CoChat

Tissue response to chronically stimulated implanted epimysial and intramuscular electrodes

Akers, J.M., Peckham, P.H., Keith, M.W., Merritt, K. · 1997 · IEEE Transactions on Rehabilitation Engineering
Cited by 77

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

Twenty-four epimysial and 16 intramuscular electrodes were implanted in five adult dogs for periods ranging from 11 to 50 months. Chronic stimulation was applied to half of the electrodes for eight weeks near the end of the implantation period. The tissue response was rated by the amount and appearance of the fibrous tissue and inflammatory cells seen in the capsule lining the region of the electrode. The encapsulation tissues were composed primarily of collagen and fibroblasts and some macrophages and few other inflammatory cells. The epimysial electrodes exhibited more variation between and within electrodes, but had more of the better scores than the intramuscular electrodes. No difference in the distribution of scores was measured between the control and stimulated groups for the epimysial electrodes. While the scores for the intra-muscular electrodes varied very little, variance was sufficient to indicate a trend for poorer ratings with the application of chronic stimulation. Fibrous capsules were generally thinner under the epimysial electrodes than around the intramuscular electrodes. For both electrode types, the thickness was not correlated with the application or level of chronic stimulation. Thickness was shown to be positively correlated to the degree of loss of the sutures used to anchor the epimysial electrodes.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1109/86.593301

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 3 references

Cited by

See all 77

Metadata

Authors
Akers, J. M., Peckham, P. H., Keith, M. W., & Merritt, K.
Venue
IEEE Transactions on Rehabilitation Engineering
Published
1997-06
Volume
5
Issue
2
Pages
207-220
ISSN
  • 1063-6528 (ISSN-L)
  • 1558-0024
OpenAlex
W2163310646
Cited by
77
References
3
FWCI
0.89
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Tissue response to chronically stimulated implanted epimysial and intramuscular electrodes. https://app.cochat.ai/papers/10.1109/86.593301
The paper
Akers, J. M., Peckham, P. H., Keith, M. W., & Merritt, K. (1997). Tissue response to chronically stimulated implanted epimysial and intramuscular electrodes. IEEE Transactions on Rehabilitation Engineering. https://doi.org/10.1109/86.593301

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.