Stimulation Stability and Selectivity of Chronically Implanted Multicontact Nerve Cuff Electrodes in the Human Upper Extremity
Open accessAbstract
Nine spiral nerve cuff electrodes were implanted in two human subjects for up to three years with no adverse functional effects. The objective of this study was to look at the long term nerve and muscle response to stimulation through nerve cuff electrodes. The nerve conduction velocity remained within the clinically accepted range for the entire testing period. The stimulation thresholds stabilized after approximately 20 weeks. The variability in the activation over time was not different from muscle-based electrodes used in implanted functional electrical stimulation systems. Three electrodes had multiple, independent contacts to evaluate selective recruitment of muscles. A single muscle could be selectively activated from each electrode using single-contact stimulation and the selectivity was increased with the use of field steering techniques. The selectivity after three years was consistent with selectivity measured during the implant surgery. Nerve cuff electrodes are effective for chronic muscle activation and multichannel functional electrical stimulation in humans.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Helical and spiral extra-neural interfaces have been stable for decades in clinical applications.
“The helical (Agnew et al., 1989) and spiral (Naples et al., 1988) type interfaces have been proven to be stable for decades in clinical applications (Fisher et al., 2009; Polasek et al., 2009).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Polasek et al., 2009
The Discussion reports stable recruitment from spiral nerve cuff electrodes in two human subjects for up to three years; the Introduction also describes a spiral cuff implanted in one person for over six years. These passages support clinical stability of spiral interfaces, but not stability for decades. They do not establish clinical stability of helical interfaces. (LLM, conf=0.90)
Experimental — accuracy not yet measuredElectrical stimulation in the form of spikes has been given to the residual stump nerve of a patient through typically implanted electrodes, and psychophysical judgments and verbal reports were gathered.
“Electrical stimulation in the form of spikes has been given to the residual stump nerve of a patient through typically implanted electrodes, and psychophysical judgments and verbal reports were gathered (Polasek et al., 2009; Schiefer et al., 2016).”
From Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review. (2017) · cites this paper as Polasek et al., 2009
Methods A and D describe implanted nerve-cuff electrodes delivering electrical pulses. But Methods C identifies the subjects as two people with tetraplegia, not a patient with a residual stump nerve; Methods D describes EMG measurements, not psychophysical judgments or verbal reports. No page numbers are available in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measuredIn a 3-year clinical study with two subjects, spiral cuff electrodes provided stable and selective stimulation without serious adverse reactions, after achieving stable stimulation thresholds around the 20th week.
“In a 3-year clinical study with two subjects, spiral cuff electrodes were shown to provide stable and selective stimulation without serious adverse reactions, after achieving stable stimulation thresholds around the 20th week [72].”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [72]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Zagrean, L. (2014) · Acta Endocrinologica (Bucharest) · doi:10.4183/aeb.2014.529 · Cited by 1,926 · see citing works
The adaptive response of skeletal muscle to increased use
Salmons, S., & Henriksson, J. (1981) · Muscle & Nerve · doi:10.1002/mus.880040204 · Cited by 765 · see citing works
Katz, B., & Miledi, R. (1965) · Proceedings of the Royal Society B Biological Sciences · doi:10.1098/rspb.1965.0016 · Cited by 611 · see citing works
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415 · see citing works
Veraart, C., Raftopoulos, C., Mortimer, J. T., Delbeke, J., Pins, D., Michaux, G., Vanlierde, A., Parrini, S., & Wanet-Defalque, M.-C. (1998) · Brain Research · doi:10.1016/s0006-8993(98)00977-9 · Cited by 407 · see citing works
Functionally selective peripheral nerve stimulation with a flat interface nerve electrode
Tyler, D. J., & Durand, D. M. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806840 · Cited by 398 · cited 3× · see citing works
“electrodes with up to 12 stimulation positions could selectively activate all four functions without using field steering [13].”
