“Let the Algorithm Do the Work”: Reduction of Hypoglycemia Using Sensor-Augmented Pump Therapy with Predictive Insulin Suspension (SmartGuard) in Pediatric Type 1 Diabetes Patients

Open access
Biester, Torben, Kordonouri, Olga, Holder, Martin, Remus, Kerstin, Kieninger-Baum, Dorothee, Wadien, Tanja, Danne, Thomas · 2017 · Diabetes Technology & Therapeutics
Cited by 128

The DOI matches its Crossref record; 9 of the 9 references it cites were found in a registry or an index.

Abstract

BACKGROUND: 640G system with SmartGuard™ technology allows an automatic stop of insulin delivery based on prediction of low glucose levels. Since pediatric patients are particularly prone to hypoglycemia, this device may offer additional protection beyond conventional sensor-augmented therapy. METHODS: This prospective, pediatric multicenter user evaluation assessed 6 weeks of SAP with SmartGuard (threshold setting for hypoglycemia: 70 mg/dL) compared to a preceding period of 2 weeks with SAP only. The primary outcome was the potential reduction in the frequency of hypoglycemic episodes and hypoglycemic intensity (area under the curve [AUC] and time <70 mg/dL). RESULTS: The study included 24 patients with at least 3 months of insulin pump use (average age: 11.6 ± 5.1 years, 15 female, average type 1 diabetes duration: 7.5 ± 4.2 years, mean ± SD) who had on average 3.2 ± 1.0 predictive suspensions/patient/day. The mean sensor glucose minimum during suspension was 78 ± 6 mg/dL and the average suspension time was 155 ± 47 min/day. Use of SmartGuard in patients treated as per the protocol (n = 18) reduced the number of instances in which the glucose level was <70 mg/dL (1.02 ± 0.52 to 0.72 ± 0.36; P = 0.027), as well as AUC <70 mg/dL (0.76 ± 0.73 to 0.38 ± 0.24; P = 0.027) and the time/day the level fell below 70 mg/dL (73 ± 56 to 31 ± 22 min). The reduction of hypoglycemia was not associated with a significant change in mean glucose concentration (171 ± 26 to 180 ± 19 mg/dL, P = 0.111) and HbA1c (7.5% ± 0.5% to 7.6% ± 0.7%, (P = 0.329). Manual resumption of insulin delivery followed by carbohydrate intake resulted in significantly higher glucose levels 1 h after suspension compared to SmartGuard suspensions with automatic resume (190.8 ± 26.5 vs. 138.7 ± 10.3 mg/dL; P < 0.001). CONCLUSIONS: SmartGuard technology significantly reduced the risk for hypoglycemia in pediatric type 1 diabetes patients without increasing HbA1c. Patients must be educated that when using combining predictive low-glucose insulin suspension technology, extra carbohydrate intake in response to an alarm combined with manual resumption is likely to cause rebound hyperglycemia. The best results were achieved when the user did not interfere with pump operation.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/dia.2016.0349

Cites

9 of 9 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 9 references

Cited by

Metadata

Authors
Biester, T., Kordonouri, O., Holder, M., Remus, K., Kieninger-Baum, D., Wadien, T., & Danne, T.
Venue
Diabetes Technology & Therapeutics
Published
2017-03
Volume
19
Issue
3
Pages
173-182
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2580312941
Cited by
128
References
9
FWCI
13.18
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.