Development of real-time motion artifact reduction algorithm for a wearable photoplethysmography
Abstract
This paper presents a motion artifact reduction algorithm for a real-time, wireless and wearable photoplethysmography (PPG) device for measuring heart beats. A wearable finger band PPG device consists of a 3-axis accelerometer, infrared LED, photo diode, a microprocessor and wireless module. Sources of the motion artifacts were investigated from the hand motions, through computing the correlations between the three directional finger motions and distorted PPG signals. A two-dimensional active noise cancellation algorithm was applied to compensate the distorted signals by motions, using the directional accelerometer data. NLMS (Normalized Least Mean Square) adaptive filter (4th order) was employed in the algorithm. As a result, the signals' distortion rates were reduced from 52.34% to 3.53%, at frequencies between 1 and 2.5 Hz, which representing daily motions such walking and jogging. The wearable health monitoring device equipped with the motion artifact reduction algorithm can be integrated as a terminal in a so-called ubiquitous healthcare system, which provides a continuous health monitoring without interrupting a daily life.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Secondary sensors can be integrated into PPG sensors to detect motion, including accelerometers, gyroscopes, piezoelectric transducers, or another wavelength of light as a motion reference.
“As implemented by consumer-grade HR monitors, secondary sensors can also be integrated into PPG sensors to detect motion. These include accelerometers [175,176,177,178,179], gyroscopes [179,180], piezoelectric transducers [181], or the utilization of another wavelength of light as a motion reference [182,183].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [177]
Cites
Active Noise Control Systems: Algorithms and DSP Implementations
Kuo, S. M., & Morgan, D. R. (1996) · Medical Entomology and Zoology · Cited by 1,560 · see citing works
Mobile monitoring with wearable photoplethysmographic biosensors
Asada, H. H., Shaltis, P., Reisner, A., Rhee, S., & Hutchinson, R. C. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213624 · Cited by 466 · see citing works
(1997) · Institute of Physics eBooks · doi:10.1887/0750304677 · Cited by 455 · see citing works
Motion Artifact Reduction in Photoplethysmography Using Independent Component Analysis
Kim, B., & Yoo, S. K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.869784 · Cited by 400 · see citing works
Reduction of motion artifact in pulse oximetry by smoothed pseudo Wigner-Ville distribution
Yan, Y. S., Poon, C. C. Y., & Zhang, Y. (2005) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-2-3 · Cited by 157 · see citing works
Invariant temporal characteristics of manipulative hand movements
Kunesch, E., Binkofski, F., & Freund, H. (1989) · Experimental Brain Research · doi:10.1007/bf00230241 · Cited by 132 · see citing works
Active noise cancellation using MEMS accelerometers for motion-tolerant wearable bio-sensors
Asada, H. H., Jiang, H.-H., & Gibbs, P. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2004.1403631 · Cited by 113 · see citing works
Chan, K. W., & Zhang, Y. (2003) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2002.1037314 · Cited by 104 · see citing works
Hand Motion Gesture Frequency Properties and Multimodal Discourse Analysis
Xiong, Y., & Quek, F. K. H. (2006) · International Journal of Computer Vision · doi:10.1007/s11263-006-8112-5 · Cited by 65 · see citing works
Wrist-located pulse detection using IR signals, activity and nonlinear artifact cancellation
Renevey, P., Vetter, R., Krauss, J., Celka, P., & Depeursinge, Y. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2001.1017437 · Cited by 63 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · see citing works
“In another study, a two-dimensional (2-D) active noise cancellation algorithm was applied to compensate for motion-distorted signals using directional accelerometer data [55]; a NLMS adaptive filter (fourth-order) was used in the algorithm, resulting in a reduction in signal distortion from 52.”
MethodSources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Cuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226 · see citing works
Organic Multi-Channel Optoelectronic Sensors for Wearable Health Monitoring
Khan, Y., Han, D., Ting, J., Ahmed, M., Nagisetty, R., & Arias, A. C. (2019) · IEEE Access · doi:10.1109/access.2019.2939798 · Cited by 111 · see citing works
Zong, C., & Jafari, R. (2015) · 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7320268 · Cited by 39 · see citing works
Couceiro, R., Carvalho, P., Paiva, R. P., Henriques, J., & Muehlsteff, J. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6346497 · Cited by 33 · see citing works
- Authors
- Han, H., Kim, M.-J., & Kim, J.
- Affiliations
- Korea Advanced Institute of Science and Technology (KR)
- Venue
- 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society
- Publisher
- IEEE
- Published
- 2007-08
- Volume
- 2007
- Pages
- 1538-1541
- ISSN
- 1557-170X (ISSN-L)
- OpenAlex
- W2111481778
- Cited by
- 101
- References
- 10
- Citations by year
- 2025: 1 · 2024: 2 · 2023: 3 · 2022: 5 · 2021: 7 · 2020: 9 · 2019: 12 · 2018: 9 · 2017: 11 · 2016: 6 · 2015: 8 · 2014: 7 · 2013: 3 · 2012: 3
- FWCI
- 8.42
- Type
- conference paper
- Language
- en
- Open access
- closed
- Locations
- Conference proceedings (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- photoplethysmography, motion artifact reduction, wearable sensors, accelerometer, active noise control, real-time signal processing, heart rate monitoring, wireless health monitoring, ubiquitous healthcare, walking and jogging
- MeSH terms
- Algorithms, Diagnosis, Computer-Assisted, Heart Rate, Humans, Motion, Sensitivity and Specificity, Reproducibility of Results, Artifacts, Photoplethysmography, Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review