Early Detection of Infusion Set Failure During Insulin Pump Therapy in Type 1 Diabetes
Open accessAbstract
Background: Insulin infusion set failure resulting in prolonged hyperglycemia or diabetic ketoacidosis can occur with pump therapy in type 1 diabetes. Set failures are frequently characterized by variable and unpredictable patterns of increasing glucose values despite increased insulin infusion. Early detection may minimize the risk of prolonged hyperglycemia, an important consideration for automated insulin delivery and closed-loop applications. Methods: A novel algorithm designed to alert the patient to the onset of infusion set failure was developed based upon continuous glucose sensor values and insulin delivered from an insulin pump. The method was calibrated on 12 weeks of infusion set wear without failures recorded by 4 patients in ambulatory conditions and prospectively validated on 18 weeks of infusion set wear with and without failures belonging to 9 other subjects in ambulatory conditions. Results: The algorithm, evaluated retrospectively, identified a failure 2.52 ± 1.91 days ahead of the actual event as recorded by the clinical team, corresponding to 50% sensitivity, 66% specificity and 55% accuracy. If set failure alarms had been activated in real time, the average time >180 mg/dl would be reduced from 82.7 ± 40.9 hours/week/subject (without alarm) to 58.8 ± 31.1 hours/week/subject (with alarm), corresponding to a potential 29% reduction in time spent >180mg/dl. Conclusion: The proposed method for early detection of infusion set failure based on glucose sensor and insulin data demonstrated favorable results on retrospective data and may be implemented as an additional safeguard in a future fully automated closed-loop system.
AI snapshot
Written from the paper’s full text.
TL;DR
The paper proposes a data-driven algorithm to anticipate infusion set failure during insulin pump therapy in type 1 diabetes using CGM, glucose trend, and insulin intake data. In retrospective validation, it detected failures 2.52 ± 1.91 days ahead but achieved 50% sensitivity, 66% specificity, and 55% accuracy.
Key findings
- The algorithm achieved 50% sensitivity, 66% specificity, and 55% accuracy, with 6 true positives, 2 false positives, 6 false negatives, and 4 true negatives.
- Failures were identified 2.52 ± 1.91 days ahead of the actual event recorded by the clinical team.
- If the alarm had been activated in real time and patients reacted, the authors estimated a potential 29% reduction in time spent above 180 mg/dl, from 82.7 ± 40.9 hours/week/patient to 58.8 ± 31.1 hours/week/patient.
- The method does not explicitly use a glucose regulatory model, meal announcement, or physical activity inputs, and aims to anticipate rather than detect faults; the authors state it is the first time such a method has been applied to actual outpatient data.
Methods
- Twenty-three subjects with type 1 diabetes at 2 clinical sites wore teflon infusion sets for 7 days or until failure; CGM was measured with a Dexcom G4 Platinum sensor. Of 69 patient-weeks collected, 30 weeks from 15 subjects were suitable; 12 failure-free weeks from 4 patients were used for calibration, and 18 weeks (13 with verified failures, 5 without) from 9 patients were used for validation.
- The algorithm used three components: CGM data segmentation via a time-varying autoregressive model with sum-of-norms regularization, linear trend slope assessment, and insulin counting comparing bolus number and total insulin infused in the past τ hours versus the same interval 24 hours earlier. User parameters were n = 3, λ = 0.007 λmax, α = 1.5, and τ = 4.
- The algorithm was evaluated retrospectively and blind to failures; sensitivity, specificity, and accuracy were computed, and time spent above 180 mg/dl was compared with and without the alert.
Limitations
- The paper states the method requires an uninterrupted stream of CGM data to detect an increasing glucose trend accurately; it is not suitable for abrupt failures such as tubing caught on a door handle or suddenly ripped off; patients changing the set and glucose sensor simultaneously at week 2 fed the algorithm a poorly calibrated data stream, negatively affecting performance, and no mitigation was taken.
- The paper states long-term clinical studies are needed to minimize the false positive rate, and that additional data and clinical work may improve specificity and accuracy.
- Evident from the text: validation was retrospective and small (18 weeks from 9 patients), and the 29% reduction in hyperglycemia was “potential,” assuming patients would acknowledge and react to alerts; real-time clinical utility remains to be established.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
2 of 6 cited sources checked · 2 supported · 0 flagged
- Premisecitation
In all these cases, metabolic decompensation and development of ketoacidosis can occur due to insufficient insulin delivery.
“In all these cases, metabolic decompensation and development of ketoacidosis can occur due to insufficient insulin delivery.5”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [5] source
- Findingnone given
Infusion set failures are frequently characterized by variable and unpredictable patterns of increasing glucose values despite increased insulin infusion.
“Set failures are frequently characterized by variable and unpredictable patterns of increasing glucose values despite increased insulin infusion.”
Quote in the paperSubstantial claim - Findingown data
Retrospective analysis of the collected data highlighted a persistent ramp-like increase in glycemia levels for weeks with an infusion set failure, as opposed to no increase for fault-free weeks.
“Retrospective analysis of the collected data highlighted a persistent ramp-like increase in glycemia levels for those weeks presenting an infusion set failure, as opposed to no increase for weeks fault-free”
Quote in the paperSubstantial claim - Findingnone given
The main predictor of length of wear was the individual subject.
