Observed Glycemic and Psychosocial Benefits in the Prospective Bigfoot Unity Real World Study: A 6-Month Analysis
Open accessAbstract
Context: The Bigfoot Unity Diabetes Management System integrates Abbott FreeStyle Libre 2 continuous glucose monitoring (CGM) data into a smart insulin pen cap and mobile app, enabling clinician-directed insulin dose recommendations and real-time alerts. Objective: The objective was to analyze real-world 6-month glycemic control in a prospective study for individuals using the System for multiple daily insulin injections (MDI). Methods: We conducted a 6-month analysis from the BURST study (NCT05088265) of individuals with type 1 or type 2 diabetes (T2D). Participants reported baseline demographics, adverse events, and other survey data electronically. Either at-home kit or electronic medical record glycated hemoglobin A1c (HbA1c) data were collected. Results: Of 102 participants in the per protocol cohort, median age was 59 years, 87% had T2D, 42% used CGM previously, 62% were White non-Hispanic, and 59% female. Mean HbA1c decreased from 9.1 ± 1.7% at baseline to 8.0 ± 1.2% at 6 months (mean difference −1.1%, 95% CI −1.4 to −0.8, P < .001). At 6 months, time in range (70-180 mg/dL), time at < 70 mg/dL, and time at < 54 mg/dL were 56 ± 23%, 1.0 ± 1.4%, and 0.04 ± 0.14%, respectively. Six severe hypoglycemia events occurred in 4 participants (none System-related) and no diabetic ketoacidosis events occurred in the per protocol cohort. Conclusion: In this study primarily of older adults with T2D using MDI, durable glycemic improvement occurred using the System at 6 months, with the frequency of hypoglycemia being substantially below established targets of < 4% and < 1% for time below 70 and 54 mg/dL, respectively.
AI snapshot
Written from the paper’s full text.
TL;DR
In a prospective, single-arm 6-month analysis of the Bigfoot Unity Diabetes Management System in adults using multiple daily insulin injections, mean HbA1c fell from 9.1% to 8.0% in the per protocol cohort. Glycemic improvement was accompanied by low rates of hypoglycemia and significant improvements in diabetes distress and hypoglycemia confidence.
Key findings
- In the per protocol cohort (N = 102), mean HbA1c decreased from 9.1 ± 1.7% at baseline to 8.0 ± 1.2% at 6 months (mean difference −1.1%, 95% CI −1.4 to −0.8, P < .001). In the expanded cohort (N = 171), mean HbA1c decreased from 9.4 ± 1.9% to 8.1 ± 1.4% (mean change −1.3%, 95% CI −1.6 to −1.0, P < .001).
- At 6 months in the per protocol cohort, time in range (70-180 mg/dL) was 56 ± 23%, time < 70 mg/dL was 1.0 ± 1.4%, and time < 54 mg/dL was 0.04 ± 0.14%. Six severe hypoglycemia events occurred in 4 participants (none System-related), and no diabetic ketoacidosis events occurred in the per protocol cohort.
- Patient-reported outcomes improved: mean T1-DDS total score fell from 2.4 ± 0.9 to 1.7 ± 0.6 (mean difference −0.7, 95% CI −0.9 to −0.5, P < .001), mean HCS rose from 2.8 ± 0.7 to 3.1 ± 0.7 (mean difference 0.4, 95% CI 0.2 to 0.5, P < .001), and mean SUS score was 80.7 at 6 months.
- In the safety cohort (N = 232), severe hypoglycemia incidence was 9.0 events per 100 person-years and DKA incidence was 0.6 events per 100 person-years; there were no CGM-related adverse reactions requiring treatment, and no serious adverse events were considered related to System use by the medical monitor.
Methods
- Prospective, single-arm observational study (BURST; NCT05088265) enrolling participants with type 1 or type 2 diabetes prescribed Bigfoot Unity for multiple daily insulin injections between October 2021 and July 2023. The per protocol cohort included participants with baseline and 6-month HbA1c and at least 50% CGM data availability at 6 months (N = 102); the expanded cohort included those with baseline HbA1c and any HbA1c or glucose management indicator after day 68 (N = 171); the safety cohort included all who enrolled and completed training (N = 232).
- Data included electronic questionnaires, at-home capillary HbA1c kits or electronic medical record HbA1c, CGM and System use data, patient-reported outcomes (T1-DDS, HCS, SUS), and safety events. Main analyses used longitudinal linear regression for HbA1c and PRO change, with robust regression and multiple imputation for skewed changes, adaptive Benjamini-Hochberg false discovery rate control, and P < .05 considered statistically significant.
Limitations
- The paper states: single-arm design with no concurrent control group; relatively small per protocol sample; no baseline CGM metrics; primarily self-reported data; incomplete System data for some participants (45 of 216, 21%, who consented, started using the System, and were not screening failures had very limited data to analyze); and a small type 1 diabetes subgroup that limits interpretation.
- Evident from the text: the pre-post observational design cannot establish causality, and the authors note that the larger improvement in those with baseline HbA1c ≥ 8% may reflect regression to the mean.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
34 of 50 cited sources checked · 15 supported · 1 flagged · accuracy not yet measured
- Interpretationcitation
Occurrences of severe hypoglycemia and DKA (9.0 and 0.6 events per 100 person-years, respectively, in the safety cohort) observed during the study period are in line with other literature with estimates ranging from 7 to 130 and 4 to 8 events per 100 person-years, respectively, in insulin-requiring diabetes populations using MDI therapy.
“Occurrences of severe hypoglycemia and DKA (9.0 and 0.6 events per 100 person-years, respectively, in the safety cohort) observed during the study period are in line with other literature with estimates ranging from 7 to 130 and 4 to 8 events per 100 person-years, respectively, in insulin-requiring diabetes populations using MDI therapy (39-43).”
Quote in the paperNot a substantial claim3 of 5 sources checked- cites [39] UK Hypoglycaemia Study Group, 2007, Risk of hypoglycaemia in types 1 and 2 diabetes: effects of treatment modalities and their duration, Diabetologia source
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured - cites [40] Karges, 2017, Association of insulin pump therapy vs insulin injection therapy with severe hypoglycemia, ketoacidosis, and glycemic control among children, adolescents, and young adults with type 1 diabetes, JAMA source
- cites [41] Donnelly, 2005, Frequency and predictors of hypoglycaemia in type 1 and insulin-treated type 2 diabetes: a population-based study, Diabet Med source
- cites [42] Khunti, 2016, Rates and predictors of hypoglycaemia in 27 585 people from 24 countries with insulin-treated type 1 and type 2 diabetes: the global HAT study, Diabetes Obes Metab source
Split into 2 atoms — 0 confirmed, 1 partially supported, 1 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured - cites [43] Beck, 2012, The T1D exchange clinic registry, J Clin Endocrinol Metab source
- Premisecitation
Injection therapy is used by more than 80% of people with insulin-requiring diabetes.
“As injection therapy is used by more than 80% of people with insulin-requiring diabetes, there is a clear need for intuitive, integrated systems to help multiple daily injection (MDI) patients interpret CGM data into actionable management steps to reduce the burden of demanding self-care directives and support telemedicine (3, 6, 10-13).”
Quote in the paperSubstantial claim4 of 6 sources checked- cites [3] Peyrot, 2012, Insulin adherence behaviours and barriers in the multinational global attitudes of patients and physicians in insulin therapy study, Diabet Med source
Table 1 (Participants and methods > Statistical analysis; page unavailable) reports that 98.5% of the 1,530 surveyed insulin users used a pen, syringe, or both, supporting the claim for this sample. The Study design section says participants were recruited from research-consumer panels; the paper does not establish that more than 80% of all people with insulin-requiring diabetes use injection therapy. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [6] Tanenbaum, 2017, Optimal use of diabetes devices: clinician perspectives on barriers and adherence to device use, J Diabetes Sci Technol source
The Results (Sample Characteristics; page unavailable) report clinicians’ estimates of insulin-pump use among their patients with type 1 diabetes, not the prevalence of injection therapy among all people with insulin-requiring diabetes. Searches for “more than 80%,” injection therapy, and related terms found no passage establishing the claimed figure. (LLM, conf=0.72)
Experimental — accuracy not yet measured - cites [10] Polonsky, 2005, Assessing psychosocial distress in diabetes: development of the diabetes distress scale, Diabetes Care source
The Results section reports that 50.4% of patients in three study samples used insulin, but does not say whether they used injections. Searches for injection therapy, insulin-requiring diabetes, and the claimed 80% figure found no passage addressing the claim. Page numbers were not available in our copy. (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [11] Steenkamp, 2022, Adherence and persistence to insulin therapy in people with diabetes: impact of connected insulin pen delivery ecosystem, J Diabetes Sci Technol source
- cites [12] Szmuilowicz, 2022, Stepwise approach to continuous glucose monitoring interpretation for internists and family physicians, Postgrad Med source
- cites [13] Klonoff, 2018, Smart pens will improve insulin therapy, J Diabetes Sci Technol source
The paper’s discussion of insulin delivery (page not provided) says pens are the most widely used devices worldwide. It reports pump uptake of 12%–40% among people with type 1 diabetes in specified regions and says uptake is exceedingly low among people with type 2 diabetes. This supports the general point that injection therapy is common, but the paper does not establish the claimed figure of more than 80% of all people with insulin-requiring diabetes. (LLM, conf=0.84)
Experimental — accuracy not yet measured
- Premisecitation
There is evidence that patients may not know how to interpret what to do with CGM data.
“While recent guidelines call for continuous glucose monitoring (CGM) use for all who require insulin, regardless of diabetes type (7), there is also evidence that patients may not know how to interpret what to do with the CGM data (6, 8, 9).”
Quote in the paperSubstantial claim2 of 3 sources checked- cites [6] Tanenbaum, 2017, Optimal use of diabetes devices: clinician perspectives on barriers and adherence to device use, J Diabetes Sci Technol source
- cites [8] Goldman, 2022, Continuous glucose monitoring metrics: interpretation and integrations, Am J Manag Care source
- cites [9] Lanning, 2020, Barriers to continuous glucose monitoring in people with type 1 diabetes: clinician perspectives, Diabetes Spectr source
- Interpretationcitation
The mean HbA1c, GMI, TIR, and TBR results (8.0%, 7.6%, 56%, and 1.0% at 6 months, respectively) demonstrate close adherence to established glycemic targets.
“Considering the older age of participants and comorbid conditions reported, the mean HbA1c, GMI, TIR, and TBR results (8.0%, 7.6%, 56%, and 1.0% at 6 months, respectively) demonstrate close adherence to established glycemic targets (23, 24, 38).”
