Nocturnal Continuous Glucose Monitoring: Accuracy and Reliability of Hypoglycemia Detection in Patients with Type 1 Diabetes at High Risk of Severe Hypoglycemia

Bay, Christiane, Kristensen, Peter Lommer, Pedersen-Bjergaard, Ulrik, Tarnow, Lise, Thorsteinsson, Birger · 2013 · Diabetes Technology & Therapeutics
Cited by 51

The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.

Abstract

BACKGROUND: A reliable method to detect biochemical nocturnal hypoglycemia is highly needed, especially in patients with recurrent severe hypoglycemia. We evaluated reliability of nocturnal continuous glucose monitoring (CGM) in patients with type 1 diabetes at high risk of severe hypoglycemia. PATIENTS AND METHODS: Seventy-two type 1 diabetes patients with recurrent severe hypoglycemia (two or more events within the last year) participated for 4 nights in blinded CGM recordings (Guardian(®) REAL-Time CGMS and Sof-Sensor(®); Medtronic MiniMed, Northridge, CA). Blood was drawn hourly from 23:00 to 07:00 h for plasma glucose (PG) measurements (gold standard). RESULTS: Valid data were obtained in 217 nights. The sensitivity of CGM was 65% (95% confidence interval, 53-77%) below 4 mmol/L, 40% (24-56%) below 3 mmol/L, and 17% (0-47%) below 2.2 mmol/L. PG and CGM readings correlated in the total measurement range (Spearman's ρ=0.82; P<0.001). In the normo- and hyperglycemic ranges CGM underestimated PG by 1.1 mmol/L (0.9-1.2 mmol/L) (P<0.001); in contrast, in the hypoglycemic range (PG<4 mmol/L) CGM overestimated PG levels by 1.0 mmol/L (P<0.001). The mean absolute relative differences in the hypo- (≤3.9 mmol/L), normo- (4-9.9 mmol/L), and hyperglycemic (≥10 mmol/L) ranges were 45% (37-53%), 23% (22-25%), and 20% (19-21%), respectively. Continuous glucose error grid analysis indicated a clinical accuracy of 56%, 99%, and 93% in the hypo-, normo-, and hyperglycemic ranges, respectively. CONCLUSIONS: The accuracy in the hypoglycemic range of nocturnal CGM data using Sof-Sensor is suboptimal in type 1 diabetes patients at high risk of severe hypoglycemia. To ensure clinical useful sensitivity in detection of nocturnal hypoglycemic episodes, an alarm threshold should not be lower than 4 mmol/L.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/dia.2013.0004

Cites

5 of 5 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 5 references

Cited by

See all 51

Metadata

Authors
Bay, C., Kristensen, P. L., Pedersen-Bjergaard, U., Tarnow, L., & Thorsteinsson, B.
Venue
Diabetes Technology & Therapeutics
Published
2013-05
Volume
15
Issue
5
Pages
371-377
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2063298884
Cited by
51
References
5
FWCI
5.90
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Nocturnal Continuous Glucose Monitoring: Accuracy and Reliability of Hypoglycemia Detection in Patients with Type 1 Diabetes at High Risk of Severe Hypoglycemia. https://app.cochat.ai/papers/10.1089/dia.2013.0004
The paper
Bay, C., Kristensen, P. L., Pedersen-Bjergaard, U., Tarnow, L., & Thorsteinsson, B. (2013). Nocturnal Continuous Glucose Monitoring: Accuracy and Reliability of Hypoglycemia Detection in Patients with Type 1 Diabetes at High Risk of Severe Hypoglycemia. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2013.0004

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.