Self-citationSelective control of muscle activation with a multipolar nerve cuff electrode
Veraart, C., Grill, W. M., & Mortimer, J. T. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237694 · Cited by 320 · see citing works
The Effect of Stimulus Parameters on the Recruitment Characteristics of Direct Nerve Stimulation
Gorman, P. H., & Mortimer, J. T. (1983) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1983.325041 · Cited by 313 · see citing works
Smith, B., Tang, Z., Johnson, M. W., Pourmehdi, S., Gazdik, M. M., Buckett, J. R., & Peckham, P. H. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.664202 · Cited by 270 · see citing works
Pattern recognition with the optic nerve visual prosthesis
Veraart, C., Wanet‐Defalque, M., Gérard, B., Vanlierde, A., & Delbeke, J. (2003) · Artificial Organs · doi:10.1046/j.1525-1594.2003.07305.x · Cited by 259 · see citing works
Procházka, A., Gauthier, M., Wieler, M., & Kenwell, Z. (1997) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(97)90426-3 · Cited by 250 · see citing works
A nerve cuff technique for selective excitation of peripheral nerve trunk regions
Sweeney, J. D., Ksienski, D. A., & Mortimer, J. T. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.55681 · Cited by 196 · see citing works
An Implanted Upper-Extremity Neuroprosthesis. Follow-up of Five Patients*
KILGORE, K. L., PECKHAM, P. H., KEITH, M. W., THROPE, G. B., WUOLLE, K. S., BRYDEN, A. M., & HART, R. L. (1997) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199704000-00008 · Cited by 192 · see citing works
Self-citationPeckham, P. H., MORTIMER, J. T., & Marsolais, E. B. (1976) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-197601000-00041 · Cited by 164 · see citing works
Methods for estimating isometric recruitment curves of electrically stimulated muscle
Durfee, W. K., & MacLean, K. E. (1989) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.32097 · Cited by 155 · see citing works
Position, size and luminosity of phosphenes generated by direct optic nerve stimulation
Delbeke, J., Oozeer, M., & Veraart, C. (2003) · Vision Research · doi:10.1016/s0042-6989(03)00013-0 · Cited by 138 · see citing works
Quantification of recruitment properties of multiple contact cuff electrodes
Grill, W. M., & Mortimer, J. T. (1996) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.506402 · Cited by 136 · see citing works
Recruitment properties of monopolar and bipolar epimysial electrodes
Grandjean, P. A., & Mortimer, J. T. (1986) · Annals of Biomedical Engineering · doi:10.1007/bf02364648 · Cited by 134 · see citing works
Creating a meaningful visual perception in blind volunteers by optic nerve stimulation
Brelen, M. E., Duret, F., Gérard, B., Delbeke, J., & Veraart, C. (2005) · Journal of Neural Engineering · doi:10.1088/1741-2560/2/1/004 · Cited by 115 · see citing works
Human nerve stimulation thresholds and selectivity using a multi-contact nerve cuff electrode
Polasek, K. H., Hoyen, H. A., Keith, M. W., & Tyler, D. J. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.891383 · Cited by 111 · see citing works
Self-citationA surgically-implanted intramuscular electrode for an implantable neuromuscular stimulation system
Memberg, W. D., Peckham, P. H., & Keith, M. W. (1994) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.313149 · Cited by 111 · see citing works
Self-citationTarler, M. D., & Mortimer, J. T. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.828415 · Cited by 110 · see citing works
Grill, W. M., & Mortimer, J. T. (1998) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.736150 · Cited by 103 · see citing works
A musculoskeletal model of the upper extremity for use in the development of neuroprosthetic systems
Blana, D., Hincapie, J. G., Chadwick, E. K., & Kirsch, R. F. (2008) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2008.03.001 · Cited by 99 · see citing works
Self-citationDavis, J. A., Triolo, R. J., Uhlir, J. P., Bieri, C., Rohde, L., Lissy, D. A., & Kukke, S. N. (2001) · PubMed · Cited by 99 · see citing works
Combined feedforward and feedback control of a redundant, nonlinear, dynamic musculoskeletal system
Blana, D., Kirsch, R. F., & Chadwick, E. K. (2009) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-009-0479-3 · Cited by 90 · see citing works
Self-citationTissue response to chronically stimulated implanted epimysial and intramuscular electrodes
Akers, J. M., Peckham, P. H., Keith, M. W., & Merritt, K. (1997) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.593301 · Cited by 77 · see citing works
Self-citationMotor and sensory conduction in different segments of the radial nerve in normal subjects.
