Shining the Spotlight on Multiple Daily Insulin Therapy: Real-World Evidence of the InPen Smart Insulin Pen
Open accessAbstract
Objective: Connected insulin pens are creating opportunities for the millions of individuals with diabetes using multiple daily injections (MDI) therapy across the globe. Continuous glucose monitoring (CGM) data from connected insulin pens are revealing gaps and opportunities to significantly improve care for this population. In this article, we report real-world findings of the InPen™ smart insulin pen paired with CGM (InPen system), used by persons with type 1 diabetes (T1D) and type 2 diabetes (T2D). Methods: A retrospective cohort analysis was conducted with the real-world data collected from the InPen system of individuals ( N = 3793 with T1D, N = 552 with T2D, and N = 808 unidentified) who used the system from January 01, 2020, to December 31, 2021. Diabetes management (e.g., missed and mistimed insulin dosing, mismatched food intake, and correction dose delivery) and glycemic outcomes were assessed. Results: In the overall and T1D populations, a dosing frequency of ≥3 doses per day and a missed dose frequency of <20% was associated with improved glycemia. In adults with T2D, missing <20% of doses was the significant factor determining improved glycemia. Conclusion: This analysis, integrating data from a smart insulin pen and CGM, provides insights into the impact of dosing behavior on glycemic outcomes and informs counseling strategies for the diabetes care team, through technologically advanced insulin management for those using MDI therapy.
AI snapshot
Written from the paper’s full text.
TL;DR
In a retrospective real-world analysis of 5,153 InPen smart insulin pen users paired with CGM, dosing ≥3 rapid-acting insulin doses/day and missing <20% of doses were associated with better glycemia in the overall and type 1 diabetes populations, while missing <20% of doses was the significant factor in adults with type 2 diabetes. On-time dosing and fewer missed correction opportunities correlated with higher time-in-range.
Key findings
- Overall, ≥3 doses/day plus missed dose rate <20% (N = 962) was associated with TIR 70.3 ± 19.2%, TAR 26.9 ± 19.8%, and GMI 7.0 ± 0.8, compared with <3 doses/day (N = 1909; TIR 52.0 ± 26.0%, TAR 46.3 ± 26.4%, GMI 7.9 ± 1.3) and ≥3 doses/day with missed dose rate ≥20% (N = 2282; TIR 50.1 ± 20.2%, TAR 48.2 ± 20.8%, GMI 7.8 ± 0.9).
- In adults with type 2 diabetes, missing <20% of doses was the significant factor determining improved glycemia.
- Cluster analysis showed Cluster 3 had the highest on-time dosing (72.9 ± 1.0%) and TIR 69.0 ± 19.8%, while Cluster 1 had the lowest on-time dosing (27.9 ± 8.5%) and TIR 37.4 ± 20.5%; mean TIR and TAR differed significantly across clusters (P < 0.0001). On-time dose frequency correlated with TIR (r = 0.570, P < 0.0001).
- Higher dose-count groups tended to have lower TAR and GMI and higher TIR; for example, ≥6+ doses/day (N = 382) had TIR 61.0% and GMI 7.33 versus ≥1 dose/day (N = 462) with TIR 50.8% and GMI 8.01, with significant difference in mean TIR (P < 0.05). Missed correction opportunities per day showed strong negative correlation with TIR (r = −0.852, P < 0.0001).
Methods
- Retrospective cohort analysis of real-world InPen system data (InPen smart insulin pen paired with personal CGM) from January 1, 2020, to December 31, 2021. Users were adults and pediatrics who used the system ≥30 days; the analyzed dataset included the last 14 days of bolus dosing and CGM data. Users without at least one bolus record and <70% sensor glucose data were excluded, and the first 30 days were excluded. Overall N = 5153 (3793 type 1 diabetes, 552 type 2 diabetes, 808 unidentified).
- Dosing behaviors included average rapid-acting doses/day, correction-only versus meal/meal+correction doses, and missed/mistimed doses. Meal events were detected by sensor glucose rate-of-change; on-time bolus was 30 min before to ≤15 min after the start of glucose rise, late was >15 to ≤60 min after, and missed was no bolus from ≤30 min before to ≥60 min after. Missed correction opportunities were defined as elevated sensor glucose where dose recommendations were no less than each user’s correction threshold.
