Continuous Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio
Abstract
Pulse transit time (PTT) has attracted much interest for cuffless blood pressure (BP) measurement. However, its limited accuracy is one of the main problems preventing its widespread acceptance. Arterial BP oscillates mainly at high frequency (HF) because of respiratory activity, and at low frequency (LF) because of vasomotor tone. Prior studies suggested that PTT can track BP variation in HF range, but was inadequate to follow the LF variation, which is probably the main reason for its unsatisfactory accuracy. This paper presents a new indicator, the photoplethysmogram intensity ratio (PIR), which can be affected by changes in the arterial diameter, and, thus, trace the LF variation of BP. Spectral analysis of BP, PTT, PIR, and respiratory signal confirmed that PTT was related to BP in HF at the respiratory frequency, while PIR was associated with BP in LF range. We, therefore, develop a novel BP estimation algorithm by using both PTT and PIR. The proposed algorithm was validated on 27 healthy subjects with continuous Finapres BP as reference. The results showed that the mean ± standard deviation (SD) for the estimated systolic, diastolic, and mean BP with the proposed method against reference were -0.37 ±5.21, -0.08 ±4.06, -0.18 ±4.13 mmHg, and mean absolute difference (MAD) were 4.09, 3.18, 3.18 mmHg, respectively. Furthermore, the proposed method outperformed the two most cited PTT algorithms for about 2 mmHg in SD and MAD. These results demonstrated that the proposed BP model using PIR and PTT can estimate continuous BP with improved accuracy.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cardiovascular neural regulation explored in the frequency domain.
Malliani, A., Pagani, M., Lombardi, F., & Cerutti, S. (1991) · Circulation · doi:10.1161/01.cir.84.2.482 · Cited by 3,569
Chow, C. K. (2013) · JAMA · doi:10.1001/jama.2013.184182 · Cited by 1,991
Westerhof, N., Lankhaar, J.-W., & Westerhof, B. E. (2009) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-008-0359-2 · Cited by 1,181
The dynamic elastic properties of the arterial wall
Bergel, D. H. (1961) · The Journal of Physiology · doi:10.1113/jphysiol.1961.sp006687 · Cited by 904
Assessment and management of blood-pressure variability
Parati, G., Ochoa, J. E., Lombardi, C., & Bilo, G. (2013) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2013.1 · Cited by 853
Hemodynamic fluctuations and baroreflex sensitivity in humans: a beat-to-beat model
deBoer, R. W., Karemaker, J. M., & Strackee, J. (1987) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1987.253.3.h680 · Cited by 851
Parati, G., Saul, J. P., Di Rienzo, M., & Mancia, G. (1995) · Hypertension · doi:10.1161/01.hyp.25.6.1276 · Cited by 849
Unobtrusive Sensing and Wearable Devices for Health Informatics
Zheng, Y.-L., Ding, X.-R., Poon, C. C. Y., Lo, B. P. L., Zhang, H., Zhou, X.-L., Yang, G.-Z., Zhao, N., & Zhang, Y.-T. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2309951 · Cited by 722
Self-citationGesche, H., Grosskurth, D., Küchler, G., & Patzak, A. (2012) · European Journal of Applied Physiology · doi:10.1007/s00421-011-1983-3 · Cited by 470
Cited by
Flexible Piezoresistive Sensor Patch Enabling Ultralow Power Cuffless Blood Pressure Measurement
Luo, N., Dai, W., Li, C., Zhou, Z., Lu, L., Poon, C. C. Y., Chen, S., Zhang, Y., & Zhao, N. (2016) · Advanced Functional Materials · doi:10.1002/adfm.201504560 · Cited by 467
Self-citationBroza, Y. Y., Zhou, X., Yuan, M., Qu, D., Zheng, Y., Vishinkin, R., Khatib, M., Wu, W., & Haick, H. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.9b00437 · Cited by 461
Xu, H., Liu, J., Zhang, J., Zhou, G., Luo, N., & Zhao, N. (2017) · Advanced Materials · doi:10.1002/adma.201700975 · Cited by 298
Self-citationWearable Sensing and Telehealth Technology with Potential Applications in the Coronavirus Pandemic
Ding, X., Clifton, D., Ji, N., Lovell, N. H., Bonato, P., Chen, W., Yu, X., Xue, Z., Xiang, T., Long, X., et al. (2021) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2020.2992838 · Cited by 281
Self-citationOptical blood pressure estimation with photoplethysmography and FFT-based neural networks
Xing, X., & Sun, M. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.003007 · Cited by 265
Cuff-Less and Continuous Blood Pressure Monitoring: A Methodological Review
Sharma, M., Barbosa, K., Ho, V., Griggs, D., Ghirmai, T., Krishnan, S., Hsiai, T., Chiao, J.-C., & Cao, H. (2017) · Technologies · doi:10.3390/technologies5020021 · Cited by 226
Blood Pressure Assessment in Adults in Clinical Practice and Clinic-Based Research
Muntner, P., Einhorn, P. T., Cushman, W. C., Whelton, P. K., Bello, N. A., Drawz, P. E., Green, B. B., Jones, D. W., Juraschek, S. P., Margolis, K. L., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.069 · Cited by 199
(2022) · PubMed Central · doi:10.6515/acs.202205_38(3).20220321a · Cited by 117
Scardulla, F., D’Acquisto, L., Colombarini, R., Hu, S., Pasta, S., & Bellavia, D. (2020) · Sensors · doi:10.3390/s20185052 · Cited by 72
The use of pulse transit time in pediatric sleep studies: A systematic review
Smith, L. A., Dawes, P. J., & Galland, B. C. (2018) · Sleep Medicine Reviews · doi:10.1016/j.smrv.2016.11.006 · Cited by 22
Blood Pressure Estimation Based on Photoplethysmography: Finger Versus Wrist
Paliakaite, B., Charlton, P. H., Rapalis, A., Plusciauskaite, V., Piartli, P., Kaniusas, E., & Marozas, V. (2021) · 2021 Computing in Cardiology (CinC) · doi:10.23919/cinc53138.2021.9662716 · Cited by 11
Metadata
- Authors
- Ding, X.-R., Zhang, Y.-T., Liu, J., Dai, W.-X., & Tsang, H. K.
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2016-05
- Volume
- 63
- Issue
- 5
- Pages
- 964-972
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- OpenAlex
- W2338236284
- Cited by
- 330
- References
- 47
- FWCI
- 11.17
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.