Quote in the paperSubstantial claim3 of 3 sources checked- cites [23] American Diabetes Association Professional Practice Committee, 2021, 13. Older adults: standards of medical care in diabetes—2022, Diabetes Care source
The Treatment Goals recommendations (section 13.3, page unavailable) set an individualized A1C goal of less than 8.0% for older adults with multiple coexisting conditions, so the reported mean of 8.0% is close to that target. The cited paper does not establish corresponding GMI, TIR, or TBR targets that would support describing all four results as close to established targets. The six-month results are the citing authors’ data, not results attributed to this source. (LLM, conf=0.85)
Experimental — accuracy not yet measured - cites [24] American Diabetes Association Professional Practice Committee, 2021, 6. Glycemic targets: standards of medical care in diabetes—2022, Diabetes Care source
The Recommendations (p. 5) establish glycemic targets, and a TBR of 1.0% is within the stated <4% target. But the cited paper gives a TIR target of >70% for many nonpregnant adults, well above the reported 56%, and describes an A1C goal of <8% as a possible less stringent goal for selected patients—not a general target met by an A1C of 8.0%. It does not set a numerical GMI target or report the citing study’s six-month results, so it does not support the assessment that all four results closely adhere to its targets. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites [38] Battelino, 2019, Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range, Diabetes Care source
In the discussion of older and high-risk individuals (page not provided), the consensus sets a TIR target of >50% and a TBR target of <1%; the cited 56% TIR meets the first, while 1.0% TBR is close to but does not meet the second. The Relationship Between TIR and A1C section associates roughly 50% TIR with an A1C of approximately 8%, but the paper does not establish a GMI target that would substantiate the claim for all four measures. (LLM, conf=0.86)
Experimental — accuracy not yet measured
- Interpretationreasoning
Comparing the average improvement in the current study with the MCID suggests that for the majority of participants, there may have been clinically meaningful improvement.
“To place context around these results, note that the estimated minimally clinically important difference (MCID) for the 28-item T1-DDS total score is 0.19 in an insulin-requiring T1D population (19); comparing the average improvement in the current study with the MCID suggests that for the majority of participants, there may have been clinically meaningful improvement.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [19] Fisher, 2016, Diabetes distress in adults with type 1 diabetes: prevalence, incidence and change over time, J Diabetes Complications source
- Rests on an assumption: The MCID for the T1-DDS total score derived in an insulin-requiring T1D population applies to the current mixed T1D/T2D cohort.
- Interpretationcitation
The observed reduction in HbA1c did not compromise the population from meeting established hypoglycemia targets of <4% and <1% for percent time <70 mg/dL (mean 1.0%) and <54 mg/dL (mean 0.04%), respectively.
“The observed reduction in HbA1c did not compromise the population from meeting established hypoglycemia targets of < 4% and < 1% for percent time < 70 mg/dL (mean 1.0%) and < 54 mg/dL (mean 0.04%), respectively (22-24).”
Quote in the paperSubstantial claim2 of 3 sources checked- cites [22] Battelino, 2023, Continuous glucose monitoring and metrics for clinical trials: an international consensus statement, Lancet Diabetes Endocrinol source
- cites [23] American Diabetes Association Professional Practice Committee, 2021, 13. Older adults: standards of medical care in diabetes—2022, Diabetes Care source
The Hypoglycemia and 13.3 sections discuss reducing hypoglycemia risk and report a separate trial’s time below 70 mg/dL. They do not state the <4% and <1% time-below-range targets or report the citing population’s means of 1.0% and 0.04%. I searched for those targets and for time below 70 and 54 mg/dL. (LLM, conf=0.76)
Experimental — accuracy not yet measured - cites [24] American Diabetes Association Professional Practice Committee, 2021, 6. Glycemic targets: standards of medical care in diabetes—2022, Diabetes Care source
- Findingown data
In a subset of participants with baseline HbA1c ≥ 8%, the magnitude of the observed difference was larger (approximately 1.5%).
“Similar to other investigations (34, 37), in a subset of participants with suboptimal glycemic control (those with baseline HbA1c ≥ 8%), the magnitude of the observed difference was larger (approximately 1.5% here) which may reflect regression to the mean phenomenon.”
Quote in the paperNot a substantial claim2 of 2 sources checked- cites [34] Baliga, 2023, First real-world experience with bigfoot unity: a 6-month retrospective analysis, Clin Diabetes source
- cites [37] Billings, 2018, Baseline glycated hemoglobin values predict the magnitude of glycemic improvement in patients with type 1 and type 2 diabetes: subgroup analyses from the DIAMOND study program, Diabetes Technol Ther source
- Premisecitation
Diabetes distress has been shown to be relatively stable over time, unless some form of intervention takes place.
“Diabetes distress has been shown to be relatively stable over time, unless some form of intervention takes place (19).”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [19] Fisher, 2016, Diabetes distress in adults with type 1 diabetes: prevalence, incidence and change over time, J Diabetes Complications source
- Premisecitation
The estimated minimally clinically important difference (MCID) for the 28-item T1-DDS total score is 0.19 in an insulin-requiring T1D population.
“To place context around these results, note that the estimated minimally clinically important difference (MCID) for the 28-item T1-DDS total score is 0.19 in an insulin-requiring T1D population (19); comparing the average improvement in the current study with the MCID suggests that for the majority of participants, there may have been clinically meaningful improvement.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [19] Fisher, 2016, Diabetes distress in adults with type 1 diabetes: prevalence, incidence and change over time, J Diabetes Complications source
- Findingown data
In the per protocol cohort, among 13 participants with T1D, mean HbA1c decreased from 9.0 ± 1.1% to 8.3 ± 1.0% (mean change −0.7%, 95% CI −1.5% to 0.1%).
“Mean HbA1c decreased from 9.0% ± 1.1% to 8.3% ± 1.0% (mean change −0.7%, 95% CI −1.5% to 0.1%) for 13 participants with T1D and decreased from 9.1% ± 1.8% to 7.9% ± 1.3% (mean change −1.2%, 95% CI −1.5% to −0.8%) for the 89 participants with T2D.”
Quote in the paperSubstantial claim- Rests on an assumption: At-home capillary blood HbA1c values and electronic medical record HbA1c values are comparable for the analysis.
- Findingown data
In the per protocol cohort, among 89 participants with T2D, mean HbA1c decreased from 9.1 ± 1.8% to 7.9 ± 1.3% (mean change −1.2%, 95% CI −1.5% to −0.8%).
“Mean HbA1c decreased from 9.0% ± 1.1% to 8.3% ± 1.0% (mean change −0.7%, 95% CI −1.5% to 0.1%) for 13 participants with T1D and decreased from 9.1% ± 1.8% to 7.9% ± 1.3% (mean change −1.2%, 95% CI −1.5% to −0.8%) for the 89 participants with T2D.”
Quote in the paperSubstantial claim- Rests on an assumption: At-home capillary blood HbA1c values and electronic medical record HbA1c values are comparable for the analysis.
- Findingown data
In the expanded cohort, the mean HbA1c decrease was −1.4% (95% CI −1.7% to −1.1%) in 139 participants with T2D.
“The mean decrease was −0.9% (95% CI −1.5% to −0.4%) in 32 participants with T1D and −1.4% (95% CI −1.7% to −1.1%) in the 139 participants with T2D.”
Quote in the paperSubstantial claim- Rests on an assumption: At-home capillary blood HbA1c values and electronic medical record HbA1c values are comparable for the analysis.
- Findingown data
In the per protocol cohort at 6 months, time below range <70 mg/dL was 1.0 ± 1.4%.
“At 6 months, time in range (70-180 mg/dL), time at < 70 mg/dL, and time at < 54 mg/dL were 56 ± 23%, 1.0 ± 1.4%, and 0.04 ± 0.14%, respectively.”
- Also stated in: § Glycemic Outcomes · plain
“Time below range (TBR; < 70 mg/dL) was 0.6% ± 1.2%, 0.8% ± 1.2%, and 1.0% ± 1.4% and time < 54 mg/dL was 0.01% ± 0.10%, 0.03% ± 0.08%, and 0.04% ± 0.14% in the first 2 weeks, the third month, and the sixth month periods, respectively (Table 3).”
Quote in the paperNot a substantial claim - Also stated in: § Glycemic Outcomes · plain
- Findingown data
In the expanded cohort, the mean HbA1c decrease was −0.9% (95% CI −1.5% to −0.4%) in 32 participants with T1D.
“The mean decrease was −0.9% (95% CI −1.5% to −0.4%) in 32 participants with T1D and −1.4% (95% CI −1.7% to −1.1%) in the 139 participants with T2D.”
Quote in the paperSubstantial claim- Rests on an assumption: At-home capillary blood HbA1c values and electronic medical record HbA1c values are comparable for the analysis.
- Findingown data
In the per protocol cohort (N=102), mean HbA1c decreased from 9.1 ± 1.7% at baseline to 8.0 ± 1.2% at 6 months, with a mean difference of −1.1% (95% CI −1.4 to −0.8, P < .001).
“Mean HbA1c decreased from 9.1 ± 1.7% at baseline to 8.0 ± 1.2% at 6 months (mean difference −1.1%, 95% CI −1.4 to −0.8, P < .001).”
- Also stated in: § Glycemic Outcomes · plain
“In the per protocol cohort, mean HbA1c was 9.1 ± 1.7% prior to System use and decreased to 8.0% ± 1.2% at 6 months (mean difference −1.1%, 95% CI −1.4% to −0.8%, P < .001).”
Quote in the paperSubstantial claim- Rests on an assumption: At-home capillary blood HbA1c values and electronic medical record HbA1c values are comparable for the analysis.
- Also stated in: § Glycemic Outcomes · plain
- Findingown data
In the per protocol cohort, time <54 mg/dL was 0.01 ± 0.10% in the first 2 weeks, 0.03 ± 0.08% in the third month, and 0.04 ± 0.14% in the sixth month.
“Time below range (TBR; < 70 mg/dL) was 0.6% ± 1.2%, 0.8% ± 1.2%, and 1.0% ± 1.4% and time < 54 mg/dL was 0.01% ± 0.10%, 0.03% ± 0.08%, and 0.04% ± 0.14% in the first 2 weeks, the third month, and the sixth month periods, respectively (Table 3).”
Quote in the paperNot a substantial claim - Findingown data
In the per protocol cohort, no diabetic ketoacidosis events occurred.
“Six severe hypoglycemia events occurred in 4 participants (none System-related) and no diabetic ketoacidosis events occurred in the per protocol cohort.”
Quote in the paperNot a substantial claim - Findingown data
In the expanded cohort, mean HbA1c decreased from 9.4 ± 1.9% (N=171) prior to System use to 8.1 ± 1.4% (N=148) at 6 months (mean change −1.3%, 95% CI −1.6% to −1.0%, P < .001).