- Statistical analysis used ANOVA with Tukey post hoc (significance 0.05), Pearson correlation, and k-means clustering (k-means++ with standardized numerical features) in Python/SciPy and BigQuery Machine Learning. Continuous variables are mean ± SD; categorical variables are n (%).
Limitations
- The paper states that the population only included users with compatible real-time CGM alongside the smart insulin pen; a follow-up study using blood glucose monitoring would be needed for smart insulin pen users without CGM. The study period was limited to 30 days, not the entire duration of usage, and race/ethnicity data were not available.
- The paper states that the analyses compared groups split from real-world data and do not prove direct causal relationships; for example, more frequent bolus dosing was associated with higher TIR, but other factors such as on-time, late, or missed dosing and optimized therapy settings may contribute.
- Evident from the text: this was a retrospective observational analysis without a control group, and demographic information was self-reported and optional, with missing age, diabetes type, and gender data (e.g., 808 type not specified, 695 age not specified), which limits generalizability.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
4 of 12 cited sources checked · 1 supported · 0 flagged · accuracy not yet measured
- Findingown data
All three clusters had a significant difference in mean time-in-range (P < 0.0001).
“All three groups had a significant difference in mean TIR (P < 0.0001) and TAR (P < 0.0001).”
Quote in the paperSubstantial claim - Findingown data
In adults with T2D, missing <20% of doses was the significant factor determining improved glycemia.
“In adults with T2D, missing <20% of doses was the significant factor determining improved glycemia.”
Quote in the paperSubstantial claim - Findingown data
In the T1D population, a dosing frequency of ≥3 doses per day and a missed dose frequency of <20% were associated with better glycemic outcomes compared to lower dosing frequency or higher missed dose rate.
“In the overall and T1D populations, a dosing frequency of ≥3 doses per day and a missed dose frequency of <20% was associated with improved glycemia.”
- Also stated in: § Results · plain
“In the overall and T1D populations, a dosing frequency of 3+ doses/day and a missed dose frequency of <20% were associated with better glycemic outcomes compared to those with a lower dosing frequency or higher missed dose rate.”
Quote in the paperSubstantial claim - Also stated in: § Results · plain
- Findingown data
Increased correction dose frequency is associated with improved time-in-range.
“This analysis, in fact, shows that increased correction dose frequency is associated with improved TIR.”
Quote in the paperSubstantial claim - Findingown data
All three clusters had a significant difference in mean time above range (P < 0.0001).
“All three groups had a significant difference in mean TIR (P < 0.0001) and TAR (P < 0.0001).”
Quote in the paperSubstantial claim
- Findingown data
Missed and mistimed insulin doses are common in multiple daily injection therapy, as verified in this analysis of smart insulin pen users.
“Missed and mistimed insulin doses are common in MDI therapy as verified in this analysis of smart insulin pen users.”
Quote in the paperSubstantial claim - Findingown data
The average duration of missed correction opportunities per day showed a strong negative correlation with time-in-range (r = −0.852, P < 0.0001).
“The average duration of missed correction opportunities per day showed strong negative correlation with TIR (r = −0.852, P < 0.0001) (Supplementary Fig. S1).”
Quote in the paperSubstantial claim - Findingown data
The frequency of on-time dosing was significantly correlated with user time-in-range (r = 0.570, P < 0.0001).
“The frequency of on-time dosing was significantly correlated with user TIR (r = 0.570, P < 0.0001, Tukey's test), as illustrated in Figure 2.”
Quote in the paperSubstantial claim - Findingown data
In the overall study population, mean time-in-range differed significantly across groups based on the number of injections per day (P < 0.05).
“Groups had a significant difference in mean TIR (P < 0.05) based on the number of injections per day (Table 2).”
Quote in the paperSubstantial claim - Findingown data
Cluster 2 had outcomes that fell between Clusters 1 and 3.
“Cluster 2 (N = 2139) had outcomes that fell between Clusters 1 and 3 (Table 3).”
Quote in the paperNot a substantial claim - Findingown data
In the overall population, a dosing frequency of ≥3 doses per day and a missed dose frequency of <20% were associated with better glycemic outcomes compared to lower dosing frequency or higher missed dose rate.
