Cited by 26Wireless Power Transfer SystemsWireless Body Area NetworksRadio Frequency Integrated Circuit Design

A 430-MHz Wirelessly Powered Implantable Pulse Generator with Intensity/Rate Control and Sub-1 <formula> <tex>$\mu$</tex> </formula>A Quiescent Current Consumption

Lyu, Hongming, Gad, Parag, Zhong, Hui, Edgerton, V. Reggie, Babakhani, Aydin·2018·IEEE Transactions on Biomedical Circuits and Systems

Abstract

This work presents a miniaturized μW-level implantable pulse generator (IPG) inductively powered at 430 MHz. Notches are intentionally applied to the incident power, which are replicated to precisely control the timing of the output pulses. Fabricated in a 180-nm CMOS process, the concise circuitry occupies a pad-included footprint of 850 μm × 450 μm and achieves a quiescent current consumption of 950 nA. To reduce the form factor, 401-457 MHz MedRadio-band is utilized to realize the induction link. The finalized assembly achieves one of the smallest dimensions (4.6 mm × 7.0 mm) for near-field IPGs with the Rx coil size of 4.5 mm × 3.6 mm. Codesign of the rectifier and Rx coil accommodates the possible resonant frequency drifts in biological tissues. In the benchtop measurement, a 430-MHz Tx coil is demonstrated to operate the IPG at 4.5 and 4 cm proximities in the air and through water, respectively. An in vivo experiment has been performed, in which the IPG was implanted on the hindlimb muscle belly of an anesthetized rat with the connective tissue and skin sutured. The electrical stimuli induced the isolated ankle flexion at specific strengths and rates, and the experiment complies with the specific absorption rate regulations. This work shows the potential for applications requiring stringent form factors and high sensitivities.

Integrity signals

Registry

Crossref · 10.1109/tbcas.2018.2879357

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Lyu, H., Gad, P., Zhong, H., Edgerton, V. R., & Babakhani, A.
Affiliations
University of California, Los Angeles (US)
Venue
IEEE Transactions on Biomedical Circuits and Systems
Published
2018
Volume
13
Issue
1
Pages
1-1
ISSN
  • 1932-4545 (ISSN-L)
  • 1940-9990
OpenAlex
W2902722753
Cited by
26
References
7
Citations by year
2025: 1 · 2024: 3 · 2023: 2 · 2022: 3 · 2021: 4 · 2020: 8 · 2019: 5
FWCI
1.62
Type
journal article
Language
en
Open access
closed
Locations
  • IEEE Transactions on Biomedical Circuits and Systems (publishedVersion)
  • PubMed (publishedVersion)
Funders
National Science Foundation
Topics
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
  • Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
  • Physical Sciences › Engineering › Electrical and Electronic Engineering › Radio Frequency Integrated Circuit Design
Keywords
implantable pulse generator, inductive power transfer, 430 MHz, MedRadio band, rectifier design, resonant frequency drift, specific absorption rate
MeSH terms
Animals, Computer Simulation, Electricity, Models, Theoretical, Muscles, Rats, Wireless Technology

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review