Diagnostic reference levels for common computed tomography (CT) examinations: results from the first Nigerian nationwide dose survey
Abstract
Purpose . To explore doses from common adult computed tomography (CT) examinations and propose national diagnostic reference levels (nDRLs) for Nigeria. Materials and Methods . This retrospective study was approved by the Nnamdi Azikiwe University and University Teaching Hospital Institutional Review Boards (IRB: NAUTH/CS/66/Vol8/84) and involved dose surveys of adult CT examinations across the six geographical regions of Nigeria and Abuja from January 2016 to August 2017. Dose data of adult head, chest and abdomen/pelvis CT examinations were extracted from patient folders. The median, 75th and 25th percentile CT dose index volume (CTDI vol ) and dose-length-product (DLP) were computed for each of these procedures. Effective doses (E) for these examinations were estimated using the k conversion factor as described in the ICRP publication 103 (E DLP = k × DLP ). Results . The proposed 75th percentile CTDI vol for head, chest, and abdomen/pelvis are 61 mGy, 17 mGy, and 20 mGy, respectively. The corresponding DLPs are 1310 mGy.cm, 735 mGy.cm, and 1486 mGy.cm respectively. The effective doses were 2.75 mSv (head), 10.29 mSv (chest), and 22.29 mSv (abdomen/pelvis). Conclusion . Findings demonstrate wide dose variations within and across centres in Nigeria. The results also show CTDI vol comparable to international standards, but considerably higher DLP and effective doses.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The primary factors influencing CT dose variation are multi-phase scanning and institutional protocol choices.
“Other scholars have opined that CT scan doses vary considerably within and across facilities; the primary factors influencing dose variation are multi-phase scanning and institutional protocol choices.”
From Adult Computed Tomography examinations in Uganda: Towards determining the National Diagnostic Reference Levels (2022) · cites this paper as [11]
CT scan doses vary considerably within and across facilities.
“Other scholars have opined that CT scan doses vary considerably within and across facilities; the primary factors influencing dose variation are multi-phase scanning and institutional protocol choices.”
From Adult Computed Tomography examinations in Uganda: Towards determining the National Diagnostic Reference Levels (2022) · cites this paper as [11]
Cites
Computed Tomography — An Increasing Source of Radiation Exposure
Brenner, D. J., & Hall, E. J. (2007) · New England Journal of Medicine · doi:10.1056/nejmra072149 · Cited by 8,738 · see citing works
Smith‐Bindman, R. (2009) · Archives of Internal Medicine · doi:10.1001/archinternmed.2009.427 · Cited by 2,467 · see citing works
Effective Doses in Radiology and Diagnostic Nuclear Medicine: A Catalog
Mettler, F. A., Huda, W., Yoshizumi, T. T., & Mahesh, M. (2008) · Radiology · doi:10.1148/radiol.2481071451 · Cited by 2,007 · see citing works
ICRP Publication 103, The Recommendations of the International Commission on Radiological Protection
Vetter, R. J. (2008) · Health Physics · doi:10.1097/01.hp.0000324200.73903.5b · Cited by 1,263 · see citing works
Sodickson, A. D., Baeyens, P. F., Andriole, K. P., Prevedello, L. M. S., Nawfel, R. D., Hanson, R., & Khorasani, R. (2009) · Radiology · doi:10.1148/radiol.2511081296 · Cited by 993 · see citing works
Silva, A. C., Lawder, H. J., Hara, A. K., Kujak, J. L., & Pavlicek, W. (2009) · American Journal of Roentgenology · doi:10.2214/ajr.09.2953 · Cited by 561 · see citing works
Radiation Risks of Medical Imaging: Separating Fact from Fantasy
Hendee, W. R., & O’Connor, M. K. (2012) · Radiology · doi:10.1148/radiol.12112678 · Cited by 354 · see citing works
Volume CT Dose Index and Dose-Length Product Displayed during CT: What Good Are They?
