Comparison of microfluidic digital PCR and conventional quantitative PCR for measuring copy number variation
Open accessThe DOI matches its Crossref record; 50 of the 53 references it cites were found in a registry or an index; 15 of the 67 sources its claims cite were checked for support.
Abstract
One of the benefits of Digital PCR (dPCR) is the potential for unparalleled precision enabling smaller fold change measurements. An example of an assessment that could benefit from such improved precision is the measurement of tumour-associated copy number variation (CNV) in the cell free DNA (cfDNA) fraction of patient blood plasma. To investigate the potential precision of dPCR and compare it with the established technique of quantitative PCR (qPCR), we used breast cancer cell lines to investigate HER2 gene amplification and modelled a range of different CNVs. We showed that, with equal experimental replication, dPCR could measure a smaller CNV than qPCR. As dPCR precision is directly dependent upon both the number of replicate measurements and the template concentration, we also developed a method to assist the design of dPCR experiments for measuring CNV. Using an existing model (based on Poisson and binomial distributions) to derive an expression for the variance inherent in dPCR, we produced a power calculation to define the experimental size required to reliably detect a given fold change at a given template concentration. This work will facilitate any future translation of dPCR to key diagnostic applications, such as cancer diagnostics and analysis of cfDNA.
Overview
CoChat Summary
TL;DR
This study directly compares microfluidic digital PCR (dPCR) with conventional quantitative PCR (qPCR) for measuring HER2 copy number variation using breast cancer cell line DNA and spiked DNA models. It reports that dPCR can detect smaller CNVs than qPCR at equal replication and provides a Poisson/binomial variance model with power calculations to design dPCR experiments based on template concentration.
Key findings
- dPCR and qPCR gave similar HER2:RNase P ratios for normal female gDNA (1.03 vs 1.00; P = 0.39), MCF-7 (0.44 by both; P = 0.71), and T-47D (2.00 vs 1.96; P = 0.52), but SK-BR-3 was significantly lower by dPCR (7.15) than qPCR (9.43; P = 0.00005).
- In an in vitro gene-amplification model at low copy number (2.1 ng/μl; λr ≈ 0.2), dPCR detected a ratio of ≥1.17 with eight panels (P < 0.0003), whereas qPCR detected ratios of ≥1.27 with eight wells (P < 0.0005); a 1.12 ratio was not significantly different from normal female gDNA by dPCR with eight panels, and ratios ≤1.22 were not significantly different by qPCR.
- Power calculations predicted that with 95% power and λr = 0.2, a 1.2-fold change was detectable with five 770-chamber panels per assay, ratios of 1.1 and below required >15 panels, and the smallest CNV measurable with eight panels was approximately 1.15.
- For qPCR, power curves indicated ratios greater than 1.25 could be measured with 95% power and eight replicate wells, while an excess of 20 qPCR wells would be needed to measure ratios of 1.17 or fewer.
Methods
- Genomic DNA from three breast cancer cell lines with different HER2 status (SK-BR-3, T-47D, MCF-7) and pooled normal female gDNA was analyzed with HER2 and RNase P assays by qPCR and microfluidic dPCR on 48.770 arrays.
- An in vitro CNV model was created by spiking T-47D gDNA into normal female gDNA to generate theoretical HER2:RNase P ratios between 1.00 and 2.00 at low copy number.
- dPCR quantification used Poisson estimation of molecules per chamber (λ = −ln(1 − H/C)); qPCR used standard curves. Statistical comparisons used ANOVA and two-tailed Student’s t-tests, and power calculations were performed in R based on a derived variance model for the log ratio.
Limitations
- Stated: the proof-of-principle experiments used high-quality DNA template to model cfDNA; translation will require consideration of template integrity, PCR-assay efficiency, and matrix effects.
- Stated: dPCR may not accurately quantify concatameric HER2 amplifications such as those in SK-BR-3, and pre-amplification to separate linked copies requires careful validation because of potential bias.
- Evident from the text: the findings are tied to one microfluidic dPCR platform (Fluidigm 48.770), specific HER2/RNase P assays, and a one-tailed power model focused on HER2 amplification, so generalizability to other platforms, assays, or deletion/two-tailed CNV settings is not established here.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures



Verification
Claims the paper makes
13 Valid · 1 Not valid · 16 Not checked · 62 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Analysis of the SK-BR-3 gDNA identified high levels of HER2 amplification that corresponds with the published copy number range (14–24 copies per diploid genome) (41,43–45,47).”
§ Comparison of microfluidic digital PCR and conventional quantitative PCR for measuring copy number variation > DISCUSSION
What the sources say
- Not checked — no full text of the sourcecites [41] Konigshoff, 2003, HER-2/neu gene copy number quantified by real-time PCR: comparison of gene amplification, heterozygosity, and immunohistochemical status in breast cancer tissue, Clin. Chem. source
- Not checked — no full text of the sourcecites [43] Sestini, 1995, Gene amplification for c-erbB-2, c-myc, epidermal growth factor receptor, int-2, and N-myc measured by quantitative PCR with a multiple competitor template, Clin. Chem. source
- Not checked — no full text of the sourcecites [44] Lyon, 2001, Quantification of HER2/neu gene amplification by competitive pcr using fluorescent melting curve analysis, Clin. Chem. source
- Partly supportedcites [45] Millson, 2003, Comparison of two quantitative polymerase chain reaction methods for detecting HER2/neu amplification, J. Mol. Diagn. sourceExperimental — accuracy not yet measured
The Results (p. 4) report high HER2/neu amplification in SKBR3, with a gene dose of 10 relative to the single-copy MRC-5 control; the competitive PCR results (p. 4) give a dose of 11. Our copy does not state the claimed published range of 14–24 copies per diploid genome or explicitly equate its measurements with that range. (LLM, conf=0.84)
- Not checked — no full text of the sourcecites [47] Hynes, 1989, Overexpression of the c-erbB-2 protein in human breast tumor cell lines, J. Cell Biochem. source
plainstates an associationSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
Every source the paper cites for this claim was checked and supports it.
In the paper
“it is also essential that some estimation of efficiency is made (25).”
