A Robust Motion Artifact Detection Algorithm for Accurate Detection of Heart Rates From Photoplethysmographic Signals Using Time–Frequency Spectral Features
Open accessThe DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.
Abstract
Motion and noise artifacts (MNAs) impose limits on the usability of the photoplethysmogram (PPG), particularly in the context of ambulatory monitoring. MNAs can distort PPG, causing erroneous estimation of physiological parameters such as heart rate (HR) and arterial oxygen saturation (SpO2). In this study, we present a novel approach, "TifMA," based on using the time-frequency spectrum of PPG to first detect the MNA-corrupted data and next discard the nonusable part of the corrupted data. The term "nonusable" refers to segments of PPG data from which the HR signal cannot be recovered accurately. Two sequential classification procedures were included in the TifMA algorithm. The first classifier distinguishes between MNA-corrupted and MNA-free PPG data. Once a segment of data is deemed MNA-corrupted, the next classifier determines whether the HR can be recovered from the corrupted segment or not. A support vector machine (SVM) classifier was used to build a decision boundary for the first classification task using data segments from a training dataset. Features from time-frequency spectra of PPG were extracted to build the detection model. Five datasets were considered for evaluating TifMA performance: (1) and (2) were laboratory-controlled PPG recordings from forehead and finger pulse oximeter sensors with subjects making random movements, (3) and (4) were actual patient PPG recordings from UMass Memorial Medical Center with random free movements and (5) was a laboratory-controlled PPG recording dataset measured at the forehead while the subjects ran on a treadmill. The first dataset was used to analyze the noise sensitivity of the algorithm. Datasets 2-4 were used to evaluate the MNA detection phase of the algorithm. The results from the first phase of the algorithm (MNA detection) were compared to results from three existing MNA detection algorithms: the Hjorth, kurtosis-Shannon entropy, and time-domain variability-SVM approaches. This last is an approach recently developed in our laboratory. The proposed TifMA algorithm consistently provided higher detection rates than the other three methods, with accuracies greater than 95% for all data. Moreover, our algorithm was able to pinpoint the start and end times of the MNA with an error of less than 1 s in duration, whereas the next-best algorithm had a detection error of more than 2.2 s. The final, most challenging, dataset was collected to verify the performance of the algorithm in discriminating between corrupted data that were usable for accurate HR estimations and data that were nonusable. It was found that on average 48% of the data segments were found to have MNA, and of these, 38% could be used to provide reliable HR estimation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
EEG analysis based on time domain properties
Hjorth, B. (1970) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(70)90143-4 · Cited by 1,504
Algorithmic Stability and Sanity-Check Bounds for Leave-One-Out Cross-Validation
Kearns, M., & Ron, D. (1999) · Neural Computation · doi:10.1162/089976699300016304 · Cited by 407
Signal quality measures for pulse oximetry through waveform morphology analysis
Sukor, J. A., Redmond, S. J., & Lovell, N. H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/3/008 · Cited by 166
Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation
Karlen, W., Kobayashi, K., Ansermino, J. M., & Dumont, G. A. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/10/1617 · Cited by 123 · cited 1×
“Some approaches calculate the SQI using waveform morphologies [15]–[17] or filtered output [18], [19], while others derive the SQI with the help of additional hardware such as an accelerometer and electrocardiogram (ECG) [20], [21].”
MethodStatistical approach for the detection of motion/noise artifacts in Photoplethysmogram
Selvaraj, N., Mendelson, Y., Shelley, K. H., Silverman, D. G., & Chon, K. H. (2011) · 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2011.6091232 · Cited by 119 · cited 4×
“Statistical measures, such as skewness, kurtosis (K), and quadratic phase coupling [22], Shannon entropy (SE), and Renyi’s entropy [23], have been shown to be helpful in determining the SQI.”
InfluentialSelf-citationChong, J. W., Dao, D. K., Salehizadeh, S. M. A., McManus, D. D., Darling, C. E., Chon, K. H., & Mendelson, Y. (2014) · Annals of Biomedical Engineering · doi:10.1007/s10439-014-1080-y · Cited by 79 · cited 4×
“As mentioned above, the results of the MNA detection phase of TifMA were compared to three other existing MNA detection algorithms: the Hjorth [26], kurtosis-SE [23] and TDV-SVM approach, an approach recently developed in our laboratory [27].”
InfluentialKrishnan, R., Natarajan, B., & Warren, S. (2008) · 2008 IEEE International Conference on Acoustics, Speech and Signal Processing · doi:10.1109/icassp.2008.4517684 · Cited by 45
Naraharisetti, K. V. P., & Bawa, M. (2011) · 2011 IEEE INTERNATIONAL CONFERENCE ON ELECTRO/INFORMATION TECHNOLOGY · doi:10.1109/eit.2011.5978571 · Cited by 35 · cited 2×
“Several algorithm-based MNA reduction methods were proposed, such as time- and frequency-domain filtering, power spectrum analysis, and blind source separation techniques [6]–[12].”
Cited by
Atrial fibrillation detection from wrist photoplethysmography signals using smartwatches
Bashar, S. K., Han, D., Hajeb-Mohammadalipour, S., Ding, E., Whitcomb, C., McManus, D. D., & Chon, K. H. (2019) · Scientific Reports · doi:10.1038/s41598-019-49092-2 · Cited by 120
Self-citationXu, K., Jiang, X., Lin, S., Dai, C., & Chen, W. (2020) · IEEE Transactions on Industrial Informatics · doi:10.1109/tii.2020.2988097 · Cited by 37
“The representations of spectral filtering are SpaMA [19], TifMA [20], OxiMA [2], and SFST [21].”
MethodPhotoplethysmogram Signal Quality Evaluation by Unsupervised Learning Approach
Roy, M. S., Gupta, R., & Sharma, K. D. (2020) · 2020 IEEE Applied Signal Processing Conference (ASPCON) · doi:10.1109/aspcon49795.2020.9276733 · Cited by 24
“The non-segmenting approaches were illustrated in [11], [12], [13], [14], [15] and [16].”
BackgroundMethod
Metadata
- Authors
- Dao, D., Salehizadeh, S. M. A., Noh, Y., Chong, J. W., Cho, C. H., McManus, D., Darling, C. E., Mendelson, Y., & Chon, K. H.
- Venue
- IEEE Journal of Biomedical and Health Informatics
- Published
- 2017-09
- Volume
- 21
- Issue
- 5
- Pages
- 1242-1253
- ISSN
- 2168-2194 (ISSN-L)
- 2168-2208
- OpenAlex
- W2538635319
- Cited by
- 101
- References
- 8
- FWCI
- 3.95
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A Robust Motion Artifact Detection Algorithm for Accurate Detection of Heart Rates From Photoplethysmographic Signals Using Time–Frequency Spectral Features. https://app.cochat.ai/papers/10.1109/jbhi.2016.2612059
- The paper
- Dao, D., Salehizadeh, S. M. A., Noh, Y., Chong, J. W., Cho, C. H., McManus, D., Darling, C. E., Mendelson, Y., & Chon, K. H. (2017). A Robust Motion Artifact Detection Algorithm for Accurate Detection of Heart Rates From Photoplethysmographic Signals Using Time–Frequency Spectral Features. IEEE Journal of Biomedical and Health Informatics. https://doi.org/10.1109/jbhi.2016.2612059
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.