“For the expanded cohort, similar improvement was observed. Mean HbA1c was 9.4% ± 1.9% (N = 171) prior to System use, decreasing to 8.1% ± 1.4% (N = 148) at 6 months (mean change −1.3%, 95% CI −1.6% to −1.0%, P < .001).”
Quote in the paperSubstantial claim- Rests on an assumption: At-home capillary blood HbA1c values and electronic medical record HbA1c values are comparable for the analysis.
- Findingown data
In the per protocol cohort, mean GMI was 7.7 ± 1.0% in the first 2 weeks of use and 7.6 ± 1.0% in the sixth month of use.
“Mean GMI in the first 2 weeks of use was 7.7% ± 1.0% and 7.6% ± 1.0% in the sixth month of use (Fig. 2).”
Quote in the paperNot a substantial claim - Findingown data
In the per protocol cohort at 6 months, time below range <54 mg/dL was 0.04 ± 0.14%.
“At 6 months, time in range (70-180 mg/dL), time at < 70 mg/dL, and time at < 54 mg/dL were 56 ± 23%, 1.0 ± 1.4%, and 0.04 ± 0.14%, respectively.”
- Also stated in: § Glycemic Outcomes · plain
“Time below range (TBR; < 70 mg/dL) was 0.6% ± 1.2%, 0.8% ± 1.2%, and 1.0% ± 1.4% and time < 54 mg/dL was 0.01% ± 0.10%, 0.03% ± 0.08%, and 0.04% ± 0.14% in the first 2 weeks, the third month, and the sixth month periods, respectively (Table 3).”
Quote in the paperNot a substantial claim - Also stated in: § Glycemic Outcomes · plain
- Findingown data
In the per protocol cohort at 6 months, time in range (70-180 mg/dL) was 56 ± 23%.
“At 6 months, time in range (70-180 mg/dL), time at < 70 mg/dL, and time at < 54 mg/dL were 56 ± 23%, 1.0 ± 1.4%, and 0.04 ± 0.14%, respectively.”
- Also stated in: § Glycemic Outcomes · plain
“TIR (70-180 mg/dL) in the first 2 weeks, third month, and sixth month of use was 55% ± 25%, 56% ± 24%, and 56% ± 23%, respectively.”
Quote in the paperSubstantial claim - Also stated in: § Glycemic Outcomes · plain
- Findingown data
At 6 months within the expanded cohort, median CGM use was 79% (IQR 62%, 89%).
“At 6 months within the expanded cohort, median CGM use was 79% (IQR 62%, 89%), mean TIR was 54% ± 25%, mean TBR was 1.0% ± 1.5%, and median long- and rapid-acting insulin doses were 6.8 (IQR 5.5, 7.0) per week and 2.3 per day (IQR 1.4, 3.0), respectively.”
Quote in the paperNot a substantial claim - Findingown data
In the expanded cohort, at 6 months, mean difference from baseline in HCS was 0.4 (95% CI 0.2 to 0.5, P < .001).
“For the expanded cohort, similar outcomes were observed. At the 6-month assessment, the mean difference from baseline in the total T1-DDS score was −0.8 (95% CI −0.9 to −0.6, P < .001) and the mean difference in HCS was 0.4 (95% CI 0.2 to 0.5, P < .001).”
Quote in the paperSubstantial claim - Findingown data
At 6 months within the expanded cohort, median rapid-acting insulin doses were 2.3 per day.
“At 6 months within the expanded cohort, median CGM use was 79% (IQR 62%, 89%), mean TIR was 54% ± 25%, mean TBR was 1.0% ± 1.5%, and median long- and rapid-acting insulin doses were 6.8 (IQR 5.5, 7.0) per week and 2.3 per day (IQR 1.4, 3.0), respectively.”
Quote in the paperNot a substantial claim - Findingown data
In the expanded cohort, at 6 months, mean difference from baseline in total T1-DDS score was −0.8 (95% CI −0.9 to −0.6, P < .001).
“For the expanded cohort, similar outcomes were observed. At the 6-month assessment, the mean difference from baseline in the total T1-DDS score was −0.8 (95% CI −0.9 to −0.6, P < .001) and the mean difference in HCS was 0.4 (95% CI 0.2 to 0.5, P < .001).”
Quote in the paperSubstantial claim - Findingown data
In the per protocol cohort, hypoglycemia confidence rose significantly from baseline to 6 months.
“In the per protocol cohort, there was significant drop in total diabetes distress from baseline to 6 months and a corresponding rise in hypoglycemia confidence (Table 4).”
Quote in the paperSubstantial claim - Findingown data
Significant improvement was observed in all 7 T1-DDS subscales.
“Significant improvement in all 7 T1-DDS subscales were also observed; of particular interest were the drops in powerlessness, management distress, and hypoglycemic distress.”
Quote in the paperSubstantial claim - Findingown data
The mean SUS score was 80.7 at the 6-month assessment.
“The mean SUS score was 80.7 at the 6-month assessment (Table 4).”
Quote in the paperNot a substantial claim - Findingown data
After 6 months of System use, among those with elevated baseline distress (≥2.0), 49% (24 of 49) dropped below the cutpoint.
“Among participants completing the survey instrument at both baseline and 6 months, 56% (49 of 88) were found to have elevated distress (ie, above the mean T1-DDS cutpoint of ≥ 2.0) at baseline; after 6 months of System use, in those with elevated baseline distress (≥ 2.0), 49% (24 of 49) had dropped below the cutpoint, suggesting clinically relevant improvement.”
Quote in the paperSubstantial claim - Findingown data
In the per protocol cohort, mean T1-DDS score for those with T1D was 2.7 ± 0.8 at baseline and fell to 2.0 ± 0.9 at 6 months.
“In the per protocol cohort, mean T1-DDS score at baseline for those with T1D was 2.7 ± 0.8 and fell to 2.0 ± 0.9 at 6 months while baseline for T2D was 2.3 ± 0.9 and fell to 1.6 ± 0.6 at 6 months.”
Quote in the paperSubstantial claim - Findingown data
At 6 months within the expanded cohort, mean TBR was 1.0% ± 1.5%.
“At 6 months within the expanded cohort, median CGM use was 79% (IQR 62%, 89%), mean TIR was 54% ± 25%, mean TBR was 1.0% ± 1.5%, and median long- and rapid-acting insulin doses were 6.8 (IQR 5.5, 7.0) per week and 2.3 per day (IQR 1.4, 3.0), respectively.”
Quote in the paperNot a substantial claim - Findingown data
In the per protocol cohort, HCS for those with T1D was 2.9 ± 0.7 at baseline and 2.8 ± 0.8 at 6 months.
“The HCS baseline for those with T1D was 2.9 ± 0.7 and 2.8 ± 0.8 at 6 months while baseline for T2D was 2.8 ± 0.7 and 3.2 ± 0.7 at 6 months.”
Quote in the paperNot a substantial claim - Findingown data
In the per protocol cohort, HCS for those with T2D was 2.8 ± 0.7 at baseline and 3.2 ± 0.7 at 6 months.
“The HCS baseline for those with T1D was 2.9 ± 0.7 and 2.8 ± 0.8 at 6 months while baseline for T2D was 2.8 ± 0.7 and 3.2 ± 0.7 at 6 months.”
Quote in the paperNot a substantial claim - Findingown data
In the per protocol cohort, mean nocturnal CGM-measured hypoglycemia events per week were 0.2, 0.3, and 0.4 in the first 2 weeks, third month, and sixth month, respectively.
“Events occurring at night were 0.2 ± 0.5, 0.3 ± 0.6, and 0.4 ± 0.6, respectively.”
Quote in the paperNot a substantial claim - Findingown data
At 6 months, 71% of the time (median), when a long-acting dose alert was issued, a long-acting dose occurred within 2 hours of receiving the alert.
“At 6 months, 71% percent of the time (median), when a long-acting dose alert was issued, a long-acting dose occurred within 2 hours of receiving the alert (Table 3).”
Quote in the paperNot a substantial claim - Findingown data
In the per protocol cohort, total diabetes distress dropped significantly from baseline to 6 months.
“In the per protocol cohort, there was significant drop in total diabetes distress from baseline to 6 months and a corresponding rise in hypoglycemia confidence (Table 4).”
Quote in the paperSubstantial claim - Findingown data
In the per protocol cohort, HCS scores rose from 2.8 ± 0.7 at baseline to 3.1 ± 0.7 at 6 months, with a mean difference of 0.4 (95% CI 0.2 to 0.5, P < .001).
“HCS scores rose from 2.8 ± 0.7 at baseline to 3.1 ± 0.7 at 6 months (mean difference 0.4, 95% CI 0.2 to 0.5, P < .001).”
Quote in the paperSubstantial claim - Findingown data
In the per protocol cohort, mean T1-DDS score for those with T2D was 2.3 ± 0.9 at baseline and fell to 1.6 ± 0.6 at 6 months.
“In the per protocol cohort, mean T1-DDS score at baseline for those with T1D was 2.7 ± 0.8 and fell to 2.0 ± 0.9 at 6 months while baseline for T2D was 2.3 ± 0.9 and fell to 1.6 ± 0.6 at 6 months.”
Quote in the paperSubstantial claim - Findingown data
At 6 months within the expanded cohort, mean TIR was 54% ± 25%.
“At 6 months within the expanded cohort, median CGM use was 79% (IQR 62%, 89%), mean TIR was 54% ± 25%, mean TBR was 1.0% ± 1.5%, and median long- and rapid-acting insulin doses were 6.8 (IQR 5.5, 7.0) per week and 2.3 per day (IQR 1.4, 3.0), respectively.”
Quote in the paperSubstantial claim - Findingown data
At 6 months within the expanded cohort, median long-acting insulin doses were 6.8 per week.
“At 6 months within the expanded cohort, median CGM use was 79% (IQR 62%, 89%), mean TIR was 54% ± 25%, mean TBR was 1.0% ± 1.5%, and median long- and rapid-acting insulin doses were 6.8 (IQR 5.5, 7.0) per week and 2.3 per day (IQR 1.4, 3.0), respectively.”
Quote in the paperNot a substantial claim - Findingown data
After 6 months, among those with low confidence at baseline, 59% (29 of 49) rose above the 3.0 cutpoint.
“Similarly, using < 3.0 as the mean HCS cutpoint, 56% (49 of 88) indicated relatively low confidence regarding their management of hypoglycemia at baseline; after 6 months, among those with low confidence, 59% (29 of 49) had risen above the 3.0 cutpoint, again pointing to clinically relevant improvement.”
Quote in the paperSubstantial claim - Findingown data
Among participants completing the survey at both baseline and 6 months, 56% (49 of 88) had relatively low confidence (HCS < 3.0) at baseline.