“In the overall and T1D populations, a dosing frequency of ≥3 doses per day and a missed dose frequency of <20% was associated with improved glycemia.”
- Also stated in: § Results · plain
“In the overall and T1D populations, a dosing frequency of 3+ doses/day and a missed dose frequency of <20% were associated with better glycemic outcomes compared to those with a lower dosing frequency or higher missed dose rate.”
Quote in the paperSubstantial claim - Also stated in: § Results · plain
- Premisecitation
In a systematic review, 20%-45% of participants on insulin therapy reported mistiming of their insulin doses.
“In a systematic review, 20%–45% of participants on insulin therapy reported the mistiming of their insulin doses17 citing disrupted daily routines, social situations, and hypoglycemia avoidance as reasons.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [17] Robinson S, Newson RS, Liao B, et al. Missed and mistimed insulin doses in people with diabetes: A systematic literature review. Diabetes Technol Ther 2021; 23: 844– 856. 34270324 source
- Interpretationreasoning
Smart insulin pen systems may facilitate communication with the diabetes care team.
“It demonstrates that smart insulin pen systems support MDI therapy patients in daily dosing decisions and may facilitate communication with the diabetes care team, which provides precision insulin management for all individuals on insulin therapy regardless of delivery method.”
Quote in the paperNot a substantial claim - Study design and conductnone given
The population for the analysis only included those who used compatible real-time continuous glucose monitoring along with the smart insulin pen.
“First, the population for the analysis only included those who used compatible real-time CGM along with the smart insulin pen.”
Quote in the paperSubstantial claim - Interpretationown data
Smart insulin pen systems support multiple daily injection therapy patients in daily dosing decisions.
“It demonstrates that smart insulin pen systems support MDI therapy patients in daily dosing decisions and may facilitate communication with the diabetes care team, which provides precision insulin management for all individuals on insulin therapy regardless of delivery method.”
Quote in the paperSubstantial claim - Interpretationown data
Smart insulin pens support individuals on multiple daily injection therapy with intelligent algorithms and features allowing opportunities to safely deliver more frequent rapid-acting insulin doses, including correction doses.
“Smart insulin pens support individuals on MDI therapy with intelligent algorithms and features (e.g., objective active insulin tracking, differentiating priming from therapy doses, and dose recommendations) allowing opportunities to safely deliver more frequent rapid-acting insulin doses, including correction doses.”
Quote in the paperSubstantial claim - Premisecitation
A retrospective analysis of InPen system data indicated that 60% of device users required adjustments in their basal insulin dose.
“In fact, a retrospective analysis of InPen system data indicated 60% of device users required adjustments in their basal insulin dose.27”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [27] Smith MB, Gaetano A, Vigersky R, et al. Smart insulin pens uncover the need for long-acting insulin dose pptimization in multiple daily insulin therapy. Endocr Pract 2021; 27: S49– S50. source
- May combine two claims. 11
- Classified as probably not a substantial claim. 9
- The quote only partly states this claim. 1
May combine two claims.
PremisecitationIn a prospective, observational trial of 94 participants with T1D, the use of a connected insulin pen with continuous glucose monitoring resulted in significant improvements in glycemia and a significant decrease in missed doses.
“In a prospective, observational trial of 94 participants with T1D, the use of a connected insulin pen with CGM resulted in significant improvements in glycemia and a significant decrease in missed doses.13”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [13] Adolfsson P, Hartvig NV, Kaas A, et al. Increased time in range and fewer missed bolus injections after introduction of a smart connected insulin pen. Diabetes Technol Ther 2020; 22: 709– 718. 32003590 PMC7591375 source
The Methods describe a prospective, observational study, and the Results report significant improvements in glycemic measures among 94 adults with T1D using CGM. Missed bolus injections fell significantly, but that analysis included only 81 participants. The Discussion cautions that this single-arm study cannot establish that the connected pen directly caused the improvements, so “resulted in” overstates the finding. (LLM, conf=0.90)
Experimental — accuracy not yet measured
May combine two claims.
