · Studies in health technology and informatics · doi:10.3233/978-1-60750-706-2-86 · Cited by 20 · see citing works
Lee, Y., Lee, B., & Lee, M. (2010) · Telemedicine and e-Health · doi:10.1089/tmj.2009.0039 · Cited by 19 · see citing works
Preparation of Disposable Saliva α-Amylase Biosensor
Chao-Shi, Z. O. U., Mi, Z. H. O. U., Xie, G., Peng, L. U. O., Xiong, X., Xu, H., & Zheng, J. (2008) · Chinese Journal of Analytical Chemistry · doi:10.1016/s1872-2040(08)60068-7 · Cited by 17 · see citing works
Disposable blood potassium sensors based on screen-printed thick film electrodes
Xu, H., Yang, X., Wang, Y., Zheng, J., Luo, Z., & Li, G. (2010) · Measurement Science and Technology · doi:10.1088/0957-0233/21/5/055802 · Cited by 14 · see citing works
Continuous Non-Invasive Ophthalmic Glucose Sensor for Diabetics
Domschke, A. (2010) · CHIMIA International Journal for Chemistry · doi:10.2533/chimia.2010.43 · Cited by 11 · see citing works
The factors influence compatibility of pulse-pulse intervals with R-R intervals
Liu, A., Wu, H., Liu, C.-C., Hsu, C.-H., & Chen, D.-Y. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6609939 · Cited by 4 · see citing works
A Stretchable, Wearable Lactate Sensor with a Textile-Based Biofuel Cell
Kato, Y., Ogawa, Y., Kai, H., et al. (2016) · ECS Meeting Abstracts · doi:10.1149/ma2016-02/44/3210 · Cited by 2 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Smart wearable devices in cardiovascular care: Where we are and how to move forward
Bayoumy, K., Gaber, M., Elshafeey, A., Mhaimeed, O., Dineen, E. H., Marvel, F. A., Martin, S. S., Muse, E. D., Turakhia, M. P., Tarakji, K. G., et al. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-021-00522-7 · Cited by 829 · see citing works
Artificial intelligence biosensors: Challenges and prospects
Jin, X., Liu, C., Xu, T., Su, L., & Zhang, X. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112412 · Cited by 399 · see citing works
Touch‐Based Stressless Cortisol Sensing
Tang, W., Yin, L., Sempionatto, J. R., Moon, J., Teymourian, H., & Wang, J. (2021) · Advanced Materials · doi:10.1002/adma.202008465 · Cited by 302 · see citing works
Wearable Biosensors for Non-Invasive Sweat Diagnostics
Xu, J., Fang, Y., & Chen, J. (2021) · Biosensors · doi:10.3390/bios11080245 · Cited by 208 · see citing works
Quer, G., Gouda, P., Galarnyk, M., Topol, E. J., & Steinhubl, S. R. (2020) · PLOS ONE · doi:10.1371/journal.pone.0227709 · Cited by 185 · see citing works
The Current State of Optical Sensors in Medical Wearables
Vavrinsky, E., Esfahani, N. E., Hausner, M., Kuzma, A., Rezo, V., Donoval, M., & Kosnacova, H. (2022) · Biosensors · doi:10.3390/bios12040217 · Cited by 103 · see citing works
Phatak, A. A., Wieland, F.-G., Vempala, K., Volkmar, F., & Memmert, D. (2021) · Sports Medicine - Open · doi:10.1186/s40798-021-00372-0 · Cited by 74 · see citing works
Portable and wearable real-time stress monitoring: A critical review
Parlak, O. (2021) · Sensors and Actuators Reports · doi:10.1016/j.snr.2021.100036 · Cited by 60 · see citing works
Yan, T., Zhang, G., Chai, H., Qu, L., & Zhang, X. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.753692 · Cited by 59 · see citing works
Shajari, S., Salahandish, R., Zare, A., Hassani, M., Moossavi, S., Munro, E., Rashid, R., Rosenegger, D., Bains, J. S., & Sanati Nezhad, A. (2023) · Advanced Science · doi:10.1002/advs.202204171 · Cited by 55 · see citing works
- Authors
- Seshadri, D. R., Li, R. T., Voos, J. E., Rowbottom, J. R., Alfes, C. M., Zorman, C. A., & Drummond, C. K.
- Affiliations
- Case Western Reserve University (US)
- University Hospitals of Cleveland (US)
- Cleveland Clinic (US)
- Corresponding author
- Seshadri, Dhruv R.
- ORCID
- Venue
- npj Digital Medicine
- Published
- 2019-07-22
- Volume
- 2
- Issue
- 1
- Pages
- 72-72
- ISSN
- 2398-6352 (ISSN-L)
- PMC
- PMC6646404
- OpenAlex
- W2964108233
- Semantic Scholar
- f71913dd2c6c1cea045bddf24c4859be837ff4bb
- Cited by
- 537
- References
- 135
- Citations by year
- 2026: 71 · 2025: 97 · 2024: 82 · 2023: 63 · 2022: 80 · 2021: 102 · 2020: 39 · 2019: 3
- FWCI
- 4.52
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- npj Digital Medicine (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Scholarly Commons @ Case Western Reserve University (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- D.R.S and C.K.D. acknowledge financial support from the Brenda A. and Robert M. Aiken Strategic Initiative.
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Keywords
- wearable sensors, athletes, recovery protocols, salivary biomarkers, sweat biomarkers, non-invasive monitoring, flexible electronics, stretchable electronics, hydration, stress symptoms, athletic performance
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review