Over-pulsing degrades activated iridium oxide films used for intracortical neural stimulation
Abstract
Microelectrodes using activated iridium oxide (AIROF) charge-injection coatings have been pulsed in cat cortex at levels from near-threshold for neural excitation to the reported in vitro electrochemical charge-injection limits of AIROF. The microelectrodes were subjected to continuous biphasic current pulsing, using an 0.4V (versus Ag|AgCl) anodic bias with equal cathodal and anodal pulse widths, for periods up to 7h at a frequency of either 50Hz or 100Hz. At charge densities of 3mC/cm(2), histology revealed iridium-containing deposits in tissue adjacent to the charge-injection sites and scanning electron microscopy of explanted electrodes revealed a thickened and poorly adherent AIROF coating. Microelectrodes pulsed at 2mC/cm(2) or less remained intact, with no histologic evidence of non-biologic deposits in the tissue. AIROF microelectrodes challenged in vitro under the same pulsing conditions responded similarly, with electrodes pulsed at 3mC/cm(2) showing evidence of AIROF delamination after only 100s of pulsing at 100Hz (10,000 pulses total), while electrodes pulsed at 2mC/cm(2) for 7h at 50Hz (1.3 x 10(6) pulses total) showed no evidence of damage. In vitro electrochemical potential transient measurements in buffered physiologic saline indicate that polarizing the AIROF beyond the potential window for electrolysis of water (-0.6 to 0.8V versus Ag|AgCl) results in the observed degradation.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Activated iridium oxide films have uneven growth of the oxide layer that introduces strain and can result in a tendency to delaminate from the underlying iridium metal.
“For activated iridium oxide films, uneven growth of the oxide layer introduces strain in the oxide layer and can result in the tendency for the activated iridium oxide film to delaminate from the underlying iridium metal [12] [13].”
From Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer (2009) · cites this paper as [13]
The sputtered iridium oxide film on LCP was robust and could provide a good solution for stimulation electrodes that require a flexible substrate and large charge injection capability.
“Finally, the sputtered iridium oxide film on LCP was robust and could provide a good solution for stimulation electrodes that require a flexible substrate and large charge injection capability [12] [13].”
From Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer (2009) · cites this paper as [13]
Cites
Schmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672 · see citing works
In vitro electrical properties for iridium oxide versus titanium nitride stimulating electrodes
Weiland, J. D., Anderson, D. J., & Humayun, M. S. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2002.805487 · Cited by 382 · see citing works
Rose, T. L., & Robblee, L. S. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.61038 · Cited by 359 · see citing works
Neural prostheses : fundamental studies
Agnew, W. F., & McCreery, D. B. (1990) · Medical Entomology and Zoology · Cited by 316 · see citing works
Self-citationVisual sensations produced by intracortical microstimulation of the human occipital cortex
Bak, M., Girvin, J. P., Hambrecht, F. T., Kufta, C. V., Loeb, G. E., & Schmidt, E. M. (1990) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442682 · Cited by 299 · see citing works
Liu, X., McCreery, D. B., Carter, R. R., Bullara, L. A., Yuen, T. G. H., & Agnew, W. F. (1999) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.788468 · Cited by 272 · see citing works
Self-citationChronic neural stimulation with thin-film, iridium oxide electrodes
Weiland, J. D., & Anderson, D. J. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.846685 · Cited by 263 · see citing works
Batch fabricated thin-film electrodes for stimulation of the central auditory system
