Background: A new eight-wavelength pulse oximeter is designed to measure methemoglobin and carboxyhemoglobin, in addition to the usual measurements of hemoglobin oxygen saturation and pulse rate. This study examines this device's ability to measure dyshemoglobins in human volunteers in whom controlled levels of methemoglobin and carboxyhemoglobin are induced. Methods: Ten volunteers breathed 500 ppm carbon monoxide until their carboxyhemoglobin levels reached 15%, and 10 different volunteers received intravenous sodium nitrite, 300 mg, to induce methemoglobin. All were instrumented with arterial cannulas and six Masimo Rad-57 (Masimo Inc., Irvine, CA) pulse oximeter sensors. Arterial blood was analyzed by three laboratory CO-oximeters, and the resulting carboxyhemoglobin and methemoglobin measurements were compared with the corresponding pulse oximeter readings. Results: The Rad-57 measured carboxyhemoglobin with an uncertainty of +/-2% within the range of 0-15%, and it measured methemoglobin with an uncertainty of 0.5% within the range of 0-12%. Conclusion: The Masimo Rad-57 is the first commercially available pulse oximeter that can measure methemoglobin and carboxyhemoglobin, and it therefore represents an expansion of our oxygenation monitoring capability.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Conventional pulse oximetry cannot distinguish between COHb and oxyHb, and, as such, can miss signi fi cant COHb levels and profound hypoxia (44).”
From Carbon monoxide poisoning: pathogenesis, management, and future directions of therapy (2017) · cites this paper as [44]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300 · cited 2×
“Bias (mean error) and precision (SD of error) were calculated as recommended by Altman and Bland.7 Linear regression was used to calculate the slope and intercept of the best fit straight line relation between the two methods.”
MethodEffects of Methemoglobinemia on Pulse Oximetry and Mixed Venous Oximetry
Barker, S. J., Tremper, K. K., & Hyatt, J. (1989) · Anesthesiology · doi:10.1097/00000542-198901000-00021 · Cited by 382 · cited 2×
“to drive the SpO2 readings toward 85%, no matter what the actual oxygen saturation.(3) In this animal study, metUS Department of Labor report on Sago Mine disaster.”
BackgroundMethodSelf-citationThe Effect of Carbon Monoxide Inhalation on Pulse Oximetry and Transcutaneous Po 2
Barker, S. J., & Tremper, K. K. (1987) · Anesthesiology · doi:10.1097/00000542-198705000-00014 · Cited by 281
Self-citationEffects of Intravenously Administered Dyes on Pulse Oximetry Readings
SCHELLER, M. S., UNGER, R. J., & KELNER, M. J. (1986) · Anesthesiology · doi:10.1097/00000542-198611000-00023 · Cited by 194
History of blood gas analysis. VII. Pulse oximetry
Severinghaus, J. W., & Honda, Y. (1987) · Journal of Clinical Monitoring · doi:10.1007/bf00858362 · Cited by 179 · cited 1×
“Since its invention by Aoyagi in 1974, the pulse oximeter has become the most common monitor of patient oxygenation.(8) However, all pulse oximeters in clinical use to date share a serious limitation.”
BackgroundThe antidotal action of sodium nitrite and sodium thiosulfate against cyanide poisoning
Baskin, S. I., Horowitz, A. M., & Nealley, E. W. (1992) · The Journal of Clinical Pharmacology · doi:10.1002/j.1552-4604.1992.tb03849.x · Cited by 165
Pulse Oximetry in Severe Carbon Monoxide Poisoning
Hampson, N. B. (1998) · CHEST Journal · doi:10.1378/chest.114.4.1036 · Cited by 145 · cited 1×
“2,3 These errors have been confirmed in clinical case reports.(4,5) No pre-”
BackgroundPulse Oximetry in Methemoglobinemia
Watcha, M. F., Connor, M. T., & Hing, A. V. (1989) · Archives of Pediatrics and Adolescent Medicine · doi:10.1001/archpedi.1989.02150190095030 · Cited by 69 · cited 1×
“2,3 These errors have been confirmed in clinical case reports.(4,5) No pre-”
BackgroundSevere Carbon Monoxide Poisoning during Desflurane Anesthesia
Berry, P. D., Sessler, D. I., & Larson, M. D. (1999) · Anesthesiology · doi:10.1097/00000542-199902000-00036 · Cited by 65 · cited 1×
“Louis, MO) has also caused serious toxicity.(11)”
Benzocaine-Induced Methemoglobinemia in an Adult: Accuracy of Pulse Oximetry with Methemoglobinemia
Anderson, S. T., Hajduczek, J., & Barker, S. J. (1988) · Anesthesia & Analgesia · doi:10.1213/00000539-198811000-00015 · Cited by 64
Self-citation
Cited by
Carbon monoxide poisoning: pathogenesis, management, and future directions of therapy
Rose, J. J., Wang, L., Xu, Q., McTiernan, C. F., Shiva, S., Tejero, J., & Gladwin, M. T. (2017) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.201606-1275ci · Cited by 965
Jubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 491
Pulse oximetry: fundamentals and technology update
Nitzan, M., Romem, A., & Koppel, R. (2014) · Medical Devices: Evidence and Research · doi:10.2147/mder.s47319 · Cited by 438
“Some studies showed accurate measurement of COHb and MetHb, 22,40–42 while others 43,44 claim that pulse co-oximetry cannot replace standard blood COHb measurement, though it could be used as a first-line screening test.”
Methemoglobinemia Related to Local Anesthetics: A Summary of 242 Episodes
Guay, J. M. Y. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e318187c4b1 · Cited by 357
Practice recommendations in the diagnosis, management, and prevention of carbon monoxide poisoning
Hampson, N. B., Piantadosi, C. A., Thom, S. R., & Weaver, L. K. (2012) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.201207-1284ci · Cited by 331
Metadata
- Authors
- Barker, S. J., Curry, J., Redford, D., & Morgan, S.
- Venue
- Anesthesiology
- Published
- 2006-11-01
- Volume
- 105
- Issue
- 5
- Pages
- 892-897
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2123754678
- Cited by
- 238
- References
- 16
- FWCI
- 9.75
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.