A new method for pulse oximetry possessing inherent insensitivity to artifact
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
A new method for pulse oximetry is presented that possesses an inherent insensitivity to corruption by motion artifact, a primary limitation in the practical accuracy and clinical applicability of current technology. Artifact corruption of the underlying photoplethysmographic signals is reduced in real time, using an electronic processing methodology that is based upon inversion of a physical artifact model. This fundamental approach has the potential to provide uninterrupted output and superior accuracy under conditions of sustained subject motion, therefore, widening the clinical scope of this useful measurement. A new calibration technique for oxygen saturation is developed for use with these processed signals, which is shown to be a generalization of the classical interpretation. The detailed theoretical and practical issues of implementation are then explored, highlighting important engineering simplifications implicit in this new approach. A quantitative investigation of the degree of insensitivity to artifact is also undertaken, with the aid of a custom electronic system and commercial pulse oximeter probes, which is compared and contrasted with the performance of a conventional implementation. It is demonstrated that this new methodology results in a reduced sensitivity to common classes of motion artifact, while retaining the generality to be combined with conventional signal processing techniques.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Gaskin, L., & Thomas, J. (1995) · Physiotherapy · doi:10.1016/s0031-9406(05)66819-8 · Cited by 14 · cited 1×
“This implies that the oxygen saturation output of a commercial pulse oximeter would tend toward a calibration depend constant under similar conditions, and may in part explain why some authors report erroneously high oxygen saturation readings during periods of sustained subject motion, while others report readings th…”
Result
Cited by
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
“This includes the Masimo SET technology (Goldman et al 2000, Hayes and Smith 2001).”
MethodSmart wearable systems: Current status and future challenges
Chan, M., Estève, D., Fourniols, J.-Y., Escriba, C., & Campo, E. (2012) · Artificial Intelligence in Medicine · doi:10.1016/j.artmed.2012.09.003 · Cited by 953
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935
“But, in general, the more convenient the waveform measurement is, the more noise contamination can be expected [51]– [62].”
BackgroundMethodInfluentialWearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912
“The scope and power of this methodology was investigated based on the results obtained with a practical electronic system [77,78].”
MethodZhang, Z., Pi, Z., & Liu, B. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2359372 · Cited by 748
How accurate is pulse rate variability as an estimate of heart rate variability?
Schäfer, A., & Vagedes, J. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.03.119 · Cited by 661
Gil, E., Orini, M., Bailón, R., Vergara, J. M., Mainardi, L., & Laguna, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/9/015 · Cited by 461
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B. S., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 399
A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
Ram, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., & Reddy, K. A. (2012) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2011.2175832 · Cited by 345
Wearable Medical Systems for p-Health
Teng, X.-F., Zhang, Y.-T., Poon, C. C. Y., & Bonato, P. (2008) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2008.2008248 · Cited by 310
“This observation has been driving researchers to develop more realistic models for the PPG or the artifact [115].”
MethodDynamic time warping and machine learning for signal quality assessment of pulsatile signals
Li, Q., & Clifford, G. D. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/9/1491 · Cited by 274
Current progress of photoplethysmography and SPO2 for health monitoring
Tamura, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 247
Signal quality measures for pulse oximetry through waveform morphology analysis
Sukor, J. A., Redmond, S. J., & Lovell, N. H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/3/008 · Cited by 166
Reducing motion artifacts in photoplethysmograms by using relative sensor motion: Phantom study
Wijshoff, R. W. C. G. R., Mischi, M., Veen, J., & Van Der Lee, A. M. (2012) · Journal of Biomedical Optics · doi:10.1117/1.jbo.17.11.117007 · Cited by 40
“Motion artifacts in PPGs are presumably caused by hemodynamic effects, tissue deformation, and sensor movement and deformation.(1,2,4,24) It is hypothesized that a large part of these motion artifacts correlate with relative sensor motion.”
Metadata
- Authors
- Hayes, M. J., & Smith, P. R.
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2001-04
- Volume
- 48
- Issue
- 4
- Pages
- 452-461
- ISSN
- 0018-9294 (ISSN-L)
- OpenAlex
- W2124731769
- Cited by
- 122
- References
- 1
- FWCI
- 1.54
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.