Sustained safety and performance of the second-generation drug-eluting absorbable metal scaffold in patients with de novo coronary lesions: 12-month clinical results and angiographic findings of the BIOSOLVE-II first-in-man trial
Open accessThe DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index.
Abstract
AIMS: Metal absorbable scaffolds constitute a conceptually attractive alternative to polymeric scaffolds. Promising 6-month outcomes of a second-generation drug-eluting absorbable metal scaffold (DREAMS 2G), consisting of an absorbable magnesium scaffold backbone, have been reported. We assessed the 12-month safety and performance of this novel device. METHODS AND RESULTS: The prospective, international, multi-centre, first-in-man BIOSOLVE-II trial enrolled 123 patients with up to two de novo lesions with a reference diameter between 2.2 and 3.7 mm. All patients were scheduled for angiographic follow-up at 6 months, and-if subjects consented-at 12 months. Dual antiplatelet therapy was recommended for 6 months. Quantitative coronary angiography (QCA) parameters remained stable from 6 to 12 months [paired data of 42 patients: in-segment late lumen loss 0.20 ± 0.21 mm vs. 0.25 ± 0.22 mm, P = 0.117, Δ 0.05 ± 0.21 mm (95% CI: -0.01;0.12); in-scaffold late lumen loss 0.37 ± 0.25 mm vs. 0.39 ± 0.27 mm, P = 0.446, Δ 0.03 ± 0.22 (95% CI: -0.04;0.10), respectively]. Intravascular ultrasound and optical coherence tomography findings corroborated the QCA results. Target lesion failure occurred in four patients (3.4%), consisting of one death of unknown cause, one target-vessel myocardial infarction, and two clinically driven target lesion revascularization. No additional event occurred beyond the 6-month follow-up. During the entire follow-up of 12 months, none of the patients experienced a definite or probable scaffold thrombosis. CONCLUSION: The novel drug-eluting metal absorbable scaffold DREAMS 2G showed a continuous favourable safety profile up to 12 months and stable angiographic parameters between 6 and 12 months. CLINICALTRIALSGOV IDENTIFIER: NCT01960504.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Clinical End Points in Coronary Stent Trials: A Case for Standardized Definitions
Cutlip, D. E., Windecker, S., Mehran, R., Boam, A., Cohen, D. J., van Es, G.-A., Gabriel Steg, P., Morel, M.-A., Mauri, L., Vranckx, P., et al. (2007) · Circulation · doi:10.1161/circulationaha.106.685313 · Cited by 5,497
Zheng, Y. F., Gu, X. N., & Witte, F. (2014) · Materials Science and Engineering: R: Reports · doi:10.1016/j.mser.2014.01.001 · Cited by 2,436 · cited 1×
“In brief, the biodegradation of magnesium occurs by means of anions and cations exchange, magnesium hydroxide is formed, and in a second step, magnesium hydroxide is converted to amorphous calcium phosphate.(22,23)”
Erbel, R., Di Mario, C., Bartunek, J., Bonnier, J., de Bruyne, B., Eberli, F. R., Erne, P., Haude, M., Heublein, B., Horrigan, M., et al. (2007) · The Lancet · doi:10.1016/s0140-6736(07)60853-8 · Cited by 907 · cited 2×
“Additional longer-term follow-up evaluation is required to confirm luminal enlargement, as shown for other polymeric scaffolds or previous versions of the magnesium scaffold implanted in the PROGRESS and BIOSOLVE-I studies.(5,8,24)”
Self-citationMoussa, I. D., Klein, L. W., Shah, B., Mehran, R., Mack, M. J., Brilakis, E. S., Reilly, J. P., Zoghbi, G., Holper, E., & Stone, G. W. (2013) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2013.08.720 · Cited by 650
Biodegradable Metals for Cardiovascular Stent Application: Interests and New Opportunities
Moravej, M., & Mantovani, D. (2011) · International Journal of Molecular Sciences · doi:10.3390/ijms12074250 · Cited by 600 · cited 1×
“(b) Magnesium has a negatively charged surface promoting anti-thrombotic properties in vivo.(1,20) However, this advantage might be hypothetical as the scaffold has been shown to be covered and embedded in the vessel prior to the absorption of the drug–polymer coating.”