Anderson, D. J., Najafi, K., Tanghe, S., Evans, D. A., Levy, K., Hetke, J. F., Xue, X., Zappia, J. J., & Wise, K. D. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.32101 · Cited by 229 · see citing works
Activated Ir: An Electrode Suitable for Reversible Charge Injection in Saline Solution
Robblee, L. S., Lefko, J. L., & Brummer, S. B. (1983) · Journal of The Electrochemical Society · doi:10.1149/1.2119793 · Cited by 228 · see citing works
Beebe, X., & Rose, T. L. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.2122 · Cited by 220 · see citing works
Mozota, J., & Conway, B. E. (1983) · Electrochimica Acta · doi:10.1016/0013-4686(83)85079-8 · Cited by 204 · see citing works
Histopathologic evaluation of prolonged intracortical electrical stimulation
Agnew, W. F., Yuen, T. G. H., McCreery, D. B., & Bullara, L. A. (1986) · Experimental Neurology · doi:10.1016/0014-4886(86)90132-9 · Cited by 173 · see citing works
Self-citationA model for anodic hydrous oxide growth at iridium
Pickup, P. G., & Birss, V. I. (1987) · Journal of Electroanalytical Chemistry · doi:10.1016/0022-0728(87)88006-3 · Cited by 138 · see citing works
McCreery, D. B., Yuen, T. G. H., Agnew, W. F., & Bullara, L. A. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.634645 · Cited by 130 · see citing works
Self-citationOxidation State Changes and Structure of Electrochromic Iridium Oxide Films
McIntyre, J. D. E., PECK, W. F. J., & NAKAHARA, S. (1980) · Journal of The Electrochemical Society · doi:10.1149/1.2129868 · Cited by 103 · see citing works
Jezernik, S., Craggs, M., Grill, W. M., Creasey, G., & Rijkhoff, N. J. M. (2002) · Neurological Research · doi:10.1179/016164102101200294 · Cited by 102 · see citing works
Chronic microstimulation in the feline ventral cochlear nucleus: physiologic and histologic effects
McCreery, D. B., Yuen, T. G. H., & Bullara, L. A. (2000) · Hearing Research · doi:10.1016/s0378-5955(00)00190-8 · Cited by 84 · see citing works
Self-citationBladder and urethral pressures evoked by microstimulation of the sacral spinal cord in cats
Grill, W. M., Bhadra, N., & Wang, B. (1999) · Brain Research · doi:10.1016/s0006-8993(99)01581-4 · Cited by 80 · see citing works
McCreery, D. B., Agnew, W. F., & Bullara, L. A. (2002) · Annals of Biomedical Engineering · doi:10.1114/1.1430748 · Cited by 77 · see citing works
Self-citationNeuronal activity evoked by chronically implanted intracortical microelectrodes
McCreery, D. B., Bullara, L. A., & Agnew, W. F. (1986) · Experimental Neurology · doi:10.1016/0014-4886(86)90131-7 · Cited by 67 · see citing works
Self-citationVisual prostheses based on direct interfaces with the visual system.
Hambrecht, F. T. (1995) · PubMed · Cited by 46 · see citing works
Calculations of the pH changes produced in body tissue by a spherical stimulation electrode
Ballestrasse, C. L., Ruggeri, R. T., & Beck, T. R. (1985) · Annals of Biomedical Engineering · doi:10.1007/bf02407769 · Cited by 39 · see citing works
Charge-injection waveforms for iridium oxide (AIROF) microelectrodes
Cogan, S. F., Troyk, P. R., Ehrlich, J., Plante, T. D., McCreery, D. B., & Bullara, L. A. (2004) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2003.1279829 · Cited by 18 · see citing works
Self-citationElectrical Characteristics of Chronically Implanted Platinum-Irdium Electrodes
Creery, D. B. M., Agnew, W. F., & Hardy, J. M. (1987) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1987.325990 · Cited by 15 · see citing works
Self-citationThe amazing adventures of robotrat
Nicolelis, M. A. L. (2002) · Trends in Cognitive Sciences · doi:10.1016/s1364-6613(02)01991-5 · Cited by 12 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · see citing works
Self-citationCarbon nanotube coating improves neuronal recordings