PremisecitationResearch from individuals using multiple daily injection therapy with continuous glucose monitoring and connected insulin pen or cap devices revealed that missed, mistimed, and mismatched insulin bolus doses are common and occur in both older and younger populations with T1D or T2D, and are associated with higher A1C.
“Research from individuals using MDI therapy with CGM and connected insulin pen or cap devices revealed that missed, mistimed, and mismatched insulin bolus doses are common and occur in both older and younger populations with T1D or T2D, and are associated with higher A1C.13–18”
Quote in the paperSubstantial claim2 of 6 sources checked- cites [13] Adolfsson P, Hartvig NV, Kaas A, et al. Increased time in range and fewer missed bolus injections after introduction of a smart connected insulin pen. Diabetes Technol Ther 2020; 22: 709– 718. 32003590 PMC7591375 source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [14] Munshi MN, Slyne C, Greenberg JM, et al. Nonadherence to insulin therapy detected by bluetooth-enabled pen cap is associated with poor glycemic control. Diabetes Care 2019; 42: 1129– 1131. 30862650 source
- cites [15] Norlander LM, Anderson S, Levy CJ, et al. Late and missed meal boluses with multiple daily insulin injections. Diabetes 2018; 67: A259. source
- cites [16] Randlov J, Poulsen JU. How much do forgotten insulin injections matter to hemoglobin a1c in people with diabetes? A simulation study. J Diabetes Sci Technol 2008; 2: 229– 235. 19885347 PMC2771482 source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [17] Robinson S, Newson RS, Liao B, et al. Missed and mistimed insulin doses in people with diabetes: A systematic literature review. Diabetes Technol Ther 2021; 23: 844– 856. 34270324 source
- cites [18] Toschi E, Slyne C, Greenberg JM, et al. Examining the relationship between pre- and postprandial glucose levels and insulin bolus timing using bluetooth-enabled insulin pen cap technology and continuous glucose monitoring. Diabetes Technol Ther 2020; 22: 19– 24. 31483167 source
May combine two claims.
PremisecitationA study of pediatric individuals with T1D identified six habits associated with reduction of HbA1c, four of which are related to insulin delivery.
“A study of pediatric individuals with T1D identified six habits associated with reduction of HbA1c.28 Four of these six habits are related to insulin delivery, demonstrating the importance of ensuring that individuals with diabetes have the support they need to manage the complexity of determining when to deliver an insulin dose and how much insulin is needed.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [28] Lee JM, Rusnak A, Garrity A, et al. Feasibility of electronic health record assessment of 6 pediatric type 1 diabetes self-management habits and their association with glycemic outcomes. JAMA Netw Open 2021; 4: e2131278. 34709387 PMC8554640 source
May combine two claims.
Findingown dataCluster 3 had the highest adherence to on-time dosing and better glycemic outcomes.
“Cluster analysis demonstrated that Cluster 1 (N = 1125) had the lowest adherence to on-time dosing and the suboptimal glycemic outcomes, while Cluster 3 (N = 1889) demonstrated the highest adherence to on-time dosing and had better glycemic outcomes.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataUsers tended to have lower time above range and lower glucose management indicator as they dosed more often with a higher correction frequency.
“Users tended to have lower TAR and GMI, as they dosed more often with a higher correction frequency.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataCluster 1 had the lowest adherence to on-time dosing and suboptimal glycemic outcomes.
“Cluster analysis demonstrated that Cluster 1 (N = 1125) had the lowest adherence to on-time dosing and the suboptimal glycemic outcomes, while Cluster 3 (N = 1889) demonstrated the highest adherence to on-time dosing and had better glycemic outcomes.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenDemographic information on race/ethnicity was not available in this real-world analysis.
“Demographic information on race/ethnicity was not available in this real-world analysis.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductreasoningThe analyses were designed to compare between groups split from the population and do not prove a direct causal relationship between variables.
“Based on the real-world data, the analyses were designed to compare between groups split from the population. This may be useful to reveal the difference between the groups by the certain behaviors, which was an aim for this study. However, it does not prove the direct causal relationship between variables.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe study period is limited to 30 days, not during the entire duration of usage.
“Second, the study period is limited to 30 days, not during the entire duration of the usage.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). The quote only partly states this claim.