Trojaborg, W., & Sindrup, E. (1969) · Journal of Neurology Neurosurgery & Psychiatry · doi:10.1136/jnnp.32.4.354 · Cited by 70 · see citing works
Neuromuscular stimulation selectivity of multiple-contact nerve cuff electrode arrays
Sweeney, J. D., Crawford, N. R., & Brandon, T. A. (1995) · Medical & Biological Engineering & Computing · doi:10.1007/bf02510525 · Cited by 51 · see citing works
Position-selective activation of peripheral nerve fibers with a cuff electrode
Goodall, E. V., de Breij, J. F., & Holsheimer, J. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.508548 · Cited by 47 · see citing works
Platia, E. V., & Brinker, J. (1986) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1986.tb05408.x · Cited by 38 · see citing works
Measurement of external pressures generated by nerve cuff electrodes
Cuoco, F. A., & Durand, D. M. (2000) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.830947 · Cited by 36 · see citing works
Hincapie, J. G., Blana, D., Chadwick, E. K., & Kirsch, R. F. (2008) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2008.922681 · Cited by 30 · see citing works
Self-citationJeong (2000) · doi:10.1002/(sici)1097-4636(200005)50:2
Flagged first, then by citations
Cited by
Sorted by citations
Literature Review on Needs of Upper Limb Prosthesis Users
Cordella, F., Ciancio, A. L., Sacchetti, R., Davalli, A., Cutti, A. G., Guglielmelli, E., & Zollo, L. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00209 · Cited by 766 · see citing works
A neural interface provides long-term stable natural touch perception
Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008669 · Cited by 751 · see citing works
Self-citationOrtiz-Catalan, M., Håkansson, B., & Brånemark, R. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008933 · Cited by 484 · see citing works
Control of Hand Prostheses Using Peripheral Information
Micera, S., Carpaneto, J., & Raspopović, S. (2010) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2010.2085429 · Cited by 405 · see citing works
Sensory feedback for limb prostheses in amputees
Raspopovic, S., Valle, G., & Petrini, F. M. (2021) · Nature Materials · doi:10.1038/s41563-021-00966-9 · Cited by 308 · see citing works
The neural basis of perceived intensity in natural and artificial touch
Graczyk, E. L., Schiefer, M. A., Saal, H. P., Delhaye, B. P., Bensmaia, S. J., & Tyler, D. J. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf5187 · Cited by 258 · see citing works
Self-citationRestoration of sensory information via bionic hands
Bensmaı̈a, S. J., Tyler, D. J., & Micera, S. (2020) · Nature Biomedical Engineering · doi:10.1038/s41551-020-00630-8 · Cited by 256 · see citing works
Self-citationSchultz, A. E., & Kuiken, T. A. (2011) · PM&R · doi:10.1016/j.pmrj.2010.06.016 · Cited by 186 · see citing works
Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., & Tyler, D. J. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/2/026002 · Cited by 175 · see citing works
Self-citationLong-term usability and bio-integration of polyimide-based intra-neural stimulating electrodes
Wurth, S., Capogrosso, M., Raspopovic, S., Gandar, J., Federici, G., Kinany, N., Cutrone, A., Piersigilli, A., Pavlova, N., Guiet, R., et al. (2017) · Biomaterials · doi:10.1016/j.biomaterials.2017.01.014 · Cited by 176 · see citing works
Neural stimulation with optical radiation
Richter, C., Matic, A. I., Wells, J. D., Jansen, D., & Walsh, J. T. (2010) · Laser & Photonics Review · doi:10.1002/lpor.200900044 · Cited by 146 · see citing works
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · see citing works
Implanted Neuroprosthesis for Restoring Arm and Hand Function in People With High Level Tetraplegia
Memberg, W. D., Polasek, K. H., Hart, R. L., Bryden, A. M., Kilgore, K. L., Nemunaitis, G. A., Hoyen, H. A., Keith, M. W., & Kirsch, R. F. (2014) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2014.01.028 · Cited by 137 · see citing works
Self-citationCharkhkar, H., Shell, C. E., Marasco, P. D., Pinault, G. J., Tyler, D. J., & Triolo, R. J. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aac964 · Cited by 131 · see citing works
Self-citationControl of prosthetic hands via the peripheral nervous system
Ciancio, A. L., Cordella, F., Barone, R., Romeo, R. A., Bellingegni, A. D., Sacchetti, R., Davalli, A., Di Pino, G., Ranieri, F., Di Lazzaro, V., et al. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00116 · Cited by 131 · see citing works
Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113 · see citing works
Raspopović, S., Capogrosso, M., & Micera, S. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2151878 · Cited by 113 · see citing works
Neural interfaces for somatosensory feedback: bringing life to a prosthesis
Tyler, D. J. (2015) · Current Opinion in Neurology · doi:10.1097/wco.0000000000000266 · Cited by 105 · see citing works
Self-citationHigh-Frequency Electrical Nerve Block for Postamputation Pain: A Pilot Study
Soin, A., Shah, N. S., & Fang, Z.-P. (2015) · Neuromodulation Technology at the Neural Interface · doi:10.1111/ner.12266 · Cited by 98 · see citing works
Thin film multi-electrode softening cuffs for selective neuromodulation
González-González, M. A., Kanneganti, A., Joshi-Imre, A., Hernandez-Reynoso, A. G., Bendale, G., Modi, R., Ecker, M., Khurram, A., Cogan, S. F., Voit, W. E., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-34566-6 · Cited by 97 · see citing works
“Long-term stability of stimulating spiral nerve cuff electrodes on human peripheral nerves”
Christie, B. P., Freeberg, M., Memberg, W. D., Pinault, G. J. C., Hoyen, H. A., Tyler, D. J., & Triolo, R. J. (2017) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-017-0285-3 · Cited by 97 · see citing works
Self-citationSelectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review.