“Similarly, using < 3.0 as the mean HCS cutpoint, 56% (49 of 88) indicated relatively low confidence regarding their management of hypoglycemia at baseline; after 6 months, among those with low confidence, 59% (29 of 49) had risen above the 3.0 cutpoint, again pointing to clinically relevant improvement.”
Quote in the paperNot a substantial claim - Findingown data
Among participants completing the survey at both baseline and 6 months, 56% (49 of 88) had elevated distress (T1-DDS ≥ 2.0) at baseline.
“Among participants completing the survey instrument at both baseline and 6 months, 56% (49 of 88) were found to have elevated distress (ie, above the mean T1-DDS cutpoint of ≥ 2.0) at baseline; after 6 months of System use, in those with elevated baseline distress (≥ 2.0), 49% (24 of 49) had dropped below the cutpoint, suggesting clinically relevant improvement.”
Quote in the paperSubstantial claim - Findingown data
In the safety cohort, 1 DKA event occurred in a person with T2D.
“Additionally, in the safety cohort, 1 DKA event occurred in a person with T2D and 2 other serious hyperglycemic events needing treatment in a healthcare facility occurred in 2 participants, 1 with T1D and 1 with T2D; neither was using the System at the time these events occurred.”
Quote in the paperNot a substantial claim - Findingown data
Among those participants with relatively low confidence in managing hypoglycemia at baseline (mean HCS < 3.0), 59% had risen above the 3.0 cutpoint, indicating at least moderate confidence (≥ 3.0) after using the System for 6 months.
“Similarly, among those participants with relatively low confidence in managing hypoglycemia at baseline (mean HCS < 3.0), 59% had risen above the 3.0 cutpoint, indicating at least moderate confidence (≥ 3.0) after using the System for 6 months.”
Quote in the paperSubstantial claim - Findingown data
Twenty-eight other non-diabetes-related hospitalizations occurred in 20 participants.
“Twenty-eight other non-diabetes-related hospitalizations occurred in 20 participants.”
Quote in the paperNot a substantial claim - Findingown data
In the subset of participants with elevated distress at baseline (mean T1-DDS > 2.0), approximately half (49%) had fallen below the 2.0 cutpoint for elevated distress at 6 months.
“It is therefore notable that in the subset of participants with elevated distress at baseline (mean T1-DDS > 2.0), approximately half, (49%) had fallen below the 2.0 cutpoint for elevated distress at 6 months.”
Quote in the paperSubstantial claim - Findingown data
A similar 1.3% improvement in HbA1c was observed in the expanded population at 6 months where stringent CGM data availability was not required.
“A similar 1.3% improvement was observed in the expanded population at 6 months where stringent CGM data availability was not required.”
Quote in the paperSubstantial claim - Findingown data
None of the serious adverse events were considered related to System use by the medical monitor.
“None of these serious adverse events were considered related to System use by the medical monitor.”
Quote in the paperNot a substantial claim - Findingown data
In the per protocol cohort, a 1.1% improvement in HbA1c was observed after 3 and 6 months from a baseline of 9.1%.
“In the per protocol cohort, a 1.1% improvement in HbA1c was observed after 3 and 6 months from a baseline of 9.1%.”
Quote in the paperSubstantial claim - Findingown data
The incidence of severe hypoglycemia per 100 person-years was 6.7, 5.2, and 9.0 in the per protocol, expanded, and safety cohorts, respectively.
“The incidence of severe hypoglycemia per 100 person-years was 6.7, 5.2, and 9.0 in the per protocol, the expanded, and safety cohorts, respectively.”
Quote in the paperNot a substantial claim - Findingown data
There were no CGM-related adverse reactions requiring treatment reported.
“There were no CGM-related adverse reactions requiring treatment reported.”
Quote in the paperNot a substantial claim - Interpretationown data
In this study primarily of older adults with T2D using MDI, durable glycemic improvement occurred using the System at 6 months.
“In this study primarily of older adults with T2D using MDI, durable glycemic improvement occurred using the System at 6 months, with the frequency of hypoglycemia being substantially below established targets of < 4% and < 1% for time below 70 and 54 mg/dL, respectively.”
Quote in the paperSubstantial claim- Rests on an assumption: The observed glycemic improvement is attributable to use of the Bigfoot Unity System rather than to other factors.
- Interpretationown data
The rise above the HCS cutpoint points to clinically relevant improvement.
“Similarly, using < 3.0 as the mean HCS cutpoint, 56% (49 of 88) indicated relatively low confidence regarding their management of hypoglycemia at baseline; after 6 months, among those with low confidence, 59% (29 of 49) had risen above the 3.0 cutpoint, again pointing to clinically relevant improvement.”
Quote in the paperSubstantial claim - Interpretationown data
The expanded cohort had similar TIR and use outcomes to the per protocol cohort despite not having strict CGM use requirements for analysis.
“The expanded cohort had similar TIR and use outcomes despite not having strict CGM use requirements for analysis.”
Quote in the paperNot a substantial claim - Interpretationown data
The drop below the distress cutpoint suggests clinically relevant improvement.
“Among participants completing the survey instrument at both baseline and 6 months, 56% (49 of 88) were found to have elevated distress (ie, above the mean T1-DDS cutpoint of ≥ 2.0) at baseline; after 6 months of System use, in those with elevated baseline distress (≥ 2.0), 49% (24 of 49) had dropped below the cutpoint, suggesting clinically relevant improvement.”
Quote in the paperSubstantial claim - Interpretationown data
The average of 1.5 long-acting dose alerts per person per week in the sixth month may be interpreted as between 1 and 2 potentially forgotten long-acting doses.
“The average of 1.5 long-acting dose alerts per person per week in the sixth month may be interpreted as between 1 and 2 potentially forgotten long-acting doses; data that about 71% of such alerts received are followed by a dose taken within 2 hours of receiving the alert supports good engagement with the System in helping participants to remember to take their long-acting insulin.”
Quote in the paperNot a substantial claim - Interpretationown data
Data that about 71% of such alerts received are followed by a dose taken within 2 hours supports good engagement with the System in helping participants to remember to take their long-acting insulin.
“The average of 1.5 long-acting dose alerts per person per week in the sixth month may be interpreted as between 1 and 2 potentially forgotten long-acting doses; data that about 71% of such alerts received are followed by a dose taken within 2 hours of receiving the alert supports good engagement with the System in helping participants to remember to take their long-acting insulin.”
Quote in the paperSubstantial claim - Premiseown data
None of the 55 articles from the PubMed search addressed prospective collection of clinically relevant adverse events such as severe hypoglycemia or DKA.
“A PubMed search using the terms connected insulin pen or smart insulin pen on February 27, 2024, returned 55 articles, of which only 11 had relevant effectiveness or safety-related data for connected insulin pen systems (most were reviews of or perspectives on smart pen technology); surprisingly, none addressed the prospective collection of clinically relevant adverse events such as severe hypogly”
Quote in the paperSubstantial claim - Interpretationown data
These data suggest that HCP-directed meal and correction dose recommendations along with CGM contribute to the reduction from a mean baseline HbA1c of 9.1% to a mean of 8.0% after 6 months of use.
“These data suggest that HCP-directed meal and correction dose recommendations along with CGM contribute to the reduction from a mean baseline HbA1c of 9.1% to a mean of 8.0% after 6 months of use.”
Quote in the paperSubstantial claim - Interpretationcitation
These findings are consistent with other studies in similar MDI populations.
“These findings are consistent with other studies in similar MDI populations (44, 45).”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [44] Beck, 2017, Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial, JAMA source
- cites [45] Beck, 2017, Continuous glucose monitoring versus usual care in patients with type 2 diabetes receiving multiple daily insulin injections: a randomized trial, Ann Intern Med source
- Interpretationcitation
The 6-month mean SUS of 80.7 indicates that the System supports an above average user experience.
“For context for the SUS score, scores > 80 are considered excellent system use (48) and thus, the 6-month mean of 80.7 indicates that the System supports an above average user experience.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [48] Lewis, 2018, Item benchmarks for the system usability scale, J User Exp source
- Interpretationown data
The authors do not believe that the incompleteness of data for some participants was an appreciable source of bias.
“Thus, we do not believe that the incompleteness of data for some participants was an appreciable source of bias.”
Quote in the paperSubstantial claim - Premiseown data
The authors report the first prospective real-world data from individuals with diabetes using an MDI regimen and the Bigfoot Unity System for 6 months.
“In conclusion, we report the first prospective real-world data from individuals with diabetes using an MDI regimen and the Bigfoot Unity System for 6 months.”
Quote in the paperSubstantial claim - Interpretationown data
For this cohort, consisting primarily of older adults with T2D and having suboptimal glycemic control, using the System resulted in rapid and durable improvement in glycemic control.
“For this cohort, consisting primarily of older adults with T2D and having suboptimal glycemic control, using the System resulted in rapid and durable improvement in glycemic control.”
Quote in the paperSubstantial claim- Rests on an assumption: The observed improvements in glycemic control and psychosocial outcomes are attributable to use of the Bigfoot Unity System rather than other factors.
- Interpretationown data
Improvement in glycemic control, accompanied by psychosocial benefits, did not appear to compromise the cohort in meeting established standards for hypoglycemia and DKA.
“Improvement in glycemic control, accompanied by psychosocial benefits, did not appear to compromise the cohort in meeting established standards for hypoglycemia and DKA.”
Quote in the paperSubstantial claim
- Classified as probably not a substantial claim. 28
- May combine two claims. 34
- The quote only partly states this claim. 3
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
PremisereasoningThere is a clear need for intuitive, integrated systems to help MDI patients interpret CGM data into actionable management steps to reduce the burden of demanding self-care directives and support telemedicine.
“As injection therapy is used by more than 80% of people with insulin-requiring diabetes, there is a clear need for intuitive, integrated systems to help multiple daily injection (MDI) patients interpret CGM data into actionable management steps to reduce the burden of demanding self-care directives and support telemedicine (3, 6, 10-13).”