“The main predictor of length of wear was the individual subject.”
Quote in the paperNot a substantial claim - Findingown data
Failures were identified 2.52 ± 1.91 days ahead of the actual event.
“The failures were identified 2.52 ± 1.91 days ahead of the actual event.”
Quote in the paperSubstantial claim
- Premisecitation
Infusion set failure resulting in prolonged hyperglycemia or diabetic ketoacidosis can occur with pump therapy in type 1 diabetes.
“Infusion set failure resulting in prolonged hyperglycemia or diabetic ketoacidosis 1 can occur with pump therapy in type 1 diabetes.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [1]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation
It is not uncommon for an infusion set to be left in longer than 3 days, despite manufacturer recommendation, for practical and economical reasons.
“Despite the recommendation of the manufacturers (see, eg, Medtronic Diabetes 2 ), it is not an uncommon occurrence for an infusion set to be left in longer than 3 days, for both practical and economical reasons.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [2]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationown data
If the set failure alarm had been activated in real time, a potential 29% reduction in the time spent above 180 mg/dl would have been achieved, from 82.7 ± 40.9 hours/week/patient to 58.8 ± 31.1 hours/week/patient.
“If the set failure alarm had been activated in real time, a potential 29% reduction in the time spent above 180 (mg/dl) would have been achieved, from 82.7 ± 40.9 hours/week/ patient to 58.8 ± 31.1 hours/week/patient.”
Quote in the paperSubstantial claim- The authors state this as an assumption: The patient would have acknowledged the alert and reacted taking appropriate actions.
- Interpretationown data
Our study describes a new alert method capable of predicting the development of an infusion set failure event.
“Our study describes a new alert method capable of predicting the development of an infusion set failure event.”
Quote in the paperSubstantial claim - Interpretationown data
The clinical benefit of removing a set is significant and overcomes the time spent in hyperglycemia.
“The clinical benefit of removing a set is significant and overcomes the time spent in hyperglycemia.”
Quote in the paperSubstantial claim - Study design and conductnone given
A limitation of the method is represented by the need of an uninterrupted stream of CGM data from the sensor, in order to be able to detect an increasing glucose trend accurately.
“a limitation of the method is represented by the need of an uninterrupted stream of CGM data from the sensor, in order to be able to detect an increasing glucose trend accurately.”
Quote in the paperNot a substantial claim - Interpretationown data
A positive slope of the glucose trend indicates potential hyperglycemia if the same trend continues.
“A positive slope of the glucose trend indicate potential hyperglycemia if the same trend continues.”
Quote in the paperSubstantial claim
- May combine two claims. 14
- Classified as probably not a substantial claim. 11
- The quote only partly states this claim. 1
May combine two claims.
PremisecitationInfusion set failure can occur when there is local inflammation at the infusion site, when the catheter is occluded, kinked or has simply dislodged from the subcutaneous space.
“Infusion set failure can occur when there is local inflammation at the infusion site, when the catheter is occluded, kinked or has simply dislodged from the subcutaneous space.3,4”
Quote in the paperNot a substantial claim1 of 2 sources checkedMay combine two claims.
Findingnone givenIn case of a developing failure, the patient glucose-insulin dynamics is slowly changing, presenting variable patterns of increasing glucose levels, and more insulin is being infused by the patients outside of their ordinary therapy because insulin is not as effective in lowering blood glucose.
“In case of a developing failure, the patient glucose-insulin dynamics is slowly changing, presenting variable patterns of increasing glucose levels.At the same time, more insulin is being infused by the patients outside of their ordinary therapy because insulin is not as effective in lowering blood glucose (see Figure 1 for a representative subject).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe algorithm achieved 50% sensitivity, 66% specificity and 55% accuracy, with 6 TP, 2 FP, 6 FN, 4 TN.
“The algorithm achieved 50% sensitivity, 66% specificity and 55% accuracy, with 6 TP, 2 FP, 6 FN, 4 TN.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn the validation subset, 17% of sets failed due to a failed correction dose, 17% failed because of serum ketones ≥0.6 mmol/l with hyperglycemia, 28% had erythema or induration >1 cm in diameter at the infusion site, 5% were removed because of bleeding and 5% were pulled out by accident.
“With regard to the second subset, overall, 17% of the sets failed due to a failed correction dose, 17% failed because of the presence of serum ketones ≥0.6 mmol/l with hyperglycemia, 28% had erythema or induration > 1 cm in diameter at the infusion site, 5% were removed because of bleeding and 5% were pulled out by accident.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe method was evaluated retrospectively on a validation data set collected from patients in ambulatory conditions and demonstrated favorable results detecting the onset of faulty events before they become severe with enough time to react to them.
“The method was evaluated retrospectively on a validation data set collected from patients in ambulatory conditions and demonstrated favorable results detecting the onset of faulty events before they become severe with enough time to react to them.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationcitationEarly detection of infusion set failure may minimize the risk of prolonged hyperglycemia, an important consideration for automated insulin delivery and closed-loop applications.
“Early detection may minimize the risk of prolonged hyperglycemia, an important consideration for automated insulin delivery and closed-loop applications.1”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [1]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
Study design and conductnone givenThe method was developed and retrospectively validated on actual patients data collected in ambulatory conditions.