Ghafoor, U., Kim, S., & Hong, K.-S. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00059 · Cited by 88 · see citing works
“The helical (Agnew et al., 1989) and spiral (Naples et al., 1988) type interfaces have been proven to be stable for decades in clinical applications (Fisher et al., 2009; Polasek et al., 2009).”
Raspopović, S., Petrini, F. M., Żelechowski, M., & Valle, G. (2016) · Proceedings of the IEEE · doi:10.1109/jproc.2016.2600560 · Cited by 84 · see citing works
Quantification of human upper extremity nerves and fascicular anatomy
Brill, N. A., & Tyler, D. J. (2017) · Muscle & Nerve · doi:10.1002/mus.25534 · Cited by 64 · see citing works
Self-citationMotor-commands decoding using peripheral nerve signals: a review
Hong, K.-S., Aziz, N., & Ghafoor, U. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aab383 · Cited by 63 · see citing works
Raspopovic, S., Capogrosso, M., Badia, J., Navarro, X., & Micera, S. (2012) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2012.2189021 · Cited by 63 · see citing works
Schiefer, M. A., Freeberg, M., Pinault, G. J. C., Anderson, J., Hoyen, H., Tyler, D. J., & Triolo, R. J. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/5/056006 · Cited by 57 · see citing works
Self-citationA Parylene Self-Locking Cuff Electrode for Peripheral Nerve Stimulation and Recording
Yu, H., Xiong, W., Zhang, H., Wang, W., & Li, Z. (2014) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2014.2333733 · Cited by 56 · see citing works
The design of and chronic tissue response to a composite nerve electrode with patterned stiffness
Freeberg, M. J., Stone, M. A., Triolo, R. J., & Tyler, D. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6632 · Cited by 54 · see citing works
Self-citationMotor neuron activation in peripheral nerves using infrared neural stimulation
Peterson, E. J., & Tyler, D. J. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/1/016001 · Cited by 54 · see citing works
Self-citationGreenwald, E., Masters, M. R., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1429-x · Cited by 53 · see citing works
Recording and decoding for neural prostheses
Warren, D. J., Kellis, S., Nieveen, J. G., Wendelken, S. M., Dantas, H., Davis, T. S., Hutchinson, D. T., Normann, R. A., Clark, G. A., & Mathews, V. J. (2016) · Proceedings of the IEEE · doi:10.1109/jproc.2015.2507180 · Cited by 52 · see citing works
Infrared neural stimulation: Beam path in the guinea pig cochlea
Moreno, L. E., Rajguru, S. M., Matic, A. I., Yerram, N. K., Robinson, A. M., Hwang, M., Stock, S. R., & Richter, C. (2011) · Hearing Research · doi:10.1016/j.heares.2011.06.006 · Cited by 52 · see citing works
Kent, A. R., & Grill, W. M. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/3/036010 · Cited by 52 · see citing works
Herring, E. Z., Graczyk, E. L., Memberg, W. D., Adams, R., Baca-Vaca, G. F., Hutchison, B. C., Krall, J. T., Alexander, B. J., Conlan, E. C., Alfaro, K. E., et al. (2023) · Neurosurgery · doi:10.1227/neu.0000000000002769 · Cited by 42 · see citing works
Self-citationMulti-Muscle FES Force Control of the Human Arm for Arbitrary Goals
Schearer, E. M., Liao, Y.-W., Perreault, E. J., Tresch, M. C., Memberg, W. D., Kirsch, R. F., & Lynch, K. (2013) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2013.2282903 · Cited by 37 · see citing works
Self-citationMakowski, N. S., Campean, A., Lambrecht, J. M., Buckett, J. R., Coburn, J., Hart, R. L., Miller, M. E., Montague, F. W., Crish, T., Fu, M. J., et al. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3066838 · Cited by 34 · see citing works
Metcalfe, B., Chew, D. J., Clarke, C. T., Donaldson, N., & Taylor, J. T. (2015) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2015.05.003 · Cited by 34 · see citing works
