A simplified biventricular defibrillator with fixed long detection intervals reduces implantable cardioverter defibrillator (ICD) interventions and heart failure hospitalizations in patients with non-ischaemic cardiomyopathy implanted for primary prevention: the RELEVANT [Role of long dEtection window programming in patients with LEft VentriculAr dysfunction, Non-ischemic eTiology in primary prevention treated with a biventricular ICD] study
Open accessThe DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index.
Abstract
AIMS: To investigate the efficacy and safety of a cardiac resynchronization therapy with cardioverter-defibrillator (CRT-D) device with simplified ventricular tachycardia management in patients with non-ischaemic heart failure (HF) and primary prevention implantable cardioverter defibrillator (ICD) indication. METHODS AND RESULTS: Prospective, controlled, parallel, multicentre, non-randomized study enrolling 324 primary prevention non-ischaemic HF patients implanted with CRT-D devices from 2004 to 2007: Protect group, 164 patients implanted with a Medtronic Insync III Protect device and Control group, 160 patients utilizing other Medtronic CRT-D devices. Efficacy was assessed by computing appropriate and inappropriate detections and therapies during follow-up; safety compared hospitalizations and syncopal events between groups. Ninety per cent of both ventricular and supraventricular tachyarrhythmias terminated within the 13-29 beat detection interval with the Protect algorithm. The Protect group showed a significantly better event-free survival to first delivered therapy for total (P = 0.0001), appropriately treated (P = 0.002), and inappropriately treated episodes (P = 0.017). The total number of delivered shocks was significantly lower in the Protect group (22 vs. 59, P < 0.0001). In the Protect group, a significantly reduced HF hospitalization (hazard ratio 0.38, 95% CI 0.15-0.98, P = 0.044) was observed without any increase of syncope or death. CONCLUSION: A simplified CRT-D device with fixed long detection reduced overall ICD therapy burden and HF hospitalizations without entailing any additional adverse events in primary prevention non-ischaemic HF patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Moss, A. J., Hall, W. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Klein, H., Levine, J. H., Saksena, S., Waldo, A. L., Wilber, D., et al. (1996) · New England Journal of Medicine · doi:10.1056/nejm199612263352601 · Cited by 4,165
(1997) · New England Journal of Medicine · doi:10.1056/nejm199711273372202 · Cited by 3,232
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities
Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.02.032 · Cited by 2,546
Writing Committee Members, Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., et al. (2008) · Circulation · doi:10.1161/circualtionaha.108.189742 · Cited by 1,774
Young, J. B., Abraham, W. T., & Smith, A. L. (2003) · ACC Current Journal Review · doi:10.1016/j.accreview.2003.08.087 · Cited by 1,682
Prognostic Importance of Defibrillator Shocks in Patients with Heart Failure
Poole, J. E., Johnson, G. W., Hellkamp, A. S., Anderson, J., Callans, D. J., Raitt, M. H., Reddy, R. K., Marchlinski, F. E., Yee, R., Guarnieri, T., et al. (2008) · New England Journal of Medicine · doi:10.1056/nejmoa071098 · Cited by 1,462
Authors/Task Force Members, Vardas, P. E., Auricchio, A., Blanc, J.-J., Daubert, J.-C., Drexler, H., Ector, H., Gasparini, M., Linde, C., Morgado, F. B., et al. (2007) · European Heart Journal · doi:10.1093/eurheartj/ehm305 · Cited by 902
Self-citationWILKOFF, B. L., HESS, M., YOUNG, J., & ABRAHAM, W. T. (2004) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2004.03625.x · Cited by 835
Inappropriate Implantable Cardioverter-Defibrillator Shocks in MADIT II
Daubert, J. P., Zareba, W., Cannom, D. S., McNitt, S., Rosero, S. Z., Wang, P., Schuger, C., Steinberg, J. S., Higgins, S. L., Wilber, D. J., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.09.073 · Cited by 813
Wathen, M. S., DeGroot, P. J., Sweeney, M. O., Stark, A. J., Otterness, M. F., Adkisson, W. O., Canby, R. C., Khalighi, K., Machado, C., Rubenstein, D. S., et al. (2004) · Circulation · doi:10.1161/01.cir.0000145610.64014.e4 · Cited by 775
[Guidelines for cardiac pacing and cardiac resynchronization therapy].