Quote in the paperNot a substantial claim4 of 6 sources checked- cites [3] Peyrot, 2012, Insulin adherence behaviours and barriers in the multinational global attitudes of patients and physicians in insulin therapy study, Diabet Med source
The Discussion’s Clinical implications section says patients have difficulty adjusting insulin doses to daily changes and recommends reducing the burden of restrictive regimens. It does not support the specific need for integrated systems to interpret CGM data, support telemedicine, or the sentence’s more-than-80% figure. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites [6] Tanenbaum, 2017, Optimal use of diabetes devices: clinician perspectives on barriers and adherence to device use, J Diabetes Sci Technol source
The Results report that clinicians saw patients’ difficulty understanding device information as a barrier, and the Conclusions describe complex adherence tasks and suggest remote monitoring and practical support. The paper does not specifically call for intuitive, integrated systems for MDI patients that turn CGM data into management steps or support telemedicine. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [10] Polonsky, 2005, Assessing psychosocial distress in diabetes: development of the diabetes distress scale, Diabetes Care source
The introduction (section labeled “L”; page unavailable) says that complex, demanding self-care directives can leave people with diabetes overwhelmed or discouraged. It does not address MDI patients, CGM interpretation, integrated systems, or telemedicine, so it supports only the self-care-burden rationale. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites [11] Steenkamp, 2022, Adherence and persistence to insulin therapy in people with diabetes: impact of connected insulin pen delivery ecosystem, J Diabetes Sci Technol source
- cites [12] Szmuilowicz, 2022, Stepwise approach to continuous glucose monitoring interpretation for internists and family physicians, Postgrad Med source
- cites [13] Klonoff, 2018, Smart pens will improve insulin therapy, J Diabetes Sci Technol source
In “The Future for Smart Pens,” the paper envisions connected devices that integrate diabetes data and support insulin-dose decisions; its Conclusion says decision support could reduce the day-to-day burden of diabetes care. It does not specifically discuss helping MDI patients turn CGM data into actionable steps or supporting telemedicine. It also does not give the citing sentence’s more-than-80% figure for injection therapy. (LLM, conf=0.86)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S2 (checkable).
PremisecitationFor people with insulin-requiring diabetes, dosing instructions may be complex and confusing.
“For people with insulin-requiring diabetes, dosing instructions may be complex and confusing (3-6).”
Quote in the paperNot a substantial claim2 of 4 sources checked- cites [3] Peyrot, 2012, Insulin adherence behaviours and barriers in the multinational global attitudes of patients and physicians in insulin therapy study, Diabet Med source
The Discussion reports dissatisfaction with regimen simplicity, and the Clinical implications section says patients have difficulty adjusting insulin doses to daily changes. These passages support the possibility that dosing is complex, but they do not say that dosing instructions themselves are confusing. No page numbers are available in our copy. (LLM, conf=0.82)
Experimental — accuracy not yet measured - cites [4] Farsaei, 2014, Insulin adherence in patients with diabetes: risk factors for injection omission, Prim Care Diabetes source
- cites [5] Page, 1981, Patient recall of self-care recommendations in diabetes, Diabetes Care source
- cites [6] Tanenbaum, 2017, Optimal use of diabetes devices: clinician perspectives on barriers and adherence to device use, J Diabetes Sci Technol source
The Abstract (page unavailable) describes insulin-pump use as requiring complex tasks, including carbohydrate counting and administering boluses. The Results (page unavailable) report clinicians’ perceptions that some patients do not understand pump information or features. These passages support complexity and possible confusion in device use, but do not specifically say that dosing instructions are confusing for people with insulin-requiring diabetes generally. (LLM, conf=0.83)
Experimental — accuracy not yet measured
May combine two claims.
InterpretationcitationMean percent time of sensor use and mean coefficient of variation (84% and 31% at 6 months, respectively) are within established guidelines.
“Other parameters such as the mean percent time of sensor use and mean coefficient of variation (84% and 31% at 6 months, respectively) are also within established guidelines (23, 24, 38).”
Quote in the paperNot a substantial claim3 of 3 sources checked- cites [23] American Diabetes Association Professional Practice Committee, 2021, 13. Older adults: standards of medical care in diabetes—2022, Diabetes Care source
The section on continuous glucose monitoring (13.3 Screening for early detection; page unavailable) reports a six-month trial and reduced glycemic variability, but gives no guideline threshold for sensor use or coefficient of variation. Searches for the 84% and 31% figures and for those metrics’ guidelines found no passage establishing that they are within guidelines. (LLM, conf=0.73)
Experimental — accuracy not yet measured - cites [24] American Diabetes Association Professional Practice Committee, 2021, 6. Glycemic targets: standards of medical care in diabetes—2022, Diabetes Care source
- cites [38] Battelino, 2019, Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range, Diabetes Care source
May combine two claims.
FindingcitationLow and very low glucose events occurred at a low rate (mean of ∼1.3 and 0.2 events per person per week in the sixth month, respectively) and were in keeping with time below range meeting established standards.
“Low and very low glucose events occurred at a low rate (mean of ∼1.3 and 0.2 events per person per week in the sixth month, respectively) and were in keeping with time below range meeting established standards (23, 24, 38).”
Quote in the paperSubstantial claim3 of 3 sources checked- cites [23] American Diabetes Association Professional Practice Committee, 2021, 13. Older adults: standards of medical care in diabetes—2022, Diabetes Care source
The Hypoglycemia section gives a threshold for low glucose, and section 13.3 reports time spent below 70 mg/dL in a separate six-month CGM trial. Neither passage gives the claimed sixth-month event rates or a time-below-range standard against which to assess them. I searched for those rates and for time-below-range targets in our copy. (LLM, conf=0.72)
Experimental — accuracy not yet measured - cites [24] American Diabetes Association Professional Practice Committee, 2021, 6. Glycemic targets: standards of medical care in diabetes—2022, Diabetes Care source
- cites [38] Battelino, 2019, Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range, Diabetes Care source
May combine two claims.
InterpretationcitationThe prospective 6-month study reported here showed similar results to 3-month data presented previously and similar glycemic control to a retrospective cohort using Bigfoot Unity.
“In general, the prospective 6-month study reported here showed similar results to 3-month data presented previously (35) and similar glycemic control to a retrospective cohort using Bigfoot Unity (34); here, the baseline HbA1c was higher (9.1% vs 8.4%), which may reflect health status or other demographic biases in those who consent to participate in an ongoing study vs those who do not (36).”
Quote in the paperNot a substantial claim2 of 3 sources checked- cites [35] Malone, 2023, 936-P: three-month interim analysis from the prospective bigfoot unity real world study (BURST), Diabetes source
- cites [34] Baliga, 2023, First real-world experience with bigfoot unity: a 6-month retrospective analysis, Clin Diabetes source
- cites [36] Kho, 2009, Written informed consent and selection bias in observational studies using medical records: systematic review, BMJ source
May combine two claims.
Study design and conductcitation45 of 216 (21%) who consented, started using the System and were not screening failures, had very limited data to analyze; this percentage with incomplete data could introduce bias and would be considered relatively high in a controlled trial but is anticipated in the real-world setting in an older population with comorbid conditions.
“For example, 45 of 216 (21%) who consented, started using the System and were not screening failures, had very limited data to analyze; this percentage with incomplete data could introduce bias and would be considered relatively high in a controlled trial but is anticipated in the real-world setting in an older population with comorbid conditions (49).”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [49] Jacobsen, 2021, Predictors of attrition in a longitudinal population-based study of aging, Int Psychogeriatr source
Split into 2 atoms — 1 confirmed, 1 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
Experimental — accuracy not yet measured
May combine two claims.
Findingown dataIn the per protocol cohort, TIR (70-180 mg/dL) was 55 ± 25% in the first 2 weeks, 56 ± 24% in the third month, and 56 ± 23% in the sixth month.
“TIR (70-180 mg/dL) in the first 2 weeks, third month, and sixth month of use was 55% ± 25%, 56% ± 24%, and 56% ± 23%, respectively.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn the per protocol cohort, median sensor use was 87% (IQR 74-93) in the first 2 weeks, 86% (IQR 77-92) in the third month, and 84% (IQR 72-91) in the sixth month.
“Median sensor use was 87% with an interquartile range (IQR) of 74% to 93% in the first 2 weeks; median sensor use was 86% (IQR, 77% to 92%) and 84% (IQR, 72% to 91%) in the per protocol cohort in the third month and sixth month of use, respectively.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn the per protocol cohort, participants with baseline HbA1c ≥8% (N=76) had greater improvement (mean baseline 9.7 ± 1.4% to 8.3 ± 1.2% at 6 months), while those with lower baseline HbA1c had small changes in glycemic control.
“Those with higher baseline HbA1c (≥ 8%; N = 76), had greater improvement (mean baseline HbA1c of 9.7% ± 1.4% improved to 8.3% ± 1.2% at 6 months) while those with lower baseline HbA1c had small changes in glycemic control (Table 2 and Fig. 2C).”
Quote in the paperNot a substantial claim- Rests on an assumption: At-home capillary blood HbA1c values and electronic medical record HbA1c values are comparable for the analysis.
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataIn the per protocol cohort, 42% had used CGM previously.
“Of 102 participants in the per protocol cohort, median age was 59 years, 87% had T2D, 42% used CGM previously, 62% were White non-Hispanic, and 59% female.”
- Also stated in: § Participant Characteristics · plain
“CGM had been previously used by 42% (see Table 1).”
Quote in the paperNot a substantial claim- Also stated in: § Participant Characteristics · plain
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe safety cohort reflects 232 of 250 (93%) of those who enrolled.
“The safety cohort reflects 232 of 250 (93%) of those who enrolled.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn the per protocol cohort, time below range <70 mg/dL was 0.6 ± 1.2% in the first 2 weeks, 0.8 ± 1.2% in the third month, and 1.0 ± 1.4% in the sixth month.
“Time below range (TBR; < 70 mg/dL) was 0.6% ± 1.2%, 0.8% ± 1.2%, and 1.0% ± 1.4% and time < 54 mg/dL was 0.01% ± 0.10%, 0.03% ± 0.08%, and 0.04% ± 0.14% in the first 2 weeks, the third month, and the sixth month periods, respectively (Table 3).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataIn the per protocol cohort, 69% had diabetes for 10 years or more.
“Sixty-two percent (62%) selected White non-Hispanic as their race/ethnic background (see Table 1 for categories), 59% were female, and 69% had diabetes for 10 years or more.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataIn the per protocol cohort, 62% were White non-Hispanic.
“Of 102 participants in the per protocol cohort, median age was 59 years, 87% had T2D, 42% used CGM previously, 62% were White non-Hispanic, and 59% female.”
- Also stated in: § Participant Characteristics · plain
“Sixty-two percent (62%) selected White non-Hispanic as their race/ethnic background (see Table 1 for categories), 59% were female, and 69% had diabetes for 10 years or more.”
Quote in the paperNot a substantial claim- Also stated in: § Participant Characteristics · plain
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataIn the per protocol cohort, 59% were female.
“Of 102 participants in the per protocol cohort, median age was 59 years, 87% had T2D, 42% used CGM previously, 62% were White non-Hispanic, and 59% female.”
- Also stated in: § Participant Characteristics · plain
“Sixty-two percent (62%) selected White non-Hispanic as their race/ethnic background (see Table 1 for categories), 59% were female, and 69% had diabetes for 10 years or more.”