Intraoperative evaluation of the spiral nerve cuff electrode on the femoral nerve trunk
Polasek, K. H., Schiefer, M. A., Pinault, G. C. J., Triolo, R. J., & Tyler, D. J. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/6/066005 · Cited by 33 · see citing works
Self-citationRecent advances in bioelectric prostheses
Pasquina, P. F., Perry, B. N., Miller, M. E., Ling, G. S. F., & Tsao, J. W. (2015) · Neurology Clinical Practice · doi:10.1212/cpj.0000000000000132 · Cited by 33 · see citing works
Kashkoush, A. I., Gaunt, R. A., Fisher, L. E., Bruns, T. M., & Weber, D. J. (2019) · Scientific Reports · doi:10.1038/s41598-019-38924-w · Cited by 33 · see citing works
Tigra, W., Dali, M., William, L., Fattal, C., Gélis, A., Divoux, J., Coulet, B., Teissier, J., Guiraud, D., & Coste, C. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00676-4 · Cited by 32 · see citing works
Birdwell, J. A., Hargrove, L. J., Weir, R. F. F., & Kuiken, T. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2344854 · Cited by 30 · see citing works
A soft, scalable and adaptable multi-contact cuff electrode for targeted peripheral nerve modulation
Paggi, V., Fallegger, F., Serex, L., Rizzo, O., Galan, K., Giannotti, A., Furfaro, I., Zinno, C., Bernini, F., Micera, S., et al. (2024) · Bioelectronic Medicine · doi:10.1186/s42234-023-00137-y · Cited by 30 · see citing works
Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, R., & Tyler, D. J. (2013) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2013.6696070 · Cited by 26 · see citing works
Self-citationSchearer, E. M., Liao, Y.-W., Perreault, E. J., Tresch, M. C., Memberg, W. D., Kirsch, R. F., & Lynch, K. (2016) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2016.2535348 · Cited by 25 · see citing works
Self-citationRodrigues, F., Ribeiro, J. F., Anacleto, P., Fouchard, A., Olivier David, A., Sarro, P. M., & Mendes, P. M. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab4dbb · Cited by 25 · see citing works
Greenwald, E., Masters, M. R., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-016-1452-6 · Cited by 23 · see citing works
Activation of individual extrinsic thumb muscles and compartments of extrinsic finger muscles
Birdwell, J. A., Hargrove, L. J., Kuiken, T., & Weir, R. F. F. (2013) · Journal of Neurophysiology · doi:10.1152/jn.00748.2012 · Cited by 22 · see citing works
Holding Static Arm Configurations With Functional Electrical Stimulation: A Case Study
Wolf, D. N., & Schearer, E. M. (2018) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2018.2866226 · Cited by 22 · see citing works
Chronic nerve health following implantation of femoral nerve cuff electrodes
Freeberg, M. J., Pinault, G. C. J., Tyler, D. J., Triolo, R. J., & Ansari, R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00720-3 · Cited by 22 · see citing works
Self-citationCoste, C., William, L., da Fonseca, L. O., Hiairrassary, A., Andreu, D., Geffrier, A., Teissier, J., Fattal, C., & Guiraud, D. (2022) · Scientific Reports · doi:10.1038/s41598-022-19906-x · Cited by 22 · see citing works
Delianides, C., Tyler, D., Pinault, G., Ansari, R., & Triolo, R. (2020) · Neuromodulation Technology at the Neural Interface · doi:10.1111/ner.13110 · Cited by 20 · see citing works
Self-citationA Multiscale Approach to Axon and Nerve Stimulation Modeling: A Review
Stefano, M., Cordella, F., Loppini, A., Filippi, S., & Zollo, L. (2021) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2021.3054551 · Cited by 20 · see citing works
Ordelman, S. C. M. A., Kornet, L., Cornelussen, R. N., Buschman, H. P. J., & Veltink, P. H. (2012) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2012.2214058 · Cited by 20 · see citing works