§ Comparison of microfluidic digital PCR and conventional quantitative PCR for measuring copy number variation > DISCUSSION
What the sources say
- Supportedcites [25] Bustin, 2009, The MIQE guidelines: minimum information for publication of quantitative real-time PCR experiments, Clin. Chem. source
Section 7.4 lists PCR efficiency as a required assay-performance measurement. Section 7.4.1 says amplification efficiency must be established using calibration curves and explains that differences in efficiency can make comparisons of relative concentrations inaccurate. This supports the need to estimate efficiency when measuring fold changes. (LLM, conf=0.91)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“T-47D gDNA gave a HER2:RNase P ratio of < 2 (3–4 copies per diploid genome) (43–46).”
§ Comparison of microfluidic digital PCR and conventional quantitative PCR for measuring copy number variation > DISCUSSION
What the sources say
- Not checked — no full text of the sourcecites [43] Sestini, 1995, Gene amplification for c-erbB-2, c-myc, epidermal growth factor receptor, int-2, and N-myc measured by quantitative PCR with a multiple competitor template, Clin. Chem. source
- Not checked — no full text of the sourcecites [44] Lyon, 2001, Quantification of HER2/neu gene amplification by competitive pcr using fluorescent melting curve analysis, Clin. Chem. source
- Partly supportedcites [45] Millson, 2003, Comparison of two quantitative polymerase chain reaction methods for detecting HER2/neu amplification, J. Mol. Diagn. sourceExperimental — accuracy not yet measured
Sample Preparation (p. 2) and Results (p. 4) describe T47D as having approximately four HER2/neu copies, supporting the copy-number portion of the claim. The paper uses β-globin, not RNase P, as its reference gene; Results (p. 4) reports a T47D gene dose of 2, and the β-globin section (p. 4) reports 2.2 by the other method. It therefore does not support a HER2:RNase P ratio below 2. (LLM, conf=0.89)
- Not checked — no full text of the sourcecites [46] Dati, 1990, Inhibition of c-erbB-2 oncogene expression by estrogens in human breast cancer cells, Oncogene
plainSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“This is a crucial consideration when considering cfDNA as a template as its concentration can range considerably; from 2 to 30 ng/ml plasma in healthy individuals to 180–600 ng/ml plasma both during pregnancy and in cancer patients (13,16,18,22–24).”
§ Comparison of microfluidic digital PCR and conventional quantitative PCR for measuring copy number variation > INTRODUCTION
What the sources say
- Not checked — no full text of the sourcecites [13] Leon, 1977, Free DNA in the serum of cancer patients and the effect of therapy, Cancer Res.
- Not checked — no full text of the sourcecites [16] Levenson, 2007, Biomarkers for early detection of breast cancer: what, when, and where? Biochim, Biophys. Acta source
- Not checked — no full text of the sourcecites [18] Jung, 2010, Cell-free DNA in the blood as a solid tumor biomarker–a critical appraisal of the literature, Clin. Chim. Acta source
- Not checked — no full text of the sourcecites [22] Fournie, 1986, Recovery of nanogram quantities of DNA from plasma and quantitative measurement using labeling by nick translation, Anal. Biochem. source
- Not checked — no full text of the sourcecites [23] Giacona, 1998, Cell-free DNA in human blood plasma: length measurements in patients with pancreatic cancer and healthy controls, Pancreas source
- Partly supportedcites [24] Diehl, 2005, Detection and quantification of mutations in the plasma of patients with colorectal tumors, Proc. Natl Acad. Sci. USA sourceExperimental — accuracy not yet measured
The Results (page unavailable) report more circulating APC gene fragments in advanced colorectal cancer patients than in donors without tumors, supporting the general point that plasma DNA levels vary. The paper gives fragment counts per ml, not the claimed 2–30 or 180–600 ng/ml concentration ranges, and it does not report cfDNA concentrations during pregnancy. (LLM, conf=0.86)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In this study, a CNV ratio of 1.17 was significantly detected using eight panels”
§ Comparison of microfluidic digital PCR and conventional quantitative PCR for measuring copy number variation > DISCUSSION
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
43 more claims
Create a free account to see all 92 claims and to verify the 2 references still open.
- Probably not a substantial claim. 18
- It may state two separate claims as one. 31
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Since we applied the same assay and samples for both platforms, these differences most likely arise from general technical differences, causing a certain degree of measurement uncertainty that can influence the precision of estimates especially at the lower range of copy number concentrations42,53,54.”
From Comparing the precision of two digital PCR applications for copy number comparisons in protists (2025) · cites this paper as [42]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Self-check results appear here once an independent check has confirmed them.
Cited works that were retracted
Works this paper cites that were later retracted, from the retraction notices publishers deposit. Major when the notice came out before this paper was published.
A note, not an issue: for example, a cited work that was retracted only after this paper was published, which its authors could not have known.
In the paper
“Shlien, 2010, Copy number variations and cancer susceptibility, Curr. Opin. Oncol., 22, 55, 10.1097/CCO.0b013e328333dca4”
Cited work 10.1097/cco.0b013e328333dca4 · retracted 2016-09-01 · notice 10.1097/01.cco.0000491182.00738.ba · this paper published 2012-02-28
Citations
Cites
1 retracted
Copy number variations and cancer susceptibility
Shlien, A., & Malkin, D. (2010) · Current Opinion in Oncology · doi:10.1097/cco.0b013e328333dca4 · Cited by 77
RetractedA new mathematical model for relative quantification in real-time RT-PCR
Pfaffl, M. W. (2001) · Nucleic Acids Research · doi:10.1093/nar/29.9.e45 · Cited by 34,610
The MIQE guidelines: minimum information for publication of quantitative real-time PCR experiments
Bustin, S. A., Benes, V., Garson, J. A., Hellemans, J., Huggett, J., Kubista, M., Mueller, R., Nolan, T., Pfaffl, M. W., Shipley, G. L., et al. (2009) · Clinical Chemistry · doi:10.1373/clinchem.2008.112797 · Cited by 16,303 · cited 7×
“Real-time quantitative PCR was performed in accordance with the MIQE guidelines (Supplementary Table S1 and Supplementary Figure S1) (25).”
