Report of a European Society of Cardiology-European Association of Percutaneous Cardiovascular Interventions task force on the evaluation of coronary stents in Europe: executive summary
Open accessThe DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
The evaluation for European Union market approval of coronary stents falls under the Medical Device Directive that was adopted in 1993. Specific requirements for the assessment of coronary stents are laid out in supplementary advisory documents. In response to a call by the European Commission to make recommendations for a revision of the advisory document on the evaluation of coronary stents (Appendix 1 of MEDDEV 2.7.1), the European Society of Cardiology (ESC) and the European Association of Percutaneous Cardiovascular Interventions (EAPCI) established a Task Force to develop an expert advisory report. As basis for its report, the ESC-EAPCI Task Force reviewed existing processes, established a comprehensive list of all coronary drug-eluting stents that have received a CE mark to date, and undertook a systematic review of the literature of all published randomized clinical trials evaluating clinical and angiographic outcomes of coronary artery stents between 2002 and 2013. Based on these data, the TF provided recommendations to inform a new regulatory process for coronary stents. The main recommendations of the task force include implementation of a standardized non-clinical assessment of stents and a novel clinical evaluation pathway for market approval. The two-stage clinical evaluation plan includes recommendation for an initial pre-market trial with objective performance criteria (OPC) benchmarking using invasive imaging follow-up leading to conditional CE-mark approval and a subsequent mandatory, large-scale randomized trial with clinical endpoint evaluation leading to unconditional CE-mark. The data analysis from the systematic review of the Task Force may provide a basis for determination of OPC for use in future studies. This paper represents an executive summary of the Task Force's report.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Clinical End Points in Coronary Stent Trials: A Case for Standardized Definitions
Cutlip, D. E., Windecker, S., Mehran, R., Boam, A., Cohen, D. J., van Es, G.-A., Gabriel Steg, P., Morel, M.-A., Mauri, L., Vranckx, P., et al. (2007) · Circulation · doi:10.1161/circulationaha.106.685313 · Cited by 5,497
Self-citationTearney, G. J., Regar, E., Akasaka, T., Adriaenssens, T., Barlis, P., Bezerra, H. G., Bouma, B., Bruining, N., Cho, J.-M., Chowdhary, S., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2011.09.079 · Cited by 1,874
Self-citationSerruys, P. W., Ormiston, J. A., Onuma, Y., Regar, E., Gonzalo, N., Garcia-Garcia, H. M., Nieman, K., Bruining, N., Dorange, C., Miquel-Hébert, K., et al. (2009) · The Lancet · doi:10.1016/s0140-6736(09)60325-1 · Cited by 820
Self-citationThe Randomized Registry Trial — The Next Disruptive Technology in Clinical Research?
Lauer, M. S., & D'Agostino, R. B. (2013) · New England Journal of Medicine · doi:10.1056/nejmp1310102 · Cited by 595 · cited 1×
“Some groups have recently proposed to incorporate randomization within nationwide clinical registries, which may increase clinical relevance and applicability of trial results.(31) In coronary stent trials, composite endpoints that capture events clearly related to the mechanism of the study device— device-oriented co…”
BackgroundOtsuka, F., Byrne, R. A., Yahagi, K., Mori, H., Ladich, E., Fowler, D. R., Kutys, R., Xhepa, E., Kastrati, A., Virmani, R., et al. (2015) · European Heart Journal · doi:10.1093/eurheartj/ehv205 · Cited by 491 · cited 1×
“In addition, its high resolution facilitates detailed characterization of neointimal tissue as well as processes such as neoatherosclerosis at a tissue level.(25) However, thus far histopathological correlation data remain scant and the clinical implications of OCT acquired datasets are unclear.”
BackgroundSelf-citationDeveloping the Sentinel System — A National Resource for Evidence Development
Behrman, R. E., Benner, J. S., Brown, J. S., McClellan, M., Woodcock, J., & Platt, R. (2011) · New England Journal of Medicine · doi:10.1056/nejmp1014427 · Cited by 323 · cited 1×
“In particular, there is concern of underreporting of serious adverse events related to malfunction of medical devices following market approval.(34,35) The Task Force therefore highlights the importance of long-term follow-up in populations representative of routine clinical practice and proposes the concept of mandat…”
BackgroundDrug-Eluting Stents in Preclinical Studies: Recommended Evaluation From a Consensus Group
Schwartz, R. S., Edelman, E. R., Carter, A., Chronos, N., Rogers, C., Robinson, K. A., Waksman, R., Weinberger, J., Wilensky, R. L., Jensen, D. N., et al. (2002) · Circulation · doi:10.1161/01.cir.0000033485.20594.6f · Cited by 291
Durability of Antirestenotic Efficacy in Drug-Eluting Stents With and Without Permanent Polymer
Byrne, R. A., Iijima, R., Mehilli, J., Pinieck, S., Bruskina, O., Schömig, A., & Kastrati, A. (2009) · JACC: Cardiovascular Interventions · doi:10.1016/j.jcin.2008.11.015 · Cited by 187 · cited 1×