Haude, M., Erbel, R., Erne, P., Verheye, S., Degen, H., Böse, D., Vermeersch, P., Wijnbergen, I., Weissman, N., Prati, F., et al. (2013) · The Lancet · doi:10.1016/s0140-6736(12)61765-6 · Cited by 423 · cited 2×
“39 mm).(7) There are reports that found a significant relationship between LLL and TLR even in patients with a LLL of ,0.”
Self-citationCassese, S., Byrne, R. A., Ndrepepa, G., Kufner, S., Wiebe, J., Repp, J., Schunkert, H., Fusaro, M., Kimura, T., & Kastrati, A. (2016) · The Lancet · doi:10.1016/s0140-6736(15)00979-4 · Cited by 369 · cited 8×
“Correspondingly, a meta-analysis of RCTs showed that BRS had similar 12-month TLR rates as new-generation DES despite inferior LLL.(3) This might also explain the good clinical outcomes in our series despite a higher LLL.”
InfluentialHaude, M., Ince, H., Abizaid, A., Toelg, R., Lemos, P. A., von Birgelen, C., Christiansen, E. H., Wijns, W., Neumann, F.-J., Kaiser, C., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(15)00447-x · Cited by 349 · cited 13×
“Absorbable metal scaffolds constitute an attractive alternative to polymeric BRS, as they conform more to the technique of percutaneous coronary intervention with DES.(6) We assessed a novel second-generation drug-eluting absorbable metal scaffold (DREAMS 2G), which has been redesigned iteratively by (i) improving the…”
InfluentialSelf-citationBioresorbable scaffolds: rationale, current status, challenges, and future
Iqbal, J., Onuma, Y., Ormiston, J., Abizaid, A., Waksman, R., & Serruys, P. (2014) · European Heart Journal · doi:10.1093/eurheartj/eht542 · Cited by 269 · cited 1×
“Bioresorbable scaffolds (BRS) have been designed to overcome problems related to the long-term persistence of metallic stents implanted in coronary arteries.(1,2) Several randomized controlled trials (RCTs) comparing polymeric BRS with new-generation drug-eluting stents (DES) found similar clinical outcomes at 1 year.”
Self-citationPilgrim, T., Heg, D., Roffi, M., Tüller, D., Muller, O., Vuilliomenet, A., Cook, S., Weilenmann, D., Kaiser, C., Jamshidi, P., et al. (2014) · The Lancet · doi:10.1016/s0140-6736(14)61038-2 · Cited by 267 · cited 4×
“Funding to pay the Open Access publication charges for this article was provided by Biotronik AG, Buelach, Switzerland.”
InfluentialSelf-citationSerruys, P. W., Onuma, Y., Garcia-Garcia, H. M., Muramatsu, T., van Geuns, R.-J., de Bruyne, B., Dudek, D., Thuesen, L., Smits, P. C., Chevalier, B., et al. (2014) · EuroIntervention · doi:10.4244/eijv9i11a217 · Cited by 239
Self-citationByrne, R. A., Serruys, P. W., Baumbach, A., Escaned, J., Fajadet, J., James, S., Joner, M., Oktay, S., Jüni, P., Kastrati, A., et al. (2015) · European Heart Journal · doi:10.1093/eurheartj/ehv203 · Cited by 231 · cited 4×
“Secondary endpoints at 12 months were TLF, a composite of cardiac death,11 target-vessel myocardial infarction,12 and clinically TLR11; scaffold thrombosis11; in-scaffold and in-segment binary restenosis; diameter stenosis; and in-scaffold LLL.13 A clinical event committee adjudicated all adverse events, and an indepe…”
InfluentialSelf-citationOnuma, Y., Dudek, D., Thuesen, L., Webster, M., Nieman, K., Garcia-Garcia, H. M., Ormiston, J. A., & Serruys, P. W. (2013) · JACC: Cardiovascular Interventions · doi:10.1016/j.jcin.2013.05.017 · Cited by 215 · cited 2×
“So far, non-randomized studies up to 5 years showed promising outcomes.(4,5)”
Self-citationLate Loss in Lumen Diameter and Binary Restenosis for Drug-Eluting Stent Comparison
Mauri, L., Orav, E. J., & Kuntz, R. E. (2005) · Circulation · doi:10.1161/circulationaha.104.513952 · Cited by 203 · cited 1×
“5 mm has been associated with a maximal TLR rate of 5%(15) and a binary restenosis rate of 5%.(16) Furthermore, in an evaluation from 11 RCTs, Pocock et al.”