InterpretationreasoningSmart insulin pen systems provide precision insulin management for all individuals on insulin therapy regardless of delivery method.
“It demonstrates that smart insulin pen systems support MDI therapy patients in daily dosing decisions and may facilitate communication with the diabetes care team, which provides precision insulin management for all individuals on insulin therapy regardless of delivery method.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningInternal Review Board approval or exemption was not necessary because the study did not involve human subject protection requirements.
“As the present study did not involve subject recruitment, enrollment, or participation in a trial and did not fall under human subject protection requirements (per the Department of Health and Human Services, Code of Federal Regulations, Title 28, Chapter 1, Part 46), Internal Review Board approval or exemption was not necessary.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationInsulin stacking is estimated to occur in 60% of insulin bolus doses, putting the individual at risk for hypoglycemia.
“Insulin stacking (overlap of bolus doses) is estimated to occur in 60% of insulin bolus doses putting the individual at risk for hypoglycemia.22”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [22] Schmidt S, Norgaard K. Bolus calculators. J Diabetes Sci Technol 2014; 8: 1035– 1041. 24876436 PMC4455388 source
Classified as probably not a substantial claim.
Study design and conductreasoningThe first 30 days were excluded from the analysis to avoid measuring dosing behaviors or glycemic outcomes associated with onboarding of the diabetes technology.
“The first 30 days were excluded from the analysis to avoid measuring dosing behaviors or glycemic outcomes associated with the onboarding of the diabetes technology.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Study design and conductnone givenUsers without at least one bolus record and <70% of sensor glucose data available in the last 14 days between the first 30 to 59 days of InPen use were excluded.
“Users without at least one bolus record and <70% of sensor glucose (SG) data available in the last 14 days between the first 30 to 59 days of InPen use were excluded.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationIn an observational study of 50 participants with T1D on multiple daily injection therapy using a Bluetooth-enabled pen cap and continuous glucose monitoring, 37% of total boluses resulted in persistent 3-hour postprandial hyperglycemia, while 10% resulted in clinically significant 3-hour postprandial hypoglycemia (glucose <55 mg/dL).
“In an observational study of 50 participants with T1D on MDI therapy using a Bluetooth-enabled pen cap and CGM, it was noted that 37% of total boluses resulted in persistent 3-h postprandial hyperglycemia, while 10% resulted in clinically significant 3-h postprandial hypoglycemia (glucose <55 mg/dL).18”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [18] Toschi E, Slyne C, Greenberg JM, et al. Examining the relationship between pre- and postprandial glucose levels and insulin bolus timing using bluetooth-enabled insulin pen cap technology and continuous glucose monitoring. Diabetes Technol Ther 2020; 22: 19– 24. 31483167 source
Classified as probably not a substantial claim. May combine two claims.
PremisereasoningReal-world data permit analysis of a large data set that includes both sensor glucose and the time and amount of the insulin dose and reflect the actual conditions that patients encounter in daily life.
“Real-world data not only permit the analysis of a large data set that includes both sensor glucose and the time and amount of the insulin dose but also importantly reflect the actual conditions that patients encounter in daily life.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable), S4 (based). May combine two claims.
InterpretationreasoningIntelligent dosing technology can support both the person on multiple daily injection therapy and their care team, further reducing the considerable burden of insulin delivery.
“Intelligent dosing technology can support both the person on MDI therapy and their care team, further reducing the considerable burden of insulin delivery.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
This study is the first exploration and analysis of behaviors and outcomes for multiple daily injection patients using real-world data.
“A strength of this study is that it is the first exploration and analysis of behaviors and outcomes for MDI patients using real-world data.”