BackgroundMethodInfluentialSelf-citationSlamon, D. J., Clark, G. M., Wong, S. G., Levin, W. J., Ullrich, A., & McGuire, W. L. (1987) · Science · doi:10.1126/science.3798106 · Cited by 11,760
Slamon, D. J., Leyland-Jones, B., Shak, S., Fuchs, H., Paton, V., Bajamonde, A., Fleming, T., Eiermann, W., Wolter, J., Pegram, M., et al. (2001) · New England Journal of Medicine · doi:10.1056/nejm200103153441101 · Cited by 11,500
High-throughput droplet digital PCR system for absolute quantitation of DNA copy number
Hindson, B. J., Ness, K. D., Masquelier, D. A., Belgrader, P., Heredia, N. J., Makarewicz, A. J., Bright, I. J., Lucero, M. Y., Hiddessen, A. L., Legler, T. C., et al. (2011) · Analytical Chemistry · doi:10.1021/ac202028g · Cited by 2,792
(1977) · PubMed · Cited by 1,801
Copy number variation in human health, disease, and evolution
Zhang, F., Gu, W., Hurles, M. E., & Lupski, J. R. (2009) · Annual Review of Genomics and Human Genetics · doi:10.1146/annurev.genom.9.081307.164217 · Cited by 1,207
Lamon, D. E. J. S., Eyland, B. R. L., Hak, S. T. S., Uchs, H. A. F., Aton, V. I. P., Ajamonde, A. L. B., LeMing, T., Iermann, W. O. E., Olter, J. A. W., Egram, M. A. P., et al. (2009) · Cited by 1,199
Detection and quantification of mutations in the plasma of patients with colorectal tumors
Diehl, F., Li, M., Dressman, D., He, Y., Shen, D., Szabo, S., Diaz, L. A., Goodman, S. N., David, K. A., Juhl, H., et al. (2005) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0507904102 · Cited by 1,188
Cited by
Huggett, J. F., Foy, C. A., Benes, V., Emslie, K., Garson, J. A., Haynes, R., Hellemans, J., Kubista, M., Mueller, R. D., Nolan, T., et al. (2013) · Clinical Chemistry · doi:10.1373/clinchem.2013.206375 · Cited by 859
Self-citationQuan, P.-L., Sauzade, M., & Brouzes, E. (2018) · Sensors · doi:10.3390/s18041271 · Cited by 664
“In agreement with statistical considerations, dPCR precision depends on both the number of replicates and the template concentration [33].”
BackgroundInfluentialThe dMIQE Group, Whale, A. S., De Spiegelaere, W., Trypsteen, W., Nour, A. A., Bae, Y.-K., Benes, V., Burke, D., Cleveland, M., Corbisier, P., et al. (2020) · Clinical Chemistry · doi:10.1093/clinchem/hvaa125 · Cited by 516
“Restriction digestion to small linear DNA fragments may equalize template differences (36, 49, 64), and prevent underestimation of linked target molecules (18).”
BackgroundSelf-citationBaker, M. (2012) · Nature Methods · doi:10.1038/nmeth.2027 · Cited by 524
Highly Precise Measurement of HIV DNA by Droplet Digital PCR
Strain, M. C., Lada, S. M., Luong, T., Rought, S. E., Gianella, S., Terry, V. H., Spina, C. A., Woelk, C. H., & Richman, D. D. (2013) · PLoS ONE · doi:10.1371/journal.pone.0055943 · Cited by 504
“Digital PCR (dPCR) [2], in which the amplification reaction is divided into thousands of microscopic reaction volumes, is a rapidly growing alternative that is potentially more accurate [3,4,5] and more precise [6,7].”
BackgroundConsiderations for Digital PCR as an Accurate Molecular Diagnostic Tool
Huggett, J. F., Cowen, S., & Foy, C. A. (2015) · Clinical Chemistry · doi:10.1373/clinchem.2014.221366 · Cited by 444
Self-citationComparison of Droplet Digital PCR to Real-Time PCR for Quantitative Detection of Cytomegalovirus
Hayden, R. T., Gu, Z., Ingersoll, J., Abdul-Ali, D., Shi, L., Pounds, S., & Caliendo, A. M. (2013) · Journal of Clinical Microbiology · doi:10.1128/jcm.02620-12 · Cited by 335
Song, P., Wu, L. R., Yan, Y. H., Zhang, J. X., Chu, T., Kwong, L. N., Patel, A. A., & Zhang, D. Y. (2022) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00837-3 · Cited by 319
Digital Assays Part I: Partitioning Statistics and Digital PCR
Basu, A. S. (2017) · SLAS Technology · doi:10.1177/2472630317705680 · Cited by 300
“).(84) The improved resolution enables the researcher to identify not just single deletions and duplications, but more complex alterations that play a role in disease phenotypes.”
BackgroundDevonshire, A. S., Whale, A. S., Gutteridge, A., Jones, G., Cowen, S., Foy, C. A., & Huggett, J. F. (2014) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-014-7835-3 · Cited by 299
“91 nL per droplet, and confidence intervals were calculated as described previously [45].”
MethodSelf-citationAdvances in digital polymerase chain reaction (dPCR) and its emerging biomedical applications
Cao, L., Cui, X., Hu, J., Li, Z., Choi, J. R., Yang, Q., Lin, M., Ying Hui, L., & Xu, F. (2017) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2016.09.082 · Cited by 291
“…distribution (Debski et al., 2015; Dong et al., 2015; Majumdar et al., 2015) and novel applications are suggesting negative binomial distributions as the Poisson distribution is only correct in a theoretical setting with now variation in partition sizes (Kreutz et al., 2011; Whale et al., 2012).”
BackgroundBorchardt, M. A., Boehm, A. B., Salit, M., Spencer, S. K., Wigginton, K. R., & Noble, R. T. (2021) · Environmental Science & Technology · doi:10.1021/acs.est.1c01767 · Cited by 254
Quantitative Analysis of Food and Feed Samples with Droplet Digital PCR
Morisset, D., Štebih, D., Milavec, M., Gruden, K., & Žel, J. (2013) · PLoS ONE · doi:10.1371/journal.pone.0062583 · Cited by 281
“Several advantages are proposed for the use of dPCR instead of qPCR in routine GMO testing: 1) it enables absolute target copy number to be detected and avoids the amplification efficiency bias observed with qPCR [11,19], 2) it overcomes dependence on the availability of references or standards [18], 3) it provides da…”
MethodTréguier, A., Paillisson, J., Dejean, T., Valentini, A., Schlaepfer, M. A., & Roussel, J. (2014) · Journal of Applied Ecology · doi:10.1111/1365-2664.12262 · Cited by 271
“dPCR is more appropriate than classical qPCR to the detection of small amounts of DNA molecules, and it is also more tolerant to inhibitors and does not need to rely on references or standards (Whale et al. 2012).”