Lipinski, M. J., Escarcega, R. O., Baker, N. C., Benn, H. A., Gaglia, M. A., Torguson, R., & Waksman, R. (2016) · JACC: Cardiovascular Interventions · doi:10.1016/j.jcin.2015.09.024 · Cited by 168 · cited 4×
“Only one study device per lesion was allowed, although in bailout situations, a second DREAMS 2G could be used, and, in case of failure, an Orsiro DES.”
InfluentialSelf-citationAngiographic Surrogate End Points in Drug-Eluting Stent Trials
Pocock, S. J., Lansky, A. J., Mehran, R., Popma, J. J., Fahy, M. P., Na, Y., Dangas, G., Moses, J. W., Pucelikova, T., Kandzari, D. E., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.07.084 · Cited by 162
Ellis, S. G., Popma, J. J., Lasala, J. M., Koglin, J. J., Cox, D. A., Hermiller, J., O’Shaughnessy, C., Mann, J. T., Turco, M., Caputo, R., et al. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2004.11.063 · Cited by 158
Wittchow, E., Adden, N., Riedmüller, J., Savard, C., Waksman, R., & Braune, M. (2013) · EuroIntervention · doi:10.4244/eijv8i12a218 · Cited by 99 · cited 1×
“In brief, the biodegradation of magnesium occurs by means of anions and cations exchange, magnesium hydroxide is formed, and in a second step, magnesium hydroxide is converted to amorphous calcium phosphate.(22,23)”
Self-citationCurrent status of transcatheter valve therapy in Europe: results from an EAPCI survey
Petronio, A. S., Capranzano, P., Barbato, E., Piazza, N., Baumbach, A., Haude, M., & Windecker, S. (2016) · EuroIntervention · doi:10.4244/eijy16m06_01 · Cited by 96
Self-citationHaude, M., Erbel, R. M., Erné, P., Verheye, S., Degen, H., Vermeersch, P. H. M. J., Weissman, N. J., Prati, F., Bruining, N., Waksman, R., et al. (2016) · EuroIntervention · doi:10.4244/eij-d-15-00371 · Cited by 74
Self-citationRobaei, D., Back, L. M., Ooi, S., Pitney, M., & Jepson, N. S. (2016) · PubMed · Cited by 14
Cited by
Updates on the research and development of absorbable metals for biomedical applications
Hermawan, H. (2018) · Progress in Biomaterials · doi:10.1007/s40204-018-0091-4 · Cited by 260
Fully bioresorbable vascular scaffolds: lessons learned and future directions
Jinnouchi, H., Torii, S., Sakamoto, A., Kolodgie, F. D., Virmani, R., & Finn, A. V. (2019) · Nature Reviews Cardiology · doi:10.1038/s41569-018-0124-7 · Cited by 228
Long-term surveillance of zinc implant in murine artery: Surprisingly steady biocorrosion rate
Drelich, A. J., Zhao, S., Guillory, R. J., Drelich, J. W., & Goldman, J. (2017) · Acta Biomaterialia · doi:10.1016/j.actbio.2017.05.045 · Cited by 188
“However, it is still estimated that ~95% of the magnesium scaffold will be degraded within 12 months, the lower end of the 1 to 2-year ideal degradation benchmark, with a late lumen loss that is higher than comparable stent systems such as ABSORB and DESolve [35,36].”
BackgroundLin, W., Qin, L., Qi, H., Zhang, D., Zhang, G., Gao, R., Qiu, H., Xia, Y., Cao, P., Wang, X., et al. (2017) · Acta Biomaterialia · doi:10.1016/j.actbio.2017.03.020 · Cited by 183
Wei, L., & Gao, Z. (2023) · RSC Advances · doi:10.1039/d2ra07829e · Cited by 162
Metadata
- Authors
- Haude, M., Ince, H., Abizaid, A., Toelg, R., Lemos, P. A., von Birgelen, C., Christiansen, E. H., Wijns, W., Neumann, F.-J., Kaiser, C., Eeckhout, E., Lim, S. T., Escaned, J., Onuma, Y., Garcia-Garcia, H. M., & Waksman, R.
- Venue
- European Heart Journal
- Publisher
- Oxford University Press (OUP)
- Published
- 2016-05-17
- Volume
- 37
- Issue
- 35
- Pages
- 2701-2709
- ISSN
- 0195-668X (ISSN-L)
- 1522-9645
- OpenAlex
- W2401593360
- Cited by
- 189
- References
- 21
- FWCI
- 31.53
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.