Quote in the paperNot a substantial claim- Also stated in: § Participant Characteristics · plain
May combine two claims.
Findingown dataIn the per protocol cohort, there was no statistically significant interaction in HbA1c improvement from baseline in other subgroups tested, including those with and without prior CGM experience, with the exception of those living in rural vs urban areas, who had higher baseline HbA1c.
“There was no statistically significant interaction in HbA1c improvement from baseline in other subgroups tested, including those with and without prior CGM experience, with the exception of those living in rural (vs urban) areas, who had higher baseline HbA1c (Table 2).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataIn the per protocol cohort, 87% had type 2 diabetes.
“Of 102 participants in the per protocol cohort, median age was 59 years, 87% had T2D, 42% used CGM previously, 62% were White non-Hispanic, and 59% female.”
- Also stated in: § Participant Characteristics · plain
“Within the per protocol cohort (N = 102), median age was 59 years (range, 22-82) and 89 of 102 (87%) had T2D.”
Quote in the paperNot a substantial claim- Also stated in: § Participant Characteristics · plain
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataIn the per protocol cohort, median age was 59 years (range 22-82).
“Of 102 participants in the per protocol cohort, median age was 59 years, 87% had T2D, 42% used CGM previously, 62% were White non-Hispanic, and 59% female.”
- Also stated in: § Participant Characteristics · plain
“Within the per protocol cohort (N = 102), median age was 59 years (range, 22-82) and 89 of 102 (87%) had T2D.”
Quote in the paperNot a substantial claim- Also stated in: § Participant Characteristics · plain
May combine two claims.
Findingown dataIn the per protocol cohort, six severe hypoglycemia events occurred in 4 participants, and none were System-related.
“Six severe hypoglycemia events occurred in 4 participants (none System-related) and no diabetic ketoacidosis events occurred in the per protocol cohort.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe expanded cohort reflects 171 of 207 (83%) who were eligible for outcomes analysis.
“The expanded cohort reflects 171 of 207 (83%) who were eligible for outcomes analysis and the per protocol cohort represents 102 of 153 (67%) eligible for the 6-month per protocol analysis (Fig. 1).”
Quote in the paperNot a substantial claimThe quote only partly states this claim.
Findingown dataIn the per protocol cohort, median rapid-acting insulin doses per day were 2.9, 2.6, and 2.4 in the first 2 weeks, third month, and sixth month, respectively.
“The median number of rapid-acting insulin doses per day was 2.9 (IQR 2.2, 3.4), 2.6 (IQR 1.9, 3.1) and 2.4 (IQR 1.9, 3.0) in the first 2 weeks, third month, and sixth month of use, respectively (Table 3).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataOver 6 months, total System use represented 89.9 person-years in the per protocol cohort and 135.6 person-years in the expanded cohort.
“Within the per protocol and the expanded cohorts, over the course of 6 months, total System use represented 89.9 and 135.6 person-years, respectively.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S1 (asserted), S3 (load-bearing), S4 (based). The quote only partly states this claim.
Findingown dataIn the per protocol cohort, median long-acting insulin doses per week were 7.0, 7.0, and 6.9 in the first 2 weeks, third month, and sixth month, respectively.
“The average number of long-acting doses per week was 7.0 (IQR 6.5, 7.5), 7.0 (IQR 6.5, 7.3) and 6.9 (IQR 6.3, 7.0) in the first 2 weeks, third month, and sixth month of use, respectively.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn the per protocol cohort, mean T1-DDS score was 2.4 ± 0.9 at baseline and 1.7 ± 0.6 at 6 months, with a mean difference of −0.7 (95% CI −0.9 to −0.5, P < .001).
“Mean T1-DDS score at baseline was 2.4 ± 0.9 and fell to 1.7 ± 0.6 at 6 months with a mean difference of −0.7 (95% CI −0.9 to −0.5, P < .001).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe per protocol cohort represents 102 of 153 (67%) eligible for the 6-month per protocol analysis.
“The expanded cohort reflects 171 of 207 (83%) who were eligible for outcomes analysis and the per protocol cohort represents 102 of 153 (67%) eligible for the 6-month per protocol analysis (Fig. 1).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataOnly a small number of participants with low distress or moderate-to-high hypoglycemic confidence at baseline worsened after 6 months (5% and 18%, respectively).
“It should be noted that there were only a small number of participants who at baseline had relatively low distress (mean T1-DDS < 2.0) or moderate to high hypoglycemic confidence (mean HCS > 3.0) and worsened (ie, crossing their respective cutpoints) after 6 months (5% and 18% of participants, respectively).”
Quote in the paperNot a substantial claimThe quote only partly states this claim.
Findingown dataIn the per protocol cohort, mean CGM-measured hypoglycemia events per week (<70 mg/dL) were 1.0, 1.1, and 1.3 in the first 2 weeks, third month, and sixth month, respectively.
“The average number of CGM-measured hypoglycemia events per week (< 70 mg/dL or < 3.9 mmol/L) was 1.0 ± 1.9, 1.1 ± 1.6, and 1.3 ± 1.6 in the first 2 weeks, third month, and sixth month of use, respectively (Table 3).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn the HbA1c < 8% subgroup at 6 months, there was little change in glycemic control, some improvement in T1-DDS (decreased ∼0.5 points), similar improvement of HCS (increased 0.4 points), and a similar TBR (∼1.0%) compared to the overall cohort.
“However, we evaluated the HbA1c < 8% subgroup at 6 months and although there was little change in glycemic control, some improvement in T1-DDS (decreased ∼0.5 points), similar improvement of HCS (increased 0.4 points), and a similar TBR (∼1.0%) compared to the overall cohort was observed, suggesting psychosocial benefits with limited hypoglycemia.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn the safety cohort of N = 232, 15 severe hypoglycemia events occurred in 11 participants; all but 1 were in participants with T2D.
“In the safety cohort of N = 232, 15 severe hypoglycemia events occurred in 11 participants; all but 1 were in participants with T2D.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataHere, 42% of participants were current CGM users at enrollment and had similar HbA1c improvements (1.1% vs 1.2%) when switching to Bigfoot Unity compared to non-CGM users at enrollment.
“Here, 42% of participants were current CGM users at enrollment and had similar HbA1c improvements (1.1% vs 1.2%) when switching to Bigfoot Unity compared to non-CGM users at enrollment.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn the safety cohort, 2 other serious hyperglycemic events needing treatment in a healthcare facility occurred in 2 participants, 1 with T1D and 1 with T2D; neither was using the System at the time.
“Additionally, in the safety cohort, 1 DKA event occurred in a person with T2D and 2 other serious hyperglycemic events needing treatment in a healthcare facility occurred in 2 participants, 1 with T1D and 1 with T2D; neither was using the System at the time these events occurred.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingnone givenIn this prospective study of Bigfoot Unity users, most participants had T2D and about 40% had used CGM previously.
“In this prospective study of Bigfoot Unity users, most participants had T2D and about 40% had used CGM previously.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataObservations of improved glycemic control were coupled with statistically significant improvements in diabetes distress (mean change −0.7 points) and improvements in hypoglycemia confidence (mean change 0.4 points).
“Observations of improved glycemic control were coupled with statistically significant improvements in diabetes distress (mean change −0.7 points) and improvements in hypoglycemia confidence (mean change 0.4 points).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe T1D subgroup was small, but there appears to be a difference in HbA1c improvement (0.7% vs 1.2%), TIR (39% vs 58%), TBR (1.6% vs 0.9%), and time above range (59% vs 40%) between T1D and T2D subgroups, respectively.
“The T1D subgroup was small (see Figs. 1 and 3), but there appears to be a difference in HbA1c improvement (0.7% vs 1.2%), TIR (39% vs 58%), TBR (1.6% vs 0.9%), and time above range (59% vs 40%) between T1D and T2D subgroups, respectively.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataWith regard to diabetes distress, those with T1D had higher baseline distress than those with T2D (2.7 vs 2.3) with both groups having similar improvement (∼0.7 points).
“With regard to diabetes distress, those with T1D had higher baseline distress than those with T2D (2.7 vs 2.3) with both groups having similar improvement (∼0.7 points); those with T1D had similar baseline hypoglycemia confidence as those with T2D (2.9 vs 2.8) but those with T1D did not show improvement at 6 months (−0.01 points) while those with T2D improved ∼0.5 points.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThose with T1D had similar baseline hypoglycemia confidence as those with T2D (2.9 vs 2.8) but those with T1D did not show improvement at 6 months (−0.01 points) while those with T2D improved ∼0.5 points.
“With regard to diabetes distress, those with T1D had higher baseline distress than those with T2D (2.7 vs 2.3) with both groups having similar improvement (∼0.7 points); those with T1D had similar baseline hypoglycemia confidence as those with T2D (2.9 vs 2.8) but those with T1D did not show improvement at 6 months (−0.01 points) while those with T2D improved ∼0.5 points.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataThe frequency of hypoglycemia observed at 6 months was substantially below established targets of <4% for time below 70 mg/dL and <1% for time below 54 mg/dL.
“In this study primarily of older adults with T2D using MDI, durable glycemic improvement occurred using the System at 6 months, with the frequency of hypoglycemia being substantially below established targets of < 4% and < 1% for time below 70 and 54 mg/dL, respectively.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Study design and conductnone givenThe false discovery rate was controlled by using the adaptive Benjamini-Hochberg procedure.
“The false discovery rate was controlled by using the adaptive Benjamini-Hochberg procedure.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Study design and conductnone givenPrespecified outcomes for analysis were GMI, percent time in range, percent time below range, percent time above range, sensor data availability, mean sensor glucose and coefficient of variation, mean sensor scans, and dose events.
“Prespecified outcomes for analysis were GMI (an estimate of HbA1c from CGM-measured mean glucose) calculated as described (17), percent time in range (TIR; 70-180 mg/dL, 3.9-10.0 mmol/L), percent time below range (< 70 mg/dL and < 54 mg/dL or < 3.9 mmol/L and < 3.0 mmol/L, respectively), percent time above range (> 180 mg/dL and > 250 mg/dL or > 10.0 mmol/L and > 13.9 mmol/L, respectively), sensor”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningThe sample size was not based on statistical power.
“As this was an observational analysis, the sample size was not based on statistical power.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenFor reported events during the study, JCHR staff would attempt to contact participants to confirm reported events indeed met the study definitions.
“For reported events during the study, JCHR staff would attempt to contact participants to confirm reported events indeed met these definitions (it was not possible to confirm baseline information in a similar way).”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductreasoningThe T1-DDS, though initially developed for T1D, was deemed the better choice for assessing diabetes distress in the current study because the sample included a mix of adults with T1D and T2D, and all were insulin users.