Cites
State of Type 1 Diabetes Management and Outcomes from the T1D Exchange in 2016–2018
Foster, N. C., Beck, R. W., Miller, K. M., Clements, M. A., Rickels, M. R., DiMeglio, L. A., Maahs, D. M., Tamborlane, W. V., Bergenstal, R., Smith, E., et al. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0384 · Cited by 1,907 · see citing works
Garber, A. J., Handelsman, Y., Grunberger, G., Einhorn, D., Abrahamson, M. J., Barzilay, J. I., Blonde, L., Bush, M. A., DeFronzo, R. A., Garber, J. R., et al. (2020) · Endocrine Practice · doi:10.4158/cs-2019-0472 · Cited by 598 · see citing works
Association of Numeracy and Diabetes Control
Cavanaugh, K. L., Huizinga, M. M., Wallston, K. A., Gebretsadik, T. S., Shintani, A., Davis, D., Gregory, R., Fuchs, L. S., Malone, R., Cherrington, A., et al. (2008) · Annals of Internal Medicine · doi:10.7326/0003-4819-148-10-200805200-00006 · Cited by 420 · see citing works
7. Diabetes Technology: <i>Standards of Care in Diabetes</i> — <i>2023</i>
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301 · see citing works
Grunberger, G., Sherr, J., Allende, M., Blevins, T., Bode, B., Handelsman, Y., Hellman, R., Lajara, R., Roberts, V. L., Rodbard, D., et al. (2021) · Endocrine Practice · doi:10.1016/j.eprac.2021.04.008 · Cited by 288 · see citing works
Schmidt, S., & Nørgaard, K. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814532906 · Cited by 170 · see citing works
Adolfsson, P., Hartvig, N. V., Kaas, A., Møller, J. B., & Hellman, J. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0411 · Cited by 139 · see citing works
Poor numeracy skills are associated with glycaemic control in Type 1 diabetes
Marden, S. M., Thomas, P., Sheppard, Z. A., Knott, J., Lueddeke, J., & Kerr, D. E. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03466.x · Cited by 107 · see citing works
Gabbay, R. A., Kendall, D., Beebe, C., Cuddeback, J. K., Hobbs, T. M., Khan, N. D., Leal, S., Miller, E. M., Novak, L. M., Rajpathak, S. N., et al. (2020) · Clinical Diabetes · doi:10.2337/cd20-0053 · Cited by 92 · see citing works
Gilbert, T. R., Noar, A., Blalock, O., & Polonsky, W. H. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0666 · Cited by 84 · see citing works
Missed and Mistimed Insulin Doses in People with Diabetes: A Systematic Literature Review
Robinson, S., Newson, R. S., Liao, B., Kennedy-Martin, T., & Battelino, T. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0164 · Cited by 73 · see citing works
Munshi, M. N., Slyne, C., Greenberg, J. M., Greaves, T., Lee, A., Carl, S., Atakov-Castillo, A., & Toschi, E. (2019) · Diabetes Care · doi:10.2337/dc18-1631 · Cited by 66 · see citing works
Randløv, J., & Poulsen, J. U. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200209 · Cited by 61 · see citing works
Hankosky, E. R., Schapiro, D., Gunn, K. B., Lubelczyk, E. B., Mitroi, J., & Nelson, D. R. (2023) · Diabetes Therapy · doi:10.1007/s13300-023-01399-0 · Cited by 57 · see citing works
Barriers to Uptake of Insulin Technologies and Novel Solutions
PAULEY, M. E., Berget, C., Messer, L. H., & Forlenza, G. P. (2021) · Medical Devices Evidence and Research · doi:10.2147/mder.s312858 · Cited by 50 · see citing works
Diabetes Numeracy and Blood Glucose Control: Association With Type of Diabetes and Source of Care
Zaugg, S. D., Dogbey, G. Y., Collins, K., Reynolds, S., Batista, C., Brannan, G. D., & Shubrook, J. H. (2014) · Clinical Diabetes · doi:10.2337/diaclin.32.4.152 · Cited by 44 · see citing works
Toschi, E., Slyne, C., Greenberg, J. M., Greaves, T., Atakov-Castillo, A., Carl, S., Dufour, A. B., & Munshi, M. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0186 · Cited by 39 · see citing works
Smart Insulin Pens: Advancing Digital Transformation and a Connected Diabetes Care Ecosystem
Kompala, T., & Neinstein, A. B. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820984490 · Cited by 38 · see citing works
Steenkamp, D., Eby, E. L., Gulati, N., & Liao, B. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296821997923 · Cited by 33 · see citing works