Huda, W., & Mettler, F. A. (2010) · Radiology · doi:10.1148/radiol.10100297 · Cited by 209 · see citing works
Establishment of CT diagnostic reference levels in Ireland
Foley, S. J., McEntee, M. F., & Rainford, L. A. (2012) · The British Journal of Radiology · doi:10.1259/bjr/15839549 · Cited by 186 · see citing works
Treier, R., Aroua, A., Verdun, F. R., et al. (2010) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncq279 · Cited by 143 · see citing works
Radiation Doses in Consecutive CT Examinations from Five University of California Medical Centers
Smith‐Bindman, R., Moghadassi, M., Wilson, N., et al. (2015) · Radiology · doi:10.1148/radiol.2015142728 · Cited by 127 · see citing works
Establishing national diagnostic reference levels (DRLs) for computed tomography in Egypt
Salama, D. H., Vassileva, J., Mahdaly, G., Shawki, M., Salama, A., Gilley, D., & Rehani, M. M. (2017) · Physica Medica · doi:10.1016/j.ejmp.2017.05.050 · Cited by 98 · see citing works
Han, B., Grant, K. L. R., Garberich, R., Sedlmair, M., Lindberg, J. L., & Lesser, J. R. (2012) · Journal of cardiovascular computed tomography · doi:10.1016/j.jcct.2012.04.008 · Cited by 90 · see citing works
Adult exposures from MDCT including multiphase studies: first Italian nationwide survey
Palorini, F., Origgi, D., Granata, C., et al. (2013) · European Radiology · doi:10.1007/s00330-013-3031-7 · Cited by 81 · see citing works
Klink, T., Obmann, V. C., Heverhagen, J. T., et al. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2014.05.033 · Cited by 70 · see citing works
The establishment of computed tomography diagnostic reference levels in Portugal
Santos, J., Foley, S. J., Paulo, G., et al. (2013) · Radiation Protection Dosimetry · doi:10.1093/rpd/nct226 · Cited by 64 · see citing works
Variations in radiation dose between the same model of multislice CT scanner at different hospitals
Koller, C. J., Eatough, J. P., & Bettridge, A. (2003) · British Journal of Radiology · doi:10.1259/bjr/33117342 · Cited by 57 · see citing works
Diagnostic reference level of computed tomography (CT) in japan
Fukushima, Y., Tsushima, Y., Takei, H., Taketomi-Takahashi, A., Otake, H., & Endo, K. (2011) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncr441 · Cited by 53 · see citing works
NATIONAL DIAGNOSTIC REFERENCE LEVEL INITIATIVE FOR COMPUTED TOMOGRAPHY EXAMINATIONS IN KENYA
Korir, G. K., Wambani, J. S., Korir, I. K., Tries, M. A., & Boen, P. K. (2015) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncv020 · Cited by 52 · see citing works
Patient doses from CT examinations in Turkey
Ataç, G. K., Parmaksız, A., İnal, T., et al. (2015) · Diagnostic and Interventional Radiology · doi:10.5152/dir.2015.14306 · Cited by 44 · see citing works
U.S. National Diagnostic Reference Levels: Closing the Gap
Brink, J. A., & Miller, D. L. (2015) · Radiology · doi:10.1148/radiol.2015150971 · Cited by 42 · see citing works
Optimisation of direct digital chest radiography using Cu filtration
Ekpo, E. U., Hoban, A. C., & McEntee, M. F. (2014) · Radiography · doi:10.1016/j.radi.2014.07.001 · Cited by 39 · see citing works
Self-citationTack, D., Jahnen, A., Kohler, S., et al. (2013) · European Radiology · doi:10.1007/s00330-013-2994-8 · Cited by 35 · see citing works
Radiation exposure among patients with the highest CT scan utilization in the emergency department
Shah, K. H., Slovis, B. H., Runde, D. P., et al. (2013) · Emergency Radiology · doi:10.1007/s10140-013-1142-8 · Cited by 35 · see citing works
Diagnostic reference levels and patient doses in computed tomography examinations in Greece
Simantirakis, G., Hourdakis, C. J., Economides, S., et al. (2014) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncu182 · Cited by 34 · see citing works
Govindarajan, K., Saravanakumar, A., Vaideki, K., et al. (2014) · Journal of Medical Physics · doi:10.4103/0971-6203.125509 · Cited by 32 · see citing works
Nie, P., Li, H., Duan, Y., et al. (2014) · PLoS ONE · doi:10.1371/journal.pone.0091123 · Cited by 30 · see citing works