BackgroundMethodBeck, J., Bierau, S., Balzer, S., Andag, R., Kanzow, P., Schmitz, J., Gaedcke, J., Moerer, O., Slotta, J. E., Walson, P., et al. (2013) · Clinical Chemistry · doi:10.1373/clinchem.2013.210328 · Cited by 271
Schoepp, N. G., Schlappi, T. S., Curtis, M. S., Butkovich, S. S., Miller, S., Humphries, R. M., & Ismagilov, R. F. (2017) · Science Translational Medicine · doi:10.1126/scitranslmed.aal3693 · Cited by 237
Fundamentals of multiplexing with digital PCR
Whale, A. S., Huggett, J. F., & Tzonev, S. (2016) · Biomolecular Detection and Quantification · doi:10.1016/j.bdq.2016.05.002 · Cited by 229
Self-citationMouliere, F., El Messaoudi, S., Pang, D., Dritschilo, A., & Thierry, A. R. (2014) · Molecular Oncology · doi:10.1016/j.molonc.2014.02.005 · Cited by 230
Sensitivity assessment of droplet digital PCR for SARS-CoV-2 detection
Falzone, L., Musso, N., Gattuso, G., Bongiorno, D., Palermo, C., Scalia, G., Libra, M., & Stefani, S. (2020) · International Journal of Molecular Medicine · doi:10.3892/ijmm.2020.4673 · Cited by 207
“currently, ddPcR is effectively used for the absolute quantification of viral load, for the analysis of circulating dNA, gene and microRNA expression and analysis of gene copy number variation (16-19).”
MethodCurrent and Emerging Applications of Droplet Digital PCR in Oncology
Olmedillas-López, S., García-Arranz, M., & García-Olmo, D. (2017) · Molecular Diagnosis & Therapy · doi:10.1007/s40291-017-0278-8 · Cited by 201
Nathan, L. M., Simmons, M., Wegleitner, B. J., Jerde, C. L., & Mahon, A. R. (2014) · Environmental Science & Technology · doi:10.1021/es5034052 · Cited by 197
Harmful Algal Blooms in Eutrophic Marine Environments: Causes, Monitoring, and Treatment
Lan, J., Liu, P., Hu, X., & Zhu, S. (2024) · Water · doi:10.3390/w16172525 · Cited by 207
Cao, Y., Raith, M. R., & Griffith, J. F. (2015) · Water Research · doi:10.1016/j.watres.2014.12.008 · Cited by 190
Day, E., Dear, P. H., & McCaughan, F. (2013) · Methods · doi:10.1016/j.ymeth.2012.08.001 · Cited by 191
Digital PCR analysis of circulating nucleic acids
Hudecova, I. (2015) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2015.03.015 · Cited by 178
“Investigations on the breast cancer cell lines showed superior precision and sensitivity of dPCR compared to qPCR in detection of subtle fold-differences of HER2 gene copy number variation [33].”
BackgroundMethodEvaluation of Digital PCR for Absolute RNA Quantification
Sanders, R., Mason, D. J., Foy, C. A., & Huggett, J. F. (2013) · PLoS ONE · doi:10.1371/journal.pone.0075296 · Cited by 174
“Much work has already been done to meticulously evaluate this technique for DNA molecular measurement [3,4,5,6,7,8].”
BackgroundSelf-citationBian, X., Jing, F., Li, G., Fan, X., Jia, C., Zhou, H., Jin, Q., & Zhao, J. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.07.016 · Cited by 174
Digital Quantification of Proteins and mRNA in Single Mammalian Cells
Albayrak, C., Jordi, C. A., Zechner, C., Lin, J., Bichsel, C. A., Khammash, M., & Tay, S. (2016) · Molecular Cell · doi:10.1016/j.molcel.2016.02.030 · Cited by 176
Centrifugal micro-channel array droplet generation for highly parallel digital PCR
Chen, Z., Liao, P., Zhang, F., Jiang, M., Zhu, Y., & Huang, Y. (2017) · Lab on a Chip · doi:10.1039/c6lc01305h · Cited by 164
Hunter, M. E., Dorazio, R. M., Butterfield, J. S. S., Meigs‐Friend, G., Nico, L. G., & Ferrante, J. A. (2016) · Molecular Ecology Resources · doi:10.1111/1755-0998.12619 · Cited by 188
“…diagnostics (e.g. cell-free DNA and pathogen detection), GMO food analysis and environmental DNA (eDNA) analyses of animals, bacteria and pathogens (Whale et al. 2012; Olson et al. 2013; Strain et al. 2013; Ussler et al. 2013; Bass et al. 2015; McFall et al. 2015; Mendes Silva & Domingues 2015;…”
BackgroundMethodInfluentialResolving rates of mutation in the brain using single-neuron genomics
Evrony, G. D., Lee, E., Park, P. J., & Walsh, C. A. (2016) · eLife · doi:10.7554/elife.12966 · Cited by 163
“There are inevitable random and frequent systematic sources of error inherent to qPCR, including sample preparation, pipetting, and instrumentation (Fernandez-Jimenez et al., 2011; Kitchen et al., 2010; Weaver et al., 2010; Whale et al., 2012).”
BackgroundMethods for Applying Accurate Digital PCR Analysis on Low Copy DNA Samples
Whale, A. S., Cowen, S., Foy, C. A., & Huggett, J. F. (2013) · PLoS ONE · doi:10.1371/journal.pone.0058177 · Cited by 160
“might explain why it is more precise than qPCR [2].”
BackgroundMethodInfluentialSelf-citationThree-color crystal digital PCR
Madic, J., Zocevic, A., Senlis, V., Fradet, E., Andre, B., Muller, S., Dangla, R., & Droniou, M. (2016) · Biomolecular Detection and Quantification · doi:10.1016/j.bdq.2016.10.002 · Cited by 161
Detecting Genetic Mosaicism in Cultures of Human Pluripotent Stem Cells
Baker, D., Hirst, A. J., Gokhale, P. J., Juárez, M. A., Williams, S., Wheeler, M., Bean, K., Allison, T. F., Moore, H. D. M., Andrews, P. W., et al. (2016) · Stem Cell Reports · doi:10.1016/j.stemcr.2016.10.003 · Cited by 150
“Sensitivity of qPCR Assay versus Digital Droplet PCR and FISH in Detecting Mosaicism in hPSC Cultures The sensitivity of PCR-based methods is known to depend on the magnitude of the copy-number change, with a difference between zero and one copy being easier to detect than a difference between two and three copies (Wh…”
MethodKiselinova, M., Pasternak, A. O., Spiegelaere, W. D., Vogelaers, D., Berkhout, B., & Vandekerckhove, L. (2014) · PLoS ONE · doi:10.1371/journal.pone.0085999 · Cited by 159
“To date, several studies on cancer and viral infections report a higher degree of sensitivity and precision of dPCR than qPCR [13,16,18,19].”