“Of note, the T1-DDS, though it was initially developed for T1D, was deemed the better choice for assessing diabetes distress in the current study, since the sample included a mix of adults with T1D and T2D, and all were insulin users.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationRecent guidelines call for continuous glucose monitoring (CGM) use for all who require insulin, regardless of diabetes type.
“While recent guidelines call for continuous glucose monitoring (CGM) use for all who require insulin, regardless of diabetes type (7), there is also evidence that patients may not know how to interpret what to do with the CGM data (6, 8, 9).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [7] Samson, 2023, American Association of Clinical Endocrinology consensus statement: comprehensive type 2 diabetes management algorithm—2023 update, Endocr Pract source
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe analysis was prespecified to be completed when approximately 100 participants had sufficient data to be included in the per protocol analysis.
“The analysis was prespecified to be completed when approximately 100 participants had sufficient data to be included in the per protocol analysis.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe prespecified protocol, consent forms, data collection forms, and recruitment materials were approved by the JCHR institutional review board.
“The prespecified protocol, consent forms, data collection forms, and recruitment materials were approved by the JCHR institutional review board.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone givenCGM metrics were calculated over a 28-day period with the most data and required at least 50% CGM data available (at least 336 hours) in the 6-month period to be included in the per protocol analyses.
“CGM metrics were calculated over a 28-day period with the most data and required at least 50% CGM data available (at least 336 hours) in the 6-month period to be included in the per protocol analyses.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenPrespecified subgroup analyses included baseline HbA1c ≥8% vs <8%, age ≥65 vs <65 years, ethnicity, economic status, education, living areas, and previous CGM experience.
“Prespecified subgroup analyses included those with baseline HbA1c ≥ 8% vs < 8% (≥ 64 mmol/mol vs < 64 mmol/mol), those ≥ 65 years vs < 65 years, ethnicity, economic status, education, living areas, and those with and without previous CGM experience.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataGiven baseline data indicating that the vast majority took 2 (37%) or 3 (43%) meals per day, findings of a median of 2.6 and 2.4 rapid-acting doses per day in the third and sixth month (slightly down from 2.9 in the first 2 weeks of use) suggest minimal missed meal injections.
“Given baseline data indicating that the vast majority took 2 (37%) or 3 (43%) meals per day, our findings of a median of 2.6 and 2.4 rapid-acting doses per day in the third and sixth month (slightly down from 2.9 in the first 2 weeks of use) suggest minimal missed meal injections.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataPercent time < 54 mg/dL was on average 0.04% at 6 months and while this was statistically higher compared to the first 2 weeks of use (0.01%), it is likely not clinically meaningful.
“It can be noted that percent time < 54 mg/dL was on average 0.04% at 6 months and while this was statistically higher compared to the first 2 weeks of use (0.01%), it is likely not clinically meaningful.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataBased on the GMI estimate of HbA1c within the first 2 weeks of being on the System, and comparison with baseline HbA1c, there was rapid glycemic improvement which remained stable through month 6.
“Based on the GMI estimate of HbA1c within the first 2 weeks of being on the System, and comparison with the baseline HbA1c, there was rapid glycemic improvement which remained stable through month 6.”
Quote in the paperSubstantial claimMay combine two claims.
Premiseown dataA PubMed search using the terms connected insulin pen or smart insulin pen on February 27, 2024, returned 55 articles, of which only 11 had relevant effectiveness or safety-related data for connected insulin pen systems.
“A PubMed search using the terms connected insulin pen or smart insulin pen on February 27, 2024, returned 55 articles, of which only 11 had relevant effectiveness or safety-related data for connected insulin pen systems (most were reviews of or perspectives on smart pen technology); surprisingly, none addressed the prospective collection of clinically relevant adverse events such as severe hypogly”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationFor context for the SUS score, scores > 80 are considered excellent system use.
“For context for the SUS score, scores > 80 are considered excellent system use (48) and thus, the 6-month mean of 80.7 indicates that the System supports an above average user experience.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [48] Lewis, 2018, Item benchmarks for the system usability scale, J User Exp source
May combine two claims.
Study design and conductreasoningBecause the recall process for reporting severe hypoglycemia and DKA events differed between baseline and the study period and only the latter had a confirmation process by the medical monitor, it is not appropriate to compare self-reported events at baseline to those occurring during the study.
“It should be noted that because the recall process for reporting severe hypoglycemia and DKA events differed between baseline and the study period and, because only the latter had a confirmation process by the medical monitor that was not possible at baseline, we do not think it is appropriate to compare self-reported events at baseline to those occurring during the study.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThe authors found similar glycemic outcomes regardless of race, age, diabetes duration, presence of baseline diabetes complications, income, education, and prior sensor experience, which speaks to the generalizability of System use.
“We found similar glycemic outcomes regardless of race, age, diabetes duration, presence of baseline diabetes complications, income, education, and prior sensor experience, which speaks to the generalizability of System use.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe study design did not include baseline CGM metrics (only baseline HbA1c and PRO data were available).
“The study design also did not include baseline CGM metrics (only baseline HbA1c and PRO data were available).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenAn additional limitation was that the number of participants with T1D was small, which limits interpretation.
“Finally, an additional limitation was that the number of participants with T1D was small, which limits interpretation.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductnone givenStrengths of the study include the prospective collection of HbA1c, PRO, CGM and clinically relevant adverse events in a real-world and diverse population using this relatively new connected pen cap system.
“With these limits in mind, the strengths of the study include the prospective collection of HbA1c, PRO, CGM and clinically relevant adverse events in a real-world and diverse population using this relatively new connected pen cap system.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductown dataTo mitigate potential dropout bias in the per protocol analysis and maximize analyzable data without requiring stringent System use, the authors conducted an expanded cohort analysis which demonstrated similar demographics, glycemic and PRO improvement, System use, and safety outcomes.
“Nonetheless, to mitigate potential dropout bias in the per protocol analysis and maximize analyzable data without requiring stringent System use, we conducted our expanded cohort analysis which demonstrated similar demographics, glycemic and PRO improvement, System use, and safety outcomes.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductnone givenLimitations of this study include the single-arm study design with no concurrent control group and relatively small sample size in the per protocol cohort.
“Limitations of this study include the single-arm study design with no concurrent control group and relatively small sample size in the per protocol cohort.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductnone givenThe study data collected were primarily direct from participant (ie, self-reported) which likely limited collection of complete data.
“Additionally, the study data collected were primarily direct from participant (ie, self-reported) which likely limited collection of complete data.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
The observed glycemic improvement is attributable to use of the Bigfoot Unity System rather than to other factors.
“In this study primarily of older adults with T2D using MDI, durable glycemic improvement occurred using the System at 6 months, with the frequency of hypoglycemia being substantially below established targets of < 4% and < 1% for time below 70 and 54 mg/dL, respectively.”
The paper relies on this without support.
At-home capillary blood HbA1c values and electronic medical record HbA1c values are comparable for the analysis.
“If a participant did not collect the blood sample within the acceptable window (−90 to 0 days for baseline) or the blood sample was received at the central lab beyond the stability period, an electronic medical record HbA1c value was used.”
The paper relies on this without support.
The observed improvements in glycemic control and psychosocial outcomes are attributable to use of the Bigfoot Unity System rather than other factors.
“For this cohort, consisting primarily of older adults with T2D and having suboptimal glycemic control, using the System resulted in rapid and durable improvement in glycemic control.”
“Limitations of this study include the single-arm study design with no concurrent control group and relatively small sample size in the per protocol cohort.”
The paper relies on this without support.
The Bigfoot Unity System is intended for once-daily long-acting insulin dosing.
“Given that the System is intended for once-daily long-acting insulin dosing, the median of 7.0 and 6.9 long-acting doses per week in month 3 and 6 is encouraging.”
The paper relies on this without support.
The MCID for the T1-DDS total score derived in an insulin-requiring T1D population applies to the current mixed T1D/T2D cohort.
“To place context around these results, note that the estimated minimally clinically important difference (MCID) for the 28-item T1-DDS total score is 0.19 in an insulin-requiring T1D population (19); comparing the average improvement in the current study with the MCID suggests that for the majority of participants, there may have been clinically meaningful improvement.”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Per-protocol hospitalization breakdown exceeds its total
MinorTable 5 reports 13 hospitalizations for reasons other than severe hypoglycemia and DKA in the per-protocol cohort, but its using-system/not-using-system breakdown totals 22.
Reported 13 · recomputed 22
“| Hospitalizations (reasons other than SH and DKA), totalb | Hospitalizations (reasons other than SH and DKA), totalb | Hospitalizations (reasons other than SH and DKA), totalb | 13 | 24 | 28 | | | Using system/not using system | Using system/not using system | 21/1 | 21/3 | 24/4 |”
Median sensor availability is described as a mean
MinorThe discussion calls the sixth-month sensor-use value of 84% a mean, whereas Table 3 identifies 84% as the median.
“Other parameters such as the mean percent time of sensor use and mean coefficient of variation (84% and 31% at 6 months, respectively) are also within established guidelines (23, 24, 38).”