Keefer, E. W., Botterman, B. R., Romero, M. I., Rossi, A. F., & Gross, G. W. (2008) · Nature Nanotechnology · doi:10.1038/nnano.2008.174 · Cited by 721 · see citing works
Nanomaterials for Neural Interfaces
Kotov, N. A., Winter, J. O., Clements, I. P., Jan, E., Timko, B. P., Campidelli, S., Pathak, S., Mazzatenta, A., Lieber, C. M., Prato, M., et al. (2009) · Advanced Materials · doi:10.1002/adma.200801984 · Cited by 504 · see citing works
Photovoltaic retinal prosthesis with high pixel density
Mathieson, K., Loudin, J., Goetz, G. A., Huie, P., Wang, L., Kamins, T. I., Galambos, L., Smith, R., Harris, J. S., Sher, A., et al. (2012) · Nature Photonics · doi:10.1038/nphoton.2012.104 · Cited by 460 · see citing works
Neural Stimulation with a Carbon Nanotube Microelectrode Array
Wang, K., Fishman, H. A., Dai, H., & Harris, J. S. (2006) · Nano Letters · doi:10.1021/nl061241t · Cited by 416 · see citing works
A Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389 · see citing works
Tissue damage thresholds during therapeutic electrical stimulation
Cogan, S. F., Ludwig, K. A., Welle, C. G., & Takmakov, P. (2016) · Journal of Neural Engineering · doi:10.1088/1741-2560/13/2/021001 · Cited by 386 · see citing works
Self-citationCherevko, S., Geiger, S., Kasian, O. I., Mingers, A. M., & Mayrhofer, K. J. J. (2016) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2016.05.015 · Cited by 333 · see citing works
In Vitro and In Vivo Evaluation of PEDOT Microelectrodes for Neural Stimulation and Recording
Venkatraman, S., Hendricks, J., King, Z. A., Sereno, A. J., Richardson-Burns, S., Martin, D., & Carmena, J. M. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2109399 · Cited by 319 · see citing works
Neural Stimulation and Recording with Bidirectional, Soft Carbon Nanotube Fiber Microelectrodes
Vitale, F., Santacruz, S. R., Aazhang, B., Kemere, C. T., & Pasquali, M. (2015) · ACS Nano · doi:10.1021/acsnano.5b01060 · Cited by 313 · see citing works
Poly (3,4-Ethylenedioxythiophene) for Chronic Neural Stimulation
Cui, X. T., & Zhou, D. D. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.909811 · Cited by 278 · see citing works
Jovanovič, P., Hodnik, N., Ruiz‐Zepeda, F., Arčon, I., Jozinović, B., Zorko, M., Bele, M., Šala, M., Šelih, V. S., Hočevar, S. B., et al. (2017) · Journal of the American Chemical Society · doi:10.1021/jacs.7b08071 · Cited by 268 · see citing works
Highly stable carbon nanotube doped poly(3,4-ethylenedioxythiophene) for chronic neural stimulation
Luo, X., Weaver, C. L., Zhou, D. D., Greenberg, R. J., & Cui, X. T. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2011.04.051 · Cited by 255 · see citing works
Flexible Neural Electrode Array Based-on Porous Graphene for Cortical Microstimulation and Sensing
Lu, Y., Lyu, H., Richardson, A. G., Lucas, T. H., & Kuzum, D. (2016) · Scientific Reports · doi:10.1038/srep33526 · Cited by 213 · see citing works
Lu, Y., Wang, D., Li, T., Zhao, X., Cao, Y., Yang, H., & Duan, Y. Y. (2009) · Biomaterials · doi:10.1016/j.biomaterials.2009.04.030 · Cited by 210 · see citing works
High‐Performance Graphene‐Fiber‐Based Neural Recording Microelectrodes
Wang, K., Frewin, C. L., Esrafilzadeh, D., Yu, C., Wang, C., Pancrazio, J. J., Romero‐Ortega, M. I., Jalili, R., & Wallace, G. G. (2019) · Advanced Materials · doi:10.1002/adma.201805867 · Cited by 193 · see citing works
Electrodeposited polypyrrole/carbon nanotubes composite films electrodes for neural interfaces
Lu, Y., Li, T., Zhao, X., Li, M., Cao, Y., Yang, H., & Duan, Y. Y. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.03.022 · Cited by 191 · see citing works
Wilks, S. J. (2009) · Frontiers in Neuroengineering · doi:10.3389/neuro.16.007.2009 · Cited by 191 · see citing works
Sputtered iridium oxide films for neural stimulation electrodes
Cogan, S. F., Ehrlich, J., Plante, T. D., Smirnov, A., Shire, D. B., Gingerich, M. D., & Rizzo, J. F. (2008) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbm.b.31223 · Cited by 188 · see citing works