“It was developed and retrospectively validated on actual patients data collected in ambulatory conditions.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
InterpretationreasoningOne single larger insulin injection may indicate the subject had a bigger meal, while several closely spaced small injections with total insulin the same as the larger dose are more likely to indicate that a failure is developing.
“One single larger insulin injection may indicate the subject had a bigger meal, while several closelyspaced small injections with total insulin the same as the larger dose are more likely to indicate that a failure is developing.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataOverall, application of the method showed that it is possible to extend the length of wear of infusion sets beyond 3 days with associated benefits, without incurring an increase time spent in hyperglycemia.
“Overall, application of the method showed that it is possible to extend the length of wear of infusion sets beyond 3 days with associated benefits, without incurring in an increase time spent in hyperglycemia.”
Quote in the paperSubstantial claim- The authors state this as an assumption: The patient would have acknowledged the alert and reacted taking appropriate actions.
Classified as probably not a substantial claim. May combine two claims.
InterpretationreasoningOur method was designed to be proactive and to be able to deal with the harder constraint of slow occlusions in a predictive manner rather than to detect abrupt faults like catheter disconnections, which are unforeseen events and as such cannot be predicted.
“Our method was designed to be proactive and to be able to deal with the harder constraint of slow occlusions in a predictive manner rather than to detect abrupt faults like catheter disconnections, which are unforeseen events and as such cannot be predicted.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
InterpretationreasoningThe proposed method is aiming at anticipating faults rather than recognizing them after they have happened.
“Moreover, the proposed method is aiming at anticipating faults rather than recognizing them after they have happened.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenThe results obtained with the proposed failure detection algorithm are independent of the type of insulin infusion system.
“The results obtained with the proposed failure detection algorithm are independent of the type of insulin infusion system.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
InterpretationreasoningBoth the number of insulin boluses and the quantity of infused insulin need to be taken into account, under the assumption that neither the time of the meal nor the meal content are available.
“We would like to highlight that both the number of insulin boluses and the quantity of infused insulin need to be taken into account, under the assumption that neither the time of the meal nor the meal content are available.”
Quote in the paperSubstantial claim- The authors state this as an assumption: Neither the time of the meal nor the meal content are available.
Classified as probably not a substantial claim.
Study design and conductown dataThe proposed algorithm was evaluated on 18 weeks of data from 9 patients in a retrospective and blind to the failures manner.
“The proposed algorithm was evaluated on 18 weeks of data from 9 patients in a retrospective and blind to the failures manner.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Study design and conductnone givenThe 3-step procedure ensures an increased degree of robustness to the method.
“The 3-step procedure ensures an increased degree of robustness to the method.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenAlerts on an infusion site failure will reduce the complications and glucose swings associated with this type of events.
“However, the alerts on an infusion site failure will reduce the complications and glucose swings associated with this type of events.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenTo the best of the authors knowledge this is the first time that a method with these features has been applied to actual outpatient data.
“To the best of the authors knowledge this is the first time that a method with these features has been applied to actual outpatient data.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims. The quote only partly states this claim.
InterpretationreasoningThe proposed method is robust and is able to deal with inter-current illness, medication, hormone cycles and physiological stress because its design was based on machine learning tools.
“The proposed method is robust and is able to deal with this fact because its design was based on machine learning tools.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable). May combine two claims.
InterpretationreasoningEarly detection and mitigation of such failures are critical safety issues not only for a closed-loop (artificial pancreas) system, but also for current open-loop systems.
“Early detection and mitigation of such failures are critical safety issues not only for a closed-loop (artificial pancreas) system, but also for current open-loop systems.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone givenThe advantage of our algorithm compared to the cited works is that it does not require an explicit description of the glucose-insulin regulatory model, nor does it exploit user information such as meal announcement or physical activity.
“The advantage of our algorithm compared to the above cited works is that it does not require an explicit description of the glucose-insulin regulatory model, nor does it exploit user information such as meal announcement or physical activity.”
Quote in the paperSubstantial claim
The paper relies on this without support.
Patients may only change the infusion set when they notice a persistent increase in their blood glucose level, meaning that the set is not properly working.
“Patients may only change the infusion set when they notice a persistent increase in their blood glucose level meaning that the set is not properly working.”
The authors state this as an assumption.
Neither the time of the meal nor the meal content are available.
“under the assumption that neither the time of the meal nor the meal content are available.”
The authors state this as an assumption.
The patient would have acknowledged the alert and reacted taking appropriate actions.
“we assume the patient would have acknowledged the alert and reacted taking appropriate actions.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A failure detection algorithm applied to 100 in silico pump patients with type 1 diabetes detected a total occlusion or dislocation in 75% of patients within 63 minutes with a 10% false positive rate.
“A failure detection algorithm applied to 100 in silico pump patients with type 1 diabetes detected a total occlusion or dislocation in 75% of patients within 63 minutes with a 10% false positive rate.16”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [16]
The Performance Evaluation section says this paper tested its algorithm retrospectively on 18 weeks of data from 9 patients, not 100 simulated patients. The Results report 50% sensitivity and failures identified 2.52 ± 1.91 days before the recorded event, rather than 75% detected within 63 minutes; they do not report a 10% false-positive rate. The Discussion mentions a separate approach tested on 100 simulated patients, but does not give the claimed performance figures. (LLM, conf=0.89)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measuredIIS failures are the primary contributor to DKA development and the leading cause of DKA-related hospitalizations among insulin pump users.