BackgroundTriple Negative Breast Cancer: A Review of Present and Future Diagnostic Modalities
Dass, S. A., Tan, K. L., Rajan, R. S., Mokhtar, N. F., Adzemi, E. R. M., Rahman, W. F. W. A., Din, T. A. D. A. T., & Balakrishnan, V. (2021) · Medicina · doi:10.3390/medicina57010062 · Cited by 189
Digital PCR on an integrated self-priming compartmentalization chip
Zhu, Q., Qiu, L., Yu, B., Xu, Y., Gao, Y., Pan, T., Tian, Q., Song, Q., Jin, W., Jin, Q., et al. (2014) · Lab on a Chip · doi:10.1039/c3lc51327k · Cited by 149
Circular RNAs in cancer: Limitations in functional studies and diagnostic potential
Li, F., Yang, Q., He, A. T., & Yang, B. B. (2020) · Seminars in Cancer Biology · doi:10.1016/j.semcancer.2020.10.002 · Cited by 141
Digital polymerase chain reaction technology – recent advances and future perspectives
Sreejith, K. R., Ooi, C. H., Jin, J., Dao, D. V., & Nguyen, N. (2018) · Lab on a Chip · doi:10.1039/c8lc00990b · Cited by 144
One-step RT-droplet digital PCR: a breakthrough in the quantification of waterborne RNA viruses
Rački, N., Morisset, D., Gutiérrez‐Aguirre, I., & Ravnikar, M. (2013) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-013-7476-y · Cited by 142
“It is more sensitive than qPCR for rare targets [14] and provides more accurate data of metrological quality [12, 13], especially at low target copy numbers [15–17].”
BackgroundLiao, J., Yang, F., Tang, Z., Yu, W., Han, Q., Hu, L., Li, Y., Guo, J., Pan, J., Ma, F., et al. (2019) · Ecotoxicology and Environmental Safety · doi:10.1016/j.ecoenv.2019.02.069 · Cited by 140
Shoda, K., Ichikawa, D., Fujita, Y., Masuda, K., Hiramoto, H., Hamada, J., Arita, T., Konishi, H., Komatsu, S., Shiozaki, A., et al. (2016) · Gastric Cancer · doi:10.1007/s10120-016-0599-z · Cited by 141
“One method showing promise for improving the limit of detection for nucleic acid quantification is digital PCR (dPCR), with a number of studies highlighting the superior accuracy of dPCR for copy number (CN) analyses [14, 15], including HER2 amplification in circulating ctDNA in breast cancer (BC) and GC [16, 17].”
Ding, X., Yin, K., Li, Z., Sfeir, M. M., & Liu, C. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113218 · Cited by 138
Bushman, D. M., Kaeser, G. E., Siddoway, B., Westra, J. W., Rivera, R. R., Rehen, S. K., Yung, Y. C., & Chun, J. (2015) · eLife · doi:10.7554/elife.05116 · Cited by 136
“Samples were run in triplicate and assays in sextuplicate, which generated high numbers of replicate data points allowing for improved copy number resolution to an extent not possible by conventional qPCR (Weaver et al., 2010; Whale et al., 2012).”
MethodQuantification of Plasma miRNAs by Digital PCR for Cancer Diagnosis
Ma, J., Li, N., Guarnera, M., & Jiang, F. (2013) · Biomarker Insights · doi:10.4137/bmi.s13154 · Cited by 145
Digital PCR as a Novel Technology and Its Potential Implications for Molecular Diagnostics
Huggett, J. F., & Whale, A. S. (2013) · Clinical Chemistry · doi:10.1373/clinchem.2013.214742 · Cited by 145
Self-citationComparison between digital PCR and real-time PCR in detection of Salmonella typhimurium in milk
Wang, M., Yang, J., Gai, Z., et al. (2017) · International Journal of Food Microbiology · doi:10.1016/j.ijfoodmicro.2017.12.011 · Cited by 122
Emerging digital PCR technology in precision medicine
Zhang, L., Parvin, R., Fan, Q., & Ye, F. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.114344 · Cited by 127
Minimal Residual Disease in Acute Lymphoblastic Leukemia: Technical and Clinical Advances
Starza, I. D., Chiaretti, S., Propris, M. S. D., Elia, L., Cavalli, M., Novi, L. A. D., Soscia, R., Messina, M., Vitale, A., Guarini, A., et al. (2019) · Frontiers in Oncology · doi:10.3389/fonc.2019.00726 · Cited by 133
“As reported in several studies (65, 66), based on the dynamic nature of two methods, ddPCR appears more accurate than RQ-PCR since: (i) each sample is partitioned in droplets and each droplet is analyzed individually, (ii) small changes in fluorescence intensity are more readily detected, and (iii) the ratio between t…”
MethodResultmiRNAs as novel biomarkers in the management of prostate cancer
Filella, X., & Foj, L. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-1073 · Cited by 117
“In contrast, digital PCR is a direct method for quantifying nucleic acids, particularly useful for samples with low abundant miRNAs, and showing a high degree of sensitivity and precision compared to qRT-PCR [17, 18].”
MethodDigital CRISPR-based method for the rapid detection and absolute quantification of nucleic acids
Wu, X., Tay, J. K., Goh, C. K., Chan, C. S. C., Lee, Y. H., Springs, S. L., Wang, D. Y., Loh, K. S., Lu, T. K., & Yu, H. (2021) · Biomaterials · doi:10.1016/j.biomaterials.2021.120876 · Cited by 120
Droplet Digital™ PCR quantitation of HER2 expression in FFPE breast cancer samples
Heredia, N. J., Belgrader, P., Wang, S., Koehler, R. T., Regan, J. F., Cosman, A. M., Saxonov, S., Hindson, B. J., Tanner, S. C., Brown, A. S., et al. (2012) · Methods · doi:10.1016/j.ymeth.2012.09.012 · Cited by 111
Sanders, R., Mason, D. J., Foy, C. A., & Huggett, J. F. (2014) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-014-7857-x · Cited by 123
“Advances in accurate molecular quantification of RNA [111, 112] could offer enhanced analytical power for this and many similar clinical challenges, and may in the future become gold standards in clinical diagnostics.”