Self-citationSoft bioelectronics for the management of cardiovascular diseases
Sunwoo, S., Han, S. I., Park, C. S., Kim, J. H., Georgiou, J. S., Lee, S., Kim, D., & Hyeon, T. (2023) · Nature Reviews Bioengineering · doi:10.1038/s44222-023-00102-z · Cited by 176 · see citing works
Ultra-thin flexible polyimide neural probe embedded in a dissolvable maltose-coated microneedle
Xiang, Z., Yen, S.-C., Xue, N., Sun, T., Tsang, W. M., Zhang, S., Liao, L., Thakor, N. V., & Lee, C. (2014) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/24/6/065015 · Cited by 157 · see citing works
Impedance characteristics of deep brain stimulation electrodes in vitro and in vivo
Wei, X. F., & Grill, W. M. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/4/046008 · Cited by 156 · see citing works
Cogan, S. F., Troyk, P. R., Ehrlich, J., Plante, T. D., & Detlefsen, D. E. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.862572 · Cited by 138 · see citing works
Self-citationSputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes
Cogan, S. F., Plante, T. D., & Ehrlich, J. (2005) · The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2004.1404158 · Cited by 138 · see citing works
Self-citationTranslational opportunities and challenges of invasive electrodes for neural interfaces
Shen, K., Chen, O., Edmunds, J. L., Piech, D. K., & Maharbiz, M. M. (2023) · Nature Biomedical Engineering · doi:10.1038/s41551-023-01021-5 · Cited by 143 · see citing works
Optoelectronic retinal prosthesis: system design and performance
Loudin, J., Simanovskii, D. M., Vijayraghavan, K., Sramek, C., Butterwick, A., Huie, P., McLean, G. Y., & Palanker, D. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/1/s09 · Cited by 130 · see citing works
Ultra-nanocrystalline diamond electrodes: optimization towards neural stimulation applications
Garrett, D. J., Ganesan, K., Stacey, A. D., Fox, K. E., Hamish Meffin, A., & Prawer, S. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/1/016002 · Cited by 130 · see citing works
Duan, X., Tian-ming Fu, B., Jia Liu, B., & Lieber, C. M. (2013) · Nano Today · doi:10.1016/j.nantod.2013.05.001 · Cited by 125 · see citing works
Lee, J. H., Kim, H., Kim, J. H., & Lee, S. (2016) · Lab on a Chip · doi:10.1039/c5lc00842e · Cited by 124 · see citing works
Layered Carbon Nanotube-Polyelectrolyte Electrodes Outperform Traditional Neural Interface Materials
Jan, E., Hendricks, J. L., Husaini, V., Richardson-Burns, S. M., Sereno, A. J., Martin, D. C., & Kotov, N. A. (2009) · Nano Letters · doi:10.1021/nl902187z · Cited by 125 · see citing works
Hadjinicolaou, A. E., Leung, R. T., Garrett, D. J., Ganesan, K., Fox, K. E., Nayagam, D. A. X., Shivdasani, M. N., Hamish Meffin, A., Ibbotson, M. R., Prawer, S., et al. (2012) · Biomaterials · doi:10.1016/j.biomaterials.2012.04.063 · Cited by 124 · see citing works
Guidelines to Study and Develop Soft Electrode Systems for Neural Stimulation
Schiavone, G., Kang, X., Fallegger, F., Gandar, J., Courtine, G., & Lacour, S. P. (2020) · Neuron · doi:10.1016/j.neuron.2020.10.010 · Cited by 119 · see citing works
Eles, J. R., Vazquez, A. L., Snyder, N. R., Lagenaur, C. F., Murphy, M. C., Kozai, T. D. Y., & Cui, X. T. (2016) · Biomaterials · doi:10.1016/j.biomaterials.2016.10.054 · Cited by 116 · see citing works
Changes of Postoperative Impedances in Cochlear Implant Patients
Paasche, G., Bockel, F., Tasche, C., Lesinski‐Schiedat, A., & Lenarz, T. (2006) · Otology & Neurotology · doi:10.1097/01.mao.0000227662.88840.61 · Cited by 116 · see citing works
Graphene oxide doped conducting polymer nanocomposite film for electrode-tissue interface
Tian, H., Liu, J., Wei, D., Kang, X., Zhang, C., Du, J., Yang, B. T., Chen, X., Zhu, H., NuLi, Y., et al. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.11.058 · Cited by 115 · see citing works