“Correspondingly, IIS failures are the primary contributor to DKA development and the leading cause of DKA-related hospitalizations among insulin pump users.[9][10][11][12]”
From Frequency and Detection of Insulin Infusion Site Failure in the Type 1 Diabetes Exchange Online Community (2023) · cites this paper as [11]
The Abstract and Conclusions state that infusion set failures can cause DKA in people with type 1 diabetes using pumps. They do not establish that these failures are the primary contributor to DKA or the leading cause of DKA-related hospitalizations among pump users. Page numbers are unavailable in our copy. (LLM, conf=0.87)
Experimental — accuracy not yet measuredOcclusions of insulin infusion sets occur frequently.
“Occlusions do occur frequently”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [17]
The Abstract (page unavailable) identifies catheter occlusion as a cause of infusion set failure, so it supports that occlusions occur. It does not say how frequently occlusions occur; its statement that failures are ‘frequently characterized’ by rising glucose describes their pattern, not their incidence. (LLM, conf=0.83)
Experimental — accuracy not yet measuredOcclusions of insulin infusion sets often occur without an alarmed warning, or the warning is provided too late.
“often without an alarmed warning or this provided too late”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [17]
The Discussion (page not available) says pump mechanisms may detect a full occlusion only after some time and do not detect infusion-site infections, supporting the concern about late or absent warnings for some failures. It does not establish that occlusions often occur without an alarm. (LLM, conf=0.84)
Experimental — accuracy not yet measuredIn a clinical trial with nine subjects, the Cescon et al algorithm achieved 50% sensitivity and 66% specificity for predicting IIS failure.
“The algorithm achieved 50% sensitivity and 66% specificity.”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [17]
If the IIS failure alarm had been activated in real time, a potential 29% reduction in the time spent >180 mg/dl (from 82.7 hours/week/patient to 58.8 hours/week/patient) would have been achieved.
“If the IIS failure alarm had been activated in real time, a potential 29% reduction in the time spent >180 mg/dl (from 82.7 hours/week/patient to 58.8 hours/week/patient) would have been achieved, assuming that the patient would have acknowledged the alert and reacted with appropriate actions.”
From Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block (2017) · cites this paper as [17]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Subject counts exceed the two subsets combined
MinorThe two subsets contain 4 and 9 patients, so they can include at most 13 distinct subjects, not the reported 15.
Reported 15 · recomputed 13
“hence only 30 weeks from 15 subjects were selected and explored for our purposes.A first subset of 12 weeks without set failures from 4 patients were used to calibrate the algorithm and a second subset of an additional 18 weeks (13 with verified set failures and 5 without) from 9 patients were used for validation.”
Failure-reason codes conflict with their definitions
MinorTable 1 defines A as a failed correction, B as hyperglycemia with ketones, and C as an infusion-site infection. Table 2 labels a failed correction B and a tubing accident C.
“| CO01w3 | 3.37 | Failed correction before failure (B) | | CO03w3 | 4.31 | Hyperglycemia from failed correction (A) | | CO04w1 | 5.95 | Infection at infusion site (C) | | CO04w2 | 7 | Indeterminate | | CO04w3 | 3.15 | Caught tubing on door handle (C) |”
“A. Meter blood glucose > 250 mg/dl and a correction dose fails to decrease the glucose by 50 mg/dl in 1 hour B. Meter blood glucose > 250 mg/dl and serum ketones ≥ 0.6 mmol/l in the absence of infection at the infusion site C. Infection at the infusion site”
Cites
System Identification: Theory for the User
Lennart Ljung, A. (1987) · Cited by 17,690 · see citing works
Modeling by shortest data description
Rissanen, J. (1978) · Automatica · doi:10.1016/0005-1098(78)90005-5 · Cited by 6,011 · see citing works
Quantitative estimation of insulin sensitivity.
