Impedance characteristics of deep brain stimulation electrodes in vitro and in vivo
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
The objective of this study was to quantify the electrode-tissue interface impedance of electrodes used for deep brain stimulation (DBS). We measured the impedance of DBS electrodes using electrochemical impedance spectroscopy in vitro in a carbonate- and phosphate-buffered saline solution and in vivo following acute implantation in the brain. The components of the impedance, including the series resistance (R(s)), the Faradaic resistance (R(f)) and the double layer capacitance (C(dl)), were estimated using an equivalent electrical circuit. Both R(f) and C(dl) decreased as the sinusoidal frequency was increased, but the ratio of the capacitive charge transfer to the Faradaic charge transfer was relatively insensitive to the change of frequency. R(f) decreased and C(dl) increased as the current density was increased, and above a critical current density the interface impedance became nonlinear. Thus, the magnitude of the interface impedance was strongly dependent on the intensity (pulse amplitude and duration) of stimulation. The temporal dependence and spatial non-uniformity of R(f) and C(dl) suggested that a distributed network, with each element of the network having dynamics tailored to a specific stimulus waveform, is required to describe adequately the impedance of the DBS electrode-tissue interface. Voltage transients to biphasic square current pulses were measured and suggested that the electrode-tissue interface did not operate in a linear range at clinically relevant current amplitudes, and that the assumption of the DBS electrode being ideally polarizable was not valid under clinical stimulating conditions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Charge density and charge per phase as cofactors in neural injury induced by electrical stimulation
McCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.102812 · Cited by 708
Over-pulsing degrades activated iridium oxide films used for intracortical neural stimulation
Cogan, S. F., Guzelian, A. A., Agnew, W. F., Yuen, T. G. H., & McCreery, D. B. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.02.019 · Cited by 210
Comparison of neural damage induced by electrical stimulation with faradaic and capacitor electrodes
McCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. A. (1988) · Annals of Biomedical Engineering · doi:10.1007/bf02368010 · Cited by 163
Agnew, W. F., Yuen, T. G. H., & McCreery, D. B. (1983) · Experimental Neurology · doi:10.1016/0014-4886(83)90221-2 · Cited by 96
Current Density Profiles of Surface Mounted and Recessed Electrodes for Neural Prostheses
Rubinstein, J. T., Spelman, F. A., Soma, M., & Suesserman, M. F. (1987) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1987.326007 · Cited by 89
Cited by
Glial responses to implanted electrodes in the brain
Salatino, J. W., Ludwig, K. A., Kozai, T. D. Y., & Purcell, E. K. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-017-0154-1 · Cited by 642
“Even for simple surfaces, faradaic reactions such as platinum dissolution occur at low levels of stimulation 49 , and increase as a function of increasing stimulation intensity 50 .”
Lee, H.-M., Kwon, K. Y., Li, W., & Ghovanloo, M. (2015) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2014.2355814 · Cited by 161
Principles of electrical stimulation of neural tissue
(2013) · Handbook of Clinical Neurology · doi:10.1016/b978-0-444-53497-2.00001-2 · Cited by 151
Self-citationArfin, S. K., & Sarpeshkar, R. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2166072 · Cited by 133
“16(a)] and the layout of a single channel”
BackgroundCulaclii, S., Kim, B., Lo, Y.-K., Li, L., & Liu, W. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2816464 · Cited by 34
“[31] characterizes a typical DBS stimulation in-vivo with safe stimulation currents and pulse widths, and records resulting voltage amplitudes.”
Metadata
- Authors
- Wei, X. F., & Grill, W. M.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2009-07-09
- Volume
- 6
- Issue
- 4
- Pages
- 046008
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2025602507
- Cited by
- 155
- References
- 5
- FWCI
- 2.81
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Impedance characteristics of deep brain stimulation electrodes in vitro and in vivo. https://app.cochat.ai/papers/10.1088/1741-2560/6/4/046008
- The paper
- Wei, X. F., & Grill, W. M. (2009). Impedance characteristics of deep brain stimulation electrodes in vitro and in vivo. Journal of Neural Engineering. https://doi.org/10.1088/1741-2560/6/4/046008
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.