Timko, B. P., Cohen-Karni, T., Qing, Q., Tian, B., & Lieber, C. M. (2010) · IEEE Transactions on Nanotechnology · doi:10.1109/tnano.2009.2031807 · Cited by 114 · see citing works
Conducting polymer coated neural recording electrodes
Harris, A. R., Morgan, S. J., Chen, J., Kapsa, R. M. I., Wallace, G. G., & Paolini, A. G. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/1/016004 · Cited by 112 · see citing works
Retinal prostheses: current challenges and future outlook
Winter, J. O., Cogan, S. F., & Rizzo, J. F. (2007) · Journal of Biomaterials Science Polymer Edition · doi:10.1163/156856207781494403 · Cited by 109 · see citing works
Self-citationElectrical Stimuli in the Central Nervous System Microenvironment
Thompson, D. M., Koppes, A. N., Hardy, J. G., & Schmidt, C. E. (2014) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-121813-120655 · Cited by 107 · see citing works
A Review: Electrode and Packaging Materials for Neurophysiology Recording Implants
Yang, W., Gong, Y. P., & Li, W. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2020.622923 · Cited by 107 · see citing works
Lu, Y., Wang, T., Cai, Z., Cao, Y., Yang, H., & Duan, Y. Y. (2008) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2008.11.036 · Cited by 98 · see citing works
Strength–Duration Relationship for Extracellular Neural Stimulation: Numerical and Analytical Models
Boinagrov, D., Loudin, J., & Palanker, D. (2010) · Journal of Neurophysiology · doi:10.1152/jn.00343.2010 · Cited by 96 · see citing works
Zhou, H., Cheng, X., Rao, L., Li, T., & Duan, Y. Y. (2013) · Acta Biomaterialia · doi:10.1016/j.actbio.2013.01.042 · Cited by 93 · see citing works
Conducting polymers for neuronal microelectrode array recording and stimulation
Aqrawe, Z., Montgomery, J. M., Travaš‐Sejdić, J., & Svirskis, D. M. (2017) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2017.11.023 · Cited by 92 · see citing works
Davis, T. S., Parker, R. A., House, P. A., Bagley, E. E., Wendelken, S., Normann, R. A., & Greger, B. E. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/6/065003 · Cited by 91 · see citing works
Eles, J. R., Vazquez, A. L., Kozai, T. D. Y., & Cui, X. T. (2018) · Biomaterials · doi:10.1016/j.biomaterials.2018.04.043 · Cited by 85 · see citing works
Nimbalkar, S., Castagnola, E., Balasubramani, A., Scarpellini, A., Samejima, S., Khorasani, A., Boissenin, A., Thongpang, S., Moritz, C. T., & Kassegne, S. (2018) · Scientific Reports · doi:10.1038/s41598-018-25198-x · Cited by 88 · see citing works
Cogan, S. F., Troyk, P. R., Ehrlich, J., Gasbarro, C. M., & Plante, T. D. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/2/008 · Cited by 88 · see citing works
Self-citationAchieving long-term stability of thin-film electrodes for neurostimulation
Oldroyd, P., & Malliaras, G. G. (2021) · Acta Biomaterialia · doi:10.1016/j.actbio.2021.05.004 · Cited by 87 · see citing works
Electrode Materials for Chronic Electrical Microstimulation
Zheng, X. S., Tan, C., Castagnola, E., & Cui, X. T. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100119 · Cited by 86 · see citing works
Improving Cochlear Implant Properties Through Conductive Hydrogel Coatings
Hassarati, R. T., Dueck, W. F., Tasche, C., Carter, P. M., Poole-Warren, L. A., & Green, R. A. (2014) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2014.2304559 · Cited by 81 · see citing works
Nanoporous Pt Microelectrode for Neural Stimulation and Recording: In Vitro Characterization
Park, S., Song, Y. J., Boo, H., & Chung, T. D. (2010) · The Journal of Physical Chemistry C · doi:10.1021/jp911256h · Cited by 79 · see citing works
Lycke, R., Kim, R., Zolotavin, P., Montes, J., Sun, Y., Kőszeghy, Á., Altun, E., Noble, B., Yin, R., He, F., et al. (2023) · Cell Reports · doi:10.1016/j.celrep.2023.112554 · Cited by 76 · see citing works
Recent advances in materials and flexible electronics for peripheral nerve interfaces
