The influence of electrolyte composition on the in vitro charge-injection limits of activated iridium oxide (AIROF) stimulation electrodes
The DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.
Abstract
The effects of ionic conductivity and buffer concentration of electrolytes used for in vitro measurement of the charge-injection limits of activated iridium oxide (AIROF) neural stimulation electrodes have been investigated. Charge-injection limits of AIROF microelectrodes were measured in saline with a range of phosphate buffer concentrations from [PO(4)(3-)] = 0 to [PO(4)(3-)] = 103 mM and ionic conductivities from 2-28 mS cm(-1). The charge-injection limits were insensitive to the buffer concentration, but varied significantly with ionic conductivity. Using 0.4 ms cathodal current pulses at 50 Hz, the charge-injection limit increased from 0.5 mC cm(-2) to 2.1 mC cm(-2) as the conductivity was increased from 2 mS cm(-1) to 28 mS cm(-1). An explanation is proposed in which the observed dependence on ionic conductivity arises from non-uniform reduction and oxidation within the porous AIROF and from uncorrected iR-drops that result in an overestimation of the redox potential during pulsing. Conversely, slow-sweep-rate cyclic voltammograms (CVs) were sensitive to buffer concentration with the potentials of the primary Ir(3+)/Ir(4+) reduction and oxidation reactions shifting approximately 300 mV as the buffer concentration decreased from [PO(4)(3-)] = 103 mM to [PO(4)(3-)] = 0 mM. The CV response was insensitive to ionic conductivity. A comparison of in vitro AIROF charge-injection limits in commonly employed electrolyte models of extracellular fluid revealed a significant dependence on the electrolyte, with more than a factor of 4 difference under some pulsing conditions, emphasizing the need to select an electrolyte model that closely matches the conductivity and ionic composition of the in vivo environment.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · cited 1×
“The resulting high charge and current densities argue against the use of noble metal electrodes in these prostheses because of electrode dissolution or tissue damage that may occur when charge densities exceed the maximum charge-injection limits of the metal (Merrill et al 2005).”
BackgroundSchmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672 · cited 1×
“Prosthetic vision using electrodes implanted in the cortex has been investigated by several groups (Margalit et al 2002, Schmidt et al 1996, Normann et al 1999, Troyk et al 2003) and extensive studies with retinal stimulation are ongoing (Humayun et al 1999, Rizzo et al 2003a, 2003b, Weiland et al…”
MethodRizzo, J. F., Wyatt, J., Loewenstein, J., Kelly, S., & Shire, D. (2003) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.02-0817 · Cited by 478 · cited 1×
“…implanted in the cortex has been investigated by several groups (Margalit et al 2002, Schmidt et al 1996, Normann et al 1999, Troyk et al 2003) and extensive studies with retinal stimulation are ongoing (Humayun et al 1999, Rizzo et al 2003a, 2003b, Weiland et al 1999, Mahadevappa et al 2005).”
BackgroundElectrodeposited iridium oxide for neural stimulation and recording electrodes
Meyer, R. D., Cogan, S. F., Nguyen, T. H., & Rauh, R. D. (2001) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/7333.918271 · Cited by 255 · cited 14×
“At the 1:8 tc:ta ratio, the maximum Qinj values were 2.3 ± 0.8 mC cm−2, 3.5 ± 0.5 mC cm−2 and 4.7 ± 0.6 mC cm−2 (n = 5, mean ± S.D.) for the ISF, CBS/PBS and PBS, respectively.”
