An Online Failure Detection Method of the Glucose Sensor-Insulin Pump System: Improved Overnight Safety of Type-1 Diabetic Subjects
Abstract
Sensors for real-time continuous glucose monitoring (CGM) and pumps for continuous subcutaneous insulin infusion (CSII) have opened new scenarios for Type-1 diabetes treatment. However, occasional failures of either CGM or CSII may expose diabetic patients to possibly severe risks, especially overnight (e.g., inappropriate insulin administration). In this contribution, we present a method to detect in real time such failures by simultaneously using CGM and CSII data streams and a black-box model of the glucose-insulin system. First, an individualized state-space model of the glucose-insulin system is identified offline from CGM and CSII data collected during a previous monitoring. Then, this model, CGM and CSII real-time data streams are used online to obtain predictions of future glucose concentrations together with their confidence intervals by exploiting a Kalman filtering approach. If glucose values measured by the CGM sensor are not consistent with the predictions, a failure alert is generated in order to mitigate the risks for patient safety. The method is tested on 100 virtual patients created by using the UVA/Padova Type-1 diabetic simulator. Three different types of failures have been simulated: spike in the CGM profile, loss of sensitivity of glucose sensor, and failure in the pump delivery of insulin. Results show that, in all cases, the method is able to correctly generate alerts, with a very limited number of false negatives and a number of false positives, on av…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
There was a single false positive alert in one scenario after 1,257 hours of nonfaulty simulation.
“There was a single false positive alert in one scenario after 1,257 hours of nonfaulty simulation.15”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [15]
With a predefined upper safety limit for blood glucose of 300 mg/dl, all 100 in silico scenarios detected the fault below the safety limit except for two false negatives.
“With a predefined upper safety limit for blood glucose of 300 mg/dl all 100 scenarios detected the fault below the safety limit except for two false negatives.”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [15]
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559 · see citing works
Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes
Bergenstal, R. M., Tamborlane, W. V., Ahmann, A., Buse, J. B., Dailey, G., Davis, S. N., Joyce, C., Peoples, T., Perkins, B. A., Welsh, J. B., et al. (2010) · New England Journal of Medicine · doi:10.1056/nejmoa1002853 · Cited by 894 · see citing works
Insulin-pump therapy for type 1 diabetes mellitus
Pickup, J. C. (2012) · New England Journal of Medicine · doi:10.1056/nejmct1113948 · Cited by 270 · see citing works
Helton, K. L., Ratner, B. D., & Wisniewski, N. A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500318 · Cited by 102 · see citing works
Schmid, V., Hohberg, C., Borchert, M., Forst, T., & Pfützner, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400429 · Cited by 97 · see citing works
Evolution of Devices in Diabetes Management
Penfornis, A., Personeni, E., & Borot, S. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0058 · Cited by 49 · see citing works
Insulin pump failures are still frequent: a prospective study over 6 years from 2001 to 2007
Guilhem, I., Balkau, B., Lecordier, F., Malécot, J.-M., Elbadii, S., Leguerrier, A.-M., Poirier, J.-Y., Derrien, C., & Bonnet, F. (2009) · Diabetologia · doi:10.1007/s00125-009-1549-7 · Cited by 46 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Continuous glucose monitoring sensors: past, present and future algorithmic challenges
Facchinetti, A. (2016) · Sensors · doi:10.3390/s16122093 · Cited by 176 · see citing works
Self-citationResalat, N., El Youssef, J., Tyler, N., Castle, J., & Jacobs, P. G. (2019) · PLOS ONE · doi:10.1371/journal.pone.0217301 · Cited by 78 · see citing works
Early Detection of Infusion Set Failure During Insulin Pump Therapy in Type 1 Diabetes
Cescon, M., DeSalvo, D. J., Ly, T. T., Maahs, D. M., Messer, L. H., Buckingham, B. A., Doyle, F. J., & Dassau, E. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816663962 · Cited by 51 · see citing works
Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block
Klonoff, D. C., Freckmann, G., & Heinemann, L. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817700545 · Cited by 41 · see citing works
- Authors
- Facchinetti, A., Favero, S., Sparacino, G., & Cobelli, C.
- Affiliations
- University of Padua (IT)
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2013-02
- Volume
- 60
- Issue
- 2
- Pages
- 406-416
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- OpenAlex
- W1978587779
- Cited by
- 62
- References
- 7
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 9 · 2023: 3 · 2022: 1 · 2021: 5 · 2020: 5 · 2019: 4 · 2018: 7 · 2017: 7 · 2016: 6 · 2015: 1 · 2014: 5 · 2013: 4
- FWCI
- 2.57
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Padua Research Archive (University of Padova) (submittedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Control and Systems Engineering › Fault Detection and Control Systems
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- fault detection, continuous glucose monitoring, insulin pump therapy, type 1 diabetes, glucose-insulin system, state-space model identification, Kalman filter, glucose prediction, confidence intervals, sensor faults, pump failure, UVA/Padova simulator, virtual patients
- MeSH terms
- Algorithms, Computer Simulation, Diabetes Mellitus, Type 1, Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Sensitivity and Specificity, Signal Processing, Computer-Assisted, Blood Glucose Self-Monitoring, Reproducibility of Results, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review