Bettinger, C. J. (2018) · Bioelectronic Medicine · doi:10.1186/s42234-018-0007-6 · Cited by 77 · see citing works
A New Implantable Closed-Loop Clinical Neural Interface: First Application in Parkinson’s Disease
Arlotti, M., Colombo, M., Bonfanti, A. G., Mandat, T. S., Lanotte, M., Pirola, E., Borellini, L., Rampini, P. M., Eleopra, R., Rinaldo, S., et al. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.763235 · Cited by 77 · see citing works
Superior Electrochemical Performance of Carbon Nanotubes Directly Grown on Sharp Microelectrodes
Ansaldo, A., Castagnola, E., Maggiolini, E., Fadiga, L., & Ricci, D. (2011) · ACS Nano · doi:10.1021/nn103445d · Cited by 75 · see citing works
Neural interfaces engineered via micro- and nanostructured coatings
Chen, N., Tian, L., Patil, A. C., Peng, S., Yang, I. H., Thakor, N. V., & Ramakrishna, S. (2017) · Nano Today · doi:10.1016/j.nantod.2017.04.007 · Cited by 74 · see citing works
Mandal, H. S., Knaack, G. L., Charkhkar, H., McHail, D. G., Kastee, J. S., Dumas, T. C., Peixoto, N., Rubinson, J. F., & Pancrazio, J. J. (2014) · Acta Biomaterialia · doi:10.1016/j.actbio.2014.02.029 · Cited by 72 · see citing works
Hudak, E. M., Kumsa, D., Martin, H. B., & Mortimer, J. T. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6945 · Cited by 73 · see citing works
Attaining higher resolution visual prosthetics: a review of the factors and limitations
Eiber, C. D., Lovell, N. H., & Suaning, G. J. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/1/011002 · Cited by 71 · see citing works
Advances in Carbon-Based Microfiber Electrodes for Neural Interfacing
Hejazi, M., Tong, W., Ibbotson, M. R., Prawer, S., & Garrett, D. J. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.658703 · Cited by 68 · see citing works
Interfacing with the Brain: How Nanotechnology Can Contribute
Ahmed, A. A. A., Alegret, N., Almeida, B., Álvarez‐Puebla, R. A., Andrews, A. M., Ballerini, L., Barrios‐Capuchino, J. J., Becker, C., Blick, R. H., Bonakdar, S., et al. (2025) · ACS Nano · doi:10.1021/acsnano.4c10525 · Cited by 68 · see citing works
Zeng, Q., Xia, K., Sun, B., Yin, Y., Wu, T., & Humayun, M. S. (2017) · Electrochimica Acta · doi:10.1016/j.electacta.2017.03.213 · Cited by 65 · see citing works
Weremfo, A., Carter, P., Hibbert, D. B., & Zhao, C. (2015) · Langmuir · doi:10.1021/la504876n · Cited by 59 · see citing works
Flexible and Organic Neural Interfaces: A Review
Lago, N., & Cester, A. (2017) · Applied Sciences · doi:10.3390/app7121292 · Cited by 59 · see citing works
Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer
Wang, K., Liu, C.-C., & Durand, D. M. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.926691 · Cited by 57 · see citing works
Alahi, M. E. E., Liu, Y., Xu, Z., Wang, H., Wu, T., & Mukhopadhyay, S. C. (2021) · Materials Today Communications · doi:10.1016/j.mtcomm.2021.102853 · Cited by 55 · see citing works
Lu, Y., Cai, Z., Cao, Y., Yang, H., & Duan, Y. Y. (2008) · Electrochemistry Communications · doi:10.1016/j.elecom.2008.02.013 · Cited by 56 · see citing works
Arrays for chronic functional microstimulation of the lumbosacral spinal cord
McCreery, D. B., Pikov, V., Lossinsky, A. S., Bullara, L. A., & Agnew, W. F. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.827223 · Cited by 56 · see citing works
Self-citationAmorphous and crystalline IrO2 thin films as potential stimulation electrode coatings
Thanawala, S. S., Baird, R. J., Georgiev, D. G., & Auner, G. W. (2008) · Applied Surface Science · doi:10.1016/j.apsusc.2008.02.054 · Cited by 54 · see citing works
Conducting Polymers in Neural Stimulation Applications
Zhou, D. D., Cui, X. T., Hines, A., & Greenberg, R. J. (2009) · Biological and medical physics series · doi:10.1007/978-0-387-98120-8_8 · Cited by 53 · see citing works