Bergman, R. N., Ider, Y. Z., Bowden, C. R., & Cobelli, C. (1979) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1979.236.6.e667 · Cited by 1,874 · see citing works
Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes
Hovorka, R., Canonico, V., Chassin, L. J., Haueter, U., Massi-Benedetti, M., Federici, M. O., Pieber, T. R., Schaller, H. C., Schaupp, L., Vering, T., et al. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/4/010 · Cited by 1,274 · see citing works
Meal Simulation Model of the Glucose-Insulin System
Dalla Man, C., Rizza, R. A., & Cobelli, C. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.893506 · Cited by 955 · see citing works
<i>In Silico</i> Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300106 · Cited by 760 · see citing works
American Control Conference 1994
Beauchard, S. D. K. (1994) · Computing & Control Engineering Journal · doi:10.1049/cce:19940511 · Cited by 568 · see citing works
GIM, Simulation Software of Meal Glucose–Insulin Model
Dalla Man, C., Raimondo, D. M., Rizza, R. A., & Cobelli, C. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100303 · Cited by 253 · see citing works
Segmentation of ARX-models using sum-of-norms regularization
Ohlsson, H., Ljung, L., & Boyd, S. (2010) · Automatica · doi:10.1016/j.automatica.2010.03.013 · Cited by 146 · see citing works
Schmid, V., Hohberg, C., Borchert, M., Forst, T., & Pfützner, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400429 · Cited by 97 · see citing works
Facchinetti, A., Favero, S., Sparacino, G., & Cobelli, C. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2227256 · Cited by 62 · see citing works
Robust Fault Detection System for Insulin Pump Therapy Using Continuous Glucose Monitoring
Herrero, P., Calm, R., Vehí, J., Armengol, J., Georgiou, P., Oliver, N., & Tomazou, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600518 · Cited by 52 · see citing works
Insulin pump failures are still frequent: a prospective study over 6 years from 2001 to 2007
Guilhem, I., Balkau, B., Lecordier, F., Malécot, J.-M., Elbadii, S., Leguerrier, A.-M., Poirier, J.-Y., Derrien, C., & Bonnet, F. (2009) · Diabetologia · doi:10.1007/s00125-009-1549-7 · Cited by 46 · see citing works
van Bon, A. C., Dragt, D., & DeVries, J. H. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0259 · Cited by 25 · see citing works
Li, T., Evans, A. T., Chiravuri, S., Gianchandani, R. Y., & Gianchandani, Y. B. (2012) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2012.05.002 · Cited by 24 · see citing works
Real-time detection of Glucose Sensor and Insulin Pump Faults in an Artificial Pancreas.
Favero, S. D., Monaro, M., Facchinetti, A., Tagliavini, A., Sparacino, G., & Cobelli, C. (2014) · IFAC Proceedings Volumes · doi:10.3182/20140824-6-za-1003.02733 · Cited by 24 · see citing works
Increasing security in an artificial pancreas: diagnosis of actuator faults
Vega-Hernández, O., Campos-Cornejo, F., Campos-Delgado, D. U., et al. (2009) · Pan American Health Care Exchanges · doi:10.1109/pahce.2009.5158384 · Cited by 18 · see citing works
Facchinetti, A., Favero, S. D., Sparacino, G., & Cobelli, C. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091226 · Cited by 16 · see citing works
Multivariate statistical analysis to detect insulin infusion set failure
Rojas, R., García-Gabín, W., & Bequette, B. W. (2011) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2011.5990951 · Cited by 12 · see citing works
Rojas, R., García-Gabín, W., & Bequette, B. W. (2011) · IFAC Proceedings Volumes · doi:10.3182/20110828-6-it-1002.03147 · Cited by 8 · see citing works
Flagged first, then by citations
Research works printed but not linked: 7
- Williams G, 2004, The Handbook of Diabetes
- Rojas R, 18th IFAC World Congress, 2011, 14127
- Rojas R, 2011, American Control Conference, 1952
- Cameron F, Diabetes Tech Meeting, 2012, A17
- Facchinetti A, 33rd IEEE EMBS, 2011, 4947
- Del Favero S, 2014, 19th IFAC World Congress, 1941
- Bergman RN, 1979, Am J Physiol, 236, E667
Cited by
Sorted by citations
Artificial intelligence for diabetes management and decision support: Literature review
Contreras, I., & Vehi, J. (2018) · Journal of Medical Internet Research · doi:10.2196/10775 · Cited by 560 · see citing works
“Cescon et al proposed the use of a time-varying autoregressive model to develop a patient alert system [121].”
BackgroundPhillip, M., Nimri, R., Bergenstal, R. M., Barnard-Kelly, K., Danne, T., Hovorka, R., Kovatchev, B. P., Messer, L. H., Parkin, C. G., Ambler-Osborn, L., et al. (2023) · Endocrine Reviews · doi:10.1210/endrev/bnac022 · Cited by 443 · see citing works
Self-citationAdvances in technology for management of type 1 diabetes
Beck, R. W., Bergenstal, R. M., Laffel, L. M., & Pickup, J. C. (2019) · The Lancet · doi:10.1016/s0140-6736(19)31142-0 · Cited by 242 · see citing works
Danne, T., Phillip, M., Buckingham, B. A., Jarosz-Chobot, P. A., Saboo, B., Urakami, T., Battelino, T., Hanås, R., & Ethel Codner, D. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12718 · Cited by 226 · see citing works
“5 mol/L per 30 minutes).(176,177) Patients must be instructed on treatment of hyperglycemia, giving insulin with a pen or syringe in case of suspected pump failure (hyperglycemia and elevated ketone levels).”
BackgroundSelf-citationISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197 · see citing works
“Widespread adoption of this has not occurred, and many groups continue to explore how to develop improved infusion sets(77) or fault detection algorithms to advise a user of when insulin delivery may be interrupted.(78,79) Some studies have documented between a 2- and a 5-fold higher risk of DKA in those on pump thera…”
BackgroundSelf-citationContinuous glucose monitoring sensors: past, present and future algorithmic challenges
Facchinetti, A. (2016) · Sensors · doi:10.3390/s16122093 · Cited by 176 · see citing works
“Some algorithms for fault detection have been presented in the literature [17–21], but they are still prototypes, and more investigation is required on their effectiveness before porting them into AP systems.”