BackgroundSelf-citationHighly Reproducible Absolute Quantification of Mycobacterium tuberculosis Complex by Digital PCR
Devonshire, A. S., Honeyborne, I., Gutteridge, A., Whale, A. S., Nixon, G., Wilson, P., Jones, G., McHugh, T. D., Foy, C. A., & Huggett, J. F. (2015) · Analytical Chemistry · doi:10.1021/ac5041617 · Cited by 103
Self-citationHeydarian, Z., Yu, M., Gruber, M. Y., Glick, B. R., Zhou, R., & Hegedus, D. D. (2016) · Frontiers in Microbiology · doi:10.3389/fmicb.2016.01966 · Cited by 105
“The number of T-DNA insertion events was determined using droplet digital PCR (Whale et al., 2012) as follows.”
BackgroundMethodComparison of droplet digital PCR and conventional quantitative PCR for measuring EGFR gene mutation
Zhang, B., Xu, C., Shao, Y., Wang, H., WU, Y.-F., Ye-ying, S., LI, X.-B., Zhang, Z., Wang, W. P., Li, L., et al. (2015) · Experimental and Therapeutic Medicine · doi:10.3892/etm.2015.2221 · Cited by 108
“ddPCR can quantify mutations in the EGFR gene in lung cancer at the single-molecule level, making it possible to detect samples with low mutation rates (12,13).”
BackgroundMejía-Benitez, M. A., Bonnefond, A., Yengo, L., Huyvaert, M., Dechaume, A., Peralta‐Romero, J., Klünder‐Klünder, M., García-Mena, J., Moustafa, J. S. E.-S., Falchi, M., et al. (2014) · Diabetologia · doi:10.1007/s00125-014-3441-3 · Cited by 103
“We and others previously demonstrated that microfluidic digital PCR was a sensitive method to measure highly polymorphic CNVs [9, 14].”
MethodDigital PCR quantification of miRNAs in sputum for diagnosis of lung cancer
Li, N., Ma, J., Guarnera, M. A., Fang, H., Cai, L., & Jiang, F. (2013) · Journal of Cancer Research and Clinical Oncology · doi:10.1007/s00432-013-1555-5 · Cited by 102
“However, expression levels of the endogenous controls and their transcripts may differ between samples (Dodd et al. 2013; Whale et al. 2012).”
Schoepp, N. G., Khorosheva, E. M., Schlappi, T. S., Curtis, M. S., Humphries, R. M., Hindler, J. A., & Ismagilov, R. F. (2016) · Angewandte Chemie International Edition · doi:10.1002/anie.201602763 · Cited by 95
Absolute quantification of lung cancer related microRNA by droplet digital PCR
Wang, P., Jing, F., Li, G., Wu, Z., Cheng, Z., Zhang, J., Zhang, H., Jia, C., Jin, Q., Mao, H., et al. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.07.048 · Cited by 97
“It should be noted that for smaller fold changes measurement by qPCR, both the precise well-optimized assays and the PCR efficiency are important (Whale et al., 2012).”
BackgroundInfluentialMicrofluidics-based strategies for molecular diagnostics of infectious diseases
Wang, X., Hong, X., Li, Y., Li, Y., Wang, J., Chen, P., & Liu, B. (2022) · Military Medical Research · doi:10.1186/s40779-022-00374-3 · Cited by 137
Košir, A. B., Divieto, C., Pavšič, J., Pavarelli, S., Dobnik, D., Dreo, T., Bellotti, R., Sassi, M. P., & Žel, J. (2017) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-017-0625-y · Cited by 96
“assumptions of dPCR which need to be met include a random distribution of DNA molecules into the partitions and amplification of a single target in a partition [15, 16].”
BackgroundBoivin, V., Deschamps-Francoeur, G., Couture, S., Nottingham, R. M., Bouchard-Bourelle, P., Lambowitz, A. M., Scott, M. S., & Abou-Elela, S. (2018) · RNA · doi:10.1261/rna.064493.117 · Cited by 95
“Other methods such as reverse transcription quantitative PCR (RT-qPCR) (Ginzinger 2002; Shakeel et al. 2017), digital PCR (dPCR) (Whale et al. 2012; Hayden et al. 2013; Witwer et al. 2013; Morley 2014; Sager et al. 2015), and in situ hybridization (e.g., FISH) (Vera et al. 2016) are also being used…”
BackgroundMethodFrom tubes to drops: droplet-based microfluidics for ultrahigh-throughput biology
Tran, T. M., Lan, F., Thompson, C. S., & Abate, A. R. (2013) · Journal of Physics D Applied Physics · doi:10.1088/0022-3727/46/11/114004 · Cited by 86
Ultrafast multiplexed detection of SARS-CoV-2 RNA using a rapid droplet digital PCR system
Yin, H., Wu, Z., Shi, N., Qi, Y., Jian, X., Zhou, L., Tong, Y., Cheng, Z., Zhao, J., & Mao, H. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113282 · Cited by 86
“In addition, quantification of target sequences is achieved via the Poisson distribution formula, which relates the number fraction of positive fluorescent droplets after amplification and the origin copy number of the target sequence (Cao et al., 2017; Majeed et al., 2012; Whale et al., 2012).”
Current status and topical issues on the use of eDNA-based targeted detection of rare animal species
Duarte, S., Simões, L. A., & Costa, F. O. (2023) · The Science of The Total Environment · doi:10.1016/j.scitotenv.2023.166675 · Cited by 88
Microfluidic systems for rapid antibiotic susceptibility tests (ASTs) at the single-cell level
Zhang, K., Qin, S., Wu, S., Liang, Y., & Li, J. (2020) · Chemical Science · doi:10.1039/d0sc01353f · Cited by 90
Kadam, S., Vuong, T. D., Qiu, D., et al. (2015) · Plant Science · doi:10.1016/j.plantsci.2015.08.015 · Cited by 87
Copy Number Change: Evolving views on Gene Amplification
Elliott, K. T., Cuff, L. E., & Neidle, E. L. (2013) · Future Microbiology · doi:10.2217/fmb.13.53 · Cited by 84
“Digital PCR is an emerging method for DNA quantification of small variations in copy number [78,79].”