Harnack, D., Meissner, W. G., Paulat, R., Hilgenfeld, H., Müller, W., Winter, C., Morgenstern, R., & Kupsch, A. R. (2007) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2007.08.019 · Cited by 51 · see citing works
Characterization of iridium oxide thin films deposited by pulsed-direct-current reactive sputtering
Thanawala, S. S., Georgiev, D. G., Baird, R. J., & Auner, G. W. (2007) · Thin Solid Films · doi:10.1016/j.tsf.2007.02.090 · Cited by 51 · see citing works
Aqrawe, Z., Wright, B. E., Patel, N. D., Vyas, Y., Malmström, J., Montgomery, J. M., Williams, D. E., Travaš‐Sejdić, J., & Svirskis, D. M. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2018.10.099 · Cited by 49 · see citing works
Measuring the effective area and charge density of platinum electrodes for bionic devices
Harris, A. R., Newbold, C., Carter, P., Cowan, R., & Wallace, G. G. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aaba8b · Cited by 47 · see citing works
The functional consequences of chronic, physiologically effective intracortical microstimulation
Parker, R., Davis, T. S., House, P. A., Normann, R. A., & Greger, B. E. (2011) · Progress in brain research · doi:10.1016/b978-0-444-53815-4.00010-8 · Cited by 47 · see citing works
Multimodal, longitudinal assessment of intracortical microstimulation
Koivuniemi, A. S., Wilks, S. J., Woolley, A. J., & Otto, K. J. (2011) · Progress in brain research · doi:10.1016/b978-0-444-53815-4.00011-x · Cited by 45 · see citing works
Lee, S. H., Jung, J. H., Chae, Y. M., Suh, J. F., & Kang, J. Y. (2010) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/20/3/035015 · Cited by 44 · see citing works
Improved Poly(3,4-Ethylenedioxythiophene) (PEDOT) for Neural Stimulation
Mandal, H. S., Kastee, J. S., McHail, D. G., Rubinson, J. F., Pancrazio, J. J., & Dumas, T. C. (2015) · Neuromodulation Technology at the Neural Interface · doi:10.1111/ner.12285 · Cited by 43 · see citing works
Eles, J. R., & Kozai, T. D. Y. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2020.119767 · Cited by 43 · see citing works
Asymmetric Versus Symmetric Pulses for Cortical Microstimulation
Koivuniemi, A. S., & Otto, K. J. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2166563 · Cited by 43 · see citing works
Yin, P., Liu, Y., Xiao, L., & Zhang, C. (2021) · Polymers · doi:10.3390/polym13162834 · Cited by 42 · see citing works
Nanostructured Coatings for Improved Charge Delivery to Neurons
Kozai, T. D. Y., Alba, N. A., Zhang, H., Kotov, N. A., Gaunt, R. A., & Cui, X. T. (2014) · Springer eBooks · doi:10.1007/978-1-4899-8038-0_4 · Cited by 42 · see citing works
Carretero, N. M., Lichtenstein, M. P., Pérez, E., Cabana, L., Sunol, C. E., & Casañ-Pastor, N. (2014) · Acta Biomaterialia · doi:10.1016/j.actbio.2014.06.019 · Cited by 40 · see citing works
Wang, C., Brunton, E., Haghgooie, S., Cassells, K., Lowery, A. J., & Rajan, R. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/4/046010 · Cited by 39 · see citing works
Reactively Sputtered Iridium Oxide
Wessling, B., Besmehn, A., Mokwa, W., & Schnakenberg, U. (2007) · Journal of The Electrochemical Society · doi:10.1149/1.2713691 · Cited by 38 · see citing works
Bioelectronic Neural Interfaces: Improving Neuromodulation Through Organic Conductive Coatings
Duan, W., Robles, U. A. A., Poole-Warren, L. A., & Esrafilzadeh, D. (2023) · Advanced Science · doi:10.1002/advs.202306275 · Cited by 36 · see citing works
Micro-Coil Design Influences the Spatial Extent of Responses to Intracortical Magnetic Stimulation
Lee, S. W., Thyagarajan, K., & Fried, S. I. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2877713 · Cited by 37 · see citing works
Inner and outer retinal mechanisms engaged by epiretinal stimulation in normal and rd mice
Margalit, E., Babai, N. Z., Luo, J.-M., & Thoreson, W. B. (2011) · Visual Neuroscience · doi:10.1017/s0952523810000489 · Cited by 36 · see citing works
Sputtered Ir Films Evaluated for Electrochemical Performance I. Experimental Results