Sherr, J. L., Schoelwer, M., Dos Santos, T. J., Reddy, L., Biester, T., Galderisi, A., van Dyk, J. C., Hilliard, M. E., Berget, C., & DiMeglio, L. A. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13421 · Cited by 152 · see citing works
Realizing a Closed-Loop (Artificial Pancreas) System for the Treatment of Type 1 Diabetes
Lal, R., Ekhlaspour, L., Hood, K. K., & Buckingham, B. A. (2019) · Endocrine Reviews · doi:10.1210/er.2018-00174 · Cited by 97 · see citing works
Self-citationBiester, T., Berget, C., Boughton, C. K., Cudizio, L., Ekhlaspour, L., Hilliard, M. E., Reddy, L., Sap, S. N. U., Schoelwer, M. J., Sherr, J. L., et al. (2024) · Hormone Research in Paediatrics · doi:10.1159/000543034 · Cited by 110 · see citing works
“Strategies for identifying failed infusion sets include fault detection algorithms that utilize sensor glucose levels and insulin delivery data to predict potential failures have been described [52, 53].”
Data-Driven Anomaly Recognition for Unsupervised Model-Free Fault Detection in Artificial Pancreas
Meneghetti, L., Terzi, M., Favero, S. D., Susto, G. A., & Cobelli, C. (2018) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2018.2885963 · Cited by 68 · see citing works
“Methods to detect pump faults based on heuristics and threshold values were proposed in [12] and [13] and tested on real data.”
BackgroundMethodInfluentialInsulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block
Klonoff, D. C., Freckmann, G., & Heinemann, L. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817700545 · Cited by 41 · see citing works
Meneghetti, L., Facchinetti, A., & Favero, S. D. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3004270 · Cited by 36 · see citing works
“proposed a 3-step procedure algorithm designed for anticipating rather than detecting pump failures, [29].”
MethodThe automated pancreas: A review of technologies and clinical practice
Biester, T., Tauschmann, M., Chobot, A. P., Kordonouri, O., Danne, T., Kapellen, T. M., & Dovč, K. (2021) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14576 · Cited by 32 · see citing works
“7 mmol/L) within 1 hour of treatment.(88)”
Meneghetti, L., Susto, G. A., & Favero, S. D. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819881452 · Cited by 28 · see citing works
Continuous Glucose Monitoring Enables the Detection of Losses in Infusion Set Actuation (LISAs)
Howsmon, D., Cameron, F., Baysal, N., Ly, T., Forlenza, G., Maahs, D., Buckingham, B., Hahn, J., & Bequette, B. (2017) · Sensors · doi:10.3390/s17010161 · Cited by 26 · see citing works
“[17] also evaluated infusion set failure algorithms on clinical data as opposed to simulated faults, but tried to anticipate—as opposed to detect—failures.”
BackgroundSelf-citationKalaiselvan, V., Sharma, A., & Gupta, S. K. (2020) · Health and Technology · doi:10.1007/s12553-020-00495-6 · Cited by 21 · see citing works
“Ceson et al. proposed a model that can alert the patient earlier [42].”
BackgroundHughes, M. S., Douvas, J. L., Layfield-Bryan, M., Blanco, L. E., Gray, J. C., Zapotoczny, G., Espinoza, J., Wilcox, J. H., & Lal, R. A. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0005 · Cited by 16 · see citing works
“5,6,11,13,14 We present survey data from the T1D Exchange Online Community to explore real-world experience with IIS failure.”
Rabbone, I., Minuto, N., Toni, S., Lombardo, F., Iafusco, D., Marigliano, M., Schiaffini, R., Maltoni, G., Frongia, A. P., Scardapane, M., et al. (2018) · Diabetes Obesity and Metabolism · doi:10.1111/dom.13419 · Cited by 15 · see citing works
“Early detection may minimize the risk of prolonged hyperglycaemia, an important consideration for automated insulin delivery and closed-loop applications.(1) We recently described a low rate of insulin pump breakdown/ replacement (0.”
BackgroundXu, L., Huang, Y. H., Wang, D., Zhu, S., Wang, Z., Yang, Y., & Guo, Y. (2019) · Journal of Materials Science Materials in Medicine · doi:10.1007/s10856-019-6287-x · Cited by 13 · see citing works
Alshorman, J., Mehran, M. J., Bahrami, Y., Mohammadzadeh, S., Barzigar, R., Morshedi, M., Haider, K. H., Tembo, K. M., Rong, S.-J., Jadgal, N., et al. (2026) · Clinical and Experimental Medicine · doi:10.1007/s10238-026-02107-5 · Cited by 14 · see citing works
Insulin pumps in children - a systematic review
Al‐Beltagi, M., Saeed, N. K., Bediwy, A. S., & Elbeltagi, R. M. (2022) · World Journal of Clinical Pediatrics · doi:10.5409/wjcp.v11.i6.463 · Cited by 11 · see citing works
Fault Detection in Artificial Pancreas: A Model-Free approach
Meneghetti, L., Terzi, M., Susto, G. A., Favero, S. D., & Cobelli, C. (2018) · IEEE Conference on Decision and Control (CDC) · doi:10.1109/cdc.2018.8619048 · Cited by 11 · see citing works
“[9], [8]), real-data collected using AP system lack of this crucial information forcing to use labels retrospectively assigned by via visual inspection.”
