In Vitro Comparison of the Charge-Injection Limits of Activated Iridium Oxide (AIROF) and Platinum-Iridium Microelectrodes

Cogan, S.F., Troyk, P.R., Ehrlich, J., Plante, T.D.·2005·IEEE Transactions on Biomedical Engineering

Abstract

The charge-injection limits of activated iridium oxide electrodes (AIROF) and PtIr microelectrodes with similar geometric area and shape have been compared in vitro using a stimulation waveform that delivers cathodal current pulses with current-limited control of the electrode bias potential in the interpulse period. Charge-injection limits were compared over a bias range of 0.1-0.7 V (versus Ag/AgCl) and pulse frequencies of 20, 50, and 100 Hz. The AIROF was capable of injecting between 4 and 10 times the charge of the PtIr electrode, with a maximum value of 3.9 mC/cm2 obtained at a 0.7 V bias and 20 Hz frequency.

Integrity signals

Registry

Crossref · 10.1109/tbme.2005.851503

Cites

8 of 8 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Cogan, S. F., Troyk, P. R., Ehrlich, J., & Plante, T. D.
Affiliations
  • EIC Laboratories (US)
  • Illinois Institute of Technology (US)
Venue
IEEE Transactions on Biomedical Engineering
Published
2005-09
Volume
52
Issue
9
Pages
1612-1614
ISSN
0018-9294 (ISSN-L)
OpenAlex
W2159277823
Cited by
162
References
8
Citations by year
2026: 3 · 2025: 2 · 2024: 9 · 2023: 10 · 2022: 6 · 2021: 8 · 2020: 8 · 2019: 11 · 2018: 7 · 2017: 11 · 2016: 7 · 2015: 9 · 2014: 9 · 2013: 10 · 2012: 5
FWCI
1.93
Type
journal article
Language
en
Open access
closed
Locations
  • IEEE Transactions on Biomedical Engineering (publishedVersion)
  • PubMed (publishedVersion)
Funders
  • National Institutes of Health (5r44ns039714-03, 5r01eb002184-04)
  • National Institute of Neurological Disorders and Stroke (R44 NS039714-03)
  • National Institute of Biomedical Imaging and Bioengineering (R01 EB002184-02)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
Keywords
charge injection limit, neurostimulation, stimulation frequency
MeSH terms
Electric Stimulation, Equipment Design, Iridium, Microelectrodes, Electric Impedance, Platinum Compounds, Equipment Failure Analysis, Static Electricity

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review