A Quaternion Weighted Fourier Linear Combiner for Modeling Physiological Tremor
Open accessThe DOI matches its Crossref record; 45 of the 47 references it cites were found in a registry or an index; 7 of the 26 sources its claims cite were checked for support.
Abstract
GOAL: This paper offers a new approach to model physiological tremor aiming at attenuating undesired vibrations of the tip of microsurgical instruments. METHOD: Several tremor modeling algorithms, such as the weighted Fourier linear combiner (wFLC), have proved effective. They, however, treat the three-dimensional (3-D) tremor signal as three independent 1-D signals in the x-, y-, and z-axes. In addition, the force f by which a surgeon holds the instrument has never been taken into account in modeling. Hence, conventional algorithms are inherently blind to any potential multidimensional couplings. RESULTS: We first show that there exists statistically significant subject- and task-dependent coherence between data in the x-, y-, z -, and f-axes. We hypothesize that a filter that models the tremor in 4-D ( x , y, z, and f ) yields a more accurate model of tremor. We, therefore, developed a quaternion version of the wFLC algorithm and termed it QwFLC. We tested the proposed QwFLC algorithm with real physiological tremor data that were recorded from five novice subjects and four experienced microsurgeons. Although compared to wFLC, QwFLC requires six times larger computational resources, we showed that QwFLC can improve the modeling by up to 67% and that the improvement is proportional to the total coherence between the tremor in xyz and the force signal. CONCLUSION: By taking into account interactions of the 3-D tremor and the force data, the tremor modeling performance enhances significantly.
Overview
CoChat Summary
TL;DR
The paper proposes a quaternion weighted Fourier linear combiner (QwFLC) that models 3-D physiological tremor together with grip force as a 4-D quaternion process. In data from five novices and four microsurgeons, QwFLC-4D improved tremor modeling accuracy by up to 67% over conventional wFLC, with improvement proportional to xyz–force coherence.
Key findings
- Coherence analysis showed that tremor in the x, y, and z axes is not independent and that there is statistically significant, subject- and task-dependent coherence between xyz tremor and grip force f; coherence values above the threshold s = 0.1 were considered significant with p < 0.05.
- QwFLC-4D improved modeling accuracy by 67% compared with wFLC across all subjects on average; QwFLC-3D reduced modeling error by 27% versus three independent wFLCs, and QwFLC-4D reduced it further by 40%.
- Improvement in RMSE was proportional to total coherence: for 1D-to-3D improvement versus xyz coherence, R² = 0.63 and R² = 0.78 (p < 0.01); for 3D-to-4D improvement versus xyz–f coherence, R² = 0.65 and R² = 0.79 (p < 0.01).
- wFLC had a delay of 2 lags in 100% of cases (average 8 ms), whereas QwFLC-4D had 0 lag (<4 ms) for above 75% of pointing cases and above 95% of tracing cases; QwFLC was reported as requiring 6 times larger computational resources than wFLC.
Methods
- Five medical students with basic surgical skills and four experienced microsurgeons performed pointing and tracing tasks with an instrumented stylus; 3-D tip displacement was recorded by the optical micro motion sensing system M²S², grip force by a force sensor (0–15 N), digitized at 250 samples/s, and pre-filtered offline with a zero-lag 5th-order Butterworth filter (5–20 Hz). Data were shortened to 28 s, and only the 1× magnification condition was used because all subjects had data there.
- The authors extended FLC and wFLC to quaternion versions QFLC and QwFLC using quaternion LMS; QwFLC-3D used only xyz position in the quaternion, while QwFLC-4D included force as the real part. Performance was assessed by RMSE and by time delay from cross-correlation of 100 randomly selected 5-s segments per axis, subject, task, and algorithm.
- Coherence was computed with 1-s Hanning windows over D = 28 non-overlapping segments, with significance threshold s = 0.1 and p < 0.05; synthetic validation used a simulated 3-D reaching task with a 10 Hz tremor, SNR from −15 dB to 15 dB in 5 dB steps, and 10 iterations per SNR.
Limitations
- The paper states that an offline zero-lag filter was used in simulation, and that the linear filter used to extract voluntary motion would lead to up to 20 ms delay in real-time application; this delay is within a suggested range but could degrade accuracy, and multi-step prediction is proposed as a potential solution.
- The paper states that it did not determine the source of oscillations in f, leaving two competing hypotheses—neural source versus projection of xyz tremor through limb mechanics—and notes that distinguishing them would likely require EMG from hand and wrist muscles. It also notes that QwFLC still needs a reference tremor signal extracted offline from a mixture of voluntary and tremulous movement.