MethodLiu, W.-W., Zhu, Y., Feng, Y., Jin, F., & Fang, Q. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03687 · Cited by 82
Abdel-Wahab, O., Klimek, V. M., Gaskell, A. A., Viale, A., Cheng, D., Kim, E., Rampal, R., Bluth, M., Harding, J. J., Callahan, M. K., et al. (2014) · Cancer Discovery · doi:10.1158/2159-8290.cd-13-1038 · Cited by 80
Ultrahigh-throughput sorting of microfluidic drops with flow cytometry
Lim, S. W., & Abate, A. R. (2013) · Lab on a Chip · doi:10.1039/c3lc50736j · Cited by 75
Assembled Step Emulsification Device for Multiplex Droplet Digital Polymerase Chain Reaction
Nie, M., Zheng, M., Li, C., Shen, F., Liu, M., Luo, H., Song, X., Lan, Y., Pan, J., & Du, W. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.8b04313 · Cited by 75
Alikian, M., Whale, A. S., Akiki, S. J., Piechocki, K., Torrado, C., Myint, T., Cowen, S., Griffiths, M., Reid, A. G., Apperley, J. F., et al. (2016) · Clinical Chemistry · doi:10.1373/clinchem.2016.262824 · Cited by 82
Self-citationMu, D., Liang, Y., Tang, H. R., & Liao, Y. (2015) · Biotechnology Letters · doi:10.1007/s10529-015-1890-5 · Cited by 70
“However, digital PCR (dPCR) is more precise than qPCR, and this precision is applicable at very low copy numbers, without the need for any internal control as a calibration or standard curve (Whale et al. 2012).”
BackgroundMethodStarza, I. D., Nunes, V., Cavalli, M., Novi, L. A. D., Ilari, C., Apicella, V., Vitale, A., Testi, A. M., Giudice, I. D., Chiaretti, S., et al. (2016) · British Journal of Haematology · doi:10.1111/bjh.14082 · Cited by 77
“A high degree of efficiency, sensitivity and accuracy of ddPCR compared to RQ-PCR has also been reported in molecular oncology (Diehl & Diaz, 2007; Hindson et al, 2011; Sanders et al, 2011; Whale et al, 2012).”
BackgroundMethodHuang, J., Liu, Y.-J., Wang, J., Xu, Z., Yang, Y., Shen, F., Liu, X., Zhou, X., & Liu, S. (2014) · Clinical Chemistry · doi:10.1373/clinchem.2014.230227 · Cited by 68
Whale, A. S., Devonshire, A. S., Karlin-Neumann, G. A., Regan, J. F., Javier, L., Cowen, S., Fernández-González, A., Jones, G. M., Redshaw, N. M., Beck, J., et al. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03980 · Cited by 78
Self-citationDigital PCR: Principles and Applications
Kanagal‐Shamanna, R. (2016) · Methods in molecular biology · doi:10.1007/978-1-4939-3360-0_5 · Cited by 69
GMO quantification: valuable experience and insights for the future
Milavec, M., Dobnik, D., Yang, L., Zhang, D., Gruden, K., & Žel, J. (2014) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-014-8077-0 · Cited by 70
Xu, L., Qu, H., Alonso, D. G., Yu, Z., Yu, Y., Shi, Y., Hu, C., Zhu, T., Wu, N., & Shen, F. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112908 · Cited by 65
Whale, A. S., Bushell, C. A., Grant, P. R., Cowen, S., Gutiérrez‐Aguirre, I., O’Sullivan, D. M., Žel, J., Milavec, M., Foy, C. A., Nastouli, E., et al. (2015) · Journal of Clinical Microbiology · doi:10.1128/jcm.02611-15 · Cited by 65
“The data was transformed by taking the natural log (ln) of λ and % SNP as described previously (3).”
MethodSelf-citationDetection of HBV Covalently Closed Circular DNA
Li, X., Zhao, J., Yuan, Q., & Xia, N. (2017) · Viruses · doi:10.3390/v9060139 · Cited by 70
Zhao, Y., Yu, B., Mao, X., He, J., Huang, Z., Zheng, P., Yu, J., Han, G., Liang, X., & Chen, D. (2014) · British Journal Of Nutrition · doi:10.1017/s000711451400097x · Cited by 65
Complex Copy Number Variation of AMY1 does not Associate with Obesity in two East Asian Cohorts
Yong, R. Y. Y., Mustaffa, S. B., Wasan, P. S., Sheng, L., Marshall, C. R., Scherer, S. W., Teo, Y., & Yap, P. H. (2016) · Human Mutation · doi:10.1002/humu.22996 · Cited by 57
“Two main factors influence the reliability of dPCR measurements: the number of replicates analyzed and the number of template molecules in the assay [Weaver et al., 2010; Pinheiro et al., 2012; Whale et al., 2012].”
BackgroundMethodInfluentialStabley, D. L., Harris, A. W., Holbrook, J., Chubbs, N. J., Lozo, K. W., Crawford, T. O., Swoboda, K. J., Funanage, V. L., Wang, W., Mackenzie, W. G. S., et al. (2015) · Molecular Genetics & Genomic Medicine · doi:10.1002/mgg3.141 · Cited by 63
“The greater reliability of the array dPCR assays when compared against the TaqManTM qPCR assays is a result of the random distribution of template DNA molecules within the 20,000 partitions in microfluidic dPCR array (Whale et al., 2014).”
BackgroundMethodImpact of variance components on reliability of absolute quantification using digital PCR
Jacobs, B. K., Goetghebeur, E., & Clement, L. (2014) · BMC Bioinformatics · doi:10.1186/1471-2105-15-283 · Cited by 57
“Dube, S., Qin, J., Ramakrishnan, R.: Mathematical analysis of copy number variation in a DNA sample using digital PCR on a nanofluidic device.”