Vardas, P. E., Auricchio, A., Blanc, J., Daubert, J., Drexler, H., Ector, H., Gasparini, M., Linde, C. M., Morgado, F. B., Oto, A., et al. (2008) · PubMed · Cited by 550
Self-citationWilkoff, B. L., Williamson, B. D., Stern, R. S., Moore, S. L., Lu, F., Lee, S. W., Birgersdotter-Green, U. M., Wathen, M. S., Van Gelder, I. C., Heubner, B. M., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.05.011 · Cited by 548
Cost-Effectiveness of Implantable Cardioverter–Defibrillators
Sanders, G. D., Hlatky, M. A., & Owens, D. K. (2005) · New England Journal of Medicine · doi:10.1056/nejmsa051989 · Cited by 547
Wathen, M. S., Sweeney, M. O., DeGroot, P. J., Stark, A. J., Koehler, J. L., Chisner, M. B., Machado, C., Adkisson, W. O., & for the PainFREE ℞ Investigators (2001) · Circulation · doi:10.1161/hc3101.093906 · Cited by 333
A Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators
Wilkoff, B. L., Ousdigian, K. T., Sterns, L. D., Wang, Z. J., Wilson, R. D., & Morgan, J. M. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.03.037 · Cited by 242
Heart Failure Care in the Outpatient Cardiology Practice Setting: Findings From IMPROVE HF
Fonarow, G. C., Yancy, C. W., Albert, N. M., Curtis, A. B., Stough, W. G., Gheorghiade, M., Heywood, J. T., McBride, M. L., Mehra, M. R., O'Connor, C. M., et al. (2008) · Circulation: Heart Failure · doi:10.1161/circheartfailure.108.772228 · Cited by 187
Owens, D. K., Sanders, G. D., Harris, R. A., McDonald, K. M., Heidenreich, P. A., Dembitzer, A. D., & Hlatky, M. A. (1997) · Annals of Internal Medicine · doi:10.7326/0003-4819-126-1-199701010-00001 · Cited by 186
KLEIN, R. C., RAITT, M. H., WILKOFF, B. L., BECKMAN, K. J., COROMILAS, J., WYSE, D. G., FRIEDMAN, P. L., MARTINS, J. B., EPSTEIN, A. E., HALLSTROM, A. P., et al. (2003) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2003.01554.x · Cited by 179
Effect of Ventricular Shock Strength on Cardiac Hemodynamics
Tokano, T., Bach, D. S., CHANG, J., DAVIS, J., Souza, J. J., Zivin, A. H., Knight, B. P., Goyal, R., Man, K. C., Morady, F., et al. (1998) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.1998.tb00118.x · Cited by 142
Classification of death in antiarrhythmia trials
Epstein, A. E., Carlson, M. D., Fogoros, R. N., Higgins, S. L., & Venditti, F. J. (1996) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(95)00482-3 · Cited by 121
Use of Cardiac Resynchronization Therapy in Patients Hospitalized With Heart Failure
Piccini, J. P., Hernandez, A. F., Dai, D. W. H., Thomas, K. L., Lewis, W. R., Yancy, C. W., Peterson, E. D., & Fonarow, G. C. (2008) · Circulation · doi:10.1161/circulationaha.108.773838 · Cited by 98
Inappropriate therapy and fatal proarrhythmia by an implantable cardioverter-defibrillator
Vollmann, D., Lüthje, L., Vonhof, S., & Unterberg, C. (2005) · Heart Rhythm · doi:10.1016/j.hrthm.2004.11.019 · Cited by 94
Maloney, J. D., Masterson, M., Khoury, D. S., Trohman, R. G., Wilkoff, B. L., Simmons, T. W., Morant, V. A., & CASTLE, L. W. (1991) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1991.tb05107.x · Cited by 76
Hemodynamic deterioration during ICD implant: Predictors of high-risk patients
Steinbeck, G., Dorwarth, U., MATTKE, S., Hoffmann, E., Markewitz, A., Kaulbach, H. G., & Tassani, P. (1994) · American Heart Journal · doi:10.1016/0002-8703(94)90088-4 · Cited by 71
Is cardiac resynchronization therapy cost-effective?
Boriani, G., Biffi, M., Martignani, C., Valzania, C., Diemberger, I., Bertini, M., Domenichini, G., Ziacchi, M., & Branzi, A. (2009) · EP Europace · doi:10.1093/europace/eup274 · Cited by 19
Self-citation
Cited by
Safety and Efficacy of a Totally Subcutaneous Implantable-Cardioverter Defibrillator
Weiss, R., Knight, B. P., Gold, M. R., Leon, A. R., Herre, J. M., Hood, M., Rashtian, M., Kremers, M., Crozier, I., Lee, K. L., et al. (2013) · Circulation · doi:10.1161/circulationaha.113.003042 · Cited by 557
2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias
Cronin, E. M., Bogun, F. M., Maury, P., Peichl, P., Chen, M., Namboodiri, N., Aguinaga, L., Leite, L. R., Al-Khatib, S. M., Anter, E., et al. (2019) · EP Europace · doi:10.1093/europace/euz132 · Cited by 475
2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias
Cronin, E. M., Bogun, F. M., Maury, P., Peichl, P., Chen, M., Namboodiri, N., Aguinaga, L., Leite, L. R., Al-Khatib, S. M., Anter, E., et al. (2020) · Heart Rhythm · doi:10.1016/j.hrthm.2019.03.002 · Cited by 431
Gasparini, M., Proclemer, A., Klersy, C., Kloppe, A., Lunati, M., Ferrer, J. B. M., Hersi, A., Gulaj, M., Wijfels, M. C. E. F., Santi, E., et al. (2013) · JAMA · doi:10.1001/jama.2013.4598 · Cited by 407
Self-citationSchwartz, P. J., Spazzolini, C., Priori, S. G., Crotti, L., Vicentini, A., Landolina, M., Gasparini, M., Wilde, A. A. M., Knops, R. E., Denjoy, I., et al. (2010) · Circulation · doi:10.1161/circulationaha.110.950147 · Cited by 307
Self-citation
Metadata
- Authors
- Gasparini, M., Menozzi, C., Proclemer, A., Landolina, M., Iacopino, S., Carboni, A., Lombardo, E., Regoli, F., Biffi, M., Burrone, V., Denaro, A., & Boriani, G.
- Venue
- European Heart Journal
- Publisher
- Oxford University Press (OUP)
- Published
- 2009-06-29
- Volume
- 30
- Issue
- 22
- Pages
- 2758-2767
- ISSN
- 0195-668X (ISSN-L)
- 1522-9645
- OpenAlex
- W1971352825
- Cited by
- 162
- References
- 25
- FWCI
- 6.52
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.