Standard task set for evaluating rehabilitation interventions for individuals with arm paralysis
Cornwell, A. S., Liao, J., Bryden, A. M., & Kirsch, R. F. (2012) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2011.03.0040 · Cited by 20 · see citing works
Self-citationCharkhkar, H., Christie, B. P., Pinault, G. J., Tyler, D. J., & Triolo, R. J. (2019) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108414 · Cited by 19 · see citing works
Self-citationSelective stimulation of the ferret abdominal vagus nerve with multi-contact nerve cuff electrodes
Shulgach, J. A., Beam, D. W., Nanivadekar, A. C., Miller, D. M., Fulton, S. A., Sciullo, M., Ogren, J. I., Wong, L., McLaughlin, B. L., Yates, B. J., et al. (2021) · Scientific Reports · doi:10.1038/s41598-021-91900-1 · Cited by 17 · see citing works
Evaluation of a conducting elastomeric composite material for intramuscular electrode application
Zheng, X. S., Griffith, A. Y., Chang, E., Looker, M. J., Fisher, L. E., Clapsaddle, B. J., & Cui, X. T. (2019) · Acta Biomaterialia · doi:10.1016/j.actbio.2019.12.021 · Cited by 18 · see citing works
Model Learning for Control of a Paralyzed Human Arm with Functional Electrical Stimulation
Wolf, D. N., Hall, Z. A., & Schearer, E. M. (2020) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra40945.2020.9196992 · Cited by 16 · see citing works
Clinical Applications and Future Translation of Somatosensory Neuroprostheses
Graczyk, E. L., Hutchison, B. C., Valle, G., Bjånes, D. A., Gates, D. H., Raspopović, S., & Gaunt, R. A. (2024) · Journal of Neuroscience · doi:10.1523/jneurosci.1237-24.2024 · Cited by 17 · see citing works
Electrochemical modulation enhances the selectivity of peripheral neurostimulation in vivo
Flavin, M. T., Paul, M. A., Lim, A. S., Lissandrello, C. A., Ajemian, R., Lin, S. J., & Han, J. (2022) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2117764119 · Cited by 16 · see citing works
Wolf, D. N., & Schearer, E. M. (2017) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2017.8009344 · Cited by 13 · see citing works
Temporal Modulation of the Response of Sensory Fibers to Paired-Pulse Stimulation
Brunton, E., Silveira, C. R. A., Rosenberg, J. V., Schiefer, M. A., Riddell, J. S., & Nazarpour, K. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2935813 · Cited by 13 · see citing works
Karczewski, A. M., Dingle, A. M., & Poore, S. O. (2021) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2021.711028 · Cited by 13 · see citing works
Stimulation of Neurons with Infrared Radiation
Zhao, K., Tan, X., Young, H. K., & Richter, C. (2016) · Springer eBooks · doi:10.1007/978-1-4939-1758-7_17 · Cited by 13 · see citing works
Highly Flexible Precisely Braided Multielectrode Probes and Combinatorics for Future Neuroprostheses
Kim, T., Schmidt, K., Deemie, C., Wycech, J., Liang, H., & Giszter, S. F. (2019) · Frontiers in Neuroscience · doi:10.3389/fnins.2019.00613 · Cited by 9 · see citing works
Liao, Y.-W., Schearer, E. M., Hu, X., Perreault, E. J., Tresch, M. C., & Lynch, K. (2013) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2013.6610321 · Cited by 11 · see citing works
Rapid and Low Cost Manufacturing of Cuff Electrodes
Flavin, M. T., Paul, M. A., Lim, A. S., Abdul‐Hamed, S., Lissandrello, C. A., Ajemian, R., Lin, S. J., & Han, J. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.628778 · Cited by 11 · see citing works
A cable-tie-type parylene cuff electrode for peripheral nerve interfaces
Yu, H., Xiong, W., Zhang, H., Wang, W., & Li, Z. (2014) · 2014 IEEE 27th International Conference on Micro Electro Mechanical Systems (MEMS) · doi:10.1109/memsys.2014.6765560 · Cited by 10 · see citing works