“| Percent sensor data availability, median (quartiles)b | Percent sensor data availability, median (quartiles)b | 87 (74, 93) | 86 (77, 92) | 84 (72, 91) |”
Cites
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,032 · see citing works
Self-citationInternational consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069 · see citing works
Self-citationAssessing Psychosocial Distress in Diabetes
Polonsky, W. H., Fisher, L., Earles, J., Dudl, R. J., Lees, J., Mullan, J., & Jackson, R. A. (2005) · Diabetes Care · doi:10.2337/diacare.28.3.626 · Cited by 1,575 · see citing works
Self-citation6. Glycemic Targets: <i>Standards of Medical Care in Diabetes—2022</i>
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s006 · Cited by 1,403 · see citing works
Beck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185 · see citing works
Self-citationContinuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831 · see citing works
Self-citationRisk of hypoglycaemia in types 1 and 2 diabetes: Effects of treatment modalities and their duration
UK Hypoglycaemia Study Group (2007) · Diabetologia · doi:10.1007/s00125-007-0599-y · Cited by 825 · see citing works
Economic Costs of Diabetes in the U.S. in 2022
(2023) · doi:10.2337/dci23-0085 · Cited by 808 · see citing works
Glucose management indicator (GMI): a new term for estimating A1C from continuous glucose monitoring
Bergenstal, R. M., Beck, R. W., Close, K. L., Grunberger, G., Sacks, D. B., Kowalski, A., Brown, A. S., Heinemann, L., Aleppo, G., Ryan, D. B., et al. (2018) · Diabetes Care · doi:10.2337/dc18-1581 · Cited by 684 · see citing works
Self-citationPeyrot, M., Barnett, A. H., Meneghini, L. F., & Schumm‐Draeger, P. ‐. (2012) · Diabetic Medicine · doi:10.1111/j.1464-5491.2012.03605.x · Cited by 655 · see citing works
Beck, R. W., Riddlesworth, T. D., Ruedy, K., Ahmann, A., Haller, S., Kruger, D., McGill, J. B., Polonsky, W., Price, D., Aronoff, S., et al. (2017) · Annals of Internal Medicine · doi:10.7326/m16-2855 · Cited by 635 · see citing works
Self-citationDonnelly, L. A., Morris, A. D., Frier, B. M., Ellis, J. D., Donnan, P. T., Durrant, R., Band, M. M., Reekie, G., & Leese, G. P. (2005) · Diabetic Medicine · doi:10.1111/j.1464-5491.2005.01501.x · Cited by 511 · see citing works
The T1D exchange clinic registry
Beck, R. W., Tamborlane, W. V., Bergenstal, R. M., Miller, K. M., DuBose, S. N., Hall, C. A., & for the T1D Exchange Clinic Network (2012) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2012-1561 · Cited by 456 · see citing works
Self-citationKarges, B., Schwandt, A., Heidtmann, B., Kordonouri, O., Binder, E., Schierloh, U., Boettcher, C., Kapellen, T., Rosenbauer, J., & Holl, R. W. (2017) · JAMA · doi:10.1001/jama.2017.13994 · Cited by 452 · see citing works
13. Older Adults: <i>Standards of Medical Care in Diabetes—2022</i>
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s013 · Cited by 381 · see citing works
Samson, S. L., Vellanki, P., Blonde, L., Christofides, E. A., Galindo, R. J., Hirsch, I. B., Isaacs, S. D., Izuora, K. E., Low Wang, C. C., Twining, C. L., et al. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2023.02.001 · Cited by 375 · see citing works
Item benchmarks for the system usability scale
(2018) · Journal of Usability Studies archive · doi:10.5555/3294033.3294037 · Cited by 355 · see citing works
Khunti, K., Alsifri, S., Aronson, R., Cigrovski Berković, M., Enters‐Weijnen, C., Forsén, T., Galstyan, G., Geelhoed‐Duijvestijn, P., Goldfracht, M., Gydesen, H., et al. (2016) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12689 · Cited by 276 · see citing works
Kho, M. E., Duffett, M., Willison, D. J., Cook, D. J., & Brouwers, M. C. (2009) · BMJ · doi:10.1136/bmj.b866 · Cited by 269 · see citing works
Diabetes distress in adults with type 1 diabetes: prevalence, incidence and change over time
Fisher, L., Hessler, D., Polonsky, W., Strycker, L., Masharani, U., & Peters, A. (2016) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2016.03.032 · Cited by 208 · see citing works
Self-citationHaak, T., Hanaire, H., Ajjan, R., Hermanns, N., Riveline, J.-P., & Rayman, G. (2017) · Diabetes Therapy · doi:10.1007/s13300-017-0255-6 · Cited by 188 · see citing works
Optimal use of diabetes devices: clinician perspectives on barriers and adherence to device use
Tanenbaum, M. L., Adams, R. N., Hanes, S. J., Barley, R. C., Miller, K. M., Mulvaney, S. A., & Hood, K. K. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816688010 · Cited by 117 · see citing works
Investigating Hypoglycemic Confidence in Type 1 and Type 2 Diabetes
Polonsky, W. H., Fisher, L., Hessler, D., & Edelman, S. V. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0366 · Cited by 114 · see citing works
Self-citationInsulin adherence in patients with diabetes: risk factors for injection omission
Farsaei, S., Radfar, M., Heydari, Z., Abbasi, F., & Qorbani, M. (2014) · Primary care diabetes · doi:10.1016/j.pcd.2014.03.001 · Cited by 106 · see citing works
Ever-Increasing Insulin-Requiring Patients Globally
Garg, S. K., Rewers, A. H., & Aktürk, H. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0101 · Cited by 101 · see citing works
Predictors of attrition in a longitudinal population-based study of aging
Jacobsen, E., Ran, X., Liu, A., Chang, C.-C. H., & Ganguli, M. (2021) · International Psychogeriatrics · doi:10.1017/s1041610220000447 · Cited by 85 · see citing works
Smart pens will improve insulin therapy
Klonoff, D. C., & Kerr, D. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818759845 · Cited by 73 · see citing works
JENDLE, J. H., Ericsson, Å., Gundgaard, J., Møller, J. B., Valentine, W. J., & Hunt, B. (2020) · Diabetes Therapy · doi:10.1007/s13300-020-00980-1 · Cited by 68 · see citing works
Patient recall of self-care recommendations in diabetes
Page, P., Verstraete, D. G., Robb, J. R., & Etzwiler, D. D. (1981) · Diabetes Care · doi:10.2337/diacare.4.1.96 · Cited by 56 · see citing works
Can Smart Pens Help Improve Diabetes Management?
Sy, S. L., Munshi, M. N., & Toschi, E. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820965600 · Cited by 44 · see citing works
Oser, T. K., Litchman, M. L., Allen, N. A., Kwan, B. M., Fisher, L., Jortberg, B. T., Polonsky, W. H., & Oser, S. M. (2021) · Current Diabetes Reports · doi:10.1007/s11892-021-01408-1 · Cited by 43 · see citing works
Self-citationAn evaluation of two capillary sample collection kits for laboratory measurement of HbA1c
Beck, R. W., Bocchino, L. E., Lum, J. W., Kollman, C., Barnes-Lomen, V., Sulik, M., Haller, M. J., Bode, B., Cernich, J. T., Killeen, A. A., et al. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0023 · Cited by 42 · see citing works
Self-citationBarriers to continuous glucose monitoring in people with type 1 diabetes: clinician perspectives
Lanning, M. S., Tanenbaum, M. L., Wong, J. J., & Hood, K. K. (2020) · Diabetes Spectrum · doi:10.2337/ds19-0039 · Cited by 40 · see citing works
Galindo, R. J., Ramos, C., Cardona, S., Vellanki, P., Davis, G. M., Oladejo, O., Albury, B., Dhruv, N., Peng, L., & Umpierrez, G. E. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211033837 · Cited by 40 · see citing works
MacLeod, J., Im, G. H., Smith, M., & Vigersky, R. A. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0365 · Cited by 40 · see citing works
Billings, L. K., Parkin, C. G., & Price, D. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0163 · Cited by 39 · see citing works
Gomez-Peralta, F., Abreu, C., Fernández-Rubio, E., Cotovad, L., Pujante, P., Gaztambide, S., Bellido, D., Menéndez Torre, E., Ruiz-Valdepeñas, S., Bello, H., et al. (2023) · Diabetes Care · doi:10.2337/dc22-0525 · Cited by 36 · see citing works
Steenkamp, D., Eby, E. L., Gulati, N., & Liao, B. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296821997923 · Cited by 33 · see citing works
Szmuilowicz, E. D., & Aleppo, G. (2022) · Postgraduate Medicine · doi:10.1080/00325481.2022.2110507 · Cited by 27 · see citing works
Chien, A., THANASEKARAN, S., Gaetano, A., IM, G. H., Wherry, K., MacLeod, J., & Vigersky, R. A. (2023) · Journal of Managed Care & Specialty Pharmacy · doi:10.18553/jmcp.2023.22283 · Cited by 18 · see citing works
Hellman, J., Hartvig, N. V., Kaas, A., Møller, J. B., Sørensen, M. R., & JENDLE, J. H. (2023) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15316 · Cited by 18 · see citing works
First real-world experience with bigfoot unity: a 6-month retrospective analysis
Baliga, B. S., Tillman, J. B., Olson, B., Vaughan, S., Sheikh, F. N., & Malone, J. K. (2023) · Clinical Diabetes · doi:10.2337/cd22-0126 · Cited by 15 · see citing works
Li, Z., Calhoun, P. M., Ruedy, K. J., & Beck, R. W. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0094 · Cited by 6 · see citing works
Self-citationChan, K., Hansen, K., Muratov, S., Khoudigian-Sinani, S., & Lamotte, M. (2024) · Journal of Comparative Effectiveness Research · doi:10.57264/cer-2023-0124 · Cited by 5 · see citing works
Continuous glucose monitoring metrics: interpretation and integrations
(2022) · The American Journal of Managed Care · doi:10.37765/ajmc.2022.89214 · Cited by 4 · see citing works
Learning From COVID-19 and Looking to Technology Ahead
Toft, A. D. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820930280 · Cited by 3 · see citing works
936-P: three-month interim analysis from the prospective bigfoot unity real world study (BURST)
MALONE, J. K., BECK, R., TILLMAN, J., CALHOUN, P., MOUSE, T. J., BAILEY, R., OLSON, B. P., VAUGHAN, S., & NAUREEN SHEIKH, F. (2023) · Diabetes · doi:10.2337/db23-936-p · Cited by 2 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Still being checked.
- Authors
- Tillman, J. B., Beck, R. W., Polonsky, W. H., Calhoun, P., Mouse, T., Bailey, R., Jacoby, J., Nandan, R., Sheikh, F. N., & Malone, J. K.
- Affiliations
- Abbott Diabetes Care (United States)
- Jaeb Center for Health Research (US)
- Behavioral Diabetes Institute , San Diego, CA 92121 ,
- Department of Fluorescence Imaging, Intuitive , Sunnyvale, CA 94086 ,
- Corresponding author
- Tillman, John B
- Venue
- The Journal of Clinical Endocrinology & Metabolism
- Publisher
- The Endocrine Society
- Published
- 2024-11-15
- Volume
- 110
- Issue
- 8
- Pages
- 2134-2146
- ISSN
- 0021-972X (ISSN-L)
- 1945-7197
- PMC
- PMC12261103
- OpenAlex
- W4404400746
- Semantic Scholar
- 8fa27c74d834540c067ed02b06091edd786c8f4b
- Cited by
- 1
- References
- 47
- Citations by year
- 2025: 1
- FWCI
- 0.29
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- The Journal of Clinical Endocrinology & Metabolism (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion, cc-by-nc-nd)
- Funders
- Abbott
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- continuous glucose monitoring, mobile application, multiple daily insulin injections, type 2 diabetes, type 1 diabetes, glycemic control, HbA1c, time in range, hypoglycemia, severe hypoglycemia, diabetic ketoacidosis, real-world evidence, older adults
- MeSH terms
- Glycemic Control, Adult, Aged, Blood Glucose, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Female, Glycated Hemoglobin, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Male, Middle Aged, Prospective Studies, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review