BackgroundMethodInfluentialSelf-citationOver-pulsing degrades activated iridium oxide films used for intracortical neural stimulation
Cogan, S. F., Guzelian, A. A., Agnew, W. F., Yuen, T. G. H., & McCreery, D. B. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.02.019 · Cited by 210
Self-citationA model for intracortical visual prosthesis research
Troyk, P., Bak, M., Berg, J., Bradley, D., Cogan, S., Erickson, R., Kufta, C., McCreery, D., Schmidt, E., & Towle, V. (2003) · Artificial Organs · doi:10.1046/j.1525-1594.2003.07308.x · Cited by 166 · cited 1×
“…using electrodes implanted in the cortex has been investigated by several groups (Margalit et al 2002, Schmidt et al 1996, Normann et al 1999, Troyk et al 2003) and extensive studies with retinal stimulation are ongoing (Humayun et al 1999, Rizzo et al 2003a, 2003b, Weiland et al 1999,…”
MethodSelf-citationCogan, S. F., Troyk, P. R., Ehrlich, J., & Plante, T. D. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.851503 · Cited by 162 · cited 3×
“To permit higher levels of charge injection, activated iridium oxide films (AIROFs) that provide at least a factor of 5 increase in capacity compared with platinum are often employed (Beebe and Rose 1988, Rose and Robblee 1990, Cogan et al 2005).”
BackgroundMethodSelf-citationCogan, S. F., Troyk, P. R., Ehrlich, J., Plante, T. D., & Detlefsen, D. E. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.862572 · Cited by 138 · cited 3×
“The basis for the increased cathodal charge-injection capacity with positively biased AIROF using a low current, long duration anodal phase has been discussed by Cogan et al (2006).”
BackgroundMethodInfluentialSelf-citationJezernik, S., Craggs, M., Grill, W. M., Creasey, G., & Rijkhoff, N. J. M. (2002) · Neurological Research · doi:10.1179/016164102101200294 · Cited by 102 · cited 1×
“Intraspinal electrodes are showing promise for bladder management in spinal cord injury and may find use in managing other dysfunctions of the genitourinary system (Jezernik et al 2002, Grill et al 1999).”
Background
Cited by
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243
Self-citationA transverse intrafascicular multichannel electrode (TIME) to interface with the peripheral nerve
Boretius, T., Badia, J., Pascual-Font, A., Schuettler, M., Navarro, X., Yoshida, K., & Stieglitz, T. (2010) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.05.010 · Cited by 468
In Vitro and In Vivo Evaluation of PEDOT Microelectrodes for Neural Stimulation and Recording
Venkatraman, S., Hendricks, J., King, Z. A., Sereno, A. J., Richardson-Burns, S., Martin, D., & Carmena, J. M. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2109399 · Cited by 319
“The charge-injection limit is defined [29] as the quantity of charge which polarizes the electrode interface to the potential for water reduction or oxidation.”
BackgroundBoehler, C., Carli, S., Fadiga, L., Stieglitz, T., & Asplund, M. (2020) · Nature Protocols · doi:10.1038/s41596-020-0389-2 · Cited by 308
Improving the spatial resolution of epiretinal implants by increasing stimulus pulse duration
Weitz, A. C., Nanduri, D., Behrend, M. R., Gonzalez-Calle, A., Greenberg, R. J., Humayun, M. S., Chow, R. H., & Weiland, J. D. (2015) · Science Translational Medicine · doi:10.1126/scitranslmed.aac4877 · Cited by 154
“6 V, the voltage at which H2 irreversibly evolves (49).”
Background
Metadata
- Authors
- Cogan, S. F., Troyk, P. R., Ehrlich, J., Gasbarro, C. M., & Plante, T. D.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2007-03-08
- Volume
- 4
- Issue
- 2
- Pages
- 79-86
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2031272609
- Cited by
- 88
- References
- 9
- FWCI
- 1.57
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The influence of electrolyte composition on the in vitro charge-injection limits of activated iridium oxide (AIROF) stimulation electrodes. https://app.cochat.ai/papers/10.1088/1741-2560/4/2/008
- The paper
- Cogan, S. F., Troyk, P. R., Ehrlich, J., Gasbarro, C. M., & Plante, T. D. (2007). The influence of electrolyte composition on the in vitro charge-injection limits of activated iridium oxide (AIROF) stimulation electrodes. Journal of Neural Engineering. https://doi.org/10.1088/1741-2560/4/2/008
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.