Lee, J. M., Rusnak, A., Garrity, A., Hirschfeld, E., Thomas, I. H., Wichorek, M., Lee, J. E., Rioles, N. A., Ebekozien, O., & Corathers, S. D. (2021) · JAMA Network Open · doi:10.1001/jamanetworkopen.2021.31278 · Cited by 30 · see citing works
MacLeod, J., & Vigersky, R. A. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221134546 · Cited by 24 · see citing works
Self-citationBuilding a Data-Driven Multiple Daily Insulin Therapy Model Using Smart Insulin Pens
Harbison, R., Hecht, M., & MacLeod, J. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820951225 · Cited by 22 · see citing works
Self-citationKlonoff, D. C., Zhang, J. Y., Shang, T., Mehta, C., & Kerr, D. E. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820973185 · Cited by 20 · see citing works
Late and Missed Meal Boluses with Multiple Daily Insulin Injections
NORLANDER, L. M., ANDERSON, S., LEVY, C. J., EKHLASPOUR, L., LAM, D. W., HSU, L., LOEBNER, S. E., OGYAADU, S. J., O'MALLEY, G., LEVISTER, C. M., et al. (2018) · Diabetes · doi:10.2337/db18-992-p · Cited by 19 · see citing works
Smith, M. B., Gaetano, A., Vigersky, R., Cordero, T., & MacLeod, J. (2021) · Endocrine Practice · doi:10.1016/j.eprac.2021.04.576 · Cited by 5 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 2
- Vigersky RA Cordero TL MacLeod J et al. Artificial intelligence: The next frontier in diabetes therapy. Nature: Sponsor Feature. Available from: https://www.nature.com/articles/d42473-021-00265-6#:∼:text=The%20%E2%80%9Cinternet%20of%20things%E2%80%9D%20applied minimize%20the%20burden%20of%20diabetes [Last accessed: July31 2023].
- Kerr D, , Warshaw H. Smart insulin pens will address critical unmet needs for people with diabetes using insulin. Endocr Today, 2019; 17:21–22.
Cited by
Sorted by citations
Ebekozien, O., Fantasia, K., Farrokhi, F., Sabharwal, A., & Kerr, D. (2024) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.15470 · Cited by 70 · see citing works
Danne, T. P. A., Joubert, M., Hartvig, N. V., Kaas, A., Knudsen, N. N., & Mader, J. K. (2024) · Diabetes Care · doi:10.2337/dc23-2176 · Cited by 42 · see citing works
The critical elements of digital health in diabetes and cardiometabolic care
Shomali, M., Mora, P., Aleppo, G., Peeples, M., Kumbara, A., MacLeod, J., & Iyer, A. (2024) · Frontiers in Endocrinology · doi:10.3389/fendo.2024.1469471 · Cited by 10 · see citing works
Self-citationTillman, J. B., Beck, R. W., Polonsky, W. H., Calhoun, P., Mouse, T., Bailey, R., Jacoby, J., Nandan, R., Sheikh, F. N., & Malone, J. K. (2025) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgae778 · Cited by 1 · see citing works
Ruiz Gracia, T., Nattero-Chávez, L., Arranz Martín, A., de la Calle de la Villa, E., Lorenzo Moñino, M., Bayona Cebada, A., Corral, M. I., Quintero Tobar, A., Lahera Vargas, M., Jiménez Blanco, S., et al. (2026) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2026.113505
- Authors
- MacLeod, J., Im, G. H., Smith, M., & Vigersky, R. A.
- Affiliations
- Medtronic Diabetes, Northridge, California, USA.
- Corresponding author
- MacLeod, Janice
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2024-01
- Volume
- 26
- Issue
- 1
- Pages
- 33-39
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- PMC
- PMC10794824
- OpenAlex
- W4387765534
- Cited by
- 40
- References
- 25
- Citations by year
- 2026: 13 · 2025: 11 · 2024: 16
- FWCI
- 6.88
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Diabetes Technology & Therapeutics (publishedVersion, cc-by)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion, cc-by)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- smart insulin pens, connected insulin pens, continuous glucose monitoring, type 1 diabetes, type 2 diabetes, glycemic outcomes, insulin dosing frequency
- MeSH terms
- Adult, Blood Glucose, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Humans, Hypoglycemic Agents, Insulin, Retrospective Studies, Blood Glucose Self-Monitoring, Insulin, Regular, Human
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review