Wessling, B., Mokwa, W., & Schnakenberg, U. (2008) · Journal of The Electrochemical Society · doi:10.1149/1.2844805 · Cited by 36 · see citing works
Prosthetic vision: devices, patient outcomes and retinal research
Hadjinicolaou, A. E., Hamish Meffin, A., Maturana, M. I., Cloherty, S. L., & Ibbotson, M. R. (2015) · Clinical and Experimental Optometry · doi:10.1111/cxo.12342 · Cited by 35 · see citing works
Cisnal, A., Fraile, J. C., Turiel, J. P., Muñoz-Martínez, V. F., Müller, C., & Ihmig, F. R. (2018) · Sensors · doi:10.3390/s18124152 · Cited by 35 · see citing works
Three-dimensional fuzzy graphene ultra-microelectrodes for subcellular electrical recordings
Rastogi, S. K., Bliley, J. M., Matino, L., Garg, R., Santoro, F., Feinberg, A. W., & Cohen‐Karni, T. (2020) · Nano Research · doi:10.1007/s12274-020-2695-y · Cited by 35 · see citing works
Electroresponsive Hydrogels for Therapeutic Applications in the Brain
Khan, Z. M., Wilts, E. M., Vlaisavljevich, E., Long, T. E., & Verbridge, S. S. (2021) · Macromolecular Bioscience · doi:10.1002/mabi.202100355 · Cited by 36 · see citing works
Single walled carbon nanohorns composite for neural sensing and stimulation
Carli, S., Lambertini, L., Zucchini, E., Ciarpella, F., Scarpellini, A., Prato, M., Castagnola, E., Fadiga, L., & Ricci, D. (2018) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2018.05.083 · Cited by 32 · see citing works
Attenuation of the in vitro neurotoxicity of 316L SS by graphene oxide surface coating
Tasnim, N., Kumar, A., & Joddar, B. (2017) · Materials Science and Engineering C · doi:10.1016/j.msec.2016.12.123 · Cited by 31 · see citing works
Musa, S., Rand, D. R., Bartic, C., Eberle, W., Nuttin, B., & Borghs, G. (2011) · Analytical Chemistry · doi:10.1021/ac103037u · Cited by 31 · see citing works
Microelectronic Visual Prostheses
Zhou, D. D., & Greenberg, R. J. (2009) · Biological and medical physics series · doi:10.1007/978-0-387-77261-5_1 · Cited by 31 · see citing works
Winter, J. O., Gokhale, M., Jensen, R. J., Cogan, S. F., & Rizzo, J. F. (2007) · Materials Science and Engineering C · doi:10.1016/j.msec.2007.04.011 · Cited by 30 · see citing works
Self-citationActivated iridium oxide film (AIROF) electrodes for neural tissue stimulation
Frederick, R. A., Meliane, Y. I., Joshi‐Imre, A., Troyk, P. R., & Cogan, S. F. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/abb9bf · Cited by 29 · see citing works
Self-citation
- Authors
- Cogan, S. F., Guzelian, A. A., Agnew, W. F., Yuen, T. G. H., & McCreery, D. B.
- Affiliations
- EIC Laboratories (US)
- Huntington Medical Research Institutes (US)
- Corresponding author
- Cogan, Stuart F
- Venue
- Journal of Neuroscience Methods
- Publisher
- Elsevier BV
- Published
- 2004-08
- Volume
- 137
- Issue
- 2
- Pages
- 141-150
- ISSN
- 0165-0270 (ISSN-L)
- OpenAlex
- W2155993532
- Cited by
- 210
- References
- 25
- Citations by year
- 2026: 5 · 2025: 7 · 2024: 5 · 2023: 11 · 2022: 4 · 2021: 15 · 2020: 10 · 2019: 7 · 2018: 16 · 2017: 16 · 2016: 11 · 2015: 11 · 2014: 14 · 2013: 13 · 2012: 9
- FWCI
- 1.52
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Neuroscience Methods (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health
- National Institute of Neurological Disorders and Stroke (N01-NS-82388, R44 NS55673-03)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Neurology › Neurological disorders and treatments
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- microelectrodes, electrode degradation, coating delamination, electrochemical stability window, water electrolysis, cat cortex
- MeSH terms
- Animals, Cats, Cerebral Cortex, Dose-Response Relationship, Radiation, Electric Stimulation, Electrochemistry, Electrodes, Implanted, Electroplating, Iridium, Microelectrodes, Microscopy, Electron, Scanning
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review