Pellot-Cestero, J. E., Herring, E. Z., Graczyk, E. L., Memberg, W. D., Kirsch, R. F., Ajiboye, A. B., & Miller, J. P. (2023) · Neurosurgery · doi:10.1227/neu.0000000000002561 · Cited by 11 · see citing works
Self-citationCady, S. R., Lambrecht, J. M., Dsouza, K. T., Dunning, J. L., Anderson, J. R., Malone, K. J., Chepla, K. J., Graczyk, E. L., & Tyler, D. J. (2025) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-025-01613-z · Cited by 11 · see citing works
Self-citationA Switched-Capacitor Front-End for Velocity-Selective ENG Recording
Rieger, R., & Taylor, J. T. (2013) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2012.2226719 · Cited by 10 · see citing works
Identifying inverse human arm dynamics using a robotic testbed
Schearer, E. M., Liao, Y.-W., Perreault, E. J., Tresch, M. C., Memberg, W. D., Kirsch, R. F., & Lynch, K. (2014) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2014.6943064 · Cited by 10 · see citing works
Self-citationCrago, P. E. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aaeaa9 · Cited by 10 · see citing works
Blana, D., Hincapie, J. G., Chadwick, E. K., & Kirsch, R. F. (2013) · The Journal of Rehabilitation Research and Development · doi:10.1682/jrrd.2012.02.0034 · Cited by 10 · see citing works
Self-citationIncreased selectivity of clinical peripheral nerve interfaces
Tyler, D. J., Peterson, E. J., Brill, N., & White, K. (2011) · 2011 5th International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2011.5910536 · Cited by 10 · see citing works
Self-citationMuscle preservation in proximal nerve injuries: a current update
Lysak, A. S., Farnebo, S. J., Geuna, S., & Dahlin, L. B. (2024) · Journal of Hand Surgery (European Volume) · doi:10.1177/17531934231216646 · Cited by 9 · see citing works
Evaluation of a semi-parametric model for high-dimensional FES control
Schearer, E. M., Liao, Y.-W., Perreault, E. J., Tresch, M. C., Memberg, W. D., Kirsch, R. F., & Lynch, K. (2015) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2015.7146620 · Cited by 9 · see citing works
Self-citationSchearer, E. M., Liao, Y.-W., Perreault, E. J., Tresch, M. C., Memberg, W. D., Kirsch, R. F., & Lynch, K. (2012) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra.2012.6224981 · Cited by 8 · see citing works
Self-citationTaghlabi, K. M., Cruz-Garza, J. G., Hassan, T., Potnis, O., Bhenderu, L. S. S., Guerrero, J. R., Whitehead, R. E., Wu, Y. C., Luan, L., Xie, C., et al. (2024) · Journal of Neural Engineering · doi:10.1088/1741-2552/ad200f · Cited by 8 · see citing works
Askari, S., Presacco, A., Sahyouni, R., Djalilian, H. R., Shkel, A. M., & Lin, H. W. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8512778 · Cited by 8 · see citing works
Simple Quasi-Static Control of Functional Electrical Stimulation-Driven Reaching Motions
Wolf, D. N., & Schearer, E. M. (2019) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2019.8716955 · Cited by 8 · see citing works
Fattal, C., Teissier, J., Geffrier, A., da Fonseca, L. O., William, L., Andreu, D., Guiraud, D., & Coste, C. (2022) · Journal of Neurotrauma · doi:10.1089/neu.2021.0381 · Cited by 8 · see citing works
Neuroprothetik und Neuromodulation
Stieglitz, T. (2010) · Bundesgesundheitsblatt - Gesundheitsforschung - Gesundheitsschutz · doi:10.1007/s00103-010-1093-0 · Cited by 8 · see citing works
Freeberg, M. J., Ansari, R., Pinault, G. C. J., Lombardo, L. M., Miller, M. E., Tyler, D. J., & Triolo, R. J. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2951079 · Cited by 8 · see citing works
Self-citationDirected stimulation with interfascicular interfaces for peripheral nerve stimulation
Koppaka, S., Hess-Dunning, A., & Tyler, D. J. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac33e8 · Cited by 7 · see citing works