- Evident from the text: the real-data sample is small (nine subjects), limited to two laboratory tasks and one visual magnification condition, and the reported 67% improvement is an average across subjects; the authors also note that QwFLC-4D used QLMS rather than WL-QLMS even though grip force had higher power, for consistency in comparison.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 8 Not valid · 18 Not checked · 42 Own results and interpretations
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“This mismatch can push the adaptation towards the dominant components of the voluntary movement rather than that of the tremor [19].”
What the sources say
- Could not verifycites [19] sourceExperimental — accuracy not yet measured
The Results section on generalisation across BMI tasks discusses wrist movement and LFP-based firing-rate estimates, not tremor or adaptation driven by movement amplitude. Searches for ‘tremor’ and for voluntary-movement amplitude and adaptation found no passage addressing the claim. Page numbers were not provided in our copy. (LLM, conf=0.76)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“Therefore, they are not used in microsurgery in which realtime filtering is required [3].”
What the sources say
- Could not verifycites [3] sourceExperimental — accuracy not yet measured
The Experimental Methods section describes recordings of finger tremor and EMG, and the Foundations section describes spectral analysis; neither addresses microsurgery or a need for real-time filtering. Searches for microsurgery, low-pass filters, and real-time filtering found no passage stating why such filters would not be used in microsurgery. Page numbers were not available in our copy. (LLM, conf=0.78)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“The amplitude of typical physiological tremor is however approximately 50 times smaller than that of the voluntary motion.”
What the sources say
- Could not verifycites [19] sourceExperimental — accuracy not yet measured
The Methods section describes a wrist torque-controlled task, and the Results discuss wrist movement and neural signals, but neither gives a physiological-tremor amplitude. Searches for “physiological tremor,” “tremor,” and the claimed 50-fold comparison found no passage in our copy that supports the comparison with voluntary motion. Page numbers were not available. (LLM, conf=0.76)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“Real-time tremor compensation in robotics surgical instruments, requires the processes of tremor sensing and compensation to finish within a very short time, maximum 30ms [37].”
What the sources say
- Could not verifycites [37] sourceExperimental — accuracy not yet measured
The Methods (Experimental session) describe gyroscope measurements of lower-trunk motion during walking, while the Discussion mentions prior tremor-cancellation work. Searches for robotic surgical tremor compensation, latency, and 30 ms found no passage stating the claimed time limit. Page numbers were not available in our copy. (LLM, conf=0.76)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Although compared to wFLC, QwFLC requires 6 times larger computational resources”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
33 more claims
Create a free account to see all 68 claims and every source verdict.
- It may state two separate claims as one. 15
- Probably not a substantial claim. 18
- The quote only partly states this claim. 4
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
“For the pointing task we found the lowest peak power of 4 10 12 m 2 = Hz , and took our reference power level (-3dB: 2 : 8 10 12 m 2 = Hz ).”
Reported 2.8 · recomputed 2
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
“we identified the lowest peak power as 5 10 12 m 2 = Hz for the tracing task, and took our reference power level as its half power point (-3dB: 3 : 5 10 12 m 2 = Hz ).”
Reported 3.5 · recomputed 2.5
Citations
Cites
Discrete-Time Signal Processing
Oppenheim, A. V., & Schafer, R. W. (1989) · Cited by 12,572
Halliday, D. M., Rosenberg, J. R., Amjad, A. M., Breeze, P., Conway, B. A., & Farmer, S. F. (1995) · Progress in Biophysics and Molecular Biology · doi:10.1016/s0079-6107(96)00009-0 · Cited by 1,107
The Fourier approach to the identification of functional coupling between neuronal spike trains
Rosenberg, J. R., Amjad, A. M., Breeze, P., Brillinger, D. R., & Halliday, D. M. (1989) · Progress in Biophysics and Molecular Biology · doi:10.1016/0079-6107(89)90004-7 · Cited by 879
The first laparoscopic cholecystectomy.
