A combined segmenting and non-segmenting approach to signal quality estimation for ambulatory photoplethysmography
The DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.
Abstract
Continuous cardiac monitoring of healthy and unhealthy patients can help us understand the progression of heart disease and enable early treatment. Optical pulse sensing is an excellent candidate for continuous mobile monitoring of cardiovascular health indicators, but optical pulse signals are susceptible to corruption from a number of noise sources, including motion artifact. Therefore, before higher-level health indicators can be reliably computed, corrupted data must be separated from valid data. This is an especially difficult task in the presence of artifact caused by ambulation (e.g. walking or jogging), which shares significant spectral energy with the true pulsatile signal. In this manuscript, we present a machine-learning-based system for automated estimation of signal quality of optical pulse signals that performs well in the presence of periodic artifact. We hypothesized that signal processing methods that identified individual heart beats (segmenting approaches) would be more error-prone than methods that did not (non-segmenting approaches) when applied to data contaminated by periodic artifact. We further hypothesized that a fusion of segmenting and non-segmenting approaches would outperform either approach alone. Therefore, we developed a novel non-segmenting approach to signal quality estimation that we then utilized in combination with a traditional segmenting approach. Using this system we were able to robustly detect differences in signal quality as labeled by expert human raters (Pearson's r = 0.9263). We then validated our original hypotheses by demonstrating that our non-segmenting approach outperformed the segmenting approach in the presence of contaminated signal, and that the combined system outperformed either individually. Lastly, as an example, we demonstrated the utility of our signal quality estimation system in evaluating the trustworthiness of heart rate measurements derived from optical pulse signals.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Taylor, R. S., Brown, A., Ebrahim, S., Jolliffe, J., Noorani, H., Rees, K., Skidmore, B., Stone, J. A., Thompson, D. R., & Oldridge, N. (2004) · The American Journal of Medicine · doi:10.1016/j.amjmed.2004.01.009 · Cited by 2,239
Heart-rate Recovery Immediately After Exercise as a Predictor of Mortality
Cole, C. R., Blackstone, E. H., Pashkow, F. J., Snader, C. E., & Lauer, M. S. (1999) · New England Journal of Medicine · doi:10.1056/nejm199910283411804 · Cited by 1,935
How accurate is pulse rate variability as an estimate of heart rate variability?
Schäfer, A., & Vagedes, J. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.03.119 · Cited by 661
Mobile monitoring with wearable photoplethysmographic biosensors
Asada, H. H., Shaltis, P., Reisner, A., Rhee, S., & Hutchinson, R. C. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213624 · Cited by 466
Signal quality indices and data fusion for determining clinical acceptability of electrocardiograms
Clifford, G. D., Behar, J., Li, Q., & Rezek, I. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/9/1419 · Cited by 292
Dynamic time warping and machine learning for signal quality assessment of pulsatile signals
Li, Q., & Clifford, G. D. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/9/1491 · Cited by 274 · cited 1×
“…that there are multiple different signal contaminants, and that different predictors may be specific to as few as one of these contaminants, there has been a valuable effort to fuse multiple signal quality predictors in to a single quality metric (Clifford et al 2012, Li and Clifford 2012).”
BackgroundArtifact-resistant power-efficient design of finger-ring plethysmographic sensors
Rhee, S., Yang, B.-H., & Asada, H. H. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.930904 · Cited by 235
The Light–Tissue Interaction of Pulse Oximetry
Mannheimer, P. D. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000269522.84942.54 · Cited by 194
Signal quality measures for pulse oximetry through waveform morphology analysis
Sukor, J. A., Redmond, S. J., & Lovell, N. H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/3/008 · Cited by 166
Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation
Karlen, W., Kobayashi, K., Ansermino, J. M., & Dumont, G. A. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/10/1617 · Cited by 123
Cited by
Wearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250
Evaluation of the signal quality of wrist-based photoplethysmography
Pradhan, N., Rajan, S., & Adler, A. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab225a · Cited by 50
“Morris and Wander (2014) used a method combining segmenting and non-segmenting approaches to derive features for classification.”
MethodComparison of Wearable and Clinical Devices for Acquisition of Peripheral Nervous System Signals
Bizzego, A., Gabrieli, G., Furlanello, C., & Esposito, G. (2020) · Sensors · doi:10.3390/s20236778 · Cited by 25
Paliakaitė, B., Petrėnas, A., Sološenko, A., & Marozas, V. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102421 · Cited by 24
Photoplethysmogram Signal Quality Evaluation by Unsupervised Learning Approach
Roy, M. S., Gupta, R., & Sharma, K. D. (2020) · 2020 IEEE Applied Signal Processing Conference (ASPCON) · doi:10.1109/aspcon49795.2020.9276733 · Cited by 24
Metadata
- Authors
- Wander, J. D., & Morris, D.
- Venue
- Physiological Measurement
- Publisher
- IOP Publishing
- Published
- 2014-11-19
- Volume
- 35
- Issue
- 12
- Pages
- 2543-2561
- ISSN
- 0967-3334 (ISSN-L)
- 1361-6579
- OpenAlex
- W2023227878
- Cited by
- 22
- References
- 12
- FWCI
- 0.50
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A combined segmenting and non-segmenting approach to signal quality estimation for ambulatory photoplethysmography. https://app.cochat.ai/papers/10.1088/0967-3334/35/12/2543
- The paper
- Wander, J. D., & Morris, D. (2014). A combined segmenting and non-segmenting approach to signal quality estimation for ambulatory photoplethysmography. Physiological Measurement. https://doi.org/10.1088/0967-3334/35/12/2543
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.