MethodKwa, T., Zhang, G., Shepard, K., Wherry, K., & Chattaraj, S. (2021) · Journal of Medical Economics · doi:10.1080/13696998.2021.1945784 · Cited by 10 · see citing works
Paredes-Juarez, G. A., de Vos, P., & Bulte, J. W. M. (2017) · Expert Review of Precision Medicine and Drug Development · doi:10.1080/23808993.2017.1302305 · Cited by 10 · see citing works
“continuous changes in glucose-controlled insulin release is one solution, but still requires follow-up measurements to assess whether or not insulin has been infused correctly, along with recalibration of the device [14].”
BackgroundMeneghetti, L., Dassau, E., Doyle, F. J., & Favero, S. D. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296821997854 · Cited by 11 · see citing works
Self-citationDegen, I., Brown, K. R., Reeve, H. W. J., & Abdallah, Z. S. (2024) · JMIRx Med · doi:10.2196/44384 · Cited by 8 · see citing works
“[1] gave examples of problems addressed by machine learning such as blood glucose prediction, overcoming shortcomings of deterministic automatic insulin delivery algorithms such as predicting set failure [4] or sensor failure [5] and incident prediction such as predicting hypoglycemia [6].”
BackgroundBeneyto, A., Bequette, B. W., & Vehı́, J. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211029297 · Cited by 8 · see citing works
Thrasher, J. R., Polsky, S., Hovsepian, L., NOWOTNY, I. K., Pierre, S., Fer, B. B. D., Bhargava, A., Mukherjee, B., & Garg, S. K. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0446 · Cited by 8 · see citing works
Berget, C., & Wyckoff, L. (2020) · NASN School Nurse · doi:10.1177/1942602x20928914 · Cited by 8 · see citing works
Insulin infusion sets associated with adverse events: strategies for improved diabetes education
Neves, A. L. D., Martins, L. E. G., Gabbay, M. A. L., Pascali, P. M., de Oliveira, T., Martinazzo, A., Dib, S. A., Casarini, D. E., Arêdes, S. V., Tenório, F. S., et al. (2023) · Frontiers in Medicine · doi:10.3389/fmed.2023.1275394 · Cited by 6 · see citing works
“Early detection of these factors is crucial to minimize the risk of prolonged hyperglycemia, especially in automatic insulin administration methods such as CSII therapy (27).”
MethodMetabolic Models, in Silico Trials, and Algorithms
Çınar, A., Basu, A., Bequette, B. W., Breton, M. D., Buckingham, B. A., Cengiz, E., Cobelli, C., Dassau, E., Doyle, F. J., Fabris, C., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.18800.ac · Cited by 5 · see citing works
“Pressure-induced sensor attenuation, 94 imputation of missing sensor data, 95 pump and insulin delivery faults, 96,97 and errors in control logic have been addressed.”
Self-citationManzoni, E., Rampazzo, M., Facchinetti, A., Sparacino, G., & Favero, S. D. (2023) · Control Engineering Practice · doi:10.1016/j.conengprac.2023.105673 · Cited by 6 · see citing works
Autoencoder-Based Detection of Insulin Pump Faults in Type 1 Diabetes Treatment
Idi, E., Prendin, F., Sparacino, G., & Favero, S. D. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3518233 · Cited by 5 · see citing works
“In particular, the problem was investigated by developing: heuristic methods [10], model-based approaches [11], [12], [13], [14], [15] and machine learning-based strategies [16], [17].”
Fault detection in glucose control: Is it time to move beyond CGM data?
Kölle, K., Fougner, A. L., Unstad, K. A. F., et al. (2018) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2018.11.646 · Cited by 5 · see citing works
“Although a retrospective data analysis is originally reported twice (Bondia et al., 2008; Cescon et al., 2016), all methods in table 1 are generally suitable for real-time applications.”
BackgroundAhmad, A., & Mohamed, A. A. M. (2024) · Studies in computational intelligence · doi:10.1007/978-981-99-9029-0_14 · Cited by 5 · see citing works
Metabolic Models, in Silico Trials, and Algorithms
Çınar, A., Basu, A., Bequette, B. W., Breton, M. D., Buckingham, B. A., Cengiz, E., Cobelli, C., Dassau, E., Doyle, F. J., Fabris, C., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251338300 · Cited by 4 · see citing works
“Pressure-induced sensor attenuation, 94 imputation of missing sensor data, 95 pump and insulin delivery faults, 96,97 and errors in control logic have been addressed.”
Self-citationEmpfehlungen zur Diabetes-Behandlung mit automatischen Insulin-Dosierungssystemen
Berge, T. V. D., Biester, S., Biester, T., Buchmann, A.-K., Datz, N., Grosser, U., Kapitzke, K., Klusmeier, B., Remus, K., Reschke, F., et al. (2021) · Diabetologie und Stoffwechsel · doi:10.1055/a-1652-9011 · Cited by 4 · see citing works
Development of a Real-time Force-based Algorithm for Infusion Failure Detection
Blanco, L. E., Wilcox, J. H., Hughes, M. S., & Lal, R. A. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241247530 · Cited by 2 · see citing works
“3,18 19-21 Yet, the efficacy of these systems is reliant on the accuracy and connectivity of external CGM sensors.”