BackgroundMethodInfluentialNunes, B. T. D., de Mendonça, M. H. R., Durans, D. D. B. S., Moraes, A. F., Cardoso, C. C., Prazeres, I. T. E., de Aquino, A. A., Santos, A. D. C. M., Queiroz, A. L. N., Rodrigues, D. S. G., et al. (2019) · PLoS neglected tropical diseases · doi:10.1371/journal.pntd.0007884 · Cited by 61
Lin, Y., Leng, Q., Jiang, Z., Guarnera, M. A., Zhou, Y., Chen, X., Wang, H., Zhou, W., Cai, L., Fang, H.-B., et al. (2017) · International Journal of Cancer · doi:10.1002/ijc.30822 · Cited by 57
“The number of positive reactions, together with Poisson’s distribution, were used to produce a straight and high-confidence measurement of the original target concentration.(28) Therefore, ddPCR could absolutely quantify targeted nucleic acid sequences without requiring external calibrators or endogenous controls (gen…”
BackgroundMethodLou, Y., Chen, C., Long, X., Gu, J., Xiao, M., Wang, D., Zhou, X., Li, T., Hong, Z., Li, C., et al. (2020) · Journal of Molecular Diagnostics · doi:10.1016/j.jmoldx.2020.02.007 · Cited by 61
Ultra-sensitive digital quantification of proteins and mRNA in single cells
Lin, J., Jordi, C., Son, M., Phan, H. V., Drayman, N., Abasiyanik, M. F., Vistain, L. F., Tu, H., & Tay, S. (2019) · Nature Communications · doi:10.1038/s41467-019-11531-z · Cited by 58
Fang, F., Li, Z., Yu, J., Long, Y., Zhao, Q., Ding, X., Wu, L. H., Shao, S., Zhang, L., & Xiang, W. (2021) · Journal of Advanced Research · doi:10.1016/j.jare.2021.01.003 · Cited by 64
Single cell transcriptomics: moving towards multi-omics
Song, Y., Xu, X., Wang, W., Tian, T., Zhu, Z., & Yang, C. J. (2019) · The Analyst · doi:10.1039/c8an01852a · Cited by 60
Hanson, B., Scotchman, E., Chitty, L. S., & Chandler, N. J. (2022) · Clinical Science · doi:10.1042/cs20210380 · Cited by 67
Yoo, H.-B., Park, S.-R., Dong, L., Wang, J., Sui, Z., Pavšič, J., Milavec, M., Akgöz, M., Mozioğlu, E., Corbisier, P., et al. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03076 · Cited by 53
Self-citationDevonshire, A. S., O’Sullivan, D. M., Honeyborne, I., Jones, G. M., Karczmarczyk, M., Pavšič, J., Gutteridge, A., Milavec, M., Mendoza, P., Schimmel, H. G., et al. (2016) · BMC Infectious Diseases · doi:10.1186/s12879-016-1696-7 · Cited by 54
“dPCR is more precise than qPCR [30] and offers a potentially accurate method for quantifying pathogens in reference materials when coupled with an efficient extraction protocol [24].”
MethodSelf-citationThe Emerging Role of Liquid Biopsies in Revolutionising Cancer Diagnosis and Therapy
Shegekar, T., Vodithala, S., & Juganavar, A. (2023) · Cureus · doi:10.7759/cureus.43650 · Cited by 70
“Technological advances in liquid biopsy analysis with key features, advantages, limitations, sensitivity, and specificity are summarised in Table 2 [53-59].”
BackgroundThe pioneer factor PBX1 is a novel driver of metastatic progression in ERα-positive breast cancer
Magnani, L., Patten, D. K., Nguyen, V. T. M., Hong, S.-P., Steel, J. H., Patel, N. R., Lombardo, Y., Faronato, M., Gomes, A., Woodley, L., et al. (2015) · Oncotarget · doi:10.18632/oncotarget.4243 · Cited by 55
“The tumour derived fraction of circulating free-DNA (cfDNA) originating from tumour cells can be used to estimate the genomic status of breast cancer lesions [28–30].”
BackgroundSelf-citationBharuthram, A., Paximadis, M., Picton, A. C. P., & Tiemessen, C. T. (2014) · Infection Genetics and Evolution · doi:10.1016/j.meegid.2014.03.028 · Cited by 52
Dueck, M. E., Lin, R., Zayac, A., Gallagher, S., Chao, A. K., Jiang, L., Datwani, S. S., Hung, P. J., & Stieglitz, E. (2019) · Scientific Reports · doi:10.1038/s41598-019-55872-7 · Cited by 54
Metadata
- Authors
- Whale, A. S., Huggett, J. F., Cowen, S., Speirs, V., Shaw, J., Ellison, S., Foy, C. A., & Scott, D. J.
- Affiliations
- University of Leeds (GB)
- Corresponding author
- Huggett, Jim F.
- ORCID
- Venue
- Nucleic Acids Research
- Publisher
- Oxford University Press (OUP)
- Published
- 2012-02-28
- Volume
- 40
- Issue
- 11
- Pages
- e82-e82
- ISSN
- 0305-1048 (ISSN-L)
- 1362-4962
- PMC
- PMC3367212
- OpenAlex
- W2096206246
- Cited by
- 453
- References
- 53
- Citations by year
- 2026: 33 · 2025: 32 · 2024: 20 · 2023: 36 · 2022: 29 · 2021: 36 · 2020: 33 · 2019: 34 · 2018: 24 · 2017: 32 · 2016: 48 · 2015: 31 · 2014: 30 · 2013: 27 · 2012: 6
- FWCI
- 12.07
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Nucleic Acids Research (publishedVersion, cc-by-nc)
- Europe PMC (PubMed Central) (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- Aberdeen University Research Archive (Aberdeen University) (submittedVersion, other-oa)
- Figshare (submittedVersion, cc-by)
- Leicester Research Archive (University of Leicester) (submittedVersion, cc-by-nc)
- HighWire Press Open Archive (submittedVersion)
- Funders
- Government of the United Kingdom
- Cancer Research UK (13462)
- Topics
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Molecular Biology Techniques and Applications
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Cancer Research › Cancer Genomics and Diagnostics
- Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Capillary Electrophoresis Applications
- Keywords
- digital PCR, copy number variation, real-time PCR, measurement precision, cell-free DNA, breast cancer cell lines, HER2 amplification, power analysis, template concentration, cancer diagnosis
- MeSH terms
- DNA, Neoplasm, Female, Gene Amplification, Humans, Polymerase Chain Reaction, Gene Dosage, Genes, erbB-2, Cell Line, Tumor, Microfluidic Analytical Techniques, DNA Copy Number Variations
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.