“The Task Force recommends that surveillance for the determination of angiographic endpoints be undertaken at 6–9 months after stent implantation, a time point which permits benchmarking of data against the majority of existing datasets, though it should be recognized that late loss after DES seems to be an on-going dy…”
BackgroundSelf-citationRäber, L., Wohlwend, L., Wigger, M., Togni, M., Wandel, S., Wenaweser, P., Cook, S., Moschovitis, A., Vogel, R., Kalesan, B., et al. (2011) · Circulation · doi:10.1161/circulationaha.110.004762 · Cited by 186 · cited 1×
“The Task Force recommends that surveillance for the determination of angiographic endpoints be undertaken at 6–9 months after stent implantation, a time point which permits benchmarking of data against the majority of existing datasets, though it should be recognized that late loss after DES seems to be an on-going dy…”
BackgroundSelf-citationAngiographic Surrogate End Points in Drug-Eluting Stent Trials
Pocock, S. J., Lansky, A. J., Mehran, R., Popma, J. J., Fahy, M. P., Na, Y., Dangas, G., Moses, J. W., Pucelikova, T., Kandzari, D. E., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.07.084 · Cited by 162
Randomized Clinical Trials — Removing Unnecessary Obstacles
Reith, C., Landray, M., Devereaux, P. J., Bosch, J., Granger, C. B., Baigent, C., Califf, R. M., Collins, R., & Yusuf, S. (2013) · New England Journal of Medicine · doi:10.1056/nejmsb1300760 · Cited by 123 · cited 2×
“In this respect, the Task Force supports harmonization and streamlining initiatives such as Voluntary Harmonization Procedure.13 A number of obstacles in the conduct of randomized clinical trials were highlighted by members of the Sensible Guidelines Group.14 Thus, the initiation process to conduct a clinical trial re…”
BackgroundFraser, A. G., Daubert, C., de Werf, F. V., Estes, N. A. M., Smith, S. C., Krucoff, M. W., Vardas, P. E., Michel Komajda, N., participants, O. B. O. T., Anker, S. D., et al. (2011) · European Heart Journal · doi:10.1093/eurheartj/ehr171 · Cited by 94 · cited 2×
“Devices are assigned to four groups according to risk to patient and/or user (I/IIa/IIb/III).(4) Stents are in class III and therefore require ‘explicit prior authorization with regard to conformity’.”
BackgroundMethodSelf-citationProsthetic Heart Valves: Objective Performance Criteria Versus Randomized Clinical Trial
Grunkemeier, G. L., Jin, R., & STARR, A. M. (2006) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2006.06.037 · Cited by 77
Commentary: International collaboration needed on device clinical standards
Fraser, A. G., Krucoff, M. W., Brindis, R. G., Michel Komajda, N., & S., S. C. (2011) · BMJ · doi:10.1136/bmj.d2952 · Cited by 30 · cited 1×
“The European process for device regulation is usually regarded as being less onerous than the processes in some other jurisdictions.(4,11) This has advantages for patients in terms of timely access to important device innovations.”
MethodUser Fees and Beyond — The FDA Safety and Innovation Act of 2012
Kramer, D. B., & Kesselheim, A. S. (2012) · New England Journal of Medicine · doi:10.1056/nejmp1207800 · Cited by 29
Commentary: The risk of over-regulation
Mario, C. D., James, S. K., Dudek, D., Sabaté, M., & Deǧertekin, M. (2011) · BMJ · doi:10.1136/bmj.d3021 · Cited by 15 · cited 1×
“This has advantages for patients in terms of timely access to important device innovations.(12) However, some concerns have been expressed regarding the thoroughness of evaluation of devices.”
BackgroundSelf-citation
Cited by
2018 ESC/EACTS guidelines on myocardial revascularization
Neumann, F.-J., Sousa-Uva, M., Ahlsson, A., Alfonso, F., Banning, A. P., Benedetto, U., Byrne, R. A., Collet, J.-P., Falk, V., Head, S. J., et al. (2019) · European Heart Journal · doi:10.1093/eurheartj/ehy394 · Cited by 7,167
Self-citationCollet, J.-P., Thiele, H., Barbato, E., Barthélémy, O., Bauersachs, J., Bhatt, D. L., Dendale, P., Dorobantu, M., Edvardsen, T., Folliguet, T., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehaa575 · Cited by 5,037
Self-citationValgimigli, M., Bueno, H., Byrne, R. A., Collet, J.-P., Costa, F., Jeppsson, A., Jüni, P., Kastrati, A., Kolh, P., Mauri, L., et al. (2018) · European Heart Journal · doi:10.1093/eurheartj/ehx419 · Cited by 2,987
Self-citation2018 ESC/EACTS Guidelines on myocardial revascularization
Neumann, F.-J., Sousa-Uva, M., Ahlsson, A., Alfonso, F., Banning, A. P., Benedetto, U., Byrne, R. A., Collet, J.-P., Falk, V., Head, S. J., et al. (2019) · EuroIntervention · doi:10.4244/eijy19m01_01 · Cited by 1,259
Self-citationGarcia-Garcia, H. M., McFadden, E. P., Farb, A., Mehran, R., Stone, G. W., Spertus, J., Onuma, Y., Morel, M.-A., van Es, G.-A., Zuckerman, B., et al. (2018) · Circulation · doi:10.1161/circulationaha.117.029289 · Cited by 990
Self-citation
Metadata
- Authors
- Byrne, R. A., Serruys, P. W., Baumbach, A., Escaned, J., Fajadet, J., James, S., Joner, M., Oktay, S., Jüni, P., Kastrati, A., Sianos, G., Stefanini, G. G., Wijns, W., & Windecker, S.
- Venue
- European Heart Journal
- Publisher
- Oxford University Press (OUP)
- Published
- 2015-06-12
- Volume
- 36
- Issue
- 38
- Pages
- 2608-2620
- ISSN
- 0195-668X (ISSN-L)
- 1522-9645
- OpenAlex
- W2148529685
- Cited by
- 231
- References
- 16
- FWCI
- 29.69
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.