Computed tomography dose index for head CT in northern Nigeria
Garba, I., Engel-Hills, P. C., Davidson, F., et al. (2015) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncv188 · Cited by 23 · see citing works
Pediatric organ dose measurements in axial and helical multislice CT
McDermott, A., White, R. A., Mc‐Nitt‐Gray, M., et al. (2009) · Medical Physics · doi:10.1118/1.3101817 · Cited by 18 · see citing works
Head computed tomography: Dose output and relationship with anthropotechnical parameters
Adejoh, T., & Nzotta, C. C. (2016) · West African journal of radiology · doi:10.4103/1115-3474.179256 · Cited by 14 · see citing works
Self-citationDetermination of Computed Tomography Diagnostic Reference Levels in North-Central Nigeria
Abdulkadir, M. K. (2018) · doi:10.1594/ecr2018/c-0957 · Cited by 12 · see citing works
Diagnostic reference levels for computed tomography of the head in Anambra State of Nigeria
Adejoh, T., Nzotta, C., Aronu, M., et al. (2017) · West African journal of radiology · doi:10.4103/1115-3474.206806 · Cited by 10 · see citing works
Self-citationChallenges of radiological equipment management policies in some northern Nigerian hospitals
Ekpo, E. U., Egbe, N., Inyang, S. O., et al. (2013) · South African Radiographer · Cited by 9 · see citing works
Self-citationOperator eye doses during computed tomography fluoroscopic lung biopsy
Ekpo, E. U., Bakhshi, S., Ryan, E. A., Hogg, P., & McEntee, M. F. (2016) · Journal of Radiological Protection · doi:10.1088/0952-4746/36/2/290 · Cited by 8 · see citing works
Self-citationChallenges in setting up quality control in diagnostic radiology facilities in Nigeria
Inyang, S., Egbe, N., & Ekpo, E. U. (2015) · Nigerian Journal of Medicine · doi:10.4103/1115-2613.278950 · Cited by 5 · see citing works
Self-citationThe impact of diagnostic reference levels on patient doses from x-ray examinations
Leitz, W., & Almén, A. (2010) · Cited by 1 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Erem, G., Ameda, F., Otike, C., Olwit, W., Mubuuke, A. G., Schandorf, C., Kisolo, A., & Kawooya, M. G. (2022) · BMC Medical Imaging · doi:10.1186/s12880-022-00838-x · Cited by 22 · see citing works
“It is however not clear whether these are the same factors in play in resourcelimited settings like Uganda [11].”
BackgroundKumsa, M. J., Nguse, T. M., Ambessa, H. B., Gele, T. T., Fantaye, W. G., & Dellie, S. T. (2023) · BMC Medical Imaging · doi:10.1186/s12880-023-00963-1 · Cited by 18 · see citing works
Alhujaili, S. F., Alshabibi, A. S., Alafer, F., Alrowaili, Z., Salah, H., Sulieman, A., Subahi, M., Al-Raddadi, K., & Abuzaid, M. (2024) · Diagnostics · doi:10.3390/diagnostics14242882 · Cited by 7 · see citing works
- Authors
- Ekpo, E. U., Adejoh, T., Akwo, J. D., Emeka, O. C., Modu, A. A., Abba, M., Adesina, K. A., Omiyi, D. O., & Chiegwu, U. H.
- Affiliations
- The University of Sydney (AU)
- Nnamdi Azikiwe University (NG)
- University of Calabar (NG)
- National Hospital Abuja (NG)
- Bayero University Kano (NG)
- University of Lagos (NG)
- Corresponding author
- Ekpo, Ernest U
- Venue
- Journal of Radiological Protection
- Publisher
- IOP Publishing
- Published
- 2018-03-13
- Volume
- 38
- Issue
- 2
- Pages
- 525-535
- ISSN
- 0952-4746 (ISSN-L)
- 1361-6498
- OpenAlex
- W2793265424
- Cited by
- 61
- References
- 36
- Citations by year
- 2026: 2 · 2025: 7 · 2024: 8 · 2023: 5 · 2022: 11 · 2021: 13 · 2020: 10 · 2019: 4 · 2018: 1
- FWCI
- 4.83
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- Locations
- Journal of Radiological Protection (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiology practices and education
- Keywords
- diagnostic reference levels, computed tomography, radiation dose, volume CT dose index, dose-length product, effective dose, head CT, chest CT, abdomen-pelvis CT, Nigeria
- MeSH terms
- Adult, Aged, Aged, 80 and over, Female, Humans, Male, Middle Aged, Nigeria, Radiation Dosage, Reference Values, Retrospective Studies, Tomography, X-Ray Computed, Health Care Surveys, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review