Self-citationFrederick, R. A., Troyk, P. R., & Cogan, S. F. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac2bb8 · Cited by 7 · see citing works
Abiri, A., Chau, S., James, N. R., Goshtasbi, K., Birkenbeuel, J. L., Sahyouni, R., Edwards, R. A., Djalilian, H. R., & Lin, H. W. (2022) · Brain Sciences · doi:10.3390/brainsci12111457 · Cited by 7 · see citing works
Wolf, D. N., van den Bogert, A. J., & Schearer, E. M. (2023) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2023.3272929 · Cited by 7 · see citing works
Hargrove, L. J. (2017) · Springer eBooks · doi:10.1007/978-1-4939-7247-0_8 · Cited by 7 · see citing works
Liao, Y.-W., Schearer, E. M., Perreault, E. J., Tresch, M. C., & Lynch, K. (2014) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2014.6942849 · Cited by 6 · see citing works
Sahyouni, R., Haidar, Y. M., Moshtaghi, O., Wang, B. Y., Djalilian, H. R., Middlebrooks, J. C., & Lin, H. W. (2017) · Otology & Neurotology · doi:10.1097/mao.0000000000001545 · Cited by 6 · see citing works
Electrodes for the Neural Interface
Tyler, D. J. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-805353-9.00020-6 · Cited by 6 · see citing works
Self-citationZaylaa, A. J., Haidar, I. M., Doughan, Z. F., & Haidar, A. M. (2024) · BAU Journal - Science and Technology · doi:10.54729/2959-331x.1150 · Cited by 4 · see citing works
Multi-Channel Microscale Nerve Cuffs for Spatially Selective Neuromodulation
Riley, M., Tala, F., Johnson, K. J., & Johnson, B. C. (2024) · Micromachines · doi:10.3390/mi15081036 · Cited by 5 · see citing works
Hand grasp and reach in spinal cord injury
Kilgore, K. L. (2015) · Elsevier eBooks · doi:10.1016/b978-1-78242-101-6.00009-4 · Cited by 5 · see citing works
Koppaka, S. (2013) · OhioLink ETD Center (Ohio Library and Information Network) · Cited by 4 · see citing works
FTSPlot: Fast Time Series Visualization for Large Datasets
Riss, M. (2014) · PLoS ONE · doi:10.1371/journal.pone.0094694 · Cited by 4 · see citing works
M’Rad, I. B., Bailey, Z., Cuttaz, E. A., & Green, R. A. (2025) · Bioelectronic Medicine · doi:10.1186/s42234-025-00193-6 · Cited by 3 · see citing works
- Authors
- Polasek, K. H., Hoyen, H. A., Keith, M. W., Kirsch, R. F., & Tyler, D. J.
- Affiliations
- Case Western Reserve University (US)
- MetroHealth Medical Center (US)
- Louis Stokes Cleveland VA Medical Center (US)
- Venue
- IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Published
- 2009-10
- Volume
- 17
- Issue
- 5
- Pages
- 428-437
- ISSN
- 1534-4320 (ISSN-L)
- 1558-0210
- PMC
- PMC2927980
- OpenAlex
- W2128390902
- Cited by
- 142
- References
- 34
- Citations by year
- 2026: 4 · 2025: 7 · 2024: 8 · 2023: 5 · 2022: 7 · 2021: 10 · 2020: 9 · 2019: 10 · 2018: 12 · 2017: 9 · 2016: 11 · 2015: 9 · 2014: 10 · 2013: 15 · 2012: 6
- FWCI
- 2.22
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- IEEE Transactions on Neural Systems and Rehabilitation Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Case Western Reserve University
- National Institutes of Health (n01ns012333-006)
- National Institute of Neurological Disorders and Stroke (N01-NS-1-2333, R21NS058705, N01 NS012333)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- cuff electrodes, chronic implantation, nerve conduction velocity, upper limb, functional electrical stimulation, multichannel stimulation, muscle recruitment
- MeSH terms
- Electrodes, Implanted, Electric Stimulation Therapy, Humans, Muscle Contraction, Paralysis, Prosthesis Design, Sensitivity and Specificity, Reproducibility of Results, Muscle, Skeletal, Equipment Failure Analysis, Upper Extremity
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review