Reynolds, W. (2001) · PubMed Central · Cited by 578
Discrete-Time Signal Processing
Jenkins, W. K., Jones, D. L., & Hunsinger, B. J. (2002) · Reference Data for Engineers · doi:10.1016/b978-075067291-7/50030-3 · Cited by 536
A Steady-Hand Robotic System for Microsurgical Augmentation
Taylor, R., Jensen, P., Whitcomb, L., Barnes, A., Kumar, R., Stoianovici, D., Gupta, P., Wang, Z., Dejuan, E., & Kavoussi, L. (1999) · The International Journal of Robotics Research · doi:10.1177/02783649922067807 · Cited by 528
Physiological and pathological tremors and rhythmic central motor control
McAuley, J. H. (2000) · Brain · doi:10.1093/brain/123.8.1545 · Cited by 509
A telerobotic assistant for laparoscopic surgery
Taylor, R. H., Funda, J., Eldridge, B., Gomory, S., Gruben, K., LaRose, D., Talamini, M., Kavoussi, L., & Anderson, J. (1995) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.391776 · Cited by 392
Evidence from Robot-Assisted Laparoscopic Radical Prostatectomy: A Systematic Review
Ficarra, V., Cavalleri, S., Novara, G., Aragona, M., & Artibani, W. (2007) · European Urology · doi:10.1016/j.eururo.2006.06.017 · Cited by 378
The Quaternion LMS Algorithm for Adaptive Filtering of Hypercomplex Processes
Took, C. C., & Mandic, D. P. (2009) · IEEE Transactions on Signal Processing · doi:10.1109/tsp.2008.2010600 · Cited by 332
Research works printed but not linked: 1
- jacobus, 1992, Implementation issues for telerobotic handcontrollers: Human-robot ergonomics, Human Robot Interaction, 284
Cited by
Quality of Experience Framework for Cloud Computing (QoC)
Laghari, A. A., He, H. X., Khan, A., Kumar, N., & Kharel, R. (2018) · IEEE Access · doi:10.1109/access.2018.2865967 · Cited by 89
“The user experience is based on the user’s level of satisfaction, learning ability, enjoyment and engagement [33]–[35], this feedback given for the service qualitywhich end user received at the destination.”
Performance Analysis of Quaternion-Valued Adaptive Filters in Nonstationary Environments
Xiang, M., Kanna, S., & Mandic, D. P. (2017) · IEEE Transactions on Signal Processing · doi:10.1109/tsp.2017.2787102 · Cited by 48
“Quaternions have subsequently found new applications in communications, motion tracking, and biomedical engineering [12]–[15].”
BackgroundMaximum Correntropy Quaternion Kalman Filter
Lin, D., Zhang, Q., Chen, X., et al. (2023) · IEEE Transactions on Signal Processing · doi:10.1109/tsp.2023.3300631 · Cited by 44
“Some current applications, such as communication, biomedical engineering, and motion tracking, are also emerging [3], [13], [14], [15].”
BackgroundMultiple-Model Adaptive Estimation for 3-D and 4-D Signals: A Widely Linear Quaternion Approach
Xiang, M., Dees, B. S., & Mandic, D. P. (2018) · IEEE Transactions on Neural Networks and Learning Systems · doi:10.1109/tnnls.2018.2829526 · Cited by 36
“These developments have enabled quaternions to find applications in communications, motion tracking, and biomedical signal processing [13]–[16].”
BackgroundTatinati, S., Nazarpour, K., Ang, W. T., & Veluvolu, K. C. (2016) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2016.2597119 · Cited by 36
“limited multiple Fourier linear combiner (BMFLC) [12]–[15]) have been popular to model the tremulous motion without any phase delay.”
BackgroundMethodSelf-citation
Metadata
- Authors
- Adhikari, K., Tatinati, S., Ang, W. T., Veluvolu, K. C., & Nazarpour, K.
- Affiliations
- University of Newcastle Australia (AU)
- Kyungpook National University (KR)
- Nanyang Technological University (SG)
- Newcastle University (GB)
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2016-11
- Volume
- 63
- Issue
- 11
- Pages
- 2336-2346
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- OpenAlex
- W2344336171
- Semantic Scholar
- 24033a572d28cd8fbeb1779f37fcd37dd2479fbe
- Cited by
- 48
- References
- 47
- Citations by year
- 2026: 2 · 2025: 4 · 2024: 3 · 2023: 8 · 2022: 5 · 2021: 3 · 2020: 4 · 2019: 8 · 2018: 5 · 2017: 3 · 2016: 3
- FWCI
- 4.02
- Type
- journal article
- Language
- en
- Open access
- green
- doi.org
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Newcastle University ePrints (Newcastle Univesity)
- Funders
- Newcastle University
- Basic Science Research Program
- National Research Foundation of Korea
- Ministry of Education, Science and Technology (2014R1A1A2A10056145)
- EPSRC (EP/M025977/1, EP/M025594/1)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Health Sciences › Medicine › Surgery › Surgical Simulation and Training
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Effects of Vibration on Health
- Keywords
- physiological tremor, quaternion signal processing, microsurgical instruments, coherence analysis, microsurgery
- MeSH terms
- Algorithms, Computer Simulation, Fourier Analysis, Humans, Microsurgery, Models, Biological, Essential Tremor, Robotic Surgical Procedures
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.