Pachołek, A., Oniszczuk, E. R. S. , Mierzwa, J., Wróbel, A., Piotrowicz, K., Gąsowski, J., & Tomasik, T. (2023) · Folia Medica Cracoviensia · doi:10.24425/fmc.2023.145426 · Cited by 2 · see citing works
An Occlusion Detection System for Intravenous Infusion
Zhou, L., Li, Q., Wang, L., et al. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3317074 · Cited by 1 · see citing works
“Therefore, hydraulic pressure sensors [24], [25], [26], [27], [28], [29], [30], [31] with a priority of noninvasiveness, lower power consumption, and lower cost have a promising future in IV lines occlusion detection.”
BackgroundKetone-Based Alert System for Insulin Pump Failures
Aiello, E. M., Laffel, L. M., Patti, M., & Doyle, F. J. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231209339 · Cited by 1 · see citing works
“Cescon et al 39 proposed a combination of both a rising trend in average daily CGM readings along with increasing daily doses of insulin.”
Self-citationMulti-field modeling and computational optimization of a subcutaneous insulin injection port
Zoboli, L., Luppino, F., Bianchi, D., et al. (2024) · Materials & Design · doi:10.1016/j.matdes.2024.113284 · Cited by 1 · see citing works
Enlarging the loop: closed-loop insulin delivery for type 1 diabetes
Galderisi, A., & Sherr, J. L. (2018) · The Lancet · doi:10.1016/s0140-6736(18)32329-8 · Cited by 1 · see citing works
Chappel, E. (2023) · Global Journal of Engineering and Technology Advances · doi:10.30574/gjeta.2023.16.3.0181 · Cited by 1 · see citing works
“Occlusion detection in a MEMS-based micropump Occlusion detection is crucial to maintain the blood glucose level in an acceptable range [28].”
MethodZelinska, N. (2020) · Ukrainian Journal of Pediatric Endocrinology · doi:10.30978/ujpe2020-4-39 · Cited by 1 · see citing works
The dawn of automated insulin delivery: from promise to product
Nally, L. M., & Sherr, J. L. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00016-8
Empfehlungen zur Diabetes-Behandlung mit automatischen Insulin-Dosierungssystemen
Berge, T. V. D., Biester, S., Biester, T., Buchmann, A.-K., Datz, N., Grosser, U., Kapitzke, K., Klusmeier, B., Remus, K., Reschke, F., et al. (2022) · Diabetes aktuell · doi:10.1055/a-1851-2883
Ucieklak, D., Mrozińska, S., Wojnarska, A., Malecki, M. T., Klupa, T., & Matejko, B. (2023) · Folia Medica Cracoviensia · doi:10.24425/fmc.2022.144079
Chappel, E. (2024) · Encyclopedia · doi:10.3390/encyclopedia4020052
“Occlusion detection is crucial to maintain the blood glucose level in an acceptable range [25].”
BackgroundEnhancing Insulin Delivery in Type 1 Diabetes with the Detection of Insulin Flow Interruption
Hamdi, R., & Cescon, M. (2025) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2025.12.204
Self-citationNieto de la Marca, M. D. L. O., Fernández Velasco, P., Del Amo Simón, S., Sánchez Ibáñez, M., Collantes Matallana, A., Gómez Hoyos, E., Torres Torres, B., Ortolá Buigues, A., de Luis Román, D. A., & Díaz Soto, G. (2026) · Endocrine · doi:10.1007/s12020-026-04743-z
“Catheter wear beyond day three has been associated with a higher risk of diabetic ketoacidosis, unexplained hyperglycemia, severe hypoglycemia, and other adverse events related to infusion set and catheter use [11].”
- Authors
- Cescon, M., DeSalvo, D. J., Ly, T. T., Maahs, D. M., Messer, L. H., Buckingham, B. A., Doyle, F. J., & Dassau, E.
- Affiliations
- University of California, Santa Barbara (US)
- Baylor College of Medicine (US)
- Stanford Medicine (US)
- University of Colorado Denver (US)
- Harvard University (US)
- Corresponding author
- Dassau, Eyal
- ORCID
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2016-09-25
- Volume
- 10
- Issue
- 6
- Pages
- 1268-1276
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC5094340
- OpenAlex
- W2519750058
- Semantic Scholar
- b01354bd500d2f07bf5246563b2e5c1b44f454d4
- Cited by
- 51
- References
- 16
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 7 · 2023: 8 · 2022: 4 · 2021: 5 · 2020: 5 · 2019: 5 · 2018: 8 · 2017: 3 · 2016: 1
- FWCI
- 4.93
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (submittedVersion)
- PubMed Central (submittedVersion)
- Funders
- National Institutes of Health
- National Institute of Diabetes and Digestive and Kidney Diseases (DP3DK 094331)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- type 1 diabetes, insulin pump therapy, continuous glucose monitoring, hyperglycemia, diabetic ketoacidosis, artificial pancreas, alarm system
- MeSH terms
- Algorithms, Blood Glucose, Diabetes Mellitus, Type 1, Equipment Failure, Female, Humans, Hyperglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Sensitivity and Specificity, Blood Glucose Self-Monitoring, Clinical Alarms
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-13.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-13.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review