Clinical Experience with 118 Brain Tissue Oxygen Partial Pressure Catheter Probes

Dings, Jim, Meixensberger, Jürgen, Jäger, Andreas, Roosen, Klaus·1998·Neurosurgery

Abstract

OBJECTIVE: We assessed the technical and diagnostic reliability of partial pressure of oxygen (PO2) of brain tissue (P(ti)O2) monitoring. The monitoring system and the catheter probes were tested in vitro, and clinical experiences obtained with 118 brain P(ti)O2 catheter probes, used in 101 patients, are reported. METHODS: The polarographic (LICOX; Medical Systems Corp., Greenvale, NY) P(ti)O2 catheter probe lies 22 to 27 mm below the dura level; its PO2-sensitive surface is 7.1 mm2. For 10 patients, the adaptation time (with initially unreliable signals after insertion) was determined. For 27 patients, the probe was removed in a stepwise fashion (three increments of 5 mm) and the heterogeneity of brain P(ti)O2 levels was investigated. After removal of the catheter probes, their PO2 and zero display error values were determined and compared with probe performance data obtained in vitro with unused PO2 catheter probes. RESULTS: Small iatrogenic hematomas were observed for two patients (1.7%). No infection occurred after 6.7 +/- 3.9 days (mean +/- standard deviation) of monitoring. The technical complication (dislocation or defect) rate was 13.6%. The mean adaptation time was 79.0 +/- 51.7 min. A flow chart is presented, which helps to rule out artifacts. The mean P(ti)O2 measured at 22 to 27 mm below the dura was 23.8 +/- 8.1 mm Hg, at 17 to 22 mm was 25.7 +/- 8.3 mm Hg, at 12 to 17 mm was 33.0 +/- 13.3 mm Hg (P < 0.01, compared with the initial value), and at 7 to 12 mm was 33.3 +/- 13.3 mm Hg (P < 0.01). Recent catheter probe versions exhibited a PO2 display error of -1.2 +/- 5.1% (mean +/- standard deviation, n = 38) and a mean zero display error of 1.1 +/- 0.9 mm Hg (n = 34). The greatest PO2 display errors were measured during the first 4 days of continuous monitoring. In the in vitro test (of 12 unused catheter probes), the maximal probe display error was 1.07 +/- 2.14%, tested at temperatures between 22 degrees C and 37 degrees C and tested at oxygen pressures of 0, 44, and 150 mm Hg. In vitro, the zero display error was -0.21 +/- 0.25 mm Hg. CONCLUSION: Brain P(ti)O2 monitoring, reflecting an area 17 to 27 mm below the dura, is a safe and reliable technique for monitoring cerebral oxygenation. Excluding the first 1 hour after insertion, data are reliable, with almost 100% good data quality.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00006123-199811000-00045

Cites

28 of 28 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 28 references

Cited by

See all 323

Metadata

Authors
Dings, J., Meixensberger, J., Jäger, A., & Roosen, K.
Venue
Neurosurgery
Published
1998-11
Volume
43
Issue
5
Pages
1082-1094
ISSN
  • 0148-396X (ISSN-L)
  • 1524-4040
OpenAlex
W2077067643
Cited by
323
References
28
FWCI
4.04
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Clinical Experience with 118 Brain Tissue Oxygen Partial Pressure Catheter Probes. https://app.cochat.ai/papers/10.1097/00006123-199811000-00045
The paper
Dings, J., Meixensberger, J., Jäger, A., & Roosen, K. (1998). Clinical Experience with 118 Brain Tissue Oxygen Partial Pressure Catheter Probes. Neurosurgery. https://doi.org/10.1097/00006123-199811000-00045

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.